Bacharach 50 User Manual

CO SENSOR
GASKET
FILTER
WARNING!
Remove jumper-wire (if installed) from pins before installation
3. Push the Calibration Cup over the sensor housing.
4. Attach a CO Span Gas Cylinder to the regulator. Then apply span gas to the sensor by adjusting the regulator knob for a fl ow rate of 2 SCFH.
5. Allow span gas to fl ow until the displayed reading stabilizes – approx. 3 minutes. Then, using Calibra­tion Tool 0006-9453 (supplied in Calibration Tool / Cup Kit 0019-3242), turn the Span Adjust poten­tiometer until the displayed reading matches the concentration stamped on the Span Gas Cylinder.
6. Calibration is now complete. Turn off the Regulator and remove the calibration equipment.
Maintenance
Battery Replacement
Remove slotted screw from rear of instrument; then lift off front case.
Replace 9V Alkaline battery (Duracell MN1604 or equiv.) in location shown in Figure 3; then reinstall front case.
Sensor Replacement
The sensor needs replaced when it can no longer be calibrated to the Span Gas Cylinder value using the Span Adjust potentiometer, or if it no longer responds to a CO source.
1. See Figure 4. Remove the screw from the rear of the instrument; then lift off the front case and remove the printed circuit board assembly.
2. Unplug the old sensor and remove its gasket. Dis­card both items properly.
3. Inspect the sensor fi lter. Replace the fi lter if it has become contaminated with dirt.
- 5 - - 6 -
Figure 3. Battery
Location
4. Remove wire-jumper (if installed) from pins of new sensor; plug in sensor and install the new gasket supplied; then reassemble the instrument.
5. Wait 30 minutes; then calibrate the new sensor.
Figure 4. Sensor Installation
Replacement Parts
Item Part No.
CO Sensor and Gasket .................................... 0019-7061
Filter ................................................................ 0019-3244
Accessories
Item Part No.
Calibration Kit ................................................ 0024-7059
Calibration Tool / Cup Kit .............................. 0019-3242
Span Gas Cylinder, 100 ppm CO in air ......... 0051-1994
Zero Gas Cylinder, 20.9% O2 in Nitrogen ...... 0051-7131
Bacharach Service Center
Replacement parts and accessories can be obtained by contacting the following Bacharach Service Center:
Bacharach, Inc., 621 Hunt Valley Circle New Kensington, PA 15068 Phone: 1-800-736-4666 • FAX: 724-334-5723
Instruction 0019-9201
Rev.4 – May 2010
®
Sni t Model 50
Carbon Monoxide Analyzer
Introduction
Designed for HVAC professionals and utility person­nel, the Snifi t Model 50 Analyzer is ideal for measuring low levels of CO in ambient air such as in rooms and garages, or around registers, furnaces, stoves, water heaters, and other types of combustion appliances. The Snifi t samples the surrounding air and shows the detected concentration of CO on its Liquid Crystal Dis­play. The Snifi t is not intended to be used in fl ue gases or in temperatures exceeding 104 °F (40 °C).
Features
• Measures and displays 0 to 1999 ppm CO in room air
• Sensitive CO sensor will last up to 2 years
• Backlight for viewing in dark areas
• Compact pocket size
• Low battery indication
• Manual zero adjust
• Factory calibrated on 100 ppm CO
• Simple fi eld calibration
• Auto power-off after 35 minutes
• Single 9V battery (included) provides at least 1500 hours of operation
621 Hunt Valley Circle, New Kensington, PA 15068
Ph: 724-334-5000 • Fax: 724-334-5001 • Toll Free: 800-736-4666
Website: www.mybacharach.com • E-mail: help@mybacharach.com
Printed in U.S.A. ® Registered trademarks
Bacharach, Inc.
MO N OXID E ANALYZER MONO XIDE ANALYZER
ppm CO
CARBONCARBON
Snif it
model 50model 50
BacklightOnOff
Span Adjust
-Power-
ZERO
ADJUST
(Recessed)
SPAN ADJUST POT.
BACKLIGHT
ON/OFF
POWER ON
POWER OFF
DISPLAY
LO BAT
LOW
BATTERY
INDICATION
1
234
Parts Shown:
1. CO Span Gas Cylinder, 51-1994
2. Regulator, 03-4318*
3. Tubing, 03-6351*
4. Flowmeter, 06-6163*
5. Calibration Cup Assembly, 23-2156** * Part of Calibration Kit 24-7059
** Part of Tool / Cup Kit 19-3242
5
PUSH CALIBRATION CUP OVER SENSOR HOUSING
LO BAT
Specifi cations
Gas Monitored ............. Carbon Monoxide (CO)
Range ........................... 0 to 1999 ppm
Resolution .................... 1 ppm
Accuracy ......................± (5% of reading + 5 ppm)
Response ...................... 30 sec. to 90% of fi nal value
Power ........................... 9V battery
Battery Life .................1500 hours minimum
Temperature Range ....32 to 104 °F (0 to 40 °C)
Case Material .............. High impact plastic
Weight .......................... 0.25 lb (0.11 kg) without battery
Dimensions .................. 5.5" L x 2.0" W x 1.6" H
(140 x 51 x 41 mm)
Operation
Turning the Snifi t On & Off
The Snifi t is turned on and off by pressing the
Power On and Power Off buttons, respec-
tively. If the instru­ment is not manually turned off, it will au­tomatically shut itself off after approximately 35 minutes.
Zeroing the Sensor
Turn on the Snifi t and allow the displayed reading to stabilize before proceeding – approximately 30 seconds. Then with the instrument sampling fresh air (air that is free of CO), use a screwdriver to adjust the Zero Adjust potentiometer at the top of the instrument for a reading of 0 ±1 ppm.
- 2 -
Figure 1. Snifi t 50
If you’re not sure about the quality of the surrounding air, you can apply a blend of Oxygen/Nitrogen gas to the sensor as described under Calibration.
If rapid temperature changes are encountered (such as from freezing to room temperatures) during the operation of the analyzer, the operator must allow the analyzer to stabilize at the ambient temperature for at least 2 minutes before taking a measurement.
Checking for CO
Important! Ensure that the sensor grille at the rear of the instrument is unobstructed and open to the atmo­sphere. A quick instrument check can be performed by allowing the smoke of a blown-out match to enter into the sensor grille. This should cause the indicated CO level to increase.
After zeroing the Snifi t, simply hold the instrument in the area that you suspect the presence of CO gas. If CO is present, the instrument will indicate the concentrA­tion of CO in ppm on its display.
For more extensive CO investigations and testing in areas of higher temperatures, the Monoxor® II is the preferred instrument.
Backlight
Pressing the Backlight button will illuminate the display for a period of 8 minutes, or until it is manually turned off by again pressing the Backlight button.
Overrange
If the CO level exceeds 1999 ppm, the displayed reading is replaced by the number “1”.
Low Battery
When the Snifi t’s 9V battery is nearing the end of its useful life, LO BAT will
- 3 -
appear in the upper left hand corner of the display. Although you can continue using the instrument under this condition, you should replace the battery as soon as possible to ensure accurate CO readings.
Figure 2. Calibration Equipment Setup
Calibration
Important! For accurate Snifi t operation, calibration every 6 months is recommended.
To calibrate the CO sensor, you will need the equip­ment listed under Accessories. Note that the Zero Gas Cylinder is needed only if you’re unsure about the qual­ity of the surrounding air for zeroing purpose.
Calibrate the sensor to a known concentration of CO gas as follows:
1. Assemble the calibration equipment per Figure 2.
2. Zero the instrument as previously described under
Zeroing the Sensor.
If necessary, you can use the calibration cup to apply
a blend of Oxygen/Nitrogen gas directly over the sen­sor by attaching a zero gas cylinder to the regulator and adjusting the regulator knob for a fl ow rate of 2 SCFH.
- 4 -
Loading...