Bacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in
mater ial and manufac ture and will conform substan tially to Bach arach Inc.’s ap plicable spec i fi ca tions.
Bacharach’s li a bil i ty and Buyer’s remedy under this warranty are lim it ed to the repair or replacement, at
Ba charach’s option, of this Product or parts thereof returned to Seller at the factory of manufac ture and
shown to Bacharach Inc.’s reasonable satisfaction to have been defective; provided that written notice of
the defect shall have been given by Buyer to Bacharach Inc. within one (1) year after the date of delivery
of this Product by Bacharach, Inc.
Bacharach, Inc. warrants to Buyer that it will convey good title to this Product. Bacharach’s liability and
Buyer’s remedy under this warranty of title are limited to the removal of any title defects or, at the election of Bacha rach, to the replacement of this Product or parts thereof that are defective in title.
THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN AND ACCEPTED IN LIEU OF
(I) ANY AND ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, IN CLUD ING WITH OUT LIMITATION THE IMPLIED WARRANTIES OF MERCHAN TABILITY AND FITNESS FOR A PARTICULAR
PURPOSE: AND (II) ANY OBLIGA TION, LIABILITY, RIGHT, CLAIM OR REM E DY IN CON TRACT OR
TORT, WHETHER OR NOT ARISING FROM BACH A RACH’S NEGLIGENCE, ACTUAL OR IM PLIED.
The remedies of the Buyer shall be limited to those provided herein to the exclusion of any and all other
remedies includ ing, without limita tion incidental or consequen tial dam ag es. No agree ment varying or
ex tend ing the foregoing warranties, remedies or this limitation will be binding upon Bacharach, Inc. unless in writing, signed by a duly au tho rized officer of Bacharach.
Register Your Warranty by Visiting
www.mybacharach.com
Notice:
Product improvements and enhancements are continuous, therefore the spec i fi ca tions and in for ma tion
contained in this document may change without notice.
Bacharach, Inc. shall not be liable for errors contained herein or for incidental or consequential damages
in connection with the furnishing, performance, or use of this material.
No part of this document may be photocopied, reproduced, or translated to another language without the
prior written consent of Bacharach, Inc.
13.0 Emportant Notice to All Customers ......................................... 50
14.0 Glossary of Terms ..................................................................... 51
15.0 Replacement Parts .................................................................... 53
15.0 Service Centers ......................................................................... 54
Instruction 19-93395
Page 6
CO2 Analyzers 3100 & 3150
1.0 Manual Guidelines
1.1 Safety and Related Information
Information in this manual that may affect the safety of the users and
others is preceeded by the following symbol:
WARNING
Failure to follow this information may result in physical injury which in
some cases could be fatal.
1.2 Notes
Important / useful information and instructions are shown clearly
throughout the manual in a note format.
For example:
NOTE: For further information, please contact your equipment
supplier.
NOTE: Actual screenshots will vary dependent on model.
Instruction 19-93396
Page 7
CO2 Analyzers 3100 & 3150
2.0 Introduction
This manual explains how to use the instruments below:
• CO
• CO
• CO
• CO
• CO
• CO
• CO
• CO
Analyzer 3100 (0 - 20%)
2
P/N 19-7076
2
(with O2) P/N 19-7077
2
(with O2 and Relative Humidity) P/N 19-7078
2
Analyzer 3150 (0 - 10,000 PPM)
2
P/N 19-7136
2
(with O2) P/N 19-7137
2
(with O2 and Relative Humidity) P/N 19-7138
2
NOTE: The instrument is a sensitive piece of scientifi c equipment,
and should be treated as such.
The CO2 analyzer is specifi cally designed to monitor CO2 for the
verifi cation of incubators in research and pharmaceutical markets.
This unit has been developed to incorporate the latest technology and
specifi cation requirements, which provide the user with a fast, simple-
to-use and accurate piece of laboratory instrumentation.
2.1 Features of the CO2 Analyzer
The analyzer has the following features:
• CO
• CO
0 to 20% (3100)
2
0 to 10,000 PPM (3150)
2
• Options for:
• O
0 to 100%
2
• Dual Temperature probes 32 to 122°F (0 to 50 °C)
• Humidity Sensor 0 to 100%
• Improved accuracy on CO
readings
2
• Time saving dual temperature probes
• Large data storage and user friendly software and download
• Easy-to-read, large well-lit display
• Built-in gas moisture removal
Applications
• IVF/Medical • Research
• Laboratories • IAQ (Indoor Air Quality)
• Brewing • Atmosphere Control
• Incubators • Gas mixing
Instruction 19-93397
Page 8
CO2 Analyzers 3100 & 3150
2.2 Instrument Components
The following items are included in the Analyzer Kit. Please check to
make sure all items have been included.
Analyzer (One of the following - refer to your model number):
CO
CO
CO
CO
CO
CO
Analyzer 31000019-7076
2
/ O2 Analyzer 31000019-7077
2
/ O2 Analyzer 3100 w/RH0019-7078
2
Analyzer 31500019-7136
2
/ O2 Analyzer 31500019-7137
2
/ O2 Analyzer 3150 w/RH0019-7138
2
Instruction Manual (on CD)0019-9339
Battery Charger with adapters0019-0610
Sample Tube Kit (includes sample fi lter and sample tube)0019-0613
Quick Start Guide0019-9338
Instruction 19-93398
Page 9
CO2 Analyzers 3100 & 3150
3.0 CO2 Analyzer Accessories
3.1 Optional Products
The CO2 Analyzer has a number of optional products for purchase
which enhance the usability and enable further analysis of data and
reading information.
NOTE: For more information on the features listed in this section
please contact your nearest Bacharach Service Center.
3.1.1 PC Software (Optional)
The software enables the user to maximize the operation of the
incubator analyzer. Instrument readings and event log data may
be uploaded to a PC for further analysis and exported to other
applications such as MS Excel. It enables direct communication with
the unit, features a simple download facility and is fully compatible with
the latest Microsoft operating systems (32-bit only). A USB cable is
provided with the software.
3.1.1.1 Event Log
The instrument incorporates the ability to log signifi cant events via
the “Event Log.” This can be used as an aid to monitoring the use of
the instrument. It can also be used as a diagnostic tool if there is a
problem with the instrument.
The event log can only be viewed via the optional software. It cannot
be viewed on the analyzer screen. Applicable events are stored in the
event log automatiaclly, meaning no user intervention is required.
The event log can hold approximately 270 events. If the log becomes
full then it begins to overwrite the older events. This can be identifi ed
by the index fi eld which starts from event number 1. The log is cleared
when the instrument is reset.
NOTE: Please refer to Section 12.0 Event Log of this operating
manual for further information.
Instruction 19-93399
Page 10
CO2 Analyzers 3100 & 3150
3.1.2 Temperature Probe Reading (Optional)
The instrument has the ability to read and display two temperature
readings via optional temperature probes (P/N 0019-0611). When
a temperature probe is fi tted to one of the temperature ports on the
top of the instrument, the display will automatically show the current
reading and a temperature probe icon will be displayed.
The display can also be changed to show a T
- T2 calculation by
1
pressing the appropriate soft-key. The current mode of operation can
be identifi ed by normal or inverse status of the soft-key, where inverse
indicates that the option is active.
NOTE: The T
- T2 result is not stored as part of the reading.
1
The operator can also choose to display the reading in either
Centrigrade or Fahrenheit using the temperature option accessed from
the Settings Menu.
Instruction 19-933910
Page 11
CO2 Analyzers 3100 & 3150
4.0 Instrument Features
4.1 Physical Characteristics
Front View Back View
Reference:
A = Main Read Screen
B = Soft-Keys
C = ON/OFF Key
D = Pump Key
E = Key 4 - Scroll LEFT
F = Key 8 - Scroll DOWN
G = Menu Key
H = Enter Key
I = Key 2 - Scroll UP
J = Key 6 - Scroll RIGHT
K = Moisture Removal Tube
L = Serial Number
M = Instrument Stand
Instruction 19-933911
Page 12
CO2 Analyzers 3100 & 3150
4.2 Panel Key Functions
KEYSFUNCTION
AMain Read ScreenStart and end screen when using the instrument.
BSoft KeysThe function of the three “soft-keys” on the front of
the instrument panel are determined by menu options taken. Functions vary from screen to screen.
CON/OFF KeyPress the “ON/OFF” key briefl y to switch the
instrument on and off.
DPump KeyPress the “Pump” key to start or stop the pump.
EScroll LEFT KeyAlso “Key 4”. Enables the operator to scroll left to
display more information.
FScroll DOWN KeyAlso “Key 8.” Enables the operator to scroll down
to display more information.
GMenu KeyPress the “Menu” key to go to the MAIN MENU.
Enables the operator to pre-set values and settings. Select options from the MAIN MENU to also
view data and readings stored or held.
HEnter KeyThe “Enter” key accepts/confi rms choices made by
the operator to various functions and operations.
Also, required to confi rm numeric data entry.
IScroll UP KeyAlso “Key 2.” Press scroll up to view further infor-
mation on the instrument read screen.
JScroll RIGHT KeyAlso “Key 6.” Press scroll right to view further
information on the instrument read screen.
KMoisture Removal TubeRemoves the moisture from the sample gas.
LSerial NumberUnique identifi cation for the instrument. Verifi ca-
tion of the serial number will be required if Technical Support assistance is needed.
MInstrument StandInstrument stand.
NOTE: Do NOT attempt to remove the cover off the back of the
analyzer which houses the moisture removal tube.
NOTE: Do NOT cover the moisture removal tube with your hand
DGas OutletGas outlet port used to exhaust the gas.
EGas InletGas inlet point used to attach the sample
FUSB PortUsed to connect the analyzer to a PC via a
GBattery Charger PortUsed to attach the mains charger to the
TypeDescription
ential temperature - incubator and ambient
temperatures.
ential temperature - incubator and ambient
temperatures.
tube and fi lter in order to take the gas read-
ing.
USB cable to download data.
analyzer for charging.
NOTE: Depending on the confi guration purchased, certain connectors
may not be present. Reference you model number for this
information.
Instruction 19-933913
Page 14
CO2 Analyzers 3100 & 3150
5.0 General Operational Instructions
5.1 Switching the Instrument ON
1) To switch on the instrument, press the “ON/OFF” key briefl y. There
will be a short beep and a slight pause followed by the Bacharach
logo.
2) The power on self-test will then commence (approximately 60
seconds).
3) If there are no warnings to display the instrument will continue to
the MAIN READ SCREEN.
5.2 Switching the Instrument OFF
1) Purge the analyzer with fresh air. Run the pump for approximately
30 seconds or until the readings have returned to normal levels.
NOTE: Before the instrument is switched off, a clean air purge should
be performed. This ensures that the instrument is free from
gas and ready for the next measurement. This fi nal purge is
especially important for the oxygen sensor as it may degrade
if stored when contaminated with gas.
2) To switch off the instrument, press the “ON/OFF” key briefl y.
NOTE: If the “AUTO OFF” setting is set to “YES”, the analyzer will
switch off automatically after 10 minutes if not in use.
Instruction 19-933914
Page 15
CO2 Analyzers 3100 & 3150
5.3 Instrument Main Read Screen
After the analyzer has been switched on and the warm-up self-tests
have completed, the analyzer will display the “MAIN READ SCREEN.”
CO2 Analyzer 3100 w/ O2 shown
Soft-Keys:
Peak - Enables the operator to display the peak reading.
Store - Enables the operator to store the reading for viewing/download
later.
Hold - Enables the operator to hold the current reading being taken.
Instruction 19-933915
Page 16
CO2 Analyzers 3100 & 3150
s
5.4 Instrument Status Icons
The following icons may be displayed on the instrument read screens:
IconDescription
Battery Charge State
(fl ashing)< 1 Hour Remaining
(fl ashing)Battery Charging
Charged
Pump Running
(fl ashing)Pump Stalled
Alarm Set
(fl ashing)Alarm Active
USB Connected to PC (fl ickers when transferring data)
Logging Mode Active (fl ashes when memory nearly full)
Temperature Probe(s) Connected
Humidity Probe Connected
(fl a
(flashing)Service Due (every 12 months)
Service Overdue
Fault / Repair (i.e. O
Waiting
cell, battery, or pump may need replacing)
2
5.5 Entering Data
During normal operation the user may be prompted to enter data or
information via the keypad, i.e. entering an ID code or setting an alarm
level.
When entering data into the instrument, all fi elds are fi xed format and
are populated from the right. For example, to enter the new time,
09:25:00, the user would type in 092500 using the numeric keypad in
the following sequence:
• : : 0
• : :09
• : 0:92
• :09:25
• 0:92:50
• 09:25:00
Press the “ENTER” key to confi rm/accept the data.
Instruction 19-933916
Page 17
CO2 Analyzers 3100 & 3150
The soft-key “DELETE” will delete the last character typed.
Alternatively, the sequence can be re-typed before the “ENTER” key
is pressed and the existing numbers will be replaced by the new
sequence.
NOTE: The instrument will not allow invalid data to be entered; this
should be deleted and re-entered.
5.6 Changing Between Parameters
By default, the instrument displays the MAIN READ SCREEN (for gas
measurement). This shows the CO
readings (O
, %RH, etc). The instrument will return to this screen after
2
power on or when returning from the menus. The “SCROLL” keys can
be used to switch to another measurement screen, i.e. temperature or
humidity. Continue to press the “SCROLL” key to return to the MAIN
READ SCREEN.
reading along with the optional
2
5.7 Memory
The memory SHOULD NOT be used as a permanent storage medium
and any important data should be transferred to a more permanent
storage medium as soon as possible. The instrument should not be
stored for prolonged periods with valuable data in its memory.
5.8 Storage
When not in use, the instrument should be kept in a clean, dry, and
warm environment such as an offi ce. It should be stored fl at in order
to help prolong the life of the O
cell.
2
Instruction 19-933917
Page 18
CO2 Analyzers 3100 & 3150
5.9 Main Menu
The “MAIN MENU” enables the operator to select options to set up
specifi c parameters and perform operational tasks prior to sample
readings being taken or to view data/information stored in the
instrument.
1) Press the “MENU” key on the front of the instrument panel and the
following screen will be displayed:
2) Press the soft-key “EXIT” to exit the MAIN MENU.
5.9.1 Information
The “INFORMATION” option enables the operator to display
information such as instrument type, serial number, current software
version, service due date, and the dates of the last factory and user
calibrations. Note that the dates are in DD/MM/YY format.
1) From the “MAIN READ SCREEN” press the “MENU” key on the
instrument panel.
2) Press “KEY 1” to display general information about the instrument.
Instruction 19-933918
Page 19
CO2 Analyzers 3100 & 3150
5.9.2 Utilities
The “UTILITIES” option enables the operator to confi gure the
instrument settings prior to taking readings.
1) From the “MAIN READ SCREEN” press the “MENU” key on the
instrument panel.
2) Press “KEY 2” to display the “UTILITIES” menu and the following
screen will be displayed:
5.9.2.1 Time & Date
The “TIME AND DATE” option enables the operator to check or set the
instrument’s internal clock. The current time/date will appear on every
stored reading.
1) From the “MAIN READ SCREEN” press the “MENU” key on the
instrument panel.
2) Press “KEY 2” to display the “UTILITIES” menu.
3) Press “KEY 1 - TIME & DATE” and the “SET TIME & DATE” menu
is displayed as follows:
Instruction 19-933919
Page 20
CO2 Analyzers 3100 & 3150
4) Press “KEY 1” to change the time or press “KEY 2” to change the
date. Type the time or date using the numeric keypad followed by
the “ENTER” key. The instrument will not allow invalid times or
dates to be entered.
NOTE: The clock will need to be manually adjusted to cope with
daylight saving changes or changes when crossing time
zones.
5.9.2.2 Contrast
The “CONTRAST” option enables the operator to adjust the
instrument screen contrast to compensate for changes in ambient
temperature. The default setting is 0.
1) From the “MAIN READ SCREEN” press the “MENU” key on the
instrument panel.
2) Press “KEY 2” to display the “UTILITIES” menu.
3) Press “KEY 2” to select the instrument panel contrast settings and
the following screen will be displayed:
4) Press “KEY 4 - SCROLL LEFT” and “KEY 6 - SCROLL RIGHT” to
adjust the value displayed.
5) Press the soft-key “ACCEPT” or “REJECT” accordingly to accept or
reject the changes.
NOTE: The manually-set contrast setting is retained when the
instrument is switched OFF.
Instruction 19-933920
Page 21
CO2 Analyzers 3100 & 3150
5.9.2.3 Alarms
The analyzer has the ability to set rising or falling alarms for the two
main gas channels, CO
and O2 (if selected as an option). The alarms
2
for each channel can be enabled or disabled independently via the
“ALARMS” menu option.
Once enabled these alarms become active in the “MAIN READ
SCREEN”; this is indicated by a bell icon. If an alarm is triggered the
screen turns red and a fl ashing bell icon is displayed. The beeper is
sounded until the gas level has recovered beyond the trigger point.
• Rising alarms are triggered when the gas level exceeds the
maximum value entered by the user.
• Falling alarms are triggered when the gas level falls below the
minimum value entered by the user.
1) From the “MAIN READ SCREEN” press the “MENU” key on the
instrument panel.
2) Press “KEY 2” to display the “UTILITIES” menu.
3) Press “KEY 3” to select alarm settings and the following screen will
be displayed:
4) Press “KEY 1” to select the CO
alarm and “KEY 2” to select the O2
2
alarm.
5) Press “KEY 1” to enable/disable the alarm, “KEY 2” to set a
maximum value, or “KEY 3” to set a minimum value.
6) Press ENTER to store the selected choice.
Instruction 19-933921
Page 22
CO2 Analyzers 3100 & 3150
5.9.2.4 Settings
The “SETTINGS” option enables the operator to maintain information
with regards to taking samples and readings.
1) From the “MAIN READ SCREEN” press the “MENU” key on the
instrument panel.
2) Press “KEY 2” to display the “UTILITIES” menu.
3) Press “KEY 4” to select settings and the following screen will be
displayed:
4) The following instrument settings man be maintained:
SettingUse
1Prompt IDPress “KEY 1” to prompt for ID code for each sample reading,
answer YES or NO accordingly.
2TemperaturePress “KEY 2” to enter the default unit temperature, choosing
from °C or °F.
3DatePress “KEY 3” to switch the date format between DD/MM/YY and
MM/DD/YY formats.
4Auto OffPress “KEY 4” to auto switch off the instrument when not in use.
Toggle between Auto Off: YES or NO. If set to YES, the instrument will switch off after 10 minutes if not in use.
5.9.2.5 Flow Fail
The “FLOW FAIL” option enables the operator to adjust the instrument
fl ow fail detection point should it fail in normal operation with a clean
fi lter.
The instrument’s internal pump can be stalled when pulling against
a vacuum or through a blocked fi lter. This is indicated by a fl ashing
pump icon; to prevent damage to the pump, the pump will switch off
after a few seconds.
Instruction 19-933922
Page 23
CO2 Analyzers 3100 & 3150
Press the ‘Pump’ key again to remove the fl ashing pump icon
NOTE: Dirty or discolored fi lters should be changed before
use. Filters that have drawn in water should be changed
immediately to prevent damage to the instrument.
1) From the “MAIN READ SCREEN” press the “MENU” key on the
instrument panel.
2) Press “KEY 2” to display the “UTILITIES” menu.
3) Press “KEY 5” to select fl ow fail and the following screen will be
displayed:
4) Use the Scroll keys to adjust the value displayed. The larger the
value, the less sensitive the fl ow fail detection will be.
5) Press the soft-key “ACCEPT” or “REJECT” accordingly.
Follow the following fl ow chart to determine the proper setting for the
Flow Fail Alarm.
Instruction 19-933923
Page 24
CO2 Analyzers 3100 & 3150
Instruction 19-933924
Page 25
CO2 Analyzers 3100 & 3150
5.9.2.6 Logging
Data logging mode can be started or stopped via “KEY 6 - LOGGING”
which is accessed from the “UTILITIES” menu. Press “KEY 4 START/STOP LOGGING” to start and stop data logging.
While in data logging mode, the instrument will automatically record
data at the pre-set intervals including running the pump for a preset
time period. Active logging mode is indicated on the “MAIN READ
SCREEN” by its icon. The operator is able to edit the default ID,
pump run-time, interval, and start/stop logging.
1) From the “MAIN READ SCREEN” press the “MENU” key on the
instrument panel.
2) Press “KEY 6” to display the “LOGGING” menu and the following
screen will be displayed:
3) Select the desired option by pressing keys “1” through “4”. Then
enter the appropriate setting using the keypad followed by the
“ENTER” key.
SettingUse
1Every 00 mins.Press “KEY 1” to enter the time in minutes for the timeframe
between sample readings. The interval controls the reading
frequency in minutes, i.e. every 10 minutes.
2Pump 00 secs.Press “KEY 2” to enter the time in seconds for the length of
time you wish the pump to run when taking a sample reading. The pump run-time is the time in seconds for which the
pump runs prior to the reading being stored. This fi gure will
also need to take into account the length of sample tube and
the volume of the sample gas. For example, there is little
point setting a pump run-time of 10 seconds if it takes 30
seconds to draw in a new sample.
3ID 00000000Press “KEY 3” to create an 8-digit numeric ID code.
4Start/Stop LoggingPress “KEY 4” to start and stop data logging.
NOTE: Data logging mode is automatically stopped when the
instrument is switched off or the logging parameters are
edited.
Instruction 19-933925
Page 26
CO2 Analyzers 3100 & 3150
5.9.2.7 Reset
The instrument can be reset by pressing “KEY 7 - RESET” accessed
via the “UTILITIES” menu. Selecting this option will clear all user
settings and any stored data including the event log.
A confi rmation code (12345678) must be entered to confi rm that a
reset is initiated.
5.9.3 Calibration
The instrument is fully calibrated during manufacture and when
returned for service. However, to improve accuracy between services,
a user/fi eld calibration can be performed.
NOTE: For further information please refer to Section 7.0 Calibration.
5.9.4 View Data
The “VIEW DATA” option enables the operator to view the stored
readings.
1) From the “MAIN READ SCREEN” press the “MENU” key on the
instrument panel.
2) Press “KEY 4” to view stored data readings and the following
screens will be displayed:
3) Press KEY 4 - SCROLL LEFT” and “KEY 6 - SCROLL RIGHT” to
move through the stored readings either forwards or backwards.
Press “KEY 2 - SCROLL UP” and “KEY 8 - SCROLL DOWN” to
switch between the fi rst (CO
, O2, & Baro) and seconds (T1, T2, &
2
humidity) group of reading parameters.
Instruction 19-933926
Page 27
CO2 Analyzers 3100 & 3150
4) Press the soft-key “MORE” to refi ne or fi lter the readings to view.
SettingUse
1Delete AllEnables the operator to delete all the readings stored, (see below).
2FilterUsed to refi ne/fi lter the range of readings displayed by ID or date
ranges. Press between two dates, after a date, before a date, or all
dates.
3Go ToEnables the operator to jump to the fi rst or last reading in the memory
or any other reading.
5.9.4.1 Clear Reading Memory
The “CLEAR READING MEMORY” function enables the user to check
how many readings have previously been taken and clear them if
necessary.
NOTE: Before readings are actually deleted, a caution message is
displayed; once readings have been deleted they cannot be
recovered.
The instrument can store up to 1000 readings. The reading structure
is fi xed and may contain optional parameters not activated for your
particular instrument confi guration, i.e. oxygen, temperature, and
humidity.
Once the reading memory is full it is not possible to store any more
readings. When the memory is full and the “STORE” key is pressed
or data logging is activated, the instrument will show a brief message
stating that the memory is full and that no further data will be recorded.
1) From the “MAIN READ SCREEN” press the “MENU” key on the
instrument panel.
Instruction 19-933927
Page 28
CO2 Analyzers 3100 & 3150
2) Press “KEY 4” to view data.
3) To clear the readings press the soft-key “MORE” followed by “KEY
1 - DELETE ALL.”
5.9.5 Diagnostics
The “DIAGNOSTICS” option enables remote technical support to
identify and resolve issues with the instrument and readings. If
required, the operator may be asked to confi rm the diagnostics
displayed.
1) From the “MAIN READ SCREEN” press the “MENU” key.
2) Press “KEY 5” to view diagnostics.
NOTE: For further information contact your nearest Bacharach
Service Center.
5.10 Warning and Error Codes
When switched on, the instrument will perform a predetermined
self-test sequence taking approximately 60 seconds. During this
time, many of the instrument’s working parameters and settings are
checked. If any operational parameters are out of specifi cation or
if the pre-programmed recommended calibration / service date has
passed, errors or warnings may be displayed.
NOTE: For further information please refer to Section 8.0 Problem
Solving.
5.11 Battery / Charging
The battery used in the instrument is a 2 Ah Lithium-Ion cell. The
instrument must be charged using the power supply supplied with your
instrument.
NOTE: Although the instrument can be powered via the USB
connecter, it cannot be charged via USB.
Instruction 19-933928
Page 29
CO2 Analyzers 3100 & 3150
When plugged into the charger, the instrument light will power on and
display charging. When complete, the display will change to show
that the instrument is charged. To switch the instrument on while
charging or charged is displayed, the operator will need to switch the
instrument off and then on again.
NOTE: A full charge will take approximately 3 hours. Typically, a fully
charged battery will last 8-10 hours.
When the instrument is already powered on the battery icon changes
to a fl ashing plug symbol. This will stop fl ashing when the charge is
complete.
Instruction 19-933929
Page 30
CO2 Analyzers 3100 & 3150
6.0 Taking Readings
6.1 Preliminary Checks - Best Practice
Prior to use, it is good practice to ensure that:
Step 1The instrument has the correct time and date set.
Step 2The sample fi lter is fi tted and that it is clean and dry.
Step 3The battery has a good charge (minimum 25% charge, even if only a few
readings are required).
Step 4The memory has suffi cient space available.
Step 5The main gases have been auto-zeroed, without gas concentration present.
Step 6If necessary, check the span calibration with a known concentration calibra-
tion-check gas.
Step 7Take readings.
• DO protect the instrument from strong direct sunlight which could
raise the temperature of the instrument beyond its operating range.
• DO remember to always use the sample fi lter! If the sample fi lter
becomes fl ooded, change it and ensure all sample tubes are clean
and dry before re-use.
• DO NOT place the instrument against anything hot as this may
cause excessive internal temperatures which can lead to erroneous
readings.
• DO NOT get the instrument wet.
6.2 Gas Measurement Process
Depending on preferences, the exact reading procedure can change.
The following method is considered best practice and when followed
correctly will allow quick and consistent readings to be recorded.
1) When the instrument is fi rst switched on it should be purged with
fresh air and allowed to stabilize for a few minutes.
2) At this point the user must zero the CO
available via the “CALIBRATION” menu. The instrument is now
ready to take the fi rst reading.
Instruction 19-933930
channel. This option is
2
Page 31
CO2 Analyzers 3100 & 3150
3) Connect the sample tube, if relevent (always use the sample fi lter)
from the sample point to the inlet port of the instrument, ensuring
that the fi lter is seated correctly.
4) Press the “PUMP” key to draw a sample into the instrument. Notice
the main gas readings start to change. It is recommended to run
the pump until the gas readings have stabilized (approximately 30
seconds) then press the soft-key “STORE” if desired.
5) The pump will stop and the operator will be prompted to enter an
ID code to identify the reading. Press the “ENTER” button. A
“READING STORED” confi rmation message will be displayed
briefl y before returning to the “MAIN READ SCREEN.”
NOTE: The ID code prompt can be switched on or off. This option
is accessed via the “SETTINGS” menu, then press “KEY 1
- PROMPT FOR ID: YES OR NO.” If the ID prompt is set to
“NO” the reading is stored using the default logging ID.
6) After each reading the instrument should be purged with fresh air.
7) Disconnect the sample tube from the instrument. Turn the pump
ON. Then run the pump for a minimum of 30 seconds. The gas
readings should return to nominal values for fresh air.
Regardless of the instrument confi guration, the following data will be
stored for each reading:
• ID code (8characters)
• Reading Type (0=User, 1=Auto, 2=Peak, 3=Hold)
• Current time / date
• Gas readings (CO
, O2)
2
• Sample Pressure (for indication only)
• Temperature (Optional)
• Humidity (Optional)
Instruction 19-933931
Page 32
CO2 Analyzers 3100 & 3150
6.3 Alternative Reading Methods
There are three other reading types or methods which each require
slightly different operating procedures.
• Logged reading
• Peak reading
• Hold reading
6.3.1 Logged Reading
Logged readings need to be confi gured and initiated via the
“UTILITIES” menu by pressing “KEY 6.” During confi guration, the user
will be asked to supply an ID, reading interval, and pump run-time.
These parameters are used to control the reading frequency in
logging mode. Once logging mode is activated, the instrument will
automatically record a reading at every interval until stopped by the
operator or the memory becomes full. Logging is also suspended
temporarily while the user is accessing the menu options.
While the logging mode is active, both the “PUMP” and “STORE” key
will be deactivated. Only logged readings can be stored.
6.3.2 Peak Reading
The operator can toggle the reading mode between normal (current)
and peak readings. While in peak reading mode, the instrument will
only display peak values for each of the channels. These values can
then be stored by pressing the “STORE” key or automatically at the
appropriate logging interval (if logging is enabled).
The peak value is reset after a reading is stored or by exiting the peak
mode using the appropriate soft-key. The current mode of operation
can be identifi ed by the status of the soft-key, either “NORMAL” or
“INVERSE”, where inverse indicates peak mode is active.
6.3.3 Hold Reading
The “HOLD READING” option allows the user to freeze the currently
displayed reading. This allows it to be manually recorded or moved
away from the sample point. Once activated, press the soft-key
“HOLD” and the readings are fi xed until the “HOLD” key is pressed
again or by storing the reading.
Instruction 19-933932
Page 33
CO2 Analyzers 3100 & 3150
7.0 Calibration
7.1 User Calibration
The instrument is fully calibrated during manufacture and when
returned for service. However, to improve accuracy between services,
a user / fi eld calibration can be performed.
User calibration is a means of optimizing the performance of the
instrument to the current operating conditions, such as ambient
temperature and pressure, as well as correcting for instrument drift
caused by lamp and fi lter settling.
In general, the instrument should not require calibration more than
once a month, but we do recommend verifying the instrument’s
operation each day
User calibration has two operations and each may be performed
individually, however, for a complete user calibration both must be
completed.
This section describes the correct procedures to achieve an accurate
user calibration.
NOTE: If the calibration is completed incorrectly it may decrease the
accuracy of the instrument.
Two important terms that are used within this section are “Zero” and
“Span.”
Zero: The point at which the instrument is calibrated when there
is no gas present.
Span: The point at which the instrument is calibrated when a
known quantity of the target gas is present.
Instruction 19-933933
Page 34
CO2 Analyzers 3100 & 3150
7.2 Calibration Gases
User calibration of the instrument will improve the data accuracy in
the range of the calibrated gases used. However, it may cause less
accurate readings of concentrations outside this calibrated range.
Users should select the correct calibration gas for the expected gas
levels on their particular application. Only use gases with a known
certifi ed gas concentration.
NOTE: Certifi ed calibration gases can be supplied by Bacharach.
Warning: For each gas used the appropriate material safety
data sheet must be read and understood before proceeding.
Calibration gases and the use of pressure regulators can be
dangerous.
7.3 Calibration Set-Up
The regulator supplied with the calibration kit has been confi gured
to deliver a fi xed fl ow. It only requires a few turns to open and no
adjustment is necessary.
Warning: When the instrument is being calibrated there are
two possible exits for the gas; via the usual manner out of the
exhaust port of the instrument or via the port on the pressure
relief valve. It is recommended that both ports have exhaust
tubing attached. The exhaust tubing must emerge in a wellvented area. Ensure that there are no leaks in the tubing
connections. The calibration should always be carried out in a
safe area with all necessary precautions taken as all pressurized
gases are potentially dangerous.
Instruction 19-933934
Page 35
CO2 Analyzers 3100 & 3150
7.4 Calibration Equipment
The diagram below displays the regulator and tubing equipment for
user calibration:
Calibration Kits are available through Bacharach Inc.
Calibration Kit: P/N 19-8027
Cylinder of N
Cylinder of 5% CO
Cylinder of 2500 ppm CO
Gas: P/N 23-4003
2
Gas: P/N 24-1126 (Model 3100)
2
Gas: P/N 24-1130 ( Model 3150)
2
• Contact your nearest Bacharach Service Center for further
information.
• The regulator supplied with the calibration kit is recommended as
fl ow and pressure rates are factory set.
• A relief valve is not supplied with the kit but is recommended in
addition to the supplied regulator for improved fl ow.
Instruction 19-933935
Page 36
CO2 Analyzers 3100 & 3150
7.5 Calibration Method
Before you begin, ensure that the unit is stabilized at its working
temperature before performing any of the calibration operations.
To achieve the processes set out in this section, press “KEY 3 CALIBRATION” from the “MAIN” menu. The fi rst screen displayed
provides the option to select the gas that requires calibration.
The exact calibration method can vary depending on the gases used.
7.5.1 Zero CO
For maximum accuracy it is recommended that the CO
zeroed using bottled gas (certifi ed 100% N
Channel
2
2
).
channel is
2
If nitrogen is not available the user can select the option to perform an
air calibration. This option assumes that the user has access to fresh
air at around 390 ppm CO
. Generally, this can be found outside or in
2
a well ventilated corridor (typically, an offi ce lab would have a higher
CO
concentration).
2
To zero the CO
1) From the “CALIBRATION” menu press “KEY 2 - CO
channel, perfom the following:
2
CHANNEL.”
2
Instruction 19-933936
Page 37
CO2 Analyzers 3100 & 3150
2) Press either “KEY 1 - ZERO WITH N2” (recommended) or “KEY 2 ZERO WITH AIR” from the menu. Then, either attach the 100% N
2
or sample pipe to allow access to fresh air.
3) Press the “START” key. The instrument will now wait approximately
60 seconds for the gas reading to stabilize at the correct level.
Ensure the pump is on (pump icon appears in lower middle of
screen when on).
4) The instrument will then indicate a successful zero has been
completed. Press the soft-key “ACCEPT” to confi rm the calibration
and “STORE” the new user offset. Alternatively, select the soft-key
“REJECT” to exit without change.
NOTE: Typical zero calibration, although displaying zero, needs to be
given time to settle. We recommend accepting the calibration
at leastfi ve minutes after the display concentration stabilizes.
NOTE: If the calibration failed, purge and try again or select a
different air source.
NOTE: If using the Analyzer 3150 - very high concentrations of CO
may take up to 30 minutes to purge completely.
7.5.2 Span CO
Channel
2
It is recommended that the instrument is spanned to target the desired
reading range (e.g. 5%); ideally this should not be a low level close to
zero.
1) If not already preset, enter the span target (i.e. certifi ed
concentration of your calibration gas) by pressing “KEY 1” and then
entering the new desired value followed by ENTER. Then attach
the gas and open the regulator valve to allow the gas to fl ow.
2) Press the “START” key and wait for the reading to stabilize. This
can take a couple of minutes. Ensure the pump is on (pump icon
appears in lower middle of screen when on).
2
Instruction 19-933937
Page 38
CO2 Analyzers 3100 & 3150
3) Once a stable reading is shown, press the soft-key “ACCEPT.”
A successful span calibration message will then be displayed.
Press the soft-key “ACCEPT” again to confi rm the calibration
and “STORE” the new user span. Alternatively, press soft-key
“REJECT” to exit without change.
NOTE: If the calibration failed then try again using a longer purge
time or different target gas.
7.5.3 Zero O
It is not necessary to zero the O
Channel
2
channel. A span calibration should
2
correct the reading across the entire range.
7.5.4 Span O
It is recommended that the O
Channel
2
channel is spanned in fresh air with
2
a target concentration of 20.9%, although other gases and target
concentrations can be used if required.
1) If not already preset, enter the span target (i.e. certifi ed
concentration of your calibration gas) by pressing “KEY 1” followed
by the desired value. Press ENTER to store.
2) Press the soft-key “START” and wait for the reading to stabilize.
Press the “PUMP” key to draw in fresh air. It can take a couple of
minutes to stabilize.
3) Once a stable reading is shown press the soft-key “ACCEPT.” A
successful span calibration message should then be displayed.
Press the soft-key “ACCEPT” again to confi rm the calibration and
“STORE” the new user span. Alternatively, press the soft-key
“REJECT” to exit without change.
NOTE: If the calibration failed then try again using a longer purge
time or different target gas.
Instruction 19-933938
Page 39
CO2 Analyzers 3100 & 3150
7.5.5 Reset Factory Settings
This option will reset the instrument to its factory programmed
calibration characteristics and will clear the user calibration points for
both gas channels.
1) To reset to factory settings, press “KEY 1 - FACTORY RESET” from
the “CALIBRATION” menu.
2) To prevent the user calibration data being accidently erased the
user must confi rm the action by pressing the soft-key “ACCEPT” or
soft-key “REJECT” to exit without change.
7.6 Last Field Calibration
This data can be found in the “INFORMATION” screen accessed via
the “UTILITIES” menu. This option displays the date that the last fi eld
calibration was performed on the instrument.
7.7 Calibration Record
The CO2 instruments have the facility to log user calibrations via
the “EVENT LOG.” This can be used as an aid in ensuring that gas
measurements are valid and accurate.
During calibration the instrument will record the following in the event
log. For each entry the time and date will be stored.
Instruction 19-933939
Page 40
EventData Recorded
CO2 Analyzers 3100 & 3150
Successful User Zero CO
Successful User Span CO
Successful User Span O
Failed User Zero CO
Failed User Span CO
Failed User Span O
2
2
2
2
2
2
Type (N2 or Air) and Readings Before and After
Target Value, Readings Before and After
Target Value, Readings Before and After
Type (N2 or Air) and Reading
Target Value, Gas Reading
Target Value, Gas Reading
Return to Factory Settings
NOTE: If the calibration failed, then try again using a longer purge
time or different target gas. This event log can only be
downloaded and viewed via the optional software. It cannot
be viewed on the analyzer screen.
Instruction 19-933940
Page 41
CO2 Analyzers 3100 & 3150
8.0 Problem Solving
This section outlines various warning and error messages which the
operator may receive during general operation of the instrument. For
further assistance please contact your nearest Bacharach Service
Center.
8.1 Warnings and Errors
When switched on, the instrument will perform a predetermined
self-test sequence taking approximately 60 seconds. During this
time many of the instrument’s working parameters and settings are
checked.
If any operational parameters are out of specifi cation or if the pre-
programmed recommended calibration / service date has passed,
errors or warnings may be displayed.
Use the “SCROLL UP” and “SCROLL DOWN” keys to move through
the list if required.
There are two types of warning that may be displayed:
• General warnings that may not affect the instrument’s function and
those where the self-test has detected a function that is outside of
the usual operating criteria, e.g. battery charge low, memory nearly
full.
• Operational parameters that could affect the performance of the
instrument, e.g. CO2 out of calibration.
The most likely reason for these errors is either an incorrect user
calibration, or may indicate sensor failure. If an incorrect user
calibration has caused the warning it should be correctable by the way
of returning the instrument to factory settings, zeroing or carrying out a
user calibration as necessary for the relevant function.
Instruction 19-933941
Page 42
CO2 Analyzers 3100 & 3150
Under and Over Range Codes
If a reading is over range (i.e. above the maximum allowed reading)
it will be displayed with ‘more than’ chevrons (>>.>). This can occur
if a channel has been incorrectly calibrated or the sample gas has
exceeded its specifi ed range (e.g. CO
>20%).
2
If a reading is under range (i.e. below zero) it will be displayed with
‘less than’ chevrons (<<.<). Refer to Section 7.0 Calibration to remedy
under-range by performing a user zero.
A number displayed as asteriks (**.*) indicates an error, usually where
the instrument has been unable to complete a particular calculation.
Typically, this will be the fi rst indication of a fault condition.
Where no data is available, dashes (--.-) are displayed. This usually
occurs when a particular reading or parameter has been skipped by
the user, or where an optional accessory is not fi tted correctly, i.e. a
temperature probe.
8.2 Self-Test Warning Messages
The following warnings may be displayed during the self-test period
when the instrument is switched ON.
WarningDescription
Check MemoryThe instrument only has space to store less than 50 readings
before it is full. The exact number can be checked using the “VIEW
READINGS” option.
Memory FullThere is no more space in memory to store readings. Both the
store and log options will be disabled until the memory is cleared.
The readings should be downloaded to PC using the optional software before memory is cleared.
Battery LowThe instrument does not have enough power to operate for a full
day. The instrument should be recharged or connected to an external power supply.
Service DueIt has been 12 months (or more) since your instrument was
returned to the manufacturer for a service. The performance and
accuracy of the instrument may be impaired.
Low FlowThe instrument’s gas inlet (or outlet) may be blocked. This warning
is most commonly caused by a water-logged or dirty sample fi lter.
Change the sample fi lter and check for obvious blockages in the
sample tubes.
Instruction 19-933942
Page 43
CO2 Analyzers 3100 & 3150
Check CO2 Cal.This warning is most commonly caused by an incorrect user cali-
bration. Try recalibrating the sensor or press “RETURN TO FACTORY SETTINGS.” If the warning persists it may be caused by
dirt or damage to the infrared sensor. The instrument will need to
be returned to your nearest Bacharach Service Center for service
/ repair.
Check O
Ref. FaultThis may be caused by dirt or damage to the infrared sensor in the
*Invalid Confi g.The instrument has detected a problem with the confi guration
Change O
Change PumpThe pump has exceeded its recommended run-time and should be
Change BatteryThe Lithium Ion battery has exceeded its recommende life-time or
Cal.This warning is most commonly caused by an incorrect user cali-
2
CellThe oxygen cell has not been changed for at least 2 years; its
2
bration. Try re-calibrating the sensor or press “RETURN TO FACTORY SETTINGS.” If the warning persists it may be caused by a
damaged or faulty sensor. The instrument will need to be returned
to your nearest Bacharach Service Center for service / repair.
instrument. The instrument will need to be returned to your nearest
Bacharach Service Center for repair.
parameters. The instrument will need to be returned to your nearest Bacharach Service Center for service / repair. Most likely to be
caused after fi rmware update.
performance and accuracy may be impaired. The instrument will
need to be returned to your nearest Bacharach Service Center for
service.
changed. The instrument will need to be returned to your nearest
Bacharach Service Center for service.
number of charge cycles and should be changed. The instrument
will need to be returned to your nearest Bacharach Service Center
for service.
User Cal. DueIf has been over a month since the instrument was last user cali-
brated. For optimal performance and accuracy it is recommended
that the instrument is user calibrated each time it is used.
Invalid TimeThe instrument has an invalid time. This is most likely to occur af-
ter a reset. The correct time should be enterd using the set “TIME
& DATE” option via the “UTILITIES” menu.
Invalid DateThe instrument has an invalid date. This is most likely to occur af-
ter a reset. The correct date should be entered using the set “TIME
& DATE” option via the “UTILITIES” menu.
Baro. FaultThe instrument has detected a fault with the barometric sensor
of its calibration. This will have an effect on the accuracy of the
readings as they are pressure compensated. The instrument will
need to be returned to your nearest Bacharach Service Center for
service.
NOTE: Certain confi guration problems can be corrected remotely.
Using the PC software, it is possible to export the current
confi guration and e-mail it to a Bacharach Service Center.
Depending on the error it may be possible to correct the
confi guration fi le and import it back into the instrument.
Instruction 19-933943
Page 44
CO2 Analyzers 3100 & 3150
8.3 User Calibration Trouble Shooting
ErrorRemedy
User Zero FailedA possible reason for this is because the instrument is trying to
zero to a level which is outside the predetermined range set when
the unit was calibrated at the factory. To rectify this, fi rst ensure
the unit contains absolutely none of the gas which is being zeroed
by fl ushing thoroughly with nitrogen. If it will not zero, then refer to
the instructions given in the “FACTORY SETTINGS” section. If the
instrument continues to fail to zero then the unit must be returned to
your nearest Bacharach Service Center for investigation.
Calibration FailedCheck that the span target is set to the correct value. If not, cor-
rect and retry spanning the channel. Repeat the entire procedure,
including zeroing the channel, and then calibrate the span. Ensure
that the reading is stable before spanning the channel.
8.4 Cross-Gas Effects
Carbon Dioxide is measured by infrared absorption at a wavelength
specifi c to carbon dioxide. Therefore, the carbon dioxide reading will
not be affected by any other gases. The oxygen sensor is a galvanic
cell type and suffers virtually no infl uence from CO
SO
, or H2, unlike many other types of oxygen cell.
2
, CO, H2S, NO2,
2
8.5 Hardware Reset
If for any reason the instrument ‘locks up’ and will not switch off, it is
possible to force a hardware reset. Press and hold the “ON/OFF” key
for 10 seconds; wait at least 15 seconds and the instrument should
restart automatically.
NOTE: Performing a hardware reset may cause loss or corruption of
currently stored data including the time and date.
Instruction 19-933944
Page 45
CO2 Analyzers 3100 & 3150
9.0 Service
The instrument should be regularly serviced to ensure correct
operation and accurate readings. Bacharach recommends a full
service and factory calibration every 12 months.
Depending on usage, the O
cell should be replaced every 2-3 years.
2
User Serviceable Parts
NOTE: There are no user serviceable parts inside of the instrument.
Please do not attempt any repair as this may void the
warranty.
The following parts are supplied and can be user serviced:
Sample FilterThis should be regularly inspected for damage or discoloration
and changed if needed. The instrument should never be operated
without the sample fi lter as this may result in water or dust entering
the instrument. The fi lter should be changed immediately if water
can be seen.
Sample TubingAlways ensure that the sample tubes are not contaminated or dam-
aged.
Instruction 19-933945
Page 46
CO2 Analyzers 3100 & 3150
10.0 Warranty Policy
This instrument is guaranteed, to the original end user purchaser,
against defect in materials and workmanship for a period of 12 months
from the date of the shipment.
This period covers repair or replacement of defective parts on an
exchange basis.
The decision to repair or replace will be determined by Bacharach Inc.
To maintain this warranty, the purchaser must perform maintenance
and calibration as prescribed in the operating manual.
Normal wear and tear, and parts damaged by abuse, misuse,
negligence, or accidents are specifi cally excluded from the warranty.
NOTE: Please contact your nearest Bacharach Service Center for
further information.
Instruction 19-933946
Page 47
CO2 Analyzers 3100 & 3150
11.0 Technical Specifi cation
Power Supply
Battery TypeLithium Ion
Battery Life10 Hours (8 hours with pump)
Battery Lifetime>300 Cycles
Battery Charger5V DC external power supply and internal charging circuit
Charge Time3 Hours
Alternative PowerUSB connector (run only)
DC Power Supply (run + charge)
Gas Ranges
Gases MeasuredCO2 By custom dual wavelength infrared cell
and pressure
±2.0% Full Scale over operating temperature range
2
Accuracy: ±(1.5% of range +2% of reading) at reference conditions
1
Temperature dependence: ± 0.2% °C
(typical at 1000 ppm CO2)
Pressure Dependence: ± 0.02% of
reading/hPa (typical at 1000 ppm CO
2
±1.0% Full Scale at constant tempera-
)
2
ture and pressure
±2.0% Full Scale over operating temperature range
2
2
≤20 seconds
≤60 seconds
1
Conditions during factory calibration, typically 68 °F, 29.5 inHg
Instruction 19-933947
Page 48
CO2 Analyzers 3100 & 3150
Characteristics
Temperature (optional)32 to 122 °F (0 to 50.0 °C)
Temperature accuracy, typical±0.36°F from 89.6 to 111.2 °F, ±0.9°F over the rest of
the range (±0.2°C from 32 to 44 °C, ±0.5°C over the
rest of the range)
Barometric Pressure23.6 to 35.4 inHg (800 to 1200 mbar)
RH Measurement (Optional)0 to 100% non-condensing
RH Accuracy±1.5% across range
Visual and Audible AlarmUser-selectable CO
CommunicationsUSB type B mini-connector, HID device class
Data Storage1000 readings + 270 events
and O2 alarm levels
2
Pump
Flow100 cc/min (typical); 60 cc/min (minimum)
Environmental Conditions
Operating Temperature range41 to 104°F (5 to 40°C)
Relative Humidity0 to 95% non-condensing (RH probe 0 to 100% non-
condensing)
Barometric Pressure±14.8 inHg from calibration pressure (±500 mbar)
Physical Features
Weight18 oz (495 grams)
SizeL 6.5” x W 4” x D 2.2” (L 165mm x W 100mm x D 55mm)
Case MaterialABS / Polypropylene with Silicone Rubber Inserts
Keys17 Resin capped Silicone rubber keys
DisplayLiquid crystal display, 128 x 64 pixel with RGB LED back-light
Gas sample fi ltersBuilt-in gas dryer tube to remove moisture
NOTE: Specifi cations are subject to change without notice.
Instruction 19-933948
User-replaceable PTFE water fi lter
Page 49
CO2 Analyzers 3100 & 3150
12.0 Event Log
The following events are recorded in the instrument’s event log. The
event log can only be downloaded using the additional PC software.
Please refer to the PC software manual for further details.
EventData
Cold Start / ResetType of start (MCUSR, boot_key)
Firmware VersionDescription
Set TimeBefore and After
Set DateBefore and After
Re-Flash RequestedType
Restore to Factory SettingsType
Comms Clear MemoryType 0=Readings, 1=Event Log
Change ContrastBefore and After
Change Flow Fail Current LimitBefore and After
Battery Less Than Critical VoltageCritical, Actual
RTC Date/Time InvalidNone
Factory Calibration Invalid or OverdueDate, Difference
Service Invalid or OverdueDate, Difference
Power ON Self-Test, Sensor Out of RangeChannel, Reading, Limit
User Calibration Set Zero OKBefore, After
User Calibration Set Span OKTarget, Before, After
User Calibration Set Zero FailedTarget, Reading
User Calibration Set Span FailedTarget, Reading
Attempt to Store when Readings Memory
Full
Readings Memory Nearly FullLimit, Actual
Change Logging ModeStatus, Interval, Pump time
Change Logging Mode IDID
Flow Fail Current Limit ExceededLimit, Actual
Max
Instruction 19-933949
Page 50
CO2 Analyzers 3100 & 3150
13.0 Important Notice to All Customers
The trash can symbol now displayed on equipment supplied signifi es
that the apparatus MUST NOT be disposed of through the normal
municipal waste stream but through a registered recycling scheme.
The Waste Electrical and Electronic Equipment directive (WEEE)
makes producers responsible from July 1st 2007 in meeting their
obligations, with the fundamental aim of reducing the environmental
impact of electrical and electronic equipment at the end of its life.
The OEM is now registered as a producer and has joined a recycling
scheme provider who will manage and report on our electrical waste
on our behalf.
Producer Registration Number is WEE/GB0052TQ
So when your instrument is at the end of its life, please contact you
nearest Bacharach Service Center who will advise you on the next
step in order to help us meet our obligations.
Instruction 19-933950
Page 51
CO2 Analyzers 3100 & 3150
14.0 Glossary of Terms
Analyzer error messagesFor a list of standard error codes and for more
information, please refer to Section 8.0 Problem
Solving.
Analyzer warningsThere are two types of warning messages dis-
played; general warnings that may not necessarily affect the instrument’s function, for example
battery power low, and operational parameters that
could affect the performance of the analyzer, for
example CO
Battery chargeA full battery charge will take approximately three
hours.
CalibrationThe process that an instrument will undergo to
enable it to measure and display the various
parameters in accordance with the manufacturer’s
specifi cation.
Chemical cellType of gas detector which can be fi tted internally
to the analyzer at the time of manufacture.
out of calibration.
2
Clean air purgeProcess used to clear out gas from the inlet pipe
and the analyzer’s gas sensors prior to taking a
new reading.
CO
2
Carbon dioxide gas.
Contrast adjustmentAdjustable setting which darkens or lightens the
text displayed on the screen. Typically, this is
used to compensate for different environmental
temperatures. High temperature causes the
display to darken and low temperature causes the
display to lighten.
Data loggingA mode of operation that enables the user to
leave the analyzer unattended to take readings
automatically at predetermined times. The reading
interval and pump run-time may be adjusted prior
to commencing the logging cycle.
DownloadTerminology used to describe the transfer of
data from the analyzer to a PC via the optional
software.
PC softwareThe optional software enables the user to
maximize the operation of the incubator analyzer.
Instrument readings and event log data may be
installed to a PC for further analysis and exported
to other applications such as MS Excel.
Instruction 19-933951
Page 52
CO2 Analyzers 3100 & 3150
Event logRecord of signifi cant events in the life of the ana-
lyzer. Used as an aid to monitoring the use of the
analyzer. It can also be used as a diagnostic tool
if there is a problem with the analyzer. The event
log can be viewed via the optional software. It
cannot be viewed on the analyzer screen.
Exhaust portThe point at which the gas exits the analyzer. This
is located on the top of the analyzer. It is threaded
to take an M5 hose-barb to allow an exhaust tube
to be attached if required.
Exhaust tubeClear plastic tubing used to route gases from the
exhaust port.
Factory settingsDefault calibration settings preset at time of factory
calibration.
FirmwareFirmware is the name given to the analyzer’s inter-
nal software. This can be programmed using the
re-fl ash command via the optional software. The
latest software release can be downloaded from
the Geotech website. It is also automatically updated when the analyzer is returned for servicing.
General warningsDisplayed throughout the documentation with a
warning symbol. Warning information may affect
the safety of users.
Inlet portPort located on the top of the analyzer to which the
inlet tube is attached.
LCD displayLiquid Crystal Display. Fitted to the front panel of
the analyzer.
Main read screenThe main analyzer screen for normal operations
and all operations are carried out from this screen.
MemoryTha analyzer memory should not be used as a
permanent storage medium and data collected
should be transferred using the optional PC
software.
Moisture removal tubeDevice used to remove water vapor from the
sample gas. This is located at the back of the
analyzer and should not be covered or partially
covered. This is not user serviceable and should
NOT be removed by the user.
O
2
Oxygen gas typically 20.9% in air, measured via
electro-chemical cell.
Over range codesOver range codes are errors above the maximum
allowed reading and will be displayed with more
than chevrons (>>.>).
PPMParts per million
PredeterminedSet up prior to use via the optional PC software.
Instruction 19-933952
Page 53
CO2 Analyzers 3100 & 3150
PumpThe device used to transfer a gas sample into the
instrument where that sample is not under pressure. Select the “PUMP” key on the analyzer to
activate.
Regulator fl owThe regulator’s fl ow is factory set. It only requires
a few turns to open, no adjustment is available.
Sample tubeTube used to transfer sample gas from the source
to the analyzer.
SpanThe point at which the gas analyzer is calibrated
when a known concentration of the target gas is
present.
Temperature probeExternal device to enable the instrument to display
data and record the temperature readings. This is
an optional feature.
Under range codesUnder range codes are errors below zero and will
be displayed with less than chevrons (<<.<).
User calibrationUsers have the facility to calibrate the analyzer
between services. User calibration of the gas analyzer will improve the data accuracy in the range of
the calibration gases used.
Warm-up self testPredetermined self-test sequence to test the
analyzer functions which takes place after the
analyzer is switched on and lasts approximately
30-40 seconds.
WarrantyThe instrument is guaranteed against defect in
materials and workmanship for a period of 12
months from the date of shipment to the user and
is subject to the recommended service and recalibration requirements.
ZeroThe point at which the gas analyzer is calibrated
when there is non of the target gas present.
15.0 Replacement Parts
Item DescriptionPart Number
Battery Charger with Adapters19-0610
Sample Tube Kit19-0613
Temperature Probe (5 mm)19-0611
Sample Filters (Pack of 5)19-0612
PC Software19-0614
Calibration Kit19-8027
Cylinder of N
Cylinder of 5% CO
Gas23-4003
2
(for Model 3100)24-1126
2
Cylinder of 2,500 ppm CO2 (for Model 3150)24-1130
Instruction 19-933953
Page 54
16.0 Service Centers
United States
Bacharach, Inc.
621 Hunt Valley Circle
New Kensington, PA 15068
Phone: 724-334-5051
Toll Free: 1-800-736-4666
Fax: 724-334-5723
Email: help@mybacharach.com
Canada
Bacharach of Canada, Inc.
250 Shields Court Unit #3
Markham, Ontario L3R 9W7 Canada
Phone: 905-470-8985
Fax: 905-470-8963
Email: bachcan@idirect.com
CO2 Analyzers 3100 & 3150
México
Bacharach de México
Playa Regatas No. 473 Tercer Piso
Col. Militar Marte
Delegación Iztacalco, 08830
México D.F. México
Phones:+52-555-634-7740
+52-555-634-7741
FAX: +52-555-634-7738
Email: bacharach@prodigy.net.mx