Bacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in material and
manufacture and will conform substantially to Bacharach Inc.'s applicable specifications. Bacharach's liability and Buyer's
remedy under this warranty are limited to the repair or replacement, at Bacharach's option, of this Product or parts thereof
returned to Seller at the factory of manufacture and shown to Bacharach Inc.'s reasonable satisfaction to have been defective;
provided that written notice of the defect shall have been given by Buyer to Bacharach Inc. within two (2) years after the date
of delivery of this Product by Bacharach, Inc.
Bacharach, Inc. warrants to Buyer that it will convey good title to this Product. Bacharach's liability and Buyer's remedy under
this warranty of title are limited to the removal of any title defects or, at the election of Bacharach, to the replacement of
this Product or parts thereof that are defective in title.
The warranty set forth in paragraph 1 does not apply to parts the operating instructions designate as having a limited shelflife or as being expended in normal use (e.g., filters).
THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN AND ACCEPTED IN LIEU OF:
(I) ANY AND ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE: AND
(II) ANY OBLIGATION, LIABILITY, RIGHT, CLAIM OR REMEDY IN CONTRACT OR TORT, WHETHER OR NOT ARISING
FROM BACHARACH'S NEGLIGENCE, ACTUAL OR IMPLIED.
The remedies of the Buyer shall be limited to those provided herein to the exclusion of any and all other remedies including,
without limitation incidental or consequential damages. No agreement varying or extending the foregoing warranties,
remedies or this limitation will be binding upon Bacharach, Inc. unless in writing, signed by a duly authorized officer of
Bacharach.
NOTICE
Product improvements and enhancements are continuous, therefore the specifications and information contained in this
document may change without notice. Bacharach, Inc. shall not be liable for errors contained herein or for incidental or
consequential damages in connection with the furnishing, performance, or use of this material. No part of this document may
be photocopied, reproduced, or translated to another language without the prior written consent of Bacharach, Inc.
Revision 6 of this manual corresponds to firmware version 1.20 or newer.
1.2.Conventions Used in This Manual ............................................................................................................................... 1
1.4.Features and Capabilities ........................................................................................................................................... 3
1.5.1.Key System Components .............................................................................................................................. 3
1.5.2.Operation Modes: Search vs. Measure......................................................................................................... 4
2.3.Probes and Hoses .................................................................................................................................................... 12
2.3.1.Standard Probe with Flexible Tubing ........................................................................................................... 12
2.3.2.Standard Probe with LED and Push Button.................................................................................................. 13
CHAPTER 3: MENU SYSTEM OVERVIEW ............................................................................................................................ 18
3.2.3.Changing Values of Menu Items ................................................................................................................. 22
3.2.4.Configuration-Dependent Menu Items ....................................................................................................... 23
5.2.Turning on the H25-IR PRO ...................................................................................................................................... 36
5.5.1.Gas Type .................................................................................................................................................... 46
5.6.4.Date/Time Set ............................................................................................................................................ 65
5.6.7.Serial Data ................................................................................................................................................. 70
5.7.2.Sensor Data Screen .................................................................................................................................... 79
5.8.Adjustment to a Known External Leak Source .......................................................................................................... 81
5.9.Custom Gas Type Configuration ............................................................................................................................... 83
5.10.Factory Gas Type Configuration ............................................................................................................................... 84
6.7.Spare Parts and Accessories ..................................................................................................................................... 95
APPENDIX C: SUPPORTED GAS TYPES.............................................................................................................................. 106
C.1.Basic Gas Types ...................................................................................................................................................... 106
C.2.Factory Gas Types .................................................................................................................................................. 108
APPENDIX D: DECLARATION OF CONFORMITY ................................................................ ERROR! BOOKMARK NOT DEFINED.
INDEX .............................................................................................................................................................................. 111
Figure 1-1. Toggling Between Search and Measure Modes ................................................................................................... 4
Figure 1-3. Adjusting the H25-IR Pro to a Recognized Leak Standard .................................................................................... 8
Figure 2-1. Front Panel Components of the H25-IR PRO ..................................................................................................... 11
Figure 2-2. Back Panel Components of the H25-IR PRO ...................................................................................................... 12
Figure 2-3. Components of the Standard Probe with Flexible Tubing .................................................................................. 12
Figure 2-4. Components of the Standard Probe with LED and Push Button ......................................................................... 13
Figure 2-5. Components of the Smart Probe ...................................................................................................................... 14
Figure 2-7. Connecting the Probe to the H25-IR PRO .......................................................................................................... 16
Figure 2-8. Connecting the Power Cord to the H25-IR PRO ................................................................................................. 16
Figure 3-1. The Menu Interface (Display and Buttons): A Key User Interface with a Menu-Driven Architecture .................. 18
Figure 3-2. Arrows Indicating Additional Menus, Menu Items, and Parameter Editing Options ........................................... 20
Figure 3-3. Menu Architecture Summary ........................................................................................................................... 21
Figure 3-4. Front Panel Components with Sample Display ................................................................................................... 21
Figure 3-5. Examples of Value Types .................................................................................................................................. 23
Figure 3-6. PROBE BUTTON Menu Item Based on Probe Style Ordered ............................................................................... 23
Figure 3-7. Version Screen at Power Up ............................................................................................................................. 24
Figure 5-8. Graphical Representation of the Measurement Log .......................................................................................... 43
Figure 5-9. Samples of Graphic and Numeric Display Styles ................................................................................................ 51
Figure 5-10. Options of the Advanced Setup Screen ........................................................................................................... 54
Figure 5-11. Weighing Your Application Needs: When to Use User Setup 0 (Default) vs. User Setups 1-4 ........................... 55
Figure 5-12. Sample Search Screen (with User Setup Active) Showing User Name “FREEZER 1” .......................................... 57
Figure 5-13. Advanced Setup Screen Without (Left) and With (Right) a Detected Smart Probe ........................................... 72
Figure 5-14. Locked Message for Subsequent Attempts at External Adjustment ................................................................. 83
Figure 6-1. Location of Fuse Holder .................................................................................................................................... 89
Figure 6-2. Probe Filter and “O” Ring Locations .................................................................................................................. 90
Figure 6-3. Back Panel of the H25-IR Pro Showing Proper Orientation of the External Filter ................................................ 91
Figure 6-4. Removing the Top Cover .................................................................................................................................. 92
Figure 6-5. NDIR Sensor Showing Three Cable Ties Removed ............................................................................................. 93
Table 1-1. H25-IR PRO Functions and Features ..................................................................................................................... 5
Table 2-1. H25-IR PRO Standard Probe Components .......................................................................................................... 13
Table 2-2. H25-IR PRO Standard Probe Components .......................................................................................................... 13
Table 2-3. H25-IR PRO Smart Probe Components ............................................................................................................... 15
Table 2-4. Connecting the Flexible Probe Tube ................................................................................................................... 15
Table 3-1. Key Components of the Menu Structure ............................................................................................................ 19
Table 3-1. Front Panel Buttons........................................................................................................................................... 22
Table 3-2. Common Key Combinations............................................................................................................................... 24
Table 4-2. Procedure to Change Factory-Set Probe Configuration....................................................................................... 33
Table 5-1. Function Reference ........................................................................................................................................... 35
Table 5-2. Turning on the H25-IR PRO ................................................................................................................................ 36
Table 5-3. Changing the Volume Setting ............................................................................................................................ 38
Table 5-4. Accessing the Main Menu Screens Using the Setup Button ................................................................................ 40
Table 5-5. Logging Data ..................................................................................................................................................... 44
Table 5-6. Accessing the BASIC SETUP Screen..................................................................................................................... 45
Table 5-7. Setting the Gas Type from the List of Standard Gases ........................................................................................ 46
Table 5-8. Gas Type Menus and Selection .......................................................................................................................... 47
Table 5-9. Display Units Available in the H25-IR PRO .......................................................................................................... 48
Table 5-11. Changing the Unit Display Setting .................................................................................................................... 49
Table 5-12. Changing the Alarm Setpoint ........................................................................................................................... 50
Table 5-13. Changing the Display Style ............................................................................................................................... 52
Table 5-14. Accessing the ADVANCED SETUP Screen .......................................................................................................... 53
Table 5-15. Accessing the USER SETUP Screens 1-4 ........................................................................................................... 55
Table 5-16. Changing a User Setup Name ........................................................................................................................... 56
Table 5-17. Changing a User Setup Gas Type ...................................................................................................................... 57
Table 5-18. Changing a User Setup Alarm Setpoint ............................................................................................................. 58
Table 5-19. Changing a User Setup Feedback Mode ........................................................................................................... 60
Table 5-20. Accessing the RELAY SETTING Screen ............................................................................................................... 61
Table 5-22. Accessing the Measurement Log ..................................................................................................................... 64
Table 5-23. Setting the Date and Time ............................................................................................................................... 65
Table 5-24. Setting the Log Limit ........................................................................................................................................ 66
Table 5-25. Default Loop Factor Mapping Example (0.016 mA = 1 PPM) ............................................................................. 68
Table 5-26. Changing the Loop Factor ................................................................................................................................ 68
Table 5-27. Serial Data Options .......................................................................................................................................... 70
Table 5-28. Changing the Serial Data Setting ...................................................................................................................... 70
Table 5-30. Changing the Probe Button Setting .................................................................................................................. 73
Table 5-31. Accessing the Diagnostics Screen (Method #1)................................................................................................. 74
Table 5-32. Accessing the DIAGNOSTICS Screen (Method #2) ............................................................................................. 75
Table 5-33. Accessing the Current Faults Screen ................................................................................................................ 77
Table 5-35. Accessing the Sensor Data Screen .................................................................................................................... 79
Table 5-36. Sensor Data Parameters and Sample Values .................................................................................................... 80
Table 5-37. Adjusting the H25-IR PRO to a Known External Leak Source ............................................................................. 81
Table 5-38. Configuring the H25-IR PRO to use a Factory Gas Type ...................................................................................... 84
Table 6-1. Items Required for Fuse Replacement ............................................................................................................... 88
Table 6-4. Probe Tip Filter Replacement Procedure ........................................................................................................... 89
Table 6-5. Items Required for “O” Ring Replacement ......................................................................................................... 90
Table 6-6. Probe “O” Ring Replacement Procedure ............................................................................................................ 90
Table 6-7. Items Required for External Filter Replacement ................................................................................................. 90
Table 6-12. Spare Parts and Accessories ............................................................................................................................ 95
Table 6-13. Items Required for Upgrading Firmware .......................................................................................................... 96
WARNING: This symbol and/or the use of the word WARNING indicates a potential hazard associated with
the use of this equipment. Failure to correctly perform or adhere to the procedure, practice, or condition
could result in death or serious injury.
WARNING: This symbol and/or the use of the word WARNING indicates a potential hazard from electrical
shock. Failure to correctly perform or adhere to the procedure, practice, or condition could result in death
or serious injury from electrical shock.
CAUTION: This symbol and/or the use of the word CAUTION indicates a potential hazard associated with the
use of this equipment. Failure to correctly perform or adhere to the procedure, practice, or condition could
result in minor or moderate injury. This may also be used to alert readers of unsafe practices.
IMPORTANT: This symbol and the use of the word IMPORTANT calls attention to components that are
sensitive to the effects of electro-static discharge (ESD). Equipment damage will likely occur if proper ESD
mitigation procedures are not followed.
CHAPTER 1: INTRODUCTION
1.1. Product Overview
The H25-IR PRO is the next generation of industrial-grade refrigerant leak detector and gas analyzer for the detection of CFCs,
HCFCs, HFCs and halogen gas compounds including R600a. This instrument can be used to locate and then quantify gas leaks,
as well as log and totalize a group of leaks in a system. Building on the success of the H25-IR, the H25-IR PRO adds new
groundbreaking capabilities that ensure accurate and reliable operation.
The H25-IR PRO from Bacharach continues the tradition of rugged industrial gas detection and analysis that the H25-IR
established. New ALQä technology raises the bar for gas detection instruments, and has real benefits to the end user. The
new menu system in the H25-IR PRO simplifies setup and operation. Recognized and valued for its digital numeric leak rate
display (versus other solutions featuring only light bars), the Bacharach H25-IR PRO continues to offer the features requested
by industry.
If you have a working knowledge of your leak detector, you will find this manual useful as a reference tool. If you are new to
the use of leak detectors, this document is educational in the principles of gas detection and the proper operation of this
device.
1.2. Conventions Used in This Manual
The following symbols and statements may be used within this manual.
IMPORTANT: The use of the word IMPORTANT in this manual calls attention to a procedure, practice,
condition, or the like, which if not correctly performed or adhered to, could result in incorrect performance
of or damage to the equipment and may void the warranty.
NOTE: The use of the word NOTE in this manual provides emphasis of a feature, operation, etc. Notes are
also useful to address potential service calls/issues.
WARNING: This instrument has not been designed to be intrinsically safe for use in areas classified as
hazardous locations. For your safety, DO NOT use it in hazardous (classified) locations.
WARNING: This is NOT a safety device. Some gases which this instrument can detect may be
combustible/flammable. When properly configured, this instrument is designed to alarm at concentrations
that are lower than the explosive limit of the gas. As such, it is the buyer’s responsibility to initiate an
immediate planned response to any gas leaks as soon as they are detected. This equipment should NEVER
be used to measure or sample gases at or above their respective lower explosive limits.
WARNING: The leak detector must be operated with a connection to a protective ground. Failure to do
poses a potential shock hazard and is also a violation of electrical safety standards applicable to this type of
equipment.
WARNING: Do not operate this equipment in the presence of flammable liquids, vapors, or aerosols.
Operation of any electrical instrument in such an environment constitutes a safety hazard.
WARNING: It is imperative that the exhaust port on this instrument be properly vented as described in this
manual. Failure to do so may constitute a safety hazard.
WARNING: Extreme care should be exercised when accessing the interior of the leak detector. Only
qualified electrical maintenance personnel should make connections and perform adjustments. Always
remove AC power before opening the leak detector’s enclosure.
WARNING: The protection provided by the leak detector may be impaired if the leak detector is used in a
manner not specified by Bacharach, Inc. Modifications to this leak detector, not expressly approved, will
void the warranty.
WARNING: Do not continue to use this equipment if there are any symptoms of malfunction or failure. In
the case of such occurrence, de-energize the power supply and contact a qualified repair technician or the
nearest Bacharach Service Center.
WARNING: This device uses type “F” fuses rated at 1.0 A, 250 VAC. Replace ONLY with Bacharach-approved
fuses (see Spare Parts and Accessories on page 95 for the part number).
1.3. Safety Precautions
To assure operator safety and the proper use of the leak detector, please read this manual. It provides important information
on the configuration, operation, and maintenance of the leak detector.
IMPORTANT: Operating this device above the altitude limit of 6,562 ft (2,000 m) will affect the expected
results and specifications.
IMPORTANT: The leak detector uses a universal power supply that is capable of accepting inputs of 100 to
240 VAC, 50/60 Hz. The leak detector’s power consumption is 16 Watts. It is highly recommended that the
leak detector be connected directly to the AC power source, preferably on its own circuit with UPS or surge
protection.
NOTE: Use ONLY the provided connectors for electrical and communications wiring. Drilling into the box
will void the warranty.
NOTE: To clean the outside of the case use a dry cloth. To avoid shock hazard and/or equipment damage,
DO NOT use soap and water.
1.4. Features and Capabilities
Below are some of the key the features and capabilities of the H25-IR PRO.
•Automatic calculation of gas concentration regardless of flow rate (not effected by dilution effects that would
normally cause reading inaccuracy)
• No periodic calibration required
• Available in various gas models to detect and measure:
- Halogen Refrigerants - SF6
- R600a and R290 - CO2
• Infrared detector
• Provision for external adjustment, including adjustment to other refrigerant types
• User-selectable leak rate measurement in:
- oz/yr - mL/s ´ 10-5
- g/yr - PaM3/s ´ 10-5
- ppm
• Easy to read LED backlit display
• Excellent background gas rejection (3 times improved versus H25-IR)
• 4-20 mA current loop output
• Relay outputs for external equipment control
• On-screen menu system for system setup, diagnostics, and data logging
• Probe and hose assemblies available in two lengths:
- 6 foot (1.8 m) - 12 foot (3.7 m)
• 8 inch (200 mm) flexible probe tube
• Various probe styles to meet your needs:
- Standard
- Standard with push button
- Smart Probe
1.5. Functional Overview
1.5.1. Key System Components
The front panel of the H25-IR PRO contains 7 membrane push buttons, probe connector, and a 128x64 dot LCD (Liquid Crystal
Display) that provides a bright, high-contrast display for easy viewing. The instrument is housed in an all-metal case with
adjustable handle. A hand-held probe is used to draw in a gas sample from the area under inspection. The probe is available
in various styles and lengths.
NOTE: To toggle between search mode and measure mode, use the MODE button on the front panel. This
feature is also available from the probe itself. The push button on the standard probe with push button can
be programmed to toggle between modes. The smartprobe has a MODE button that functions the same as
the front panel MODE button.
NOTE: The function of the button on the standard probe with push button is defined by the user. It can be
programmed to zero the instrument or to toggle the instrument between search mode and measure mode.
NOTE: The user may easily set an alarm setpoint, which automatically scales the instrument’s response.
This feature is useful in avoiding small nuisance leaks when searching for larger ones.
1.5.2. Operation Modes: Search vs. Measure
The instrument has two modes of operation: Search and Measure.
• Search mode is used to locate the source of a leak.
• Measure mode is used to quantify the size of the leak.
Figure 1-1. Toggling Between Search and Measure Modes
In search mode, the location of a leak is pinpointed by means of visual and audio indicators positioned on both the instrument
and probe (probe indication, when equipped). As the probe tip approaches the leak point:
• The length of a bar graph on the front panel display becomes longer (or the digital reading increases).
• The frequency of an audible tone increases (depends on Feedback Mode setting, described later).
• The probe with LED (when equipped) also indicates a system response with an LED blinking faster as the response
A leak measurement can be logged in memory with a time/date stamp. Up to 50 leaks can be logged.
This feature is useful in testing an installation that has a maximum permissible leak rate. Logged data
can be recalled for viewing on the display. Works in Measure Mode only.
Setup Presets
There are four user-defined setups that allow the operator to quickly switch between frequently
used combinations of Gas Type, Range, Feedback Mode, Setpoint, and other custom capabilities.
Diagnostics
Extensive diagnostics keep track of several system parameters. When an error is detected, the i
symbol along with a brief error message is shown at the top of the display, and an audible beep is
heard every 2 seconds. Pressing the i button on the instrument keypad takes the user to additional
information, making diagnosis simple and easy. Text descriptions of all current error conditions can
be displayed from the instrument’s Diagnostics menu.
Relay Outputs
Four SPDT relays provide control over a variety of external equipment (e.g., horns, bells, strobe
lights) that can be automatically triggered when their associated relays are energized by any one of
several different operating conditions.
Current Loop
Output
A 4-20 mA current loop output can be used for the connection of external monitoring equipment
(e.g., chart recorder). An RS-232 serial port can output leak rate data as ASCII text at 10 times per
second.
In the measure mode, the probe tip is held over the leak point until the displayed value of the leak rate stabilizes. Leak rates
can be displayed in units of:
• oz/yr (ounces per year)
• g/yr (grams per year)
• mL/s-5 (milliliters per second ´ 10-5)
• Pa·m3/s-5 (Pascal meters cubed per second ´ 10-5)
• ppm (parts per million)
The measure mode can also be used to display the concentration of gas contained in a confined space with the instrument
set up in its ppm mode after first zeroing on fresh air. Measure mode automatically changes the display from a bar graph to
a digital reading.
The leak rate is measured by first capturing all the gas leaking from the component under test. Then, by accurately measuring
the flow rate of the sampling stream and the gas concentration within that stream, the leak rate can be calculated using the
equation listed below. The instrument then converts this data into the desired leak rate units of measure and displays the
leak rate on the instrument’s front panel.
where: Leak Rate = Rate of gas leak from component under test
Mass Flow Rate = Sample mass flow rate
Gas Concentration = Concentration of gas in the sample (ppm)
ALQä technology eliminates the need to periodically adjust the instrument to a calibrated reference leak source. The result
is an instrument that maintains its accuracy in real time, even under changing conditions. ALQä technology:
• eliminates down time for periodic adjustments or instrument checks
• extends the working life of the probe tip filters
• makes the instrument impervious to problems that normally could cause inaccurate results.
For example, small leaks that normally could dilute the sample and cause the measured concentration to change have no
effect on the H25-IR PRO. The H25-IR PRO comes standard with ALQä technology.
ALQä technology eliminates the need to periodically adjust the instrument, so the H25-IR PRO contains no internal leak
reference. For users who require a “bump test” to verify normal operation, any compatible gas source or reference leak
source can be used. For users who are required to validate instruments against a known reference leak source (due to
company quality policies or practices, for example), an easy-to-use adjustment option is available to set a leak rate multiplier
for the selected target gas (see Adjustment to a Known External Leak Source on page 81). An optional NIST-traceable,
calibrated leak reference is available from Bacharach that may be used for bump testing and/or validation.
1.5.5. Using a Reference Leak Source
An optional Bacharach leak reference may be purchased and established as a reference standard, and used to adjust the H25IR PRO’s indicated reading. It features a calibrated output (NIST traceable certification provided) and may be refilled as
needed. The optional Bacharach leak reference is available with most industry-standard refrigerants. Special calibrations on
other refrigerants are available at extra cost. Contact Bacharach for additional information.
Reference leaks provide reliable output when used correctly. However, even the best reference leaks are typically only
accurate to within ±15% (absolute) of the indicated reading. For example, if the indicated leak rate is 5 g/yr, the actual leak
rate could be anywhere from 5.75 g/yr to 4.25 g/yr. For leak references that do not include a pressure gauge, the tolerance
can be even larger, as the indicated leak rate (usually on the certificate of calibration) is only true at one reference
temperature. The same ±15% output tolerance combined with varying temperatures means these devices have to be used
very carefully for meaningful results. Refer to the example shown in Figure 1-2 on page 7.
The H25-IR PRO responds to within ±5% of reading (absolute) at reference conditions and maintains this across the specified
operating temperature range (see Figure 1-2 on page 7). Users who need to adjust the H25-IR PRO to their recognized leak
standard may do so within a ± 25% (absolute) tolerance band, to allow for the temperature-affected output tolerance that is
present in reference leaks. Refer to Figure 1-3 on page 8. The user simply applies his leak reference, and uses the H25-IR
PRO keypad to adjust the indicated measurement to match the leak reference value. After the adjustment is applied, the
H25-IR PRO will use its built-in calculations to provide relative leak rates, while ensuring conformance to the required
standard.
Your H25-IR PRO Refrigerant Gas Leak Detector and all accessories were carefully inspected at the factory before shipment,
and packed to be highly resistant to damage while in transport.
When you receive the instrument, please unpack it promptly and make a visual inspection to make sure that no damage has
occurred during shipment and that all ordered items have been received. Your order should consist of:
§ H25-IR PRO Instrument
§ Probe and Hose Assembly (as ordered) with 8 inch Flexible Probe Tube
§ AC Power Cord (as ordered)
§ Instruction Manual
If damage was found, immediately file a claim with the carrier. If an item is missing, please contact Bacharach’s Service
Department for assistance.
2.2. External Components
2.2.1. Front Panel
Front panel components are shown in Figure 2-1.
Figure 2-1. Front Panel Components of the H25-IR PRO
IMPORTANT: The H25-IR PRO is factory configured to use the probe with which it is purchased. This factory
configuration includes internal wiring connections. If you plan to use a different probe than the type that
was shipped with your H25-IR PRO, you must move an internal cable to a different connector. For additional
information, refer to Standard Versus Smart Probe Wiring on page 32.
2.2.2. Back Panel
Back panel components are shown in Figure 2-2.
Figure 2-2. Back Panel Components of the H25-IR PRO
2.3. Probes and Hoses
2.3.1. Standard Probe with Flexible Tubing
The standard probe is available in a 6-foot hose model and a 12-foot hose model. Components of the standard probe are
shown in Figure 2-3 and described in Table 2-1.
Figure 2-3. Components of the Standard Probe with Flexible Tubing
NOTE: When ordered with the instrument, the probe type is configured at the factory. If a standard probe
is to be used as a Smart Probe replacement, additional internal changes are required.
Component
Function/Notes
Probe Tube
• Flexible
• Length: 8 in (203 mm)
Probe Tip
•Contains filter
Probe Connector
•Align with red dot on instrument’s front panel PROBE receptacle
Hose
•Available in 6-foot (1.8 m) length and 12-foot (3.7 m) length
Component
Function/Notes
Probe Tube
• Flexible
• Length: 8 in (203 mm)
Probe Tip
•Contains filter
Push Button
•Can be programmed to either zero the instrument, or switch between the search and measure
mode of operation
LED
•Flash rate increases as the detected gas level increases, or lights steady when a user selected
leak rate is detected
Probe Connector
•Align with red dot on instrument’s front panel PROBE receptacle
Hose
•Available in 6-foot (1.8 m) length and 12-foot (3.7 m) length
Table 2-1. H25-IR PRO Standard Probe Components
2.3.2. Standard Probe with LED and Push Button
The standard probe with LED and push button has an LED indicator and a programmable button (to either zero the instrument
or switch between search mode and measurement mode). The standard probe with LED and push button is available in a 6foot hose model and a 12-foot hose model. Components are shown in Figure 2-3 and described in Table 2-1.
Figure 2-4. Components of the Standard Probe with LED and Push Button
NOTE: The function of the button on the standard probe is defined by the user. It can be programmed to
zero the instrument or to toggle the instrument between search mode and measure mode.
NOTE: When ordered with the instrument, the probe type is configured at the factory. If a standard probe
with push button and LED is to be used as a Smart Probe replacement, additional internal changes are
required.
2.3.3. Smart Probe
The smart probe has an LCD display screen and control panel buttons similar in look and function to those on the front panel.
The smart probe is available in a 6-foot hose model and a 12-foot hose model. Components of the standard probe are shown
in Figure 2-5 and described in Table 2-3.
• Same functions as front panel buttons on instrument
• MODE and SETUP buttons can be used to lock/unlock the keypad (see Figure 2-6)
Display Screen
•Same function as front panel display on instrument
Probe Connector
•Align with red dot on the PROBE receptacle of the instrument’s front panel.
Hose
•Available in 6-foot (1.8 m) length and 12-foot (3.7 m) length
NOTE: The keypad on the smart probe can be locked to prevent accidental key presses (for example, when
working in tight spaces). Press the MODE and SETUP buttons simultaneously to toggle this feature. Refer to
Figure 2-6.
Step
Description
1
Screw the flexible probe tube onto the probe assembly.
2
Plug the probe connector into the front panel PROBE receptacle by aligning the red dots on both the connector
and receptacle.
Table 2-3. H25-IR PRO Smart Probe Components
Figure 2-6. H25-IR PRO Smart Probe Key Lock Feature
NOTE: The H25-IR PRO is supplied with a 6 foot grounded AC power cord. Make sure that a compatible AC
power receptacle (socket) is available nearby.
Figure 2-7. Connecting the Probe to the H25-IR PRO
2.4. Power Cord
The H25-IR PRO uses a universal AC power supply that accepts inputs of 100 to 240 VAC, 50/60 Hz. Power consumption is
approximately 16 Watts (maximum).
Before connecting the AC power cord, first make sure the instrument’s power switch is OFF. Next, connect the power cord to
the instrument’s rear panel AC power receptacle, and then plug the power cord into a nearby AC power receptacle.
Figure 2-8. Connecting the Power Cord to the H25-IR PRO
The NDIR sensor is an internal component that sends infrared light through the sampled gas. The amount of infrared light
that passes through the gas (vs. the amount that is absorbed by the gas) provides a complex electrical signature that is used
to quantify the target gas. This component is an 11.5” long cylinder that can be removed and installed in the field, should it
ever need to be replaced. For information on replacing the NDIR sensor yourself, refer to Chapter 6: Maintenance (page 88)
for more information. Alternatively, you can send your H25-IR PRO to the factory or service center to have the NDIR sensor
replaced.
The H25-IR PRO is used to locate and quantify gas leaks. Interfaces provide external information, configuration, operation,
evaluation results, and even diagnostic information, among other things. Interfaces (i.e., the “borders” between the outside
world and the H25-IR PRO) are those key boundaries where the information is exchanged. The main interfaces of the H25-IR
PRO are:
• Speaker (used to audibly notify the operator of analysis information)
• The gas probe (used by the operator to draw gas into the detector for analysis)
• Front panel LCD and buttons (used to exchange configuration, operation, evaluation, and diagnostic information).
• Relay outputs (used as switches to control electrical devices based on input gas and programmable conditions)
• 4-20mA output (used to drive a graphical chart recorder or similar device)
• RS-232 communications port (PC interface to receive live flow rate data).
• Automatically changes to Search/Measure mode screen after countdown reaches zero.
Search/
Measure Mode
• Main monitoring screens
• Two default displays for each operation mode: Search Screen and Measure Screen
• Search screen for finding leaks, and Measure screen for quantifying the leak
• Toggle between displays using the MODE button
• Can return to this screen from any menu by pressing the ESC button one or more times
• Timeouts cause this screen to be displayed automatically after a predefined inactivity period
• Press the VOL button to access the volume control screen
• Press the RANGE button to toggle the display among three display ranges
• The menus are accessible from this screen by using the SETUP button
Basic Setup Menu
• Contains basic setup parameters needed to configure the H25-IR PRO for basic operation
• Accessed from the Search/Measure screen using the SETUP button
• Use the Up and Down Arrow buttons to select menu items. Use the ENT button to edit.
• Press the ESC button to return to the Search/Measure Mode screen
Advanced Setup
Menu
•Contains configuration parameters needed to configure the more advanced functions of the
H25-IR PRO
•Accessed from the Basic Setup Menu screen using the Right Arrow (SETUP) button (a second
time from the Search/Measure screen) (must be at the top of the menu)
• Use the Up and Down Arrow buttons to select menu items. Use the ENT button to edit.
• Press the ESC button to return to the Search/Measure Mode screen
Diagnostics Menu
• Contains troubleshooting and diagnostic information about the H25-IR PRO
• Accessed from the Advanced Setup Menu screen using the Right Arrow (SETUP) button (a third
time from the Search/Measure screen)
• Accessed directly from the Search/Measure screen using the Info (i) button
• Use the Up and Down Arrow buttons to select menu items. Use the ENT button to edit.
• Press the ESC button to return to the Search/Measure Mode screen
Though all of these “interfaces” may be used during the operation of the H25-IR PRO, the front panel LCD and buttons provide
an easy-to-use interface consisting of a robust suite of configuration, customization, operation, evaluation, and diagnostic
functions. These features are grouped logically and functionally in a menu architecture.
An operator selects items from SETUP, ADVANCED SETUP, and DIAGNOSTICS menus. Each category contains related items
from which you may choose using the Arrow keys (the active field is highlighted). Some of these “choices” provide feedback
about the sampled gas or the status of the H25-IR PRO, while some set important parameters that define how you would like
the H25-IR PRO to operate. Regardless, it is important to become familiar with the menu architecture of the H25-IR so that
you can easily locate options, properly configure them, and take advantage of the customizable features.
Table 3-1 shows key components of the H25-IR PRO’s menu structure.
Menu System Overview H25-IR PRO Instruction Manual
BASIC SETUP >>
GAS TYPE: R134A
UNIT DISPLAY: G/YR
ALARM SETPOINT: 10.0
DISPLAY STYLE: GRAPHIC
qp SELECT ± EDIT
<< ADVANCED SETUP >>
USER SETUPS 1-4
RELAY SETTINGS
MEASUREMENT LOG
DATE/TIME SET
LOG LIMIT
¯
qp SELECT ± EDIT
SELECT GAS TYPE
p R22
R134A
q R22
± SELECT ESC=QUIT
EDIT ALARM SETPOINT
qp TO CHANGE
10.0 PPM
tu TO MOVE
± SAVE ESC=QUIT
3.2. Menu Navigation
3.2.1. Overview
Navigating the H25-IR PRO menus is intuitive. Furthermore, it is likely that operators will spend only a proportionally small
amount of time configuring the unit versus a greater amount of time at the search and measure screens (locating leaks and
quantifying them once they are found). Nevertheless, understanding how to navigate all of the menus promotes efficient
operation and the best use of your time.
Consider the following general rules for menu navigation.
• The SEARCH MODE or MEASURE MODE screen is typically your home screen
• The VOL, MODE, and RANGE buttons are typically used from that screen during normal operation
• The SETUP button is typically used to access the menu system (starting with the BASIC SETUP menu)
• The BASIC SETUP menu contains all of the options that are required to setup and operate the instrument
• From the BASIC SETUP menu, the VOL, RANGE, MODE, and SETUP buttons are used primarily as UP, DOWN, LEFT,
and RIGHT navigation controls (respectively)
• The ESC key is used to cancel an operation or “back out” of a menu
• The ENT key (¿) is used to confirm the selection of a menu or value
In addition to these general rules, be sure to watch for clues such as arrows on the LCD display. Arrows are used to indicate
additional menus, additional menu items, and/or parameter editing options. Examples are shown below in Figure 3-2.
Figure 3-2. Arrows Indicating Additional Menus, Menu Items, and Parameter Editing Options
Press the SETUP button to access the main menu system. The menu system (see Figure 3-3) enables the operator to perform
such functions as setting up the instrument, viewing logged readings, configuring the four user-defined setups, and viewing
the instrument’s diagnostic readings. The buttons on the instrument’s front panel (see Figure 3-4) and Smart Probe are
mapped the same and can be used inter-changeably, with exception of the i button.
H25-IR PRO Instruction Manual Menu System Overview
After pressing SETUP, select a menu item by first using the Up and Down Arrow buttons to highlight the desired item, and
then pressing ENT to select that item and display its data screen. After an item has been selected, use the Left, Right, Up, and
Down Arrow buttons to highlight or change data within a screen. Pressing the ENT button saves the new data, while pressing
the ESC button aborts the operation and displays the previous screen.
Figure 3-3. Menu Architecture Summary
Figure 3-4. Front Panel Components with Sample Display
Menu System Overview H25-IR PRO Instruction Manual
Button
Function
• Selects a highlighted menu item
• Saves a selection or manually entered data (e.g. date and time)
• Zeros the gas reading (in the Measure and Search screens)
• Displays a previously viewed menu or data screen
• Cancels a data entry function; retains old data; and displays previous screen
• Logs the currently displayed gas measurement in memory when in the Measure mode, where it
can later be reviewed using the Main menu’s MEASUREMENT LOG function
• Directly accesses the Diagnostics menu.
• Used to enable instrument adjustment to an external leak reference (when pressed and held—
refer to Adjustment to a Known External Leak Source on page 81).
•Displays the Volume
Adjustment screen (then
Up Arrow button slowly
raises volume, Down
Arrow button slowly
lowers volume)
NOTE: These four buttons
also serve as up, down, left,
and right navigation and
value adjustment buttons
based on menu context.
The front panel shows
corresponding arrows next
to each button’s alternate
function.
•Accesses the
instrument’s setup
menus
•Toggles the operating
mode between Search
and Measure
•Menu selection and value adjustment arrow buttons:
VOL Up button: Moves selected (highlighted) menu item
up through screen, or increments an alphanumerical
value when entering data.
RANGEDown button: Moves selected (highlighted) menu
item down through screen, or decrements an
alphanumerical value when entering data.
MODE Left button: Moves cursor left across screen.
SETUP Right button: Moves cursor right across screen.
•Adjusts the bar graph
range (2X, 3X, 4X), the
resulting display, and
instrument response
NOTE: Button images are used extensively in Chapter 5. When the MODE, VOL, SETUP, and RANGE buttons
are used as “arrow” buttons (to explain navigation and value adjustment procedures), each button’s
corresponding arrow is shown with the button image to illustrate more clearly its function in that particular
context.
3.2.2. Front Panel Buttons
Table 3-1. Front Panel Buttons
3.2.3. Changing Values of Menu Items
Use the arrow buttons to change the values of menu items. After being selected for editing, menu items generally take one
of two forms:
H25-IR PRO Instruction Manual Menu System Overview
NOTE: A timeout monitors front panel keys for regular activity. After approximately 60 seconds of no key
presses, the H25-IR PRO returns to the main display (either the SEARCH MODE or MEASURE MODE screen—
whichever was last active).
SELECT GAS TYPE
p R22
R134A
q R404A
± SELECT ESC=QUIT
EDIT ALARM SETPOINT
qp TO CHANGE
10.0 PPM
tu TO MOVE
± SAVE ESC=QUIT
<< ADVANCED SETUP >>
DATE/TIME SET
'
LOG LIMIT
LOOP FACTOR
SERIAL DATA
PROBE BUTTON
qp SELECT ± EDIT
<< ADVANCED SETUP >>
MEASUREMENT LOG
'
DATE/TIME SET
LOG LIMIT
LOOP FACTOR
SERIAL DATA
qp SELECT ± EDIT
In a scrolling list, simply use the Up and Down Arrow buttons to select the desired item from the list. In the SELECT GAS TYPE
example below, use the Up and Down Arrow buttons to select the desired GAS TYPE. The current selection is highlighted and
displayed larger.
In a scrolling numeric value, use the Up and Down Arrow buttons to increase and decrease (respectively) the desired value
from the highlighted position. For numeric values, the Left and Right Arrow buttons move the highlighted numeral to provide
a larger increment/decrement base. For example, in the EDIT ALARM SETPOINT example below, the highlighted numeral is
in the “units” or “ones” position, so the Up and Down Arrow buttons increment the PPM value by 1. To increment or
decrement the value in larger steps, use the Left Arrow button to position the active (highlighted) digit one place to the left
(the “tens” position). Now, the Up and Down Arrow buttons adjust the PPM value by increments (or decrements) of 10
(versus increments of 1).
Figure 3-5. Examples of Value Types
Regardless of value type, once the desired value has been achieved, use the ENT (¿) button to save the new value. To ignore
the changes you have made to this item and to return it to its previous value, simply press the ESC button.
3.2.4. Configuration-Dependent Menu Items
Not all menu items may be available on your H25-IR PRO. The device displays certain menu options based on your particular
hardware. In particular, the choice of probe type that is ordered (standard or Smart probe) determines whether or not the
PROBE BUTTON option is shown in the ADVANCED SETUP menu. Refer to Figure 3-6 below. For detailed information on the
function of the PROBE BUTTON menu item, refer to Section 5.6.8: Probe Button on page 71 and Figure 5-13 on page 72.
Figure 3-6. PROBE BUTTON Menu Item Based on Probe Style Ordered
www.GlobalTestSupply.com
Menu System Overview H25-IR PRO Instruction Manual
Key Combination
Description
+
User Setup Selection (1-4)
+
Locks the Smart Probe keys (from Smart Probe)
Also used to clear the measurement log (from Measurement Log)
Also used to Lock/Unlock access to External Reference menu
+
Show Setup Summary screen (from Startup)
WARNING: Explosion Hazard. Do not operate this instrument in the presence of flammable liquids, vapors or
aerosols. Operation of this instrument in such an environment constitutes a safety hazard.
NOTE: If your H25-IR PRO has the Smart Probe option, its LCD will also show a version number after startup.
Note that the version shown on the Smart Probe is the version of the Smart Probe firmware, while the version
displayed on the main LCD (Figure 3-7) refers to the firmware version of the H25-IR PRO unit.
FIRMWARE
VERSION V1.20
3.2.5. Special Key Combinations
Buttons on the front panel have unique associated functions as explained earlier. Additional functions are available through
the use of key combinations. Such functions are enabled by simultaneously pressing two or more of the front panel buttons.
A summary is shown in Table 3-2.
Table 3-2. Common Key Combinations
3.3. Version Screen (at Power Up)
After the H25-IR PRO is turned on, the FIRMWARE VERSION screen is displayed for approximately 15 seconds. Afterwards,
the SETUP SUMMARY screen is automatically displayed. A sample FIRMWARE VERSION screen is shown in Figure 3-7
H25-IR PRO Instruction Manual Menu System Overview
NOTE: Unless otherwise changed (via the MODE button), the Search Mode screen is the default display of the
H25-IR PRO. That is, inactivity timeouts will automatically return the display to the SEARCH MODE screen, as will
successive presses of the ESC button.
NOTE: Unlike the SEARCH MODE screen which offers two display styles (graphic and numeric), the MEASURE
MODE screen only appears in numeric display style.
SETUP SUMMARY
GAS TYPE: R134A
UNIT DISPLAY: G/YR
ALARM SETPOINT: 10.0
DISPLAY STYLE: GRAPHIC
FIRMWARE REVISION 1.0
59 INITIALIZING…
GAS TYPE = R134A
0 40.0
ALARM >= 20.0 G/YR
SEARCH {---------}
GAS TYPE = R134A
ALARM
>=
20.0
RANGE = 0-40.0 G/YR
SEARCH {---------}
3.4. Setup Summary Screen (Warm-up)
The SETUP SUMMARY screen is displayed automatically after the FIRMWARE VERSION screen. This screen displays key setup
information (shown in Figure 3-8) and a 60-second initialization countdown. After the 60-second countdown, the H25-IR PRO
displays the SEARCH MODE screen and is ready for operation.
After the SETUP SUMMARY screen and the 60-second initialization, the H25-IR PRO automatically displays the SEARCH MODE
screen. Search mode is used to locate the source of a refrigerant leak.
The SEARCH MODE screen contains the following key elements:
• Gas reading in bar graph or numerical format based on the selected display style (see Figure 3-9)
• Gas type shown in the banner line of the SEARCH MODE screen
• Mode shown in the footer line of the SEARCH MODE screen
• Active USER SETUP name 1, 2, 3, 4, or “----------“ if none selected
Menu System Overview H25-IR PRO Instruction Manual
NOTE: Unlike the Search Mode screen which offers two display styles (graphic and numeric), the Measure Mode
screen only appears in numeric display style.
NOTE: Use the MODE button to toggle between the SEARCH MODE screen and the MEASURE MODE screen.
GAS TYPE = R134A
ALARM
>=
10.0
RANGE = 0-20.0 G/YR
MEASURE {---------}
BASIC SETUP >>
GAS TYPE: R134A
UNIT DISPLAY: G/YR
ALARM SETPOINT: 10.0
DISPLAY STYLE: GRAPHIC
qp SELECT ± EDIT
3.6. Measure Mode Screen
After locating a gas leak using the SEARCH MODE screen, switch to the MEASURE MODE screen to measure the extent of that
leak. Use the MODE button to toggle between the MEASURE MODE screen and the SEARCH MODE screen.
Figure 3-10. Sample Measure Mode Screen
The MEASURE MODE screen contains the following key elements:
• Gas reading in numerical format
• Gas type shown in the banner line of the SEARCH MODE screen
• Mode shown in the footer line of the MEASURE MODE screen
• Active USER SETUP name 1, 2, 3, 4, or “----------“ if none selected
• Programmed alarm setpoint
• Range display shown below the gas reading.
3.7. Basic Setup Screen
The Basic Setup screen contains configuration settings that you define before searching for or measuring a gas leak. To
access the Basic Setup screen from the Search Mode screen or Measure Mode screen, press the SETUP button.
H25-IR PRO Instruction Manual Menu System Overview
NOTE: The right arrows in the header of the BASIC SETUP screen indicate the existence of an additional “tabbed”
menu to the right (and the key to press in order to access it).
<< ADVANCED SETUP >>
USER SETUPS 1-4
RELAY SETTINGS
MEASUREMENT LOG
DATE/TIME SET
LOG LIMIT
¯
qp SELECT ± EDIT
<< ADVANCED SETUP >>
DATE/TIME SET
'
LOG LIMIT
LOOP FACTOR
SERIAL DATA
PROBE BUTTON
qp SELECT ± EDIT
The BASIC SETUP screen contains the following key elements:
• Menu identifier located in the banner line (BASIC SETUP in this example)
• Right arrows in the banner line indicate another menu to the right (use the right arrow key to access)
• GAS TYPE menu item (selects the type of refrigerant gas you are checking)
• UNIT DISPLAY menu item (selects your preference for gas leak rate from a list of standard units)
• ALARM SETPOINT menu item (used to define the gas leak rate setpoint above which an alarm sounds)
• DISPLAY STYLE menu item (selects your preference of graphic or numeric display for the SEARCH MODE screen)
• Command options in the footer line (in this case, the keys to press to select and edit menu items)
To select an item from the BASIC SETUP, scroll through the menu using the Up and Down Arrow buttons until the desired
menu item is highlighted and then press the ENT button to select the item for possible editing.
3.8. Advanced Setup Screen
The Advanced Setup screen contains configuration options that allow you to take advantage of some of the more complex or
infrequently-used features of the H25-IR PRO. To access the Advanced Setup screen from the Search Mode screen or Measure
Mode screen, press the SETUP button once to reach the Basic Setup menu, then press the SETUP button a second time (as
the Right Arrow button).
The Advanced Setup screen contains more menu items than will fit on the front panel display, so small up and down arrows
on the far right side of the menu items indicate additional menu items. Use the Up and Down Arrow buttons on the front
panel to scroll through the list of menu items. All of the menu items from the Advanced Setup screen are shown in the two
images of Figure 3-12. Note the arrows indicating additional menu items in the scrollable list.
Figure 3-12. Advanced Setup Screen
The Advanced Setup screen contains the following options.
• User Setups 1-4 (optional user-based pre-sets)
• Relay Settings (conditional controls for the four output relays)
• Measurement Log
• Date/Time Set
• Log Limit (an upper limit logged leak rate value, above which one or more output relays can be activated)
• Loop Factor (current loop scaling for the 4-20mA output)
• Serial Data (None, Text Packet, or Text Only option for RS-232 communications)
• Probe Button (Only present if Smart Probe is not auto-detected. Selects function of button on standard probe.)
• Menu identifier located in the banner line (Advanced Setup in this example)
• Right arrows in the banner line indicate another menu to the right (use the Right Arrow button to access)
• Left arrows in the banner line indicate another menu to the left (use the Left Arrow button to access)
• Command options in the footer line (in this case, the keys to press to select and edit menu items)
Menu System Overview H25-IR PRO Instruction Manual
NOTE: The right arrows in the header of the Advanced Setup screen indicate the existence of an additional
“tabbed” menu to the right (and the key to press in order to access it). Similarly, the left arrows show how to
return to the Basic Setup screen.
NOTE: The function of the probe button located on the standard probe can be set through the Probe Button
setting of the Advanced Setup menu. The Probe Button setting (see Figure 5-10 above) is only present (at the
bottom of the Advanced Setup menu) if a Smart Probe is not automatically detected. Refer to section 5.6.8:
Probe Button (on page 71) for more information.
NOTE: Pressing the ENT button when the Current Faults menu item is selected interrogates the faults and
provides you with fault details.
<< DIAGNOSTICS
CURRENT FAULTS 0000
SENSOR DATA
qp SELECT ± EDIT
3.9. Diagnostics Screen
The Diagnostics screen contains advanced system details used for identifying faults, troubleshooting, and maintenance. To
access the Diagnostics screen from the Search Mode screen or Measure Mode screen:
•Press the SETUP button once to reach the Basic Setup menu, then press the SETUP button two more times (as the
Right Arrow button), or
•Press the Info (i) button.
Figure 3-13. Diagnostics Screen
The Diagnostics screen contains the following options:
• Current Faults menu (lists currently active fault codes) with 4-digit fault code summary)
• Sensor Data menu (parameters for the NDIR sensor, used for troubleshooting and maintenance)
• Menu identifier located in the banner line (DIAGNOSTICS in this example)
• Left arrows in the banner line that indicate one or more additional menus (use the left arrow key to access)
• Command options in the footer line (in this case, the keys to press to select and edit menu items)
• Standard Versus Smart Probe Wiring ........................ 32
NOTE: The external connections and optional hardware devices discussed in this section are NOT required for
the operation of the H25-IR PRO. You may skip any sections that do not apply to your particular configuration of
the H25-IR PRO.
CHAPTER 4: WIRING
4.1. Introduction
This section contains information on wiring the following ports to optional hardware devices:
• Relays to control external alarm devices such as horns and strobes
• 4-20mA current output to control an external device such as a chart recorder
• RS-232 communications port to export leak data to a computer.
4.2. Relay Connections
The H25-IR PRO includes four single pole double throw (SPDT) relay contacts. These relay contacts can be used to control
external alarm devices such as:
• Horns
• Bells
• Strobe lights.
The external alarm devices can be connected to these relays and activated when their associated relay is energized. Each
relay contact has a 2 A rating at 250 VAC (inductive) and 5 A at 250 VAC (resistive).
Locate the external relay connector on the rear of the unit. Secure the leads from the external device and its power source
to the external relay connector as shown in the example diagram below. This diagram shows a typical alarm device being
connected to Relay 1. Repeat this procedure, as necessary, for the remaining relays.
NOTE: Be sure to configure the relays in the ADVANCED SETUP menu. Refer to Relay Settings on page 61.
NOTE: Ensure the circuit load does not exceed the relay ratings.
IMPORTANT: The monitoring device must be isolated from ground (i.e., a floating ground). For non-
isolated equipment, a loop-isolator must be between the H25-IR PRO and the monitoring device.
Bacharach recommends DataForth model DSCL21 at www.dataforth.com.
NOTE: A shorting wire must be connected to the 4–20 mA connector if no external monitoring device is used.
Failure to install this resistor will cause a “Loop Open Error” to occur.
CAUTION: Never apply power to the 4–20 mA Loop Connector from an external power supply. In addition,
connect only a load resistor or a floating measurement device.
NOTE: Wiring to the monitoring device should consist of 24 AWG, twisted pair up to 2000 feet in length.
• Each relay may be connected as normally open (NO) or normally closed (NC).
• The relay contacts are rated 2 A at 250 VAC (inductive) and 5 A at 250 VAC (resistive).
• Power for the external alarm devices is supplied from an external power source.
• Jumper the “neutral” line of an external power source to the “common” terminals on the relay connector.
• Connect one end of the alarm device to the appropriate terminal of the desired relay (NO or NC).
• The other end of the alarm device is connected to the other leg of the external power source. For protection, install
an in-line fuse of the appropriate size and design for the external alarm device being used.
4.3. 4-20mA Current Loop Connection
An external 4-20 mA monitoring device (e.g., chart recorder) can be connected to the H25-IR PRO using a shielded-twistedpair cable. Locate the 4-20 mA connector on the rear of the unit, and then remove the resistor or shorting wire from its
terminals.
Secure the wire leads from the external monitoring device to the connector as shown in the diagram below, making sure that
the polarity at this connector matches the wiring at the monitoring device.
NOTE: Be sure to configure/verify the Loop Factor setting (see Loop Factor on page 68).
NOTE: Be sure to configure your H25-IR PRO to send the ASCII data in the appropriate format (text packet or text
only). Also, be sure that the communications settings on your computer match those listed below.
Setting
Value
Baud Rate
19200
Parity
None
Stop Bit(s)
1
Flow Control
None
NOTE: USB-to-Serial adapters may work, but are unsupported.
WARNING: SHOCK HAZARD. When performing this procedure, turn OFF the instrument and disconnect its
AC power cord.
The default current-to-ppm factor is set to 0.016 mA = 1 ppm, providing a measurement range of 0 ppm (4 mA) to 1,000 ppm
(20 mA). Note that the current-to-ppm factor can be changed using the instrument’s Loop Factor function. Refer to Loop
Factor on page 68.
4.4. RS-232 Communications Connection
Leak rate data can be output as ASCII text via the 9-pin RS-232 connector on the back panel of the H25-IR PRO. Using a
standard RS-232 cable, connect the H25-IR PRO to a computer. To export ASCII formatted live data to a computer, you need:
• Serial communications software package (e.g., HyperTerminalä)
• Personal computer with an available COM port and running Windows® 95/98/2000/NT/XP/Vista/7 or later
• 9-pin-female to 9-pin-male serial cable (for example, P/N 104-4027)
Table 4-1. RS-232 Communications Settings
4.5. Standard Versus Smart Probe Wiring
When the H25-IR PRO is shipped from the factory, it has been configured to use the probe & hose assembly that was supplied
with the instrument.
If a different probe style is used, then the ribbon cable from the front panel PROBE connector must be reconnected to either
J2 or J7 of the LCD printed circuit board, located behind the front panel, as follows.
Remove both the front and rear bezels by first disengaging the plastic tabs on the right and left sides of the
bezel, and then pulling the bezel straight out.
3
Remove the top two screws that secure the front panel, and the top two screws that secure the rear panel.
4
Find the seam at the side of the enclosure nearest the top. Using your fingers gently pry up on one side of the
enclosure lid. The lid will disengage and swing upward, leaving the opposite side of the lid engaged. With the lid
freely moving in the opposite side, gently lift up the front and rear edges of the lid to fully disengage it. Set it to
one side and continue with the replacement.
5
Connect the ribbon cable from the front panel PROBE connector to either J2 or J7 according to the probe to be
connected:
• J7 – Standard probe with LED and push button (and Standard probe with flex tubing)
• J2 – Smart probe
IMPORTANT: Note the orientation of Pin 1 on both the ribbon cable and connectors J2 and J7.
The red stripe on the ribbon cable denotes pin 1.
6
Reassemble the case, then connect the power cord.
Table 4-2. Procedure to Change Factory-Set Probe Configuration
Figure 4-3. Front Panel Printed Circuit Board Containing Probe Connectors J2 and J7
• Adjustment to a Known External Leak Source ........... 81
• Custom Gas Type Configuration ................................ 83
• Factory Gas Type Configuration ................................ 84
WARNING: Explosion Hazard. Do not operate this instrument in the presence of flammable liquids, vapors
or aerosols. Operation of this instrument in such an environment constitutes a safety hazard.
NOTE: Button images are used extensively in this chapter. When the MODE, VOLUME, SETUP, and RANGE
buttons are used as “arrow” buttons (to explain navigation and value adjustment procedures), each button’s
corresponding arrow is shown with the button image to illustrate more clearly its function in that particular
context.
CHAPTER 5: OPERATION
5.1. Introduction
This section describes the operation of the H25-IR PRO. The instructions in this section assume the following:
• You read and understood the previous sections.
• You assembled and connected your gas probe.
• You connected the power cord to an appropriate power supply receptacle.
• You wired optional external devices to the relays, current output, and/or RS-232 computer connector as desired.
For basic applications, your H25-IR PRO is ready to use after you turn it on. If your application takes advantage of one or
more of the optional features, you should review the settings to ensure that any customizations you desire are properly
configured. Note that some features require both hardware configuration (e.g., wiring from relay outputs to external
notification devices) and firmware configuration (e.g., setup of menu parameters).
Table 5-1 on page 35 shows common application options and where to look for configuration information.
Ensure appropriate wiring has been completed and probe is connected (see Hardware Overview on page 11 and
Wiring on page 30).
2
Flip power switch (see Figure 2-2 on page 12) to the “ON" ( | ) position and wait for the Search Mode screen to
appear.
NOTE: Unless otherwise changed (via the MODE button), the Search Mode screen is the default display of the
H25-IR PRO. That is, inactivity timeouts will automatically return the display to the Search Mode screen, as will
successive presses of the ESC button.
GAS TYPE = R134A
0 40.0
ALARM >= 20.0 G/YR
SEARCH {---------}
GAS TYPE = R134A
ALARM
>=
20.0
RANGE = 0-40 G/YR
SEARCH {---------}
5.2. Turning on the H25-IR PRO
Table 5-2. Turning on the H25-IR PRO
5.3. Search Mode Screen
5.3.1. Overview
After the SETUP SUMMARY screen and the 60-second initialization, the H25-IR PRO automatically displays the SEARCH MODE
screen. Search mode is used to locate the source of a refrigerant leak.
The SEARCH MODE screen contains the following key elements:
• Gas reading in bar graph or numerical format based on the selected display style (see Display Style on page 51)
• Gas type shown in the banner line of the SEARCH MODE screen (see Gas Type on page 46)
• Mode shown in the footer line of the SEARCH MODE screen (also see Measure Mode Screen on page 42)
• Active USER SETUP name 1, 2, 3, 4, or “----------“ if none selected (see User Setups 1-4 on page 54)
• Programmed alarm setpoint (see Alarm Setpoint on page 49)
• Graphic or numeric range display based on the display style (see Display Styles (Graphic vs. Numeric) on page 36).
5.3.2. Display Styles (Graphic vs. Numeric)
Based on your desired preference, you can choose between two display styles (see Figure 5-1) for the Search Mode screen:
• Graphic display style (left)
• Numeric display style (right).
The display style is a user preference that is selected in the DISPLAY STYLE option (Display Style on page 51) of the Basic Setup
menu.
During search mode operation, it may be desirable (after locating the source of the refrigerant gas
leak) to quantify or measure the extent of that leak. This is accomplished by switching the H25-IR
PRO from the Search Mode screen to the Measure Mode screen using the MODE button.
NOTE: The MODE button toggles the H25-IR PRO between search mode (locating the
leak source) and measure mode (quantifying the leak). The Search Mode screen and
Measure Mode screen are displayed, respectively.
NOTE: Regardless of the display style that is selected (graphic display or numeric display), numeric display is
always used exclusively in measure mode.
GAS TYPE = R134A
0 20.0
ALARM >= 10.0 G/YR
SEARCH {---------}
GAS TYPE = R134A
ALARM
>=
10.0
RANGE = 0-20 G/YR
MEASURE {---------}
GAS TYPE = R134A
ALARM
>=
10.0
RANGE = 0-20 G/YR
SEARCH {---------}
GAS TYPE = R134A
ALARM
>=
10.0
RANGE = 0-20 G/YR
MEASURE {---------}
5.3.3. Mode Button
In Figure 5-2, notice the differences between the Search Mode screen and the Measure Mode screen. The H25-IR PRO in
Figure 5-2 is configured for graphic display style. The H25-IR PRO in Figure 5-3 is configured for numeric display style. Refer
to Display Style on page 51 for additional information.
Figure 5-2. Toggling Between Search Mode and Measure Mode (Example 1)
Figure 5-3. Toggling Between Search Mode and Measure Mode (Example 2)
The H25-IR PRO can be configured to produce audible tones under the following conditions:
•When the quantity of a sampled gas exceeds a programmed setpoint limit (a steady,
constant-frequency tone)
•When the detected gas level increases (an audible tone whose frequency increases as
the detected level of gas increases).
The audible tones emanate from either the speaker (at the back of the H25-IR PRO unit) or from
headphones (or an amplified speaker) connected to the headphones jack on the back panel of
the unit.
NOTE: An earphone plugged into the instrument’s earphone jack automatically disconnects the speaker and allows
the operator to listen to the unit’s audible tone in high-noise environments. If additional volume is required, connect
an amplified speaker to the earphone jack.
Step
Button
Description
Sample Screen
1
Press the VOL button to display
the Speaker Volume screen.
2
Press the Up Arrow button to
increase volume. Press the Down
Arrow button to decrease
volume. A visual indication of the
volume level is provided by a bar
graph at the bottom of the Data
Display screen.
3 Press ENT to save the new volume
level.
In addition to display style, another option is the ability to change the range for gas readings.
This is accomplished using the RANGE button on the front panel and is useful during search
mode to better monitor larger swings in detected gas. The range feature is best illustrated
using the graphic display style of the Search Mode screen. Note that it contains a hash mark
delineating the alarm setpoint (refer to the graphic display style in Figure 5-1).
By default, the display range maximum is set to twice the alarm setpoint (e.g., 40.0 g/yr for
a 20.0 g/yr alarm setpoint). The RANGE button toggles between 2x (the default), 3x, and 4x
(that is, two times, three times, and four times the alarm setpoint setting), and then returns
to the default 2x range. These examples, shown as 3 successive presses of the RANGE
button, are illustrated in Figure 5-4.
SPEAKER VOLUME
qp ADJUST ESC=QUIT
5.3.4. Volume Button
The VOL button is used to access the Adjust Speaker Volume screen. From there, the volume of the audible tones can be increased
or decreased (using the Up Arrow and Down Arrow buttons). Adjust the speaker (and earphones) volume as follows.
NOTE: When using the graphic display style, the maximum range value is shown above the right-most limit of
the bar graph. When using the numeric display style, the maximum range value is shown in text format below
the gas reading. Refer to Figure 5-1.
5.3.6. Setup (and Right Arrow) Button
The SETUP button is used to access and scroll forward through the three main menu screens
from the Search Mode or Measure Mode screen:
• BASIC SETUP screen (see Basic Setup Screen on page 45)
• ADVANCED SETUP screen (see Advanced Setup Screen on page 53)
• DIAGNOSTICS screen (see Diagnostics Screen on page 74)
Refer to these individual screens later in this section for more information.
GAS TYPE = R134A
0 40.0
ALARM >= 20.0 G/YR
SEARCH {---------}
GAS TYPE = R134A
0 60.0
ALARM >= 20.0 G/YR
SEARCH {---------}
GAS TYPE = R134A
0 80.0
ALARM >= 20.0 G/YR
SEARCH {---------}
GAS TYPE = R134A
0 40.0
ALARM >= 20.0 G/YR
SEARCH {---------}
2X RANGE (DEFAULT DISPLAY) 3X RANGE
4X RANGE 2X RANGE
Figure 5-4. Sample 2X, 3X, and 4X Range Displays in the Graphic SEARCH MODE Screen
To provide the ability to scale up the range from the default of 2 times (2X) the alarm SETPOINT, the user may press the
RANGE button. Doing this will cycle through the following scaling:
• First press = 3X scaling
• Second press = 4X scaling
• Third press = return to 2X scaling
The scaling repeats in this loop. This feature operates for both data presentation methods (Graphic and Numeric) and for
both Search and Measure modes. The range values displayed on the screen change accordingly to show the scale. The
placement of the vertical line delineating the alarm setpoint is moved to the left to show the change in scale.
NOTE: From any point in the menu structure, press the ESC button until you return to the SEARCH MODE screen
or the MEASURE MODE screen.
Step
Button
Description
Sample Screen(s)
1 Press the SETUP button once to
access the Basic Setup screen.
2 Press the SETUP button (Right
Arrow button) again to access
the Advanced Setup screen.
3 Press the SETUP button (Right
Arrow button) again to access
the Diagnostics screen.
NOTE: Consider using the Information (i) button as a shortcut to the Diagnostics screen. Refer to the
Information (i) Button section on page 41 for additional information.
BASIC SETUP >>
GAS TYPE: R134A
UNIT DISPLAY: G/YR
ALARM SETPOINT: 10.0
DISPLAY STYLE: GRAPHIC
qp SELECT ± EDIT
<< ADVANCED SETUP >>
USER SETUPS 1-4
RELAY SETTINGS
MEASUREMENT LOG
DATE/TIME SET
LOG LIMIT
¯
qp SELECT ± EDIT
<< DIAGNOSTICS
CURRENT FAULTS 0000
SENSOR DATA
qp SELECT ± EDIT
Table 5-4. Accessing the Main Menu Screens Using the Setup Button
The ZERO button manually resets the current gas level to zero. It is also used as the ENT button when
changing configuration values within the menus.
NOTE: Automatic zeroing occurs at startup during initialization. It is therefore
recommended to perform startup of the H25-IR PRO in a refrigerant-free clean-air
environment.
NOTE: In Search Mode, the H25-IR PRO performs an auto zero process to remove the
effects of background refrigerant in the air. Measure Mode does not automatically zero.
Figure 5-5 and Figure 5-6 show a sample Search screen and Measure screen both before and shortly
after the ZERO button is pressed.
The Information (i) button provides a shortcut to the DIAGNOSTICS screen. For more information
on the Diagnostics screen, refer to the Diagnostics Screen section on page 74. Faults and other
diagnostics are displayed in the header portion of the main screen. The i symbol is shown along
with any fault condition to prompt the user to use the i keypad button.
NOTE: The Information button is also used to setup a leak reference. For more information, refer to the sections
on Adjustment to a Known External Leak Source (on page 81) and Using a Reference Leak Source (on page 6).
GAS TYPE = R134A
0 20.0
ALARM >= 10.0 G/YR
SEARCH {---------}
GAS TYPE = R134A
ALARM
>=
10.0
RANGE = 0-20 G/YR
MEASURE {---------}
GAS TYPE = R134A
0 20.0
ALARM >= 10.0 G/YR
SEARCH {---------}
GAS TYPE = R134A
ALARM
>=
10.0
RANGE = 0-20 G/YR
MEASURE {---------}
5.3.7. Zero (ENT) Button
Figure 5-5. Sample Search and Measure Screens Before Zeroing
Figure 5-6. Sample Search and Measure Screens Shortly After Zeroing
NOTE: Unlike the SEARCH MODE screen which offers two display styles (graphic and numeric), the MEASURE
MODE screen only appears in numeric display style.
NOTE: Use the MODE button to toggle between the Search Mode screen and the Measure Mode screen.
NOTE: With the exception of the LOG button (discussed in the next section), most of the front panel buttons on
the H25-IR PRO behave similarly in the Measure Mode screen as they do in the Search Mode screen. For addition
information on the ENT (ZERO), Information (i), VOL, SETUP, RANGE, and MODE buttons, refer to the
corresponding subsections in the Search Mode Screen section starting on page 36.
NOTE: In Search Mode, the H25-IR PRO performs an auto zero process to remove the effects of background
refrigerant in the air. Measure Mode does not automatically zero.
GAS TYPE = R134A
ALARM
>=
24.0
RANGE = 0-12 G/YR
MEASURE {---------}
5.4. Measure Mode Screen
5.4.1. Overview
After locating a gas leak using the SEARCH MODE screen, switch to the MEASURE MODE screen to measure the extent of that
leak. Use the MODE button to toggle between the MEASURE MODE screen and the SEARCH MODE screen.
Figure 5-7. Sample Measure Mode Screen
The MEASUREMENT MODE screen contains the following key elements:
• Gas reading in numerical format
• Gas type shown in the banner line of the SEARCH MODE screen (see Gas Type on page 46)
• Mode shown in the footer line of the MEASURE MODE screen (also see Search Mode Screen on page 36)
• Active USER SETUP name 1, 2, 3, 4, or “----------“ if none selected (see User Setups 1-4 on page 54)
• Programmed alarm setpoint (see Alarm Setpoint on page 49)
• Range display shown below the gas reading (refer to RANGE button description in the Search Mode Screen section
While in Measure Mode, the LOG (ESC) button adds the current non-zero gas measurement to the next
data storage location of the measurement log. The measurement log contains 50 data locations in which
leak rate data may be stored. Refer to Figure 5-8 and Table 5-5 on page 44.
NOTE: The H25-IR PRO must be in Measure Mode and must be sampling a non-zero leak
rate in order to log leak rate data to the measurement log.
NOTE: Use the Log Limit setting in conjunction with the programmable relay control settings to configure your
H25-IR PRO for totalizer applications. For example, consider a repetitive compressor test function at an air
conditioner manufacturer for inspecting up to 50 solder joints on each production unit for gas leaks. Consider a
corporate or environmental restriction that limits each unit to a leak rate of not more than 20 grams/yr. The
H25-IR PRO can be configured to measure all 50 points, quantify each leak rate, calculate the total of all 50 leak
rates, and activate an alert device (light, buzzer, alarm) via one or more of the output relays based on a
programmable Log Limit value. The measurement log can then be reset and the test repeated for the next device.
5.4.2. Log (Esc) Button
Figure 5-8. Graphical Representation of the Measurement Log
1 Ensure that the H25-IR PRO is in
Measure Mode. Use the MODE
button if needed.
2 Press the LOG button to capture
the current (non-zero) leak rate.
You are prompted to press the
ENT button to confirm that you
want to store this value in the
measurement log.
Do nothing (i.e., wait for the
inactivity timeout to occur) or
press the ESC button to return to
the main Measure screen.
3 To log the leak rate value, press
the ENT button when
confirmation is requested. An
acknowledgement screen is
displayed for approximately
5 seconds, followed by the main
Measure screen.
NOTE: With the exception of the LOG button (discussed in the next section), most of the front panel buttons on
the H25-IR PRO behave similarly in the Measure Mode screen as they do in the SEARCH MODE screen. For
additional information on the ENT (ZERO), Information (i), VOL, SETUP, RANGE, and MODE buttons, refer to
the corresponding subsections in the Search Mode Screen section starting on page 36.
NOTE: After changing the gas type, the screen heading on the main Measure screen or Search Screen is changed
to reflect the name of the newly selected menu option.
NOTE: Configuration values are retained (even after power is cycled) in non-volatile memory.
NOTE: Configuration settings under the Basic Setup menu can be thought of as “User Setup 0” and remain active
unless one of the four User Setups has been selected. See User Setups 1-4 on page 54 for more information.
NOTE: If “Custom” is selected, the user is expected to use the External Reference feature to adjust the
instrument to his needs. For “Custom” gas configurations, the upper and lower external reference adjustment
bounds must be set by the user. Also, the user is responsible for verifying instrument response if “custom” is
used.
Menu Option
Location, Description
Gases Included
GAS TYPE
Primary List
Presented in the standard menu
(see Table 5-7 on page 46).
When equipped with the halogen
NDIR sensor this list includes the
most popular refrigerant gases; a
“CUSTOM" refrigerant option is
also shown.
CUSTOM: The user adjusts the instrument response to a known
concentration via the “external reference" feature.
Refer to Adjustment to a Known External Leak Source
on page 81.
GAS TYPE
Secondary “Factory” List
Presented only in the Factory
Menu.
When equipped with the halogen
NDIR sensor the instrument may
be set to one of the gases in this
extended gas list. The selected
gas becomes the only gas listed in
the standard Gas Type menu; no
other gases will be listed.
R290, R600a, CO2_A, SF6
NOTE: A dedicated NDIR Sensor is required; if so equipped, the
instrument is dedicated to detect the gas shown in the GAS TYPE
menu; no other gases will be listed.
Refer to Factory Gas Type Configuration on page 84 to set one of
the gases listed above.
NOTE: If your application uses different units than those provided in Table 5-9, use Table 5-10 to convert your
units to one of the available H25-IR PRO units.
From/To
cfm3/s
Pa M3/s
Pa L/s
mbar L/s
Torr L/s
Kg Mole/s
Molecules
of Air/s
Air oz/yr
Air mg/s
cfm3/s
1
0.10
101.33
1.01
0.76
4.46x10-8
2.69x10
-19
1.44x103
1.29
Pa M3/s
9.87
1
100x103
10.00
7.50
4.40x10-7
2.65x1020
1.42x104
12.75
Pa L/s
9.87x10-3
100x10-3
1
.001
7.50x10-3
4.40x10
-10
2.65x1017
14.19
1.28x10-2
mbar L/s
0.99
0.10
100.00
1
0.75
4.40x
10-8
2.65x10
-19
1.42x103
1.28
Torr L/s
1.32
0.13
133.32
1.33
1
5.87x10-8
3.54x1019
1.89x103
1.70
kg Mole/s
2.24x107
2.27x108
2.27x109
2.27x107
1.70x107
1
6.02x1026
3.22x1010
2.90x107
Molecules
of air/s
3.72x1020
3.77x10
-21
3.77x10
-18
3.77x10
-20
2.83x10
-20
1.66x10
-27
1
5.35x10
-17
4.81x10
-20
Air oz/yr
6.96x10-4
7.05x10-5
7.05x10-2
7.05x
10-4
5.29x10-4
3.10x10
-11
1.87x1016
1
8.99x10-4
Air mg/s
0.77
7.84x10-2
78.41
0.78
0.59
3.45x10-8
2.08x1019
1.11x103
1
5.5.2. Unit Display
The unit display function sets the engineering units used in leak flow rate (and other related) calculations. Available display
units are listed in Table 5-9.
Table 5-9. Display Units Available in the H25-IR PRO
Follow the steps in Table 5-6 on
page 45 to access the Display
Units menu option and to display
the Select Display Units screen.
The active selection is twice the
size of the previous and next
items (shown above and below
the active selection).
2
Press the Up Arrow and Down
Arrow buttons to scroll through
the list of display units.
Pressing and holding each button
scrolls through the list as well.
3 Press the ENTER button to
configure the H25-IR PRO for the
selected unit display.
Pressing ENT returns the user to
the previous screen with the
selected value in place.
Pressing the ESC button aborts
the change and maintain the
previous unit display.
NOTE: The default detection range is calculated as 2X the ALARM setpoint. This is done to standardize the
GRAPHIC display mode.
NOTE: The audio output scales to the detected gas concentration when the alarm setpoint is set to 0 in the Basic
Setup menu. Otherwise, its sound level is either ON or OFF (at the volume setting that was defined by the user),
using the alarm setpoint as the trigger for activation. Similarly, if User Setup 1-4 is used, the alarm can be set to
“continuous” or “setpoint”.
SELECT DISPLAY UNITS
p -----
oz
/YR
q g/YR
± SELECT ESC=QUIT
SELECT DISPLAY UNITS
p OZ/YR
g
/YR
q mL/SEC-5
± SELECT ESC=QUIT
BASIC SETUP >>
GAS TYPE: R404A
UNIT DISPLAY: G/YR
ALARM SETPOINT: 10.0
DISPLAY STYLE: GRAPHIC
qp SELECT ± EDIT
Table 5-11. Changing the Unit Display Setting
5.5.3. Alarm Setpoint
The Alarm Setpoint defines a gas leak setpoint value above which the alarm sounds. Below this value there is no alarm.
Setting it to 0 means that the instrument alarms on any gas detected.
Follow the steps in Table 5-6 on
page 45 to access the Alarm
Setpoint menu option and to
display the Edit Alarm Setpoint
screen.
2
Press the Up Arrow and Down
Arrow buttons to increase/
decrease (respectively) the
highlighted position of the value.
Pressing and holding each button
increments (or decrements) the
value of the selected position.
Values rollover to (or from) the
adjacent next highest significant
digit(s).
3
Press the Left Arrow and Right
Arrow buttons to change position
of the selected digit in the
setpoint. The selected digit is the
one that is highlighted.
4 Press the ENT button to configure
the H25-IR PRO to use the defined
setpoint.
Pressing the ENT button returns
the user to the previous screen
with the selected value in place.
Pressing the ESC button aborts
the change and maintain the
previous setpoint.
NOTE: Alarm messages are shown in the header of the main Search/Monitor screen. If 0 is entered for the Alarm
Setpoint, no alarm message will be shown.
BASIC SETUP >>
GAS TYPE: R134A
UNIT DISPLAY: G/YR
ALARM SETPOINT: 12.0
DISPLAY STYLE: GRAPHIC
qp SELECT ± EDIT
EDIT ALARM SETPOINT
qp TO CHANGE
12.0 PPM
tu TO MOVE
± SAVE ESC=QUIT
EDIT ALARM SETPOINT
qp TO CHANGE
20.0 PPM
tu TO MOVE
± SAVE ESC=QUIT
BASIC SETUP >>
GAS TYPE: R404A
UNIT DISPLAY: G/YR
ALARM SETPOINT: 20.0
DISPLAY STYLE: GRAPHIC
NOTE: If you set the alarm setpoint to a non-zero value, the alarm “beep” sounds only when the gas sample
exceeds the alarm setpoint. If you set the alarm setpoint to zero, the alarm will sound for any detected gas, and
will increase in frequency based on the concentration of the gas.
NOTE: Display Style only affects the Search mode display. Measure mode always shows a numeric display.
NOTE: The dark area shown in the bar graph is a real time display of the detected gas. The bar fills to the right
as the gas level increases. When the alarm setpoint is reached or exceeded the speaker sounds.
GAS TYPE = R134A
0 20.0
ALARM >= 10.0 G/YR
SEARCH {---------}
GAS TYPE = R134A
ALARM
>=
10.0
RANGE = 0-20 G/YR
MEASURE {---------}
5.5.4. Display Style
The Display Style setting defines the format of the main Search screen. Options are Graphic and Numeric. You choose your
preference from the Select Display Style screen.
•Graphic
o Bar graph and basic setup information
o The Alarm Setpoint setting is used to establish the bar graph’s mid-point and scaling
o Default setting
• Numeric
o Large digital numeric display of the real-time gas concentration and basic setup information
Figure 5-9. Samples of Graphic and Numeric Display Styles
1 From the main Search screen
or Measure screen, press the
SETUP button once to access
the Basic Setup screen.
2 Press the Setup (Right Arrow)
button again to access the
Advanced Setup screen.
3 Press the Up Arrow (VOL) and
Down Arrow (RANGE) buttons
to highlight the desired menu
item.
Note the small arrow to the
right of the menu indicating
that additional menu items are
available. Use the Down Arrow
button to access these menu
items.
4 Press the ENT button to select
that item for editing.
Note that time is given in
24-hour military format.
NOTE: For detailed instructions on editing menu items in the ADVANCED SETUP screen, refer to the sections
that follow.
BASIC SETUP >>
GAS TYPE: R134A
UNIT DISPLAY: G/YR
ALARM SETPOINT: 10.0
DISPLAY STYLE: GRAPHIC
qp SELECT ± EDIT
<< ADVANCED SETUP >>
USER SETUPS 1-4
RELAY SETTINGS
MEASUREMENT LOG
DATE/TIME SET
LOG LIMIT
NOTE: The function of the probe button located on the standard probe can be set through the Probe Button
setting of the Advanced Setup menu. The Probe Button setting (see Figure 5-10 above) is only present (at the
bottom of the Advanced Setup menu) if a Smart Probe is not automatically detected. Refer to section 5.6.8:
Probe Button (on page 71) for more information.
NOTE: For detailed instructions on editing menu items in the ADVANCED SETUP screen, refer to the sections
that follow.
NOTE: Gas Type, Alarm Setpoint, and Feedback Mode settings for User Setups 1-4 are defined in the User Setup
1-4 menus. User Setup 0 (the default user) is defined through settings in the Basic Setup screen.
NOTE: If you change settings on the H25-IR PRO every time it is used, and those settings may vary greatly from
one application to the next, then the User Setups feature may not offer any significant time savings. Likewise, if
your application is always the same (i.e., same gas type, same alarm setpoint, same preference for display style)
then the User Setups feature may not offer any significant time savings.
However, if your applications have you searching for and measuring the same 2, 3, or 4 gas types, or have you
using the same 2, 3, or 4 alarm setpoints, then setting these once and switching between them very quickly could
be useful. In such cases, the User Setups feature may be very convenient and save you time.
IMPORTANT: Use the ENT+SETUP button combination to activate one of the pre-defined User Setups (1-4).
Change either the Gas Type or Alarm Setpoint value of the User Setup 0 menu items (from the Basic Setup menu)
to cancel the User Setup (1-4) and to begin using User Setup 0 again. In this case, the initial values of Gas Type
and Alarm Setpoint will be from the last active User Setup (1-4), over-writing the original User Setup 0 (Basic
Setup) values. There is no provision for storing or recalling User Setup 0 values.
To avoid potential confusion, use the diagram in Figure 5-11 to determine whether your applications are best
served using either User Setup 0 or User Setup (1-4). Bacharach recommends using either User Setup 0 (Basic
Setup) or User Setup (1-4), but not both. Using, or switching between User Setup 0 and User Setup (1-4) may
cause confusion, since the Basic Setup Gas Type and Alarm Setpoint are over-written whenever a User Setup (1-
4) is activated.
<< ADVANCED SETUP >>
USER SETUPS 1-4
RELAY SETTINGS
MEASUREMENT LOG
DATE/TIME SET
LOG LIMIT
¯
qp SELECT ± EDIT
<< ADVANCED SETUP >>
DATE/TIME SET
'
LOG LIMIT
LOOP FACTOR
SERIAL DATA
PROBE BUTTON
qp SELECT ± EDIT
Figure 5-10. Options of the Advanced Setup Screen
5.6.1. User Setups 1-4
User Setups provide a method of customizing up to four distinct configurations of your H25-IR PRO. You can define up to
four configurations once and then switch between them much quicker than re-configuring those settings. User Setups are
most useful if you change gas types, alarm setpoints, and/or feedback modes frequently.
NOTE: When Feedback is set to SETPOINT, the audio alarm only occurs when the sampled gas level is at or above
the defined Alarm Setpoint value. When set to CONTINUOUS, the audio alarm will sound for any non-zero gas
sample. In CONTINUOUS mode, the pitch of the audio alarm will increase and decrease based on the leak rate
of the sampled gas.
NOTE: To return to the Advanced Setup, press the ESC button.
NOTE: From the main Search or Measure screen, use the ENT + SETUP button combination to activate one of
the pre-defined User Setups (1-4).
NOTE: When a custom setup is active, the name of the setup appears at the bottom of the screen next to the
SEARCH / MEASURE indication.
NOTE: Be sure to read the Relay Connections section in Chapter 4, and properly wire your relay connection(s)
for the desired results.
Step
Button
Description
Sample Screen(s)
1
See Table 5-14
on page 53.
Follow the steps in Table 5-14
on page 53 to access the Relay
Settings option.
2
Press the Up Arrow and Down
Arrow buttons to select the
desired relay (1-4).
<< ADVANCED SETUP >>
USER SETUPS 1-4
RELAY SETTINGS
MEASUREMENT LOG
DATE/TIME SET
LOG LIMIT
¯
qp SELECT ± EDIT
RELAY SETUP
CLOSE 1 NEVER
CLOSE 2 NEVER
CLOSE 3 NEVER
CLOSE 4 NEVER
qp SELECT ± EDIT
5.6.2. Relay Settings Screen
The relay settings screen defines how each of the H25-IR Pro’s four relays function—specifically, when (if) they energize.
If current reading is greater than or equal to the corresponding ALARM SETPOINT value
< SETPOINT
If current reading is less than the corresponding ALARM SETPOINT value
ON FAULT
If a fault occurs (see Fault Codes on page 78)
IN MEASURE
When the H25-IR PRO is in measure mode
AT LOG LIMIT
If the current reading reaches or exceeds the programmed LOG LIMIT value
SETUP 1 ON
User Setup #1 is active
SETUP 2 ON
User Setup #2 is active
SETUP 3 ON
User Setup #3 is active
SETUP 4 ON
User Setup #4 is active
IF TESTING
The relay is closed if the instrument is operating normally (Search or Measure modes).
The relay opens if the menu system is accessed or the instrument experiences a fault.
NOTE: Logging is only possible when the H25-IR PRO is in Measure mode.
NOTE: Use the SETUP button to clear individual log points.
NOTE: Use the MODE+SETUP button combination to clear the entire Measurement Log.
Table 5-21. Relay Setting Options
5.6.3. Measurement Log
The Measurement Log is a 50-element array that contains up to 50 log points that you choose. Each log point contains:
• Log point number (e.g., 1 of 50, 2 of 50, etc.)
• Date and time stamp when the log point was collected
• Measured sample rate at that date and time
• The total of all log data samples (used for totalization applications).
The user can scroll through the available data by using the Up Arrow and Down Arrow buttons. For additional information
on logging data, refer to Figure 5-8 on page43 and Table 5-5 on page 44.
5 Press the ENT button to save
the new date and time and
return to the previous User
Setup screen.
NOTE: The Date and Time are used in the Log Point data of the Measurement Log.
NOTE: For additional information on logging data and totalization applications, refer to Figure 5-8 on page 43
and Table 5-5 on page 44.
NOTE: For relay configuration information, refer to the AT LOG LIMIT option in Table 5-21 (on page 63).
Step
Button
Description
Sample Screen(s)
1
See Table 5-14
on page 53.
Follow the steps in Table 5-14
on page 53 to highlight the
Log Limit option.
SET DATE & TIME
qp TO CHANGE
12/09/11 09:30
tu TO MOVE
± SAVE ESC=QUIT
<< ADVANCED SETUP >>
USER SETUPS 1-4
RELAY SETTINGS
MEASUREMENT LOG
DATE/TIME SET
LOG LIMIT
¯
qp SELECT ± EDIT
5.6.5. Log Limit
The Log Limit value is used in totalization applications and in conjunction with the relay outputs. Every time a measurement
is logged, its value is accumulated (until the Measurement Log is reset). Relay outputs can be energized if this accumulated
value ever exceeds a programmed log limit. The Log Limit is that setpoint.
Note that the unit of measure
is set in the Basic Setup menu,
and is not independently
configurable.
3
Press the Up Arrow and Down
Arrow buttons to change the
value of the selected digit
(note that increasing the value
will increment the next
significant digit).
Press and hold the button to
scroll quickly through the list.
4
Press the Left Arrow and Right
Arrow buttons to select a
different digit to change. The
selected element is the one
that is highlighted.
Repeat these last 2 steps until
the desired date and time are
properly set.
5 Press the ENT button to save
the defined Log Limit and
return to the previous User
Setup screen.
EDIT LOG LIMIT
qp TO CHANGE
099.0 PPM
tu TO MOVE
± SAVE ESC=QUIT
EDIT LOG LIMIT
qp TO CHANGE
125.0 PPM
tu TO MOVE
± SAVE ESC=QUIT
EDIT LOG LIMIT
qp TO CHANGE
525.0 PPM
tu TO MOVE
± SAVE ESC=QUIT
<< ADVANCED SETUP >>
USER SETUPS 1-4
RELAY SETTINGS
MEASUREMENT LOG
DATE/TIME SET
NOTE: The Loop Factor setting configures the 4-20mA current output for the sample range. By default, the
standard 4-20mA output is mapped linearly to the input reading range (for example, with “ppm” set as the unit
of measure, and a range of 0 ppm to 1000 ppm, each of the 1000 ppm steps corresponds to the default
0.016 mA). This can be adjusted to your particular application using the Loop Factor setting and the following
equation.
𝐋𝐨𝐨𝐩%𝐎𝐮𝐭𝐩𝐮𝐭%𝐢𝐧%𝐦𝐀 =(𝐋𝐨𝐨𝐩%𝐅𝐚𝐜𝐭𝐨𝐫% × %𝐏𝐏𝐌)+ %𝟒𝐦𝐀%%
Output
PPM (Default)
Output
PPM (Default)
4 mA DC
0 PPM
13 mA DC
563 PPM
5 mA DC
63 PPM
14 mA DC
625 PPM
6 mA DC
125 PPM
15 mA DC
688 PPM
7 mA DC
188 PPM
16 mA DC
750 PPM
8 mA DC
250 PPM
17 mA DC
813 PPM
9 mA DC
313 PPM
18 mA DC
875 PPM
10 mA DC
375 PPM
19 mA DC
937 PPM
11 mA DC
438 PPM
20 mA DC
1000 PPM
12 mA DC
500 PPM
Step
Button
Description
Sample Screen(s)
1
See Table 5-14
on page 53.
Follow the steps in Table 5-14
on page 53 to highlight the
Loop Factor option.
<< ADVANCED SETUP >>
DATE/TIME SET
'
LOG LIMIT
LOOP FACTOR
SERIAL DATA
PROBE BUTTON
qp SELECT ± EDIT
5.6.6. Loop Factor
Table 5-25. Default Loop Factor Mapping Example (0.016 mA = 1 PPM)
2 Press ENT to view the Current
Loop Factor screen. The
default is set to 0.016.
3
Press the Up Arrow and Down
Arrow buttons to change the
value of the selected digit
(note that increasing the value
will increment the next
significant digit).
Press and hold the button to
scroll quickly through the list.
4
Press the Left Arrow and Right
Arrow buttons to select a
different digit to change. The
selected digit is the one that is
highlighted.
Repeat these last 2 steps until
the desired Current Loop
Factor is set.
5 Press the ENT button to save
the Current Loop Factor and
return to the previous User
Setup screen.
• A 16-character ASCII string is output approximately 10 times per second.
• The string is formatted as follows:
<STX> XX.X,-p.pp<ETX>cc
where XX.X is the leak rate as displayed on the front panel
-p.pp is the internal pressure drop in PSI
cc is a two-byte checksum.
Text Only
• A 10-character ASCII string is output approximately 10 times per second.
• The string is formatted as follows:
XX.X,-p.pp <CR><LF>
where XX.X is the leak rate as displayed on the front panel
-p.pp is the internal pressure drop in PSI
CR & LF stand for carriage return and line feed
•This option is convenient for capture to file using HyperTerminal or similar software.
NOTE: Be sure to read the RS-232 Communications Connection section in Chapter 4 (page 32), properly connect
a computer, and configure data capture utility software. See RS-232 Communications Settings on page 32.
Step
Button
Description
Sample Screen(s)
1
See Table 5-14
on page 53.
Follow the steps in Table 5-14
on page 53 to access the Serial
Data option.
<< ADVANCED SETUP >>
DATE/TIME SET
'
LOG LIMIT
LOOP FACTOR
SERIAL DATA
PROBE BUTTON
qp SELECT ± EDIT
5.6.7. Serial Data
The Serial Data option provides an automated, high-speed data logging tool. When enabled, the H25-IR PRO sends leak rate
data (in ASCII text format) to the RS-232 port 10 times per second. The Serial Data option defines the format of the text data:
Press the Up Arrow and Down
Arrow buttons to scroll
through the Serial Output Type
options.
5 Press the ENT button to select
that output type.
NOTE: The buttons on the instrument’s front panel and the Smart Probe are mapped the same and can be
used inter-changeably.
NOTE: The function of the probe button can be set through the Probe Button setting of the Advanced Setup
menu. The Probe Button setting (see Figure 5-10 above) is only present (at the bottom of the Advanced
Setup menu) if a Smart Probe is not automatically detected. See Figure 5-13 below.
Options
Function of Button on Standard Probe
DISABLED
Button is disabled
MODE KEY
Button toggles mode function between Search and Measurement modes
ENTER KEY
Button has the same function as the ENTER (¿) key on the front panel
SERIAL OUTPUT TYPE
p ---
NONE
q TEXT PACKET
± SELECT ESC=QUIT
SERIAL OUTPUT TYPE
p NONE
TEXT PACKET
q TEXT ONLY
± SELECT ESC=QUIT
5.6.8. Probe Button (Optional)
The H25-IR PRO supports three types of probes:
• Standard Probe with Flex Tubing
• Standard Probe with LED and Push Button
• Smart Probe.
The Smart Probe contains buttons that function the same as the buttons on the front panel. The standard probe (with LED
and push button) contains a single programmable button. The Probe Button option defines the function of that single button.
IMPORTANT: The H25-IR PRO is factory configured to use the probe with which it is purchased. This factory
configuration includes internal wiring connections. If you plan to use a different probe than was shipped with
your H25-IR PRO, you must move an internal cable to a different connector. For additional information, refer
to Standard Versus Smart Probe Wiring on page 32.
<< ADVANCED SETUP >>
DATE/TIME SET
'
LOG LIMIT
LOOP FACTOR
SERIAL DATA
PROBE BUTTON
qp SELECT ± EDIT
<< ADVANCED SETUP >>
DATE/TIME SET
'
LOG LIMIT
LOOP FACTOR
SERIAL DATA
qp SELECT ± EDIT
Figure 5-13. Advanced Setup Screen Without (Left) and With (Right) a Detected Smart Probe
1 Access the main Search or
Measure screen by pressing
the ESC button.
Check the header for a fault
notification.
Note that multiple faults may
exist, but only the first is
displayed.
2 Press the Info (i) button
when the i icon appears in
the header of the main Search
or Measure screen. The
Current Faults option of the
Diagnostics menu is
highlighted.
3 Press the ENT button to select
the Current Faults option and
display the first fault.
4
Press the Right Arrow button
to view the next fault (if one
exists).
i 0810: PRESSUREOUTOFRANGE
ALARM
>=
10.0
RANGE = 0-20 G/YR
MEASURE {---------}
<< DIAGNOSTICS
CURRENT FAULTS 0810
SENSOR DATA
qp SELECT ± EDIT
FAULT 0800 >>
PRESSURE TRANSDUCER OUTPUT
IS OUT OF RANGE. CHECK
TUBING.
<< FAULT 0010
ERROR 0010: 4-20 MA
CURRENT LOOP IS OPEN,
OR CIRCUIT HAS HIGH
RESISTANCE.
CHECK JUMPER OR WIRES
¯
5.7. Diagnostics Screen
Table 5-31. Accessing the Diagnostics Screen (Method #1)
NOTE: A fault code of 0000 indicates that there are no current faults.
Step
Button
Description
Sample Screen
1
See Table 5-32
on page 75.
Follow the steps in Table 5-32
on page 75 to access the
Current Faults.
2
Press the ENT button to view
details of the current faults.
3
Press the Right Arrow button
to view the next fault (if one
exists).
Use the Left Arrow button to
return to the previous fault.
NOTE: The Info (i) button can also be used to access the Diagnostics Menu and the CURRENT FAULTS screen.
NOTE: The Info symbol (i) may also be displayed at the banner line on the display when a maintenance
condition warranting the users attention occurs.
NOTE: The 4-digit fault code is an additive code that can represent multiple fault conditions. The first detected
fault is shown on the Search/Measure screen. The Diagnostics screen shows the fault “total”.
<< DIAGNOSTICS
CURRENT FAULTS 0810
SENSOR DATA
qp SELECT ± EDIT
FAULT 0800 >>
LOW AIR FLOW.
CHECK TIP FILTER
AND REPLACE, CHECK
HOSES FOR BLOCKAGE
<< FAULT 0010
4-20MA CURRENT LOOP
IS OPEN OR CIRCUIT
HAS HIGH RESISTANCE.
CHECK JUMPER OR WIRES
ON REAR OF UNIT
4 Press ENT to select the desired
reference gas type. This displays
the reference screen.
At this point, hold the probe to the
reference leak source.
The timeout for this screen is 5
minutes.
5
Press the Up Arrow and Down
Arrow buttons until the displayed
(sampled) leak rate matches the
leak rate of your external leak
source certificate.
As the leak rate changes, so does
the adjustment coefficient (1.075x
in this case).
IMPORTANT: While using this feature, the user is responsible for ascertaining the temperature of the leak source
and compensating for the temperature-induced change to the leak reference output.
Step
Button
Description
Sample Screen(s)
6
Press the ENT button when the
display rate and leak source rate
match.
You are then prompted to Lock the
Adjustment.
7
Press the ESC button to exit without
locking.
Note the crosshair icon that appears
on the main screen indicating that
an unlocked adjustment is in place.
Simultaneously press the MODE
button and the SETUP button to lock
the adjustment. This prevents
inadvertent adjustment and
tampering.
Note the icon that appears on the
main screen indicating that a locked
adjustment is in place.
NOTE: When an adjustment is in place and locked, an “ADJUSTMENT IS LOCKED” message will be displayed the
next time the operator enters the External Adjustment option by holding the i button. The user at this screen
must unlock the adjustment by pressing MODE + SETUP for 3 seconds to unlock the adjustment. The operator
has 1 minute to do this before the screen times out. Pressing ESC will cancel the adjustment routine and return
to normal operation.
If the operator correctly unlocks the adjustment, he is presented the normal EXTERNAL REFERENCE start screen,
followed by the gas selection screen. If the adjustment wasn’t locked to begin with, this same screen would be
presented directly. The user may clear the applied adjustment. Any icons set on the main screens are cleared.
Alternatively, the user may continue and set the adjustment to 1.00 to clear the setting.
IMPORTANT: The customer and/or user is responsible for verifying the response of the H25-IR PRO if the custom
gas type option is used.
GAS TYPE = R404A
0 40.0
ALARM >= 20.0 G/YR
SEARCH {---------}
i
EXTERNAL REFERENCE
ADJUSTMENT IS LOCKED
Figure 5-14. Locked Message for Subsequent Attempts at External Adjustment
5.9. Custom Gas Type Configuration
For a custom gas type, select CUSTOM from the Basic Gas List (refer to Gas Type Menus and Selection on page 47), and
configure the gas using the same procedure used to adjust the H25-IR PRO to a known external leak source that contains the
custom gas. Refer to the previous section (Adjustment to a Known External Leak Source) on page 81.
NOTE: The Factory menu has restricted access and is not visible during normal operations. The procedure below
details how qualified operators may access the Factory menu. To return the Factory menu to restricted access
status, simply cycle the power to the H25-IR PRO.
IMPORTANT: Configuring the H25-IR PRO for a sample factory gas type requires (a) an appropriate NDIR sensor
installed in the instrument, and (b) gas type configuration in the Factory menu. Be sure you ordered an H25-IR
PRO with an appropriate NDIR sensor for the corresponding gas you wish to detect.
IMPORTANT: Though setting your factory gas type is very straightforward, it does involve accessing the Factory
menu. Arbitrarily changing settings in the Factory menu or deviating from the procedure listed below may
produce non-standard operation, cause system failures, and void your factory warranty. Use caution when
making changes to the Factory menu.
Step
Button
Description
Sample Screen(s)
1
+
Enable the factory menu. With
the instrument OFF, press and
hold the ENT button while turning
ON the instrument.
You will hear a long beep, then the
Firmware Version screen is
displayed for approximately 15
seconds.
FIRMWARE
VERSION V1.09
5.10. Factory Gas Type Configuration
Gases listed as “factory gas types” are used less frequently than the more common basic gas types listed in Gas Type Menus
and Selection on page 47. Though used less frequently, the large list of supported factory gas types are available as H25-IR
PRO configuration options, but in a special Factory menu (versus the standard Gas Type menu) to minimize the sheer volume
of options in the Gas Type menu.
If you have chosen to use one of these factory gas types (refer to Gas Type Menus and Selection on page 47), the gas type
shown in the Gas Type menu will reflect the your gas choice from the Factory menu. Four NDIR sensors are used to support
the listed factory gas types. These four NDIR sensors support the following factory gas types:
• R600A and R290.
• CO2.
• SF6.
Table 5-38. Configuring the H25-IR PRO to use a Factory Gas Type
The Setup Summary screen is
displayed automatically after the
Firmware Version screen. This
screen displays key setup
information and a 60-second
initialization countdown.
3
n/a
After the 60-second countdown,
the H25-IR PRO automatically
displays the SEARCH MODE screen
and is ready for operation with the
Factory menu enabled.
4 From the main Search screen or
Measure screen, press the SETUP
button once to access the Basic
Setup screen.
(Notice the right arrows in the
header, indicating an additional
menu.)
5 Press the SETUP (Right Arrow)
button again to access the
Advanced Setup screen.
(Notice the right arrows in the
header, indicating an additional
menu.)
SETUP SUMMARY
GAS TYPE: R134A
UNIT DISPLAY: G/YR
ALARM SETPOINT: 10.0
DISPLAY STYLE: GRAPHIC
FIRMWARE REVISION 1.0
59 INITIALIZING…
GAS TYPE = R134A
0 40.0
ALARM >= 20.0 G/YR
SEARCH {---------}
BASIC SETUP >>
GAS TYPE: R134A
UNIT DISPLAY: G/YR
ALARM SETPOINT: 10.0
DISPLAY STYLE: GRAPHIC
qp SELECT ± EDIT
<< ADVANCED SETUP >>
USER SETUPS 1-4
RELAY SETTINGS
MEASUREMENT LOG
DATE/TIME SET
LOG LIMIT
Press the Up Arrow and Down
Arrow buttons to scroll through
the list of factory gases.
Pressing and holding each button
scrolls through the list as well.
11 Press the ENT button to configure
the H25-IR PRO for the selected
gas.
Pressing the ESC button returns
the user to the previous screen
with the selected factory gas type
in place.
Pressing the ESC button while in
the GAS TYPE menu aborts the
change and maintain the previous
factory gas type.
12
®
Cycle power. This returns the factory menu to restricted access.
NOTE: The Factory Gas Type menu includes a “DISABLED” option. When this is selected, the primary gas list is
used in the Gas Type option of the Basic Setup menu. This is the default condition for the halogen NDIR sensor
equipped product.
Inspect “O” ring and, if defective, remove and replace with new “O” ring.
3
Replace metal tip.
4
Unscrew flexible probe from probe housing.
5
Inspect “O” ring and, if defective, remove and replace with new “O” ring.
6
Re-attach flexible probe to probe housing.
Item
Description
1
Replacement filter
Step
Description
1
Turn off power.
2
Grasp the filter housing and rotate the top back and down, twisting the filter out of the mounting clip. Use
caution as the clip has high force. Do not pull the filter away further than 2” (50 mm) from the back of the
instrument. If the tubing becomes taut, move the filter closer to the instrument.
3
With the filter free of the clip, disconnect the top tubing from the fitting.
4
Noting its orientation, install the new filter for proper flow of sample gas (the internal white filter membrane is
at the bottom of the filter assembly).
Figure 6-2. Probe Filter and “O” Ring Locations
6.3. Probe “O” Ring Replacement
Table 6-5. Items Required for “O” Ring Replacement
Table 6-6. Probe “O” Ring Replacement Procedure
6.4. External Filter Replacement
Table 6-7. Items Required for External Filter Replacement