AZ Instrument NICEFLOW Users Manual

Page 1
Operation Manual
10cm Vane
Wireless Air Flow Meter
with Airflow cone
Model:8903/8904/8906
Model:
89161 89171 89191
Page 2
Thank you for purchasing this wireless
INTRODUCTION
air flow meter. This meter is specially designed for HVAC&R application that combines those required measurements, like Air Velocity, Air volume, Temp., RH, WBT, DP, and CO2 in one handy meter.
Optional airflow cone helps in quick volume measurement, just install the compatible airflow cone, meter will recognize the area for air volume calculations automatically. A hassle free way to take the value of air volume .
Features :
Wireless transfer with free App
available for both iOS & Android. Available for air velocity, air volume, temp., RH, WBT, DP, and CO2.
Big LCD digital display Multi-points and Timed average Metric& Imperial units selectable Optional Round & Square airflow cone Fast response & Low battery indication Blue back light for dark places use Handheld size, easy to carry
Function Table
Air Velocity Air volume
Air temp. Humidity Dew Point Wet Bulb CO2
Accurate reading
89161 8919189171
1
Page 3
MATERIAL SUPPLIED
This package contains:
The meter x 1 Battery x 4 (AAA size ) Operation manual x1
Hard carry case x1
Optional accessory:
Airflow cone set: Round conex1 (dia.210mm) Square conex1 (346x346mm)
CONTROLS AND INDICATORS
LCD DISPLAY
HOLD
inch
cm
Upper display: air velocity, RH, CO2 value.
(Depends on model) duration time in average.
Lower display: air volume, temp., WBT, DP
(Depends on model) value.
HOLD MAXMIN:maximum & minimum reading
AVG: multi-points & timed average
: battery low voltage indicator
ppm
m/s,fpm: air velocity unit %RH : relative humidity unit
OO
OO
: temperature unit
F
C
CMM,CFM: air volume unit WBT, DP : wet bulb, dew point temp.
2
2
cm
inch
: area dimension unit : air flow cone indicator : vain icon in this meter
MIN
2
2
MAX
AVG
m/s fpm
ppm
%RH
OO
C
CMM CFM WBT
DP
: freezes the current reading.
:CO2 unit
2
OO
F
Page 4
KEY PAD
-Power button
-Long press to enter or exit setup mode.
-Freeze current reading on LCD
-Long press to review MIN and MAX
-In average mode, press to record data or start timed recording.
-In setup mode, press to enter data settings.
-Switch for upper display.
-Long press to enter average mode.
-In average mode, press to select option of average mode, press again to return to normal mode.
-In setup mode, press to select category or increase value.
-Switch for lower display.
-In average mode, press to view recorded and average reading of all parameters.
-In setup mode, press to select category or decrease value.
+ -Disable auto power off.
3
Page 5
MEASUREMENT
POWER ON/OFF
Press to turn on the meter. The screen shows full display with blue backlight on and enter normal mode. (Model 89161/89171). For model 89191,it proceeds 30 seconds warm up after turned on the meter, then it enter normal mode. Press again in any mode to turn off the meter.
The meter is preset auto sleep in 20 minutes of inactivity. To disable this function, hold down and for 2 seconds to turn on the meter while it is off. It shows “n” on the LCD (see Fig.A) and goes into normal mode(or 30 sec. warm up for model 89191) and won’t turn off itself until is pressed or out of voltage.
Fig.A
SWITCH DISPLAY
When the meter is turned on, it shows the current readings of air velocity on the upper display and air temp. on the lower(see Fig.B). To switch display of other measurements, press or .
m/s
OO
C
Fig.B
4
Page 6
Short press , the upper display will cycle among air velocity, RH, and CO2 (89191 only). And short press , the lower display cycles among air temp. air volume WBT(wet bulb temp) DP(dew point temp)(Fig.C).
%RH
CMM
Fig.C
AIR VOLUME
To measure air volume, it is necessary to get the outlet dimensions. Go to Setup mode and input the AREA value first before taking measurement. (See page 9 , P2.0)
After AREA setting is completed, press to switch the display to air volume and the measured air volume will be displayed on the lower LCD.
Measure with airflow cones
Two optional airflow cones are available for using with this anemometer and are sold as a set.
Square 346x346mm
Round dia. 210mm
5
Page 7
To measure air volume with use of an airflow cone. Just plug on the cone and the meter will automatically recognize it and shows a on the display(Fig.C). Then the meter will calculate the air volume by the default dimensions despite of any AREA value is input.
Note:
Be sure the airflow cone is well mounted and locked tight. (See pic.1)
pic.1
Lock
RH(89171/89191 only)
The capacitive humidity sensor is built in the center of the vane to measure air flow relative humidity and is used for calculating dew point and wet bulb temperatures.
Press to switch to RH display. It shows measured RH readings on the upper display. And it is also possible to display with RH offset. Go to Setup mode (see page 9, P3.0) and input offset value, then the displayed readings will be the measured data with offset.
6
Page 8
CO2 (89191 only)
Model 89191 features additional CO2 measurement for HVAC application. The NDIR CO2 sensor is built in the rear side of the meter. Press to switch to CO2 display and the current reading will show on the upper LCD.
For places in high altitude, the pressure compensation needs to be considered in order to get accurate CO2 readings. Go to Setup mode P4.0 and input the pressure value according to your location before taking measurements.
DATA HOLD
In normal mode, press and the current readings are fixed on the LCD with HOLD icon on the top and won’t change until a second press on .
MIN/MAX
In normal mode, long press for 3 seconds and it displays the minimum readings of all parameters since power on with MIN icon on the top. Long press this button again, it shows the maximum readings with MAX on the top. (Fig.D) Long press again and it returns to the normal mode.
While reviewing the minimum and maximum data, press and and it switches to display MIN and MAX readings of other parameters.
MAX
MAX
m/s
m/s
OO
OO
C
C
Fig.D
7
Page 9
AVERAGE
This meter calculates multi-points average and timed average of all parameters.
Multi-points average
In normal mode, press for 2 seconds to enter multi-point average mode.(See
Fig.E, the upper display shows the record number and the lower is the measured reading)
- Press to record readings.
- Press to see the multi-point average once all data is recorded. It shows the average readings with flashing” “.
AVGAVG
- Press to view the average reading of each parameter.
(See Fig.F, the upper display means total 20 recorded data; and the lower display is the average reading of RH.)
- Press to return to normal mode.
m/sAVG m/sAVG m/sAVG m/sAVG
AVG
AVG
%RH
%RH
Fig.E
Fig.F
Timed average
In normal mode, press for 2 seconds and it goes into multi-point average. Press it again to enter timed average mode. (See Fig.G, the upper display shows
the duration time in second, the maximum time is 19999 seconds, and the lower is the measured reading)
- Press to start recording, the upper display starts the counting clock.
- Press to stop and calculate the timed average. It shows the duration seconds on the upper display and the average readings at the lower display with flashing” “. (See Fig.H)
AVG
AVG
8
Page 10
AVGAVGAVGAVG
AVGAVG
ppmp pm
OOCOO
C
Fig.G
Fig.H
- Press to view average reading of other parameters.
- Press to return to normal mode.
SETUP
The setup mode is for those related parameters setting that can ensure more accurate measurements. It includes: P1.0: Unit selection P2.0: AREA size P3.0: RH offset P4.0: Pressure input (model 89191 only)
In normal mode, press for 2 seconds to enter SETUP mode. It starts with P1.0 unit selection. Press or to select among P1.0 to P4.0. Long press again to return to normal mode.
P1.0: Unit selection
In P1.0 (Fig.I), press to enter units selection (Fig.J). Then press or to switch from Metric to Imperial unit. The units include air velocity(m/s, fpm), temp.(C, F), air volume(cmm,cfm), and area size(cm ,inch ). After finishing the setting, press to return to P1.0.
2 2
2
2
cm
cm
m/s
OO
OO
C
C
CMM
CMM
Fig.I
Fig.J
9
Page 11
P2.0: AREA Size
In P2.0(Fig.K), press to enter AREA size setting. It shows “99999” on the lower display with the first digit flashing. (Fig.L) The flashing digit means it is ready to adjust. Press to change the number and it will cycle from 0 to 9. Press to move to next digit. After completed, press again to return to P2.0.
2
cm2cm
Fig.K
Fig.L
P3.0: RH offset
In P3.0(Fig.M), press to enter RH offset setting. It shows blinking 0.0 RH reading on the upper display(Fig.N). Press to increase the offset value and to decrease it. When finished the setting, press again to return to P3.0.
%RH%RH
Fig.M
Fig.N
P4.0:Pressure input (89191 only)
In P4.0 (Fig.O), press to enter pressure input. It shows default 1013 hpa on the display with blinking value at the upper (Fig.P). Press to increase the pressure value and to decrease it. Once finished, press again to return to P3.0.
10
Page 12
Fig.O
Fig.P
CALIBRATION
RH CALIBRATION
The humidity calibration of this meter requires specific fixture and cannot be done by end users. Please contact the
dealer for calibration service.
CO2 CALIBRATION(89191 only)
The meter is calibrated at standard 400ppm CO concentration in factory. It’s suggested to do manual calibration regularly to maintain good accuracy.
Note:
When the accuracy becomes a concern after a long time use, return to dealers for standard calibration.
400ppm calibration
400ppm calibration is suggested to be done in outdoor fresh air that is well ventilated and in sunny weather.
CAUTION:
Do not calibrate the meter in the air with unknown CO concentration. Otherwise, it will be calibrated as default and leads to inaccurate measurements.
2
2
400ppm
Place the meter in the calibration site and follow the below procedure to complete the 400ppm calibration.
11
Page 13
1. While the meter is turned off, hold down , and simultaneously to enter CO2 calibration mode. It runs 30 seconds countdown with CAL at lower display(Fig.T), and then starts 400ppm calibration.
2. During calibration, a CO2 reading in
the range of 380 to 420ppm will be flashing on the upper display.(Fig.U)
3. Wait for 10 minutes until the reading stops flashing, the 400ppm calibration is completed and it returns to normal mode automatically. . To abort the calibration, turn off the meter at any time.
ppm
Fig.T
Fig.U
NOTE:
Be sure the batteries are with full voltage during the calibration to prevent from interruption and fail in the calibration.
LOW-BATTERY
When the battery voltage falls below the measurement required range, shows on the LCD. Replace with fresh batteries to ensure accurate measurements.
12
Page 14
Nice Flow App installation
Smart device system requirement iOS version above iOS Android version above 5.0
Please always remember to enable the wireless function on smart device before using this App.
App function
Main Page
Preference display
User can choose which parameters to be
displayed at the main page. Default setting is to display all parameters, press “ star ” button next to the parameter to remove the one you don’t want to see at the main page, and press again to add it back.
Main Page Press to return to the main page and
view the current measuring values of different meters.
(Maximum display 5 meters at one time)
Files Press to check the history data of
Avg-Min-Max data.
About
Information of privacy and version.
Return
Return to previous page.
13
Page 15
Avg-Min-Max by App
This App is able to calculates multi-points average / timed average, and display together with Minimum / Maximum values of all parameters after starts the recording process.
User can determine when to start and when to stop the calculation, or when to take the measurements under Point Avg. Mode by App.
The data will be saved automatically after stop recording, it can be output in csv format for further analyze.
Timed average
- Select one device at one time on main page, you will see the detail of current measurements of the specific meter.
- Press “Time Avg.” to enter the Timed­ average mode.
- Press “Start” to start recording, the upper right corner display the time duration of recording.
- Press “Stop” to stop recording, the data will be saved automatically.
-Check the recording detail at “Files”
Multi-points average
- Select one device at one time on main page, you will see the detail of current measurements of the specific meter.
- Press “Point Avg.” to enter the Point Average mode.
- Press “Start” to record, the upper right corner display the total record point
- Press “Stop” to stop recording, the data will be saved automatically.
-Check the recording detail at “Files” 14
Page 16
Data Review & Output by App
Search Recorded Data
- Press “File” to enter the search page
- Enter serial number or file name at the upper blank area, press “icon ” to search.
- The search result will be listed, click “File Name” to see the detail.
Edit File Name
- Select the file and you will see the “Rename” option appear.
- Click “Rename”, click edit, enter file name.
- Press Save to save the file name.
Send in CSV format
- Select the file and you will see the “CSV” option appear
- You can select multiple files to send them in CSV format for further analyze.
Delete File
- Select the file and you will see the “delete” option appear.
- Click “delete” to delete the file, the delete file will not be able to resume, please be sure before press click.
15
Page 17
Setting by App
User is able to do the setting by App.
- Select one meter on main page
- Press “ SET ” on upper left corner for
general setting list as below:
Device Name
- Press Device Name to start.
- Enter maximum 6 characters for
device name setting.
- Press ok to confirm.
Unit Selection
- Press Unit to start.
- Select Metric or Imperial.
- Press ok to confirm.
RH off set(89171/89191 only)
- Press RH off set to start.
- Enter the number between 0~12 for
RH off set setting.
- Press ok to confirm.
Pressure input (89191 only)
- Press Absolute pressure to start.
- Enter number between 700~1990 for
pressure compensation.
- Press ok to confirm.
Outlet Area input
User can also input the outlet area via App, please follow below steps:
- Ensure airflow cone is not installed
- Press the meter you would like to input
outlet area on main page
- Press “Volume Flow” for outlet area
input, 3 options available:
*Area: Enter the calculated outlet area *Diameter: If the outlet is in round
shape, enter the diameter *LxW: Enter the Length and Width of the outlet area
16
Page 18
TROUBLESHOOTING
1. Power on but no display
1) Make sure the time of pressing key is more than 200mS.
2) Check the batteries are in good contact with correct polarity.
3) Replace new batteries and try again.
2. Display disappeared Check whether the low battery indicator shows before display disappears, if yes, replace with new batteries.
3. Error code and solution
1)Air Temp. E02: Temp. is under low range limit. Put the meter in room temp for 30 minutes. If E02 still shows, send to the dealer for repair. E03: Temp. is over high range limit. Put the meter in room temp for 30 minutes. If E03 still shows, send to the dealer for repair. E31: temperature measuring AD damaged. Send to the dealer for repair.
2) RH
E04: temperature error caused the RH error. Check the temp. error solution. E11: RH calibration error. Try recalibration or send for repair if it still can’t work. E33: RH measurement circuit error. Send to the dealer for repair.
17
Page 19
3) DP and WBT
E04: Temp. or RH error. Check temp. or RH error solution.
4) Air velocity
E03: Velocity is over high range limit. Try in range measurement. If it still shows in range velocity, recalibrate it send for repair.
5) Air volume
E03: Reading exceeds display limit. Check if the outlet area setting is correct. E04: Air velocity error. Try recalibration or send for repair if it still can’t work.
6)CO2
E03: Temp. is over high range limit. Put the meter in room temp. for 10 minutes and if E03 still shows, recalibrate it or send for repair.
E01/E33: Voltage too low or CO2 module error. Check if battery low icon showed on the LCD, if yes, replace with new batteries. Try recalibration, or send for repair if it still can’t work.
7) Others
E32: Memory IC error. Turn off the meter and turn it on again. If E32 still shows, send to the dealer for repair.
18
Page 20
TROUBLESHOOTING FOR APP
1. App no display
1) Make sure meter is power on with display.
2) Turn off other BLE device.
3) Close the app and switch it on again.
FCC Radiation Exposure Statement
This equipment complies with FCC RF radiation exposure limits set forth for an uncontrolled environment. This equipment should be installed and operated with a minimum distance of 0.5 centimeters between the radiator and your body. This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter. The antennas used for this transmitter must be installed to provide a separation distance of at least 0.5 cm from all persons and must not be co-located or operating in conjunction with any other
antenna or transmitter.
19
Page 21
SPECIFICATION
Measuring r ange
Air velocity 0.20~30.00m/s Air volume
Temperature Relative Humidity DP(Dew point temp.)
WBT(Wet bulb temp.) CO2
Resolution
Accuracy
Air velocity +/-(1.5% of reading +0.3m/s) for under 20m/s
Temperature
Relative Humidity
CO2
CO2 Warm-up time
Vane / Meter siz e (mm)
Air flow cone size(mm) Round Dia. 210 Square 346x346
Operating condition (avoid condensation) Storage c ondition
Power supply
Standard package Optional accessory Air flow cone set ( w /soft carry case ),
89161 89191
0.1m/s, 0.1(0~9999.9) or 1(10000~99999), 0.1 C/ F, 0.1%RH
Meter, manual ,AAA bat.,Hard carry case
Extension set
Certification
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including interference that may cause undesired operation.
89171
3
0 to 99,999 m /m
o o
-20.0~60.0 C (-5~140 F) N/A N/A N/A
N/A N/A
+/-(3% of reading +0.3m/s) for above 20m/s
N/A N/A
N/A
Dia. 100, 269(L)x106(W)x51(H)
-10~ 50 C, <90%RH(avoid condensation)
0.1%~99.9%RH
o
-5.0~59.9 C
o
-20.0~59.9 C 0~9999 ppm,
(5001~9999 ppm out o f scale range)
o o
o
+0.6 C
o
+3%RH(a t 25 C, 10~90%RH) ; +5%RH( oth ers )
+30ppm+5% of rdg.(0~5000ppm)
N/A
N/A
0 to 50 °C, <80%RH
o
AAA x 4pcs
30 seconds
20
Page 22
WARRANTY
The meter is warranted to be free from defects in material and workmanship for a period of one year from the date of purchase. This warranty covers normal operation and does not cover battery, misuse, abuse, alteration, tampering, neglect, improper maintenance or damage resulting from leaking batteries.
Proof of purchase is required for warranty repairs. Warranty is void if the meter has been opened .
RETURN AUTHORIZATION
Authorization must be obtained from the supplier before returning items for any reason. When requiring a RA (Return Authorization) , please include data regarding the defective reason, the meters are returned along with good packing to prevent any damage in shipment and insured against possible damage or loss.
21
Page 23
Accuracy, the Zenith of
Measuring / Testing Instruments !
Anemometer Data logger Datalogging printer Gas Leak Detector CO meter CO2 meter Hygrometer/Psychrometer RTD probe Thermometer K.J.T.R.S.E. type Thermometer Heat Index Meter / WBGT Meter Infrared Thermometer Manometer Tachometer Sound Level Meter pH Meter Conductivity Meter T.D.S. Meter D.O. Meter ORP Meter
More products available !
2018.09 v.01
Loading...