Bacharach Comfort Chek 400 Operating Manual

Page 1
Instruction 1509-9000 Operation Manual
Rev. 4 - June 2013
Product Leadership • Training • Service • Reliability
Page 2
Comfort Chek 400 Operation Manual
ii 1509-9000 Rev 4
Page 3
Important Note ............................................................................................................ iv
Warranty Policy .......................................................................................................... iv
Service Policy .............................................................................................................. v
Notices ........................................................................................................................ v
Disclaimer .................................................................................................................. vi
Revisions .................................................................................................................... vi
1.
OVERVIEW ................................................................................................................ 1
1.1. Welcome ............................................................................................................ 1
1.2. Comfort Chek 400 Shipping Checklist ............................................................... 2
1.3. Key Features ...................................................................................................... 2
1.4. General CO2 Guidelines...................................................................................... 2
2. SPECIFICATIONS ..................................................................................................... 3
3. COMPONENTS .......................................................................................................... 5
3.1. Power Supply ..................................................................................................... 5
3.2. LCD Display........................................................................................................ 5
3.3. Keypad ............................................................................................................... 6
4. OPERATION .............................................................................................................. 7
4.1. Power On/Off ...................................................................................................... 7
4.2. Taking Measurements ........................................................................................ 7
4.3. Back Light ........................................................................................................... 7
4.4. Data Hold ........................................................................................................... 8
4.5. MIN, MAX, TWA, and STEL ................................................................................ 8
4.6. Audible Alarm ..................................................................................................... 9
4.7. Auto Power Off ................................................................................................... 9
4.8. Setup Mode ........................................................................................................ 9
4.9. P1.0 CO2 Alarm Level ......................................................................................... 9
4.10. P3.0 Temperature Units .................................................................................... 9
4.11. CO2 Calibration ................................................................................................10
4.12. RH Calibration Overview ..................................................................................11
4.13. 33% Calibration ...............................................................................................11
4.14. 75% Calibration ...............................................................................................11
5. TROUBLESHOOTING ............................................................................................. 12
5.1. Cannot Power On ..............................................................................................12
5.2. Fixed (Unchanging) Readings ............................................................................12
5.3. Slow Response ..................................................................................................12
5.4. Error Codes and Descriptions ............................................................................12
5.5. Service Centers .................................................................................................12
6. CO2 LEVELS AND GUIDELINES ............................................................................. 13
6.1. Enforceable and/or Regulatory Levels ...............................................................13
6.2. Non-Enforced Guidelines and Reference Levels ................................................13
7. NOTES ..................................................................................................................... 13
TABLE OF CONTENTS
Comfort Chek 400 Operation Manual
1509-9000 Rev 1 iii
Page 4
INSTRUMENT SERIAL NUMBER:
PURCHASE DATE:
PURCHASED FROM:
IMPORTANT NOTE
Comfort Chek 400 Operation Manual
READ AND UNDERSTAND THIS MANUAL PRIOR TO USING THIS INSTRUMENT. CAREFULLY READ THE WARRANTY POLICY, SERVICE POLICY, NOTICES, DISCLAIMERS AND REVISIONS ON THE FOLLOWING PAGES.
THIS INSTRUMENT SHOULD BE INSPECTED AND CALIBRATED REGULARLY BY A QUALIFIED AND TRAINED TECHNICIAN.
THIS INSTRUMENT HAS NOT BEEN DESIGNED TO BE INTRINSICALLY SAFE IN HAZARDOUS OR EXPLOSION-RATED ENVIRONMENTS. FOR YOUR SAFETY,
DO NOT
USE IT IN AREAS CLASSIFIED AS HAZARDOUS AREAS (E.G.,
EXPLOSION-RATED ENVIRONMENTS).
WARRANTY POLICY
BACHARACH, INC. WARRANTS THIS INSTRUMENT, EXCLUDING SENSORS, TO BE FREE FROM DEFECTS IN MATERIALS AND WORKMANSHIP FOR A PERIOD OF TWO YEARS FROM THE DATE OF PURCHASE BY THE ORIGINAL OWNER. THE SENSORS HAVE A WARRANTY PERIOD OF ONE YEAR FROM THE DATE OF PURCHASE. IF THE PRODUCT SHOULD BECOME DEFECTIVE WITHIN THIS WARRANTY PERIOD, WE WILL REPAIR OR REPLACE IT AT OUR DISCRETION.
BATTERIES ARE NOT COVERED UNDER WARRANTY. WARRANTY DOES NOT COVER DAMAGE FROM MISUSE.
THE WARRANTY STATUS MAY BE AFFECTED IF THE INSTRUMENT HAS NOT BEEN USED AND MAINTAINED PER THE INSTRUCTIONS IN THIS MANUAL OR HAS BEEN ABUSED, DAMAGED, OR MODIFIED IN ANY WAY. THIS INSTRUMENT IS ONLY TO BE USED FOR PURPOSES STATED HEREIN. THE MANUFACTURER IS NOT LIABLE FOR AUXILIARY INTERFACED EQUIPMENT OR CONSEQUENTIAL DAMAGE.
DUE TO ONGOING RESEARCH, DEVELOPMENT, AND PRODUCT TESTING, THE MANUFACTURER RESERVES THE RIGHT TO CHANGE SPECIFICATIONS WITHOUT NOTICE. THE INFORMATION CONTAINED HEREIN IS BASED ON DATA CONSIDERED ACCURATE. HOWEVER, NO WARRANTY IS EXPRESSED OR IMPLIED REGARDING THE ACCURACY OF THIS DATA.
ALL GOODS MUST BE SHIPPED TO THE MANUFACTURER BY PREPAID FREIGHT. ALL RETURNED GOODS MUST BE PRE-AUTHORIZED BY OBTAINING A RETURN MERCHANDISE AUTHORIZATION (RMA) NUMBER. CONTACT THE MANUFACTURER FOR A NUMBER AND PROCEDURES REQUIRED FOR PRODUCT TRANSPORT.
1509-9000 Rev 2 iv
Page 5
SERVICE POLICY
Comfort Chek 400 Operation Manual
BACHARACH, INC. MAINTAINS AN INSTRUMENT SERVICE FACILITY AT THE FACTORY. SOME BACHARACH DISTRIBUTORS/AGENTS MAY ALSO HAVE REPAIR FACILITIES, HOWEVER, BACHARACH ASSUMES NO LIABILITY FOR SERVICE PERFORMED BY ANYONE OTHER THAN BACHARACH PERSONNEL. REPAIRS ARE WARRANTED FOR 90 DAYS AFTER DATE OF SHIPMENT (SENSORS, PUMPS, FILTERS AND BATTERIES HAVE INDIVIDUAL WARRANTIES). SHOULD YOUR INSTRUMENT REQUIRE NON-WARRANTY REPAIR, YOU MAY CONTACT THE DISTRIBUTOR FROM WHOM IT WAS PURCHASED OR YOU MAY CONTACT BACHARACH DIRECTLY.
IF BACHARACH IS TO DO THE REPAIR WORK, SEND THE INSTRUMENT, PREPAID, TO BACHARACH, INC. AT THE FOLLOWING ADDRESS.
BACHARACH, INC. 621 HUNT VALLEY CIRCLE NEW KENSINGTON, PA 15068 ATTENTION: SERVICE DEPARTMENT
ALWAYS INCLUDE YOUR RMA #, ADDRESS, TELEPHONE NUMBER, CONTACT NAME, SHIPPING/BILLING INFORMATION AND A DESCRIPTION OF THE DEFECT AS YOU PERCEIVE IT. YOU WILL BE CONTACTED WITH A COST ESTIMATE FOR EXPECTED REPAIRS PRIOR TO THE PERFORMANCE OF ANY SERVICE WORK. FOR LIABILITY REASONS, BACHARACH HAS A POLICY OF PERFORMING ALL NEEDED REPAIRS TO RESTORE THE INSTRUMENT TO FULL OPERATING CONDITION.
PRIOR TO SHIPPING EQUIPMENT TO BACHARACH, CONTACT OUR OFFICE FOR AN RMA # (RETURNED MERCHANDISE AUTHORIZATION). ALL RETURNED GOODS MUST BE ACCOMPANIED WITH AN RMA NUMBER.
PACK THE EQUIPMENT WELL (IN ITS ORIGINAL PACKING IF POSSIBLE), AS BACHARACH CANNOT BE HELD RESPONSIBLE FOR ANY DAMAGE INCURRED DURING SHIPPING TO OUR FACILITY.
NOTICES
COPYRIGHTS: THIS MANUAL IS SUBJECT TO COPYRIGHT PROTECTION; ALL RIGHTS ARE RESERVED UNDER INTERNATIONAL AND DOMESTIC COPYRIGHT LAWS. THIS MANUAL MAY NOT BE COPIED OR TRANSLATED, IN WHOLE OR IN PART, IN ANY MANNER OR FORMAT, WITHOUT THE WRITTEN PERMISSION OF BACHARACH, INC.
ALL SOFTWARE WHICH BACHARACH UTILIZES AND/OR DISTRIBUTES, HOLDS A PROPRIETARY INTEREST AND IS ALSO SUBJECT TO COPYRIGHT PROTECTION AND ALL RIGHTS ARE RESERVED. NO PARTY MAY USE OR COPY SUCH SOFTWARE IN ANY MANNER OR FORMAT, EXCEPT TO THE EXTENT THAT BACHARACH GRANTS THEM A LICENSE TO DO SO. IF THIS SOFTWARE IS BEING LOADED ONTO MORE THAN ONE COMPUTER, EXTRA SOFTWARE LICENSES MUST BE PURCHASED.
1509-9000 Rev 2 v
Page 6
Comfort Chek 400 Operation Manual
DISCLAIMER
UNDER NO CIRCUMSTANCES WILL BACHARACH, INC. BE LIABLE FOR ANY CLAIMS, LOSSES, OR DAMAGES RESULTING FROM OR ARISING OUT OF THE REPAIR OR MODIFICATION OF THIS EQUIPMENT BY A PARTY OTHER THAN BACHARACH SERVICE TECHNICIANS, OR BY OPERATION OR USE OF THE EQUIPMENT OTHER THAN IN ACCORDANCE WITH THE PRINTED INSTRUCTIONS CONTAINED WITHIN THIS MANUAL OR IF THE EQUIPMENT HAS BEEN IMPROPERLY MAINTAINED OR SUBJECTED TO NEGLECT OR ACCIDENT. ANY OF THE FORGOING WILL VOID THE WARRANTY.
REVISIONS
BACHARACH, INC. MAKES NO WARRANTY OR REPRESENTATION, EXPRESSED OR IMPLIED INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR PURPOSE, WITH RESPECT TO THIS MANUAL. ALL INFORMATION CONTAINED IN THIS MANUAL IS BELIEVED TO BE TRUE AND ACCURATE AT THE TIME OF PRINTING. HOWEVER, BACHARACH RESERVES THE RIGHT TO MAKE CHANGES AT ANY TIME WITHOUT NOTICE. REVISED COPIES OF THIS MANUAL CAN BE OBTAINED BY CONTACTING BACHARACH, INC. OR DOWNLOADING IT FROM OUR WEB SITE WWW.MYBACHARACH.COM
SHOULD YOU DETECT ANY ERRORS OR OMISSIONS IN THIS MANUAL, PLEASE CONTACT THE COMPANY AT THE FOLLOWING ADDRESS.
BACHARACH, INC. 621 HUNT VALLEY CIRCLE NEW KENSINGTON, PA 15068-7074 USA
TOLL FREE: 1-800-736-4666 FAX: 724-334-5001 E-MAIL: help@MyBacharach.com WEBSITE: www.MyBacharach.com
IN NO EVENT WILL BACHARACH, INC., OR ITS OFFICERS OR EMPLOYEES BE LIABLE FOR ANY DIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES RESULTING FROM ANY DEFECT IN ANY MANUAL, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGES.
vi 1509-9000 Rev 2
Page 7
Thank you for purchasing the Comfort
indoor environmental quality
powered IAQ instrument for
measuring and displaying carbon dioxide
), Temperature and RH indoors. It
aid in verifying readings providing an
1. OVERVIEW
1.1. Welcome
Chek 400 monitor. The Comfort Chek 400 is an easy­to-use, portable device for monitoring indoor air quality. It utilizes an infrared CO plus temperature and relative humidity (RH) sensors.
The Comfort Chek 400 is a rugged, hand­held, battery-
(CO
2
operates from internal alkaline batteries or wall adapter.
The Comfort Chek 400 is equipped with integral CO
, temperature and RH sensors to
2
indication of air quality.
NOTE: Calibration and repair are available at our manufacturing facility and through some of our authorized and trained distributors.
If after reading the manual you have any questions, please contact our customer service department for technical support.
2
Comfort Chek
sensor
Figure 1-1. Comfort Chek 400
400 Operation Manual
Figure 1-2. Comfort Chek 400 Kit
1509-9000 Rev 2 1
Page 8
Comfort Chek 400 Operation Manual
1.2. Comfort Chek 400 Shipping Checklist
This checklist ensures that you have received everything required to run your Comfort Chek400. If you do not receive any of the items listed below, contact the factory immediately.
Comfort Chek 400 instrument
4 x “AA” alkaline batteries
AC wall adapter (with North American plug)
Operation manual
Black plastic carrying/storage case
1.3. Key Features
Features of the Comfort Chek 400 include the following.
Displays CO2, temperature and RH simultaneously
Stable NDIR sensor for CO
Statistics for STEL and TWA
Backlit LCD display for dark areas
Audible alarm for CO
Battery (operates up to 10 hours) or continuously with wall adapter power supply
Easy user calibration for CO
1.4. General CO2 Guidelines
General CO2 indoor air quality guidelines are summarized in the chart below. Aside from the OSHA level of 5000 ppm for unsafe CO normal (300-400 ppm) and poorly ventilated (1000-1100 ppm) are generally accepted standards for indoor air quality. Your guidelines may vary based on local practices and personal preferences. Refer to Section 7 for additional information.
2
readings
2
detection
2
levels, the other guidelines of
2
Figure 1-3. General CO
Guidelines
2
2 1509-9000 Rev 4
Page 9
2. SPECIFICATIONS
Categories/Subcategories Descriptions
Comfort Chek 400 Operation Manual
Enclosure
Dimensions
Weight
Battery
Power
External
Display
User Interface
Audio
Warm-up
Environ­mental
Temperature
Relative Humidity
Standards Certifications
Method
Range
Resolution
Accuracy
Pressure
Sensors
Temperature
Humidity
H x W x D
6.63” x 2.75” x 1.25” (17 cm x 7 cm x 3 cm)
7.1 ounces (200 grams) including batteries
Four alkaline batteries (4 x “AA”)
9 VDC from external AC adapter which is included in kit (North American wall configuration only).
The end of the wall adapter that plugs into the instrument is 1.3 mm diameter, center positive.
CAUTION: DO NOT use rechargeable batteries in
Comfort Chek
the wall adapter to charge them.
Three parameters, backlit, numerical LCD display
Alarms will sound at 1000 PPM, the ASHRAE limit. The alarm level can be modified and the alarm sound can be turned off through the instrument menu.
30 seconds
Operating: 0° to 50° C (32° to 122° F) Storage: -20° to 700° C (-4° to 158° F)
Operating: 0 to 95% RH non-condensing
CE
Diffusion
0 - 5000 ppm
1 ppm
±30 ppm ±5% of reading (0 - 5000 ppm)
Dependence: +1.6% reading per kPa deviation from normal pressure, 100 kPa
Range: -10.0° C to 60.0° C (14° F to 140° F) Resolution: 0.1° C (0.1° F) Accuracy: ±0.6°
Range: 0.0% to 99.9% Resolution: 0.1% Accuracy: ±3% (10% to 90% range),
±5% (other ranges)
400 and DO NOT try to use the
C (±0.9° F)
1509-9000 Rev 4 3
Page 10
Comfort Chek 400 Operation Manual
Figure 2-1. Comfort Chek 400 Dimensions
4 1509-9000 Rev 4
Page 11
NOTE: If the AC adapter is plugged
lose power.
CO2 concentration in PPM
Comfort Chek 400 Operation Manual
3. COMPONENTS
3.1. Power Supply
The instrument is powered by either 4 “AA” alkaline batteries or a supplied DC wall adapter (9V/1A output). Install batteries into the battery compartment on the rear of the instrument and make sure they are in correct polarity and making good contact. When the wall adapter is used it will cut off the power supply from the batteries. The wall adapter cannot
When battery voltage gets low, “Lob” will appear on the LCD and the audible alarm sounds. The CO beeps to indicate failed CO beeps and at that time the sensor readings will not be displayed. To continue, replace the alkaline batteries with fresh ones or connect the wall adapter.
in while the instrument is powered on, the Comfort CHEK 400 will
be used as a battery charger.
infrared sensor cannot operate under low voltage, so the instrument
2
measurement. The user can press any button to stop the
2
3.2. LCD Display
Air temperature, dew point,
or wet bulb temperature in
degrees C or F.
Figure 3-1. Sample Low Battery Screen
Relative humidity in %
Figure 3-2. LCD Components
1509-9000 Rev 4 5
Page 12
Comfort Chek 400 Operation Manual
Switches instrument on and off
Sets as non-sleep mode along with “HOLD”
LCD Symbol Description
TWA
STEL
HOLD
MIN/MAX
DP
AIR
WBT
%
°T (C/F)
3.3. Keypad
Keypad Button Description
Time Weighted Average (8 hours)
Short-Term Exposure Limit (15 minute weighted average)
Readings are frozen and remain unchanged
MINimum/MAXimum readings
Low battery indicator
Dew Point temperature
AIR temperature
Wet Bulb Temperature
Unit of measurement of relative humidity
Celsius/Fahrenheit temperature unit of measure
Enters set up mode
Exits set up page/mode Enters CO Enters RH calibration along with “DP/WBT”
calibration along with “MODE”
2
Freezes the current readings Cancels data hold function
Activates or cancels the backlight Selects unit or increases value in set up
Selects AIR, DP, WBT temps display Selects unit or decreases value in set up
Activates MIN, MAX, STEL, TWA functions Saves and finishes settings
6 1509-9000 Rev 4
Page 13
4. OPERATION
4.1. Power On/Off
Press the SET button to switch instrument on. At power up, it emits a short beep and
Comfort Chek 400 Operation Manual
performs a 30-second countdown while the instrument warms up. It then enters normal operating mode displaying real time CO switch the instrument off, press the SET button.
, temperature, and RH readings. To
2
Figure 4-1. Warm-up Countdown
Figure 4-2. Normal Operating Mode
NOTE: For the purpose of saving battery life, the unit will switch itself off after 20
minutes of operation. For continuous operation, switch the unit on using both SET and HOLD buttons together.
4.2. Taking Measurements
The instrument starts taking measurements upon power up and updates information on the LCD every second. When the environment to be monitored changes (example: from high temperature to low temperature), it takes 30 seconds to update the CO reading and 30 minutes to update the RH reading.
NOTE: Do not hold the instrument close to your face because your exhalation will affect the CO
, temperature and RH readings.
2
Press DP/WBT (down arrow) button to switch between the three different temperature units of measure. The lower left portion of the LCD will cycle from air temperature to dew point temperature to wet bulb temperature.
Figure 4-3. Dew Point Temperature
4.3. Back Light
Figure 4-4. Wet Bulb Temperature
Hold down MODE for more than 1 second to activate or cancel the back light function.
2
1509-9000 Rev 4 7
Page 14
Comfort Chek 400 Operation Manual
4.4. Data Hold
Press the HOLD button to freeze the current readings. The HOLD icon is indicated at the top left corner of the LCD. All current readings remain unchanged, except TWA and STEL. Press the HOLD button again to cancel this function.
Figure 4-5. HOLD Mode
4.5. MIN, MAX, TWA, and STEL
While in normal operating mode, press MNX/AV to see the minimum, maximum, and time weighted average readings. Each press of this button displays another readings in sequence and finally returns to the normal operating mode.
In the MIN/MAX modes, the instrument indicates the minimum and maximum readings for CO left corner of the LCD and RH in the lower right corner of the LCD (Figure a).
In STEL and TWA modes, the main part of the LCD indicates the time weighted average for readings for CO The lower portions of the LCD are the current AIR, DP/WB temperatures readings and current RH readings (Figure b).
NOTE: If the instrument is switched on for less than 15 minutes, the STEL value will be the time weighted average of the readings taken since switch on. The same calculations apply to the TWA readings indicated prior to 8 hours of operation.
NOTE: It takes at least 5 minutes of recorded readings to calculate the STEL and TWA. The display indicates “- - - -” (Figure c) during the first 5 minutes of operation from power on.
NOTE: While all the other readings are held unchanged, the STEL and TWA are updated every 5 minutes.
on the upper half of the LCD, air/DP/WB temperatures in the lower
2
for the past 15 minutes (STEL) and past 8 hours (TWA).
2
(a) (b) (c)
Figure 4-6. Statistics Screens
8 1509-9000 Rev 4
Page 15
4.6. Audible Alarm
Comfort Chek 400 Operation Manual
The instrument has an audible alarm to provide warnings to the user when CO concentrations exceed the limit (see P1.0 in setup for setting alarm threshold). It emits beeps (about 80 dB) when measured CO
concentration exceeds the preset value
2
and stops when any key (except SET) is pressed or readings fall below the preset alarm value. It beeps again when the value once again exceeds the preset limit. Restart the instrument if the beeping cannot be stopped.
4.7. Auto Power Off
The instruments automatically switches off after 20 minutes of inactivity to save on battery life. To override this function, power on the instrument by holding down the SET and HOLD buttons for 2 seconds until “n” appears on the LCD.
NOTE: Auto sleep function is disabled during calibration mode.
4.8. Setup Mode
Hold down the “SET” button for more than 1 second while in normal operation mode to enter the set up mode. To exit set up mode, press the CAL Esc button in P1.0 or P3.0 and it returns to normal operating mode.
4.9. P1.0 CO2 Alarm Level
When entering the set up mode, P1.0 and AL are indicated on the LCD. Press MNX/AV button to go into P1.1 value will be blinking on the LCD. Press the MODE (up arrow) button or DP/WBT (down arrow) button to change the set point in intervals of 100 ppm. Press the MNX/AV button to save changes, or the CAL Esc button to exit without saving.
for setting the CO2 alarm threshold. The current set
2
Figure 4-7. Entering Setup Mode Figure 4-8. Current Set Value (Blinking)
4.10. P3.0 Temperature Units
Press the MODE (up arrow) button or DP/WBT (down arrow) button in P1.0 to access
for setting up the temperature scale. Press the MNX/AV button and it enters P3.1
P3.0 mode with blinking switch between down) button. Then press the MNX/AV button to save the setting or CAL Esc button to exit without saving and return to P3.0.
1509-9000 Rev 4 9
°C or °F current settings on the lower left area of the LCD. To
°C and °F, press the MODE (arrow up) button and DP/WBT (arrow
Page 16
Comfort Chek 400 Operation Manual
Figure 4-9. P3.1 Temperature Scale Figure 4-10. Temperature Units
4.11. CO2 Calibration
The instrument defaults to be calibrated manually in ambient air where CO concentration outdoors is around 400 ppm.
NOTE: Though not required, it is recommended that you calibrate the instrument prior to each use. It is also recommended that the instrument be calibrated annually by a trained technician or the factory using calibration gas.
NOTE: Do not calibrate the instrument in air with unknown CO2 concentration otherwise it will be calibrated as 400 ppm by default and the CO will be inaccurate.
It is suggested to calibrate the instrument in fresh outdoor air that is well ventilated and it is better in sunny weather. Do not calibrate the instrument in places crowed with people or close to areas where CO
concentrations may be high such as ventilation
2
outputs, near exhaust, idling cars, garages, transportation depots, traffic, etc.
Set the instrument in the calibration site, switch it on and hold down the CAL Esc and MODE (arrow up) buttons simultaneously to enter the CO
calibration mode. CAL and
2
400 ppm are blinking on the LCD while performing calibration.
readings displayed
2
2
Figure 4-11. CO2 Calibration Mode
NOTE: Ensure either the wall adapter or fresh batteries are used for the calibration
procedure because the infrared CO2 sensor draws a fairly high current. Low battery voltage may result in an interruption or failed calibration.
Wait about 10 minutes until the blinking stops and the calibration is completed automatically. The instrument returns to normal operation mode. To abort the calibration, switch the instrument off at any time.
10 1509-9000 Rev 4
Page 17
4.12. RH Calibration Overview
Comfort Chek 400 Operation Manual
The meter defaults to be calibrated at 33% RH and 75% salt solution. The ambient condition is recommended to be at 25° C with stable humidity (it’s better to be close to the calibration value). To abort the calibration, switch the instrument off at any time.
CAUTION: Do not calibrate the humidity without the default calibration salt solution as it will cause permanent damage to the RH sensor.
4.13. 33% Calibration
Plug the sensor probe into a 33% salt solution bottle. Hold down the CAL Esc button and DP/WBT button while in normal operating mode to enter the 33% calibrating mode. CAL and calibration value (32.7% if at 25°
C) are blinking on the LCD with
current temperature displayed on the lower left side.
The instrument is now calibrating and will finish in approximately 60 minutes when CAL and humidity stop blinking.
Figure 4-12. 33% Cal Mode Figure 4-13. Cal Mode Finished
4.14. 75% Calibration
After 33% calibration, plug the sensor probe into a 75% salt solution bottle, then press MNX/AV to enter 75% calibration.
CAL and calibrating value (75.2% if at 25° temperature indicated at the lower left corner. The instrument is now calibrating and will finish in approximately 60 minutes when CAL and humidity stop blinking and the instrument returns to normal operating mode.
NOTE: To calibrate 33% only, press CAL Esc and exit when 33% calibration is complete. To calibrate 75% only, press MODE (arrow up) or DP/WBT (arrow down) within 5 minutes of initializing 33% calibration.
C) are blinking on the LCD with current
1509-9000 Rev 4 11
Page 18
Comfort Chek 400 Operation Manual
United States
Email: help@MyBacharach.com
Canada
Email: bachcan@idirect.com
Figure 4-14. 75% Cal Mode
5. TROUBLESHOOTING
5.1. Cannot Power On
Press the SET button for more than 1 second and try again. In using batteries, ensure that the batteries are making good contact and that they are installed with the correct polarity. Otherwise, ensure that the wall adapter is connected and plugged in.
5.2. Fixed (Unchanging) Readings
Verify that the data hold function is not activated. In data hold mode, the HOLD icon is displayed at top left corner of LCD.
5.3. Slow Response
Ensure that the air flow vents on the rear of the instrument are not blocked.
5.4. Error Codes and Descriptions
Error Code Description
E01 CO2 sensor damage
E02 The value is under range
E03 The value is over range
E04 The original data error result in this error (DP, WB)
E07
E11 Retry humidity calibration
E17 Retry CO2 calibration
E31 Temperature sensor damaged
E34 Humidity sensor damaged
5.5. Service Centers
Battery voltage too low to measure CO2. Replace batteries or use wall adapter.
Bacharach, Inc. 621 Hunt Valley Circle New Kensington, PA 15068 USA Phone: 724-334-5051 Fax: 724-334-5723
12 1509-9000 Rev 4
Bacharach of Canada, Inc. 20 Amber St. Unit# 7 Markham, Ontario L3R SP4 Canada Phone: 905-470-8985 Fax: 905-470-8963
Loading...