
(1) Turn meter on by pressing ON/OFF
button.
(2) Press F/C button to convert reading
to desired range.Both temperature
and relative humidity measurement
will display simultaneously.(8708)
NORMAL MEASUREMENT
INTRODUCTION
8708
Therm-
Hygrometer
O
N
/O
FF
HOLD
MIN/MAX
TEMP
(F/C)
RESET
td
WB
MIN
MAX
C
F
%
HLD
8
70
8
ThermoHygrometer
O
N
/O
FF
HOLD
MIN/MAX
TEMP
(F/C)
RESET
td
WB
MIN
MAX
C
F
%
HLD
8709
Therm-
Hygrometer
Congratulation on your purchase of
the Pen type Thermo-Hygromter .
This is a low cost but easy use to get
accurate measurement .
The difference between model 8708
and 8709 is the latter one features
self-calibration by using 0033RH/
0075RH salt solution to maintain the
meter to get a reading .
The front panel of 8708 with TEMP
(F/C) printing , instead 8709 without
a button with printing F/C,CAL .
ccurate
td
WB
MIN
MAX
C
F
%
HLD

(1)Power on the meter by pressing
ON/OFF button. You will see a full
display on the screen ( though you
might have seen "td","WB" shown
on the screen, however these icons
are only available for other models )
, then "HLD" icon will disappear in
one second ,then after 3 seconds,
the meter will start measuring the
temperature and humidity and
display both readings at the same
time.
POWER ON THE METER
(1) Turn meter on by pressing ON/OFF
button.
(2) Use a pen tip to press F/C key
whenever you need to.
Press F/C to convert reading to
desired range. Both temperature
and relative humidity measurement
will display simultaneously.(8709)
READ BEFORE USE
td
WB
MIN
MAX
C
F
%
HLD
td
WB
MIN
MAX
C
F
%
C
%

(2) The default setting of temperature
oo
display can be in F or C. You can
press F/C button for 2 seconds to
change the temperature display
from C to F or vice versa. You
may use a pen tip to press F/C
button whenever convenient.(8709)
(1) Press and hold MIN/MAX button
until MIN appears on display.
The LCD screen is now showing
the minimum humidity and tem perature readings in memory.
MIN/MAX FUNCTION
2
C
%
F
%
C
%
MIN
(2) Press and hold MIN/MAX button
again until MAX appears on
display. The LCD screen is now
showing the maximum humidity
and temperature readings in
memory.

MAX
C
%
(3) To return to measuring mode of
current temperature and humidity,
press and hold MIN/MAX button
until MIN or MAX disappears from
the display.
(4) To clear current MIN and MAX
readings from memory, press and
hold RESET button until the full
display flashes .The meter reading
turns to the normal measurement
with either MIN or MAX icon , after
about 10 second, the MIN/MAX icon
will disappear automatically .
C
%
DATA HOLD FUNCTION
Press hold button until "HLD" appears
in display. The current reading is now
being held and will not change until
Hold function is cancelled. Press
RESET button to cancel Hold function.
Hold function can be used for Humidity
and Temperature .
C
%
HLD

AUTO POWER OFF
3
Unit will turn itself off after 20 minutes.
(Sleep mode).To override Auto Power
Off function (Non-Sleep Mode) press
ON/OFF ,MIN/MAX keys simultaneously.
After the full display appears release
ON/OFF button, the display shows
"n" on the screen. When "n" appears
then release MIN/MAX button. Unit
will now be in Non-sleep mode .
1. Turn off the meter.
2. Press ON/OFF button. When LCD
displays full readings, press F/C
button with a pen tip and the LCD
shows original temperature default
C or F right down the corner.
3. Press MAX/MIN button to switch
the default C or F. Press F/C button
to save the new temperature scale
default and LCD will first show "S"
then show "%MIN" right up the corner.
Turn off the meter.
The procedure is completed.
CHANGE DEFAULT OF
TEMPERATURE SCALE
(8709 only)

CHECK AND CALIBRATION
The carrying case is for not only
storage (see FIG1)but also calibra tion (see FIG2).
4
C
MIN
%
F
FIG 1
FIG 2
HR 33 MgCl
TEMP
RH%
ooooo
15 C 20 C 25 C 30 C 35 C
33.30 33.07 32.78 32.44 32.05
(8709 only)

NOTE:
A convenient look-up table has been
built in to aid calibration.
HR 33 MgCl
TEMP
RH%
ooooo
15 C 20 C 25 C 30 C 35 C
33.30 33.07 32.78 32.44 32.05
o
If different temperatures, say 25 C,
o
the RH of MgCl is 32.78 and in 30 C,
the RH is 32.44.
o
The same reference says 30 C , the
RH of Nacl is 75.09 RH% humidity
o
and in 35 C , the RH is 74.87% .
(See the table)
HR75 NaCl
TEMP
RH%
ooooo
15 C 20 C 25 C 30 C 35 C
75.61 75.47 75.29 75.09 74.87
The relations are recorded in the look
-up table.
Once the calibration mode in ON, the
meter will search the right humidity
value according to the current temperature to get more accurate calibration.
Before checking the humidity 33% or
75% , make sure you purchased model
8709 and salt solution 33% ( dark
brown bottle) , 75% ( blue bottle ) as
well as the carry case, if you purchased
model 8708 , it is the model not available for checking or re-calibration .

1.1. CHECK HUMIDITY 33%
1.1.1. Turn on the meter. Plug unit
in 33% salt reference bottle and
put unit into the carrying case as
shown FIG2. Check the reading
on LCD after 40 minutes. If the
display reads between 31.5% and
34.5%, that means the meter is
accurate. Otherwise, it needs
to be recalibrated.
Please calibrate the meter by
following calibration instruction
1.3.A.
1.2.CHECK HUMIDITY 75%
1.2.1. Turn on the meter.Plug unit
in 75% salt reference bottle as you
do in step1.1.1. Check the reading
on the LCD after 40 minutes. If the
display reads between 73.5% and
76.5%, that means the meter is
accurate. Otherwise, it needs to
be calibrated. Please calibrate the
meter by following calibration
instruction 1.3.B
(8709 only)
1.3.CALIBRATION
A. CALIBRATION FOR
RELATIVE HUMIDITY 33%
1. Turn off the meter.
2. Press ON/OFF button. When LCD
displays full readings, press F/C
button after plugging the meter in
the 33% salt reference bottle for 40
minutes(you have done in 1.a).
3. Press RESET button and the LCD
shows " %MIN ".

4. Press F/C button to save the cali-
bration value of 33%. The LCD
shows "S" for 2 seconds and then
the "%MAX" shows right up the
corner of LCD.
5. Turn off the meter. The calibration
of 33% is completed.
5
MIN
%
MAX
%
B .CALIBRATION FOR
RELATIVE HUMIDITY 75%
1. Turn off the meter.
2. Press ON/OFF button. When LCD
displays full readings, press F/C
button after plugging the meter in
the 75% salt reference bottle for 40
minutes.( you have done in 1.2.1)
3. Press RESET button twice and the
LCD shows "%MAX" .

6
MAX
%
AUTO CALIBRATION
a.
If you want to calibrate the meter
without checking the accuracy,
Please plug the meter probe into
the 33% salt bottle and then power
on the meter by pressing ON/OFF
button and the F/C button at the
same time.
4. Press F/C button to save the
calibration value of 75%. The LCD
will show " S " for 2 seconds and
then resume normal measurement
5. Turn off the meter. The calibration
of 75% is completed.
When LCD screen shows a full
display, release the ON/OFF button.
You will see the display of "C"
and " F" at the lower right corner of
LCD screen, then release the F/C
button. The LCD will display only
the default setting of temperature
unit, C or F.
Press RESET key and you will see
a "% MIN" display on the right
upper corner. Please leave the
meter in the bottle for at least 40
minutes and the meter will start the
auto calibration process.
b.
(8709 only)

When the calibration for 33% is
done, you will see a "MIN " and
"MAX" display on the LCD screen.
Please remove the meter probe
from the 33% salt bottle and plug
it into the 75% salt bottle.
Then press the MIN/MAX button,
you will see a "% MAX" display
on the LCD. Then leave the meter
in the 75% bottle for another 40
minutes. When the calibration for
75% is completed, the meter will
be back to normal measuring
mode for temperature and humidity.
You can then remove the meter
from the 75% bottle.
c.
CHANGE BATTERY
(1) Turn off the meter , then use a coin
to open the battery cover.
(2) Put the battery (CR2032) whose
"+" faces out into the battery
compartment and place the battery
cover back and lock it with a coin.
8
MATERIAL SUPPLIED
The package supplied with a standard :
1) The unit
2) Battery CR2031 x1
3) Operation manual
Optional accessory : ( model 8708,8709)
a) 33% Salt solution
b) 75% Salt solution
c) Carrying Case

Size: 17.5cm 4.2cm 1.6cm
Battery: (1) 2032 Life-150hrs.
Sensor: Electronic Capacitance
Ploymer Film Sensor
Accuracy: 5% from 5 to 95%RH
oo
at 25 C 1 C from~
o
10 to 45 C
o
o
Resolution:0.1 C/ 0.1 F
Remember Low Batteries will
tend to give inaccurate readings,
so make sure you have good
batteries.
Please do not immerse probe
into liquids as this will cause
permanent damage to the
sensor.
SPECIFICATION
9
The meter is warranted to be free from
defects in material and workmanship
for a period of one years from the date
of purchase.
This warranty covers normal operation
and does not cover batteries, misuse,
abuse, alteration, tampering, neglect,
improper maintenance, or damage
resulting from leaking batteries.
Proof of purchase is required for
warranty repairs.
WARRANTY

RETURN AUTHORIZATION
Authorization must be obtained from
the supplier before returning items for
any reason .
When requiring a RA (Return
Authorization) , please include data
regarding the defective reason, the
meters are to be returned along with
good packing to prevent any damage
in shipment and insured against
possible damage or loss .
The meter conforms to the following
standards:
* EN 50081-1/1992 : EN 55022
* EN 50082-1/1997 : EN 55024
(EN 61000-4-2/-3/-8, ENV 50204)
, the meter complies with the essential
protection requirements of Council
Directive 89/336/EEC on the approximation of the laws of the Member
States relating to electromagnetic
compatibility.
CE APPROVAL

Accuracy, the Zenith of
Measuring / Testing Instruments !
More products available !
Hygrometer / Psychrometer
Thermometer
Anemometer
Sound Level Meter
Air Flow meter
Infrared Thermometer
K type Thermometer
K.J.T. type Thermometer
K.J.T.R.S.E. type Thermometer
pH Meter
Conductivity Meter
T.D.S. Meter
D.O. Meter
Saccharimeter
Manometer
Tacho Meter
Lux / Light Meter
Moisture Meter
Data logger
Temp. / RH transmitter
Wireless Transmitter ..........
2002/08,1000