Bacharach DIOXOR II, MONOXOR II, 0019-7038, NONOXOR II, MONOXOR II H Combined Service Manual

...
INSTRUCTION 0019-9166
COMBINED SERVICE MANUAL for:
DIOXOR
®
II
MONOXOR® II
MONOXOR® II H (High Range)
NONOXOR® II
621 Hunt Valley Circle, New Kensington, PA 15068
Phone: 1-800-736-4666 • Fax: 724-334-5001 • Web: www.mybacharach.com
Printed in the U,S.A.
Bacharach, Inc.
® Bacharach, Dioxor, Monoxor, and Nonoxor are registered trademarks of Bacharach, Inc. Copyright © Bacharach 1997-2010, all rights reserved.
A
TABLE OF CONTENTS
1.0 SERVICE MANUAL CONTENTS.................................................................4
2.0 MAINTENANCE ............................................................................................5
2.1 Opening the Case..................................................................................5
2.2 Closing the Case....................................................................................6
2.3 Cleaning........................................................................................7
2.4 Sensor Replacement..............................................................................9
2.5 Printed Circuit Board Replacement.....................................................10
2.6 Chamber Assembly Servicing..............................................................11
3.0 SENSOR CALIBRATION............................................................................12
3.1 Monoxer II CO Calibration.................................................................12
3.2 Dioxor II SO2, Monoxor II H CO, and Nonoxor II NOx Calibration...13
4.0 TROUBLESHOOTING CHART ................................................................14
5.0 REPLACEMENT PARTS.............................................................................17
5.1 Probe Assembly .................................................................................17
5.2 Instrument Assembly ..........................................................................18
5.3 Miscellaneous Parts and Accessories.................................................20
5.4 BACHARACH Sales/Service Centers...............................................20
Single Gas Instrument Service Manual
1.0 SERVICE MANUAL CONTENTS
This manual contains the following service information for the . . .
DIOXOR II (Part No. 0019-7038 & 0019-7043)
MONOXOR II (Part No. 0019-7034 & 0019-7039)
MONOXOR II H (Part No. 0019-8015)
NONOXOR II (Part No. 0019-7036 & 0019-7042).
NOTE: The 0019-7039, 0019-7042, and 0019-7043 versions of these instruments do not contain a probe, all non-probe related functions remain the same in this instruction.
Instrument maintenance, which includes information on how to:
• Open and close the case
• Clean the Fan Assembly and chamber assembly area
• Replace the sensor
• Replace the Printed Circuit Board (P.C.B.)
• Service the gas-sensing chamber assembly (chamber assembly)
Sensor calibration
Troubleshooting chart
Replacement parts list
GENERAL NOTE
If excessive torque is applied to the instrument's fasteners, poor performance or damage to components may occur.
The torque specifi cations are as follows:
Screws Torque
Sensor Mounting 4 - 8 oz. in. Motor Mounting 6-10 oz. in. Chamber Sealing 16-20 oz. in. Rear Cover/Case 18-24 oz. in. Fan Assembly 32-40 oz. in.
Page 4 Bacharach, Inc. Instruction 0019-9166
2.0 MAINTENANCE
2.1 Opening the Case
Equipment Needed:
Phillips-head screwdriver, No. 1
Procedure:
1. Lay the instrument face down on a workbench; then detach the metal clip from the bottom of case and fold the elastic strap over top of the instrument (Fig 2-1).
2. Slide off battery cover.
3. Remove batteries.
4. Using a No. 1 Phillips-head screwdriver, remove the two screws from the rear of the case.
Single Gas Instrument Service Manual
Figure 2-1.
Opening and Closing the Case
5. Lift up the bottom portion of the rear case and slide it forward until the case clears the inlet port.
6. To gain better access to the components, remove the gas-sensing chamber as­sembly (Chamber Assembly) and the Printed Circuit Board (P.C.B.) from the front case by pulling them straight up from the front case (Fig 2 -2); then lay these items on your workbench.
Instruction 0019-9166 Bacharach, Inc. Page 5
Single Gas Instrument Service Manual
2.2 Closing the Case
Equipment Needed:
• Phillips-Head Screwdriver, No. 1
Procedure:
When replacing the P.C.B. and chamber assembly, the P.C.B. must be installed FIRST. The P.C.B. switches could be damaged if the P.C.B. is installed after the chamber assembly is in place.
1. Ensure that the P.C.B. is properly positioned face down over its locating-tab on the front case (Fig 2-2).
2. Properly position the chamber assembly between the ribs of the front case.
CAUTION
Figure 2-2.
P.C.B. and Chamber Assem­bly Removal and Installation
3. Slide the rear case over the inlet port and properly join the two case halves together.
4. Secure the case halves to each other using the two screws removed in Sec­tion 2.1, Step 4 (Tighten to 18–24 oz. in.).
5. Reinstall the batteries (observe proper battery polarity); replace the battery cover; and attach the elastic strap’s metal clip to the bottom of the instru­ment.
Page 6 Bacharach, Inc. Instruction 0019-9166
2.3 Cleaning
The inside of the chamber assembly is the only part of the instrument that needs to be cleaned. To clean the chamber assembly, proceed as follows:
Equipment Needed:
Screwdriver, No. 1 Phillips head Clean rags
Screwdriver, 1/8" at-blade Cotton Swabs
Warm soapy water Pipe cleaners
Allen Wrench, .05 in.
Procedure:
1. Open the case as described in Section “2.1 Opening the Case.”
2. Using a fl at-blade screwdriver, remove the three screws and nuts that secure the Sensor to the chamber assembly; DO NOT remove the sensor wires. Then remove the Sensor along with its O-ring and Membrane from the chamber assembly.
Single Gas Instrument Service Manual
3. Using a No. 1 Phillips-head screwdriver, disassemble the chamber assembly by removing the two screws and nuts that secure the Chamber to the Motor Assembly
(Fig 2-3).
NOTE: Do not remove the Fan Assembly unless it's absolutely neces­sary for cleaning or repair purposes.
4. If it is necessary to remove the Fan Assembly, fi rst loosen the set screw on the Fan Assembly with the .05 in. Allen wrench, and then pull off the assem­bly.
NOTE: It is not necessary to disassemble the Motor Assembly. If motor replacement is required, refer to Section 2.6, Step 5.
5. Clean the Fan Assembly, Chamber & Motor Assembly (see Fig 2-3) using cotton swabs and clean rags dampened with warm, soapy water. After clean­ing, dry all parts with clean dry rags and cotton swabs.
NOTE: Replace, DO NOT CLEAN, the Membrane (Fig 2-2) if dirty.
6. Clean the Chamber’s inlet and outlet ports using pipe cleaners dipped in warm, soapy water. After cleaning, dry them with clean dry pipe cleaners.
7. Reassemble the chamber assembly as follows:
NOTE: DO NOT overtighten any screws; damage to plastic parts will result.
Instruction 0019-9166 Bacharach, Inc. Page 7
Single Gas Instrument Service Manual
a. If disassembled, push the Fan Assembly onto the Motor shaft until it seats
(Fig 2-3). Be sure the set screw (Fig 2-3A) is not screwed in so far that it prevents the adapter from seating fully. Tighten the set screw to approxi­mately 32–40 ounce-inches using the .05 in. Allen wrench. Spin the Fan Assembly to be sure it will run in the chamber.
b. Mate and secure the Chamber and Motor Assembly together with the
two screws and nuts removed in Step 3 (Tighten to 16–20 oz. in.).
8. Reinstall the Membrane (mesh side toward the chamber assembly), O-ring, and Sensor on the chamber assembly (Fig 2-2); then secure the Sensor using the three screws and nuts removed in Step 2 (Tighten to 4–8 oz. in.).
9. Close the case as per Section “2.2 Closing the Case.”
Figure 2-3.
Chamber As­sembly, Exploded View
Figure 2-3A.
Fan-blade assem­bly set screw
Page 8 Bacharach, Inc. Instruction 0019-9166
2.4 Sensor Replacement
When the instrument can no longer be spanned, as described in Section “3.0 Sen­sor Calibration,” the Sensor has reached the end of its useful life, and should be replaced. The life of a Sensor is typically two years. The Sensor may also need to be replaced if the instrument will not zero properly, or if erratic gas readings occur. Refer to Section “4.0 Troubleshooting Chart.”
Equipment Needed (Refer to Section “5.0 Replacement Parts” for Part Nos.):
Phillips-head screwdriver, No. 1 Replacement Sensor:
Nutdriver, 4mm (5/32") Dioxor II - SO
Screwdriver, 1/8" at blade Monoxor II - CO
O-Ring and Membrane, if required Nonoxor II - NO
Procedure:
Burn hazard! The Sensors contain corrosive chemicals, DO NOT puncture or take the sensors apart.
Single Gas Instrument Service Manual
Monoxor II H - CO
WARNING!
2
X
1. Open the case as described in Section “2.1 Opening the Case.”
2. Using a nutdriver, loosen the nuts that secure the sensor wires to the Sen­sor terminals R, S and C; then remove the wires. If the Sensor has pin and socket connections, simply pull the sockets from the sensor pins.
3. Using a fl at-blade screwdriver, remove the three screws and nuts that secure the Sensor to the chamber assembly; then remove the Sensor along with its O-ring and Membrane from the chamber assembly.
4. Remove and discard of the old Sensor in accordance with local and federal regulations. DO NOT discard the sensor’s O-ring and Membrane, unless you intend to replace these items.
NOTE: Replace the O-ring if it looks deformed, and replace the Membrane if it is contaminated with oil or dirt. We recommend that both the O-ring and Membrane be replaced at the same time.
5. Remove the new Sensor from its shipping canister. Do not remove the short­ing wire (across the R & S terminals) from the Sensor (SO2 & CO only, the NOX does not have a shorting wire) until you are ready to make the electrical connection.
6. Position the Membrane, O-ring, and new Sensor on the chamber assembly as per Fig 2-2. Ensure that the sensor terminals are positioned as shown. Secure these items to the chamber assembly using the screws and nuts re­moved in Step 3 (Torque to 4–8 oz. in.). DO NOT overtighten the screws!
7. Remove the shorting wire from the R & S terminals (SO2 & CO only). Con­nect the blue (S), red (R), and black (C) wires from the P.C.B. to their as­sociated sensor terminals (See Fig 2-4). If the Sensor uses nuts to secure the
Sensor wires, DO NOT overtighten these connections!
8. Close the case as per Section “2.2 Closing the Case.”
9. Calibrate the new Sensor as per Section “3.0 Sensor Calibration.”
Instruction 0019-9166 Bacharach, Inc. Page 9
Single Gas Instrument Service Manual
2.5 Printed Circuit Board (P.C.B.) Replacement
Equipment Needed (Refer to Section “5.0 Replacement Parts” for Part Nos.):
Phillips-head screwdriver, No. 1
Printed Circuit Board (P.C.B.)
Procedure:
1. Open the case as described in Section “2.1 Opening the Case.”
2. Unplug the motor connector from the P.C.B.
3. Unplug the sensor connector from the P.C.B.
NOTE: (For SO2 & CO Sensors only) If the Sensor is not going to be immediately plugged into the new P.C.B., place a shorting wire between sensor terminals R and S.
4. Remove the Switch Covers from the old P.C.B. and place them on the new
5. Plug the motor connector into the P.C.B.
6. Remove the shorting wire from sensor terminals R & S (if installed); then
7. Close the case as per Section 2.2 Closing the Case.”
8. Calibrate the Sensor with the new P.C.B. as per Section “3.0 Sensor Calibra-
9. Discard the old P.C.B.
Figure 2-4.
Sensor & Motor Wiring Diagram
P.C.B., then replace the old P.C.B. with the new one.
plug the sensor connector into the P.C.B.
tion.”
Page 10 Bacharach, Inc. Instruction 0019-9166
2.6 Chamber Assembly Servicing
To disassemble the chamber assembly, proceed as follows:
Equipment Needed (Refer to Section “5.0 Replacement Parts” for Part Nos.):
Screwdriver, Phillips head, No. 1
Screwdriver, 1/8" at-blade
Allen Wrench, .05 in.
Replacement pump parts, as required
Procedure:
1. Open the case as described in Section “2.1 Opening the Case.”
2. Using a fl at-blade screwdriver, remove the three screws and nuts that secure the Sensor to the chamber assembly; DO NOT remove the sensor wires. Then remove the Sensor along with its O-ring and Membrane from the chamber assembly.
Single Gas Instrument Service Manual
3. Using a No. 1 Phillips-head screwdriver, disassemble the Chamber from the Motor Assembly (Fig 2-3).
4. To remove the Fan Assembly, refer to Section 2.3, Step 4.
5. If Motor replacement is required:
a. Unplug the motor connector from the circuit board and discard the Motor
Assembly.
b. Install a new Motor Assembly and connect the motor connector to the
circuit board.
NOTE: A newly installed Motor must be broken in by running it for at least 30 minutes. A Motor not broken in may cause erratic display readings.
6. Reassemble the chamber assembly with any replacement parts using Section
2.3, Step 7.
7. Reinstall the Membrane, O-ring, and Sensor using Section 2.3, Step 8.
8. Close the case as per Section “2.2 Closing the Case.”
Instruction 0019-9166 Bacharach, Inc. Page 11
Single Gas Instrument Service Manual
3.0 SENSOR CALIBRATION
3.1 Monoxor II CO Calibration
The following calibrates the MONOXOR II to a known concentration of carbon monoxide.
Equipment Needed (Refer to Section “5.0 Replacement Parts” for Part Nos.):
Calibration Kit
Carbon Monoxide Gas Cylinder, 500 ppm
Potentiometer adjustment tool
Procedure:
Perform this procedure in a well ventilated area. CO is toxic!
1. Assemble the calibration equipment by connecting the Carbon Monoxide Gas Cylinder and the components of the Calibration Kit together as shown in Fig 3-1.
WARNING!
Figure 3-1.
Calibration Kit Setup
NOTE: DO NOT attach the calibration setup to the instrument at this
time.
2. Zero the instrument as follows:
a. Ensure instrument will be sampling CO-free (fresh) air in the following
steps.
NOTE: An indication that the instrument was not zeroed in fresh air
is when a large negative CO display appears when the instrument is moved into another area.
b. Set front panel POWER switch to ON.
Page 12 Bacharach, Inc. Instruction 0019-9166
Single Gas Instrument Service Manual
Important! Do not proceed with this calibration procedure if the instrument’s “LO BAT”
indicator is on. We recommend that the instrument be calibrated with a fresh set of batteries.
c. Wait until the instrument’s CO display settles down (approximately 1 minute); then, using
a potentiometer adjustment tool, turn the front panel ADJUST potentiometer (Fig 3-2) until the display shows 0 ±3 ppm. The display shows negative numbers for zeroing purposes.
Figure 3-2.
ADJUST & CO Span Adjustments
3. Connect the CO discharge hose of the calibration equipment to the Monoxor's probe tip; then open the CO cylinder regulator valve and adjust the regulator for a fl owmeter indication of 1 to 2 SCFH.
4. Allow gas to fl ow for 2 minutes.
5. Read the detected CO level on the instrument’s LCD. Then using a potentiometer adjustment tool, adjust the CO SPAN potentiometer until the LCD reading matches the gas cylinder’s concentration value within ±5 ppm.
6. Shut off the regulator and remove the CO discharge hose from the gas inlet port. Then while observing the displayed CO level, allow the instrument to run until the CO reading drops to zero.
NOTE: If after 5 minutes the CO reading stabilizes at some value other than zero, readjust the
ADJUST potentiometer for an LCD indica tion of ±3 ppm.
7. Set front panel POWER switch to OFF.
8. Disassemble the calibration-equipment setup and store the components in the calibration kit’s carrying case.
3.2 Dioxor SO2, Monoxor II H CO, and Nonoxor NOX Calibration
The presence of SO2 and NOX gas in small amounts, and CO gas in larger amounts, can be very toxic. Calibration at a Bacharach Sales/Service Center is recommended (See Section “5.4 Bacharach Sales/ Service Centers” to fi nd the nearest center). If you have a facility that is properly equipped to handle toxic gas (i.e.. fume hoods, self contained breathing apparatus etc.), we would be glad to suggest a calibration method.
Instruction 0019-9166 Bacharach, Inc. Page 13
Single Gas Instrument Service Manual
4.0 TROUBLESHOOTING CHART
Symptom Cause Remedy
No display when unit is switched on
Backlight won't light
Unit doesn't respond to gas (e.g., from calibration gas)
Batteries dead
Batteries installed backwards
P.C.B. defective
Backlight or its switch is defec­tive
Sensor connections loose or broken
Sensor Membrane blocked with water or dirt
Replace batteries with four fresh “C” Alkaline cells.
Insert batteries correctly. Wait at least one minute be­fore switching on unit to allow its PTC resistor to cool.
Replace P.C.B. Refer to Section “2.5 Printed Circuit Board Replacement.”
Replace P.C.B. Refer to Section “2.5 Printed Circuit Board Replacement.”
Check Sensor connections (Fig. 2-2).
Remove Sensor and replace its Membrane. See Fig. 2-2.
LCD segments missing
Sensor depleted or defective
P.C.B. defective
LCD loose or defective
P.C.B. defective
Replace Sensor. Refer to Section “2.4 Sensor Replace­ment.”
Replace P.C.B. Refer to Section “2.5 Printed Circuit Board Replacement.”
Carefully re-seat LCD in its socket. If segments are still missing, replace LCD.
Replace P.C.B. Refer to Section “2.5 Printed Circuit Board Replacement.”
Page 14 Bacharach, Inc. Instruction 0019-9166
Single Gas Instrument Service Manual
Symptom Cause Remedy
Erratic display
Unit won't zero and/or span
Low batteries
Faulty Motor
Sensor connections loose or broken
Sensor depleted or defective
P.C.B. defective
Too much gas present to zero
Replace batteries with four fresh “C” Alkaline cells.
Break in a new Motor for at 30 minutes.
Replace Motor Assembly. Refer to Section “2.6 Chamber Assembly Servicing.”
Check Sensor connections. See Fig 2-2.
Replace Sensor. Refer to Section “2.4 Sensor Replace­ment.”
Replace P.C.B. Refer to Section “2.5 Printed Circuit Board Replacement.”
Move unit into an area known to be gas free.
Probe kinked or blocked pre­venting gas fl ow
Sensor connections loose or broken
Sensor depleted or defective
P.C.B. defective
Unkink or Replace Probe.
Check Sensor connections. See Fig 2-2.
Replace Sensor. Refer to Section “2.4 Sensor Replace­ment.”
Replace P.C.B. Refer to Section “2.5 Printed Circuit Board Replacement.”
Instruction 0019-9166 Bacharach, Inc. Page 15
Single Gas Instrument Service Manual
Symptom Cause Remedy
Bouncing display > ± 3 counts
Very little (<0.25" H2O) or no draft indication on manometer dead headed into instrument
Sensor connections loose
Motor has excessive arcing in its carbon brush/commutator interface causing RFI
Chamber assembly screws loose
Fan-blade Assembly loose
Inlet and/or outlet port(s) blocked
Check Sensor connections. See Fig 2-2.
Run Motor for at least 3 hours to wear away arc spots. If display still bounces, replace Motor Assembly. Refer to Sec­tion “2.6 Chamber Assembly Servicing.”
Tighten screws. Refer to Sec­tion “2.6 Chamber Assembly Servicing.”
Open chamber assembly and ensure that Fan-blade Assem­bly is mounted on Motor shaft. Refer to Section “2.6 Chamber Assembly Servicing.”
Clean ports. Refer to Section “2.3 Cleaning.”
Motor connections loose
Motor defective
P.C.B. defective
Reconnect Motor leads. Refer to Section "2.6 Chamber As­sembly Servicing."
Check operation of Motor. If motor is receiving power but isn’t turning, replace Motor Assembly. Refer to Section “2.6 Chamber Assembly Servicing.”
Replace P.C.B. Refer to Sec­tion “2.5 Printed Circuit Board Replacement.”
Page 16 Bacharach, Inc. Instruction 0019-9166
5.0 REPLACEMENT PARTS
Replacement parts for the instrument are listed below. Please refer to the ap­propriate fi gure(s) for part locations.
5.1 Probe Assembly (Fig 5-1)
Item Description Part No.
1 Probe Assembly, Standard 0019-3076
-- Probe Assembly, Flexible (not shown) 0019-3104 2 Probe Tube 0019-3073 3 Probe Handle 0019-3017 4 Screw, Phillips head, M3 x .5 x 8 mm 0501-3806 5 Tubing, 5/16 OD X 3/16 ID 0003-6109 6 Filter Saturator Bushing 0011-0074 7 Filter Packing 0011-0121 8 Filter Tube 0011-0130 9 Front Ferrule 0003-2024 10 Rear Ferrule 0003-2025 11 Knurled Nut 0003-2023 12 Compression Fitting 0003-2021 13 Barbed Fitting 0003-6164 14 Extended Adapter Block 0019-3072 15 Connector Tube 0011-0070
Single Gas Instrument Service Manual
Figure 5-1. Standard Probe Assembly
Instruction 0019-9166 Bacharach, Inc. Page 17
Single Gas Instrument Service Manual
5.2 Instrument Assembly (Figs 5-2 & 5-3)
Item Description Part No.
1 Adapter, Motor Shaft 0024-0412 2 Battery Clip, push on 0004-3113 3 Battery Cover with Foam Pad 0019-3029 4 Chamber 0019-3090 5 Front Cover w/Display Window (Dioxor II) 0019-3079 Front Cover w/Display Window (Monoxor II) 0019-3055 Front Cover w/Display Window (Nonoxor II) 0019-3078 Front Cover (Monoxor II H) (order with 0019-0437) 0019-3014 Display Window (Monoxor II H) (order with 0019-3014) 0019-0437 6 (Not Used) 7 Fan-Blade Assembly 0019-3013 8 LCD 0004-3700 9 Motor Assembly 0019-3063 10 Nut, Hex M2 x .4 0502-0007 11 O-Ring and Membrane 0024-0520 12 Printed Circuit Board Assy. (Dioxor II) 0019-3022 Printed Circuit Board Assy. (Monoxor II) 0019-3012 Printed Circuit Board Assy. (Nonoxor II) 0019-3020 Printed Circuit Board Assy. (Monoxor II H) 0019-3094 13 PTC Resistor 0204-9427 14 Rear Case w/ Label (Dioxor II) 0019-3082 Rear Case w/ Label (Monoxor II) 0019-3053 Rear Case w/ Label (Nonoxor II) 0019-3081 Rear Case (Monoxor II H) (order with 51-2347) 0019-3008 Label (Monoxor II H) (order with 0019-3008) 0051-2347 15 Retaining Ring 0002-2675 16 Screw, Cheese head, M2 x .4 x 30 mm 0501-3807 17 Screw, Phillips head, M2 x .4 x 6 mm 0501-3803 18 Screw, Phillips head, M3 x .5 x 8 mm 0501-3806 19 Sensor, SO2 (Dioxor II) 0019-8002 Sensor, CO (Monoxor II) 0024-0498 Sensor, ,NOX (Nonoxor II) 0019-8001 Sensor, CO (Monoxor II H) 0019-8016 20 Set Screw, #4-40 x 1/8 LG 0002-0001 21 Strap 0019-3028 22 Strap Clip 0019-3030 23 Strap Retainer 0019-3027 24 Switch Cover 0019-3021
Figure 5-2. Replacement Parts, Chamber Assem­bly Exploded View
Page 18 Bacharach, Inc. Instruction 0019-9166
Figure 5-3.
Replacement Parts, Back and Front Exploded Views
Single Gas Instrument Service Manual
Instruction 0019-9166 Bacharach, Inc. Page 19
5.3 Miscellaneous Parts and Accessories
Description Part No.
Calibration Kit 0024-7059 Carbon Monoxide Gas Cylinder, 500 ppm 0024-0492 Potentiometer adjustment tool 0006-9456 Custom Carrying Case (1 Instrument) 0019-0377 Custom Carrying Case (2 Instrument) 0019-3040 Instruction Manual, (Dioxor II) 0019-9121 Instruction Manual, (Monoxor II) 0019-9113 Instruction Manual, (Nonoxor II) 0019-9120 Instruction Manual, (Monoxor II H) 0019-9180 Calibration Card 0019-9150 Service Manual, (This Manual) 0019-9166 Label, Carrying Case 0019-3077 Probe Stop (for Securing ¼ Probe in a Flue or Stack) 0024-0434 Warranty Card 0006-8825
5.4 BACHARACH Sales/Service Centers
United States
Bacharach, Inc. 621 Hunt Valley Circle New Kensington, PA 15068 Phone: 724-334-5051 Fax: 724-334-5723 E-mail: help@mybacharach.com
Canada
Bacharach of Canada, Inc. 20 Amber St. Unit #7 Markham, Ontario L3R SP4 Canada Phone: 905-470-8985 Fax: 905-470-8963 Email: bachcan@idirect.com
Loading...