Azbil MagneW FLEX+, MagneW PLUS+, MGG11, MGG19, MGF11 User Manual

...
CM2-MGG200-2001
MagneW FLEX+/PLUS+
Electromagnetic Flowmeter Detector
Model MGG11/18/MGG12/19/MGF11/MGS11U/28U
User’s Manual
NOTICE
While the information in this manual is presented in good faith and believed to be accurate, Azbil Corporation disclaims any implied warranty of merchantability or tness for a particular purpose and makes no express warranty except as may be stated in its written agreement with and for its customer.
In no event shall Azbil Corporation be liable to anyone for any indirect, special or consequential damages. is information and specications in this document are subject to change without notice. MagneW is a trademark of Azbil Corporation in Japan and/or other countries.
© 1997-2019 Azbil Corporation. All Rights Reserved.
Safety
About This Manual
This manual contains information and warnings that must be observed to keep the model MGG/MGS Electromagnetic Flowmeter Detector operating safely. Correct installation, correct operation and regular maintenance are essential to ensure safety while using this device.
In addition to this model MGG/MGS detector user manual, please refer to model MGG10C/14C converter user’s manual for start-up, calibration, trouble-shooting, operation, and maintenance of the complete Electromagnetic Flowmeter.
For the correct and safe use of this device it is essential that both operating and service personnel follow generally accepted safety procedures in addition to the safety
precautions specied in this manual.
The following symbols are used in this manual to alert you to possible hazards:
Denotes a potentially hazardous situation which, if not avoided, could result in death or serious injury.
WARNING
Failure to observe these precautions may produce dangerous conditions that could result in operator injury or in physical damage to the device.
Safety Messages
Carefully read this section before in stalling or operating this device.
To prevent the gas or liquid in the pipe from escaping, do not remove the electrode cover or the electrodes when the detector is installed on a pipe.
ELECTRIC SHOCK HAZARD! Turn the power supply o before wiring.
CAUTION
WARNING
WARNING
WARNING
This detector is heavy. Dropping it could cause personal injury or damage to the device.
MagneW FLEX+/PLUS+ Electromagnetic Flowmeter Detector i
Safety Azbil Corporation
WARNING
Before removing the unit, make sure that there is no residual liquid or pressure inside the piping and the detector to avoid personal injury or damage to the unit.
CAUTION
The detector must be grounded (grounding reference is <100 W) to avoid output
uctuation, zero point instability or output drift.
CAUTION
The installation location must satisfy the following conditions to avoid output
errors and uctuations.
At the installation location:
The electrical conductivity of the uid to be measured must match the stated
specication (specs vary according to the converter used) and be more or less
constant.
The uid to be measured must be electrochemically homogenized. For example,
if two uids are mixed at an upstream point, the device should be installed at a point so that the two uids can be evenly mixed by the time they reach the
measuring point.
• The distribution of additive matter, if any, must be nearly uniform.
CAUTION
To avoid measurement problems, do not use the MagneW
FLEX+/PLUS+ to measure
the following uids, even if their electrical conductivity, temperature and pressure fall within the specications.
Fluids that have sucient electrical conductivity at high temperatures but do not satisfy the conductance requirements at room temperature (about 20°C (68°F)).
For example, fatty acids or soap.
Certain uids that contain surfectants. For example, rinses, shampoos, CWM.
Conductive adherents. For example, deposition of rosin or conductive material.
Insulating adherents. For example, oil, kaolinite, kaolin, or calcium stearate.
CAUTION
This model MGG/MGS detector can only be connected to Azbil Corporation's model MGG10C/14C converter.
No other types of converters can be used with the device.
ii MagneW FLEX+/PLUS+ Electromagnetic Flowmeter Detector
Azbil Corporation Safety
CAUTION
If the pipe is not lled, output errors will occur.
CAUTION
Do not rotate the unit more than 180° (one half rotation). Any further rotation can disconnect wiring.
After removing the screws, do not pull on the terminal box. You can break the
lead wire.
When you move the terminal box, make sure the O-ring remains in the groove to
insure an air-tight seal when the unit is reassembled.
CAUTION
Before installing the detector, make sure any foreign matter is ushed from
the interior passage of the detector. Residual foreign matter could cause output
uctuations.
Do not touch the electrodes or allow oil or fat to come into contact with them to
avoid output uctuations.
Align the direction mark on the detector in the direction of the liquid ow.
Misalignment can result in a negative output.
CAUTION
The correct fastening torque must be used to prevent leakage. To avoid damage to
the detector, do not exceed the listed fastening torque.
CAUTION
Before installing the detector, make sure that the pipe is exactly straight and
centered. Any irregularity could cause leakage or other hazards.
CAUTION
Never attempt to force the detector between two anges if the space is too narrow.
It can damage the detector.
CAUTION
Make certain the bore diameters of the pipe and the detector are exactly the same
and install the detector so that the gasket does not protrude into the internal bore of the pipe, as this could result in leakage or other hazards.
MagneW FLEX+/PLUS+ Electromagnetic Flowmeter Detector iii
Safety Azbil Corporation
CAUTION
Tighten each bolt a little at a time and apply uniform pressure to all the bolts.
If leakage continues after tightening the bolts, make sure that the pipe is not o
center, then continue to tighten each bolt a little at a time.
CAUTION
Using a gasket with too small a diameter can aect the ow velocity distribution,
resulting in inaccurate measurement.
Using a gasket with too large a diameter can cause leakage. Also, any solid
substance in the uid to be measured could accumulate between the gasket and the ange, resulting in inaccurate measurement.
CAUTION
The necessary materials and the installation method vary according to the material of the ring and that of the pipe on which the detector is to be installed. Select the
appropriate method of installation from the following table after conrming the specications of the detector to be installed and the conditions of installation. Improper installation can result in leakage or damage to the pipe anges.
CAUTION
The use of anything other than the PTFE gaskets provided can result in leakage.
The use of rubber gaskets is not recommended and can cause leakage.
CAUTION
The use of rubber gaskets and a lower fastening torque is not recommended and
can cause insucient surface pressure between the lining and the grounding ring,
resulting in leakage.
CAUTION
Insucient grounding can cause output uctuations, instability of the zero point,
or output drift. Secure single point grounding with a grounding resistance of 100 W or less is recommended.
Do not ground a welder to the detector to avoid damaging the detector.
iv MagneW FLEX+/PLUS+ Electromagnetic Flowmeter Detector

Table of Contents

Chapter 1 : Introduction
1.1 : Principle of operation ......................................................................................... 1-1
1.2 : System operation ............................................................................................... 1-2
1.2.1: Integral type ........................................................................................................ 1-2
1.2.2: Remote type ........................................................................................................ 1-3
1.3 : Main components .............................................................................................. 1-4
1.3.1: Wafer type detector ............................................................................................. 1-5
1.3.2: Flange type detector ........................................................................................... 1-6
1.3.3: Union / Hose / Clamp type detector .................................................................... 1-7
1.3.4: Sanitary type detector ......................................................................................... 1-8
1.3.5: Detector terminal box .......................................................................................... 1-9
1.3.6: Approval of this device ........................................................................................ 1-10
Chapter 2 : Installation
2.1 : Unpacking .......................................................................................................... 2-1
2.1.1: Verifying specications ........................................................................................ 2-1
2.2 : Storage .............................................................................................................. 2-2
2.3 : Site selection ..................................................................................................... 2-3
2.3.1: Detector position ................................................................................................. 2-4
2.3.2: Changing the position of the terminal box ........................................................... 2-5
2.4 : Installing a wafer detector .................................................................................. 2-6
2.4.1: Determining the fastening torque ........................................................................ 2-7
2.4.2: Selecting the ange shape .................................................................................. 2-7
2.4.3: Necessary parts .................................................................................................. 2-9
2.4.4: Selecting an installation method ......................................................................... 2-12
2.4.5: Installation on a horizontal pipe ........................................................................... 2-13
2.4.6: Installation on a vertical pipe ............................................................................... 2-14
2.4.7: Installation on a metal pipe using SUS grounding ring material .......................... 2-15
2.4.8: Installation on a metal pipe using non-SUS grounding ring material .................. 2-16
2.4.9: Installation on a PVC pipe using SUS grounding ring material ........................... 2-19
2.4.10: Installation on a PVC pipe using non-SUS grounding ring material .................. 2-22
2.5 : Installing a anged detector ............................................................................... 2-25
2.5.1: Determining the fastening torque ........................................................................ 2-26
2.5.2: Necessary parts .................................................................................................. 2-31
2.5.3: Selecting an installation method ......................................................................... 2-32
2.5.4: Installation on a metal pipe using SUS grounding ring material .......................... 2-33
2.5.5: Installation on a metal pipe using non-SUS grounding ring material .................. 2-34
2.5.6: Installation on a PVC pipe using SUS grounding ring material ........................... 2-35
2.5.7: Installation on a PVC pipe using non-SUS grounding ring material .................... 2-39
2.6 : Installing a Union / Hose / Clamp Detector ........................................................ 2-43
2.6.1: Installation of Union and Hose assemblies basic installation method ................. 2-43
2.6.2: Installation of ISO-clamp/ Tri-clamp assembles basic installation method ......... 2-44
2.7 : Wiring the detector ............................................................................................. 2-46
2.7.1: Connecting the detector and converter (Remote models) .................................. 2-46
2.7.2: Grounding the detector (Remote models) ........................................................... 2-47
2.7.3: Installing the dedicated cable .............................................................................. 2-47
2.8 : Troubleshooting and maintenance .................................................................... 2-56
Table of Contents
Table of Contents
List of Figure
Figure 1-1 Faraday’s law..................................................................................................... 1-1
Figure 1-2 Integral conguration......................................................................................... 1-2
Figure 1-3 Remote Conguration......................................................................................... 1-3
Figure 1-4 Wafer detector main parts .................................................................................. 1-5
Figure 1-5 Flange detector main parts ................................................................................ 1-6
Figure 1-6 Union / Hose / Clamp detector main parts......................................................... 1-7
Figure 1-7 Details of the clamp type detector ...................................................................... 1-8
Figure 1-8 Detector terminal box main parts ...................................................................... 1-9
Figure 2-1 Detector placement ............................................................................................. 2-4
Figure 2-2 Straight pipe section on upstream side of detector ............................................ 2-4
Figure 2-3 Space allowance required for maintenance ....................................................... 2-5
Figure 2-4 Repositioning the terminal box........................................................................... 2-5
Figure 2-5 Installing a wafer detector ................................................................................. 2-6
Figure 2-6 Flange Shapes .................................................................................................... 2-7
Figure 2-7 Unacceptable positioning ................................................................................... 2-8
Figure 2-8 Unacceptable placement of detector between anges ........................................ 2-8
Figure 2-9 Centering nut positioning for horizontal centering ............................................ 2-9
Figure 2-10 Centering nut positioning for vertical centering ................................................ 2-9
Figure 2-11 Installation on a metal pipe using SUS grounding ring material and a specied
torque................................................................................................................ 2-15
Figure 2-12 Installation on a metal pipe using non-SUS grounding ring material and a
specied torque................................................................................................. 2-16
Figure 2-13 Installation on a metal pipe using non-SUS grounding ring material with rubber
gaskets .............................................................................................................. 2-17
Figure 2-14 Incorrect installation ........................................................................................ 2-18
Figure 2-15 Installation on a PVC pipe using SUS grounding ring material ...................... 2-19
Figure 2-16 Installation on a PVC pipe using SUS grounding ring material and a protective
plate .................................................................................................................. 2-20
Figure 2-17 Installation on a PVC pipe using SUS grounding ring material with rubber
gaskets (Not recommended) ............................................................................. 2-21
Figure 2-18 Installation on a PVC pipe using non-SUS grounding ring material .............. 2-22
Figure 2-19 Installation on a PVC pipe using non-SUS grounding ring material and a
protective plate ................................................................................................. 2-23
Figure 2-20 Installation on a PVC pipe using non-SUS grounding ring material with rubber
gaskets .............................................................................................................. 2-24
Figure 2-21 Installing a anged detector ............................................................................. 2-25
Figure 2-22 Flange Shapes .................................................................................................. 2-30
Figure 2-23 Unacceptable Placement of Detector Between Flanges .................................. 2-30
Figure 2-24 Installation on a metal pipe using SUS grounding ring material and a specied
torque................................................................................................................ 2-33
Figure 2-25 Installation on a metal pipe using non-SUS grounding ring material and a
specied torque................................................................................................. 2-34
Figure 2-26 Incorrect installation ........................................................................................ 2-35
Figure 2-27 Installation on a PVC pipe using SUS grounding ring material and a specied
torque................................................................................................................ 2-36
List of Figure
Figure 2-28 Installation on a PVC pipe using SUS grounding ring material and a protective
plate .................................................................................................................. 2-37
Figure 2-29 Installation on a PVC pipe using SUS grounding ring material with rubber
gaskets (not recommended) .............................................................................. 2-38
Figure 2-30 Installation on a PVC pipe using Non-SUS grounding ring material and a
specied torque................................................................................................. 2-40
Figure 2-31 Installation on a PVC pipe using non-SUS grounding ring material and a
protective plate ................................................................................................. 2-41
Figure 2-32 Installation on a PVC pipe using non-SUS grounding ring material with rubber
gaskets .............................................................................................................. 2-42
Figure 2-33 Example of installation (union assembly) ........................................................ 2-44
Figure 2-34 Example of installation ..................................................................................... 2-44
Figure 2-35 Connection using a dedicated cable ................................................................. 2-46
Figure 2-36 Grounding via the external grounding terminal .............................................. 2-47
Figure 2-37 Connection using a dedicated cable ................................................................. 2-47
Figure 2-38 Dedicated cable structure ................................................................................ 2-48
Figure 2-39 Trimming the outside sheath ............................................................................ 2-48
Figure 2-40 Attaching the C-wire ........................................................................................ 2-49
Figure 2-41 Trimming the outside shield ............................................................................. 2-49
Figure 2-42 Trimming the inside sheath .............................................................................. 2-50
Figure 2-43 Attaching the lead wires (SA and SB) .............................................................. 2-50
Figure 2-44 Trimming the inside shield ............................................................................... 2-51
Figure 2-45 Removing the conductive tube .......................................................................... 2-51
Figure 2-46 Trimming the insulator ..................................................................................... 2-51
Figure 2-47 Terminal connections ....................................................................................... 2-52
Figure 2-48 Trimming the outside sheath ............................................................................ 2-53
Figure 2-49 Attaching the C-wire ........................................................................................ 2-53
Figure 2-50 Trimming the outside shield ............................................................................. 2-54
Figure 2-51 Trimming the inside sheath .............................................................................. 2-54
Figure 2-52 Attaching the lead wires (SA and SB) .............................................................. 2-54
Figure 2-53 Attaching the lead wires (SA and SB) .............................................................. 2-55
Figure 2-54 Terminal connections ....................................................................................... 2-56
List of Figure
List of Table
Table 2-1 Fastening torque for wafer detectors ............................................................................... 2-7
Table 2-2 Recommended inner diameters of gaskets ...................................................................... 2-10
Table 2-3 Table 2-4
Table 2-5 Wafer detector installation methods ................................................................................ 2-12
Table 2-6 Fastening torque for anged detectors ............................................................................ 2-26
Table 2-7 Recommended inner diameters of gaskets ...................................................................... 2-31
Table 2-8 Flange detector installation methods ............................................................................... 2-32
Table 2-9 Fastening torque .............................................................................................................. 2-43
Table 2-10 Cable continuity .............................................................................................................. 2-52
Table 2-11 Cable continuity .............................................................................................................. 2-52
Table 2-12 Cable continuity .............................................................................................................. 2-56
Table 2-13 Cable continuity .............................................................................................................. 2-56
Inner and outer diameters of rubber gaskets of thickness 0.5 to 1 mm (0.02 to 0.04 inches) Inner and outer diameters of rubber gaskets of thickness 3 to 4 mm (0.12 to 0.16 inches)
... 2-11
...... 2-11
List of Table

Chapter 1 : Introduction

Induced voltage (E)
Excitation

1.1 : Principle of operation

The owmeter consists of two parts: a detector which is mounted in the pipeline and through which the measured liquid ows, and a converter which may be mounted
either integral with the detector or separately. The converter conditions and outputs the electrical signal from the detector or separately. The converter conditions and outputs the electrical signal from the detector.
The owmeter works based on the principle of Faraday’s law of electromagnetic
induction, which states that if an electrical conductor, in this case the measured
conductive liquid, passes through a magnetic eld a small electromotive force (EMF) is induced perpendicular to the eld and ow (refer to Figure 1-1).
Faraday’s law:
E = k BDV when
E = Induced voltage (EMF)
B = Strength of the magnetic eld
D = Conductor width (electrode spacing)
V = Velocity of the conductor
k = Correction factor
This induced EMF is proportional to the average ow rate and is detected by
two electrodes (E converter.
The only variable in this application of Faraday’s law is the ow rate (V) of the conductive measured liquid, because eld strength (B), is controlled constant and electrode spacing (D) is xed.
Therefore, the output electromotive force is directly proportional to liquid ow rate, resulting in the linear output of a magnetic owmeter.
and E2) mounted in the wall of the detector and then fed to the
1
Magnetic field (B)
Flow rate (V)
voltage
supply
Measuring fluid
E1
E2
Figure 1-1 Faraday’s law
MagneW FLEX+/PLUS+ Electromagnetic Flowmeter Detector 1-1
Introduction Azbil Corporation

1.2 : System operation

Introduction
Depending on how it is combined with the converter, this product is available in two
congurations, integral and remote.
Integral: Detector and converter are installed as an integrated unit on a pipe.
Remote: Detector and converter are installed connected by cable.

1.2.1: Integral type

Figure 1-2 and Figure 1-3 show examples of measurement systems using the device.


/


Figure 1-2 Integral conguration



1-2 MagneW FLEX+/PLUS+ Electromagnetic Flowmeter Detector
Azbil Corporation Introduction
Contact input/output

1.2.2: Remote type

Converter
AC power supply
Pulse output,
Analog output/
Digital output
Dedicated cable
Detector
Excitation output
Flow signal input
Direction of fluid
Figure 1-3 Remote Conguration
MagneW FLEX+/PLUS+ Electromagnetic Flowmeter Detector 1-3
Introduction Azbil Corporation

1.3 : Main components

Detector
There are three types of model MGG/MGS Electromagnetic Flowmeter Detectors: Wafer detectors, Flange detector, and Union / Hose / Clamp detector. The type of
detector you choose depends on your specic installation requirements. All types of detectors function such that when a conductive uid passes through the detectors, an electromotive force signal proportional to the ow rate is generated.
WARNING
To prevent the gas or liquid in the pipe from escaping, do not remove the electrode cover or the electrodes when the detector is installed on a pipe.
1-4 MagneW FLEX+/PLUS+ Electromagnetic Flowmeter Detector
Azbil Corporation Introduction
Grounding terminal
Waterproof gland
Grounding ring
Mounting screw (4 places)

1.3.1: Wafer type detector

A wafer detector consists of the components shown in the gure below.
Terminal box cover
Terminal box
Flow direction mark
Electrode cover
Figure 1-4 Wafer detector main parts
Name Function
Flow direction mark Indicates the direction of uid ow.
Mount the detector so that the measured uid ows in the
direction indicated by this mark.
Electrodes Detect an electromagnetic force signal proportional to the
ow rate of the uid passing through the detector.
The electrode material varies depending on the corrosion
characteristics of the uid to be measured.
Electrode cover Houses the electrodes. Do not remove the cover with the
detector installed on a pipe.
Grounding rings Keep reference voltage as zero by grounding the unit.
The grounding ring material varies depending on the
corrosion characteristics of the uid to be measured.
Terminal box (Remote model only)
Houses the connection terminals used for applying
a standard voltage. Houses the excitation and signal
terminals.
Terminal box cover (Remote model only)
MagneW FLEX+/PLUS+ Electromagnetic Flowmeter Detector 1-5
Keeps the terminals dry and protected. Keep the terminal
box cover on during operation.
Introduction Azbil Corporation
Mounting screw
(4 places)
Grounding terminal
Grounding ring

1.3.2: Flange type detector

A Flange detector consists of the components shown in the gure below.
Waterproof gland
Terminal box cover
Terminal box
Flow direction mark
Flange
Electrode cover
Figure 1-5 Flange detector main parts
Flow direction mark Indicates the direction of uid ow. Mount the detector
Electrodes Detect an electromagnetic force signal proportional to
Electrode cover Houses the electrodes. Do not remove the cover with the
Grounding rings Keep reference voltage as zero by grounding the unit.
Terminal box (Remote model only)
Terminal box cover (Remote model only)
Name Function
so that the measured uid ows in the direction indicated
by this mark.
the ow rate of the uid passing through the detector.
The electrode material varies depending on the corrosion
characteristics of the uid to be measured.
detector installed on a pipe.
The grounding ring material varies depending on the
corrosion characteristics of the uid to be measured.
Houses the connection terminals used for applying
a standard voltage. Houses the excitation and signal
terminals.
Keeps the terminals dry and protected. Keeps the
terminal box cover on during operation.
1-6 MagneW FLEX+/PLUS+ Electromagnetic Flowmeter Detector
Azbil Corporation Introduction
Grounding terminal
Terminal box cover

1.3.3: Union / Hose / Clamp type detector

Waterproof gland
Terminal box
Mounting screw(4 places)
Flow direction mark
Union
Electrode cover
Clamp
Hose
Figure 1-6 Union / Hose / Clamp detector main parts
Name Function
Flow direction mark Indicates the direction of uid ow.
Mount the detector so that the measured uid ows in the
direction indicated by this mark.
Electrodes The electrodes generate an electromotive force signal
proportional to the ow rate of the uid passing through
the detector. The electrode material varies depending on
the corrosion characteristics of the uid to be measured.
Electrode cover Houses the electrodes. Do not remove the cover with the
detector installed on a pipe.
Union (Connected by screws) Hose
The connection uses a union, hose, and clamp. The material is SUS316. Applicable for detector bore diameters of 2.5 to 15 mm.
Clamp
Terminal box (Remote model only)
Houses the connection terminals used to apply a standard voltage.
Houses excitation and signal terminals.
Terminal box cover
Keeps the terminals box cover on during operation.
(Remote model only)
MagneW FLEX+/PLUS+ Electromagnetic Flowmeter Detector 1-7
Introduction Azbil Corporation
Terminal box
Grounding
Mounting screw
Flow direction mark

1.3.4: Sanitary type detector

Figure 1-7 shows the structure of the detector and the names of its major parts.
Waterproof gland
Electrode cover
terminal
cover
Terminal box
(4 places)
Ferrule
Figure 1-7 Details of the clamp type detector
Name Function
Flow direction mark Indicates the direction of uid ow.
Mount the detector so that the measured uid ows in the
direction indicated by this mark.
Electrodes The electrodes generate an electromotive force signal
proportional to the ow rate of the uid passing through
the detector. The electrode material varies depending on the corrosion
characteristics of the uid to be measured.
Electrode cover Houses the electrodes. Do not remove the cover with the
detector installed on a pipe.
Clamp Clamp structure varies IDF clamp and tri-clamp.
Terminal box (Remote model only)
Houses the connection terminals used to apply a standard voltage.
Houses excitation and signal terminals.
Terminal box cover
Keeps the terminals box cover on during operation.
(Remote model only)
1-8 MagneW FLEX+/PLUS+ Electromagnetic Flowmeter Detector
Azbil Corporation Introduction
Grounding terminal
Excitation terminals

1.3.5: Detector terminal box

The detector terminal box houses the connection terminals used for applying a standard voltage. Houses the excitation and signal terminals.
WARNING
ELECTRIC SHOCK HAZARD! Turn o power to the converter side before
wiring.
CAUTION
The detector must be grounded (grounding resistance is <100 W) to avoid output
uctuation, zero point instability or output drift.
Signal terminals
Conduit wiring connectors
Figure 1-8 Detector terminal box main parts
Name Function
Signal terminals Marked A, B and C.
Excitation terminals Marked X and Y.
Conduit wiring connectors
The excitation cable and the signal cable are wired
through these connectors.
Grounding terminal This terminal is used to ground the detector (grounding
resistance is <100 W).
MagneW FLEX+/PLUS+ Electromagnetic Flowmeter Detector 1-9
Introduction Azbil Corporation

1.3.6: Approval of this device

Overview
If 1/2 NPT wiring connection is selected, this device functions as an FM/CSA, non­incendive -approved model. In this case, the installation standards described in this section must be followed.
Installation of this device
FM/CSA Nonincendive model
THIS EQUIPMENT IS SUITABLE FOR USE IN CLASS I, DIVISION 2, GROUPS (A, B, C, D), CLASS II/III, DIVISION 2, GROUPS (F, G), OR NON-HAZARDOUS
LOCATIONS ONLY.
CAUTION
(1) Power supply and internal voltage of ordinary equipment to the earth shall not
exceed 250 V AC 50/60 Hz, 250 V DC in case on normal / formal conditions.
(2) Ambient temperature is from -25 to 60°C
(3) Process temperature is from -40 to 160°C (Remote model).
Process temperature is from -40 to 120°C (Integral model)
1-10 MagneW FLEX+/PLUS+ Electromagnetic Flowmeter Detector
Azbil Corporation Introduction
If an MGG18/19 detector is used with an MGG14C converter as an FM-approved nonincendive product, both the detector and the converter should be FM-approved nonincendive products.
If they are not, the MGG18/19 detector cannot be used as an FM-approved nonincendive product.
MagneW FLEX+/PLUS+ Electromagnetic Flowmeter Detector 1-11
Introduction Azbil Corporation
1-12 MagneW FLEX+/PLUS+ Electromagnetic Flowmeter Detector

Chapter 2 : Installation

This section describes unpacking, installation and wiring of the MagneW FLEX+/
.
PLUS+
Depending on the grounding ring or piping material, the required parts or the method of device installation vary.

2.1 : Unpacking

This device is a precision instrument and should be handled with care to prevent damage or breakage.
WARNING
This detector is heavy. Dropping it could cause personal injury or damage to the device.
After unpacking the unit, verify that the following items are present:
MagneW
Standard accessories
Precautions for installation sheet
FLEX+/PLUS+ Detector
2.1.1: Verifying specications
The specications for this device are written on its attached identication plate. Compare these specications with those listed in Appendix A, device standard specications and model numbers, and verify that all specications on the plate are
correct, paying special attention to the following:
Detector size
Electrode material
Lining material
Flange rating
Grounding ring material
If you have questions regarding the specications of your device, contact your nearest Azbil Corporation oce or the azbil Group representative. When making an inquiry,
be sure to provide the model number and product number of this device.
MagneW FLEX+/PLUS+ Electromagnetic Flowmeter Detector 2-1
Loading...
+ 61 hidden pages