Bacharach 23-9613 Instruction Manual

Page 1
hIstruction 23-9613
Rev. No. 2
January, 1990
:
-..._
L
INSTRUCTION MANUAL
.
Tl!! SNIFFER”
z
.%.
.,
I
A
‘A.-
:
s
WARNING!
Because this instrument is used to detect and monitor materials and conditions which are listed by OSHA or others as potentially hazardous to personnel and property, the information in this manual must be fully understood and utilized to ensure that the instrument is operating properly and is both used and maintained in the proper manner by qualified personnel. An instrument that is not properly calibrated, operated and maintained by qualifiedpersonnel is likely to provide erroneous information, which could prevent
user
awareness
of a potentially hazardous situation for the instrument user, other
personnel and property.
If, after reading the information in this manual, the user has questions regarding the operation, application or maintenance of the instrument, supervisory or training assistance should be obtained before use. Factory assistance is available by tailing (4 12) 963-2000.
Bacharach, Inc.
625
Alpha Drive, Pittsburgh, PA 15238 (412) 963-2000
Printed in U.S.A.
@Registered Trademarks
Page 2
Bacharach, Inc., warrants to Buyer that at the time of delivery this Product will be free from defects in material and manufacture and will conform substantially to Bacharach Inc.'s applicable
specifications. Bacharach's liability and Buyer's remedy under this
warranty are limited to the repair or replacement, at Bacharach's
option, of this Product or parts thereof returned to Seller at the factory of manufacture and shown to Bacharach Inc.'s reasonable satisfaction to have been defective; provided that written notice of the defect shall have been given by Buyer to Bacharach Inc. within one (1) year after the date of delivery of this Product by Bacharach, Inc.
Bacharach, Inc. warrants to Buyer that it will convey good title to
this Product.
Bacharach's liability and Buyer's remedy under this
warranty of title are limited to the removal of any title defects or,
at the election of Bacharach, to the replacement of this Product or part-s thereof that are defective in title.
The warranty set forth in paragraph 1 does not apply to parts the Operating Instructions designate as having a limited shelf-life or as being expended in normal use.
THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN AND ACCEPTED IN LIEU OF (i) ANY AND ALL OTHER WARRANTIES, EXPRESS OR IMPLIED,
INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE: AND (ii> ANY OBLIGATION, LIABILITY, RIGHT, CLAIM OR REMEDY IN CONTRACT OR TORT, WHETHER OR NOT ARISING FROM BACHARACH'S NEGLIGENCE, ACTUAL OR
IMPLIED.
The remedies of the Buyer shall be limited to those
provided herein to the exclusion of any and all other remedies
including,
without limitation incidental or consequential damages. No agreement varying or extending the foregoing warranties, remedies or this limitation will be binding upon Bacharach, Inc. unless in
writing, signed by a duly authorized officer of Bacharach.
NOTE: Fuses,
batteries and sensors are expendable items not covered
by the warranty.
NOTICE
This manual contains proprietary information that
shall not be reproduced or copied in whole or in
part as the basis for manufacture of items herein
described.
Page 3
.
ADDENDUM ,i
Instruction 23-9613
TLV Sniffer@
Rev. No. 2
January, 1990
USE OF DILUTION PROBE 23-7355 WITH TLV SNIFFER Dilution probe 23-7355 has an O-ring covering holes designed to admit fresh
air to dilute, by a factor of 10, any gas sample passing through the probe. With the O-ring moved to expose the dilution holes, the 1O:l dilution ratio
enables the TLV Sniffer to indicate concentrations up to 100,000 ppm, thus
accommodating sample concentrations above the normal 10,000 ppm limit of the
Sniffer.
DETENT WITH 10 TO 1 AIR DILUTION HOLES.
MOVABLE O-RING SHOWN IN AIR-DILUTION
POSITION. SLIDE O-RING TO COVER DILUTION HOLES TO OBTAIN UNDILUTED SAMPLES.
It must be remembered that, when using this probe in the dilution mode, all meter indications should be considered as approximate, only. This is true because of several variable factors which can affect the probe operation,
such as:
1. The actual dilution ratio is dependent upon the flow rate of the sample. The flow rate is affected by battery charge, speed of the pump, cleanliness of the probe,
condition of the probe filter, etc.
2. The actual dilution
ratio
is dependent upon the quality of the air
which provides the dilution.
If the dilution holes are ingesting combustible
gases rather than fresh air,
the ppm indication of the Sniffer will vary significantly from that of an indication obtained with fresh air entering the dilution holes.
WARNING: At the extreme, of course, if the dilution holes and
the probe tip are sampling from the same atmosphere, no dilution actually occurs.
In that instance,
if perfect dilution is assumed
when interpreting meter indications, an error of 1000% is incurred.
Addendum 1 to Instruction 23-9613 Rev.
1,
January 1990
@RegIstered Trademarks
Page 4
REMRMBER:
@ For proper dilution,
only fresh air should enter the dilution holes.
0 Multiply all meter indications by a factor of 10 when using the probe
in the dilution mode.
0 All meter readings in the dilution mode should be considered as
approximate, only.
0 Don't forget to re-cover the dilution holes with the movable O-ring
to obtain undiluted gas samples.
-
-
Bacharach, Inc.
625
Alpha Drive, Pittsburgh, PA 15238 (412) 963-2000
Page 5
.
NICKEL-CADMIUM
BATTERY
MAINTENANCE
DO:
1.
Recharge new NiCads before use.
It is a good idea to cycle them once or
twice before normal operation.
This requires three steps: (1) fully charging
the NiCads, (2) discharging until the unit ceases functioning or indicates low battery power, and (3) fully recharging (12-16 hours).
2. Put NiCads through a charge cycle at least once a month when they are not in use.
NiCads function best when they function frequently.
3:
Prevent
"memory" from occurring by running the NiCads through a full
discharge/charge cycle every week or two (once a month at least).
4.
Match charge time to use on about a two to one ratio (e.g., if you use
the NiCads for four hours
,
.charge them for about eight hours).
DON'T:
1.
Do not use new NiCads without putting them through a full charge.
2.
Do not attempt to use NiCads (or leave the instrument "0N")when the cells
' are fully discharged.
To obtain proper instrument operation, and to prevent
NiCad damage,
the cells must be recharged before further use.
3. Do not overcharge: a.
Never leave NiCads charging for more than 16 hours.
b.
Do not charge over an entire weekend.
4.
Do not attempt to charge other types of batteries in a NiCad charger or
to charge NiCads in a charger other than one designed specifically for NiCads.
Issue Date: October 1981
P/N 51-9480
Page 6
TECH NOTE
EXPLANATION: The Nickel-Cadmium battery (NiCad) can best be thought of as an energy
reservoir.
Provided that it is properly filled with energy, it will-supply
that energy at a useful rate for many applications.
However, underfilling or overfilling the reservoir can cause serious problems, including the destruc­tion of the system.
It is the purpose of this bulletin to explain how NiCads should be mafntained to insure that they will last for the hundreds and even thousands of energy transfers for which they are designed.
Most of the NiCads used in portable instruments are designed to carry an
energy supply sufficient to power the instrument for about 6-8 hours.
Because
of electrical and chemical losses, more energy must be put into the NiCads than is drawn from them as they discharge while being used.
Therefore,
it
normally takes from 12-16 hours of charging time to refill depleted NiCads completely.
If you use an instrument for 6-8 hours a day and then charge it
each night for 12-16 hours , you should obtain hundreds of cycles of NiCad use.
However,
if you use an instrument for varying periods of time each day, or if
you use it only occasionally, then the charging requirements ,become more
complex.
Due to some peculiar characteristics of NiCads,
the manner in &ich they.are
charged is crucial in .detertnining how well and how long they will function.
There are several things that can go wrong with NiCads, and it is important to know what they are:
1.
If NiCads are used when they are almost totally out of energy, they can
be destroyed. This happens because the internal ionic structure of one of the
NiCad cells reverses polarity. Consequently, do not continue to use an
instrument with NiCads when the battery monitoring system indicates low or depleted batteries. The instrument should be turned off and recharged before
being used again.
2. If NiCads are overcharged, they can be destroyed. NiCads work by means of an internal chemical-electrical process, involving interaction of gases and
metals.
When NiCads are charged, gases are built up in the cells. If over-
charging occurs,
the gas pressure in the cell can become too high, causing
' internal damage in the cell and the possibility of an explosion. None of
Bacharach's NiCad systems should be left in a. charging condition for more than 16 hours; and, in general, 12-14 hours of charging should be sufficient. Do
not leave an instrument charging for an entire weekend, as it will almost certainly cause overcharging, a reduction of the NiCad life, and even a safety hazard.
Also, the charge time required
to
refill the NiCad reservoir is re-
lated to how
far
discharged the NiCad is. Thus,
if you have a NiCad system
designed to power an instrument for eight hours,
and you use it for only two
hours, you do not need to recharge it for a full 12-14 hour period. Instead,
3-4 hours of charging should be sufficient to return the unit to a full charge.
3.
NiCads 'lose energy when not in use.
NiCads have a shorter shelf life
than other kinds of batteries.
They will lose 25% of their charge in one
month of non-use, 50% in three months, and 85% in five months. This means two
things:
(1) you should apply a full charge to the NiCads supplied with a new instrument before you use the instrument for the first time, and (2) if an instrument remains idle for one month or more,
it should be charged before
use.
*
9-2
Page 7
TECH NOTE
NiCads tend to develop a "memory." For example, if a set of Nicads is
for three hours every day and then recharged, the internal ionic
4. used structure will eventually tend to "memorize"
that cycle and the NiCads will
readily yield only three hours of service.
Once this "memory" occurs, no
amount of recharging will overcome it.
It will be necessary to discharge the
batteries almost completely (which may take considerably longer than normal)
. and then to recharge them in order to defeat the "memorized" ionic structure.
Sometimes this cycle (depletion/recharge) will have to be repeated two or three times before the "memory" can be overcome.
.
The best way to insure against "memory"
is to fully cycle the NiCads periodi-
cally. That is, once every week or two,
leave the instrument on until the battery indication shows a low reading or a recharge requirement (be careful not to leave the unit on for too long),
and then charge the unit fully over­night (12-16 hours). This will revitalize the entire internal ionic structure of the NiCads and will prevent "memory" from occurring.
9-3
Page 8
a
SECTION 1
CONTENTS
PAGE
iz-
.
DESCRIPTION
PURPOSE AND FUNCTIONS.
...................... l-l
Toxicity Warning Function
...................
l-l
PPMMeter...........................l- 1
.
Search Function
........................ l-1
Continuous Self-Monitoring.
.................. l-l
PRINCIPLE OF OPERATION
...................... l&l
SENSITIVITY RANGES.
......................
.1-2
ACCURACYANDRELIABILITY
.....................
1-2
PHYSICAL DESCRIPTION
...................... : 1-3
Instrument and Case
....................... l-3
Air Sampling Mechanisms
.................... l-3
Combustible Gas Detector.
................... 1-4
AUXILIARY EQUIPMENT.
....................... 1-4
Probe and Hose.
........................ 1-4
Earphone.
...........................
l-4
Batteries
...........................
l-4
Battery Charger
........................
l-5
Recorder.
........................... l-5
CONTROLS .............................
1-5
Mode Selector
......................... l-5
ADJUSTMENTS ............................
1-5
Meter Zero Mechanical Adjustment Screw.
............ l-5
Meter Zero Coarse Adjustment Screw.
..............
l-6
Meter Zero Fine Adjustment Knob
................ l-6
Gain Potentiometer Adjustment Screws.
.............
l-6
Alarm Level Potentiometer Adjustment Screw.
..........
l-6
Recorder Level Potentiometer Adjustment Screw
......... 1-6
.
SECTION 2
2-l
PRE-OPERATION CALIBRATIONS AND ADJUSTMENTS
BATTERY TEST
........................... 2-l
SETTING METER POINTER TO ZERO.
.................. 2-1
SETTING METER POINTER DEFLECTION (GAIN CALIBRATION)
........
2-2-5
RESETTING ALARM RESPONSE
..................... 2-6
SETTING RECORDING.LEVEL.
..................... 2-7
SECTION &
OPERATION
MONITORING TOXICITP. . . . . . . . . . . . . . . . . . . . . . . . 3
-1
Direct Readings and Alarm . . . . . . . . . . . . . . . . . . . 3-l
Converting Hexane-calibrated Meter ppm Readings to ppm
Readings for Other Gases. . . . . . . . . . . . . . . . . . . 3-l
Converting ppm Readings to Percent Lower Explosive Limit. . . . 3-3
LOCATING GAS LEAK SOURCES. . . . . . . . . . . . . . . . . . . . . 3-4
Page 9
SECTION 5
4-l
MAINTENANCE AND REPAIRS
SYMPTOMS ANDREMEDIES (TABLE). . . . . . . . . . . . . . . . . . .
4-1
REPLACING FAULTY COMPONENTS AND PARTS. . . . . . . . . . . . . . :4-5
Replacing Batteries . . . . . . . . . . . . . . . . . . . . . . 4-5
Replacing Detector. . . . . . . . . . . . . . . . . . . . . . . 4-5
Replacing Printed Circuit Board . . . . . . . . . . . . . . . . 4-5
Replacing Sample Pump . . . . . . . . . . . . . . . . . . . . .
4-6
SECTION 5
.5-l
REPLACEMENT PARTS LIST . . . . . . . . . . . . . . . . . . . . . . i .
5-1
ILLUSTRATIONS
FIGURE
l-l 2-l
2-2 2-3
2-4
3-l
3-2
4-l
TABLE 3-l
4-l
Troubleshooting Procedures. . . . . . . . . . . . . . . . . . . 4-l
TLV Circuitry Adjustment Potentiometer Locations.
.......
l-7
Bacharach Code 51-7199 Gas Calibration Kit.
..........
2-3
Calibration Gas Transfer Assembly
...............
2-3
Locating and adjusting R-13 on PPM X 100 Position
.......
2-5
Locating and adjusting R-3 for Certified Gas Indication
....
2-6
Use of Dilution Probe and Hose Assembly and I In-line
Filter and Trap Assembly.
..................
3-2
Conversion Curves Showing Relationship of PPM
Concentrations of Various Gases to Percent
L.E.L. Equivalents. . . . . . . . . . . . . . . . . . . . a . 3-4
TLV Sniffer Chassis Subassembly Showing Miniature
Pump Assembly in Disassembled Position. . . . . . . . . . . . 4-7
TABLES
.
Multiplying Factors for Converting PPM Meter Readings
of Hexane-calibrated Instruments to PPM
Concentrations of Other Gases . . . . c . . . . . . . . . . . 3-3
Page 10
l-l
SECTION 1
DESCRIPTION
.
PURPOSE AND FUNCTIONS Toxicity Warning Function
The portable TLV Sniffer is an extremely sensitive combustible gas and vapor sensing instrument with an overall detecting sensitivity range,of from 1 to 10,000 parts of gas or volatile flammable vapor per million parts of air (by
volume).
The instrument is equipped with an audible (beep) alarm that can be set to sound at any desired level of gas concentration to warn automatically of toxic concentrations of gas.
Connections are provided for
an auxiliary earphone for use in noisy areas to assure that audible warnings
will be heard.
PPM Meter
-­A meter reading directly in parts per million (ppm) allows instant visual
determination of toxicity danger as readings are compared to published
standard Threshold Limit Values. The instrument can be calibrated to read directly in parts per million for any one of many kinds of combustible gases.
Factory calibration is for hexane unless otherwise specified, though readings from other gases and vapors may be interpreted easily by means of reading conversion cur(7es (in OPERATION Section, this manual).
Search Function In addition to its primary function of indicating ppm combustible vapor
concentrations and toxicity levels, an ultrasensitive "search" range of from 1 to 100 ppm makes the TLV Sniffer especially useful for locating hard-to-
find gas leaks. A hand-held probe attached to the instrument can be moved
in the direction indicated by rising meter readings to point out quickly the source of escaping gas.
Continuous ~Self-Monitoring Other functions provided in the TLV Sniffer circuitry for continuous self-
monitoring purposes include a continuous,
audible note of warning in
response to failing batteries and to excessive negative drift in signal
voltage due to malfunctioning detector circuit components.
A 45-second time
delay circuit is incorporated to prevent the sounding of a false alarm due to temporary
voltage
imbalances during thebrief warmup period after the
instrument mode selector switch has been set to operate the instrument.
PRINCIPLE OF OPERATION
-
To detect and measure concentrations of combustible gas in the air, the TLV
Sniffer catalytically oxidizes gas in a pumped-in sample of air by means of a catalyst-coated resistance element.
The resistance of this element
changes with changes in heat that are proportional to the amount of oxidized
gas,
thereby altering the electrical balance of the catalytic element as
Page 11
1-2
compared to the resistance of a reference element.
Both the catalyst-coated
("active") element and the reference element are
incorporated in a Wheatstone Bridge circuit in such a way as to produce an electrical output proportional to their differences in resistance.
Since any changes in air sample temperature and humidity affect both active and reference elements equally,
the electrical signal output is proportional to the concentrations of combustible gas or vapor in the sample of air (expressed involumetric terms as ppm). However,
sudden changes in humidity may affect the zero
reading
on
the
X-l range. The instrument should, therefore, be zeroed at
the same R.H. prevailing during use.
The audible alarm response at the desired gas concentration ppm level is
accomplished by comparing the gas concentration signal level with an internal reference voltage.
Amplification of the difference between signal
and reference voltages will, at a preset level, operate the audible alarm.
SENSITIVITY RANGES
The TLV Sniffer circuitry and meter provide readings from 0 - 10,000 ppm in three range settings.
The first, with MODE SELECTOR switch set at PPM X
100, indicates from 0 to 10,000 ppm
on
the meter. The second range, with
MODE SELECTOR
set
at
PPMX 10, gives readings between 0 to 1000 ppm. The
third range setting,
at MODE SELECTOR position PPM X 1, provides readings
from 0 to 100 ppm.
Each range setting requires an easily made adjustment of
the ZERO ADJUST control knob on the instrument front panel to set the meter
indicating pointer to zero.
Span adjustments for full-scale pointer
deflection within each range are made periodically as necessary by means of
three gain potentiometers located within the instrument.
ACCURACY AND RELIABILITY The extent to which meter readings correspond to actual parts per million of
combustible gases in sampled air depend upon (1) the internal electrical stability of the instrument, (2) proper calibration of the instrument on gas
mixtures of known concentration,
and (3) purity of the air sample used for
meter pointer zero setting.
The electrical stability of TLV Sniffer circuitry is best demonstrated with the MODE SELECTOR switch set
at
PPM X 1. In this most sensitive range of the instrument, all gas signal voltages are multiplied by 100, so that a O­to l-millivolt signal actually drives the meter pointer with the necessary 0 to 100 millivolts to achieve full-scale deflection. Thus a pointer deflection of l/100 sensitivity of the TLV circuitry to minute electrical changes, the meter needle shows practically no deflection as power is applied to the speaker for the audible beep signal.
Given this electrical stability,
any lack of correspondence between actual
parts per million of combustible gas and the meter readings would be much
more Likely to arise from improper calibration of the instrument, or from
setting meter zero in
the
presence of impure air
than
from instability of
the TLV Sniffer circuitry.
Page 12
Calibrated for hexane gas (a relatively high-energy gas) at the factory, the TLV Sniffer should also provide accurate ppm readings for benzene, toluene, and other gases of similar combustion rates. Conversion curves aid
in
correctly interpreting readings from combustible gases that release energy
at different rates. Thus, purity of the air sample used for meter pointer zero settings'is by
far the most important factor in obtaining accurate ppm readings. Ideally, the pointer should be zeroed in the PPM X 1 mode of operation, where
deviations from zero would be more apparent because of maximum deflection.
In this range, however,
extreme sensitivity causes the instrument to respond to the slightest traces of gas. Wisps of cigarette smoke, fumes from passing autos,
and subtle air contaminations from many other sourcesmay
affect the zero setting.
The extent to which zero readings are biased because of air sample impurities will be reflected in less accurate readings of actual gas concentration ppm.
Frequently,
an apparent negative drift of
the meter pointer may be caused by carrying the instrument
to
an area of fresher air after zeroing the meter inadvertently on air that was not as pure.
PHYSICAL DESCRIPTION
Instrument and Case
--
The TLV Sniffer is housed in a sturdy, brushed-aluminum and blue, wrinkle­finished plastic case.
An attractively designed carrying handle serves also as an adjustable support stand and shoulder strap holder. Weighing 5-l/2 lbs.,
the compact 8-x6-l/2-x3-l/4-inch instrument is as convenient to carry
as a transistor radio.
A front panel
,
easily visible with the instrument in
carrying position, contains a meter reading directly in ppm, and control
knob for range selection and zero setting of the indicating meter pointer. The side panels of the instrument provide plugs and connectors for an air
sample probe, earphones, battery charger, and recorder. Removal of ten screws holding the rugged plastic cover to the rigid aluminum case gives access to batteries and calibration adjustment controls located within the interior of the case.
TLV Sniffer Models
The TLV Sniffer is available as a standard model 0023-7350, for use in areas known
to
be free of combustible gases and vapors, or as an intrinsically
safe model 0023-7356,
for use in hazardous areas designated Class I,
Divisions 1 or 2 by the National Electrical Code.
Model 0023-7356 has been
certified to be intrinsically safe by the Factory Mutual Laboratories.
Air Sampling Mechanisms The TLV Sniffer air sampling system consists of a short intake connection
leading directly into an interior aluminum air chamber holding the detector, a six-inch length of tubing connecting the chamber and a miniature sample-
drawing pump, and a three-inch length of tubing to an exhaust port on the
left side of the instrument
*
case.
Page 13
1-4
Combustible Gas Detector
The combustible gas detector consists of an "active" catalytic-coated
-\. resistance element to oxidize combustible gas, and an identical second resistance element without the catalyst coating which provides a "reference"
0
resistance value unaltered by the oxidation of combustible gas.
Since both
operate at approximately equal temperatures,
only changes in gas content of
sampled air cause differences in resistance between the two to produce
4
signals to the meter and gas alarm circuits.
Both active and reference elements are protected within a porous bronze cylinder that plugs into a mounting block attached with four screws to the air chamber block.
The
.
three-pronged mounting block connects inturnto a cylindrical plug with
wires that connect to circuitry on the instrument circuit board.
AUXILIARY EQUIPMENT Probe and Hose, 0023-7243, and Dilution Probe and Hose, 0023-7355
-­A five-foot-long flexible hose and probe assembly, for sampling air at
specific points to find exact locations of gas leaks, attaches to a fitting on the left side of the case by means of a snap-fitting, spring-loaded collar that can be drawn back with one hand, A cottondustfilter within the probe protects the sample drawing lines and chamber from intrusion of dust and dirt.
The dilution probe 0023-7355 has an O-ring covering holes designed to admit air to dilute the sample 10 times. With O-ring moved to expose dilution holes,
the instrument reads up to 100,000 ppm to accommodate
sample concentrations above the normal 10,000 ppm limit.
(See Figure 3-l.)
An in-line filter and trap assembly 23,-7341 is available for use in dust-
or moisture-laden atmospheres.
Earphones An earphone set of lOO-ohm impedance may be attached to an earphone jack
plug on the right side of the TLV Sniffer case wherever noise in an area to be tested obscures the audible signal from the speaker of the instrument. The earphone circuit does not cutout the audible speaker signal when the
earphone is attached, but rather provides a second way to hear the signal
where noise conditions so require. Batteries
For standard model TLV Sniffer:
The portable TLV Sniffer standard model is powered by six D-size dry-cell batteries inserted in tubes and spring clamps within
the
instrument case
at
the
rear.
Though
any
kind of D cells maybe
used, nickel-cadmium rechargeable batteries are recommended for longest
uninterrupted service (six to nine hours of continuous service).
The
instrument circuitry is so designed that an audible tone sounds when battery power falls below that required to sustain gas detection operations.
Page 14
1-5
For intrinsically safe model:
The Intrinsically Safe battery pack is similar to above, but the battery consists of six nickel-cadmium cells connected by welded links to prevent arcing.
The cell assembly is enclosed
0
in a rivet-sealed, high-impact,
plastic
wire-wound resistors that limit current
.-.
conditions to levels below that which
mixtures of combustible gas and air.
Battery Charger
W
housing with protective fuses and
capacity even under short-circuit
would ignite the worst possible
A compact battery charger (230 VAC:
#0023-7353; 115 VAC: #0023-7230) with
a plug-in connector may be attached to a jack provided in the right side of
the instrument case.
For stationary testing, the TLV Sniffer may be
operated with the charger connected.
For portable operations, the charger
may be used to rejuvenate batteries within the instrument overnight, The
intrinsically safe battery packmay be recharged either in or outside the instrument.
Recorder A recorder (range: O-100 mv; impedance:
10,000 ohms or greater) may be connected to the TLV Sniffer for use if variations in combustible gas levels are to be recorded for study or legal data.
Recording is more suitable for
the PPM X 100 and PPM X 10 mode ranges.
The ultrasensitive PPM X 1 "search" range, responsive from 1 to 100 ppm of gas, is not as suitable for an accessory recorder because of the small magnitude of the signal and the potentially rapid fluctuations of the signal within the range in response to
small changes in detected quantities of combustible gas.
CONTROLS Mode Selector The MODE SELECTOR control knob at lower right on the instrument panel
operates a rotating gang switch with an OFF position, a battery test switch, and three range-selector switch positions that allow a choice among three
sensitivity ranges for readings of 0 - 100 ppm, 0 - 1000 ppm, or 0 - 10,000 ppm on the meter scale.
ADJUSTMENTS
Meter Zero Mechanical Adjustment Screw
-­The meter pointer mechanical adjustment screw is located within the
instrument on the back side of the meter barrel near the - terminal. This
adjustment screw is factory set,
and does not ordinarily require subsequent attention. Any adjustment of this screw should be done with the MODE SELECTOR knob turned to the OFF position.
Page 15
l-6
Meter Zero Coarse Adiustment Screw (Figure l-1)
--
The met er pointer coarse adjustment is accomplished by means of a small
potenti
.ometer adjustment screw located immediately under the ZERO ADJUST
knob at
lower left on the instrument panel. In fresh air, with power on,
0
with the MODE SELECTOR control knob set to PPM X 100, and with the ZERO
ADJUST knob turned to midpoint,
this screw should be slowly turned to
position the pointer to zero on the meter scale.
Meter Zero Fine Adiustment Knob (Figure l-1)
--­In the presence of fresh air, the ZERO ADJUST knob at lower left on the
.
instrument panel may be turned to.adjust the meter pointer to zero -- first
for the PPM X 100 range, then the PPM X 10 range, and lastly for the PPM X 1 range --
just
before
using the instrument to detect combustible gas. If
adjustment of this control does notbringthe pointer
to
zero, the coarse zero-adjustment screw directly below may be turned to provide a greater latitude of fine-adjust control. If the coarse adjustment eventually fails to provide latitude for the fine-adjust control to operate,
in all
probability the replacement of an exhausted detector element is indicated. Gain Potentiometer Adiustment Screws (Figure l-1)
Three gain potentiometer adjustment screws accessible with the instrument
cover removed, are located
on
the
circuit
board
at
lower right within the
instrument case.
With sample calibration gas of known concentration applied
to the air sample intake of the instrument, these screws are turned to
adjust amplifier gain for the correct amount of pointer deflection across
the meter scale.
(See OPERATION Section.) Separate potentiometer screws,
marked X10, X100, and Xl,
are provided for each of the three scales
available by means.of the MODE SELECTOR knob.
Alarm Level Potentiometer Adjustment Screw (Figure 1-l)
The alarm level adjustment screw may be turned to set the audible alarm to respond at any desired degree of pointer deflection across the meter dial. The standard factory setting,
which causes the alarm to respond at a meter
reading of 50,
may be reset to correspond to the ppm Threshold Limit Value,
or to any arbitrary figure.
Record Level Potentiometer Adiustment Screw (Figure l-1) The level of signal outputs to an accessory recorder may be set by turning
the adjustment screw marked
"RECORDER" on the potentiometer assembly within
the instrument case at bottom right.
Page 16
Alarm
Trigger
Point
.
-Adjuster (x2) (w/ 1
(NOTE: Standard
model fflustra.ted.
\.
Adjuster locations
typical
for intrinsically safe
\/f Fine
Zero
Adjust
(R26\
/ i’
model also.
)
,;
Voltage
Adjuster (R20)
Offset Adjuster
(RP3)
Figure l-l.
TLV Circuitry Adjustment Potentiometer Locations.
Page 17
Page 18
2-l
PRE-OPERATION
SECTION 2
CALIBRATIONS AND ADJUSTMENTS
Prepare the TLV Sniffer Combustible Gas Detector for operation in accordance
L
with the following steps:
BATTERY TEST
*
Test battery as follows: Turn MODE SELECTOR knob from OFF position to BATT TEST ,position, Meter pointer should come to rest in BATTERY GOOD range of meter scale.
(Both a meter reading below BATTERY GOOD range and an audible
signal warn of batteries too weak to sustain normal operation.) SETTING METER,POINTER TO ZERO
--
Set pointer to meter zero as follows:
1.
2.
3.
4.
5.
6.
i
7.
t
Attach air sampling probe connector to instrument intake on left side of
case by pulling back spring collar of connector , pressing connector
over
intake, and releasing spring collar,
Place TLV Sniffer inpositionin whichmeter indications willberead
(usually in meter-up position).
NOTE:
Heat distribution from active and reference filaments of the
detector sensor changes from vertical to horizontal position.
The
resulting change in electrical balance between elements causes a shift
in pointer zero from one position to the other.
Set MODE SELECTOR switch to PPM X 100 and operate instrument for 10
minutes to allow circuits to stabilize.
In fresh air, set ZERO ADJUST knob
at
midpoint (five full
turns
from
either extreme position). If fresh air is not available, use Bacharach
Kit #51-7199 to apply known pure air to the Sniffer intake (instructions
in kit).
Turn coarse adjustment screw,
located under
ZERO
ADJUST
knob,
to move
meter pointer to zero on the meter scale.
Turn MODE SELECTOR to PPM X 10 position and turn ZERO ADJUST knob to set
pointer to zero.
Turn MODE SELECTOR to PPM X 1 position and turn ZERO ADJUST knob to set pointer to zero.
NOTE:
The TLV Sniffer is extremely sensitive in the PPM X 1 range.
~02 from breath too close to the intake,
cigarette smoke, auto fumes,
etc.,
can interfere with accurate setting of the pointer to meter zero,
Page 19
2-2
SETTING METER POINTER
Quantitative Gas Test (Refer to Figures 2-1, 2-2, 2-3 and 2-4)
DEFLECTION (GAIN CALIBRATION)
To ensure proper operation and to check calibration, it is necessary periodically check the instrument against a known, standard blend calibrating gas.
to
a
of
i
The Bacharach Code 51-7199 Gas Calibration Kit and optionally available Code
51-1120 Certified Gas Cylinder containing 500 PPM Hexane-in-air are readily available to meet this requirement.
I
Refer to Figure 2-1, the Calibration Kit consists of the following:
Item
No.
Code No.
Description
Qtre
1 51-1201 2
06-6163
3 03-5393
4
03-4318
5 03-5532
6
03-6351
7 51-1127 Optional items may be
8
51-1120
9.
Not
51-7131
Shown
Flowmeter Mtg. Bracket Flowmeter Quick Connect Ftg. Regulator Tee 3/16" (plastic) Rubber Tubing Barbed Hose Ftg. (connected to
Regulator Assembly)
ordered separately:
Cylinder Hexane-Air mixture
(certified 500 PPM)
Zero Calibration Gas (Compressed Air)
1. 1 1 1 1 5 ft.
1 1 1
Refer to Figure 2-2 and connect the gas transfer assembly as shown, making
:
certain all connections are air-tight.
Use the retaining clips (2 each) to mount Flowmeter (06-6163) to its Mounting Bracket (51-1201). Make certain to connect rubber tubing at the base inlet connection on the Flowmeter, then to the barbed fitting on the Regulator and to the Quick Connect fitting previously installed on the TLV Sample-In (inlet fitting).
Turn Regulator Valve (03-4318) fully counterclockwise (closed position) before attempting' to screw regulator into calibration gas tank.
NOTE :
DO NOT OPEN REGULATOR VALVE AT THIS TIME.
Page 20
-
.
Figure Z-l. Bacharach Code 51-7199 Gas Calibration Kit
Figure 2-2. Calibration Gas Transfer Assembly
Properly Hooked Up.
2-3
Page 21
2-4
This test is to be performed in a clean,
fresh air (combustible-free)
environment.
If this is not possible, refer to Figure 2-2 and substitute
Code 51-7131 Zero Calibration Gas for the Code 51-1120 Cylinder of Hexane-
Air mixture. Disregard Step 1) and connect the gas transfer assembly at the TLV Sample-In
(inlet> fitting before performing Steps 4) and 5).
Open the Regulator Valve (clockwise) and adjust for flowmeter indication of
(1) cfh to ensure adequate pump flow.
Remove Code 51-7131 Zero Calibration Gas and
substitute
the Code 51-1120
Cylinder of Hexane-Air mixture before proceeding with Step 6). To calibrate the instrument in fresh air (combustible-free) environment,
proceed as follows:
1) Remove case cover for access to internal adjustments and temporarily break gas transfer assembly connection at the TLV Sample-In (inlet)
fitting.
2)
Turn FINE ZERO ADJUST (pot) full clockwise and then five turns
counterclockwise to mid-range. Then turn COARSE ADJUST (pot) full
clockwise and then ten turns counterclockwise to mid-range.
3)
Turn MODE SELECTOR TO BATT. TEST position. The meter pointer must
indicate within BATTERY GOOD range, if not recharge.
Refer to Figure 2-4 locating TP-3 and connect a Voltmeter between TP-3 (+I and ground (-1, check for 6VDC.
If not, refer to Figure 2-3
locating R-20 and adjust for 6VDC 2 .Ol VDC.
4)
After allowing for five-minute warm-up, turn MODE SELECTOR switch to
PPMXlOO position and adjust R-13 (see Figure 2-3) for meter pointer indication of scale zero.
5)
Turn MODE SELECTOR switch to PPMXlO position and adjust COARSE ADJUST
for meter pointer indication of scale zero. Readjust per Steps 4 and 5 until meter pointer indicates a relatively constant scale zero when MODE SELECTOR is switched between PPMXlO and PPMXlOO range.
6)
Turn MODE SELECTOR switch to PPMXlO position.
Reconnect Gas Transfer
Assembly to TLV Sample-In (Inlet> fitting.
Open Regulator Valve
(clockwise) and adjust for Flowmeter indications of (1) cfh to ensure adequate pump flow. Allow one minute for meter pointer to achieve maximum indication (refer to Figure 2-41, adjust R-3 the X10 Span
Adjuster until meter pointer indicates mid-scale (SO> or 500 ppm.
Remove gas,
close Regulator Valve (fully CCW) and allow about two
minutes for meter pointer to return to zero.
Page 22
2-5
- R-13
. R-20
l
Figure
2-3.
Locating and adjusting R-13 on PPMXlOO position
7)
8)
9)
10)
for meter pointer indication of scale zero.
Turn MODE SELECT switch to PPMXlO position. Then turn the FINE ZERO ADJUST until
meter
pointer indicates
full
scale1000 ppm. Turn MODE SELECT switch to PPMXlOO position and adjust R4 the Xl00 Span Adjuster until meter pointer indicates (10) or 1000 ppm. Turn FINE ZERO ADJUST until meter pointer indicates scale zero.
Turn MODE SELECT switch to PPMXlO position, then turn FINE ZERO ADJUST
until meter pointer indicates 10 on the scale or 100 ppm.
Turn MODE SELECT switch to PPMXl position and adjust the Xl Span Adjuster until meter pointer indicates 100 (full scale) or 100 ppm.
Turn FINE ZERO ADJUST
until
meter pointer indicates scale zero. The TLV is now calibrated and ready for use on the low range O-100 ppm as a gas leak detector.
,
Page 23
2-6
FINE ZERO
ADJUST
R2Xl
R3XlO
- R4XlOO
- TP-3
COARSE
/
ADJUST 1
Figure
2-4.
Locating and Adjusting R-3 for certified Gas Indication of 500 ppm on the meter.
RESETTING ALARM RESPONSE
If factory set alarm response at midpoint of the meter scale is not
suitable, reset alarm response level as follows:
1.
Turn meter zero coarse adjustment screw (located under ZERO ADJUST
control knob at lower left on instrument panel) to set meter pointer to
desired alarm point on meter scale.
2. Turn ALARM potentiometer adjustment screw until audible alarm sounds.
3. Turn meter zero coarse adjustment screw to return pointer to zero on meter scale.
Page 24
2-7
SETTING RECORDING LEVEL If recorder (range: O-100 mv; impedance:
18,000 ohms or greater) is to be
used,
attach accessory recorder jack to RECORDER plug in right side of
instrument case and set recording level as follows:
1.
Set MODE SELECTOR knob to PPM X 100 or PPM X 10 as desired and apply combustible gas to instrument intake.
2.
Turn RECORDER potentiometer adjustment screw until accessory recorder
response corresponds with meter readings as desired.
Page 25
Page 26
3-1
SECTION 2
0
OPERATION
MONITORING TOXICITY
Monitor combustible gas and vapors to determine concentrations with respect
b
to Threshold Limit Values as follows: Direct Readings and Alarm
--
1.
2.
3.
4.
*
5.
6.
7.
Turn MODE SELECTOR control knob to BATT TEST position and read condition of battery on meter dial. Install new or recharged batteries if necessary.
Turn MODE SELECTOR control to desired operating range, selected in accordance with the Threshold Limit Value for the toxic gas to be monitored (PPM X 1 for TLV from ,O to 100 ppm; PPM X 10 for TLV from 0 to 1000 ppm; PPM X 100 for TLV from 0 to 10,000 ppm>.
Allow ten-minute warmup period with instrument in same position as it is
to be used in service (meter facing up or meter facing to the side). In fresh air before entering monitoring area, turn ZERO ADJUST
knob until meter pointer resets on zero. For monitoring in noisy areas,
insert jack of accessory earphone
on right side of instrument case.
in plug
Enter monitoring area and read ppm gas concentrations on meter. Audible warning sounds if gas concentration causes readings at mid-point of
scale or above,
or if toxic Threshold Limit Value has been exceeded,
provided the alarm has been set for this response.
control
For readings .above 10,000 ppm: Replace probe assembly 0023-7243 with dilution probe 0023-7355 and slide dilution probe O-ring to expose dilutionholes of probe (extends range 10X to read
up
to
100,000 ppm).
See Figure 3-l.
Add in-line filter and trap assembly if sampling in
dust- or moisture-laden areas.
Converting Hexane-calibrated Meter PDF Readings to ppm Readings for Other
--
Gases Hexane gas is commonly used for factory calibration and subsequent in-
service recalibration8 of the TLV Sniffer.
To determine ppm concentrations
of gases other than hexane with instruments calibrated for hexane, multiply
the ppm meter reading by the factor for the gas detected as listed in the table below.
(Note meter range setting in making ppm readings and
calculations.)
Page 27
To install in-line filter: Pull back spring-loaded retainer ring and press fitting over connector on
side of sniffer
TLV SNIFFER 23-7350 or 23-7356
To install probe
II
and hose assembly
spring-loaded r
and press fitting
over filter
connector.
Use in dust-or moisture-laden
atmospheres .
Movable O-ring shown in air-dilution
position.
Slide O-ring to cover dilution holes
to obtain undiluted samples.
DILUTION PROBE AND HOSE ASSEMBLY
23-7355
Page 28
3-3
Table 3-l.
Multiplying Factors for Converting ppm Meter Readings of Hexane-
calibrated Instruments to ppm Concentrations of Other Gases.
0
(Approximations)
.
.
Gas Detected
Factor
Gas Detected
Factor
r
Acetone Acetylene Acrylonitrile
.
Benzene
1,3-Butadiene Butane
Butyl Acetate
Carbon Disulfide Carbon Monoxide Cyclohexane Ethane Ethanol Ethyl Acetate Ethyl Ether Ethylene
Ethylene Oxide Heptane Hexane Hydrogen
1.50
1.78
1.54
1.02
1.52
1.04
2.08
5.92
5.11
1.02
1.36
1.90
2.22
1.30
1.38
2.05
1.05
1.00
1.45
Hydrogen Sulfide
18.60
Isopropanol
1.59
M.E.K.
1.60
Methane
1.58
Methanol
3.71
Methyl Acrylate
3.37
Methyl Chloride
4.02
Methyl Chloroform
4.44
Pentane
1,04
Perchlorethylene
13.66
Propane
1.14
Propylene
1.30
Styrene
2.25
Tetrahydrofuran
1.41
Toluene
1.03
Trichloroethylene
6.40
Vinyl Acetate
2.00
Vinyl Chloride
2.24
o-Xylene
1.64
Convertinqppm Readings to Percent Level of Lower Exnlosive Limit (XL.E.L.1
-
---
To determine gas concentration levels in terms of percent of lower explosive
limit (% L.E.L.) from direct ppm readings for hexane or from calculated ppm
concentration levels for other gases:
1.
Read ppm on TLV Sniffer indicating meter.
2.
On O-to-lO,OOO-"PPM CONCENTRATION IN SAMPLE" horizontal scale at bottom of % L.E.L. Conversion Chart, Figure 3-2,
locate position left-to-right
representing ppm reading.
3.
On slanted chart line representing kind of gas detected, find the'point
in vertical alignment over ppm reading point on horizontal scale.
4.
On vertical scale at left labeled % L.E.L. EQUIVALENT," read the percent-of-lower-limit equivalent found in horizontal alignment with the
.
point located on the slanted line.
Page 29
3-4
90-
70-
t 2
60
s
// //
i
so-
2
ETHYLENE
s
0
ETHANOL VINYL CHLORIDE
METHANE
PPM CONCENTRATION IN SAMPLE
Figure 3-2.
Conversion Curves Showing Relationship of PPM
Concentrations
of Various Gases to Percent L.E.L.Approximate
Equivalents.
LOCATING GAS LEAX SOURCES
--
To utilize the TLV Sniffer in searching for gas leaks hoses,
containers, etc.:
1.
Set MODE SELECTOR control knob to PPM X 1 position.
in tanks, pipes,
2.
Search
for exact location of leak with probe.
Meter reading will
increase as leak is approached and decrease as probe moves away from leak.
Page 30
4-1
SECTION 4
MAINTENANCE AND REPAIRS
.
SYMPTOM POSSIBLE CAUSE
REMEDY
SYMPTOMS AND REMEDIES Gradual exhaustion of the detector element,
eventual wearing of air-sample pump parts,batteryrecharging orreplacement,etc., require an amount of maintenance and repair work over an extended period of time.
The following
table describes symptoms, possible causes,
and remedial actions relating to
conditions requiring maintenance and repairs.
TABLE 4-1
TROUBLESHOOTING PROCEDURES
Meter pointer drifts Traces of combustible
Adjust pointer to zero
from zero.
gas in air sample
in fresh air.
Use Bach-
thought to be pure.
arach Kit 51-7199 if fresh air is otherwise unavailable.
Change of position of
Keep instrument
instrument from upright
consistently in either
to horizontal or vice
vertical or horizontal
versa,
thus redistribu-
position while in use.
ting heat between active and reference elements of the detector.
Insufficient warmup
period.
Allow lo-minute warmup
period before zeroing meter pointer.
Erratic monolithic
Consult factory or
chip.
replace circuit board.
Erratic sensing element.
Replace detector.
Page 31
4-2
TABLE 4-1 (continued)
SYMPTOM POSSIBLE CAUSE
REMEDY
False sounding of Change of position of audible gas alarm.
instrument distributes
more heat to detector
active element.
(Meter reads in
alarm zone)
Intrusion of cigarette smoke, auto fumes, or other common combustible
gases or vapors.
Keep instrument consist-
ently in one position
while in use-
Wait for intruding
combustibles to pass;
move away from source; remove source.
Incorrect setting of
alarm level.
Reset alarm level
consistent with appro­priate Threshold
Limit Value.
Faulty monolithic chip.
Consult factory or re-
place printed circuit board.
Faulty detector or
Replace detector; make
1
detector lead
lead connections clean
connections. and tight.
Continuous tone
Run-down battery.
audible signal; meter
Recharge or replace
batteries.
pointer to left of BATTERY GOOD zone
of meter with control set to BATT TEST.
Negative (downscale)
Meter pointer set to
Set pointer to zero
drift of meter
zero on impure air
pointer (audible alarm
with pure air sample.
sounds at minus 10%)
sample, then responding
to purer air.
Faulty detector.
Replace detector.
Faulty monolithic chip.
Consult factory or replace circuit board.
Page 32
4-3
TABLE 4-l (continued)
0
SYMPTOM
POSSIBLE CAUSE
REMEDY
No audible alarm in
Operation of 45-second Wait approximately 45
presence of strong
delay circuit after seconds for delay period
. .
',
concentration of
turning MODE SELECTOR to expire.
combustible gas.
control to operate instrument.
-L
Improper setting of
Reset alarm to desired
alarm response level.
level.
Faulty monolithic
chip.
Consult factory or replace circuit board.
Faulty alarm circuit.
Replace printed circuit board.
Loose lead to speaker;
Inspect for loose leads.
damaged speaker.
Using 0-15V range voltmeter, check for
excitation voltage across speaker terminals while meter
indicates in alarm zone.
Dead batteries. Confirm battery
exhaustion at meter
with MODE SELECTOR
switch in BATT TEST position. Recharge or replace batteries
if necessary.
Page 33
4-4
TABLE 4-l (continued)
SYMPTOM POSSIBLE CAUSE
REMEDY
0
Inability to set
calibration for proper reading on known gas sample.
Error in sample;
Apply
another
known'
sample cylinder empty. gas sample; compare
sample response with
that of a known good
instrument.
Inadequate sample
flow.
Check for air leak
in sample hose, probe,
or external connections.
(Plugging probe inlet
will stall pump if no
leak exists.) Examine
pump piston and diaphragm. Repair or replace faulty components and parts.
If in only one range:
Check potentiometer
faulty calibration. resistance with
ohmmeter.
Replace
if necessary.
Faulty sensing
element.
Replace detector.
Page 34
4-5
Replacing Batteries Replace batteries, if 'necessary, as follows:
1,
Remove ten screws used to hold cover to case and remove cover.
2. Standard model:
Remove two battery assemblies (three batteries in each
of two cardboard tubes) from spring clips in back of case.
*
Intrinsically safe model:
Replace entire sealed battery pack. The
following steps 3 and 4 do not apply.
1
3. Standard model:
Remove old batteries from tubes and replace with new batteries. Orient plus and minus battery terminals as indicated by arrows with polarity signs on battery mounting panel.
4.
Standard model:
Snap replacement batteries with tubes into battery
holding clips.
Spring terminals at one end may be held back with a
small
screwdriver
or similar
tool
to
facilitate insertion of battery
assemblies in clips.
5. Replace cover and attach with ten screws. Replacing Detector Replace detector when necessary as follows:
a
2.
3.
4.
5.
6.
Remove instrument case cover to gain access to detector housing. With instrument in speaker-down position, remove four screws used to
attach detector socket assembly to cubical aluminum air sample chamber.
Lift detector socket assembly from air chamber block.
Unplug old detector from socket assembly.
Insert new detector in socket. New detector part number should correspond to that listed on label affixed to cover of instrument (#800-
080.90).
Reinstall socket assembly with new detector in air chamber and replace
instrument cover.
Tighten screws uniformly and snugly.
E
Replacing Printed Circuit Board
Replace printed circuit board, when necessary, as follows:
1. Remove instrument case cover to gain access to printed circuit board.
2. Using Allen wrench,
remove two small setscrews used to hold MODE'
SELECTOR knob to shaft and remove knob from shaft.
3.
Remove hex head nut used to hold MODE SELECTOR switch shaft to
instrument panel.
Page 35
4-6
4.
Using hex head nut driver,
remove plugs and insulating
recorder and earphone jacks on outside of instrument case.
washers for
5.
Remove two screws used to attach battery charger plug to case.
6. Disconnect terminal clips and soldered connections of all wires leading from printed circuit board to battery terminal block, speaker, meter, coarse zero adjustment potentiometer, and fine zero adjust
potentiometer.
4
7,
Remove four screws used to hold speaker mounting panel
to
instrument
s
case.
8. Slide speaker mounting panel as necessary to free recorder and earphone plugs and MODE SELECTOR shaft from case. Remove circuit board from
case.
9.
Install new circuit board, part #23-4441.
Replacing Sample
Pump
Replace air sample pump, when necessary, as follows:
1. Remove cover of instrument case to gain access to air sample pump.
2. Turn fastener stud #0002-8059 l/4 turn to release miniature pump
assembly from chassis panel.
3. Disconnect pump motor leadwire terminals from battery terminal strip.
4. Disconnect flexible plastic sample tubes from intake and exhaust ports of pump air chamber.
5. Remove pump assembly from instrument.
6. Install replacement pump and connect motor leadwires and flexible plastic
tubes as shown in Figure 4-1.
7.
Operate instrument to test new pump.
Replace cover.
Page 36
I. S. (intrhsically safe) Battery Pa&
MINIATURE PUMP ASSEMBLY
0023-4491
\
,TUBING, l/4
\
Standard Model: 6
Ni-Cad
3 14-070-00
SPEAKER Cells in Tubes
0023-4150\
\
04-0696
0002-3885
CHASSIS
SUBASSEMBLY
0023-4633
Page 37
5-l
REPLACEMENT PARTS LIST
--
z SNIFFER (Standard Model 0023-7350: Intiinsicallv
Safe Model 0023-7356)
b
c
*
*
.
*
*
**
*
*
ITEM
Calibration Kit Connector, Tubing Cover, Instrument
Dry Cell (alternate
to
Ni-Cad Cell; 6 required) Earphone Element, Detector Filter, In-line Dust and Moisture Trap Filter, Probe (box of 24 refills) Fuse (for Intrinsically Safe battery pack)
Handle Hexane Tank (500 ppm)
I.S. (Intrinsically Safe) Battery Pack
Knob, Control
Knob, Handle
Ni-Cad Cell (preferred for standard model; 6 required) Printed Circuit Board Assembly-Std. Model Printed Circuit Board Assembly-Intrinsically Safe Probe and Hose Assembly Probe and Hose Assembly, Dilution PumpAssembly Pump Bellows (replacement) Pump Check Valve (replacement; 2 required) Recorder Plug,
Switchcraft #755
Screw, Cover Mounting (10 required)
Finishing Washer (10 required)
Spacer, Handle Speaker
Tube, Battery (standard model; 2 required)
Tubing, l/4 O.D. x l/16 wall
*
Standard model only
**
Intrinsically safe model only
STOCK'NO.
0051-7199
0003-6152
0023-4454
3600-3100
0023-0605
8000-8090 0023-7341 5500-7000 0204-2703 0023-4445 0051-1120 0023-7354 0023-4443 0023-4444
0004-0030 0023-4441
0023-4778 0023-7243 0023-7355 0023-4773
0023-0018 0023-4218
---
01-3461
102-4062 0023-4647
0023-4150
23-2313
03-6105
Page 38
Loading...