Bacharach Fyrite INSIGHT User Manual

Fyrite® INSIGHT
Combustion Gas Analyzer
Instruction 24-9460
Operation & Maintenance
Rev. 0 – Oct. 2008
RUN
ESC
HOLD
Product Leadership • Training • Service • Reliability
PWR
Fyrite® INSIGHT
WARRANTY
Bacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in mater ial and manufac ture and will conform substan tially to Bach arach Inc.’s ap plicable spec i fi ca tions. Bacharach’s li a bil i ty and Buyer’s remedy under this warranty are lim it ed to the repair or replacement, at Ba charach’s option, of this Product or parts thereof returned to Seller at the factory of manufac ture and shown to Bacharach Inc.’s reasonable satisfaction to have been defective; provided that written notice of the defect shall have been given by Buyer to Bacharach Inc. within two (2) years after the date of delivery on Product and all gas sensors except the O2 sensor which has a one (1) year warranty by Bacharach, Inc.
Bacharach, Inc. warrants to Buyer that it will convey good title to this Product. Bacharach’s liability and Buyer’s remedy under this warranty of title are limited to the removal of any title defects or, at the election of Bacha rach, to the replacement of this Product or parts thereof that are defective in title.
THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN AND ACCEPTED IN LIEU OF (I) ANY AND ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, IN CLUD ING WITH OUT LIMITATION THE IM­PLIED WARRANTIES OF MERCHAN TABILITY AND FITNESS FOR A PARTICULAR PURPOSE: AND (II) ANY OBLIGA TION, LIABILITY, RIGHT, CLAIM OR REM E DY IN CON TRACT OR TORT, WHETHER OR NOT ARIS­ING FROM BACH A RACH’S NEGLIGENCE, ACTUAL OR IM PLIED. The remedies of the Buyer shall be limited to those provided herein to the exclusion of any and all other remedies includ ing, without limita tion incidental or consequen tial dam ag es. No agree ment varying or ex tend ing the foregoing warranties, remedies or this limita­tion will be binding upon Bacharach, Inc. unless in writing, signed by a duly au tho rized officer of Bacharach.
Register your warranty by visiting
www.bacharach-inc.com
TABLE OF CONTENTS
1.0 Introduction ................................................................................. 1
1.1 The Fyrite® INSIGHT ...............................................................1
1.2 Fyrite
1.3 Operational Overview ................................................................4
2.0 Technical Characteristics ......................................................... 5
3.0 Setting Up the Analyzer ............................................................ 7
3.1 Preliminary Steps ......................................................................7
3.2 Power ..........................................................................................7
3.2.1 Installing or Replacing Batteries...........................................7
3.2.2 Using the AC Power Adapter .................................................7
3.3 Connecting the Probe Hose & Thermocouple ...........................7
3.4 Front Panel Push Buttons .......................................................10
3.5 Setup Mode ..............................................................................11
3.5.1 Entering Setup ......................................................................11
3.5.2 Temperature Units ...............................................................11
3.5.3 Pressure Units ......................................................................11
3.5.4 Clock ......................................................................................12
3.5.5 O2 Reference ..........................................................................13
3.5.6 Print Pressure ......................................................................13
3.5.7 Zoom ......................................................................................13
3.5.8 Username ..............................................................................14
3.5.9 Run/Hold Format ..................................................................15
3.5.10 Language Selection ..............................................................17
3.5.11 Button Sound .......................................................................17
3.5.12 CAL Reminder Period ......................................................... 17
3.5.13 Inactivity Timeout ...............................................................18
3.5.14 Post-Purge Period ................................................................19
3.5.15 Date Format .........................................................................19
®
INSIGHT Features & Benefits ....................................2
Notice:
Product improvements and enhancements are continuous, therefore the spec i fi ca tions and in for ma tion contained in this document may change without notice.
Bacharach, Inc. shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of Bacharach, Inc.
Copyright © 2002–2008, Bacharach, Inc., all rights reserved. BACHARACH, Fyrite, INSIGHT & B-SMART are registered trademarks of Bacharach, Inc. All other trademarks, trade names, service marks and logos referenced herein belong to their respective companies.
A
Instruction 24-9460
4.0 Operation .................................................................................... 20
4.1 Operating Tips ........................................................................ 20
4.2 Analyzer Power ON and Warm Up .........................................21
4.3 Fuel Selection...........................................................................21
4.4 Sampling Point ....................................................................... 22
4.5 Performing a Combustion Test .............................................. 23
4.6 Making a Draft/Pressure Measurement ............................... 25
4.7 Temperature Measurement .................................................... 26
4.8 Saving Test Data .....................................................................27
4.9 Ending a Combustion Test ......................................................27
Instruction 24-9460
i
Fyrite® INSIGHT
Fyrite® INSIGHT
4.10 Emptying the Water Trap.......................................................27
4.11 Powering OFF the Analyzer & Purging ............................... 28
4.12 Low Battery Alarm ................................................................ 28
4.13 Memory ................................................................................... 28
4.14 Printing Test Data ................................................................. 30
®
4.15 Fyrite
User Software Installation ..................................... 31
4.16 Fyrite® User Software Description ....................................... 34
5.0 Calibration & Maintenance ......................................................35
5.1 B-Smart Sensors ...................................................................... 35
5.2 Starting a Calibration ............................................................. 35
5.3 B-Smart Sensor Replacement & Calibration ......................... 37
5.4 Pressure Sensor Calibration ................................................... 37
5.5 T-Stack Calibration.................................................................. 39
5.6 T-Air Calibration ..................................................................... 41
5.7 CO Sensor Calibration ............................................................. 43
6.0 Maintenance ................................................................................. 44
6.1 Fyrite® INSIGHT Disassembly ............................................. 44
6.2 Water Trap / Filter .................................................................46
6.2.1 Emptying the Water Trap Chamber .................................... 46
6.2.2 Replacing the Filter Element ............................................... 46
6.3 O2 Sensor Replacement .......................................................... 47
6.4 CO Sensor Replacement ......................................................... 48
6.5 Thermocouple Replacement .................................................... 49
6.6 Gas Pump Replacement .......................................................... 50
6.7 Cleaning the Probe ................................................................. 50
6.8 Error Messages ....................................................................... 51
6.9 Diagnostics Screen ................................................................. 52
6.10 Status Screen ......................................................................... 53
7.0 Parts & Service ............................................................................ 54
7.1 Replacement Parts .................................................................. 54
7.2 Accessories .............................................................................. 54
7.3 Service Centers ....................................................................... 55
1.0 INTRODUCTION
1.1 The Fyrite® INSIGHT
The Fyrite® INSIGHT is a hand-held residential and commercial grade portable combustion an a lyz er that is designed for on-demand sampling of light industrial combustion equipment, residential furnaces, and appli­ances.
The Fyrite® INSIGHT is upgradeable from the Basic model to the Record­ing model which includes additional features such as software compatible for downloading data and the IrDA printer.
This manual contains information on all models of the Fyrite INSIGHT. Please disregard any information that does not pertain to your model.
WARNINGS!
This analyzer is not intended to be used on a continuous basis.
This analyzer is not in tend ed to be used as a safety device.
Except for sensor and battery replacement, this analyzer should only be opened and ser viced by au tho rized person­nel.
When testing an appliance, a full visual in spec tion of the appliance should be carried out to ensure its safe op er a tion.
Basic
24-8250 24-8251 HARD CARRY CASE X X PROBE ASSEMBLY X X PRINTER IRDA X BOOT X X PC SOFTWARE X USB CABLE X BATTERIES X X QUICK START GUIDE X X
Reporting
ii
Instruction 24-9460
Instruction 24-9460
1
Fyrite
®
INSIGHT
Fyrite® INSIGHT
1.2 Fyrite® INSIGHT Features & Benefi ts
AUTOMATIC ZERO - The instrument will automatically zero all sensing channels on ambient air when it is powered on. If a particular sensor is in error during warm-up, the instrument will automatically display the error and continue to operate with the sensor in error. However, all information dependent on the sensor in error will not be presented.
TEMPERATURE UNITS - Temperatures can be displayed in Centigrade or Fahrenheit.
PRESSURE UNITS - Pressures can be displayed in milliBar (mB), Pascal (Pa), hectoPascal (hPa), Millimeters of Water Column (mmwc), or Inches of Water Column (inwc).
LANGUAGE - The display options are English, French or Spanish.
®
MEMORY - The Fyrite play, and print a minimum of 100 sets of time and date coded combustion, differential pressure, or differential temperature test records.
DOWNLOAD - Test records can be downloaded to a personal computer via the USB port. Saved records are transferred using PC based software and will be automatically loaded into an Excel spreadsheet. PC based software will also have the capability to update the instrument time and date and upload custom fuels, username, customer logo, and future code updates. The INSIGHT can also be set up from PC software.
USER INFORMATION - The instrument will accept up to three lines of 20 characters of user information. This information will appear with the test records when they are printed or downloaded.
INSIGHT has the ability to store, recall to dis-
DISPL AY ZOOM - Combustion test data in the Run/Hold screens can be shown with enlarged characters to make viewing easier. The complete list of Run/Hold data can be scrolled through for all zoom levels. Available zoom levels and information display are as follows:
• Standard - 4 lines of information
• 2X - 3 lines of enlarged characters
• 3X - 2 lines of enlarged characters
STATUS - A status menu displays the instrument software version, boot version, model number, and serial number.
DIAGNOSTIC S - The diagnostics menu displays the O2 sensor life, CO sensor information, and hours of operation for the instrument and the pump.
KEYPAD SOUND - Keypad sounds can be enabled or disabled with soft­ware menu selections.
LOW BATTERY - A tone will sound every 10 seconds when a low battery condition exists. A low battery message will also be displayed.
AUTOM ATIC POW ER DOWN - The instrument can be set to automati­cally power off after a period of inactivity unless the user presses any but­ton; the CO is greater than 50 ppm, or the O2 is less than 16.0%. Auto­matic power down can be set for None, 30, or 60 minutes.
PURGE - The sensors will be purged when the instrument is shutdown until sensor outputs indicate that they have been exposed to fresh air. Pressing the On/Off key a seconds time can defeat the purge routine. The minimum purge time can be set to None, 5 seconds, 1 , 5, or 10 minutes.
CALI BRATION - Calibration can be performed by applying gas and enter­ing calibration gas concentrations through software menu selections.
SMART SENSORS - B-Smart sensor technology is employed for the CO channel. CO sensors are calibrated prior to shipment and provide data that can be entered through the software menu selections for easy calibration without applying gas.
CALIBRATION REMINDERS - Calibration reminders can be set for Never, 6, 8, 10, 12, or 15 months.
DISPL AY FORMAT - The Run/Hold display format can be presented in default or custom ordered for specifi c needs. Up to 8 lines can be custom- ized in the Run/Hold screen.
2
Instruction 24-9460
SENSOR ACCESS - Sensor replacement can be accomplished in the fi eld. Easy access to the O2 and CO sensors is provided through the back of the instrument.
POWER - The Fyrite
®
INSIGHT is powered by 4 AA alkaline batteries. A new set of alkaline batteries will provide a minimum of 20 hours of con­tinuous operation with the pump running and the back light on. Option­ally, the analyzer can operate with a universal AC adapter. When the AC adapter is plugged into the unit, batteries do not have to be removed.
Instruction 24-9460
3
Fyrite
®
INSIGHT
Fyrite® INSIGHT
1.3 Operational Overview
Pressing the PWR button turns the analyzer ON. Note that there is a
warm-up period of 60 seconds that must elapse before the analyzer can be used.
To perform a combustion test, choose a fuel code that corresponds to the fuel being burned by the appliance being tested (Section 4.3), and then
press the RUN/HOLD button to place the analyzer into its Run Mode.
Begin testing by inserting the analyzer’s probe tube into the flue-gas stream of the appliance under test. The analyzer will monitor the flue gas and display measured and calculated values that are relative to the combustion process. These values are dis played on the analyzer’s LCD, and are chosen for display by pressing the In cre ment () and Dec re ment () but tons.
Pressing the RUN/HOLD button during a test will freeze all measured and calculated values at their current levels. Pressing the RUN/HOLD button
again resumes testing.
A backlight enables an operator to read the display in dimly-lit areas.
Pressing the PWR button turns the backlight ON and OFF.
An optional power saver function will shut the analyzer OFF after a set time period of inactivity. This feature is dis abled if the CO value is greater than 50 ppm or the O
Press the PWR button for at least two seconds to power the analyzer
OFF. Note that there is a 5 second delay before the analyzer actually turns OFF, during which time the unit can be kept ON by pressing the
RUN/HOLD button. There is a CO purge feature that keeps the analyzer’s
pump running until the measured CO level drops below 50 ppm.
value is less than 16.0%.
2
2.0 TECHNICAL CHARACTERISTICS
The Analyzer measures & displays:
Primary/Ambient Air Temperature -4 to 600 °F (-20 to 316°C) Oxygen 0.1 to 20.9% O
Stack Temperature -4 to 1202 °F (–20 to 650 °C) Carbon Monoxide 0 to 2000 ppm CO
Differential Pressure/Draft ± 100 inwc
The Analyzer computes & displays:
Combustion Efficiency 0.1 to 100% Carbon Dioxide
0.1 to a fuel dependent
maximum value in % CO referenced to %O2 0 to 9,999 ppm Excess Air 0 to 250%
Fuel Selection
• Natural Gas
• Oil #2
• Oil #4
• Oil #6
• Propane
• Kerosene
• Wood
• Coal
Normal Operating Conditions:
Temperature
• Analyzer
• Probe
32 to 104 °F (0 to 40 °C)
1,000 °F max (538 °C) at
5" insertion
2
Humidity 15 to 90% RH, Non-Condensing
Dimensions Weight
7.9 in (20.1 cm) Height
1lb. (0.45 kg)
3.6 in (9.1 cm) Width
2.3 in (5.8 cm) Depth
4
Instruction 24-9460
Instruction 24-9460
5
Fyrite
®
INSIGHT
Fyrite® INSIGHT
General Characteristics:
Power Requirements Four disposable ‘AA’ Alkaline batteries (op-
tional AC adapter is available)
Operating Time Minimum of 20 hours con tin u ous operation
(pump run ning and backlight ON)
Warm Up Time 60 seconds Memory Up to 100 complete sets of
combustion, temperature, & pressure tests can be saved in memory.
Display 128 x 64, LCD Graphic Dis play Front Panel Controls Eleven front panel push buttons
(Refer to Section 3.4)
Interfaces IrDA printer & USB connectivity Accuracy:
Oxygen ±0.3% O
concentration of CO2
Carbon Monoxide ±5% of reading or ±10 ppm, whichever is
greater*, in the range of 0 to 1000 ppm.
Flue Gas Temperature ±4 °F between 32 & 255 °F
(±2 °C between 0 & 124 °C) ±6 °F between 256 & 480 °F (±3 °C between 125 & 249 °C) ±8 °F between 481 & 752 °F (±4 °C between 250 & 400 °C)
Ambient/Primary Temperature
Draft ±2% of reading or ±0.02 InWC, whichever is
* Tighter CO accuracy in the lower ranges, up to ±2 ppm, may be
attained if a lower range calibration gas (e.g. 100 ppm CO) is used.
±2 °F between 32 & 212 °F (±1 °C between 0 & 100 °C
greater in the range of -1 0 to + 1 0 InW C ± 3% in the range of -1 0 to -40 InWC ± 3% in the range of 10 to 40 InWC
with a typical flue gas
2
3.0 SETTING UP THE ANALYZER
3.1 Preliminary Steps
Before using the analyzer . . .
• Check batteries (Section 3.2)
• Connect probe to analyzer (Section 3.3)
• Check setup (Section 3.5)
3.2 Power
3.2.1 Installing or Replacing Batteries
Install fresh batteries as described below. Check the analyzer for sufficient charge prior to each use. Replace the batteries if the low-battery symbol appears in the upper right corner of the screen. To re place the bat ter ies:
1. Remove battery cover from back of analyzer.
2. If old batteries are installed, remove them and properly discard.
3. Observing the polarity markings inside the battery com part ment, install four ‘AA’ Al ka line bat ter ies.
4. Replace battery cover.
3.2.2 Using the AC Power Adapter
The optional AC power adapter is capable of powering the analyzer on a continuous basis. The adapter plugs into an appropriate 100-240 VAC wall outlet and produces an output of +5 VDC. The adapter's output connector plugs into the analyzer's POWER jack located on the bottom of the unit. The batteries do not need to be removed when using the AC adapter.
3.3 Connecting the Probe Hose & Thermo­couple
A rigid stainless steel probe with handle is connected to a flexible hose with integral water-trap / filter used to draw a gas sample into the ana­lyzer from the room, grilles, diffusers, and furnace flues.
6
Instruction 24-9460
Instruction 24-9460
7
Fyrite
®
INSIGHT
Fyrite® INSIGHT
Inspect the flue-gas hose for cracks. If a hose is defective, replace the en­tire probe assembly.
Before using the analyzer, check that the Water Trap / Filter is dry and not dirty. If necessary, dry out the trap and replace the filter el e ment per Section 6.2.
Perform the following steps to connect the probe assembly to the Fyrite INSIGHT:
1. Push the probe’s sample-gas hose onto the GAS inlet fitting.
2. Push the probe’s draft hose onto the “+” pressure fitting.
3. Push the probe’s thermocouple into the T-STACK jack
Important: DO NOT force the thermocouple connector into
its jack. The con nec tion tabs are dif fer ent sizes, allowing the connector to fit in only one way.
4. Push the optional ambient/primary-air thermocouple into the T-AIR jack.
AC Power Adapter Jack (Power)
Differential Pressure Hose (Optional)
Fyrite® INSIGHT Connectors
Draft Hose
Sample Gas Thermocouple Connector (T-Stack)
Probe Tube
USB Cable
Primary Ambient Air Thermocouple (T-Air) (Optional)
Sample Gas Hose
Probe Stop
Probe Handle
Water Trap/ Filter Assembly
Fig. 3.1
8
Instruction 24-9460
Instruction 24-9460
9
Fyrite
®
INSIGHT
Fyrite® INSIGHT
3.4 Front Panel Push Buttons
Note that a push button may perform several functions, depending on the analyzer’s model number and what screen is being displayed at the time.
Powers the analyzer ON and OFF. Hold this button down for at least 2 seconds to turn the power OFF.
Toggles the backlight ON and OFF while the analyzer is turned ON.
Scrolls up, down, left, and right through the display screen options.
• Up and Down arrows cause the displayed value to in- crease or decrease correspondingly.
Acts as an ENTER button. Performs the action selected.
• While in the HOLD screen, turns the sample pump on, dis- plays the RUN screen, and begins a combustion test.
• While in the RUN screen, turns the sample pump off, displays the HOLD screen and the last set of combustion data.
• Displays the HOLD screen while pressing it from most menus.
• Return the display to the HOLD screen while pressing it during the 5 second power down sequence.
Cancels most operations and displays the previous screen.
Pressing function keys accepts the corresponding function
defined at the bottom of the display, including PRINT , SA VE, MENU, differential pressure ZERO, differential temperature ZERO, P AGE UP, PAGE DOWN and CLEAR data.
3.5 Setup Mode
The analyzer is preset at the factory for the parameters shown below, but can be changed as described in their corresponding sec tions.
Function Parameter To Change
Fuel Natural Gas Section 4.3 Temperature Unit °F Section 3.5.2
Pressure Unit InWC Section 3.5.3 Language English Section 3.5.10 CAL Reminder Never Section 3.5.12 Clock Not Initialized Section 3.5.4
3.5.1 Entering Setup
1. Display the MAIN MENU by pressing the
MENU (F2) button.
2. Use the ▲▼ buttons to highlight SETUP and press ENTER to display the SETUP
MENU.
3.5.2 Temperature Units
Select to display temperature in either ºF or ºC as follows:
1. Enter the Setup Menu per Section 3.5.1. If necessary, press ESC
until MENU appears above F2.
2. From the SETUP MENU, use the ▲▼ but­tons to select TEMPERATURE UNITS.
Press ENTER to display the Temperature
Units Menu.
3. Use the ▲▼ buttons to select the desired temperature unit.
10
Instruction 24-9460
4. Press ENTER to save selection and return to the SETUP MENU or ESC to revert the
previous setting.
Instruction 24-9460
11
®
Fyrite
INSIGHT
3.5.3 Pressure Units
Select to display pressure in inches of water column (inwc), millibar (mb), Pascals (Pa), hectoPascals (hPa), or millimeter of water column (mmwc) as follows:
Fyrite® INSIGHT
3.5.5 O2 Reference
The measured value of CO can be referenced to a specific O2 percentage of 0% to 15%. Set up the reference value as follows:
1. Enter the Setup Menu per Section 3.5.1. If necessary, press ESC until MENU appears above F2.
2. From the SETUP MENU, use the ▲▼ but­tons to select PRESSURE UNITS. Press
ENTER to display the Pressure Units Menu.
3. Use the ▲▼ buttons to selecte the desired pressure unit.
4. Press ENTER to save selection and return to the SETUP Menu or ESC to revert to the
previous setting.
3.5.4 Clock
The clock settings (12 hr. AM/PM or 24 hr.) and date format options (MM/ DD/Y Y or DD/MM/Y Y) are interdependent. (See DATE FORMAT, Section
3.5.15).
1. Enter the Setup Menu per Section 3.5.1. If necessary, press ESC
until MENU appears above F2.
2. From the SETUP MENU, use the▲▼ but-
tons to select CLOCK. Press ENTER to
display the Clock Setup Screen
3. Use the ▲▼ buttons to enter the current date and time. Use the the ◄► buttons to move the cursor across the screen.
4. Press ENTER to save the selection and re­turn to the SETUP MENU or ESC to revert
to the previous setting.
1. Enter the Setup Menu per Section 3.5.1. If
necessary, press ESC until MENU appears above F2.
2. From the SETUP MENU, use the ▲▼ buttons to select O2 REFERENCE. Press
ENTER to display the O2 Reference Screen.
3. Use the ▲▼ buttons to enter the O2 refer­ence value.
4. Press ENTER to save the selection and re­turn to the SETUP MENU or ESC to revert
to the previous setting.
3.5.6 Print Pressure
Select whether to print or not print the pressure measurement on the com­bustion test printout as follows:
1. Enter the Setup Menu per Section 3.5.1. If necessary, press ESC
until MENU appears above F2.
2. From the SETUP MENU, use the ▲▼ but­tons to select PRINT PRESSURE. Press
ENTER to display the O2 Reference Screen.
3. Use the ▲▼ buttons to select YES or NO to print the pressure measurement.
4. Press ENTER to save the selection and re­turn to the SETUP MENU or ESC to revert
to the previous setting.
3.5.7 Zoom
Combustion test data in the Run/Hold screen can be shown with enlarged characters to make viewing easier. The operator can set zoom levels to Standard, 2X or 3X. The Standard zoom setting will display 4 lines of combustion test data at one time; 2X will display 3 lines of data with en­larged characters; 3X will display 2 lines of data with enlarged characters. The operator can scroll through the complete list of measured and calcu-
12
Instruction 24-9460
Instruction 24-9460
13
Fyrite
®
INSIGHT
Fyrite® INSIGHT
lated data no matter what zoom level has been selected. Select the desired zoom level as follows:
1. Enter the Setup Menu per Section 3.5.1. If necessary, press ESC until MENU appears above F2.
2. From the SETUP MENU, use the ▲▼
buttons to select ZOOM. Press ENTER to
display the Zoom Menu.
3. Use the ▲▼ buttons to select the desired zoom level.
4. Press ENTER to save the selection and dis­play the SETUP MENU or ESC to revert to
the previous setting.
3.5.8 Username
The name of the user or owner (e.g. company name, address, phone num­ber) can be stored in memory by manually entering up to three lines of text, with each line containing up to 20 alphanumeric characters. This information will appear at the top of each printout, until it is cleared or new information is entered.
Username Information can be entered as follows:
1. Enter the Setup Menu per Section 3.5.1. If necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to select USERNAME.
Press ENTER to display the most recently
saved information.
6. Repeat Steps 3-5 to edit the remaining lines.
7. Once the input is complete, use the ▲▼ buttons to select EDIT COMPLETE.
8. Press ENTER to accept the entered information or ESC to revert to
the previous information and display the SETUP MENU.
Helpful Hint: Use the Fyrite® user software to upload this informa-
tion from your computer.
To Clear a Username proceed as follows:
1. Enter the SETUP MENU per Section 3.5.1. If necessary, press ESC
until MENU appears above F2.
2. Use the ▲▼ buttons to select USERNAME. Press ENTER to dis-
play the most recently saved information.
3. Clear the information by selecting the CLEAR option (F3).
3.5.9 Run/Hold Format
Fyrite INSIGHT test data is located in the Run/Hold screen. By press-
ing the RUN/HOLD key, you should hear the pump running and see the
word RUN at the upper-left hand corner of the display. The instrument is continuously measuring and calculating the data that is shown in the
Run/Hold screen. Press the RUN/HOLD key again. The pump should stop
running and the word HOLD should be shown at the upper-left hand cor­ner of the display. The instrument will now show the last measured and calculated data taken before the instrument was placed in HOLD. Use the ▲▼ buttons to scroll through the complete list of measured and calculated values when the instrument is running or in the HOLD mode.
3. Use the ▲▼ buttons to select the line of in­formation which you would like to edit and
press ENTER. A single cursor will appear
denoting the character you are changing.
4. Use the ▲▼ buttons to scroll the the list of alphanumeric characters and the ◄► but­tons to move across the screen.
5. Press ENTER to save the selection or ESC to
revert to the previous input.
14
Instruction 24-9460
The order in which data appears in the RUN/HOLD screen is as follows:
O2 = Oxygen CO = Carbon Monoxide Eff = Combustion Effi ciency CO2 = Carbon Dioxide T-STK = Stack Temperature T-AIR = Ambient/Primary Air Temperature EA = Excess Air CO(n) = Carbon Monoxide content referenced to an Oxygen percentage
NOTE: (n) is the current O2 Reference selected.
Instruction 24-9460
15
Loading...
+ 21 hidden pages