Product Leadership • Training • Service • Reliability
ENTER
HOLD
LIGHT
I/O
WARRANTY
Bacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in mater ial
and manufac ture and will conform substan tially to Bach arach Inc.’s ap plicable spec i fi ca tions. Bacharach’s
li a bil i ty and Buyer’s remedy under this warranty are lim it ed to the repair or replacement, at Ba charach’s option,
of this Product or parts thereof returned to Seller at the factory of manufac ture and shown to Bacharach Inc.’s
reasonable satisfaction to have been defective; provided that written notice of the defect shall have been given
by Buyer to Bacharach Inc. within two (2) years after the date of delivery of this Product by Bacharach, Inc.
Bacharach, Inc. warrants to Buyer that it will convey good title to this Product. Bacharach’s liability and Buyer’s
remedy under this warranty of title are limited to the removal of any title defects or, at the election of Bacha rach,
to the replacement of this Product or parts thereof that are defective in title.
THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN AND ACCEPTED IN LIEU OF (I) ANY
AND ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, IN CLUD ING WITH OUT LIMITATION THE IMPLIED WARRANTIES OF MERCHAN TABILITY AND FITNESS FOR A PARTICULAR PURPOSE: AND (II) ANY
OBLIGA TION, LIABILITY, RIGHT, CLAIM OR REM E DY IN CON TRACT OR TORT, WHETHER OR NOT ARISING FROM BACH A RACH’S NEGLIGENCE, ACTUAL OR IM PLIED. The remedies of the Buyer shall be limited
to those provided herein to the exclusion of any and all other remedies includ ing, without limita tion incidental or
consequen tial dam ag es. No agree ment varying or ex tend ing the foregoing warranties, remedies or this limitation will be binding upon Bacharach, Inc. unless in writing, signed by a duly au tho rized officer of Bacharach.
Register Your Warranty by Visiting
www.MyBacharach.com
Notice:
Product improvements and enhancements are continuous, therefore the spec i fi ca tions and in for ma tion
contained in this document may change without notice.
Bacharach, Inc. shall not be liable for errors contained herein or for incidental or consequential damages in
connection with the furnishing, performance, or use of this material.
No part of this document may be photocopied, reproduced, or translated to another language without the prior
written consent of Bacharach, Inc.
6.3 Service Centers .............................................................................. 57
ii
Instruction 0024-9395
Fyrite Pro
Confi gurations
Fyrite Pro
Ordering Information
Model100100
Printer
Kit
24-
Part Number
Measurements:
Oxygen
Stack
Temperature
Ambient
Temperature
Second
Thermocouple
Channel
(Combustion Air
Temperature)
8200
24-
8201
1051 10120125125
Printer
Kit
24-
8202
24-
8103
24-
8104
24-
8105
24-
8205
125
Printer/
Oil Kit
24-
8206
Differential
Temperature
Carbon Monoxide
(CO)
Draft
Differential
Presssure*
Instruction 0024-9395
iii
Fyrite Pro
Notes:
iv
Instruction 0024-9395
Fyrite Pro
1.0 INTRODUCTION
1.1 The Fyrite Pro
The Fyrite Pro series of instruments are hand-held an a lyz ers that are
designed for on-demand sampling of residential furnaces and appliances.
The more advanced models can also to used to check combustion efficiency
by measuring the stack temperature and the percent oxygen level in the
flue-gas stream.
The Fyrite Pro is completely upgradable from the very simple Model 100
that measures only carbon monoxide, to the Model 125 that measures
carbon monoxide (CO), oxygen (O
air temperature, stack temperature, and combustion air temperature, as
well as calculating combustion efficiency, excess air, carbon dioxide (CO
differential temperature, and CO air free.
), draft, differential pressure, ambient
2
),
2
This manual contains information on all models of the Fyrite Pro. Please
disregard any information that does not pertain to your model.
WARNING S!
This analyzer should not be used on a continuous basis.
This analyzer does not have an audible alarm, and should
not be used as a safety device.
Except for battery replacement, this analyzer should only be
opened and ser viced by au tho rized personnel.
When testing an appliance, a full visual in spec tion of the
appliance should be carried out to ensure its safe op er a tion.
CAUTION!
When the instrument is used in an ineffi cient furnace or
boiler application where there is a high emission of soot,
probe sample fi lter may become clogged. Check fi lter before
every use to confi rm that it is clean for use or replace with
a new fi lter. To prevent this from occurring, a smoke test
should be performed before operating under such conditions
to ensure that the furnace is burning at a level appropriate
for the use of this instrument.
Instruction 0024-9395
1
Fyrite Pro
1.2 Operational Overview
Pressing the I/O button turns the analyzer ON. Note that there is a warm-
up period of either 10 or 60 seconds, dependent on the analyzer’s model and
setup, that must elapse before the analyzer can be used.
When using the analyzer to just measure CO or draft, insert the analyzer’s probe into the flue-gas stream of the appliance under test and begin
testing. The measured values of draft and/or CO will appear on the analyzer’s LCD.
For models that calculate combustion efficiency, choose a fuel code that corresponds to the fuel being burned by the appliance being tested, and then
press the ENTER button to place the analyzer into its Run Mode. Begin
testing by inserting the analyzer’s probe tube into the flue-gas stream
of the appliance under test. The analyzer will monitor the flue gas and
display measured and calculated values that are relative to the combustion
process. These values are dis played on the analyzer’s LCD, and are chosen
for display by pressing the In cre ment (▲) and Dec re ment (▼) but tons.
Pressing the HOLD button during a test will freeze all measured and calculated values at their current levels. Pressing the ENTER button resumes testing.
For analyzers with memory, pressing the HOLD button twice during a
test will display the Print/Memory Menu Screen. Using the ▲▼ and
ENTER buttons the operator can choose to print the current readings,
save the current readings in memory, recall and view saved readings from
memory, clear all the memory, or print all saved readings.
A backlight enables an operator to read the display in dimly-lit areas.
Pressing the LIGHT button turns the backlight ON and OFF. The backlight
will automatically turn OFF after 10 minutes of keyboard inactivity.
A power saver function will shut the analyzer OFF after 20 min utes of keyboard inactivity. For models with an O
saver function is dis abled if the detected O
and CO sensor, however, the power
2
value is less than 17.9% or the
2
CO value is more than 50 ppm.
Press the I/O button to turn the analyzer OFF. Note that there is a 5 sec-
ond delay before the analyzer actually turns OFF, during which time the
unit can be kept ON by pressing the ENTER button. For analyzers that
contain a CO sensor, there is a CO purge feature that keeps the analyzer’s
pump running until the measured CO level drops below 50 ppm.
2
Instruction 0024-9395
Fyrite Pro
2.0 TECHNICAL CHARACTERISTICS
The Analyzer Directly Measures and Displays:
• Ambient Temperature ............................32 to 212 °F (0 to 100 °C)
• Flue Gas O
• Flue Gas Temperature ...........................0 to 999 °F (–18 to 537 °C)
• Flue Gas CO content ..............................0 to 2,000 ppm CO
• Draft ........................................................–9.99 to 27 inWC
The Analyzer Computes and Displays:
(when the measured oxygen level is below 17.9%)
• Combustion Efficiency ............................0.1 to 99.9%
• Flue Gas CO
• Flue Gas CO Air-Free content ...............0 to 9,999 ppm
• Excess Air ...............................................0 to 400%
content ................................ 0.0 to 20.9% O
2
2
(–9.99 to 67 mb/HPa)
content .............................0.1 to a fuel dependent
2
maximum value in %
Fuel Selection:
(F1 thru F4 denotes the fuel selected as displayed on the LCD)
• Natural Gas (F1)
• Oil #2 (F2)
• LPG (F3)
• Kerosene (F4)
Normal Operating Conditions:
Temperature:
• Analyzer .................................................32 to 104 °F (0 to 40 °C)
• Probe .......................................................1,000 °F max (538 °C) at
5" insertion
Humidity .......................................................15 to 90% RH,
Non-Condensing
Power Requirements ....................................Four disposable ‘AA’ Alkaline
batteries (an optional AC adapter is available that eliminates
the need for batteries)
Operating Time ................................. Up to 18 hours con tin u ous (pump
run ning and backlight off)
Warm Up Time ................................... 10 or 60 seconds, dependent on
Instruction 0024-9395
analyzer’s model and setup (Refer
to Section 4.13.1)
3
Fyrite Pro
Memory
(Model 105/110/120/125) ..................... Up to 10 sets of readings can be
saved in memory. When memory
is full, the next reading saved will
overwrite the oldest reading.
Front Panel Controls ....................... Six push button switches
(Refer to Section 3.4)
Accuracy:
• Oxygen............................................. ±0.3% O
concentration of CO
with a typical flue gas
2
2
• Carbon Monoxide ............................ ±5% of reading or ±10 ppm, which-
ever is greater*
• Flue Gas Temperature .................... ±4 °F between 32 & 255 °F
(±2 °C between 0 & 124 °C)
±6 °F between 256 & 480 °F
(±3 °C between 125 & 249 °C) ±8 °F
between 481 & 752 °F
(±4 °C between 250 & 400 °C)
• Second Thermocouple Channel ...... ±4 °F between 32 & 212 °F
(±2 °C between 0 & 100 °C)
• Ambient Temperature ..................... ±4 °F between 32 & 104 °F
(±2 °C between 0 & 40 °C)
• Draft ................................................ ±1% of reading or ±0.02 inWC
(±0.05 mB), whichever is greater
* Tighter CO accuracy in the lower ranges, up to ±2 ppm, may be
attained if a lower range calibration gas (e.g. 100 ppm CO) is used.
4
Instruction 0024-9395
Fyrite Pro
3.0 SETTING UP THE ANALYZER
3.1 Preliminary Steps
Before using the analyzer . . .
• Check batteries (Section 3.2)
• Connect probe to analyzer (Section 3.3)
• Check setup (Section 3.5)
3.2 Checking & Replacing the Bat ter ies
Install fresh batteries as described below. Check the analyzer for sufficient
charge prior to each use. Replace the batteries if the low-battery symbol
appears in the lower right corner of the screen. To re place the bat-
ter ies:
1. Remove battery cover from back of analyzer.
2. If old batteries are installed, remove them and properly discard.
3. Observing the polarity markings inside the battery com part ment,
install four ‘A A’ Al ka line bat ter ies as shown in Fig ure 3-1.
4. Replace battery cover.
Remove battery cover by
pushing down on cover
and sliding it outward
Figure 3-1. Battery Installation
Instruction 0024-9395
5
Fyrite Pro
3.3 Connecting the Probe Hose and
Thermocouple
The analyzer’s model number determines which hose and thermocouple
connections are made at the bottom of the unit. Perform the following steps
as required by your particular analyzer model (see Figures 3-2A thru D):
1. Push the probe’s sample-gas hose onto the GAS inlet fitting.
2. Push the probe’s draft hose onto the “+” pressure fitting.
3. Push the probe’s thermocouple into the T-STACK jack
Important: DO NOT force the thermocouple connector into
its jack. The con nec tion tabs are dif fer ent sizes, allowing the
connector to fit in only one way.
4. Push the optional room-air/primary-air thermocouple into the
T-AIR jack.
Inspect the flue-gas hose for cracks. If a hose is defective, replace the entire probe assembly.
Before using the analyzer, check that the Water Trap / Filter is dry and
not dirty. If necessary, dry out the trap and replace the filter el e ment per
Section 5.8.
6
Instruction 0024-9395
PROBE TUBE
Fyrite Pro
AC ADAPTER
JACK
GAS INLET
SAMPLE-GAS
HOSE
PROBE STOP
(Optional)
WATER TRAP /
FILTER ASSEMBLY
Figure 3-2A. Connecting the 0019-7110 Probe (Model 100)
Instruction 0024-9395
7
AC ADAPTER
JACK
Fyrite Pro
GASDRAFT
DIFFERENTIAL
PRESSURE HOSE
(Optional)
THERMOCOUPLE
CONNECTOR
(Not used on the
Model 105)
SAMPLE-GAS
HOSE
WATER
TRAP / FILTER
ASSEMBLY
DRAFT HOSE
PROBE TUBE
PROBE STOP
PROBE
HANDLE
Figure 3-2B. Connecting the 0024-3004 Probe (Model 105)
8
Instruction 0024-9395
Fyrite Pro
PROBE TUBE
PROBE STOP
(Optional)
AC ADAPTER
JACK
GAS
INLET
SAMPLE-GAS
HOSE
T-STACK
THERMOCOUPLE
CONNECTOR
(T-STACK)
WATER TRAP /
FILTER ASSEMBLY
Figure 3-2C. Connecting the 0019-7111 Probe (Model 110/120)
Instruction 0024-9395
9
AC ADAPTER
JACK
Fyrite Pro
GASDRAFT
T-STACK
T-AIR
ROOM AIR /
PRIMARY AIR
THERMOCOUPLE
CONNECTOR
(T-AIR)
DIFFERENTIAL
PRESSURE HOSE
(Optional)
SAMPLE GAS
THERMOCOUPLE
CONNECTOR
(T-STACK)
SAMPLE-GAS
HOSE
WATER
TRAP / FILTER
ASSEMBLY
DRAFT HOSE
PROBE TUBE
PROBE STOP
PROBE
HANDLE
Figure 3-2D. Connecting the 0024-3004 Probe (Model 125)
10
Instruction 0024-9395
Fyrite Pro
3.4 Front Panel Push Buttons
Note that a push button may perform several functions, depending on the
analyzer’s model number and what screen is being displayed at the time.
I/O• Toggles the analyzer ON and OFF.
• Places the analyzer into either its Setup or Calibration Mode
when used in conjunction with the ENTER or HOLD button.
▲• Scrolls up through the display screens during a test.
• Causes the displayed value to increase or change while in the
Calibration, Setup, or Print/Memory Menu Screen.
▼• Scrolls down through the display screens during a test.
• Causes the displayed value to decrease or change while in the
Calibration, Setup, or Print/Memory Menu Screen.
ENTER • Enters the Run Mode (starts a combustion test) from the Fuel
Select Screen.
• Unfreezes the display after pressing the HOLD button.
• Displays the Fuel Select Screen when held down for 2 seconds
while in the Run Mode.
• Performs the action selected in the Print/Memory Menu Screen.
• Stores the displayed value and automatically steps to the next
screen when pressed during calibration or setup.
• Places the unit into its Run Mode when held down for 2 seconds
while in the Calibration Mode.
• Aborts turn-off and keeps the analyzer turned ON when
pressed during the 5 second turn-off-delay period.
• Aborts the CO purge function at turn-off when the measured
CO level is above 50 ppm.
• Sets up the analyzer to be placed into its Calibration Mode
when held down with the analyzer OFF. (Used in conjunction
with the I/O button.)
HOLD• Places the analyzer on hold and freezes the values in all Run
Mode Screens during a test, allowing the op er a tor to scroll
through the displays and view all test values at that point in
time. Pressing ENTER resumes testing.
• Model 100: Starts a printout when pressed twice from the Run Mode.
• Model 105/110/120/125: Displays the Print/Memory Menu
Screen when pressed twice from the Run Mode.
• Sets up the analyzer to be placed into its Setup Mode when held down
with the analyzer OFF. (Used in conjunction with the I/O button.)
LIGHT
Instruction 0024-9395
Toggles the back light ON and OFF.
11
Fyrite Pro
3.5 Setup Mode
The analyzer is preset at the factory for the parameters shown below, but
can be changed as described in their associated sec tions.
Function Parameter To Change
Fuel Natural Gas (F1) Section 4.13.5
Temperature Unit °F Section 3.5.2
CO Channel Auto Zero (A2) Section 3.5.3
Draft (Pressure) WC Section 3.5.4
Clock Not Initialized Section 3.5.5 thru 3.5.7
Printer Protocol IrDA Section 3.5.8
3.5.1 Entering Setup
1. With the analyzer turned OFF, press and hold down the HOLD
button.
2. Press the I/O button.
3. Release both buttons.
4. The analyzer is now in its Setup Mode. Refer to Sec tions 3.5.2 thru
3.5.8 for information on how to set up each analyzer parameter.
3.5.2 Selecting Temperature Units
The Temperature Setup Screen is labeled “Unit.”
1. Enter the Setup Mode per Section 3.5.1. If necessary, repeatedly
press the EN TER button until “Unit” is dis played.
2. Press the ▲ or ▼ button until the desired temperature unit (°F or
°C) is displayed.
3. Press ENTER to move to the next Setup Screen, or I/O to exit setup.
12
Instruction 0024-9395
Fyrite Pro
3.5.3 CO ChannelSetup
The CO Channel Setup Screen is labeled “CO.”
1. Enter the Setup Mode per Section 3.5.1. If necessary, repeatedly
press the ENTER button until “CO” is displayed.
2. Press the ▲ or ▼ button until the desired parameter is dis played.
- no Sensor disabled. Use this function to turn off the CO channel if the sensor is missing or known to be bad to avoid inaccurate CO readings from appearing.
- 2Ero Manual Zero. When the CO channel is set to man u al zero,
the analyzer does not zero the CO sensor to ambient conditions during start up.
- A2 Auto Zero. With the CO channel set to auto zero, the CO
sensor is zeroed to the ambient CO level during start up.
Important: When using this mode the analyzer must
be turned ON in fresh air; otherwise, in cor rect CO
readings will occur.
3. Press ENTER to move to the next Setup Screen, or I/O to exit setup.
3.5.4 Draft Channel Setup
The Draft Setup Screen is labeled “PrES Unit”.
1. Enter the Setup Mode per Section 3.5.1. If necessary, repeatedly
press the ENTER button until “PrES” is displayed.
2. Press the ▲ or ▼ button until the desired parameter is dis played.
- no Pressure sensor disabled or not installed.
- WC inches of Water Column
- mB Millibars
- HPa Hecto Pascals
3. Press ENTER to move to the next Setup Screen, or I/O to exit setup.
Instruction 0024-9395
13
Fyrite Pro
3.5.5 Time Setup
There are two Time Setup Screens, one for hours and the other for minutes. Two bars appear above the segments being changed. The clock is in
a 24 hour format, but will appear as AM/PM on the printout.
1. Enter the Setup Mode per Section 3.5.1. If necessary, repeatedly
press ENTER until the first Time Setup Screen is displayed—the
one with two bars over the hour digits.
2. Press the ▲ or ▼ button until the correct hour value is displayed.
3. Press ENTER to move the selection bars over the minute digits.
4. Press the ▲ or ▼ button until the correct minute value is displayed.
5. Press ENTER to move to the next Setup Screen, or I/O to exit setup.
3.5.6 Date Setup
There are two Date Setup Screens, each labeled “DAtE”. The first screen
sets the month while the second sets the day.
1. Enter the Setup Mode per Section 3.5.1. If necessary, repeatedly
press ENTER until the first Date Setup Screen is displayed.
2. Press the ▲ or ▼ button until the correct month is displayed.
3. Press ENTER to change the ‘Day’ value.
4. Press the ▲ or ▼ key until the correct day is displayed.
5. Press ENTER to move to the next Setup Screen, or I/O to exit setup.
14
Instruction 0024-9395
Loading...
+ 44 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.