Bacharach 3100, 3150 Operation & Maintenance Instructions Manual

Page 1
CO2 Analyzer Models 3100 & 3150
Operation & Maintenance
Instruction 0019-9339
Rev. 3 – May 2010
Page 2
CO2 Analyzers 3100 & 3150
WARRANTY
Bacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in mater ial and manufac ture and will conform substan tially to Bach arach Inc.’s ap plicable spec i fi ca tions. Bacharach’s li a bil i ty and Buyer’s remedy under this warranty are lim it ed to the repair or replacement, at Ba charach’s option, of this Product or parts thereof returned to Seller at the factory of manufac ture and shown to Bacharach Inc.’s reasonable satisfaction to have been defective; provided that written notice of the defect shall have been given by Buyer to Bacharach Inc. within one (1) year after the date of delivery of this Product by Bacharach, Inc.
Bacharach, Inc. warrants to Buyer that it will convey good title to this Product. Bacharach’s liability and Buyer’s remedy under this warranty of title are limited to the removal of any title defects or, at the elec­tion of Bacha rach, to the replacement of this Product or parts thereof that are defective in title.
THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN AND ACCEPTED IN LIEU OF (I) ANY AND ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, IN CLUD ING WITH OUT LIMITA­TION THE IMPLIED WARRANTIES OF MERCHAN TABILITY AND FITNESS FOR A PARTICULAR PURPOSE: AND (II) ANY OBLIGA TION, LIABILITY, RIGHT, CLAIM OR REM E DY IN CON TRACT OR TORT, WHETHER OR NOT ARISING FROM BACH A RACH’S NEGLIGENCE, ACTUAL OR IM PLIED. The remedies of the Buyer shall be limited to those provided herein to the exclusion of any and all other remedies includ ing, without limita tion incidental or consequen tial dam ag es. No agree ment varying or ex tend ing the foregoing warranties, remedies or this limitation will be binding upon Bacharach, Inc. un­less in writing, signed by a duly au tho rized officer of Bacharach.
Register Your Warranty by Visiting
www.mybacharach.com
Notice:
Product improvements and enhancements are continuous, therefore the spec i fi ca tions and in for ma tion contained in this document may change without notice.
Bacharach, Inc. shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of Bacharach, Inc.
Copyright © 2009, Bacharach, Inc., all rights reserved. BACHARACH is a registered trademark of Bacharach, Inc. All other trademarks, trade names, service marks and logos referenced herein belong to their respective companies.
Page 3
NOTES:
CO2 Analyzers 3100 & 3150
Page 4
CO2 Analyzers 3100 & 3150
1.0 Manual Guidelines .........................................................................6
1.1 Safety Related Information ...................................................... 6
1.2 Notes ....................................................................................... 6
2.0 Introduction ....................................................................................7
2.1 Features of the CO
Analyzer 3100/3150 ................................ 7
2
2.2 Instrument Components - Standard Product ........................... 8
3.0 Optional Products and Accessories ............................................ 9
3.1 Optional Products .................................................................... 9
3.1.1 PC Software (Optional) ................................................ 9
3.1.1.1 Event Log ......................................................... 9
3.1.2 Temperature Probe Reading (Optional) ..................... 10
3.1.3 Humidity Probe Reading (Optional) ........................... 10
4.0 Instrument Features .................................................................... 11
4.1 Physical Characteristics of the Instrument Panel ...................11
4.2 Panel Key Functions ............................................................. 12
4.3 Instrument Connection Points ............................................... 13
5.0 General Operational Instructions............................................... 14
5.1 Switching the Instrument ON ................................................ 14
5.2 Switching the Instrument OFF ............................................... 14
5.3 Instrument Main Read Screen .............................................. 15
5.4 Instrument Status Icons ........................................................ 16
5.5 Entering Data ........................................................................ 16
5.6 Changing Between Parameters ............................................ 17
5.7 Memory ................................................................................. 17
5.8 Storage .................................................................................. 17
5.9 Main Menu ............................................................................ 18
5.9.1 Information ................................................................. 18
5.9.2 Utilities ....................................................................... 19
5.9.2.1 Time & Date ................................................... 19
5.9.2.2 Contrast ......................................................... 20
5.9.2.3 Alarms ............................................................ 21
5.9.2.4 Settings .......................................................... 22
5.9.2.5 Flow Fail ......................................................... 22
5.9.2.6 Logging .......................................................... 25
5.9.2.7 Reset .............................................................. 26
5.9.3 Calibration ................................................................. 26
5.9.4 View Data .................................................................. 26
5.9.4.1 Clear Reading Memory .................................. 27
5.9.5 Diagnostics ................................................................ 28
5.10 Warning and Error Codes .................................................... 28
5.11 Battery / Charging................................................................ 28
6.0 Taking Readings .......................................................................... 30
6.1 Preliminary Checks - Best Practice ....................................... 30
6.2 Gas Measurement Process - Best Practice .......................... 30
6.3 Alternative Reading Methods ................................................ 32
Page 5
CO2 Analyzers 3100 & 3150
6.3.1 Logged Reading ........................................................ 32
6.3.2 Peak Reading ............................................................ 32
6.3.3 Hold Reading ............................................................. 32
7.0 Calibration ................................................................................... 33
7.1 User Calibration .................................................................... 33
7.2 Calibration Gases .................................................................. 34
7.3 Calibration Set-Up ................................................................. 34
7.4 Calibration Equipment ........................................................... 35
7.5 Calibration Method ................................................................ 36
7.5.1 Zero CO
7.5.2 Span CO
7.5.3 Zero O
7.5.4 Span O
Channel ..................................................... 36
2
Channel .................................................... 37
2
Channel ........................................................ 38
2
Channel ....................................................... 38
2
7.5.5 Reset Factory Settings .............................................. 39
7.6 Last Field Calibration ............................................................ 39
7.7 Calibration Record ................................................................ 39
8.0 Problem Solving .......................................................................... 41
8.1 Warnings and Errors ............................................................. 41
8.2 Self-test Warning Messages ................................................. 42
8.3 User Calibration Trouble Shooting ........................................ 44
8.4 Cross-Gas Effects ................................................................. 44
8.5 Switching Off the Instrument ................................................. 44
9.0 Service .......................................................................................... 45
10.0 Warranty Policy ......................................................................... 46
11.0 Technical Specifi cation ............................................................. 47
12.0 Event Log ................................................................................... 49
13.0 Emportant Notice to All Customers ......................................... 50
14.0 Glossary of Terms ..................................................................... 51
15.0 Replacement Parts .................................................................... 53
15.0 Service Centers ......................................................................... 54
Page 6
CO2 Analyzers 3100 & 3150
1.0 Manual Guidelines
1.1 Safety and Related Information
Information in this manual that may affect the safety of the users and others is preceeded by the following symbol:
WARNING
Failure to follow this information may result in physical injury which in some cases could be fatal.
1.2 Notes
Important / useful information and instructions are shown clearly throughout the manual in a note format.
For example:
NOTE: For further information, please contact your equipment
supplier.
NOTE: Actual screenshots will vary dependent on model.
Page 7
CO2 Analyzers 3100 & 3150
2.0 Introduction
This manual explains how to use the instruments below:
• CO
CO
CO
CO
• CO
CO
CO
CO
Analyzer 3100 (0 - 20%)
2
P/N 19-7076
2
(with O2) P/N 19-7077
2
(with O2 and Relative Humidity) P/N 19-7078
2
Analyzer 3150 (0 - 10,000 PPM)
2
P/N 19-7136
2
(with O2) P/N 19-7137
2
(with O2 and Relative Humidity) P/N 19-7138
2
NOTE: The instrument is a sensitive piece of scientifi c equipment,
and should be treated as such.
The CO2 analyzer is specifi cally designed to monitor CO2 for the verifi cation of incubators in research and pharmaceutical markets. This unit has been developed to incorporate the latest technology and specifi cation requirements, which provide the user with a fast, simple- to-use and accurate piece of laboratory instrumentation.
2.1 Features of the CO2 Analyzer
The analyzer has the following features:
• CO
• CO
0 to 20% (3100)
2
0 to 10,000 PPM (3150)
2
• Options for:
O
0 to 100%
2
• Dual Temperature probes 32 to 122°F (0 to 50 °C)
• Humidity Sensor 0 to 100%
• Improved accuracy on CO
readings
2
• Time saving dual temperature probes
• Large data storage and user friendly software and download
• Easy-to-read, large well-lit display
• Built-in gas moisture removal
Applications
IVF/Medical Research
• Laboratories • IAQ (Indoor Air Quality)
Brewing Atmosphere Control
Incubators Gas mixing
Page 8
CO2 Analyzers 3100 & 3150
2.2 Instrument Components
The following items are included in the Analyzer Kit. Please check to make sure all items have been included.
Analyzer (One of the following - refer to your model number):
CO
CO
CO
CO
CO
CO
Analyzer 3100 0019-7076
2
/ O2 Analyzer 3100 0019-7077
2
/ O2 Analyzer 3100 w/RH 0019-7078
2
Analyzer 3150 0019-7136
2
/ O2 Analyzer 3150 0019-7137
2
/ O2 Analyzer 3150 w/RH 0019-7138
2
Instruction Manual (on CD) 0019-9339
Battery Charger with adapters 0019-0610
Sample Tube Kit (includes sample fi lter and sample tube) 0019-0613
Quick Start Guide 0019-9338
Page 9
CO2 Analyzers 3100 & 3150
3.0 CO2 Analyzer Accessories
3.1 Optional Products
The CO2 Analyzer has a number of optional products for purchase which enhance the usability and enable further analysis of data and reading information.
NOTE: For more information on the features listed in this section
please contact your nearest Bacharach Service Center.
3.1.1 PC Software (Optional)
The software enables the user to maximize the operation of the incubator analyzer. Instrument readings and event log data may be uploaded to a PC for further analysis and exported to other applications such as MS Excel. It enables direct communication with the unit, features a simple download facility and is fully compatible with
the latest Microsoft operating systems (32-bit only). A USB cable is
provided with the software.
3.1.1.1 Event Log
The instrument incorporates the ability to log signifi cant events via the “Event Log.” This can be used as an aid to monitoring the use of the instrument. It can also be used as a diagnostic tool if there is a problem with the instrument.
The event log can only be viewed via the optional software. It cannot be viewed on the analyzer screen. Applicable events are stored in the event log automatiaclly, meaning no user intervention is required.
The event log can hold approximately 270 events. If the log becomes full then it begins to overwrite the older events. This can be identifi ed by the index fi eld which starts from event number 1. The log is cleared when the instrument is reset.
NOTE: Please refer to Section 12.0 Event Log of this operating
manual for further information.
Page 10
CO2 Analyzers 3100 & 3150
3.1.2 Temperature Probe Reading (Optional)
The instrument has the ability to read and display two temperature readings via optional temperature probes (P/N 0019-0611). When a temperature probe is fi tted to one of the temperature ports on the top of the instrument, the display will automatically show the current reading and a temperature probe icon will be displayed.
The display can also be changed to show a T
- T2 calculation by
1
pressing the appropriate soft-key. The current mode of operation can be identifi ed by normal or inverse status of the soft-key, where inverse indicates that the option is active.
NOTE: The T
- T2 result is not stored as part of the reading.
1
The operator can also choose to display the reading in either Centrigrade or Fahrenheit using the temperature option accessed from
the Settings Menu.
Instruction 19-9339 10
Page 11
CO2 Analyzers 3100 & 3150
4.0 Instrument Features
4.1 Physical Characteristics
Front View Back View
Reference:
A = Main Read Screen B = Soft-Keys C = ON/OFF Key D = Pump Key E = Key 4 - Scroll LEFT F = Key 8 - Scroll DOWN G = Menu Key H = Enter Key I = Key 2 - Scroll UP J = Key 6 - Scroll RIGHT K = Moisture Removal Tube L = Serial Number M = Instrument Stand
Instruction 19-9339 11
Page 12
CO2 Analyzers 3100 & 3150
4.2 Panel Key Functions
KEYS FUNCTION
A Main Read Screen Start and end screen when using the instrument.
B Soft Keys The function of the three “soft-keys” on the front of
the instrument panel are determined by menu op­tions taken. Functions vary from screen to screen.
C ON/OFF Key Press the “ON/OFF” key briefl y to switch the
instrument on and off.
D Pump Key Press the “Pump” key to start or stop the pump.
E Scroll LEFT Key Also “Key 4”. Enables the operator to scroll left to
display more information.
F Scroll DOWN Key Also “Key 8.” Enables the operator to scroll down
to display more information.
G Menu Key Press the “Menu” key to go to the MAIN MENU.
Enables the operator to pre-set values and set­tings. Select options from the MAIN MENU to also view data and readings stored or held.
H Enter Key The “Enter” key accepts/confi rms choices made by
the operator to various functions and operations. Also, required to confi rm numeric data entry.
I Scroll UP Key Also “Key 2.” Press scroll up to view further infor-
mation on the instrument read screen.
J Scroll RIGHT Key Also “Key 6.” Press scroll right to view further
information on the instrument read screen.
K Moisture Removal Tube Removes the moisture from the sample gas.
L Serial Number Unique identifi cation for the instrument. Verifi ca-
tion of the serial number will be required if Techni­cal Support assistance is needed.
M Instrument Stand Instrument stand.
NOTE: Do NOT attempt to remove the cover off the back of the
analyzer which houses the moisture removal tube.
NOTE: Do NOT cover the moisture removal tube with your hand
when holding the analyzer to take readings.
Instruction 19-9339 12
Page 13
CO2 Analyzers 3100 & 3150
4.3 Instrument Connections
Top View Side View
Refer­ence
A Temperature 1 Temperature 1 connector measures differ-
B Temperature 2 Temperature 2 connector measures differ-
C Humidity Probe (Optional) Humidity probe attachment point.
D Gas Outlet Gas outlet port used to exhaust the gas.
E Gas Inlet Gas inlet point used to attach the sample
F USB Port Used to connect the analyzer to a PC via a
G Battery Charger Port Used to attach the mains charger to the
Type Description
ential temperature - incubator and ambient temperatures.
ential temperature - incubator and ambient temperatures.
tube and fi lter in order to take the gas read- ing.
USB cable to download data.
analyzer for charging.
NOTE: Depending on the confi guration purchased, certain connectors
may not be present. Reference you model number for this information.
Instruction 19-9339 13
Page 14
CO2 Analyzers 3100 & 3150
5.0 General Operational Instructions
5.1 Switching the Instrument ON
1) To switch on the instrument, press the “ON/OFF” key briefl y. There will be a short beep and a slight pause followed by the Bacharach logo.
2) The power on self-test will then commence (approximately 60 seconds).
3) If there are no warnings to display the instrument will continue to the MAIN READ SCREEN.
5.2 Switching the Instrument OFF
1) Purge the analyzer with fresh air. Run the pump for approximately 30 seconds or until the readings have returned to normal levels.
NOTE: Before the instrument is switched off, a clean air purge should
be performed. This ensures that the instrument is free from gas and ready for the next measurement. This fi nal purge is especially important for the oxygen sensor as it may degrade if stored when contaminated with gas.
2) To switch off the instrument, press the “ON/OFF” key briefl y.
NOTE: If the “AUTO OFF” setting is set to “YES”, the analyzer will
switch off automatically after 10 minutes if not in use.
Instruction 19-9339 14
Page 15
CO2 Analyzers 3100 & 3150
5.3 Instrument Main Read Screen
After the analyzer has been switched on and the warm-up self-tests have completed, the analyzer will display the “MAIN READ SCREEN.”
CO2 Analyzer 3100 w/ O2 shown
Soft-Keys:
Peak - Enables the operator to display the peak reading. Store - Enables the operator to store the reading for viewing/download
later.
Hold - Enables the operator to hold the current reading being taken.
Instruction 19-9339 15
Page 16
CO2 Analyzers 3100 & 3150
s
5.4 Instrument Status Icons
The following icons may be displayed on the instrument read screens:
Icon Description
Battery Charge State
(fl ashing) < 1 Hour Remaining
(fl ashing) Battery Charging
Charged
Pump Running
(fl ashing) Pump Stalled
Alarm Set
(fl ashing) Alarm Active
USB Connected to PC (fl ickers when transferring data)
Logging Mode Active (fl ashes when memory nearly full)
Temperature Probe(s) Connected
Humidity Probe Connected
( a
( ashing) Service Due (every 12 months)
Service Overdue
Fault / Repair (i.e. O
Waiting
cell, battery, or pump may need replacing)
2
5.5 Entering Data
During normal operation the user may be prompted to enter data or information via the keypad, i.e. entering an ID code or setting an alarm level.
When entering data into the instrument, all fi elds are fi xed format and are populated from the right. For example, to enter the new time, 09:25:00, the user would type in 092500 using the numeric keypad in the following sequence:
• : : 0
• : :09
• : 0:92
• :09:25
• 0:92:50
• 09:25:00
Press the “ENTER” key to confi rm/accept the data.
Instruction 19-9339 16
Page 17
CO2 Analyzers 3100 & 3150
The soft-key “DELETE” will delete the last character typed. Alternatively, the sequence can be re-typed before the “ENTER” key is pressed and the existing numbers will be replaced by the new sequence.
NOTE: The instrument will not allow invalid data to be entered; this
should be deleted and re-entered.
5.6 Changing Between Parameters
By default, the instrument displays the MAIN READ SCREEN (for gas measurement). This shows the CO readings (O
, %RH, etc). The instrument will return to this screen after
2
power on or when returning from the menus. The “SCROLL” keys can be used to switch to another measurement screen, i.e. temperature or humidity. Continue to press the “SCROLL” key to return to the MAIN READ SCREEN.
reading along with the optional
2
5.7 Memory
The memory SHOULD NOT be used as a permanent storage medium and any important data should be transferred to a more permanent storage medium as soon as possible. The instrument should not be stored for prolonged periods with valuable data in its memory.
5.8 Storage
When not in use, the instrument should be kept in a clean, dry, and warm environment such as an offi ce. It should be stored fl at in order to help prolong the life of the O
cell.
2
Instruction 19-9339 17
Page 18
CO2 Analyzers 3100 & 3150
5.9 Main Menu
The “MAIN MENU” enables the operator to select options to set up specifi c parameters and perform operational tasks prior to sample readings being taken or to view data/information stored in the instrument.
1) Press the “MENU” key on the front of the instrument panel and the
following screen will be displayed:
2) Press the soft-key “EXIT” to exit the MAIN MENU.
5.9.1 Information
The “INFORMATION” option enables the operator to display information such as instrument type, serial number, current software version, service due date, and the dates of the last factory and user calibrations. Note that the dates are in DD/MM/YY format.
1) From the “MAIN READ SCREEN” press the “MENU” key on the
instrument panel.
2) Press “KEY 1” to display general information about the instrument.
Instruction 19-9339 18
Page 19
CO2 Analyzers 3100 & 3150
5.9.2 Utilities
The “UTILITIES” option enables the operator to confi gure the instrument settings prior to taking readings.
1) From the “MAIN READ SCREEN” press the “MENU” key on the
instrument panel.
2) Press “KEY 2” to display the “UTILITIES” menu and the following
screen will be displayed:
5.9.2.1 Time & Date
The “TIME AND DATE” option enables the operator to check or set the instrument’s internal clock. The current time/date will appear on every stored reading.
1) From the “MAIN READ SCREEN” press the “MENU” key on the
instrument panel.
2) Press “KEY 2” to display the “UTILITIES” menu.
3) Press “KEY 1 - TIME & DATE” and the “SET TIME & DATE” menu
is displayed as follows:
Instruction 19-9339 19
Page 20
CO2 Analyzers 3100 & 3150
4) Press “KEY 1” to change the time or press “KEY 2” to change the
date. Type the time or date using the numeric keypad followed by the “ENTER” key. The instrument will not allow invalid times or dates to be entered.
NOTE: The clock will need to be manually adjusted to cope with
daylight saving changes or changes when crossing time zones.
5.9.2.2 Contrast
The “CONTRAST” option enables the operator to adjust the instrument screen contrast to compensate for changes in ambient temperature. The default setting is 0.
1) From the “MAIN READ SCREEN” press the “MENU” key on the
instrument panel.
2) Press “KEY 2” to display the “UTILITIES” menu.
3) Press “KEY 2” to select the instrument panel contrast settings and
the following screen will be displayed:
4) Press “KEY 4 - SCROLL LEFT” and “KEY 6 - SCROLL RIGHT” to
adjust the value displayed.
5) Press the soft-key “ACCEPT” or “REJECT” accordingly to accept or
reject the changes.
NOTE: The manually-set contrast setting is retained when the
instrument is switched OFF.
Instruction 19-9339 20
Page 21
CO2 Analyzers 3100 & 3150
5.9.2.3 Alarms
The analyzer has the ability to set rising or falling alarms for the two main gas channels, CO
and O2 (if selected as an option). The alarms
2
for each channel can be enabled or disabled independently via the “ALARMS” menu option.
Once enabled these alarms become active in the “MAIN READ SCREEN”; this is indicated by a bell icon. If an alarm is triggered the screen turns red and a fl ashing bell icon is displayed. The beeper is sounded until the gas level has recovered beyond the trigger point.
• Rising alarms are triggered when the gas level exceeds the
maximum value entered by the user.
• Falling alarms are triggered when the gas level falls below the
minimum value entered by the user.
1) From the “MAIN READ SCREEN” press the “MENU” key on the
instrument panel.
2) Press “KEY 2” to display the “UTILITIES” menu.
3) Press “KEY 3” to select alarm settings and the following screen will
be displayed:
4) Press “KEY 1” to select the CO
alarm and “KEY 2” to select the O2
2
alarm.
5) Press “KEY 1” to enable/disable the alarm, “KEY 2” to set a
maximum value, or “KEY 3” to set a minimum value.
6) Press ENTER to store the selected choice.
Instruction 19-9339 21
Page 22
CO2 Analyzers 3100 & 3150
5.9.2.4 Settings
The “SETTINGS” option enables the operator to maintain information with regards to taking samples and readings.
1) From the “MAIN READ SCREEN” press the “MENU” key on the
instrument panel.
2) Press “KEY 2” to display the “UTILITIES” menu.
3) Press “KEY 4” to select settings and the following screen will be
displayed:
4) The following instrument settings man be maintained:
Setting Use
1 Prompt ID Press “KEY 1” to prompt for ID code for each sample reading,
answer YES or NO accordingly.
2 Temperature Press “KEY 2” to enter the default unit temperature, choosing
from °C or °F.
3 Date Press “KEY 3” to switch the date format between DD/MM/YY and
MM/DD/YY formats.
4 Auto Off Press “KEY 4” to auto switch off the instrument when not in use.
Toggle between Auto Off: YES or NO. If set to YES, the instru­ment will switch off after 10 minutes if not in use.
5.9.2.5 Flow Fail
The “FLOW FAIL” option enables the operator to adjust the instrument
ow fail detection point should it fail in normal operation with a clean lter.
The instrument’s internal pump can be stalled when pulling against a vacuum or through a blocked fi lter. This is indicated by a fl ashing pump icon; to prevent damage to the pump, the pump will switch off after a few seconds.
Instruction 19-9339 22
Page 23
CO2 Analyzers 3100 & 3150
Press the ‘Pump’ key again to remove the fl ashing pump icon
NOTE: Dirty or discolored fi lters should be changed before
use. Filters that have drawn in water should be changed immediately to prevent damage to the instrument.
1) From the “MAIN READ SCREEN” press the “MENU” key on the
instrument panel.
2) Press “KEY 2” to display the “UTILITIES” menu.
3) Press “KEY 5” to select fl ow fail and the following screen will be
displayed:
4) Use the Scroll keys to adjust the value displayed. The larger the
value, the less sensitive the fl ow fail detection will be.
5) Press the soft-key “ACCEPT” or “REJECT” accordingly.
Follow the following fl ow chart to determine the proper setting for the Flow Fail Alarm.
Instruction 19-9339 23
Page 24
CO2 Analyzers 3100 & 3150
Instruction 19-9339 24
Page 25
CO2 Analyzers 3100 & 3150
5.9.2.6 Logging
Data logging mode can be started or stopped via “KEY 6 - LOGGING” which is accessed from the “UTILITIES” menu. Press “KEY 4 ­START/STOP LOGGING” to start and stop data logging.
While in data logging mode, the instrument will automatically record data at the pre-set intervals including running the pump for a preset time period. Active logging mode is indicated on the “MAIN READ SCREEN” by its icon. The operator is able to edit the default ID, pump run-time, interval, and start/stop logging.
1) From the “MAIN READ SCREEN” press the “MENU” key on the
instrument panel.
2) Press “KEY 6” to display the “LOGGING” menu and the following
screen will be displayed:
3) Select the desired option by pressing keys “1” through “4”. Then
enter the appropriate setting using the keypad followed by the “ENTER” key.
Setting Use
1 Every 00 mins. Press “KEY 1” to enter the time in minutes for the timeframe
between sample readings. The interval controls the reading frequency in minutes, i.e. every 10 minutes.
2 Pump 00 secs. Press “KEY 2” to enter the time in seconds for the length of
time you wish the pump to run when taking a sample read­ing. The pump run-time is the time in seconds for which the pump runs prior to the reading being stored. This fi gure will also need to take into account the length of sample tube and the volume of the sample gas. For example, there is little point setting a pump run-time of 10 seconds if it takes 30 seconds to draw in a new sample.
3 ID 00000000 Press “KEY 3” to create an 8-digit numeric ID code.
4 Start/Stop Logging Press “KEY 4” to start and stop data logging.
NOTE: Data logging mode is automatically stopped when the
instrument is switched off or the logging parameters are edited.
Instruction 19-9339 25
Page 26
CO2 Analyzers 3100 & 3150
5.9.2.7 Reset
The instrument can be reset by pressing “KEY 7 - RESET” accessed
via the “UTILITIES” menu. Selecting this option will clear all user
settings and any stored data including the event log.
A confi rmation code (12345678) must be entered to confi rm that a reset is initiated.
5.9.3 Calibration
The instrument is fully calibrated during manufacture and when returned for service. However, to improve accuracy between services, a user/fi eld calibration can be performed.
NOTE: For further information please refer to Section 7.0 Calibration.
5.9.4 View Data
The “VIEW DATA” option enables the operator to view the stored readings.
1) From the “MAIN READ SCREEN” press the “MENU” key on the
instrument panel.
2) Press “KEY 4” to view stored data readings and the following
screens will be displayed:
3) Press KEY 4 - SCROLL LEFT” and “KEY 6 - SCROLL RIGHT” to
move through the stored readings either forwards or backwards. Press “KEY 2 - SCROLL UP” and “KEY 8 - SCROLL DOWN” to switch between the fi rst (CO
, O2, & Baro) and seconds (T1, T2, &
2
humidity) group of reading parameters.
Instruction 19-9339 26
Page 27
CO2 Analyzers 3100 & 3150
4) Press the soft-key “MORE” to refi ne or fi lter the readings to view.
Setting Use
1 Delete All Enables the operator to delete all the readings stored, (see below).
2 Filter Used to refi ne/fi lter the range of readings displayed by ID or date
ranges. Press between two dates, after a date, before a date, or all dates.
3 Go To Enables the operator to jump to the fi rst or last reading in the memory
or any other reading.
5.9.4.1 Clear Reading Memory
The “CLEAR READING MEMORY” function enables the user to check how many readings have previously been taken and clear them if necessary.
NOTE: Before readings are actually deleted, a caution message is
displayed; once readings have been deleted they cannot be recovered.
The instrument can store up to 1000 readings. The reading structure is fi xed and may contain optional parameters not activated for your particular instrument confi guration, i.e. oxygen, temperature, and humidity.
Once the reading memory is full it is not possible to store any more readings. When the memory is full and the “STORE” key is pressed or data logging is activated, the instrument will show a brief message stating that the memory is full and that no further data will be recorded.
1) From the “MAIN READ SCREEN” press the “MENU” key on the
instrument panel.
Instruction 19-9339 27
Page 28
CO2 Analyzers 3100 & 3150
2) Press “KEY 4” to view data.
3) To clear the readings press the soft-key “MORE” followed by “KEY
1 - DELETE ALL.”
5.9.5 Diagnostics
The “DIAGNOSTICS” option enables remote technical support to identify and resolve issues with the instrument and readings. If required, the operator may be asked to confi rm the diagnostics displayed.
1) From the “MAIN READ SCREEN” press the “MENU” key.
2) Press “KEY 5” to view diagnostics.
NOTE: For further information contact your nearest Bacharach
Service Center.
5.10 Warning and Error Codes
When switched on, the instrument will perform a predetermined self-test sequence taking approximately 60 seconds. During this time, many of the instrument’s working parameters and settings are checked. If any operational parameters are out of specifi cation or if the pre-programmed recommended calibration / service date has passed, errors or warnings may be displayed.
NOTE: For further information please refer to Section 8.0 Problem
Solving.
5.11 Battery / Charging
The battery used in the instrument is a 2 Ah Lithium-Ion cell. The instrument must be charged using the power supply supplied with your instrument.
NOTE: Although the instrument can be powered via the USB
connecter, it cannot be charged via USB.
Instruction 19-9339 28
Page 29
CO2 Analyzers 3100 & 3150
When plugged into the charger, the instrument light will power on and display charging. When complete, the display will change to show that the instrument is charged. To switch the instrument on while charging or charged is displayed, the operator will need to switch the instrument off and then on again.
NOTE: A full charge will take approximately 3 hours. Typically, a fully
charged battery will last 8-10 hours.
When the instrument is already powered on the battery icon changes to a fl ashing plug symbol. This will stop fl ashing when the charge is complete.
Instruction 19-9339 29
Page 30
CO2 Analyzers 3100 & 3150
6.0 Taking Readings
6.1 Preliminary Checks - Best Practice
Prior to use, it is good practice to ensure that:
Step 1 The instrument has the correct time and date set. Step 2 The sample lter is tted and that it is clean and dry. Step 3 The battery has a good charge (minimum 25% charge, even if only a few
readings are required).
Step 4 The memory has suf cient space available. Step 5 The main gases have been auto-zeroed, without gas concentration present. Step 6 If necessary, check the span calibration with a known concentration calibra-
tion-check gas.
Step 7 Take readings.
• DO protect the instrument from strong direct sunlight which could
raise the temperature of the instrument beyond its operating range.
• DO remember to always use the sample fi lter! If the sample fi lter
becomes fl ooded, change it and ensure all sample tubes are clean and dry before re-use.
• DO NOT place the instrument against anything hot as this may
cause excessive internal temperatures which can lead to erroneous readings.
• DO NOT get the instrument wet.
6.2 Gas Measurement Process
Depending on preferences, the exact reading procedure can change. The following method is considered best practice and when followed correctly will allow quick and consistent readings to be recorded.
1) When the instrument is fi rst switched on it should be purged with
fresh air and allowed to stabilize for a few minutes.
2) At this point the user must zero the CO
available via the “CALIBRATION” menu. The instrument is now ready to take the fi rst reading.
Instruction 19-9339 30
channel. This option is
2
Page 31
CO2 Analyzers 3100 & 3150
3) Connect the sample tube, if relevent (always use the sample fi lter)
from the sample point to the inlet port of the instrument, ensuring that the fi lter is seated correctly.
4) Press the “PUMP” key to draw a sample into the instrument. Notice
the main gas readings start to change. It is recommended to run the pump until the gas readings have stabilized (approximately 30 seconds) then press the soft-key “STORE” if desired.
5) The pump will stop and the operator will be prompted to enter an
ID code to identify the reading. Press the “ENTER” button. A “READING STORED” confi rmation message will be displayed briefl y before returning to the “MAIN READ SCREEN.”
NOTE: The ID code prompt can be switched on or off. This option
is accessed via the “SETTINGS” menu, then press “KEY 1
- PROMPT FOR ID: YES OR NO.” If the ID prompt is set to “NO” the reading is stored using the default logging ID.
6) After each reading the instrument should be purged with fresh air.
7) Disconnect the sample tube from the instrument. Turn the pump
ON. Then run the pump for a minimum of 30 seconds. The gas readings should return to nominal values for fresh air.
Regardless of the instrument confi guration, the following data will be stored for each reading:
• ID code (8characters)
• Reading Type (0=User, 1=Auto, 2=Peak, 3=Hold)
• Current time / date
• Gas readings (CO
, O2)
2
• Sample Pressure (for indication only)
• Temperature (Optional)
• Humidity (Optional)
Instruction 19-9339 31
Page 32
CO2 Analyzers 3100 & 3150
6.3 Alternative Reading Methods
There are three other reading types or methods which each require slightly different operating procedures.
• Logged reading
• Peak reading
• Hold reading
6.3.1 Logged Reading
Logged readings need to be confi gured and initiated via the “UTILITIES” menu by pressing “KEY 6.” During confi guration, the user will be asked to supply an ID, reading interval, and pump run-time.
These parameters are used to control the reading frequency in logging mode. Once logging mode is activated, the instrument will automatically record a reading at every interval until stopped by the operator or the memory becomes full. Logging is also suspended temporarily while the user is accessing the menu options.
While the logging mode is active, both the “PUMP” and “STORE” key will be deactivated. Only logged readings can be stored.
6.3.2 Peak Reading
The operator can toggle the reading mode between normal (current) and peak readings. While in peak reading mode, the instrument will only display peak values for each of the channels. These values can then be stored by pressing the “STORE” key or automatically at the appropriate logging interval (if logging is enabled).
The peak value is reset after a reading is stored or by exiting the peak mode using the appropriate soft-key. The current mode of operation can be identifi ed by the status of the soft-key, either “NORMAL” or “INVERSE”, where inverse indicates peak mode is active.
6.3.3 Hold Reading
The “HOLD READING” option allows the user to freeze the currently displayed reading. This allows it to be manually recorded or moved away from the sample point. Once activated, press the soft-key “HOLD” and the readings are fi xed until the “HOLD” key is pressed again or by storing the reading.
Instruction 19-9339 32
Page 33
CO2 Analyzers 3100 & 3150
7.0 Calibration
7.1 User Calibration
The instrument is fully calibrated during manufacture and when returned for service. However, to improve accuracy between services, a user / fi eld calibration can be performed.
User calibration is a means of optimizing the performance of the instrument to the current operating conditions, such as ambient temperature and pressure, as well as correcting for instrument drift caused by lamp and fi lter settling.
In general, the instrument should not require calibration more than once a month, but we do recommend verifying the instrument’s operation each day
User calibration has two operations and each may be performed individually, however, for a complete user calibration both must be completed.
This section describes the correct procedures to achieve an accurate user calibration.
NOTE: If the calibration is completed incorrectly it may decrease the
accuracy of the instrument.
Two important terms that are used within this section are “Zero” and “Span.”
Zero: The point at which the instrument is calibrated when there
is no gas present.
Span: The point at which the instrument is calibrated when a
known quantity of the target gas is present.
Instruction 19-9339 33
Page 34
CO2 Analyzers 3100 & 3150
7.2 Calibration Gases
User calibration of the instrument will improve the data accuracy in the range of the calibrated gases used. However, it may cause less accurate readings of concentrations outside this calibrated range. Users should select the correct calibration gas for the expected gas levels on their particular application. Only use gases with a known certifi ed gas concentration.
NOTE: Certifi ed calibration gases can be supplied by Bacharach.
Warning: For each gas used the appropriate material safety
data sheet must be read and understood before proceeding. Calibration gases and the use of pressure regulators can be dangerous.
7.3 Calibration Set-Up
The regulator supplied with the calibration kit has been confi gured to deliver a fi xed fl ow. It only requires a few turns to open and no adjustment is necessary.
Warning: When the instrument is being calibrated there are
two possible exits for the gas; via the usual manner out of the exhaust port of the instrument or via the port on the pressure relief valve. It is recommended that both ports have exhaust tubing attached. The exhaust tubing must emerge in a well­vented area. Ensure that there are no leaks in the tubing connections. The calibration should always be carried out in a safe area with all necessary precautions taken as all pressurized gases are potentially dangerous.
Instruction 19-9339 34
Page 35
CO2 Analyzers 3100 & 3150
7.4 Calibration Equipment
The diagram below displays the regulator and tubing equipment for user calibration:
Calibration Kits are available through Bacharach Inc.
Calibration Kit: P/N 19-8027 Cylinder of N Cylinder of 5% CO Cylinder of 2500 ppm CO
Gas: P/N 23-4003
2
Gas: P/N 24-1126 (Model 3100)
2
Gas: P/N 24-1130 ( Model 3150)
2
• Contact your nearest Bacharach Service Center for further information.
• The regulator supplied with the calibration kit is recommended as ow and pressure rates are factory set.
• A relief valve is not supplied with the kit but is recommended in addition to the supplied regulator for improved fl ow.
Instruction 19-9339 35
Page 36
CO2 Analyzers 3100 & 3150
7.5 Calibration Method
Before you begin, ensure that the unit is stabilized at its working temperature before performing any of the calibration operations.
To achieve the processes set out in this section, press “KEY 3 ­CALIBRATION” from the “MAIN” menu. The fi rst screen displayed provides the option to select the gas that requires calibration.
The exact calibration method can vary depending on the gases used.
7.5.1 Zero CO
For maximum accuracy it is recommended that the CO zeroed using bottled gas (certifi ed 100% N
Channel
2
2
).
channel is
2
If nitrogen is not available the user can select the option to perform an air calibration. This option assumes that the user has access to fresh air at around 390 ppm CO
. Generally, this can be found outside or in
2
a well ventilated corridor (typically, an offi ce lab would have a higher CO
concentration).
2
To zero the CO
1) From the “CALIBRATION” menu press “KEY 2 - CO
channel, perfom the following:
2
CHANNEL.”
2
Instruction 19-9339 36
Page 37
CO2 Analyzers 3100 & 3150
2) Press either “KEY 1 - ZERO WITH N2” (recommended) or “KEY 2 ­ZERO WITH AIR” from the menu. Then, either attach the 100% N
2
or sample pipe to allow access to fresh air.
3) Press the “START” key. The instrument will now wait approximately 60 seconds for the gas reading to stabilize at the correct level. Ensure the pump is on (pump icon appears in lower middle of screen when on).
4) The instrument will then indicate a successful zero has been completed. Press the soft-key “ACCEPT” to confi rm the calibration and “STORE” the new user offset. Alternatively, select the soft-key “REJECT” to exit without change.
NOTE: Typical zero calibration, although displaying zero, needs to be
given time to settle. We recommend accepting the calibration
at least fi ve minutes after the display concentration stabilizes.
NOTE: If the calibration failed, purge and try again or select a
different air source.
NOTE: If using the Analyzer 3150 - very high concentrations of CO
may take up to 30 minutes to purge completely.
7.5.2 Span CO
Channel
2
It is recommended that the instrument is spanned to target the desired
reading range (e.g. 5%); ideally this should not be a low level close to
zero.
1) If not already preset, enter the span target (i.e. certifi ed concentration of your calibration gas) by pressing “KEY 1” and then entering the new desired value followed by ENTER. Then attach the gas and open the regulator valve to allow the gas to fl ow.
2) Press the “START” key and wait for the reading to stabilize. This can take a couple of minutes. Ensure the pump is on (pump icon appears in lower middle of screen when on).
2
Instruction 19-9339 37
Page 38
CO2 Analyzers 3100 & 3150
3) Once a stable reading is shown, press the soft-key “ACCEPT.” A successful span calibration message will then be displayed. Press the soft-key “ACCEPT” again to confi rm the calibration and “STORE” the new user span. Alternatively, press soft-key “REJECT” to exit without change.
NOTE: If the calibration failed then try again using a longer purge
time or different target gas.
7.5.3 Zero O
It is not necessary to zero the O
Channel
2
channel. A span calibration should
2
correct the reading across the entire range.
7.5.4 Span O
It is recommended that the O
Channel
2
channel is spanned in fresh air with
2
a target concentration of 20.9%, although other gases and target concentrations can be used if required.
1) If not already preset, enter the span target (i.e. certifi ed concentration of your calibration gas) by pressing “KEY 1” followed by the desired value. Press ENTER to store.
2) Press the soft-key “START” and wait for the reading to stabilize. Press the “PUMP” key to draw in fresh air. It can take a couple of minutes to stabilize.
3) Once a stable reading is shown press the soft-key “ACCEPT.” A successful span calibration message should then be displayed. Press the soft-key “ACCEPT” again to confi rm the calibration and “STORE” the new user span. Alternatively, press the soft-key “REJECT” to exit without change.
NOTE: If the calibration failed then try again using a longer purge
time or different target gas.
Instruction 19-9339 38
Page 39
CO2 Analyzers 3100 & 3150
7.5.5 Reset Factory Settings
This option will reset the instrument to its factory programmed calibration characteristics and will clear the user calibration points for both gas channels.
1) To reset to factory settings, press “KEY 1 - FACTORY RESET” from the “CALIBRATION” menu.
2) To prevent the user calibration data being accidently erased the user must confi rm the action by pressing the soft-key “ACCEPT” or soft-key “REJECT” to exit without change.
7.6 Last Field Calibration
This data can be found in the “INFORMATION” screen accessed via the “UTILITIES” menu. This option displays the date that the last fi eld calibration was performed on the instrument.
7.7 Calibration Record
The CO2 instruments have the facility to log user calibrations via the “EVENT LOG.” This can be used as an aid in ensuring that gas measurements are valid and accurate.
During calibration the instrument will record the following in the event log. For each entry the time and date will be stored.
Instruction 19-9339 39
Page 40
Event Data Recorded
CO2 Analyzers 3100 & 3150
Successful User Zero CO
Successful User Span CO
Successful User Span O
Failed User Zero CO
Failed User Span CO
Failed User Span O
2
2
2
2
2
2
Type (N2 or Air) and Readings Before and After
Target Value, Readings Before and After
Target Value, Readings Before and After
Type (N2 or Air) and Reading
Target Value, Gas Reading
Target Value, Gas Reading
Return to Factory Settings
NOTE: If the calibration failed, then try again using a longer purge
time or different target gas. This event log can only be downloaded and viewed via the optional software. It cannot be viewed on the analyzer screen.
Instruction 19-9339 40
Page 41
CO2 Analyzers 3100 & 3150
8.0 Problem Solving
This section outlines various warning and error messages which the operator may receive during general operation of the instrument. For further assistance please contact your nearest Bacharach Service Center.
8.1 Warnings and Errors
When switched on, the instrument will perform a predetermined self-test sequence taking approximately 60 seconds. During this time many of the instrument’s working parameters and settings are checked.
If any operational parameters are out of specifi cation or if the pre- programmed recommended calibration / service date has passed, errors or warnings may be displayed.
Use the “SCROLL UP” and “SCROLL DOWN” keys to move through the list if required.
There are two types of warning that may be displayed:
• General warnings that may not affect the instrument’s function and
those where the self-test has detected a function that is outside of the usual operating criteria, e.g. battery charge low, memory nearly full.
• Operational parameters that could affect the performance of the
instrument, e.g. CO2 out of calibration.
The most likely reason for these errors is either an incorrect user calibration, or may indicate sensor failure. If an incorrect user calibration has caused the warning it should be correctable by the way of returning the instrument to factory settings, zeroing or carrying out a user calibration as necessary for the relevant function.
Instruction 19-9339 41
Page 42
CO2 Analyzers 3100 & 3150
Under and Over Range Codes
If a reading is over range (i.e. above the maximum allowed reading) it will be displayed with ‘more than’ chevrons (>>.>). This can occur if a channel has been incorrectly calibrated or the sample gas has exceeded its specifi ed range (e.g. CO
>20%).
2
If a reading is under range (i.e. below zero) it will be displayed with ‘less than’ chevrons (<<.<). Refer to Section 7.0 Calibration to remedy under-range by performing a user zero.
A number displayed as asteriks (**.*) indicates an error, usually where the instrument has been unable to complete a particular calculation. Typically, this will be the fi rst indication of a fault condition.
Where no data is available, dashes (--.-) are displayed. This usually occurs when a particular reading or parameter has been skipped by the user, or where an optional accessory is not fi tted correctly, i.e. a temperature probe.
8.2 Self-Test Warning Messages
The following warnings may be displayed during the self-test period when the instrument is switched ON.
Warning Description
Check Memory The instrument only has space to store less than 50 readings
before it is full. The exact number can be checked using the “VIEW READINGS” option.
Memory Full There is no more space in memory to store readings. Both the
store and log options will be disabled until the memory is cleared. The readings should be downloaded to PC using the optional soft­ware before memory is cleared.
Battery Low The instrument does not have enough power to operate for a full
day. The instrument should be recharged or connected to an exter­nal power supply.
Service Due It has been 12 months (or more) since your instrument was
returned to the manufacturer for a service. The performance and accuracy of the instrument may be impaired.
Low Flow The instrument’s gas inlet (or outlet) may be blocked. This warning
is most commonly caused by a water-logged or dirty sample fi lter. Change the sample fi lter and check for obvious blockages in the sample tubes.
Instruction 19-9339 42
Page 43
CO2 Analyzers 3100 & 3150
Check CO2 Cal. This warning is most commonly caused by an incorrect user cali-
bration. Try recalibrating the sensor or press “RETURN TO FAC­TORY SETTINGS.” If the warning persists it may be caused by dirt or damage to the infrared sensor. The instrument will need to be returned to your nearest Bacharach Service Center for service / repair.
Check O
Ref. Fault This may be caused by dirt or damage to the infrared sensor in the
*Invalid Confi g. The instrument has detected a problem with the confi guration
Change O
Change Pump The pump has exceeded its recommended run-time and should be
Change Battery The Lithium Ion battery has exceeded its recommende life-time or
Cal. This warning is most commonly caused by an incorrect user cali-
2
Cell The oxygen cell has not been changed for at least 2 years; its
2
bration. Try re-calibrating the sensor or press “RETURN TO FAC­TORY SETTINGS.” If the warning persists it may be caused by a damaged or faulty sensor. The instrument will need to be returned to your nearest Bacharach Service Center for service / repair.
instrument. The instrument will need to be returned to your nearest Bacharach Service Center for repair.
parameters. The instrument will need to be returned to your near­est Bacharach Service Center for service / repair. Most likely to be caused after fi rmware update.
performance and accuracy may be impaired. The instrument will need to be returned to your nearest Bacharach Service Center for service.
changed. The instrument will need to be returned to your nearest Bacharach Service Center for service.
number of charge cycles and should be changed. The instrument will need to be returned to your nearest Bacharach Service Center for service.
User Cal. Due If has been over a month since the instrument was last user cali-
brated. For optimal performance and accuracy it is recommended that the instrument is user calibrated each time it is used.
Invalid Time The instrument has an invalid time. This is most likely to occur af-
ter a reset. The correct time should be enterd using the set “TIME & DATE” option via the “UTILITIES” menu.
Invalid Date The instrument has an invalid date. This is most likely to occur af-
ter a reset. The correct date should be entered using the set “TIME & DATE” option via the “UTILITIES” menu.
Baro. Fault The instrument has detected a fault with the barometric sensor
of its calibration. This will have an effect on the accuracy of the readings as they are pressure compensated. The instrument will need to be returned to your nearest Bacharach Service Center for service.
NOTE: Certain confi guration problems can be corrected remotely.
Using the PC software, it is possible to export the current confi guration and e-mail it to a Bacharach Service Center. Depending on the error it may be possible to correct the confi guration fi le and import it back into the instrument.
Instruction 19-9339 43
Page 44
CO2 Analyzers 3100 & 3150
8.3 User Calibration Trouble Shooting
Error Remedy
User Zero Failed A possible reason for this is because the instrument is trying to
zero to a level which is outside the predetermined range set when the unit was calibrated at the factory. To rectify this, fi rst ensure the unit contains absolutely none of the gas which is being zeroed by fl ushing thoroughly with nitrogen. If it will not zero, then refer to the instructions given in the “FACTORY SETTINGS” section. If the instrument continues to fail to zero then the unit must be returned to your nearest Bacharach Service Center for investigation.
Calibration Failed Check that the span target is set to the correct value. If not, cor-
rect and retry spanning the channel. Repeat the entire procedure, including zeroing the channel, and then calibrate the span. Ensure that the reading is stable before spanning the channel.
8.4 Cross-Gas Effects
Carbon Dioxide is measured by infrared absorption at a wavelength specifi c to carbon dioxide. Therefore, the carbon dioxide reading will not be affected by any other gases. The oxygen sensor is a galvanic cell type and suffers virtually no infl uence from CO SO
, or H2, unlike many other types of oxygen cell.
2
, CO, H2S, NO2,
2
8.5 Hardware Reset
If for any reason the instrument ‘locks up’ and will not switch off, it is possible to force a hardware reset. Press and hold the “ON/OFF” key for 10 seconds; wait at least 15 seconds and the instrument should restart automatically.
NOTE: Performing a hardware reset may cause loss or corruption of
currently stored data including the time and date.
Instruction 19-9339 44
Page 45
CO2 Analyzers 3100 & 3150
9.0 Service
The instrument should be regularly serviced to ensure correct operation and accurate readings. Bacharach recommends a full service and factory calibration every 12 months.
Depending on usage, the O
cell should be replaced every 2-3 years.
2
User Serviceable Parts
NOTE: There are no user serviceable parts inside of the instrument.
Please do not attempt any repair as this may void the warranty.
The following parts are supplied and can be user serviced:
Sample Filter This should be regularly inspected for damage or discoloration
and changed if needed. The instrument should never be operated without the sample fi lter as this may result in water or dust entering the instrument. The fi lter should be changed immediately if water can be seen.
Sample Tubing Always ensure that the sample tubes are not contaminated or dam-
aged.
Instruction 19-9339 45
Page 46
CO2 Analyzers 3100 & 3150
10.0 Warranty Policy
This instrument is guaranteed, to the original end user purchaser, against defect in materials and workmanship for a period of 12 months from the date of the shipment.
This period covers repair or replacement of defective parts on an exchange basis.
The decision to repair or replace will be determined by Bacharach Inc.
To maintain this warranty, the purchaser must perform maintenance and calibration as prescribed in the operating manual.
Normal wear and tear, and parts damaged by abuse, misuse, negligence, or accidents are specifi cally excluded from the warranty.
NOTE: Please contact your nearest Bacharach Service Center for
further information.
Instruction 19-9339 46
Page 47
CO2 Analyzers 3100 & 3150
11.0 Technical Specifi cation
Power Supply
Battery Type Lithium Ion
Battery Life 10 Hours (8 hours with pump)
Battery Lifetime >300 Cycles
Battery Charger 5V DC external power supply and internal charging circuit
Charge Time 3 Hours
Alternative Power USB connector (run only)
DC Power Supply (run + charge)
Gas Ranges
Gases Measured CO2 By custom dual wavelength infrared cell
with reference channel
(Optional) By internal electrochemical cell
O
2
Oxygen Cell Lifetime Approximately 3 Years in Air
Range CO
CO2 O2
Measurement Accuracy
CO
(Model 3100)
O
Measurement Accuracy
CO
(Model 3150)
O
Response Time, T
90
CO O
2
0 - 20% (Model 3100) 0 - 10,000 PPM (Model 3150) 0 - 100%
2
Accuracy: ±(1% of range +2% of read-
ing) at reference conditions
1
Temperature dependence: ±0.2% read-
ing /°C (typical at 5% CO2)
Pressure Dependence: ±0.02% of read-
ing/hPa (typical at 5% CO
2
±1.0% Full Scale at constant temperature
)
2
and pressure ±2.0% Full Scale over operating tem­perature range
2
Accuracy: ±(1.5% of range +2% of read­ing) at reference conditions
1
Temperature dependence: ± 0.2% °C
(typical at 1000 ppm CO2)
Pressure Dependence: ± 0.02% of
reading/hPa (typical at 1000 ppm CO
2
±1.0% Full Scale at constant tempera-
)
2
ture and pressure ±2.0% Full Scale over operating tem­perature range
2
2
20 seconds60 seconds
1
Conditions during factory calibration, typically 68 °F, 29.5 inHg
Instruction 19-9339 47
Page 48
CO2 Analyzers 3100 & 3150
Characteristics
Temperature (optional) 32 to 122 °F (0 to 50.0 °C)
Temperature accuracy, typical ±0.36°F from 89.6 to 111.2 °F, ±0.9°F over the rest of
the range (±0.2°C from 32 to 44 °C, ±0.5°C over the rest of the range)
Barometric Pressure 23.6 to 35.4 inHg (800 to 1200 mbar)
RH Measurement (Optional) 0 to 100% non-condensing
RH Accuracy ±1.5% across range
Visual and Audible Alarm User-selectable CO
Communications USB type B mini-connector, HID device class
Data Storage 1000 readings + 270 events
and O2 alarm levels
2
Pump
Flow 100 cc/min (typical); 60 cc/min (minimum)
Environmental Conditions
Operating Temperature range 41 to 104°F (5 to 40°C)
Relative Humidity 0 to 95% non-condensing (RH probe 0 to 100% non-
condensing)
Barometric Pressure ±14.8 inHg from calibration pressure (±500 mbar)
Physical Features
Weight 18 oz (495 grams)
Size L 6.5” x W 4” x D 2.2” (L 165mm x W 100mm x D 55mm)
Case Material ABS / Polypropylene with Silicone Rubber Inserts
Keys 17 Resin capped Silicone rubber keys
Display Liquid crystal display, 128 x 64 pixel with RGB LED back-light
Gas sample fi lters Built-in gas dryer tube to remove moisture
NOTE: Specifi cations are subject to change without notice.
Instruction 19-9339 48
User-replaceable PTFE water fi lter
Page 49
CO2 Analyzers 3100 & 3150
12.0 Event Log
The following events are recorded in the instrument’s event log. The event log can only be downloaded using the additional PC software. Please refer to the PC software manual for further details.
Event Data
Cold Start / Reset Type of start (MCUSR, boot_key)
Firmware Version Description
Set Time Before and After
Set Date Before and After
Re-Flash Requested Type
Restore to Factory Settings Type
Comms Clear Memory Type 0=Readings, 1=Event Log
Change Contrast Before and After
Change Flow Fail Current Limit Before and After
Battery Less Than Critical Voltage Critical, Actual
RTC Date/Time Invalid None
Factory Calibration Invalid or Overdue Date, Difference
Service Invalid or Overdue Date, Difference
Power ON Self-Test, Sensor Out of Range Channel, Reading, Limit
User Calibration Set Zero OK Before, After
User Calibration Set Span OK Target, Before, After
User Calibration Set Zero Failed Target, Reading
User Calibration Set Span Failed Target, Reading
Attempt to Store when Readings Memory Full
Readings Memory Nearly Full Limit, Actual
Change Logging Mode Status, Interval, Pump time
Change Logging Mode ID ID
Flow Fail Current Limit Exceeded Limit, Actual
Max
Instruction 19-9339 49
Page 50
CO2 Analyzers 3100 & 3150
13.0 Important Notice to All Customers
The trash can symbol now displayed on equipment supplied signifi es that the apparatus MUST NOT be disposed of through the normal municipal waste stream but through a registered recycling scheme.
The Waste Electrical and Electronic Equipment directive (WEEE) makes producers responsible from July 1st 2007 in meeting their obligations, with the fundamental aim of reducing the environmental impact of electrical and electronic equipment at the end of its life.
The OEM is now registered as a producer and has joined a recycling scheme provider who will manage and report on our electrical waste on our behalf.
Producer Registration Number is WEE/GB0052TQ
So when your instrument is at the end of its life, please contact you nearest Bacharach Service Center who will advise you on the next step in order to help us meet our obligations.
Instruction 19-9339 50
Page 51
CO2 Analyzers 3100 & 3150
14.0 Glossary of Terms
Analyzer error messages For a list of standard error codes and for more
information, please refer to Section 8.0 Problem
Solving.
Analyzer warnings There are two types of warning messages dis-
played; general warnings that may not necessar­ily affect the instrument’s function, for example battery power low, and operational parameters that could affect the performance of the analyzer, for example CO
Battery charge A full battery charge will take approximately three
hours.
Calibration The process that an instrument will undergo to
enable it to measure and display the various parameters in accordance with the manufacturer’s specifi cation.
Chemical cell Type of gas detector which can be fi tted internally
to the analyzer at the time of manufacture.
out of calibration.
2
Clean air purge Process used to clear out gas from the inlet pipe
and the analyzer’s gas sensors prior to taking a new reading.
CO
2
Carbon dioxide gas.
Contrast adjustment Adjustable setting which darkens or lightens the
text displayed on the screen. Typically, this is used to compensate for different environmental temperatures. High temperature causes the display to darken and low temperature causes the display to lighten.
Data logging A mode of operation that enables the user to
leave the analyzer unattended to take readings automatically at predetermined times. The reading interval and pump run-time may be adjusted prior to commencing the logging cycle.
Download Terminology used to describe the transfer of
data from the analyzer to a PC via the optional software.
PC software The optional software enables the user to
maximize the operation of the incubator analyzer. Instrument readings and event log data may be installed to a PC for further analysis and exported to other applications such as MS Excel.
Instruction 19-9339 51
Page 52
CO2 Analyzers 3100 & 3150
Event log Record of signifi cant events in the life of the ana-
lyzer. Used as an aid to monitoring the use of the analyzer. It can also be used as a diagnostic tool if there is a problem with the analyzer. The event log can be viewed via the optional software. It cannot be viewed on the analyzer screen.
Exhaust port The point at which the gas exits the analyzer. This
is located on the top of the analyzer. It is threaded to take an M5 hose-barb to allow an exhaust tube to be attached if required.
Exhaust tube Clear plastic tubing used to route gases from the
exhaust port.
Factory settings Default calibration settings preset at time of factory
calibration.
Firmware Firmware is the name given to the analyzer’s inter-
nal software. This can be programmed using the re-fl ash command via the optional software. The latest software release can be downloaded from the Geotech website. It is also automatically up­dated when the analyzer is returned for servicing.
General warnings Displayed throughout the documentation with a
warning symbol. Warning information may affect the safety of users.
Inlet port Port located on the top of the analyzer to which the
inlet tube is attached.
LCD display Liquid Crystal Display. Fitted to the front panel of
the analyzer.
Main read screen The main analyzer screen for normal operations
and all operations are carried out from this screen.
Memory Tha analyzer memory should not be used as a
permanent storage medium and data collected should be transferred using the optional PC software.
Moisture removal tube Device used to remove water vapor from the
sample gas. This is located at the back of the analyzer and should not be covered or partially covered. This is not user serviceable and should NOT be removed by the user.
O
2
Oxygen gas typically 20.9% in air, measured via electro-chemical cell.
Over range codes Over range codes are errors above the maximum
allowed reading and will be displayed with more than chevrons (>>.>).
PPM Parts per million
Predetermined Set up prior to use via the optional PC software.
Instruction 19-9339 52
Page 53
CO2 Analyzers 3100 & 3150
Pump The device used to transfer a gas sample into the
instrument where that sample is not under pres­sure. Select the “PUMP” key on the analyzer to activate.
Regulator fl ow The regulator’s fl ow is factory set. It only requires
a few turns to open, no adjustment is available.
Sample tube Tube used to transfer sample gas from the source
to the analyzer.
Span The point at which the gas analyzer is calibrated
when a known concentration of the target gas is present.
Temperature probe External device to enable the instrument to display
data and record the temperature readings. This is an optional feature.
Under range codes Under range codes are errors below zero and will
be displayed with less than chevrons (<<.<).
User calibration Users have the facility to calibrate the analyzer
between services. User calibration of the gas ana­lyzer will improve the data accuracy in the range of the calibration gases used.
Warm-up self test Predetermined self-test sequence to test the
analyzer functions which takes place after the analyzer is switched on and lasts approximately 30-40 seconds.
Warranty The instrument is guaranteed against defect in
materials and workmanship for a period of 12 months from the date of shipment to the user and is subject to the recommended service and recali­bration requirements.
Zero The point at which the gas analyzer is calibrated
when there is non of the target gas present.
15.0 Replacement Parts
Item Description Part Number
Battery Charger with Adapters 19-0610
Sample Tube Kit 19-0613
Temperature Probe (5 mm) 19-0611
Sample Filters (Pack of 5) 19-0612
PC Software 19-0614
Calibration Kit 19-8027
Cylinder of N
Cylinder of 5% CO
Gas 23-4003
2
(for Model 3100) 24-1126
2
Cylinder of 2,500 ppm CO2 (for Model 3150) 24-1130
Instruction 19-9339 53
Page 54
16.0 Service Centers
United States
Bacharach, Inc. 621 Hunt Valley Circle New Kensington, PA 15068 Phone: 724-334-5051 Toll Free: 1-800-736-4666 Fax: 724-334-5723 Email: help@mybacharach.com
Canada
Bacharach of Canada, Inc. 250 Shields Court Unit #3 Markham, Ontario L3R 9W7 Canada Phone: 905-470-8985 Fax: 905-470-8963 Email: bachcan@idirect.com
CO2 Analyzers 3100 & 3150
México
Bacharach de México Playa Regatas No. 473 Tercer Piso Col. Militar Marte Delegación Iztacalco, 08830 México D.F. México Phones:+52-555-634-7740 +52-555-634-7741 FAX: +52-555-634-7738 Email: bacharach@prodigy.net.mx
Instruction 19-9339 54
Page 55
CO2 Analyzers 3100 & 3150
Instruction 19-9339 55
Page 56
621 Hunt Valley Circle, New Kensington, PA 15068
Ph: 724-334-5000 • Fax: 724-334-5001 • Toll Free: 800-736-4666
Website: www.mybacharach.com • E-mail: help@mybacharach.com
Loading...