Bacharach Fyrite Operating Manual

Fyrite® Tech
Model 50 & 60
Combustion Gas Analyzer
Instruction 24-9428
Operation & Maintenance
Rev. 6 – August 2009
WARRANTY
Bacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in material
and manufature and will onform substantially to Baharah In.’s aliable seiations. Baharah’s
liability and Buyer’s remedy under this warranty are limited to the repair or replacement, at Bacharach’s option, of this Product or parts thereof returned to Seller at the factory of manufacture and shown to Bacharach Inc.’s reasonable satisfaction to have been defective; provided that written notice of the defect shall have been given by Buyer to Bacharach Inc. within two (2) years after the date of delivery of this Product by Bacharach, Inc.
Bacharach, Inc. warrants to Buyer that it will convey good title to this Product. Bacharach’s liability and Buyer’s remedy under this warranty of title are limited to the removal of any title defects or, at the election of Bacharach, to the replacement of this Product or parts thereof that are defective in title.
THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN AND ACCEPTED IN LIEU OF (I) ANY AND ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION THE IM­PLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE: AND (II) ANY OBLIGATION, LIABILITY, RIGHT, CLAIM OR REMEDY IN CONTRACT OR TORT, WHETHER OR NOT ARIS­ING FROM BACHARACH’S NEGLIGENCE, ACTUAL OR IMPLIED. The remedies of the Buyer shall be limited to those provided herein to the exclusion of any and all other remedies including, without limitation incidental or consequential damages. No agreement varying or extending the foregoing warranties, remedies or this limita-
tion will be binding uon Baharah, In. unless in writing, signed by a duly authorized ofer of Baharah.
Register Your Warranty by Visiting
www.mybacharach.com
Declaration of Conformity
Manufacturer’s name: Bacharach, Inc.
Manufacturer’ address: 621 Hunt Valley Circle New Kensington, PA 15068
Product name: Fyrite Tech
onforms to the following seiations:
European Directive 89/336/EEC EN50082-1: 1997 (Electrostatic Discharge) EN50082-1: 1997 (Immunity) EN55022; Group 1, Class B (Emmission)
Notice:
Produt imrovements and enhanements are ontinuous, therefore the seiations and information on­tained in this document may change without notice.
Bacharach, Inc. shall not be liable for errors contained herein or for incidental or consequential damages in con­nection with the furnishing, performance, or use of this material.
No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of Bacharach, Inc.
Copyright © 2009, Bacharach, Inc., all rights reserved. BACHARACH and Fyrite are registered trademarks of Bacharach, Inc. All other trademarks, trade names, service marks and logos referenced herein belong to their respective companies.
A
Instruction 24-9435
Fyrite Tech
Contents
Ordering Information ...............................................................................iii
1.0 INTRODUCTION .................................................................................. 1
1.1 The Fyrite Tech ................................................................................1
1.2 Operation Overview ......................................................................... 2
2.0 TECHNICAL CHARACTERISTICS...................................................3
3.0 SETTING UP THE ANALYZER .........................................................5
3.1 Scope ................................................................................................. 5
3.2 Checking & Replacing the Batteries ...............................................5
3.3 Connecting the Probe .......................................................................6
3.4 Front Panel Pushbuttons ................................................................. 8
3.5 Setup Mode .......................................................................................9
3.5.1 Entering Setup ......................................................................9
3.5.2 Selecting Temperature Units................................................9
3.5.3 CO Channel Setup (Model 60 only) ....................................10
3.5.4 Time Setup (Model 60 only) ................................................ 10
3.5.5 Date Setup (Model 60 only) ................................................ 11
3.5.6 Year Setup (Model 60 only) ................................................11
3.5.7 Selecting Printer Protocol (Model 60 only) ........................12
3.5.8 Exiting Setup .......................................................................12
4.0 OPERATION ........................................................................................13
4.1 Performing a Test ........................................................................... 13
4.1.1 Analyzer Turn On and Warm Up .......................................13
4.1.2 Selecting a Fuel ................................................................... 13
4.1.3 Manual Zero of the CO Channel (Model 60 only) ..............13
4.1.4 Sampling Point .................................................................... 14
4.1.5 Performing a Combustion Efciency Test ..........................16
4.1.6 Printing Test Results (Model 60 only) ............................... 16
4.1.7 Ending a Combustion Test ..................................................18
4.1.8 Turning OFF the Analyzer .................................................18
4.2 CO Purge During Turn OFF (Model 60 only) .............................. 18
4.3 Operating Tips ...............................................................................19
4.4 Using the Backlight .......................................................................20
4.5 Using the Probe ............................................................................. 20
4.6 Resetting the Microprocessor ........................................................20
4.7 Run Mode Screens .........................................................................21
Instruction 24-9435
i
Fyrite Tech
4.8 Screen Descriptions .......................................................................22
4.8.1 Warm-Up Screens................................................................22
4.8.2 Fuel Screen .......................................................................... 22
4.8.3 O2 Screen (Model 50 only) ...................................................23
4.8.4 CO & O2 Screen (Model 60 only) ......................................... 23
4.8.5 CO Air Free Screen (Model 60 only) ................................... 24
4.8.6 Ambient Air Temperature Screen ...................................... 24
4.8.7 Stack Temperature & Efciency Screen ............................ 25
4.8.8 CO2 & Excess Air Screen ....................................................25
5.0 CALIBRATION & MAINTENANCE ................................................27
5.1 Entering the Calibration Mode and Testing the Display Seg-
ments .....................................................................................................27
5.2 Ambient Temperature Calibration ................................................28
5.3 Stack Thermocouple Channel Calibration ...................................28
5.4 O2 Sensor Zero ...............................................................................29
5.5 CO Sensor Zero & Calibration (Model 60 only) ............................30
5.6 Water Trap / Filter Maintenance .................................................. 32
5.7 O2 Sensor Replacement .................................................................. 33
5.8 CO Sensor Replacement (Model 60 only) ...................................... 34
5.9 Pump Assembly Replacement .......................................................35
6.0 PARTS & SERVICE ............................................................................38
6.1 Replacement Parts .........................................................................38
6.2 Accessories ......................................................................................38
6.3 Service Centers ...............................................................................39
ii
Instruction 24-9435
Fyrite Tech
Ordering Information
Model No. Part No.
Measurements:
Oxygen Stack Temperature Ambient Temperature Carbon Monoxide
Calculations:
Combustion Efieny
Excess Air Carbon Dioxide CO air free
Print Capability*
Fyrite Tech 50
24-8216
X X X
X X X
Fyrite Tech 60
24-8217
X X X X
X X X X
X
Instruction 24-9435
iii
Fyrite Tech
Notes:
iv
Instruction 24-9435
Fyrite Tech
1.0 INTRODUCTION
1.1 The Fyrite Tech
Available in two models, the Fyrite Tech 50 and 60 are hand-held ana-
lyzers that are designed to calculate the efciency of residential furnaces
and combustion appliances by measuring the stack temperature and the
percent oxygen level in the ue-gas stream.
While both analyzers are capable of combustion testing, the Fyrite Tech 60 has the added capability of measuring carbon monoxide and calculating CO air-free, and also has the capability of sending combustion test data to an optional printer via an infrared port using either HP or IrDA protocol.
Fyrite Tech 50 Fyrite Tech 60
Measures: Measures:
• Oxygen (O2)
• Stack Temperature • Stack temperature
• Ambient Temperature • Ambient Temperature
• Carbon Monoxide (CO)
Calculates: Calculates:
• Combustion Efciency • Combustion Efciency
• Carbon Dioxide (CO2) • Carbon Dioxide (CO2)
• Excess Air • Excess Air
• CO air-free
• Oxygen (O2)
WARNINGS!
This analyzer is not intended to be used on a continuous basis.
The analyzer does not have an audible alarm, and is not intended to be used as a safety device.
Except for battery replacement, this analyzer should only be opened and serviced by authorized personnel.
When testing an appliance, a full visual inspection of the ap­pliance should be carried out to ensure its safe operation.
Instruction 24-9435
1
Fyrite Tech
1.2 Operation Overview
Pressing the I/O button turns the analyzer ON. Note that there is a 60 sec­ond warm-up period which must elapse before the analyzer can be used.
After warm-up is complete, choose the fuel code that corresponds to the fuel being burned by the appliance being tested. Then press the ENTER button to place the analyzer into its Run Mode and begin the combustion testing process.
Begin testing by inserting the analyzer’s probe tube into the ug-gas
stream of the appliance under test. Each analyzer model will continuously
monitor the ue gas and display measured and calculated values that
are relative to the combustion process. These values are displayed on the analyzer’s LCD, and are chosen for display by pressing the Increment () and Decrement () buttons.
The Fyrite Tech 60 has the capability of sending the current combustion test to an optional printer by pressing the HOLD button twice.
Pressing the HOLD button during a test will freeze all measured and calculated values at their current values. Pressing the ENTER button resumes testing.
A backlight enables an operator to read the display in dimly-lit areas. Pressing the LIGHT button turns the backlight ON and OFF. The backlight will automatically turn OFF after 10 minutes of keyboard inactivity.
A power-saver function will shut the analyzer OFF after 20 minutes of keyboard inactivity. Note, however, that the power-saver function is dis­abled if the detected O2 value is less than 17.9% or the CO value is more than 50 ppm.
Press the I/O button to turn the analyzer OFF. Note, that there is a 5 sec­ond delay before the analyzer actually turns OFF, during which time the unit can be kept ON by pressing the ENTER button.
If the analyzer is turned OFF while a high level of CO is still present within the unit, the pump remains running and the unit will not turn OFF until the detected CO level drops below 50 ppm.
2
Instruction 24-9435
Fyrite Tech
2.0 TECHNICAL CHARACTERISTICS
The Analyzer Directly Measures and Displays:
• Flue Gas O2 content ..................................... 0.0 to 20.9% O
2
• Stack Temperature ....................................... 0 to 999 °F (–18 to 537 °C)
• Ambient Temperature .................................. 32 to 212 °F (0 to 100 °C)
• Flue Gas CO content (Model 60) ................. 0 to 2,000 ppm CO
The Analyzer Computes and Displays:
(when the measured oxygen level is below 17.9 %)
• Combustion Efciency .................................. 0.1 to 99.9%
• Flue Gas CO2 content ................................... 0.0 to a fuel dependent
maximum value in %
• Flue Gas CO air-free content (Model 60) ..... 0 to 9,999 ppm
• Excess Air .................................................... 0 to 400%
Fuel Selection: (F1 thru F4 denotes the fuel selected as displayed on the LCD)
• Natural Gas (F1)
• Oil #2 (F2)
• LPG (F3)
• Kerosene (F4)
Normal Operating Conditions:
Temperature:
• Analyzer ....................................................... 32 to 104 °F (0 to 40 °C)
• Probe ............................................................. 1,000 °F max (538 °C) at
5" insertion
Humidity ............................................................. 15 to 90% RH,
Non-Condensing
Power Requirements .................................... Four ‘AA’ Alkaline batteries
Operating Time ........................................up to 18 hours continuous
(pump running and backlight off)
Warm Up Time ..........................................60 seconds
Instruction 24-9435
3
Fyrite Tech
Display ........................................................4 Digit, 2 Line, 7-segment
Liquid Crystal Display
Front Panel Controls ..............................Six pushbutton switches
(Refer to Section 3.4)
Accuracy:
• Oxygen....................................................±0.3% O2 w/typical ue gas
concentration of CO2
• Carbon Monoxide (Model 60) ................±5% of reading or ±10 ppm,
whichever is greater*
• Stack Temperature ................................±4 °F between 32 & 255 °F
(±2 °C between 0 & 124 °C) ±6 °F between 256 & 480 °F (±3 °C between 125 & 249 °C) ±8 °F between 481 & 752 °F (±4 °C between 250 & 400 °C)
• Ambient Temperature............................±4 °F between 32 & 104 °F
(±2 °C between 0 & 40 °C)
Size .............................................................. 7.5"H x 3.1"W x 2.1"D
(190.5 x 78.7 x 53.3 mm)
Weight .........................................................16 oz (0.454 kg) with batteries
Agency Approval ......................................CE Mark
* Tighter CO accuracy in the lower ranges, up to ±2 ppm, may be attained
if a lower range calibration gas (e.g. 100 ppm CO) is used.
4
Instruction 24-9435
Fyrite Tech
Remove battery cover by pushing down here and sliding cover outward
3.0 SETTING UP THE ANALYZER
3.1 Scope
Before using the analyzer. . .
• Check batteries (Section 3.2)
• Connect probe to analyzer (Section 3.3)
• Check setup (Section 3.5)
3.2 Checking & Replacing the Batteries
Install fresh batteries as described below. Check the analyzer for sufcient
charge prior to each use. Replace the batteries if the low-battery symbol
appears in the lower right corner of the screen. To replace the bat-
teries:
1. Remove battery cover from back of analyzer.
2. If old batteries are installed, remove them and properly discard.
3. Observing the polarity markings inside the battery compartment, install four ‘AA’ Alkaline batteries as shown in Figure 3-1.
4. Replace battery cover.
Figure 3-1. Battery Installation
Instruction 24-9435
5
Fyrite Tech
3.3 Connecting the Probe
Connect the probe assembly’s ue-gas hose and thermocouple connector to
the analyzer as follows (see Figure 3-2):
1. Push the ue gas hose onto the analyzer’s GAS inlet tting.
2. Push the ue gas thermocouple into the T-STACK jack
Important: DO NOT force thermocouple connector. The con-
nection tabs are different sizes, allowing the connector to t
in only one way.
Inspect the ue-gas hose for cracks. Replace the hose if it is defective.
Before using the analyzer, check that the Water Trap / Filter is dry and
not dirty. If necessary, dry out the trap and replace the lter element per
Section 5.6.
6
Instruction 24-9435
T-STACK
GAS
THERMOCOUPLE
PROBE STOP (Optional)
WATER TRAP / FILTER ASSEMBLY
Fyrite Tech
Instruction 24-9435
Figure 3-2. Connecting the Probe
7
Fyrite Tech
3.4 Front Panel Pushbuttons
Note that a pushbutton may perform several functions, depending on what screen is being displayed at the time.
I/O Turns analyzer ON/OFF. There is a 60 second warm-up and
a 5 second turn-off-delay period.
Places the analyzer into either its Setup or Calibration Mode when used in conjunction with the ENTER or HOLD button.
Aborts the CO purge function during turn-off.
Scrolls up through the display screens during a test.
Causes the displayed value to increase or change while in the Calibration or Setup Mode.
Scrolls down through the display screens during a test.
Causes the displayed value to decrease or change while in the Calibration or Setup Mode.
ENTER Enters the Run Mode (starts a combustion test) from the
Fuel Select Screen.
Displays the Fuel Select Screen when held down for 2 sec­onds while in the Run Mode.
Sets up the analyzer to be placed into its Calibration Mode when held down with the analyzer OFF. (Used in conjunc­tion with the I/O button.)
Stores the displayed value and automatically steps to the next screen when pressed during calibration or setup.
Places the unit into its Run Mode when held down for 2 sec­onds while in the Calibration Mode.
Aborts turn-off and keeps the analyzer turned ON when pressed during the 5 second turn-off-delay period.
HOLD Places the analyzer on hold and freezes the values in all
Run Mode Screens during a test, allowing the operator to scroll through the displays and view all test values at that point in time. Pressing ENTER resumes testing.
Sets up the analyzer to be placed into its Setup Mode when
held down with the analyzer OFF. (Used in conjunction with
the I/O button.)
Sends combustion test data to a printer when pressed twice with the pump running (Fyrite Tech 60 only).
LIGHT Toggles the backlight ON and OFF.
8
Instruction 24-9435
Fyrite Tech
3.5 Setup Mode
The analyzer is preset at the factory for the parameters shown below, but can be changed as described in their associated sections.
Function Parameter To Change
Fuel Natural Gas (F1) Section 4.1.2 Temperature Unit °F Section 3.5.2 CO Channel Auto Zero (A2) Section 3.5.3 Clock Not initialized Section 3.5.4 thru 3.5.6 Printer Protocol IrDA Section 3.5.7
3.5.1 Entering Setup
1. With the analyzer turned OFF, press and hold down the HOLD button.
2. Press the I/O button.
3. Release both buttons.
4. The analyzer is now in its Setup Mode. Refer to Sections 3.5.2 thru
3.5.7 for information on how to set up the analyzer.
3.5.2 Selecting Temperature Units
The Temperature Setup Screen is labeled “Unit.”
1. Enter the Setup Mode per Section 3.5.1. If necessary, repeatedly press the ENTER button until “Unit” is displayed.
2. Press the or button until the desired temperature unit (°F or °C) is displayed.
3. Press ENTER to move to the next Setup Screen, or press I/O to exit setup.
Instruction 24-9435
9
Loading...
+ 32 hidden pages