Bacharach 0012-0266 User manual

Find Quality Products Online at: sales@GlobalTestSupply.com
PARTS LIST
Part No. Description Qty. per Assembly
0012-7011 0012-7012 0012-7013 0012-7043
0012-0180 Plug and Swivel Assembly ...................1 1 1 1
0012-0176 Thermometer Clip ................................. 2 2 2 2
0012-0007 Wick ....................................................... 1 1 1 1
0002-1740 Screw ..................................................... 2 2 2 2
0002-6858 Roll Pin .................................................. 1 1 1 1
0006-5456 Cap ......................................................... 1 1 1 1
0012-0012 Friction Ring ......................................... 2 2 2 2
0012-0018 Body (°C) ............................................... - - 1 1
0012-0006 Tube (°F) ................................................ 1 1 - -
0012-0019 Tube (°C) ............................................... - - 1 1
www.GlobalTestSupply.com
0012-0267 Thermometer, 30 to 110°F, Mercury ...2 - - ­0012-0266 Thermometer, 30 to 110°F, Red Spirit - 2 - -
0012-0265 Thermometer, –5 to 50°C, Mercury ..... - - 2 -
0012-0327 Thermometer, –5 to 50°C, Red Spirit .. - - - 2
Accessories
0012-0011 Wick Replacement Kit containing ................................................ 4 Wicks
INSTRUCTION 0012-9015
SLING PSYCHROMETER
Part Numbers:
0012-7011 30° to 110°F Mercury Thermometer 0012-7013 –5° to 50°C Mercury Thermometer 0012-7012 30° to 110°F Red Spirit Thermometer 0012-7043 –5° to 50°C Red Spirit Thermometer
Rev. 5 – May 2010
CAUTION: Damage to thermometers or
plastic case will result if instrument is ex­posed to direct sun light or ex ces sive heat for prolonged periods.
WARRANTY
Bacharach, Inc. war rants to Buy er that at the time of deliv­ery this Prod uct will be free from defects in ma te ri al and manufac ture and will con form sub stan tial ly to Bacha rach Inc.’s applica ble spec i fi ca tions. Ba charach’s li a bil i ty and Buyer’s remedy un der this warranty are lim it ed to the repair or replace ment, at Ba charach’s op tion, of this Prod uct or parts thereof returned to Seller at the factory of manu facture and shown to Bacharach Inc.’s reason able sat is fac tion to have been de fec tive; pro vid ed that written notice of the defect shall have been given by Buyer to Bach arach Inc. with in one (1) year after the date of delivery of this Product by Bacharach, Inc. For full details con cern ing this war ran ty, contact Bacharach Inc.
Register Your Warranty On-line
Printed in U.S.A. ®Registered Trademark
4
In struc tion 0012-9015
Find Quality Products Online at: sales@GlobalTestSupply.com
1
Remove cap; im­merse end of body to saturate wick.
2
Fill psy­chrometer reservoir with water. Replace cap.
www.GlobalTestSupply.com
3
Be sure wick covers wet” bulb (the other bulb is dry).
4
Pull body from tube until body hangs free.
5
Use tube as handle; whirl body about 1-½ minutes.
6
Read wet & dry bulb ther­ mom e ters; replace body in
A D G H I
B G H C F E
7
Set thermometer readings on upper two calculator scales.
8
Read % relative humidity as indicated by arrow.
OPERATION
Before Using:
WICK (A) should be thoroughly sat u rat ed with wa ter. Remove END CAP (B) and im­ merse PSY CHROM E TER BODY (C) up to mercury (or red spirit) reservoir on the ther­ mom e ters un til WICK is thor ough ly wet ted. Fill END CAP with water and re place; tight en just enough to pre vent leak age.
To Use:
1. Be sure WICK (A) is wet and covers mer– cury (or red spirit) reservoir on WET BULB THER MOM E TER (D). Be sure res er voir on other THER MOM E TER (F) is dry.
2. Pull TUBE (E) clear of BODY so BODY can swivel.
APPLICATION
Wet-bulb temperatures should be read fi rst and as quickly as possible for high est ac­ cu ra cy. Delay in reading may cause er ror. In addition, the wick must be kept clean, sat u ­rat ed with water, and whirled long enough to stabilize tem per a tures.
The range of the psychrometer is from 10% to 100% R.H., for dry-bulb tem per a tures of 30° to 100°F (–5° to 50°C).
In addition to the above instructions, baro met ric pressure and other factors will in fl u ence exact relative humidity de ter mi n-
MAINTENANCE
WICK (A) should be kept clean; when dirty, cut off WICK below WET BULB THER­ MOM E TER (D) and pull clean sec tion out of END CAP (B) and slide over bulb on WET BULB THER MOM E TER.
A WICK REPLACEMENT KIT may be pur­ chased separately (refer to parts list). One or two extra WICKS may be kept in END CAP and will help retain mois ture longer. Pack WICKS loosely to allow thor ough wa ter
3. Holding TUBE, whirl body two to three revolutions per second (120 to 180 RPM).
4. Continue whirling until temperatures sta bi lize (1-½ minutes is usually am ple).
5. Immediately read WET BULB THER­ MOM E TER (D) and then DRY BULB THER MOM E TER (F). (See ap pli ca tion in struc tions.)
6. Set wet and dry bulb temperatures op­ po site each other on side rule type cal cu ­la tor scales, sliding BODY into TUBE as re quired.
7. Read % R.H. (percent relative hu mid i ty) indicated by arrowhead on lower scale.
a tions to a very minor degree. For precise work, use a psychrometric chart or set of tables such as W.B. No. 235 “Psy chro met ric Tables for Ob tain ing the Vapor Pres sure, Relative Hu mid i ty, and Tem per a ture of Dew Point,” which can be purchased from the Su per in ten dent of Documents, United States Gov ern ment Printing Of fi ce, D.C. The accuracy of the Psy chrom e ter, how ev er, is sat is fac to ry for all ex cept the most ex act ing work.
sat u ra tion and ample wa ter supply to the WET BULB THER MOM E TER.