Bacharach Fyrite Classic Operating Manual

INSTRUCTION 11-9026
FYRITE® Gas Analyzer
CO2 and O2 Indicators
Operation/Maintenance
Rev. 8 – September 2005
20
16
%CO
2
12
8
4
0
®
Bacharach, Inc.
621 Hunt Valley Circle, New Kensington, PA 15068
Phone: 724-334-5000 • Fax: 724-334-5001
Web: www.bacharach-inc.com
Printed in U.S.A. ® Registered Trademarks
WARRANTY
Bacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in material and manufacture and will conform substantially to Bacharach Inc.’s applicable specifications. Bacharach’s liability and Buyer’s remedy under this warranty are limited to the repair or replacement, at Bacharach’s option, of this Product or parts thereof returned to Seller at the factory of manufacture and shown to Bacharach Inc.’s reasonable satisfaction to have been defective; provided that written notice of the defect shall have been given by Buyer to Bacharach Inc. within one (1) year after the date of delivery of this Product by Bacharach, Inc.
Bacharach, Inc. warrants to Buyer that it will convey good title to this Product. Bacharach’s liability and Buyer’s remedy under this warranty of title are limited to the removal of any title defects or, at the election of Bacharach, to the replacement of this Product or parts thereof that are defective in title.
The warranty set forth in paragraph 1 does not apply to parts the Operating Instructions designate as having a limited shelf-life or as being expended in normal use.
THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN AND ACCEPTED IN LIEU OF (I) ANY AND ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION THE IMPLIED WARANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE: AND (II) ANY OBLIGATION, LIABILITY, RIGHT, CLAIM OR REMEDY IN CONTACT OR TORT, WHETHER OR NOT ARISING FROM BACHARACH’S NEGLIGENCE, ACTUAL OR IMPLIED. The remedies of the Buyer shall be limited to those provided herein to the exclusion of any and all other remedies including, without limitation incidental or consequential dam­ages. No agreement varying or extending the foregoing warranties, remedies or this limitation will be binding upon Bacharach, Inc. unless in writing, signed by a duly authorized officer of Bacharach.
NOTE: Gaskets are considered consumable items and are excluded from the terms of this warranty.
Register Your W arranty by Visiting
www.bacharach-inc.com
CAUTION
Corrosive fluid is contained inside this unit. These instructions should be thoroughly read before operation is attempted.
Copyright © 1998-2005 Bacharach, Inc., all rights reserved.
A
FYRITE SAFETY PRECAUTIONS
The operator (s) of this instrument should thoroughly familiarize themselves with the applicable safety precautions before handling or using the FYRITE for gas analysis.
Review Figures 1a,1b, 1c and 1d on this page. Make certain to follow the steps outlined below and read the fluid handling precautions for your personal safety.
1) Always use protective equipment such as safety goggles, gloves, and protective clothing as illustrated in Figure 1a.
2) Read the instruction manual thoroughly. When filling the FYRITE with the applicable fluid charge, use safety equipment and perform the operation in the vicinity of running tap water (Figure 1b).
3) The FYRITE fluid should always be in the bottom reservoir before and during sample introduction to the FYRITE. When venting the FYRITE to atmosphere, hold the instrument in the vertical position and at a location away from your face (Figure 1c).
4) NEVER vent FYRITE in the inverted position (with the plunger facing down­ward). This will cause fluid to spill, which is corrosive and contains harmful elements (Figure 1d).
NOTE: In the event of fluid spill - refer to Section 2.2 (FYRITE Handling Precautions). For Maintenance - refer to Section 6.0.
20
16
12
8
%CO
2
®
20
16
12
8
4
%CO
2
0
®
Figure 1a. Protective equipment required before attempting gas analysis.
20
16
12
2
%CO
4
®
Figure 1c. Venting FYRITE to atomsphere at 45° angle.
Figure 1b. Preparing to fill FYRITE using recommended protective safety equipment.
®
0
4
%CO
2
8
12
16
20
Figure 1d. NEVER vent FYRITE in the inverted position, which will cause fluid to spill.
i
TABLE OF CONTENTS
1.0 DESCRIPTION ............................................................................................................................................. 1
2.0 FYRITE INSPECTION BEFORE AND DURING TEST............................................................................ 1
2.1 Pre-Operational Check .................................................................................................................... 1
2.2 FYRITE FLUID HANDLING PRECAUTIONS ............................................................................ 2
3.0 BASIC FYRITE OPERATION PERCENT CO2 AND 02............................................................................ 3
3.1 Determining CO2 and O2 FYRITE Fluid Strength ......................................................................... 7
3.2 Operating Precautions to Ensure Maximum Accuracy..................................................................7
3.2.1 Draining Fluid Droplets .................................................................................................. 8
3.3 FYRITE Temperature ...................................................................................................................... 8
3.3.1 Temperature of Gas Sample............................................................................................ 8
4.0 PRINCIPLE OF OPERATION ...................................................................................................................... 9
5.0 APPLICATION INFORMATION ............................................................................................................... 10
5.1 FYRITE Absorbing Fluid .............................................................................................................. 10
5.1.1 FYRITE Refill Fluid (Storage Life, etc.) ..................................................................... 10
5.2 FYRITE Operating Temperature Range ....................................................................................... 11
5.3 Typical FYRITE Applications....................................................................................................... 11
5.4 CO2 FYRITE Only Combustion Testing....................................................................................... 12
5.4.1 Percent CO2 Shows Volume of Excess Air................................................................... 12
5.4.2 Too Much Excess Air Means Inefficient Combustion .................................................13
5.4.3 What is Proper CO2 Value? ........................................................................................... 13
5.4.4 How To Calculate Combustion Efficiency .................................................................. 13
5.5 O2 FYRITE Only Combustion Testing ......................................................................................... 13
5.5.1 Percent O2 Shows Volume of Excess Combustion Air ................................................ 13
5.5.2 Too Much Excess Air Means Inefficient Combustion .................................................13
5.5.3 What is Proper O2 Value? .............................................................................................. 14
5.5.4 Measuring O2 in Addition to CO2................................................................................ 14
5.5.5 Where is Percent CO2 or O2 Measured? ....................................................................... 14
5.6 Background Gases Affecting FYRITE Readings ......................................................................... 16
5.7 General FYRITE Applications ...................................................................................................... 17
5.8 Altitude Correction Table .............................................................................................................. 18
6.0 FYRITE MAINTENANCE ........................................................................................................................ 18
6.1 Increasing FYRITE Fluid To Proper Level ...................................................................................18
6.1.1 Decreasing FYRITE Fluid to Proper Level ..................................................................19
6.2 Checking FYRITE Fluid Strength ................................................................................................ 19
6.2.1 CO2 FYRITE Fluid Only............................................................................................... 19
6.2.2 O2 FYRITE Fluid Only ................................................................................................. 21
6.3 Replacing FYRITE Fluid............................................................................................................... 21
6.4 Inspection of FYRITE for Fluid Leakage ..................................................................................... 22
6.5 Cleaning FYRITE .......................................................................................................................... 24
6.6 Replacing FYRITE Plastic and Rubber Parts............................................................................... 24
6.7 Aspirator Bulb - Sampling Assembly ........................................................................................... 26
6.8 Storing FYRITE and Refill Fluid.................................................................................................. 26
6.9 Replacing Saturator Filter Packing Part #11-0121 ...................................................................... 26
7.0 ILLUSTRATED PARTS LIST FYRITE CO2/O2........................................................................................ 29
7.1 FYRITE Illustrated Parts ............................................................................................................... 30
7.2 FYRITE Sampling Assemblies...................................................................................................... 31
ii
INSTRUCTIONS 11-9026
OPERATION/MAINTENANCE
BACHARACH FYRITE MODELS
CO2 and O2 INDICATORS
1.0 DESCRIPTION (Refer to Figure 1)
The FYRITE employs the well-known “Orsat” method of volumetric analysis involv­ing chemical absorption of a sample gas, such as carbon dioxide or oxygen. The reagent used to absorb carbon dioxide (CO2) is potassium hydroxide (dyed red), and chromous chloride (blue) is the absorbent for oxygen (O2). The unique feature of the FYRITE is that the absorbing fluid is also used as the indicating fluid so that one vessel takes the place of both measuring burette and absorption pipette.
The body of the FYRITE is molded of clear high-strength plastic, comprises top and bottom reservoirs and a center tube connecting the two reservoirs. The bottom of the lower reservoir is sealed off by a synthetic rubber diaphragm that rests on a perfo­rated metal plate. The upper reservoir is covered by a molded plastic cap that contains a double-seated plunger valve. A spring holds this valve against a carefully finished seat in the top cap providing a perfect seal which makes the instrument spill-proof in any position. When the valve is partially depressed, the entire instru­ment is open to the atmosphere and the instrument is, of course, is no longer spill­proof.
The bottom reservoir is filled with the absorbing fluid, which extends about 1/4 inch into the bore of the center tube when the instrument is held upright. The scale position, mounted to one side of the center tube, is adjustable.
2.0 FYRITE INSPECTION BEFORE AND DURING TEST
CAUTION
Corrosive fluid is contained inside unit, instructions should be thoroughly read before operation is attempted.
2.1 Pre-Operational Check
To ensure proper FYRITE operation and reliable results, check the following:
(a) With FYRITE vented and in a vertical position, it should be possible to adjust
scale zero to the top of the fluid column. Refer to FYRITE Operation for proper setup (Section 3.0 Steps 1 through 5). If this is not possible, add or remove a small amount of fluid per Section 6.1.
Instruction 11-9026 Page 1
(b) FYRITE Fluid Strength
To check fluid strength.
NOTE: When repeating procedures as outlined in Section 3.0 Steps 7 through 9 (absorbing and reading percent O absorbing and reading percent CO2) and before venting
and
2
FYRITE to atmosphere for next sample, if the reading in­creases by more than 1/2 percent for either CO2 or O2, replace the fluid.
It is often desirable to check fluid strength before taking the FYRITE to a location where it will be used. It can be tested on sample atmospheric air (which contains
20.9% O
maximum accuracy. (See Section 6.9)
O
2
samples containing 10% O2.
CO
approximately 350 samples containing 10% CO
). Be certain to wet Filter Saturator when checking atmospheric air for
2
(Oxygen) Fluid - Fresh FYRITE fluid will absorb all O2 from approximately 100
(Carbon Dioxide) Fluid - Fresh FYRITE Fluid will absorb all CO2 from
2
.
2
(c) FYRITE Sampling Assembly - Always check sampling assembly for proper
(leak-free) operation as outlined in Section 6.7 before proceeding with test.
(d) Make certain wool in Filter Saturator Tube is moistened with water if using
either O2 or CO2 FYRITE in a noncombustible application where the gas sample is not fully saturated with water vapor. Examples where filter-saturator must be wetted include measuring CO samples. Refer to Section 6.9 when replacing or moistening saturator (wool)
or O2 in ambient air or from compressed gas
2
filter is required.
FYRITE OPERATION PRECAUTION: To prevent fluid spill, never depress Plunger Valve to vent FYRITE when unit is in the inverted or any other non-upright position. When depressing Plunger Valve in the upright position to vent FYRITE, hold at a slight angle away from the operator’s face.
2.2 FYRITE FLUID HANDLING PRECAUTIONS
CAUTION
FYRITE fluids used in the CO and contain poisonous elements which must not be taken internally. In the event of a spill or accidental body contact with FYRITE fluid, read the following carefully.
Page 2 Instruction 11-9026
and O2 analyzers are corrosive
2
HAZARDS
Corrosive liquid causes burns. May cause blindness if splashed in eyes. Vapors are irritating and may be harmful.
PRECAUTIONS
Prevent contact with eyes, skin and clothing. Wear eye protection and gloves. Do not vent instrument until fluid has drained from top well. Do not vent instrument (FYRITE) in inverted position.
FIRST AID
For contact with eyes: Immediately flush eyes with water 20 minutes. Get immedi­ate medical attention.
For contact with skin: Immediately flush skin with water 20 minutes. Get immedi­ate medical attention.
For inhalation overexposure: If irritation develops, move victim to fresh air.
For swallowing: Do not induce vomiting. Give two glasses of water or milk if con­scious and not convulsing. Get immediate medical attention.
Refer to the material safety data sheet for further information.
3.0 BASIC FYRITE OPERATION PERCENT CO2 AND 0
2
Four basic steps are required when making an analysis, many being common to both the CO2 and O2 FYRITE.
FYRITE MODEL CO (Red Fluid) (Blue Fluid)
2
FYRITE MODEL O
2
(a) Vent and adjust scale zero. (a) Vent and adjust scale zero.
(b) Pump sample into FYRITE. (b) Pump sample into FYRITE.
(c) Absorb CO2 from sample. (c) Absorb O2 from sample.
(d) Read % CO2 on scale. (d) Read % O2 on scale.
NOTE: To prevent warming of FYRITE fluid during analysis, hold FYRITE by the fins only.
Instruction 11-9026 Page 3
CO2 FYRITE OPERATION O2 FYRITE OPERATION
(RED FLUID) (BLUE FLUID)
1. Hold upright (Fig. 1) and
away from face. Depress Plunger Valve (momen­tarily) to vent FYRITE, and release.
2. Invert FYRITE (Fig. 2).
Hold at slight angle to drain fluid into top reservoir.
3. Turn upright. Hold
FYRITE at 45° angle (Fig. 3) momentarily to allow fluid droplet drainage into bottom reservoir.
Figure 1
Figure 2
Figure 3
1. Hold upright (Fig. 1) and away from face. Depress Plunger Valve (momen­tarily) to vent FYRITE, and release.
2. Invert FYRITE (Fig. 2) to absorb O2 drawn into FYRITE whenever Plunger Valve is de­pressed). Hold at slight angle to drain fluid into top reservoir.
3. Turn upright. Hold FYRITE at 45° angle to drain fluid into bottom reservoir. Turn upright. Repeat sequence of Steps 1, 2, and 3 twice until fluid level does not drop more that 1/2% when Plunger Valve is de­pressed. Turn upright. Hold FYRITE at 45° angle (Fig. 3) momen­tarily until excess fluid droplets have been drained from inside surfaces.
4. Hold FYRITE in upright (Fig. 4) position and away from face. Depress Plunger Valve (momen­tarily) and release.
4. Hold FYRITE in upright (Fig. 4) position and away from face. Note fluid level in column. Depress Plunger Valve (momen­tarily), and release.
Figure 4
Page 4 Instruction 11-9026
CO2 FYRITE OPERATION O2 FYRITE OPERATION
(RED FLUID) (BLUE FLUID)
5. Holding FYRITE upright (Fig. 5), loosen locknut at rear of scale. Slide scale (Fig. 5a) until top of fluid column lines up with zero line on scale (Fig. 5b). Tighten scale locknut.
NOTE: When setting scale zero, hold FYRITE vertically as shown and level with eyes while sighting across scale to top of fluid column.
6. To pump gas sample into FYRITE (Fig. 6), insert open end of metal sampling tube into area bearing gas for analysis. Hold FYRITE in upright position and place sampling assembly rubber connector tip over the Plunger Valve.
Depress Plunger Valve firmly with connector tip.
Figure 5
Figure 5a
Figure 5b
5. Holding FYRITE upright (Fig. 5), loosen locknut at rear of scale. Slide scale (Fig. 5a) until top of fluid column lines up with zero line on scale (Fig. 5b). Tighten scale locknut.
NOTE: When setting scale zero, hold FYRITE vertically as shown and level with eyes while sighting across scale to top of fluid column.
6. To pump gas sample into FYRITE (Fig. 6), insert open end of metal sampling tube into area bearing gas for analysis. Hold FYRITE in upright position and place sampling assembly rubber connector tip over the Plunger Valve.
Depress Plunger Valve firmly with connector tip.
Pump sample by squeez­ing and releasing aspirator bulb 18 times. During 18th bulb squeeze (with bulb held deflated) release connector tip and Plunger Valve.
Pump sample by squeez­ing and releasing aspirator bulb 18 times. During 18th bulb squeeze (with bulb held deflated) release connector tip and Plunger Valve.
Figure 6
Instruction 11-9026 Page 5
CO
FYRITE OPERATION O2 FYRITE OPERATION
2
(RED FLUID) (BLUE FLUID)
7. Absorb sample gas into FYRITE by inverting until fluid drains into top reservoir (Fig. 7). Then turn upright (Fig. 8) to drain fluid into bottom reservoir. Repeat this step once.
8. Hold FYRITE at 45° angle (Fig. 9) momen­tarily to allow fluid droplets to drain into the bottom reservoir.
9. With FYRITE held upright (Fig. 10), permit fluid in column to stabilize a few seconds, then immediately read % carbon dioxide on scale at the point corresponding to top of the fluid column.
7. Absorb sample gas into FYRITE by inverting until fluid drains into top reservoir (Fig. 7). Then turn upright (Fig. 8) to drain fluid into bottom reservoir. Repeat this step three (3) more times (four complete inversions total).
Figure 7
8. Hold FYRITE at 45° angle (Fig. 9) momen­tarily to allow fluid droplets to drain into the bottom reservoir.
Figure 8
9. With FYRITE held upright (Fig. 10), permit fluid in column to stabilize a few seconds, then immediately read % oxygen on scale at the point corresponding to top of the fluid column.
Figure 9
Page 6 Instruction 11-9026
Figure 10
10. This step completes CO2 or O2 FYRITE gas sample analysis. A few points to remember when reading the FYRITE:
(a) FYRITE accuracy is within ±1/2% CO
or O2 compared to actual value.
2
(b) Always handle FYRITE by fins to ensure body heat is not absorbed by fluid.
(c) A delay in reading of 5 or 10 seconds may decrease accuracy of reading
slightly but longer delays may cause substantial error.
(d) The FYRITE is calibrated to indicate on a “dry” basis for flue gas samples,
which are normally fully saturated with moisture. Failure to artificially moisten wool packing in Filter Saturator Tube when actual sample is not fully saturated will cause FYRITE to read slightly low.
Therefore, make certain wool packing in Filter Saturator Tube is suffi­ciently moistened with water for accurate results on non-flue gases. Avoid excessive moisture which can be drawn into the Aspirator Bulb and forced into the FYRITE during sampling.
CAUTION
Never depress plunger valve to vent FYRITE in the inverted position. This will cause fluid to spill, which is corrosive and contains poisonous elements. In the event of a spill, read Instruc­tions in Section 2.2 FYRITE fluid handling precautions.
3.1 Determining CO2 and O2 FYRITE Fluid Strength
FYRITE Fluid strength can be conveniently checked immediately after the first sample reading. Without venting, repeat the absorbing operations (Steps 7 through
9) by inverting FYRITE again and positioning upright until all fluid drains to the
bottom reservoir. Observe scale reading. An increase of more than 1/2% CO the second reading as compared to the first indicates a need for fluid replacement.
or O2 in
2
3.2 Operating Precautions to Ensure Maximum Accuracy
Locate top of fluid column (refer to Fig. 11). The surface at the top of the fluid column (meniscus) in the small, center bore will be dish-shaped as shown in Fig­ure 11. Either the high or low point of this dish-shaped surface may be used to locate top of fluid column providing the same point is always used both for setting scale zero and reading percent CO operation and low point for the other will cause an error.
Best practice is to use high point of this fluid surface just at the small center bore wall. In setting scale zero or reading percent CO level with eyes and sight across scale to top of fluid column.
Instruction 11-9026 Page 7
or O2. Obviously, using high point for one
2
or 02, hold FYRITE vertically and
2
3.2.1 Draining Fluid Droplets
For maximum accuracy, it is important to form the habit of following a standard procedure in this operation and to use the same procedure both before adjusting scale zero and before reading percent CO2 or O2 .
3.3 FYRITE Temperature
The FYRITE temperature should be at or close to the temperature of the working environment where the analysis is being made and should not be subject to sudden temperature changes. If FYRITE is carried from cooler to warmer location or vice versa, expose FYRITE to new temperature condition to permit uniform temperature stabilization. Usually 15 or 20 minutes will suffice, except when extreme tempera­ture variations are involved. Temperature stabilization can also be accelerated by inverting FYRITE frequently.
3.3.1 Temperature of Gas Sample
The gas sample as it enters FYRITE should be cooled (or warmed if sampling cool gases) to same (ambient) temperature as the FYRITE.
Capacity of Standard Aspirator Bulb and Sampling Assembly (Part # 11-7029) to cool the gas sample is sufficient if the temperature of gas being sampled does not exceed 850°F (454°C), even where a large number of consecutive samples are required if there is an interval of not less than 5 minutes between samples.
FYRITE indicators operate over a wide temperature range. They may be exposed to ambient temperatures from –30° to 150°F (–34° to 66° C). Gases up to 850° F (454° C) may be tested with Standard Aspirator Sampling Assembly (11-7029). Special sampling equipment for high temperatures and unsaturated gas samples are listed in Section 7.2. FYRITE Sampling Assemblies.
20
16
12
8
%CO2
4
0
®
Top of Fluid Column
Figure 11. Locating Top of Fluid Column.
Page 8 Instruction 11-9026
Loading...
+ 26 hidden pages