Bacharach Fyrite INSIGHT User Manual

Fyrite® INSIGHT
Combustion Gas Analyzer
Instruction 24-9460
Operation & Maintenance
Rev. 0 – Oct. 2008
RUN
ESC
HOLD
Product Leadership • Training • Service • Reliability
PWR
Fyrite® INSIGHT
WARRANTY
Bacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in mater ial and manufac ture and will conform substan tially to Bach arach Inc.’s ap plicable spec i fi ca tions. Bacharach’s li a bil i ty and Buyer’s remedy under this warranty are lim it ed to the repair or replacement, at Ba charach’s option, of this Product or parts thereof returned to Seller at the factory of manufac ture and shown to Bacharach Inc.’s reasonable satisfaction to have been defective; provided that written notice of the defect shall have been given by Buyer to Bacharach Inc. within two (2) years after the date of delivery on Product and all gas sensors except the O2 sensor which has a one (1) year warranty by Bacharach, Inc.
Bacharach, Inc. warrants to Buyer that it will convey good title to this Product. Bacharach’s liability and Buyer’s remedy under this warranty of title are limited to the removal of any title defects or, at the election of Bacha rach, to the replacement of this Product or parts thereof that are defective in title.
THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN AND ACCEPTED IN LIEU OF (I) ANY AND ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, IN CLUD ING WITH OUT LIMITATION THE IM­PLIED WARRANTIES OF MERCHAN TABILITY AND FITNESS FOR A PARTICULAR PURPOSE: AND (II) ANY OBLIGA TION, LIABILITY, RIGHT, CLAIM OR REM E DY IN CON TRACT OR TORT, WHETHER OR NOT ARIS­ING FROM BACH A RACH’S NEGLIGENCE, ACTUAL OR IM PLIED. The remedies of the Buyer shall be limited to those provided herein to the exclusion of any and all other remedies includ ing, without limita tion incidental or consequen tial dam ag es. No agree ment varying or ex tend ing the foregoing warranties, remedies or this limita­tion will be binding upon Bacharach, Inc. unless in writing, signed by a duly au tho rized officer of Bacharach.
Register your warranty by visiting
www.bacharach-inc.com
TABLE OF CONTENTS
1.0 Introduction ................................................................................. 1
1.1 The Fyrite® INSIGHT ...............................................................1
1.2 Fyrite
1.3 Operational Overview ................................................................4
2.0 Technical Characteristics ......................................................... 5
3.0 Setting Up the Analyzer ............................................................ 7
3.1 Preliminary Steps ......................................................................7
3.2 Power ..........................................................................................7
3.2.1 Installing or Replacing Batteries...........................................7
3.2.2 Using the AC Power Adapter .................................................7
3.3 Connecting the Probe Hose & Thermocouple ...........................7
3.4 Front Panel Push Buttons .......................................................10
3.5 Setup Mode ..............................................................................11
3.5.1 Entering Setup ......................................................................11
3.5.2 Temperature Units ...............................................................11
3.5.3 Pressure Units ......................................................................11
3.5.4 Clock ......................................................................................12
3.5.5 O2 Reference ..........................................................................13
3.5.6 Print Pressure ......................................................................13
3.5.7 Zoom ......................................................................................13
3.5.8 Username ..............................................................................14
3.5.9 Run/Hold Format ..................................................................15
3.5.10 Language Selection ..............................................................17
3.5.11 Button Sound .......................................................................17
3.5.12 CAL Reminder Period ......................................................... 17
3.5.13 Inactivity Timeout ...............................................................18
3.5.14 Post-Purge Period ................................................................19
3.5.15 Date Format .........................................................................19
®
INSIGHT Features & Benefits ....................................2
Notice:
Product improvements and enhancements are continuous, therefore the spec i fi ca tions and in for ma tion contained in this document may change without notice.
Bacharach, Inc. shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of Bacharach, Inc.
Copyright © 2002–2008, Bacharach, Inc., all rights reserved. BACHARACH, Fyrite, INSIGHT & B-SMART are registered trademarks of Bacharach, Inc. All other trademarks, trade names, service marks and logos referenced herein belong to their respective companies.
A
Instruction 24-9460
4.0 Operation .................................................................................... 20
4.1 Operating Tips ........................................................................ 20
4.2 Analyzer Power ON and Warm Up .........................................21
4.3 Fuel Selection...........................................................................21
4.4 Sampling Point ....................................................................... 22
4.5 Performing a Combustion Test .............................................. 23
4.6 Making a Draft/Pressure Measurement ............................... 25
4.7 Temperature Measurement .................................................... 26
4.8 Saving Test Data .....................................................................27
4.9 Ending a Combustion Test ......................................................27
Instruction 24-9460
i
Fyrite® INSIGHT
Fyrite® INSIGHT
4.10 Emptying the Water Trap.......................................................27
4.11 Powering OFF the Analyzer & Purging ............................... 28
4.12 Low Battery Alarm ................................................................ 28
4.13 Memory ................................................................................... 28
4.14 Printing Test Data ................................................................. 30
®
4.15 Fyrite
User Software Installation ..................................... 31
4.16 Fyrite® User Software Description ....................................... 34
5.0 Calibration & Maintenance ......................................................35
5.1 B-Smart Sensors ...................................................................... 35
5.2 Starting a Calibration ............................................................. 35
5.3 B-Smart Sensor Replacement & Calibration ......................... 37
5.4 Pressure Sensor Calibration ................................................... 37
5.5 T-Stack Calibration.................................................................. 39
5.6 T-Air Calibration ..................................................................... 41
5.7 CO Sensor Calibration ............................................................. 43
6.0 Maintenance ................................................................................. 44
6.1 Fyrite® INSIGHT Disassembly ............................................. 44
6.2 Water Trap / Filter .................................................................46
6.2.1 Emptying the Water Trap Chamber .................................... 46
6.2.2 Replacing the Filter Element ............................................... 46
6.3 O2 Sensor Replacement .......................................................... 47
6.4 CO Sensor Replacement ......................................................... 48
6.5 Thermocouple Replacement .................................................... 49
6.6 Gas Pump Replacement .......................................................... 50
6.7 Cleaning the Probe ................................................................. 50
6.8 Error Messages ....................................................................... 51
6.9 Diagnostics Screen ................................................................. 52
6.10 Status Screen ......................................................................... 53
7.0 Parts & Service ............................................................................ 54
7.1 Replacement Parts .................................................................. 54
7.2 Accessories .............................................................................. 54
7.3 Service Centers ....................................................................... 55
1.0 INTRODUCTION
1.1 The Fyrite® INSIGHT
The Fyrite® INSIGHT is a hand-held residential and commercial grade portable combustion an a lyz er that is designed for on-demand sampling of light industrial combustion equipment, residential furnaces, and appli­ances.
The Fyrite® INSIGHT is upgradeable from the Basic model to the Record­ing model which includes additional features such as software compatible for downloading data and the IrDA printer.
This manual contains information on all models of the Fyrite INSIGHT. Please disregard any information that does not pertain to your model.
WARNINGS!
This analyzer is not intended to be used on a continuous basis.
This analyzer is not in tend ed to be used as a safety device.
Except for sensor and battery replacement, this analyzer should only be opened and ser viced by au tho rized person­nel.
When testing an appliance, a full visual in spec tion of the appliance should be carried out to ensure its safe op er a tion.
Basic
24-8250 24-8251 HARD CARRY CASE X X PROBE ASSEMBLY X X PRINTER IRDA X BOOT X X PC SOFTWARE X USB CABLE X BATTERIES X X QUICK START GUIDE X X
Reporting
ii
Instruction 24-9460
Instruction 24-9460
1
Fyrite
®
INSIGHT
Fyrite® INSIGHT
1.2 Fyrite® INSIGHT Features & Benefi ts
AUTOMATIC ZERO - The instrument will automatically zero all sensing channels on ambient air when it is powered on. If a particular sensor is in error during warm-up, the instrument will automatically display the error and continue to operate with the sensor in error. However, all information dependent on the sensor in error will not be presented.
TEMPERATURE UNITS - Temperatures can be displayed in Centigrade or Fahrenheit.
PRESSURE UNITS - Pressures can be displayed in milliBar (mB), Pascal (Pa), hectoPascal (hPa), Millimeters of Water Column (mmwc), or Inches of Water Column (inwc).
LANGUAGE - The display options are English, French or Spanish.
®
MEMORY - The Fyrite play, and print a minimum of 100 sets of time and date coded combustion, differential pressure, or differential temperature test records.
DOWNLOAD - Test records can be downloaded to a personal computer via the USB port. Saved records are transferred using PC based software and will be automatically loaded into an Excel spreadsheet. PC based software will also have the capability to update the instrument time and date and upload custom fuels, username, customer logo, and future code updates. The INSIGHT can also be set up from PC software.
USER INFORMATION - The instrument will accept up to three lines of 20 characters of user information. This information will appear with the test records when they are printed or downloaded.
INSIGHT has the ability to store, recall to dis-
DISPL AY ZOOM - Combustion test data in the Run/Hold screens can be shown with enlarged characters to make viewing easier. The complete list of Run/Hold data can be scrolled through for all zoom levels. Available zoom levels and information display are as follows:
• Standard - 4 lines of information
• 2X - 3 lines of enlarged characters
• 3X - 2 lines of enlarged characters
STATUS - A status menu displays the instrument software version, boot version, model number, and serial number.
DIAGNOSTIC S - The diagnostics menu displays the O2 sensor life, CO sensor information, and hours of operation for the instrument and the pump.
KEYPAD SOUND - Keypad sounds can be enabled or disabled with soft­ware menu selections.
LOW BATTERY - A tone will sound every 10 seconds when a low battery condition exists. A low battery message will also be displayed.
AUTOM ATIC POW ER DOWN - The instrument can be set to automati­cally power off after a period of inactivity unless the user presses any but­ton; the CO is greater than 50 ppm, or the O2 is less than 16.0%. Auto­matic power down can be set for None, 30, or 60 minutes.
PURGE - The sensors will be purged when the instrument is shutdown until sensor outputs indicate that they have been exposed to fresh air. Pressing the On/Off key a seconds time can defeat the purge routine. The minimum purge time can be set to None, 5 seconds, 1 , 5, or 10 minutes.
CALI BRATION - Calibration can be performed by applying gas and enter­ing calibration gas concentrations through software menu selections.
SMART SENSORS - B-Smart sensor technology is employed for the CO channel. CO sensors are calibrated prior to shipment and provide data that can be entered through the software menu selections for easy calibration without applying gas.
CALIBRATION REMINDERS - Calibration reminders can be set for Never, 6, 8, 10, 12, or 15 months.
DISPL AY FORMAT - The Run/Hold display format can be presented in default or custom ordered for specifi c needs. Up to 8 lines can be custom- ized in the Run/Hold screen.
2
Instruction 24-9460
SENSOR ACCESS - Sensor replacement can be accomplished in the fi eld. Easy access to the O2 and CO sensors is provided through the back of the instrument.
POWER - The Fyrite
®
INSIGHT is powered by 4 AA alkaline batteries. A new set of alkaline batteries will provide a minimum of 20 hours of con­tinuous operation with the pump running and the back light on. Option­ally, the analyzer can operate with a universal AC adapter. When the AC adapter is plugged into the unit, batteries do not have to be removed.
Instruction 24-9460
3
Fyrite
®
INSIGHT
Fyrite® INSIGHT
1.3 Operational Overview
Pressing the PWR button turns the analyzer ON. Note that there is a
warm-up period of 60 seconds that must elapse before the analyzer can be used.
To perform a combustion test, choose a fuel code that corresponds to the fuel being burned by the appliance being tested (Section 4.3), and then
press the RUN/HOLD button to place the analyzer into its Run Mode.
Begin testing by inserting the analyzer’s probe tube into the flue-gas stream of the appliance under test. The analyzer will monitor the flue gas and display measured and calculated values that are relative to the combustion process. These values are dis played on the analyzer’s LCD, and are chosen for display by pressing the In cre ment () and Dec re ment () but tons.
Pressing the RUN/HOLD button during a test will freeze all measured and calculated values at their current levels. Pressing the RUN/HOLD button
again resumes testing.
A backlight enables an operator to read the display in dimly-lit areas.
Pressing the PWR button turns the backlight ON and OFF.
An optional power saver function will shut the analyzer OFF after a set time period of inactivity. This feature is dis abled if the CO value is greater than 50 ppm or the O
Press the PWR button for at least two seconds to power the analyzer
OFF. Note that there is a 5 second delay before the analyzer actually turns OFF, during which time the unit can be kept ON by pressing the
RUN/HOLD button. There is a CO purge feature that keeps the analyzer’s
pump running until the measured CO level drops below 50 ppm.
value is less than 16.0%.
2
2.0 TECHNICAL CHARACTERISTICS
The Analyzer measures & displays:
Primary/Ambient Air Temperature -4 to 600 °F (-20 to 316°C) Oxygen 0.1 to 20.9% O
Stack Temperature -4 to 1202 °F (–20 to 650 °C) Carbon Monoxide 0 to 2000 ppm CO
Differential Pressure/Draft ± 100 inwc
The Analyzer computes & displays:
Combustion Efficiency 0.1 to 100% Carbon Dioxide
0.1 to a fuel dependent
maximum value in % CO referenced to %O2 0 to 9,999 ppm Excess Air 0 to 250%
Fuel Selection
• Natural Gas
• Oil #2
• Oil #4
• Oil #6
• Propane
• Kerosene
• Wood
• Coal
Normal Operating Conditions:
Temperature
• Analyzer
• Probe
32 to 104 °F (0 to 40 °C)
1,000 °F max (538 °C) at
5" insertion
2
Humidity 15 to 90% RH, Non-Condensing
Dimensions Weight
7.9 in (20.1 cm) Height
1lb. (0.45 kg)
3.6 in (9.1 cm) Width
2.3 in (5.8 cm) Depth
4
Instruction 24-9460
Instruction 24-9460
5
Fyrite
®
INSIGHT
Fyrite® INSIGHT
General Characteristics:
Power Requirements Four disposable ‘AA’ Alkaline batteries (op-
tional AC adapter is available)
Operating Time Minimum of 20 hours con tin u ous operation
(pump run ning and backlight ON)
Warm Up Time 60 seconds Memory Up to 100 complete sets of
combustion, temperature, & pressure tests can be saved in memory.
Display 128 x 64, LCD Graphic Dis play Front Panel Controls Eleven front panel push buttons
(Refer to Section 3.4)
Interfaces IrDA printer & USB connectivity Accuracy:
Oxygen ±0.3% O
concentration of CO2
Carbon Monoxide ±5% of reading or ±10 ppm, whichever is
greater*, in the range of 0 to 1000 ppm.
Flue Gas Temperature ±4 °F between 32 & 255 °F
(±2 °C between 0 & 124 °C) ±6 °F between 256 & 480 °F (±3 °C between 125 & 249 °C) ±8 °F between 481 & 752 °F (±4 °C between 250 & 400 °C)
Ambient/Primary Temperature
Draft ±2% of reading or ±0.02 InWC, whichever is
* Tighter CO accuracy in the lower ranges, up to ±2 ppm, may be
attained if a lower range calibration gas (e.g. 100 ppm CO) is used.
±2 °F between 32 & 212 °F (±1 °C between 0 & 100 °C
greater in the range of -1 0 to + 1 0 InW C ± 3% in the range of -1 0 to -40 InWC ± 3% in the range of 10 to 40 InWC
with a typical flue gas
2
3.0 SETTING UP THE ANALYZER
3.1 Preliminary Steps
Before using the analyzer . . .
• Check batteries (Section 3.2)
• Connect probe to analyzer (Section 3.3)
• Check setup (Section 3.5)
3.2 Power
3.2.1 Installing or Replacing Batteries
Install fresh batteries as described below. Check the analyzer for sufficient charge prior to each use. Replace the batteries if the low-battery symbol appears in the upper right corner of the screen. To re place the bat ter ies:
1. Remove battery cover from back of analyzer.
2. If old batteries are installed, remove them and properly discard.
3. Observing the polarity markings inside the battery com part ment, install four ‘AA’ Al ka line bat ter ies.
4. Replace battery cover.
3.2.2 Using the AC Power Adapter
The optional AC power adapter is capable of powering the analyzer on a continuous basis. The adapter plugs into an appropriate 100-240 VAC wall outlet and produces an output of +5 VDC. The adapter's output connector plugs into the analyzer's POWER jack located on the bottom of the unit. The batteries do not need to be removed when using the AC adapter.
3.3 Connecting the Probe Hose & Thermo­couple
A rigid stainless steel probe with handle is connected to a flexible hose with integral water-trap / filter used to draw a gas sample into the ana­lyzer from the room, grilles, diffusers, and furnace flues.
6
Instruction 24-9460
Instruction 24-9460
7
Fyrite
®
INSIGHT
Fyrite® INSIGHT
Inspect the flue-gas hose for cracks. If a hose is defective, replace the en­tire probe assembly.
Before using the analyzer, check that the Water Trap / Filter is dry and not dirty. If necessary, dry out the trap and replace the filter el e ment per Section 6.2.
Perform the following steps to connect the probe assembly to the Fyrite INSIGHT:
1. Push the probe’s sample-gas hose onto the GAS inlet fitting.
2. Push the probe’s draft hose onto the “+” pressure fitting.
3. Push the probe’s thermocouple into the T-STACK jack
Important: DO NOT force the thermocouple connector into
its jack. The con nec tion tabs are dif fer ent sizes, allowing the connector to fit in only one way.
4. Push the optional ambient/primary-air thermocouple into the T-AIR jack.
AC Power Adapter Jack (Power)
Differential Pressure Hose (Optional)
Fyrite® INSIGHT Connectors
Draft Hose
Sample Gas Thermocouple Connector (T-Stack)
Probe Tube
USB Cable
Primary Ambient Air Thermocouple (T-Air) (Optional)
Sample Gas Hose
Probe Stop
Probe Handle
Water Trap/ Filter Assembly
Fig. 3.1
8
Instruction 24-9460
Instruction 24-9460
9
Fyrite
®
INSIGHT
Fyrite® INSIGHT
3.4 Front Panel Push Buttons
Note that a push button may perform several functions, depending on the analyzer’s model number and what screen is being displayed at the time.
Powers the analyzer ON and OFF. Hold this button down for at least 2 seconds to turn the power OFF.
Toggles the backlight ON and OFF while the analyzer is turned ON.
Scrolls up, down, left, and right through the display screen options.
• Up and Down arrows cause the displayed value to in- crease or decrease correspondingly.
Acts as an ENTER button. Performs the action selected.
• While in the HOLD screen, turns the sample pump on, dis- plays the RUN screen, and begins a combustion test.
• While in the RUN screen, turns the sample pump off, displays the HOLD screen and the last set of combustion data.
• Displays the HOLD screen while pressing it from most menus.
• Return the display to the HOLD screen while pressing it during the 5 second power down sequence.
Cancels most operations and displays the previous screen.
Pressing function keys accepts the corresponding function
defined at the bottom of the display, including PRINT , SA VE, MENU, differential pressure ZERO, differential temperature ZERO, P AGE UP, PAGE DOWN and CLEAR data.
3.5 Setup Mode
The analyzer is preset at the factory for the parameters shown below, but can be changed as described in their corresponding sec tions.
Function Parameter To Change
Fuel Natural Gas Section 4.3 Temperature Unit °F Section 3.5.2
Pressure Unit InWC Section 3.5.3 Language English Section 3.5.10 CAL Reminder Never Section 3.5.12 Clock Not Initialized Section 3.5.4
3.5.1 Entering Setup
1. Display the MAIN MENU by pressing the
MENU (F2) button.
2. Use the ▲▼ buttons to highlight SETUP and press ENTER to display the SETUP
MENU.
3.5.2 Temperature Units
Select to display temperature in either ºF or ºC as follows:
1. Enter the Setup Menu per Section 3.5.1. If necessary, press ESC
until MENU appears above F2.
2. From the SETUP MENU, use the ▲▼ but­tons to select TEMPERATURE UNITS.
Press ENTER to display the Temperature
Units Menu.
3. Use the ▲▼ buttons to select the desired temperature unit.
10
Instruction 24-9460
4. Press ENTER to save selection and return to the SETUP MENU or ESC to revert the
previous setting.
Instruction 24-9460
11
®
Fyrite
INSIGHT
3.5.3 Pressure Units
Select to display pressure in inches of water column (inwc), millibar (mb), Pascals (Pa), hectoPascals (hPa), or millimeter of water column (mmwc) as follows:
Fyrite® INSIGHT
3.5.5 O2 Reference
The measured value of CO can be referenced to a specific O2 percentage of 0% to 15%. Set up the reference value as follows:
1. Enter the Setup Menu per Section 3.5.1. If necessary, press ESC until MENU appears above F2.
2. From the SETUP MENU, use the ▲▼ but­tons to select PRESSURE UNITS. Press
ENTER to display the Pressure Units Menu.
3. Use the ▲▼ buttons to selecte the desired pressure unit.
4. Press ENTER to save selection and return to the SETUP Menu or ESC to revert to the
previous setting.
3.5.4 Clock
The clock settings (12 hr. AM/PM or 24 hr.) and date format options (MM/ DD/Y Y or DD/MM/Y Y) are interdependent. (See DATE FORMAT, Section
3.5.15).
1. Enter the Setup Menu per Section 3.5.1. If necessary, press ESC
until MENU appears above F2.
2. From the SETUP MENU, use the▲▼ but-
tons to select CLOCK. Press ENTER to
display the Clock Setup Screen
3. Use the ▲▼ buttons to enter the current date and time. Use the the ◄► buttons to move the cursor across the screen.
4. Press ENTER to save the selection and re­turn to the SETUP MENU or ESC to revert
to the previous setting.
1. Enter the Setup Menu per Section 3.5.1. If
necessary, press ESC until MENU appears above F2.
2. From the SETUP MENU, use the ▲▼ buttons to select O2 REFERENCE. Press
ENTER to display the O2 Reference Screen.
3. Use the ▲▼ buttons to enter the O2 refer­ence value.
4. Press ENTER to save the selection and re­turn to the SETUP MENU or ESC to revert
to the previous setting.
3.5.6 Print Pressure
Select whether to print or not print the pressure measurement on the com­bustion test printout as follows:
1. Enter the Setup Menu per Section 3.5.1. If necessary, press ESC
until MENU appears above F2.
2. From the SETUP MENU, use the ▲▼ but­tons to select PRINT PRESSURE. Press
ENTER to display the O2 Reference Screen.
3. Use the ▲▼ buttons to select YES or NO to print the pressure measurement.
4. Press ENTER to save the selection and re­turn to the SETUP MENU or ESC to revert
to the previous setting.
3.5.7 Zoom
Combustion test data in the Run/Hold screen can be shown with enlarged characters to make viewing easier. The operator can set zoom levels to Standard, 2X or 3X. The Standard zoom setting will display 4 lines of combustion test data at one time; 2X will display 3 lines of data with en­larged characters; 3X will display 2 lines of data with enlarged characters. The operator can scroll through the complete list of measured and calcu-
12
Instruction 24-9460
Instruction 24-9460
13
Fyrite
®
INSIGHT
Fyrite® INSIGHT
lated data no matter what zoom level has been selected. Select the desired zoom level as follows:
1. Enter the Setup Menu per Section 3.5.1. If necessary, press ESC until MENU appears above F2.
2. From the SETUP MENU, use the ▲▼
buttons to select ZOOM. Press ENTER to
display the Zoom Menu.
3. Use the ▲▼ buttons to select the desired zoom level.
4. Press ENTER to save the selection and dis­play the SETUP MENU or ESC to revert to
the previous setting.
3.5.8 Username
The name of the user or owner (e.g. company name, address, phone num­ber) can be stored in memory by manually entering up to three lines of text, with each line containing up to 20 alphanumeric characters. This information will appear at the top of each printout, until it is cleared or new information is entered.
Username Information can be entered as follows:
1. Enter the Setup Menu per Section 3.5.1. If necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to select USERNAME.
Press ENTER to display the most recently
saved information.
6. Repeat Steps 3-5 to edit the remaining lines.
7. Once the input is complete, use the ▲▼ buttons to select EDIT COMPLETE.
8. Press ENTER to accept the entered information or ESC to revert to
the previous information and display the SETUP MENU.
Helpful Hint: Use the Fyrite® user software to upload this informa-
tion from your computer.
To Clear a Username proceed as follows:
1. Enter the SETUP MENU per Section 3.5.1. If necessary, press ESC
until MENU appears above F2.
2. Use the ▲▼ buttons to select USERNAME. Press ENTER to dis-
play the most recently saved information.
3. Clear the information by selecting the CLEAR option (F3).
3.5.9 Run/Hold Format
Fyrite INSIGHT test data is located in the Run/Hold screen. By press-
ing the RUN/HOLD key, you should hear the pump running and see the
word RUN at the upper-left hand corner of the display. The instrument is continuously measuring and calculating the data that is shown in the
Run/Hold screen. Press the RUN/HOLD key again. The pump should stop
running and the word HOLD should be shown at the upper-left hand cor­ner of the display. The instrument will now show the last measured and calculated data taken before the instrument was placed in HOLD. Use the ▲▼ buttons to scroll through the complete list of measured and calculated values when the instrument is running or in the HOLD mode.
3. Use the ▲▼ buttons to select the line of in­formation which you would like to edit and
press ENTER. A single cursor will appear
denoting the character you are changing.
4. Use the ▲▼ buttons to scroll the the list of alphanumeric characters and the ◄► but­tons to move across the screen.
5. Press ENTER to save the selection or ESC to
revert to the previous input.
14
Instruction 24-9460
The order in which data appears in the RUN/HOLD screen is as follows:
O2 = Oxygen CO = Carbon Monoxide Eff = Combustion Effi ciency CO2 = Carbon Dioxide T-STK = Stack Temperature T-AIR = Ambient/Primary Air Temperature EA = Excess Air CO(n) = Carbon Monoxide content referenced to an Oxygen percentage
NOTE: (n) is the current O2 Reference selected.
Instruction 24-9460
15
Fyrite
®
INSIGHT
Fyrite® INSIGHT
Change the order in which data is displayed as follows:
1. Enter the SETUP MENU per Section 3.5.1. If necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to select RUN/HOLD
FORMAT. Press ENTER to display the
Run/Hold Format options.
3. Use the ▲▼ buttons to select EDIT FOR-
MAT. Press ENTER to display the current
format.
4. Change data for a particular location by rst using the ▲▼ buttons to select the loca-
tion. Press ENTER and the cursor will start
to fl ash.
5. Use the ▲▼ buttons to scroll through and select the desired data to appear in that
location. Press ENTER to save the selection.
6. Change the data displayed at other locations by repeating steps 4 and 5
7. When fi nished, use the ▲▼ buttons to select EDIT COMPLETE,
located at the bottom of the list. Press ENTER to save the new dis-
play format and return to the Run/Hold Format options.
3.5.10 Language Selection
Information on the display screen can be shown in English, French, or Spanish.
Select the desired language as follows:
1. Enter the SETUP MENU per Section 3.5.1. If necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to select LANGUAGE
SELECTION. Press ENTER to display the
Language Selection Menu.
3. Use the ▲▼ buttons to select the desired
language. Press ENTER to activate the
selection and return to the SETUP MENU.
3.5.11 Button Sound
The audible sound used to signal when a button is pressed can be turned OFF and ON as follows:
1. Enter the SETUP MENU per Section 3.5.1. If necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to select BUTTON
SOUND. Press ENTER to display the Button
Sound Menu.
Reset Display format back to factory default settings as follows:
1. Enter the SETUP MENU per Section 3.5.1.
If necessary, press ESC until MENU ap-
pears above F2.
2. Use the ▲▼ buttons to select RESET FOR-
MAT. Press ENTER to display the Reset
Format option.
3. Use the ▲▼ buttons to select YES. Press
ENTER to reset the format and return to the
Run/Hold format options.
16
Instruction 24-9460
3. Use the ▲▼ buttons to select YES or NO button sound.
4. Press ENTER to save the selection and re­turn to the SETUP MENU or ESC to revert
to the previous setting.
3.5.12 CAL Reminder Period
The analyzer can be set to indicate a calibration reminder during the 60 second warm-up. Calibration reminders can be preset to occur never, 6, 8, 10, 12, or 15 months after the last calibration. When the preset period is exceeded the instrument will display the reminder, and how long since the sensors were last calibrated. If a calibration reminder is displayed the
operator can press the RUN/HOLD key to move to the Run/Hold Screen for
normal operation. Regular calibration periods of 6 months to 1 year are recommended.
Instruction 24-9460
17
Fyrite
®
INSIGHT
Fyrite® INSIGHT
NOTE: The default CAL Reminder period is set to NEVER.
Set the calibration reminder period as follows:
1. Enter the SETUP MENU per Section 3.5.1. If necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to select CAL RE-
MINDER PERIOD. Press ENTER to dis-
play the Cal Reminder Period Menu.
3. Use the ▲▼ buttons to select the desired time period.
4. Press ENTER to save selection and return to the SETUP MENU or ESC to revert to the
previous setting.
NOTE: The date and time settings must be correct to get ac-
curate cal reminders.
3.5.13 Inactivity Timeout
The analyzer can be placed in an inactivity timeout period of None, 30 or 60 minutes. With this time period set the analyzer will automatically shut off if there have been no button pushes made in the set period of time. The analyzer will not turn off if the CO value is greater than 50 ppm or the O2 value is less than 16%.
NOTE: The default set time is NONE which will not allow the ana-
lyzer to automatically turn off.
3.5.14 Post-Purge Period
The analyzer can be setup to purge the sensors following the combustion test. It can be programmed to purge for None, 5 seconds, 1, 5, or 10 min­utes.
Set the Post-Purge Period as follows:
1. Enter the SETUP MENU per Section 3.5.1.
If necessary, press ESC until MENU ap-
pears above F2.
2. Use the ▲▼ buttons to select POST-PURGE
PERIOD. Press ENTER to display the Post-
Purge Period Menu.
3. Use the ▲▼ buttons to select the desired time period.
4. Press ENTER to save the selection and re­turn to the SETUP MENU or ESC to revert
to the previous setting.
3.5.15 Date Format
The date format options and the clock settings are interdependent.
MM/DD/YY format - Used with a 12 hour time clock. This set up menu
includes an AM/PM option.
DD/MM/YY format - Used with a 24 hour time clock.
Set the Inactivity Timeout Period as follows:
1. Enter the SETUP MENU per Section 3.5.1. If necessary, press ESC
until MENU appears above F2.
2. Use the ▲▼ buttons to select INACTIVITY
TIMEOUT. Press ENTER to display the
Inactivity Timeout Menu.
3. Use the ▲▼ buttons to select the desired timeout period.
4. Press ENTER to save the selection and re­turn to the SETUP MENU or ESC to revert
to the previous setting.
18
Instruction 24-9460
Change the date format as follows:
1. Enter the SETUP MENU per Section 3.5.1.
If necessary, press ESC until MENU ap-
pears above F2.
2. Use the ▲▼ buttons to select DATE FOR-
MAT. Press ENTER to display the Date
Format Menu.
3. Use the ▲▼ buttons to select the desired date format.
4. Press ENTER to save the selection and re­turn to the SETUP MENU or ESC to revert
to the previous setting.
Instruction 24-9460
19
Fyrite
®
INSIGHT
Fyrite® INSIGHT
4.0 OPERATION
4.1 Operating Tips
When an analyzer is brought in from a cold vehicle, let it warm up
• slowly to minimize condensation. Temperatures below freezing will not damage the analyzer; however, bringing a cold analyzer into a warm, humid environment may cause condensate to form inside the case.
Caution: Although the analyzer itself is not damaged by an ex-
tremely cold environment, the electrochemical sensors may be damaged. The O
-20 °F and the CO sensor's at approximately -90 °F. If the analyzer is exposed to an extremely cold condition, it is strongly suggested that the sensor housings be examined for hairline cracks. Be aware that a leaking sensor can cause chemical burns to the skin and possibly damage the PCB assemblies.
Ensure that the analyzer is sampling fresh air when turned ON. Pulling
• a stack-gas sample through the analyzer during its warm-up period will not damage the analyzer, but it will result in incorrect sensor readings, and may result in sensor error messages appearing after the warm-up cycle completes.
Note that f lue -gas condensat e is acid ic and very cor r osive. It i s imp or tant
• not to allow the analyzer's internal components to come in contact with condensate for long periods of time.
sensor's electrolyte will freeze at approximately
2
4.2 Analyzer Power On & Warm Up
1. Connect the probe and make sure that the analyzer is properly set up per Sec tion 3.
IMPORTANT! DO NOT insert probe into stack before powering ON
the analyzer.
2. Place the probe in an area that contains fresh air. This ensures that the sensors will be properly zeroed during the warm-up cycle.
3. Power ON the analyzer by pressing the
PWR button for at least 1 second, or until
a single beep is heard. Observe that the analyzer's firmware version, model and se­rial numbers are briefly displayed followed by the Warm Up screen.
4. Wait for the analyzer to count down its 60 second warm-up period; after which, the in­strument will display the Combustion Test HOLD screen.
If, however, any errors were detected during warm-up, the message "ERRORS DETECT­ED" is displayed along with a list of those errors. In this example, the O2 sensor is missing and the T-Stack thermocouple is not connected. Refer to Section 6.8 for a listing and possible remedy for the errors displayed.
V1.00
24-8250
Before each use, inspect the filter element of the water-trap / filter as-
• sembly. Replace the filter if it looks dirty.
When sampling flue-gas, keep the analyzer above the water-trap, and keep
• the trap in a vertical position. This will maximize the effectiveness of the trap and keep liquid from being drawn directly into the analyzer.
When liquid condensate is seen inside the water trap, empty the trap
• before it becomes full.
When storing the analyzer, it's a good idea to empty the water trap and
• leave it open to further dry it out.
Calibrate the analyzer every 6 months - 1 year to assure its accuracy.
20
Instruction 24-9460
4.3 Fuel Selection
The top line of Combustion Test HOLD screen shows the fuel currently selected. If necessary, change the fuel as follows:
1. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2
2. Use the ▲▼ buttons to select FUEL. Press
ENTER to display the Fuel Selection screen.
3. Use the ▲▼ buttons to select the desired fuel.
Instruction 24-9460
21
Fyrite
®
INSIGHT
Fyrite® INSIGHT
4. Press ENTER to save the selection and re-
turn to the Combustion Test HOLD screen
or ESC to revert to the previous setting and
return to the MAIN MENU.
4.4 Sampling Point
Forced Air Furnace – When testing atmospheric burner or gravity vented, forced air heating equip ment with a clamshell or sectional heat exchanger design, test each of the exhaust ports at the top of the heat ex­changer. The probe should be inserted back into each of the exhaust ports to obtain a flue gas sample, before any dilution air is mixed in.
Hot Water Tank – Domestic hot water tanks with the ‘bell’ shaped draft diverter on top can be accurately tested by inserting the probe tip directly into the top of the fire tube be low the diverter.
80% Efficiency Fan Assist or Power Vented – Combustion testing of fan assist or power vented, furnaces/boilers should be done through a hole drilled in the vent approximately 12 inches above the inducer fan.
90% Efficiency Condensing – Condensing furnaces/boilers can be tested through a hole drilled in the plastic vent pipe (when allowed by the manufac­turer or local authority of jurisdiction) or taken from the exhaust termination. After testing, the hole should be sealed with high temperature silicon.
Atmospheric or Gravity Vented Boiler – Boilers, which have a ‘bell’ shaped draft diverter directly on top, should be tested directly below the diverter through a hole drilled in the vent connector.
NOTE: All drill holes should be sealed upon completion of combus-
tion testing.
1. Press the RUN/HOLD button to start the test. You should hear the
pump start running and see the word RUN appear at the top of the Combustion Test screen.
Sensor Indicators: The following indicators appear in the sensor's data
field depending on certain conditions:
∗ ∗ ∗) Sensor that is not calibrated or installed.
(
(XXX) Sensor overrange (
- - -
) The calculated data cannot be displayed because the measured data necessary to make the calculation is out of range (i.e., oxygen level above 16%).
2. Use the up and down arrow keys to scroll to the T-STK reading. Loosen the thumbscrew on probe stop and move probe in and out of the stack until the stack’s core temperature (hot spot) is located as indicated by the highest T-STK reading; then tighten thumbscrew to prevent further probe movement. Locating the highest stack tempera-
ture is very important for accurate efficiency calculations.
3. You can now begin burner-service procedures. The analyzer readings will change quickly to show changes in burner performance.
CAUTION: Position the Water Trap with its gas-flow
arrow pointing upward. Do not let water condensate go above the tip of the riser tube. The sensors could be damaged if water would enter the analyzer. Empty the Water Trap after every combustion test (refer to Section 4.10)
4. Pressing the RUN/HOLD button freezes all readings, stops the pump and
displays the Combustion-Test HOLD screen. Use the ▲▼ buttons to view
all test values at the moment the RUN/HOLD button was pressed. Press­ing RUN/HOLD again restarts the pump and resumes testing.
4.5 Performing a Com bus tion Test
Ensure that the following has been completed, and then proceed with the combustion test as described below:
• Power ON analyzer and allow it to warm up (Section 4.2).
• Select fuel being burned (Section 4.3).
• Inset probe into proper sample location (Section 4.4).
• If necessary, insert optional primary air thermocouple into combustion­air stream of burners that use an outside source of combustion air.
22
Instruction 24-9460
Display Name Description of Measurement or Calculation O
2
CO ppm Carbon Monoxide Eff % Combustion Effi ciency CO
2
T-STK Stack Temperature T-AIR Primary/Ambient Air Temperature EA % Excess Air CO(O) Carbon Monoxide ppm level referenced to a % of oxygen
Instruction 24-9460
% Oxygen
% Carbon Dioxide
23
Fyrite
Undiluted Flue GasSample Taken Under Draft Diverter inTop
ofFire Tube
®
INSIGHT
Fyrite® INSIGHT
4.6 Making a Draft / Pressure Measurement
The difference in pressure (ΔP) between two areas can be measured by using the analyzer's two pressure ports and the PRESSURE screen. By using the -ΔP port as the reference, the pressure applied to the +ΔP port will be displayed on the PRESSURE screen as the differential pressure between the two ports. Perform a draft / pressure measurement as follows:
1. Turn ON the analyzer and allow it to complete its warm-up cycle.
2. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears
above F2.
Atmospheric or Gravity Vented Boiler
Hot Water Tank
80% Eff. Fan Assist or Power Vented Furnace/Boiler
Atmospheric Burner or Gravity Vented Forced Air
3. Use the ▲▼ buttons to select PRESSURE.
Press ENTER to display the Pressure
screen.
4. Before taking a measurement, the pressure sensor may need to be re-zeroed if it is not already displaying zero with both pressure ports open to the atmosphere. If necessary, zero the pressure sensor as follows:
Press the ZERO (F2) button.
Disconnect any hoses connected to the
+ΔP and -ΔP ports, and then press EN- TER to zero the pressure sensor.
Reconnect any hoses. When measuring
draft, leave the -ΔP port open to the at- mosphere and connect the probe's draft hose to the +ΔP port.
5. Do one of the following to measure draft or differential pressure:
To measure draft, insert the probe into
the stack and observe the draft reading on the PRESSURE screen.
F1 F2 F3
24
Fig. 4.1
90% Eff. Condensing
Furnace/Boiler
Instruction 24-9460
To measure differential pressure, con-
nect two sampling hoses to the +ΔP and -ΔP ports, and place the open end of each hose into the areas being mea-
Instruction 24-9460
25
Fyrite
®
INSIGHT
Fyrite® INSIGHT
sured. The differential pressure between the two areas is now displayed on the PRESSURE screen. If the pressure at the +ΔP port is higher than the -ΔP port, the pressure reading will be positive. If it is lower, the reading will be negative.
4.7 Temperature Measurement
The difference in temperature between two areas can be measured by us­ing the analyzer's two temperature channels and the Temperature Screen. By using the T-Air channel as a reference, the temperature applied to the T-Stack channel will be displayed on the Temperature Measurement screen as differential temperature between the two channels.
Perform a differential temperature measurement as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to select TEMPERA-
TURE. Press ENTER to display the Tem-
perature Measurement screen.
3. Install thermocouples in both temperature channel connectors.
2. Press the SAVE (F3) button to save the test data in the next available memory location.
NOTE: When memory is full, the next read-
ing will not be saved until space is made available by clearing previ­ously saved data. (Section 4.13)
4.9 Ending a Combustion Test
WARNING! Burn Hazard. Do not touch the probe af-
ter removing it from the stack. Allow the probe to cool before handling (about 5 minutes).
1. Remove probe from the flue-gas stream.
2. Allow the pump to run until all combustion gases are flushed from the analyzer as indicated by the O2 reading returning to 20.9%.
4.10 Emptying the Water Trap
The Water Trap / Filter Assembly removes stack-gas condensate, and also prevents soot from contaminating the internal components of the analyzer.
4. Before taking a measurement, the tempera­ture channels may need to be zeroed, if not already displaying zero with both ther­mocouples exposed to the same condition.
Press ZERO (F2) if necessary.
.1
4.8 Saving Test Data
Up to 100 sets ("snap shots") of combustion test, pressure, or temperature data can be saved in memory, which can later be recalled for viewing from the Memory Directory.
1. First display the screen that contains the data to be saved. In the example to the right, all data associated with the Combus­tion Test RUN screen will be saved.
26
Instruction 24-9460
IMPORTANT! Use the Water Trap / Filter Assembly in a vertical posi-
tion with the gas-flow arrow pointing up.
Empty the water trap chamber after each combustion test, or stop the test and empty the chamber if the liquid condensate level approaches the tip of the riser tube.
To empty the trap, first pull apart the two halves of the Water Trap using a slight twisting motion; empty the water trap chamber, and then reas­semble the trap.
After each combustion test, also check the Water Trap's filter element. If it looks dirty, replace the filter per Section 6.2.
Instruction 24-9460
27
Fyrite
®
INSIGHT
Fyrite® INSIGHT
4.11 Powering OFF the Analyzer & Purging
Power OFF the analyzer by pressing the PWR button
for at least 2 seconds, or until two beeps are heard. The unit will count down 5 seconds before shutting down, giving the operator an opportunity to keep the analyzer
powered ON by pressing the RUN/HOLD button.
If the analyzer was not purged with fresh air then the analyzer may remain ON with its pump running and display the message "PURGING SENSORS" as the result of combustion gases still being present inside the analyzer. At this time the operator should ensure that the probe is removed from the stack, allowing the analyzer to purge itself with fresh air. The 5-second-shutdown sequence will not begin until the gas levels inside the analyzer drop below predeterminded levels.
TIP: Although not recommended, the purging process can be bypassed
by pressing the PWR button a second time.
4.12 Low Battery Alarm
When the batteries are nearly depleted, an empty battery icon appears in the upper-right corner of the display, and a short beep is sounded every 10 seconds.
After a low battery alarm occurs, the analyzer will continue to operate for only a few minutes. The amount of operating time that remains depends on many factors (e.g. pump and backlight being ON or OFF and the type and condition of the batteries).
3. Use the ▲▼ buttons to select Memory Direc-
tory. Press ENTER to display the saved data
contained in the Memory Directory.
TIP: When displaying the contents of the
memory directory, the operator can quickly page through the screens by pressing the
PAGE– (F1) and PAGE+ (F3) buttons. Or
move to the first or last memory location by pressing the
To Clear Test Data:
1. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
2. Use the ▲▼ buttons to select MEMORY.
Press ENTER to display the Memory Screen.
3. Use the ▲▼ buttons to select CLEAR
MEMORY. Press ENTER to display the
Clear Memory Options.
4. Use the ▲▼ buttons to select YES and press
ENTER. The screen will briefly display a
confirmation that the memory has been cleared.
buttons, respectively.
4. 1 3 Memory
All saved data can be retrieved for viewing through the MEMORY option.
To open and view saved test data:
1. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to select MEMORY.
Press ENTER to display the MEMORY
MENU.
28
Instruction 24-9460
Instruction 24-9460
29
Fyrite® INSIGHT
Fyrite
®
INSIGHT
4.14 Printing Test Data
Combustion, pressure, or temperature data that is currently being dis­played can be sent to a printer using IrDA protocol as described below.
Data that is stored in memory can also be printed by first displaying the stored test data as described in Section 4.13.
In addition to printing combustion, pressure, and temperature data, the
contents of any screen that shows the label “PRINT” above the F1 button
can be printed. For example, the information shown in the DIAGNOSTIC screen can be printed.
1. Power up the printer. Refer to the printer’s instruction manual for detailed operating infor­mation. If not previously done, select the follow­ing for the printer for:
• 8 bit
• No parity
• 9600 baud
• IrDA is set to IrDA-SIR
• DTR handshaking
2. Align the printer with the top of the analyzer.
3. Press the PRINT (F1)
button to begin print­ing.
F2
F1
F3
RUN
ESC
HOLD
PWR
BACHARACH, INC.
Fyrite Insight
SN: ENGR-CM03
Time: 01:28:27 PM
Date: 10/22/08
Fuel
Nat Gas
O2 4.0 % CO 12 ppm Eff CO2 T-STK 374 oF T-AIR 68.0 oF EA CO (3) 13 ppm
Comments:
82.6
9.5
21.3
%
%
%
4.15 Fyrite® User Software Installation
Windows XP Operating System
After the CD is loaded, the Setup window will be displayed. Click on OK. (If applicable, wait for the Net 3.5 Framework set up to complete.)
Click "Next" on the “Welcome” screen to begin the installation.
Respond to the prompts on the screens that follow.
Click on: “Finish”...
“Next”...
“Close”...
“OK”...
30
Fig. 4.2
Instruction 24-9460
After the installation,
• Open Fyrite® User Software via the desktop icon.
• Connect the instrument to the PC with a USB cable and power up the unit.
Instruction 24-9460
31
Fyrite
®
INSIGHT
Fyrite® INSIGHT
When first connected, a “Found New Hardware Wizard” screen will be displayed. Again, respond to the prompts to completion.
Click on: "Next"...
"Continue anyway"...
"Finish"
Windows Vista Operating System
Click on: “Finish”...
“Next”...
“Next”...
“Next”
After loading the CD, the Security window will be displayed. Click on OK. (If applicable, wait for the Net 3.5 Framework set up to complete.)
Click on: "Continue"...
"Next"...
"Install"...
"Install... software anyway"
32
Instruction 24-9460
The installation will progress... Click on: "Continue..."
Instruction 24-9460
"Close"...
"OK"
33
Fyrite
®
INSIGHT
Fyrite® INSIGHT
After the installation,
• Open Fyrite
®
User Software via the desktop icon.
• Connect the instrument to the PC with a USB cable.
• Power up the unit.
• When first connected, a “Found New Hardware Wizard” screen will be displayed.
Click on: "Next..."
"Next..."
"Continue anyway"
"Finish"
5.0 CALIBRATION & MAINTENANCE
IMPORTANT: Before performing any calibration procedure,
ensure that fresh batteries are installed or use the optional AC power adapter. Also, ensure that the analyzer is at room temperature and will be sampling fresh air when powered ON.
IMPORTANT: To main tain ac cu ra cy as listed in the Techni-
cal Characteristics Sec tion of this manual, the standards used must be at least four times as accurate as the stated
®
accuracy of the Fyrite
INSIGHT.
5.1 B-Smart Sensors
The Fyrite® INSIGHT utilizes Bacharach’s new B-Smart Sensor tech­nology for its CO sensor. The B-Smart sensor is marked with a 10-digit calibration code that can be entered in the instrument's memory via the keypad, or with Fyrite user software.
Benefits of B-Smart Sensors:
New sensors can be installed without applying gas for calibration.
• Sensors can be pre-calibrated and installed when needed.
• Sensors can be moved from one analyzer to another.
4.16 Fyrite® User Software Description
Fyrite® INSIGHT User Software
Set instrument time Sync instrument with computer's time Calibration password Change 4 digit numeric code Time meters Provides instrument & pump run time B-Smart B-Smart calibration code entry Saved measurements Downloads instrument memory to Excel User/Customer ID User ID entry Settings Change instrument set up parameter Customer logo set up Load custom logos for print out. Limited to 384 x 192. Custom fuel Bacharach can supply fuel codes for custom fuel. Consult
Update instrument software Remote upgrade of instrument software.
34
factory.
Instruction 24-9460
The B-Smart sensors should be calibrated by an authorized Bacharach Service Center every 6 months to 1 year to assure that the analyzer contin­ues to meet its published accuracy specifications. The B-Smart sensors, however, can be calibrated in the field if your facility has the necessary equipment and qualified personnel to perform the procedures described in the following sections of this instruction manual.
5.2 Starting a Calibration
Start any calibration by performing the following:
1. With the analyzer power OFF, place the unit in fresh, ambient air;
then power on the analyzer.
Instruction 24-9460
35
Fyrite
®
INSIGHT
Fyrite® INSIGHT
2. Allow the analyzer to cycle through its 60 second warm-up period. During warm-up, the analyzer's operation is checked and the sen­sors are set to the following ambient conditions:
Oxygen sensor spanned to 20.9%
CO sensor zeroed
Pressure sensor zeroed
Any errors detected during warm-up will be listed on the dis­play immediately following warm-up. Correct any errors before proceeding. Refer to Section 6.8 for a list of error messages and their meaning.
3. Display the MAIN MENU by pressing the MENU (F2) button. If necessary, press ESC until MENU appears above F2.
4. Use the ▲▼ buttons to select CALIBRA-
TION. Press ENTER to display the Calibra-
tion Password screen.
5. Before calibration can begin a 4-place nu­meric password must be entered. Use the ▲▼ and ◄► buttons to enter the password. (Note that the default password is 1111)
6. Press the ENTER button to accept the
password. If the correct password was entered the CALIBRATION Menu will be displayed.
5.3 B-Smart Sensor Replacement & Calibration
To replace a B-Smart sensor and calibrate do the fol­lowing:
1. Enter the CALIBRATION MENU per Sec­tion 5.2.
2. Use the ▲▼ buttons to select B-SMART.
Press ENTER to display the B-Smart code
screen.
3. Use the ▲▼ buttons to enter the 10 digit al­phanumeric code supplied with the B-Smart sensor. Use the the ◄► buttons to move the
cursor across the screen. Press ENTER.
NOTE: If the correct code was entered, the analyzer accepts it and
returns to the CALIBRATION MENU. If an incorrect code was entered, the screen will display "Invalid Code". Check to make sure the correct code has been entered. If problem persists, contact your nearest Bacharach Service Provider.
Bacharach also offers a convenient Exchange Program that allows the cus­tomer to regularly send in old sensors and, as scheduled, receive new re­placements that are already calibrated and that include a code that can be entered into the analyzer for a quick convenient setup. Contact Bacharach customer service for more details about this program.
7. Use the ▲▼ buttons to select the desired sensor channel to be calibrated, and then perform the calibration procedure for that sensor as described in the following sections.
36
Instruction 24-9460
5.4 Pressure Sensor Calibration
This procedure calibrates the pressure sensor to a known pressure value.
Materials Required:
Bellows
Manometer
- Range: ± 8 in. of water column (± 20 mb)
- Accuracy: ± 0.01 in. of water column (± 0.025 mb)
Instruction 24-9460
37
Fyrite
®
INSIGHT
Fyrite® INSIGHT
Procedure:
NOTE: The unit-of-measure for pressure is selected per Section 3.5.3. In
the following procedure inwc is selected, but note that any unit­of-measure can be used for calibration purposes.
1
4
5
2
4 3 2 1 0
-1
-2
-3
-4
3
Parts:
1. 3/16" I.D. Tubing
2. Tee
3. Barbed Hose Connector
4. Bellows
5. Manometer
Figure 5-1 Pressure Sensor Calibration Equipment
1. Assemble the pressure sensor calibration equipment as shown in Figure 5.1, but DO NOT connect the analyzer to the calibration equipment at this time.
2. If not already done, power ON the analyzer and display the CALI­BRATION LIST per Section 5.2.
5. Connect the hose from the manometer to the +ΔP port and apply a negative pressure to this port by adjusting the bellows for a manom­eter reading of -4.00.
6. Use the ▲▼ buttons to enter an Applied value that exactly equals the manometer reading.
The calibration range is from -6 to -2 inwc (-15 to -5 mb). An attempt to calibrate outside this range will cause the message "Applied Value High" (or Low) to appear at the bottom of the screen.
7. Wait until the Measured reading stabilizes,
and then press ENTER to calibrate the pres-
sure sensor's Measured value to that of the Applied value. The message, "Good Calibra­tion" should briefly appear, followed by the CALIBRATION LIST screen.
8. Remove calibration equipment.
5.5 T-Stack Calibration
This procedure first zeroes and then spans the stack-temperature channel to known temperature values.
The use of an electronic thermocouple simulator is the preferred method of producing the desired calibration temperatures. Alternatively, ice and boiling water baths can be used.
3. Use the ▲▼ buttons to select PRESSURE
and then press ENTER to display the CALI-
BRATE PRESSURE screen.
"Measured" is the pressure value currently being detected by the pressure sensor, while "Applied is a known value of pressure that will be applied for calibration purposes.
4. With both the -ΔP and +ΔP ports open to the atmosphere, observe that the current Measured pressure reading should be 0.00 ± 0.01 inwc. If necessary, zero the pressure sensor per Section 4.6 then repeat steps 2 through 4.
38
Instruction 24-9460
-4.02
Materials Required:
Thermocouple simulator (K-type)
- Range: 0 to 600 °F
- Accuracy: ± 0.5 °F (Alternatively) ice-water, boiling water, thermometer
TS-Zero Procedure:
1. Set the thermocouple simulator to room temperature and plug its
output into the T-STACK connector located at the bottom of the analyzer.
Alternatively: Plug the probe's thermocouple into the T-STACK con-
nector located at the bottom of the analyzer. DO NOT attach the
probe's gas hose to the analyzer's GAS port; otherwise water will be drawn into the analyzer!
Instruction 24-9460
39
Fyrite
®
INSIGHT
Fyrite® INSIGHT
2. If not already done, turn ON the analyzer and display the CALI­BRATION LIST screen per Section 5.2.
3. Use the ▲▼ buttons to highlight T-Stack,
and then press ENTER to display the CALI-
BRATE TS-ZERO screen.
"Measured" is the current temperature read­ing, while "Applied" is a known temperature that will be applied for calibration purposes.
4. Set thermocouple simulator to 32 °F (0 °C), and then use the
and buttons to enter an Applied value that exactly equals the setting of the simulator.
Alternatively: Submerge probe tip into an ice-water bath with a ther-
mometer, wait several minutes, and then use the and buttons to enter an Applied value that exactly equals the thermometer read­ing.
The calibration range is from 32 to 41 °F (0 to 5 °C). An attempt to calibrate outside this range will cause the message “Applied Value High" (or Low) to appear at the bottom of the screen.
5. Wait until the Measured reading stabilizes,
and then press ENTER to calibrate the
TS-Zero Measured value to that of the Ap­plied value; after which the message “Good Calibration” should briefly appear followed by the CALIBRATE TS-SPAN screen.
TS-Span Procedure:
6. Set thermocouple simulator to 572 °F (300 °C), and then use the and buttons to enter an Applied value that exactly equals the
setting of the simulator.
The calibration range is from 175 to 625 °F (80 to 330 °C). An at­tempt to calibrate outside this range will cause the message “Applied Value High” (or Low) to appear at the bottom of the screen.
7. Wait until the Measured reading stabilizes, and then press ENTER to
calibrate the TS-Span Measured value to that of the Applied value; after which the message “Good Calibration” should briefly appear fol­lowed by the CALIBRATION LIST screen being re-displayed.
5.6 T-Air Calibration
This procedure first zeros and then spans the ambient-temperature chan­nel to known temperature values.
The use of an electronic thermocouple simulator is the preferred method of producing the desired calibration temperatures. Alternatively, ice and boil­ing water baths can be used.
Material Required:
Thermocouple simulator (K-type)
- Range: 0 to 600 °F
- Accuracy: ±0.5 °F (Alternatively) Ice-water, boiling water, thermometer
TA-Zero Procedure:
1. Set thermocouple simulator to room temperature and plug its output
into the T-AIR connector located at the bottom of the analyzer.
Alternatively: Plug the probe’s thermocouple
into the T-AIR connector located at the bot­tom of the analyzer. DO NOT attach the
probe’s gas hose to the analyzer’s GAS port; otherwise water will be drawn into the analyzer!
40
Alternatively: Submerge probe tip into a container of boiling water
with a thermometer, wait several minutes, and then use the and
buttons to enter an Applied value that exactly equals the ther-
mometer reading.
Instruction 24-9460
2. If not already done, turn ON the analyzer
and display the CALIBRATION LIST screen per Section 5.2.
3. Use the buttons to highlight T-Air, and
then press ENTER to display the CALI-
BRATE TA-ZERO screen.
Instruction 24-9460
41
Fyrite
®
INSIGHT
Fyrite® INSIGHT
“Measured” is the current temperature reading, while “Applied” is a known temperature that will be applied for calibration purposes.
4. Set thermocouple simulator to 32 °F (0 °C), and then use the and buttons to enter an Applied value that exactly equals the
setting of the simulator.
Alternatively: Submerge probe tip into an ice-water bath with a ther-
mometer, wait several minutes, and then use the
and but­tons to enter an Applied value that exactly equals the thermometer reading.
The calibration range is from 32 to 41 °F (0 to 5 °C). An attempt to calibrate outside this range will cause the message “Applied Value High” (or Low) to appear at the bottom of the screen.
5. Wait until the Measured reading stabilizes, and then press ENTER
to calibrate the TA-Zero Measured value to that of the Applied value; after which the message “Good Calibration” should briefly appear followed by the CALIBRATE TA-SPAN screen.
TA-Span Procedure:
6. Set thermocouple simulator to 212 °F (100 °C), and then use the and buttons to enter an Applied value that exactly equals the set­ting of the simulator.
Alternatively: Submerge probe tip into a container of boiling water
with a thermometer, wait several minutes, and then use the and
buttons to enter an Applied value that exactly equals the ther-
mometer reading.
The calibration range is from 194 to 230 °F (90 to 110 °C). An at­tempt to calibrate outside this range will cause the message “Bad Calibration Wrong CAL Entry” to appear in the following step.
7. Wait until the Measured reading stabilizes, and then press ENTER
to calibrate the TA-Span Measured value to that of the Applied value; after which the message “Good Calibration” should briefly appear followed by the CALIBRATION LIST screen being re-dis­played.
5.7 CO Sensor Calibration
Material required:
Calibration kit, P/N 24-7059
Gas cylinder: 500 ppm CO in air, P/N 24-0492
To improve the accuracy of the CO reading, we suggest that if the analyzer will be primarily used for flue gas testing, then calibrate using 500 ppm CO. If the analyzer, however, will be pri ma ri ly used for ambient testing, then calibrate using 100 ppm CO.
Procedure:
1. If not already done, turn ON the analyzer and display the CALIBRATION LIST screen per Section 5.2.
2. Use the
then press ENTER to display the CALI-
BRATE CO screen.
3. Attach a 500 ppm CO cylinder to the regula­tor of the calibration fixture.
4. Use the and buttons to enter an Applied value that exactly equals the concentration stamped on the CO cylinder.
5. Wait until the Measured reading stabilizes and then press ENTER
to calibrate the CO Measured value to that of the Applied value. The message “Good Calibration” should briefly appear.
buttons to highlight CO, and
“Measured” is the current CO reading, while “Applied” is a known CO level
4
that will be applied for calibration purposes.
The calibration range is from 20 to 1,000 ppm. An attempt to calibrate outside this range will cause the message “Applied Value High” (or Low) to appear at the bottom of the screen.
If the sensor’s output is low, but still usable, then the message “Good Calibration WARNING Low Sensor” will appear. The sensor will now be marked as being Low in the DIAGNOS- TICS screen.
42
Instruction 24-9460
Instruction 24-9460
43
Fyrite
®
INSIGHT
Fyrite® INSIGHT
If the sensor’s output is too low to be usable, then the message “Bad Calibration Sensor End of Life, Entry Not Saved” will appear. The sensor will now be marked as being BAD in the DIAGNOSTICS screen.
6. Power OFF the regulator of calibration fixture and remove the CO cylinder.
6.0 Maintenance
6.1 Fyrite® INSIGHT Disassembly
The following section describes how to disassemble the Fyrite INSIGHT to perform necessary periodic maintenance. (See Figure 6.1 for diagram.)
Tools Required:
• Medium Phillips screwdriver
Procedure:
1. Unplug all thermocouples from bottom of analyzer.
2. Remove battery cover and then remove batteries.
3. Remove sensor caps, disconnect tubing, and then unplug all sensors.
4. Lay analyzer face down on a padded work surface; and then using
a medium Phillips screwdriver, remove the unit’s four rear-case screws.
Fyrite® Insight Components
2
1
To O
2
To gas
Fig. 6.1
1. O2 Sensor 4. Sample Pump
2. CO Sensor 5. LCD Screen
3. CO Sensor Base
5
J9 (Pump)
J11 (Battery)
4
3
To gas tting
To O2 sensor cap
Side ViewFront View
5. Lift rear case from analyzer and set aside.
6. Unplug electrical connector J11 from printed circuit board.
7. Lift printed circuit board from analyzer.
44
Instruction 24-9460
Fyrite® Insight End Plate Connectors
6. AC Power Adapter Jack (Power)
7. Sample Gas Thermocouple Connector (T-Stack)
8. Primary Air Thermocouple (T-Air)
9. USB Connector
10. Differential Pressure Hose (Optional)
11. Draft Hose
12. Sample Gas Hose
Instruction 24-9460
6
0
Front
7
8
Back
Fig. 6.2
9
1211
45
Fyrite
®
INSIGHT
Fyrite® INSIGHT
6.2 Water Trap / Filter
6.2.1 Emptying the Water Trap Chamber
The water trap chamber should be emptied after every test, or when the water condensate approaches the tip of the riser tube.
1. Remove water trap chamber per Figure 6.3.
2. Pour out liquid condensate, and then reassemble trap.
6.2.2 Replacing the Filter Element
Replace the filter element when it becomes visibly dirty or becomes satu­rated with water.
Material Required: • Filter element, P/N 07-1644
• Small flat blade screwdriver
1. Remove water trap chamber per Figure 6.3.
2. Pry apart filter chamber using a small flat-blade screw
driver. Remove and discard old filter.
3. Install new filter and reassemble filter chamber, making
sure that surfaces “A” and “B” contact each other.
6.3 O2 Sensor Replacement
Material Required:
• O2 Sensor (P/N 24-0788)
Pry tab upward to remove the battery door from the unit.
Fyrite® Rear Panel
Fig. 6.4
O
2
Sensor
CO
Sensor
4. Reassemble trap.
Pull apart using a twisting motion.
DO NOT rock back and forth!
46
FILTER CHAMBER
WATER TRAP CHAMBER
Pry apart filter chamber using a small flat-blade screwdriver inserted into notch
FILTER ELEMENT
Fig. 6.3
When reassem­bling filter chamber, make sure surfaces “A” and “B” contact each other
“A” “B”
Instruction 24-9460
Procedure:
1. After battery door is removed, remove connector tubing from both sensors.
2. Pull O
sensor from its socket and
2
remove the O2 cap.
3. Properly dispose of the old sensor. En­gage the nub on the new sensor within the slot on the cap's side and twist to secure the cap and sensor together.
4. Install the cap and sensor unit by
• Aligning the ribs on the sides of the
sensor with the corresponding shape in the base.
• Inserting the pins into the connec-
tors in the base.
5. Reattach tubing.
Instruction 24-9460
Slot
Nub
O2 Cap
Sensor
Fig. 6.5
47
Fyrite
®
INSIGHT
Fyrite® INSIGHT
6.4 CO Sensor Replacement
Material Required:
• CO Sensor (P/N 24-7265) or B-Smart sensor (P/N 24-1467)
CO
Sensor
Pry tab upward to remove the battery door from the unit.
Fyrite® Rear Panel
O2
Sensor
Fig. 6.6
6.5 Thermocouple Replacement
Using the appropriate thermocouple replacement kit listed below, replace the probe’s thermocouple as follows:
Each kit contains a thermocouple assembly, two O-rings, and two wire­splice connectors.
Tools Required:
• Small fl at blade screwdriver
• Wire cutter
• Wire stripper
• Slip joint pliers
Procedure:
1. Gain access to the thermocouple connections by first removing three screws from probe handle, and then separating the two handle pieces.
2. Cut wires attached to old crimp connectors, leaving behind as much of the probe’s thermocouple-connector wire as possible
Probe Handle Screws
Procedure:
1. After battery door is removed, remove the connector tubing from the CO sensor.
2. Remove CO cap by twisting counter clockwise. Gently pull CO sensor out of its socket.
3. Properly dispose of the old CO sen­sor.
4. Plug the new CO sensor into its socket.
5. Install the CO cap by aligning the arrow on the cap with the "OPEN" marker, then twisting to the "CLOSED" marker. Reattach tub­ing.
48
Sensor
Printed Circuit Board
OPEN
Fig. 6.7
Instruction 24-9460
CO Cap
CLOSED
Fig. 6.8
3. Pull old thermocouple from probe body and discard.
4. The new thermocouple has been coiled for shipping purposes. Straighten the thermocouple using your thumb and index finger.
5. If not already done, install supplied O-Rings onto thermocouple.
6. Insert thermocouple into probe body until it “bottoms out.”
Instruction 24-9460
49
Fyrite
®
INSIGHT
Fyrite® INSIGHT
7. Strip 1/4 inch of insulation from each of the probe’s thermocouple connector wires.
IMPORTANT: In Step 8, the thermocouple wires must first be twisted
together and then crimped.
8. Twist both red thermocouple wires together; insert them into the supplied wire-splice connector; and then crimp the connector using a pair of pliers. Repeat this step for the yellow thermocouple wires.
9. Reassemble the probe handle, being careful not to pinch the ther­mocouple wires between the handle pieces. In addition, ensure that the end of the thermocouple is in front of the rib molded into the bottom handle piece; otherwise, the handle pieces will not fit tightly together.
6.6 Gas Pump Replacement
Parts & Tools Required:
Replacement pump (P/N 24-3049)
Medium phillips screwdriver
Procedure:
1. Gain access to the gas pump by removing the rear case. Refer to Sec­tion 6.1.
NOTE: An analyzer that is used to test natural gas furnaces
normally requires less frequent cleaning than an ana­lyzer used for testing coal or oil fired furnaces.
Equipment Required:
Alcohol
Aerosol can of automotive carburetor cleaner
Clean rag
Source of compressed air (optional)
Procedure:
1. Remove gas-sample hose from top of water trap.
CAUTION: Carburetor cleaner attacks plastic compo-
nents! Take precautions not to spray cleaner onto the probe handle or analyzer.
2. Insert the plastic-spray tube of the carburetor cleaner into the gas­sample hose, and then liberally spray carburetor cleaner through the hose and out the probe tube.
3. After spraying, remove all the residual cleaner by repeatedly flush­ing the gas hose and probe tube with alcohol.
4. Wipe off the surfaces of the probe and tubing with a clean rag.
2. Disconnect J9 (pump) and J11 (Battery) connectors.
3. Lift the circuit board out of the unit.
4. Remove the screw that secures the pump to the board in order to separate the pump from the board.
5. Remove tubing from pump.
6. Install new pump by reversing this procedure.
6.7 Cleaning the Probe
The probe tube and gas-sample hose will become dirty under normal use. Note that the water trap’s filter element should prevent soot from reaching the analyzer’s internal components. If the probe is not kept clean, it could become clogged and restrict the flow of gas into the analyzer, resulting in incorrect combustion test readings and calculations.
50
Instruction 24-9460
5. Allow the parts to dry completely. If available, blow compressed air through the probe to expedite the drying process.
6. Reconnect gas-sample hose to top of water trap.
6.8 Error Messages
O2 Sensor Missing - The O2 sensor is not installed.
T-STK Disconnected - The probe thermocouple is not connected to the
analyzer's T-Stack connector. Plug the probe thermocouple plug into the T-Stack connector at the bottom of the instrument.
Bad Sensor - O2 sensor is too low and can not be calibrated in the instru-
ment and needs replaced.
Instruction 24-9460
51
Fyrite
®
INSIGHT
Fyrite® INSIGHT
Low Sensor- O2 or CO sensor outputs were low but still usable.
Sensor(s) may need to be replaced in the near future. Message will indi­cate which sensor(s) were in warning.
Warmup Sensor Error -
• CO sensor was not zeroed at warmup because of high output. Run instrument on fresh air then restart instrument to re-zero sensor. If message persists, CO sensor may need to be replaced.
• Stack or Air temperature channel is measuring temperature outside the range of -4 to 212 ˚F at startup. Make sure that the Stack and Air thermocouples are sampling ambient room air within the tem­perature range at startup.
• Pressure sensor is measuring pressure outside the range of ± 3 inwc at startup. Ensure that the analyzer is sampling atmospheric pres­sure and restart the instrument.
• The analyzer was turned on with the probe sampling flue gas. Move the probe to fresh air and restart the instrument.
• Message will indicate which channel is in error.
Low Battery - Battery voltage is low. Replace the batteries.
Main Diagnostics - Lists the current status of the Stack and Air Ther-
mocouple channels, Reference Temperature channel, Pressure channel, CO channel, O
channel and Battery.
2
O2 Sensor Life - Displays the approximate remaining life of the Oxygen sensor.
Fresh Air Diagnostics - Fresh air diagnostics will cycle the instrument through the 60 second warm-up to check on the status of the sensors. The probe must be in fresh air to perform a valid check.
Access the DIAGNOSTIC menu as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If neces-
sary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight DIAGNOSTICS and then press ENT to
display the DIAGNOSTIC MENU.
3. Use the ▲▼ buttons to highlight the desired diagnostic topic and then
press ENT to display the information under that topic.
4. Press the ESC key to exit back to the DIAGNOSTIC MENU or the MENU (F2) key to exit back to the MAIN MENU.
XXXX - Occurs in the number fi elds of sensors that have achieved over-
range condition.
**** - Occurs in the number fi elds of sensors and the calculated values
that depend on the sensors that were in error coming out or warmup.
- - - - Occurs in the number fields of calculated values when Oxygen is above 16.0%.
6.9 Diagnostics Screen
The DIAGNOSTICS menu provides information regarding the operation of the analyzer. Information includes the following:
Time Meters - Displays the run time of the analyzer and sample pump in hours of operation.
52
Instruction 24-9460
6.10 Status Screen
The Status Screen provides a quick reference to key items when trouble­shooting:
• Version
• ADC Version
• Boot Version
• Model
• SN
Access the Status menu as follows:
1. Display the MAIN MENU by pressing the MENU (F2) button. If neces-
sary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight STATUS and then press ENT to dis-
play the DEVICE STATUS screen.
3. Press the ESC key to exit back to the MAIN MENU.
Instruction 24-9460
53
Fyrite
®
INSIGHT
Fyrite® INSIGHT
7.0 PARTS & SERVICE
7.1 Replacement Parts
Description Part No.
O2 Sensor ............................................................................................ 24-0788
CO Sensor .............................................................................................24-7265
B-Smart CO Sensor ............................................................................. 24-1467
O2 Sensor Cover ................................................................................... 24-1421
CO Sensor Cover Assembly .................................................................24-1484
Probe and Hose Assembly ................................................................... 24-3004
Water Trap ...........................................................................................19-3265
Filter (pkg of 3) ..................................................................................... 07-1644
Thermocouple Replacement (12 in.) .....................................................24-8414
Probe Stop ............................................................................................19-3037
Gas Connector ..................................................................................... 24-0877
Draft Connector .................................................................................. 24-0878
Connector Plate Assembly ................................................................... 24-1483
O-Ring Kit (2 sets) ............................................................................... 24-1471
USB cable ...........................................................................................104-4032
Carry Case ........................................................................................... 24-0865
Replacement Pump.............................................................................. 24-3049
PC Software .......................................................................................... 24-1470
Battery / Sensor Cover ......................................................................... 24-1453
7.2 Accessories
IrDA Printer .........................................................................................24-1400
Printer Paper (pkg of 5) ....................................................................... 24-1310
Printer Paper (pkg of 1) ...................................................................... 06-8733
Boot ....................................................................................................... 24-1461
AC Adapter ...........................................................................................24-1254
Calibration Kit .....................................................................................24-7059
Calibration Gas, 500 ppm CO ............................................................. 24-0492
Calibration Gas, 100 ppm CO .............................................................. 51-1994
Thermocouple, 1 IN ............................................................................ 104-1798
Thermocouple, 10 FT.......................................................................... 104-1797
Smoke Kit .............................................................................................21-7006
Differential Pressure & Temperature Kit.......................................... 24-8242
7.3 Service Centers
Replacement parts and service can be obtained by con tact ing one of the fol­ low ing Bacharach Service Centers:
United States Pennsylvania
621 Hunt Valley Circle New Kensington, PA 15068 Phone: 724-334-5051 Fax: 724-334-5723 Email: help@bacharach-inc.com
Canada
Bacharach of Canada, Inc. 250 Shields Court Unit #3 Markham, Ontario L3R 9W7 Canada Phone: 905-470-8985 Fax: 905-470-8963 Email: bachcan@idirect.com
México
Bacharach de México Playa Regatas No. 473 Tercer Piso Col. Militar Marte Delegación Iztacalco, 08830 México D.F. México Phones: +52-555-634-7740 +52-555-634-7741 FAX: +52-555-634-7738 Email: bacharach@prodigy.net.mx
54
Instruction 24-9460
Instruction 24-9460
55
Printed in U.S.A.
621 Hunt Valley Circle, New Kensington, PA 15068
Headquarters:
Ph: 724-334-5000 • Fax: 724-334-5001 • Toll Free: 800-736-4666
Website: www.bacharach-inc.com • E-mail: help@bacharach-inc.com
Loading...