
FLORITE 800
INSTRUCTION 16-9013
Installation/Operation/Maintenance
Rev. 1 – March 1999
Bacharach, Inc.
625 Alpha Drive, Pittsburgh, PA 15238-2878
Phone: 1-800-736-4666 • Fax: (412) 963-2640 • Web: www.bacharach-inc.com
Printed in U.S.A.
® Registered Trademarks
of Bacharach, Inc.
Introduction
The Florite 800 is a battery powered, hand held instrument with a digital
LCD display capable of measuring the velocity, volume flowrate and temperature of air by means of a thermal probe. The unit allows the airstream
area to be entered for volume measurement. Measurements may be displayed in either Metric or Imperial units as selected from the keyboard.
After selection, this will be ‘remembered’ by the instrument until reselected
from the keyboard. The attached thermal anemometer probe has a tip
diameter of 1/4” (8mm) and is, therefore, suitable for measurements at
grilles, slot diffusers and in small ducts where large access holes are not
acceptable. The velocity sensor in the probe is compensated for variations in
airstream temperature by means of a second thermistor which also senses,
and displays, the actual airstream temperature in either metric or imperial
units consistent with the units selected for velocity.
The instrument is supplied with batteries. Four 1.5 volt AA size cells are
required which may be either disposable or rechargeable types. Either
‘Standard’ or Alkaline disposable cells may be used but the ‘Standard’ type
will exhibit a relatively short life and are, therefore, not recommended. Low
battery condition is indicated by ‘LOBAT’ appearing on the instrument
display. The cells must be replaced as soon as possible to maintain optimum
accuracy. The battery cells are accessible by sliding the cover (at the rear of
the instrument case) in the direction of the arrow. A tape is provided in the
battery compartment for easy removal of the battery cells. Ensure that the
battery cells are replaced with the correct polarity as indicated in each cell
compartment. Rotate the cell slightly forwards and backwards to check that
the cell is properly seated in the cell compartment.
Operating the Florite 800
After switching the instrument on, it will go through a warm-up procedure in
which the entire set of display segments will light up and then the instrument will count down from 9 to 0. If LOBAT is lit in the display after the
countdown the batteries need to be replaced.
All normal operating instructions can be seen on the rear of the instrument.

Zero the instrument
With the zeroing cap covering the velocity thermistor to isolate it from any
air movement, press ZERO. When zeroed the display ‘Press Zero’ will
disappear. For optimum accuracy it is necessary for the instrument to be
rezeroed with the probe at the operating temperature and zero velocity.
To set a duct area
Press the ‘vel /vol /area’ switch sequentially until ‘Area’ is displayed, followed by its current setting in either m
2
or ft2 (depending upon the unit of
measurement previously selected). Area units are automatically converted
when using the ‘I / M’ switch. Increment or decrement the area to its required value by means of the AREA + or AREA – key. Note that holding
either of these keys down will increment or decrement the duct area at an
increasing rate dependent upon how long the key is held down. Reselect
velocity or volume flowrate mode by means of the VEL / VOL / AREA key.
Using the telescopic probe
Pull on the smallest section, taking care not to damage or stress the probe
tip and ensure that the cable slides freely into the handle end of the probe.
Remove the zeroing cap and ensure that the ‘direction of flow indicator’ on
the probe handle is aligned with the ‘flow arrow’ on the probe cap. Place
probe head into the airstream pointing the arrow (or ‘direction of flow
indicator’ if the head is inside a duct) into the airstream in the direction of
flow and read the velocity and temperature of the airstream. To use the
probe inside ducting a 9.5mm (3/8") diameter minimum hole is required in
the duct wall. It may help to use the joints between the telescopic tubes as
markers to determine the probe head position. When closing down the
telescopic probe, ensure that the cable slides freely back through the probe.
Page 2