Be sure that the user receives this manual before the product is used.
While this information is presented in good faith and believed to
be accurate, Azbil Corporation disclaims the implied warranties of
merchantability and tness for a particular purpose and makes no
express warranties except as may be stated in its written agreement
with and for its customer.
In no event shall Azbil Corporation be liable to anyone for any
indirect, special or consequential damages. is information and
specications in this document are subject to change without notice.
MagneW™ is a trademark of Azbil Corporation.
F™ Fieldbus is a trademark of FieldComm Group.
Precaution
is device is designed and manufactured for the use of general
applications.
Do not use this device for the nuclear applications and applications
which are directly related to the human life. Also do not use this
device in the radiation controlled area.
In case the device is used in the following applications, please
make sure to use the device with the system designed as failsafe/
redundant and the system should be maintained periodically.
• Safety device for human body protection
• Directly control the transport machine
• Aircra
• Space appliance
Azbil Corporation disclaims any liability for damages caused by
any of accidents/results by the above application.
SAFETY PRECAUTIONS
About Icons
e safety precautions described in this manual are indicated by various icons.
Please be sure you read and understand the icons and their meanings described below before reading the
rest of the manual.
Safety precautions are intended to ensure the safe and correct use of this product, to prevent injury to the
operator and others, and to prevent damage to property.
Be sure to observe these safety precautions.
Warnings are indicated when mishandling this product might
WARNING
CAUTION
result in death or serious injury.
Cautions are indicated when mishandling this product might result in
minor injury to the user, or only physical damage to the product.
Examples
In describing the product, this manual uses the icons and conventions listed below
.
Use caution when handling the product.
The indicated action is prohibited.
Be sure to follow the indicated instructions.
SAFETY REQUIREMENTS
To reduce risk of electric shock which could cause personal injury, follow all safety notices in this
documentation.
WARNING
This symbol warns the user of a potential shock hazard where hazardous live voltages
may be accessible.
i
Safety messages
Carefully read this section before installing or operating this device.
Cautions for installation
Installation of the device should be done by the expert from the safety perspective.
In case of the installation in the hazardous area, follow the regulation/guidance of the
explosion-proof.
In case of the installation in the hazardous area, select the explosion-protected apparatus.
Do not use the non-explosion-protected apparatus in the hazardous area.
Install the device in the environment where the following is within in the specifications
listed in the specification sheet.
• Ambient temperature, process fluid temperature
• Process fluid pressure
• Humidity
• Vibration
• Power supply voltage
Environment which exceeds the specifications may cause fire or device damage and it may
cause injury.
WARNING
CAUTION
DO NOT use the installed device as foot hold. It may cause the damage of the device.
DO NOT hit the glass of the device by tools. It may cause the damage of the device.
Install the device in a location with an ambient temperature of -25 °C to 60°C (-13 °F to
140 °F) and an ambient humidity of 5 to 100% RH to prevent device malfunction or output
errors.
DO NOT install the device near high-current power lines, motors, or transformers to prevent damage from electromagnetic induction, which can cause device malfunction or
output errors.
DO NOT use the device to ground a welder. It can damage the device.
Be sure to ground the welding power transformer when welding near the device to avoid
output errors.
DO NOT install the device in the severe vibration area or in corrosive environment. It may
damage the device, or cause the fume of the device.
DO NOT install the device on the bridge or deck of the ship.
DO NOT install the device on in the severe vibration area on the ship.
Be sure to use the metal conduit for the cables between the remote style transmitter and
flowtube.
Be sure to install the flowtube with a distance of 500mm minimum from the other flowtube. Magnetic field generated by a flowtube may affect the other flowtube and it may
cause the output errors.
DO NOT install the device in a location subject to direct sunlight, wind, rain, severe vibration, or in a highly corrosive atmosphere. The transmitter and flowtube can be damaged.
ii
Cautions for wiring
DO NOT wire with wet hands. DO NOT wire while the circuit is alive. It may cause electrical
shock.
ELECTRIC SHOCK HAZARD! Turn the power supply OFF before opening the transmitter
cover.
Wiring of the device should be done by the expert from safety perspective.
Wire the cables to the correct terminal. Incorrect wiring may damage the device or cause
the fume of the device.
Connect the correct power supply. Incorrect power supply may damage the device or
cause the fume of the device.
Be sure to properly ground the device. Improper grounding may cause the output errors
DO NOT give impact to the device. It may cause the damage of the device.
WARNING
CAUTION
Be sure to use the power supply with the overcurrent protection function.
Be sure proper wiring. Incorrect wiring may cause the damage of the device.
Be sure to properly tighten the terminal cover on the terminal box so that water does not
penetrate in the terminal.
DO NOT connect AC power supply to the DC power supply model. The wrong power supply damages the device.
Turn off the Fieldbus power supply before connecting the Fieldbus cable to the transmitter. The transmitter or the Fieldbus power supply can be damaged. This type of damage is
not covered by Azbil Corporation’s warranty.
Be sure to plug all unused conduit connections with a water tight plug.
In case that a remote model is installed in a ship, the cables between the transmitter and
flowtube must be covered with a flexible metal conduit.
Switch the control equipment to manual control before terminating the device operation and shutting off the output to the control equipment. This action prevents the power
shut-off from directly affecting the control equipment.
6.2 Input of simulated signals from calibrator ............................................6-6
6.3 Measurement of excitation current ..................................................6-7
Appendix A. View List .................................................... A-1
Appendix B. Parameter List .............................................. B-1
Appendix C. Menu Configuration ..........................................C-1
v
vi
1. Introduction
This chapter contains an overview of the model MGG14C MagneW PLUS+ Flowmeter
and F Fieldbus system. It describes available configurations and provides
definitions for all the major parts of the transmitter.
MagneW PLUS+ Flowmeter
Thank you for purchasing the Azbil Corporation model MGG14C MagneW PLUS+
Flowmeter This system features:
• Advanced multi-variable capacity
• Digital panel display
• Intuitive, versatile operator interface with large characters and backlit liquid crystal
display
• I/O Capacity
• Flexible digital communication using the F Fieldbus
1.1 System configuration
The model MGG14C MagneW PLUS+ Flowmeter consists of a flowtube and a transmitter
which operate on the principles of Faraday’s law. The flowmeter is available in two
configurations:
Integral
Remote
– The transmitter is mounted directly on the flowtube and they are installed as an
integrated unit on the fluid pipe.
– The transmitter and flowtube are installed separately and connected together
via cables
Figure1-1 Indegral system
1-1
1.2 Foundation Fieldbus
The F Fieldbus provides communications and programmability for a single or
multiple flowmeter system. Some features of the fieldbus include:
• Complies with F Fieldbus H1 (31.25 kbps voltage mode bus)
specifications.
• Supports the standard Analog Input (AI) function block.
• Is an externally powered device – AC and DC powered models available.
• Comes with Enhanced Device Description (EDD) files and a Capability File (CF) for
automatic configuration.
The following figure shows a system in which the flow rate is output with a F
Fieldbus.
Figure1-2 Remote system
1-2
Figure1-3 System using a Foundation Fieldbus
1-3
1.3 Main components
The transmitter consists of the components shown in the figure below.
Figure1-4 Main components
(1)
Local user interface
can be accessed using the four infrared sensor keys on the panel.
(2)
Fieldbus interface board
(3)
Main board
(4)
Power supply board
(5)
Terminal
(6)
Name plate
manufacture and the flowtube constant (EX) for the flowmeter.
(7)
Tag number plate
– This board includes fundamental functions of the flow measurement.
– Encloses the input/output terminals. Contains an integrated 12 kV, 1000A isolator.
– Indicates model numbers and production numbers, power supply requirements, date of
– Indicates the instantaneous flow rate or the totalized valve. The flowmeter functions
– Includes the fieldbus communication function.
– There are two types, AC power supply, and DC power supply type.
– Indicates the tag number as specified in the product order.
1-4
1.4 Local user interface
Figure1-5 Local user interface
DisplayMEASURING MODE
Main displayIndicates a value set by PARAM_SELECTION
under the block selected by BLOCK_TAG_SEL.
Indicates maximum four values cyclically.
Auxiliary display
Upper auxiliary display
Indicates the information of [Tag, unit, or status]
set by DISPLAY_INFO_SEL cyclically.
Lower auxiliary display
Indicates percent flow rate with bar indicator.
Range is set by EU_100, and EU_0.
Other than MEASURING
MODE
Indicates flow velocity.
Indicates procedures for
the parameter setting or
adjustment.
1-5
The following table is a summary of the functions of each of the keys.
1-6
How to operate the infrared touch sensor
For best results, approach the key from below and completely cover the circle. Then move
your finger straight down to its original position. These motions ensure correct operation.
Moving sideways across the keys can accidentally activate the wrong control.
Figure1-6 Using the touch senser keys
1-7
Local user interface operation
The following table describes the functions available in each mode.
Display
MEASURING MODEThis is the normal operational mode and indicates the measuring
BASIC SETUP MODEThis mode is used to change data settings that must be recorded or
MAINTENANCE MODE
ADVANCED MODEThis mode is used to apply some specific noise immunity functions.
MEASURING MODE
Other than MEASURING MODE
status.
Each time the MEASURING MODE is selected, data is written into
memory. Settings entered in other modes are held in temporary
memory for two minutes, but will return to the previously saved
value unless the MEASURING MODE is selected to save the data.
The only exception is the counter, which is always saved into memory
immediately.
changed frequently. These settings include:
TAG NO.
Damping time constant
Auto zero
Detector data
This mode is used when adjustment or verification is required for
regular maintenance of the system or when troubleshooting the
system. This mode includes:
Shipping information
Output adjustment
Gain adjustment
This mode is further divided into the following two modes:
OUTPUT CHECK MODE
CALIBRATION MODE
This mode includes:
Damping
Manual zero
Averaging
Auto spike Cut
Coefficient
Drop out
Low-flow cut off
Flow direction change
1-8
AI – 01 PV
The screen returns to MEASURING mode
MEASURING MODE
to BASIC SETUP
Move the cursor under “Y” and
press key.
Push 3
seconds
MODE
ENTER BASIC SET
YES OR
NO
in case of no operation for 8 seconds.
Move the cursor under “N” and
* BASIC SETUP
MODE
Figure1-7 How to enter BASIC SETUP
* In the BASIC SETUP mode, flow velocity in m/s appears on the upper auxiliary display.
* After entering the BASIC SETIP mode, if do not operate anything for 10 minutes, the
screen automatically returns from BASIC SETUP mode to MEASURING MODE.
1-9
*
2 seconds later
BASIC SETUP
to Fig1-11 screen
*
PD TAG 1/2
YA – RC - 001
*
PD TAG 2/2
YA – RC - 001
*
NODE ADRESS
FFh
*
MEMO
*
DAMPING
*
AUTO ZERO
READY
MODE
XXXXXXXX
003.0 s
s
*
MODE ENTER
ADVANCED
*
MODE ENTER
MAINTENANCE
*
DENSITY
*
SVR 10.000 m/s
1.0000
70.686 t/h
to Fig1-10 screen
*
MGG DIA
M
*
DUMMY
EX 300.0
50.0
0
Figure1-8 BASIC SETUP MODE
*
PVR 10.000 m/s
70.686
m3/h
1-10
2 seconds later
& ADVANCED
MODE
& COEFFICIENT
1.0000
& MANUAL ZERO
READY
& AVARAGING
OFF
& PV SPIKE CUT
OFF
& PV SPIKE CUT
TIME 00.0 s
& PV SPIKE CUT
LEVEL 00.0 %
& SV SPIKE CUT
OFF
& MODE RETURN
BASIC SETUP
& FLOW SWITCH 2
HYSTERISIS
& FLOW SWITCH 2
SP
.00000 t/h
& FLOW SWITCH 2
MODE LOW
& FLOW SWITCH 2
SOURCE SV
& FLOW SWITCH 1
HYSTERISIS 0%
& FLOW SWITCH 1
SP
.00000 m3/h
0%
& SV SPIKE CUT
TIME 00.0S
& SV SPIKE CUT
& PV L-FLOW CUT
LEVEL
& SV L-FLOW CUT
LEVEL
LEVEL 00.0 %
10%
10%
Figure1-9 ADVANCED MODE
& FLOW SWITCH 1
MODE LOW
& FLOW SWITCH 1
SOURCE
& FLOW DIRECTION
& EX FREQUENCY
PV
FORWARD
12.5 Hz
1-11
> MAINTENANCE
MODE
> CALIBRATION
> SHIPPING INFO
> MODE RETURN
BASIC SETUP
OFF
OFF
Figure1-10 MAINTENANCE MODE
1-12
2. Overview of MagneW PLUS+ Foundation Fieldbus
2.1 Overview
This chapter contains an overview of the model MGG14C MagneW PLUS+ F
Fieldbus transmitter.
Model MGG14C MagneW
MagneW PLUS+ flowtube and measures the conductive fluid flow rate. It outputs the
following with the F Fieldbus protocol.
Primary value of Flow TB: volumetric flow rate
Secondary value of Flow TB: mass flow rate
Primary value of Diag TB: scale level by scale diagnostic
Those outputs are displayed on the following Display panel.
Fieldbus is a widely used bi-directional digital communication protocol for field devices
that enable the simultaneous output to many types of data to the process control system.
PLUS+ F Fieldbus transmitter is coupled with the
MODE
The MagneW
by the FieldComm Group, and provides interoperability between Azbil devices and those
produced by other manufacturers.
Fieldbus comes with software consisting of three AI function blocks that enable the
flexible implementation of systems.
PLUS+ Fieldbus communication type employs the specification standardized
2.2 Structure of MagneW PLUS+ Foundation Fieldbus
The MagneW PLUS+ contains two Virtual Field Devices (VFD), one is System/network
Management VFD and the other is Function block VFD.
2.2.1 System/network Management VFD
• Sets node addresses and Physical Device tags (PD Tag) necessary for communication.
• Controls the execution of function blocks.
• Manages operation parameters and communication resources (Virtual Communication
Relationship:VCR).
2-1
2.2.2 Function Block VFD
It includes some objects to execute function block applications. Resource block, Function
block, and Transducer block are one of the objects.
Transducer block
2.2.3 Function block
AI *2 (3)30
DI2301800, 1900For flow limit switches
PID1452000
AR1302100
TOT2302200, 2300For totalization volume
Resource block
Function block
Alert object
Trend object
View object
Link object
Block
name
Number
• Manages the status of The
• Automatically informs the host of any detected faults or other problems
based on the NAMUR NE107.
It links together with Transducer block or other function block and
provides arithmetic processing and transmits.
It provides interface function between the hardware and function block.
It provides processing for the event or alarm generated in function block.
Object to transmit trend data collected in short time
Object that provides information of each parameter in the function block.
Objects that links the above mentioned objects.
Execution
time (ms)
INDEXNote
1500, 1600,
(1700)
MagneW PLUS+
For volumetric flow rate and mass flow rate
PID function block execute a control
algorithm to minimize the error as the
difference between a measured process
variable and desired setpoint.
It also has functions of cascade control, feed
forward control, and alarm detection.
Arithmetic block perform an arithmetical
operation to the flow measurement value.
hardware.
* Number of AI becomes three when selecting flow signal analysis function.
Flow signal analysis function will be provided as call factory basis. Consult your azbil
representative.
2.2.4 Transducer Block
Flow transducer block
Display transducer blockBlock that controls the LCD display
Diag transducer blockBlock that executes the diagnosis
Block nameDescription
Block that calculates the flow speed, volumetric flow rate,
and mass flow rate
2-2
2.3 VFD/Object correlation diagram
Device
Basic information control
Function block
Control function
LAS
Communication stack
(FMS/FAS/DLL)
Physical layer (PHL)
H1 Fieldbus Line
/
VCR control
VCRVCRVCRVCRVCR
Function block VFD
Resource
block
AI
Function
Link object
FF standard transducer block
(FLOW/Positioner/Pressure)
View
Object
2-3
Figure2-1 VFD/Object correlation diagram
DI
Function
Senso r
Trend
Object
Display
Display transducer block
PID
Function
Alert
Object
2-3
2.4 Wiring
The number of devices that can be connected to a single bus and the cable length vary
depending on system design. When constructing systems, both the basic and overall
design must be carefully considered to achieve optimal performance.
Fig 2-2 shows typical wiring to the Fieldbus network. The following are required for use
with Fieldbus devices:
• Terminator:
Fieldbus requires two terminators. Refer to the supplier for details of terminators that are
attached to the host.
• Field devices:
Connect Fieldbus communication type MagneW
PLUS+. Two or more MagneW PLUS+
devices or other devices can be connected.
• Host:
Used for accessing field devices. A dedicated host (such as DCS) is used for an
instrumentation line while dedicated communication tools are used for experimental
purposes. For operation of the host, refer to the instruction manual for each host. No
other details on the host are given in this manual.
• Cable:
Used for connecting devices. For laboratory or other experimental use, a twisted pair cable
2
two to three meters in length with a cross section of 0.9 mm
or more and a cycle period
of within 5 cm (2 inches) may be used. Termination processing depends on the type of
device being deployed. For MagneW PLUS+, use an M4 screw terminal claw. Some hosts
require a connector. Refer to Azbil Corporation when making arrangements to purchase
the recommended equipment. Connect the devices as shown in Figure 2-2. Connect
the terminators at both ends of the trunk, with a minimum length of the spur laid for
connection.
Table 2-1 Fieldbus cable
ClassificationDescriptionSizeMaximum length
2
AType A Fieldbus cable0.8 mm
BType B Fieldbus cable0.32 mm
CType C Fieldbus cable0.13 mm
DType D Fieldbus cable1.25 mm
(18AWG)1900m
2
(22AWG)1200m
2
(26AWG)400m
2
(16AWG)200m
Table 2-2 Resistance of the Fieldbus cable
Classification Maximum resistance @ 25ºC
A22 /Km
B56 /Km
C132 /Km
D20 /Km
2-4
2.5 Others
Figure2-2 Cabling
Connect/disconnect the fieldbus devices
Before connecting/disconnecting the fieldbus device to the network, make sure this
activity does not affect the process.
Fieldbus network allows device connection/disconnection to the network while the device
circuit is alive. However from the process safety perspective, the device connection/
disconnection to the network after turn off the power is suggested.
Make sure to properly isolate the wires after disconnecting the device.
Non-fieldbus device
Non-fieldbus device is not allowed to connect to the fieldbus network.
Communication error and miss-wiring
If communication error appears, the device repeats sending the signal/message until it
is received by the host system. This kind of the communication error appears when the
wiring is improper.
Check the wiring if communication error appears.
2-5
2-6
3. Installation
Overview
Installation of the device should be done by the expert from the safety perspective.
In case of the installation in the hazardous area, follow the regulation/guidance of the
explosion-proof.
In case of the installation in the hazardous area, select the explosion-protected apparatus.
Do not use the non-explosion-protected apparatus in the hazardous area.
Install the device in the environment where the following is within in the specifications
listed in the specification sheet.
• Ambient temperature, process fluid temperature
• Process fluid pressure
• Humidity
• Vibration
• Power supply voltage
Environment which exceeds the specifications may cause fire or device damage and it may
cause injury.
This chapter describes the installation and wiring of the device in the following order:
- Criteria for selecting the installation environment
- Overview of the device installation
- Wiring of signal cables
WARNING
CAUTION
DO NOT use the installed device as foot hold. It may cause the damage of the device.
DO NOT hit the glass of the device by tools. It may cause the damage of the device.
Install the device in a location with an ambient temperature of -25 °C to 60°C (-13 °F to
140 °F) and an ambient humidity of 5 to 100% RH to prevent device malfunction or output
errors.
DO NOT install the device near high-current power lines, motors, or transformers to prevent damage from electromagnetic induction, which can cause device malfunction or
output errors.
DO NOT use the device to ground a welder. It can damage the device.
Be sure to ground the welding power transformer when welding near the device to avoid
output errors.
DO NOT install the device in the severe vibration area or in corrosive environment. It may
damage the device, or cause the fume of the device.
DO NOT install the device on the bridge or deck of the ship.
DO NOT install the device on in the severe vibration area on the ship.
Be sure to use the metal conduit for the cables between the remote style transmitter and
flowtube.
Be sure to install the flowtube with a distance of 500mm minimum from the other flowtube. Magnetic field generated by a flowtube may affect the other flowtube and it may
cause the output errors.
DO NOT install the device in a location subject to direct sunlight, wind, rain, severe vibration, or in a highly corrosive atmosphere. The transmitter and flowtube can be damaged.
3-1
3.1 Before installing electromagnetic flowmeter
Criteria for selecting the installation environment Getting started Select the optimal
location for installing the device according to the following criteria in order to maximize
its performance.
Note:
- Install the device in a location with an ambient temperature of -25 °C to 60 °C (-13 °F
to 140 °F) and an ambient humidity of 5 to 100% RH to prevent device malfunction or
output errors.
- DO NOT install the device near high-current power lines, motors, or transformers to
prevent damage from electromagnetic induction, which can cause device malfunctions
or output errors.
- DO NOT use the device to ground a welder. Doing so can damage the device.
- Be sure to ground the welding power transformer when welding near the device.
- DO NOT install the device in an area with severe vibrations or in a corrosive
environment. Doing so may cause damage the device.
- DO NOT install the device in a location subject to direct sunlight, wind, or rain.
WARNING
Use this product under the conditions as defined in the specifications (e.g., protection against explosion, pressure rating, temperature,
humidity, voltage, vibration, shock, mounting direction, and atmosphere) to avoid failure of the device or harmful physical effects such
as fire damage.
WARNING
In case of the installation in the hazardous area, select the explosionprotected apparatus. Do not use the non-explosion-protected apparatus in the hazardous area.
3-2
3.2 Installation of transmitter
Mounting bracket
Material: SPCC
(Zinc plating)
2B-pipe
2B-pipe mounting
Installation of transmitter
Basic mounting methods
The transmitter can be mounted in three ways—integrated mounting with the
flowtube, wall-mounting, and 2B-pipe mounting
Mounting plate
Terminal box
Tag number plate
Figure3-1 Wall-mounting of transmitter
Figure3-2 2B-pipe mounting of transmitter and Integrated mounting
3-3
Electrical wiring (1)
DO NOT wire with wet hands. DO NOT wire while the circuit is alive. It may cause electrical
shock.
ELECTRIC SHOCK HAZARD! Turn the power supply OFF before opening the transmitter
cover.
Wiring of the device should be done by the expert from safety perspective.
Wire the cables to the correct terminal. Incorrect wiring may damage the device or cause
the fume of the device.
Connect the correct power supply. Incorrect power supply may damage the device or
cause the fume of the device.
Be sure to properly ground the device. Improper grounding may cause the output errors
DO NOT give impact to the device. It may cause the damage of the device.
WARNING
CAUTION
Be sure to use the power supply with the overcurrent protection function.
Be sure proper wiring. Incorrect wiring may cause the damage of the device.
Be sure to properly tighten the terminal cover on the terminal box so that water does not
penetrate in the terminal.
DO NOT connect AC power supply to the DC power supply model. The wrong power supply damages the device.
Turn off the Fieldbus power supply before connecting the Fieldbus cable to the transmitter. The transmitter or the Fieldbus power supply can be damaged. This type of damage is
not covered by Azbil Corporation’s warranty.
Be sure to plug all unused conduit connections with a water tight plug.
In case that a remote model is installed in a ship, the cables between the transmitter and
flowtube must be covered with a flexible metal conduit.
Switch the control equipment to manual control before terminating the device operation and shutting off the output to the control equipment. This action prevents the power
shut-off from directly affecting the control equipment.
Introduction
The electromagnetic flowmeter needs to be connected to the main power supply (AC or
DC 24 V) for proper operation.
The following items are described in relation to the electrical wiring of the electromagnetic
flowmeter.
3-4
- Connection points on the main unit of the electromagnetic flowmeter
- Terminal layout
- Transmitter terminal table
- Cables between flowtube and transmitter
- Cable specifications
- Connection between flowtube and transmitter
- Selection of wiring cable
- Installation of wiring cable
- Connection to the Fieldbus output
Note:
- Do not directly connect the AC power supply to the main unit of the electromagnetic
flowmeter if the main power supply is designed to be 24 V DC. AC power supplied
to the main unit of the electromagnetic flowmeter causes irreversible damage in the
internal measurement circuit.
Connection points on the main unit of the electromagnetic flowmeter
Figure 3-4 presents the terminal block of the main unit of the electromagnetic flowmeter.
WARNING
· In wiring, turn off the power before opening the cover to avoid electric shock.
· DO NOT perform wiring work when current is being applied. Doing so may
cause electric shock.
Note:
- Properly align the wiring position as indicated. Improper wiring can damage the
device.
- Especially, check the wiring position of the power lines again as they carry a high
capacity current.
3-5
Terminal layout
Figure3-3 Terminal layout of remote transmitter
Integral transmitter: Terminal layout
Unlike a remote transmitter, an integral transmitter does not use the terminals X, Y, SB,
SA, A, B, C, and E. These terminal symbols are thus erased from the layout.
DC-24V transmitter: Terminal wiring diagram
The power terminal of the DC-24V remote transmitter is indicated as POWER DC
24V. Check carefully if the polarity is positive or negative.
Transmitter terminal table
Terminals for a remote transmitter Terminals for an integral transmitter
SymbolDescriptionSymbolDescription
A
B-
C
SAN
SB
FIELDBUS
X
Y
POWER AC
ENot used
Flow signal input
+
Fieldbus output
-
Excitation output
L
N
Power supply
Grounding
(grounding
resistance must be
FIELDBUS
POWER AC
<100)
+
Fieldbus output
L
Power supply
Grounding
(grounding resistance
must be <100)
3-6
Electrical wiring (2)
Cable length (m)
Cable length (m)
Cables between flowtube and transmitter
Please use a special cable (MGA12W) designated for connecting the flowtube and
transmitter.
Select a signal cable (Azbil’s special cable or a commercially-available shielded cable)
depending on the fluid conductance, cable length, and the diameter of the flowtube. Refer
to the figure below.
- Range where Azbil’s special cable can be used: A and B
- Range where a commercially-available cable can be used: Only A
2.5mm and 5mm Diameter Cable>10mm Diameter Cable
B. Usage range of dedicated
cable, model MGA12W
A. Usage range of
commercially
available cable
Fluid Conductivity (µs/cm)
B. Usage range
of dedicated
cable, model
MGA12W
Fluid Conductivity (µs/cm)
Figure3-4 Relationship between the fluid conductance and cable length
Specifications of cables between flowtube and transmitter
Cable (between remote flowtube and transmitter):
Length: Maximum 300 m (depends on the fluid conductance)
Outer diameter: φ 10–12 mm
2
Signal cable: Special cable (diameter of 11.4 mm, 0.75 mm
Note) Remove the conductive tubes (black) of conductive cables for terminals A and B up
to the edge s of the inner shields.
Figure3-5 Structural drawing of signal cable
3-8
Excitation cable
Designed for an M4 screw
(Model: MGA12W)
Designed for an M4 screw
Insulator
60
Flowtube
side
Transmitter
side
70
Designed for an M4 screw
70L60
Designed for an M4 screw
Designed for an M4 screw
Designed for an M4 screw
Cable termination on the owtube side
Black
White
Flowtube
side
50
50
(Model code: A)
Black
White
Flowtube
side
50
50
(Model code: A)
Black
White
Black
White
Flowtube
side
Flowtube
side
60
39
(Model code: B)
60
39
(Model code: B)
Sheath
Insulator
Conductive wire
Fabric tape
Figure3-6 Structural drawing of excitation cable
3-9
Electrical wiring (3)
Dedicated cable model MGA12W
Commercially available cable
Connection between flowtube and transmitter
Detector
A
Signal cable
model MGA12W
Signal cable 2-core
Single shielded cable
Converter
SA
A
C
B
X
Y
Excitation cable
model MGA12W
Signal cable 2-core
Single shielded cable
Detector
A
C
B
X
Y
Excitation cable
Figure3-7 Connection between flowtube and transmitter
C
B
SB
X
Y
Converter
A
C
B
X
Y
3-10
4. Startup and Shutoff
Overview
This chapter describes the procedure for starting up the device and performing zero
adjustment.
Follow the instructions in this chapter when you start up and operate the device for the
first time.
Zero adjustment can be made in one of the following ways:
Using the device’s LUI
Using a fieldbus configurator to communicate with the device
This chapter describes how to perform zero adjustment using the local user interface (LUI).
Refer to Chapter 5 for the alternative method using Fieldbus communication.
4.1 Startup
Start up the device
Procedure
Start up the electromagnetic flowmeter according to the following procedure.
StepProcedure
Make sure that the flowtube of the electromagnetic flowmeter is properly
1
connected to the piping.
Make sure that the flowtube and transmitter of the electromagnetic flowmeter are
2
securely wired.
Fill the flowtube of the electromagnetic flowmeter with a measurement fluid and
3
pause the flowtube.
Check for any fluid leakage from the flange, to which the flowtube of the
4
electromagnetic flowmeter is attached.
5Apply current to the transmitter of the electromagnetic flowmeter.
Make sure that the LUI display is turned on.
Image
6
This completes the startup of the electromagnetic flowmeter.
. . . . .
4-1
4.2 Steps before measurement
m3/h
m3/h
m3/h
m3/h
m3/h
Setting up flowtube data
Getting started
Select and configure the constant of the flowtube used in combination with the
transmitter, flowtube type, and diameter.
Default settings
The settings of EX 300.0, MGG, and DIA 050.0 are applied when no flowmeter is specified
for the combined use with the transmitter.
Note:
If you have purchased a combination set including a transmitter and a flowtube, the
flowtube data are already configured by real current calibration. Note that any change
will result in an output error from the flowmeter. Refer to Table 4-1.
Improper setting of flowtube information can result in
output errors from the flowmeter.
StepProcedureScreen
WARNING
Follow the procedure for entering BASIC SETUP MODE
1
to display the screen for setting up the flowtube data.
Touch the key to set up the constant for the flowtube.
Use the and keys to enter the value for the EX section
2
which is printed on the name plate of the flowtube to be
used in combination with the transmitter.
Next, touch the key to select the flowtube type.#Use
the and keys to select the MODEL number which is
3
printed on the name plate of the flowtube to be used in
combination with the transmitter.
Next, touch the key to select the diameter.#Use the
4
and keys to select the diameter of the flowtube used in
combination with the transmitter.
Touch the key to move the cursor underneath the *
5
mark.
4-2
Zero adjustment
m3/h
m3/h
m3/h
m3/h
m3/h
m3/h
Getting started
Make sure to perform zero adjustment after starting up the device. There are three ways to
do so.
Using the device’s LUI
Using the parameter list by communicating with the device
Using the menu by communicating with the device
Adjust the measured instantaneous flow rate value to zero when the fluid inside the
flowtube is sitting still.
Note:
Zero adjustment is extremely important for ensuring accurate measurement of flow
rate. Make sure to perform this adjustment when you first operate the device.
Before performing zero adjustment, make sure that Class D grounding is reliably
performed with the flowtube and that the flowtube is filled with a measurement fluid
which is sitting still.
Zero adjustment is possible when the flow velocity falls below 0.2 m/s. Wait until the
fluid comes to a complete rest (a flow velocity of 0.0 m/s) to avoid output errors.
Zero adjustment requires O/S to be selected for the Actual MODE in Flow Transducer
Block.
Configure the setting from the host system in advance.
StepProcedureScreen
Touch the MODE key for 3 seconds or longer.
Note: The screen as shown on the right is displayed only
1
for 8 seconds. Perform the following operations within 8
seconds. Touch the
Touch the
key to move the cursor underneath YES and
key four times.
touch the ke y.
2
is displayed before the screen
switches to the one shown on the
right in 2 seconds.
Touch the
3
the right.
Touch the key once to move the cursor underneath
4
READY.
key once to display the screen as shown on
Touch the key to start zero adjustment. During the
adjustment, the value of 0.0 blinks for displaying the large
7-segment flow rate. Once the adjustment is complete,
5
the value stops blinking and the screen returns from ON
to READY.
This period is about 30 seconds.
At the very end, make sure to touch the MODE key to
6
complete the adjustment.
4-3
Checking the operation of Fieldbus
Check to see if the device operates properly with Fieldbus.
Prior to operating Fieldbus, the DD file (device description) and the CF file (capability) of
the device need to be copied to the host.
Download the DD file and CF file from the official website of the FieldComm Group.
Operation of Fieldbus requires the following settings in the host.
Ensure that PD_TAG and NODE_ADRS can be set up by configuring the following
settings.
1) Network parameter settings for link active scheduler (LAS)
SymbolParameter nameDescription and setting value
V(ST)Slot timeSet a value of 5 or greater.
V(MID)Gap between minimum framesSet a value of 10 or greater.
V(MRD)Maximum response delay time
V(FUN)
V(NUN)Total number of unpolled nodes
The number of the first unpolled
node
Set the value so that V(MRD)
V(ST) is 20 or greater.
Set the value immediately next to
the address used by the host. Set
a value of 12 or greater in 16-bit
representation.
The number represents the range
of unused addresses. Set the value
of the minimum address used by
the device in the work site minus
the value of V(FUN).
1) Checking PD_TAG (physical device tag) and address
Parameter nameParameter nameSetting value
PD_TAGPhysical device tag
NODE_ADRSNode address
The same address setting as another device cannot be selected for NODE_ADRS. (Such an
entry is changed into the default address [0xF8–0xFB])
Assign a different address for each unit.
Up to 32 ASCII
characters
Set the minimum
address of the BASIC
device in hexadecimal
notation and F7 or less
Data when shipped
from the factory
Specified by customers
Hexadecimal notation
and F8
4-4
4.3 Shutoff
CAUTION
· Make sure to switch the controller to manual control mode in order to
shut off the device and stop the output to the controller.
This safeguards the controller from direct impacts caused by the suspended output from the device.
· Do not turn off the power within 60 seconds of any data being rewritten using the LUI or another communication method, to avoid possible errors in saving the data.
Procedure
Take the following procedure to shut off the device.
StepProcedure
1Switch the controller of the device you want to shut off to manual control mode.
2Turn off the power of the device.
4-5
4-6
5. Basic Settings by Fieldbus Communication
Overview
This chapter outlines zero adjustment as an operation performed using Fieldbus
communication, each Transducer block, as well as basic parameter settings.
Some parameters require O/S to be selected for the Block MODE parameter in order to
change the settings.
Mode parameters can be switched with MODE_BLK in each block.
In order to use each block after the settings are changed, select AUTO for MODE-BLK.
Note
- The device supports Version 3.8 or later versions of 475.
Version 3.6 or earlier versions cannot properly display some parameters.
- The device supports host authentication, HTK 6.1.a or later versions, and Standard
Dictionary 3.70 or later versions.
Other versions may fail to display some parameters.
- Some parameters may not be properly displayed when some hosts are used.
- METHOD may not work with some hosts.
In such cases, access the parameter from the parameter list to configure the settings.
5-1
5.1 Fieldbus communication menu
There are four types of menu structures for Fieldbus communication depending on the
host being used.
This section describes the device menu for communicators.
- Device menu for communicators
This menu displays parameters such as setup and adjustment of the positioner.
A host supporting the device menu can display this menu. (Example: 475 Field
Communicator by Emerson)
- Block menu for communicators
A communicator capable of Fieldbus communication can display this block menu. A
menu prepared for each block displays parameters such as setup and adjustment of the
positioner. (Example: 475 Field Communicator by Emerson)
- Block menu for PCs
A compatible host (PC) can display this block menu. A menu prepared for each block
displays parameters such as setup and adjustment of the positioner. (Example: Azbil's
device management system - InnovativeField Organizer)
- Parameter list
All parameters are displayed by block.
Appendix C presents a list of parameters for Flow Transducer Block, Resource Block,
and Display Transducer Block.
5-2
5.2 Zero adjustment
Getting started
Once the device has started up, make sure to perform zero adjustment in one of the
following three ways.
Adjust the measured instantaneous flow rate value to zero when the fluid inside the
flowtube is sitting still.
Note:
- Zero adjustment is extremely important for ensuring accurate measurement of flow
rate. Make sure to perform this adjustment when you first operate the device.
- Before performing zero adjustment, make sure that Class D grounding is reliably
performed with the flowtube and that the flowtube is filled with a measurement fluid
which is sitting still. Zero adjustment is possible when the flow velocity falls below 0.2
m/s. Make sure to wait until the fluid comes to a complete rest (a flow velocity of 0.0
m/s) to avoid output errors.
- Zero adjustment requires O/S to be selected for the Actual MODE in Flow Transducer
Block.
Configure the setting from the host system in advance.
Accessing the option from the menu
Note: Use METHOD in the menu to perform zero adjustment. METHOD may not
operate with some hosts.
In such cases, perform zero adjustment by accessing the option through the
parameter list.
- Device menu for communicators
Perform zero adjustment by selecting Device Maintenance Auto Zero Adjustment
(METHOD).
- Block menu for communicators or Block menu for PCs
Perform zero adjustment by selecting FLOW Transducer Block Block Maintenance
Auto Zero Adjustment (METHOD).
Accessing the option from the parameter list
Set TARGET in MODE_BLK of FLOW Transducer Block to O/S (Out of Service).
1
Set “1” (Execute) to AUTO_ZERO_CALIBRATION_CMD in FLOW Transducer
Block.
Zero adjustment starts.
2
Set “4” (Canceled) to AUTO_ZERO_CALIBRATION_CMD in FLOW Transducer
Block in order to cancel zero adjustment.
Execution results of automatic zero adjustment can be checked at
MANUAL_ZERO_CALIBRATION_CMD in FLOW Transducer Block.
3
While “1” (Executing) is displayed during execution of zero adjustment, “2” (Success) is
displayed when zero adjustment is completed successfully.
For flow measurement, set TARGET of Transducer Block in FLOW Transducer Block to
4
AUTO.
5-3
5.3 Basic settings
Sensor value
Flow Transducer Block
Overview of function
Flow Transducer Block enables calculation of flow velocity, volume flow rate, and mass
flow rate based on the sensor outputs of the electromagnetic flowmeter.
The calculated volume flow rate and mass flow rate are output to AI Function Block
and TOT Function Block. A flow switch is prepared for turning the transducer on or off
depending on the volume flow rate and mass flow rate. The value assigned to the flow
switch is output to DI Function Block.
Function Block diagram
The following diagram shows the configuration of Flow Transducer Block.
Calculation function
Self-diagnosis history
Adjustment function
SENSOR_VALUE (ow velocity)
Conguration function
PRIMARY_VALUE
(volume ow rate)
SECONDARY_VALUE
(mass ow rate)
FLSW1_VALUE
FLSW2_VALUE
Figure5-1 Function Block diagram
Parameters
Major parameters for Flow Transducer Block are described as follows.
Refer to the attached Parameter List for more descriptions of parameters.
Output
SENSOR_VALUE
Indicates the flow velocity (m/s) of the measurement fluid.
PRIMARY_VALUE
Indicates the volume flow rate and the status.
Channel 1 corresponds to this parameter.
Select 1 for CHANNEL of Function Block you want to use in order to connect to AI
Function Block or TOT Function Block.
SECONDARY_VALUE
Indicates the mass flow rate and the status.
Channel 2 corresponds to this parameter.
Select 2 for CHANNEL of Function Block you want to use in order to connect to AI
Function Block or TOT Function Block.
FLSW_1_VALUE_D
Indicates the output value of Flow Switch 1 and the status.
Channel 100 corresponds to this parameter.
Select 100 for CHANNEL of Function Block you want to use in order to connect to DI
Function Block.
FLSW_2_VALUE_D
Indicates the output value of Flow Switch 2 and the status.
Channel 101 corresponds to this parameter.
Select 101 for CHANNEL of Function Block you want to use in order to connect to DI
Function Block.
Settings
PRIMARY_VALUE_RANGE
5-4
Indicates setting values for the upper and lower range limits, unit, and the position of
the decimal point for the volume flow rate.
The unit for the flow rate must be identical to the unit assigned for XD_SCALE in AI
Block.
Refer to Table 5-1 for the units that can be set up.
SECONDARY_VALUE_RANGE
Indicates setting values for the upper and lower range limits, unit, and the position of
the decimal point for the mass flow rate.
The unit for the flow rate must be identical to the unit assigned for XD_SCALE in AI
Block.
Refer to Table 5-2 for the units that can be set up.
DENSITY_CONSTANT
Fluid density setup
The assigned value is applied to the calculation of SECONDARY_VALUE (mass flow
rate).
FLOWTUBE_SIZE
Flowtube diameter setup
FLOWTUBE_TYPE
Flowtube type setup
EMPTY_PIPE_DETECTOR
Indicates whether the empty detection function is enabled or disabled.
Calculation function
DAMPING_CONSTANT
A damping constant can be set up when a tiny variation component needs to be cut off
from the measured instantaneous flow rate value.
PRIMARY_VALUE_LOW_FLOW_CUT
A low cut-off threshold can be set for the volume flow rate.
Available range is 0 to 10% (percentage of PRIMARY_VALUE_RANGE.EU100).
SECONDARY_VALUE_LOW_FLOW_CUT
A low cut-off threshold can be set for the mass flow rate.
Available range is 0 to 10% (percentage of SECONDARY _VALUE_RANGE.EU100).
Adjustment function
AUTO_ZERO_CALIBRATION_CMD
This command executes automatic zero adjustment.
Run this command in order to perform zero adjustment when the fluid is filled
and sitting still in the flowtube.
GAIN_CALIBRATION_CMD
This command executes gain adjustment of the transmitter.
Run this command in order to perform zero adjustment of the transmitter using a
calibrator, etc.
Self-diagnosis history
FLOW_TB_STATUS_RECORD_1–10
Saves self-diagnosis history.
5-5
Table 5-1. Units that can be set up for volume flow rate
Table 5-2. Units that can be set up for mass flow rate
Unit of
volume
flow rate
g
g/d (1321)
kg/d (1325)
dhmins
g/h (1320)
kg/h (1324)
tt/d (1329)t/h (1328)t/min (1327)t/s (1326)
lblb/d (1333)lb/h (1332)lb/min (1331)lb/s (1330)
Unit of time
g/min (1319)
kg/min (1323)
g/s (1318)
kg/s (1322)
5-6
Display Transducer Block
Overview of function
Display Transducer Block controls the display of measurement values and alarm
indication in the local user interface (LUI).
Input and output signal values for function blocks installed in the device can be displayed,
such as OUT of AI Block and PRIMARY_VALUE and SECONDARY_VALUE of Flow
Transducer Block.
Parameter
Major parameters of Display Transducer Block are described as follows.
Refer to the attached Parameter List for more descriptions of parameters.
DISPLAY_PARAM_SELECTION
Select the parameter you want to display from among the parameters configured in the
display setting 1 to 4.
Up to four parameters can be displayed in chronological order.
If all of the parameters for 1 to 4 are unselected, O/S is set for MODE_BLK._Actual in
Display Transducer Block.
DISPLAY_INFO_SELECTION
Adjunct data (tag, status, and unit) that are commonly assigned for parameters in the
display setting 1 to 4 can be configured. Factory default is 0x07 (Tag, Unit, Status
are all selected)
For example, if not displaying Tag and Status, and displaying only Unit, change the
setting to 0x02.
If all the adjunct data are unselected, O/S is set for MODE_BLK._Actual in Display
Transducer Block.
DISPLAY_CYCLE
The cycle for updating the display can be set in units of seconds, in a range from 1 to 10
seconds. Factory default is 0x05(5 sec).
BLOCK_TAG_SELECTION_n
Enter BLOCK_TAG for the block to which the parameter is to be displayed in the
display setting n. Enter the BLOCK_TAG of the block that has the parameters to be
displayed into BLOCK_TAG_SELECTION_n. If an incorrect value is entered, the block
cannot be identified and display settings cannot be set up.
PARAM_SELECTION_n
Select a parameter you want to display in the display setting n.
Available values are presented in Table 5-3.
*1
*2
*2
Note: When writing is not possible for BLOCK_TAG_SELECTION_n (n=1 to 4), or
PARAMETER_SELECTION_n (n=1 to 4), switch the settings of RB FEATURE_SEL
Bit_12 (Deferral of Inter-Parameter Write Checks) to ON (enabled).
DISPLAY_TAG_n
*2
Enter the text you want to display in the indicator as a tag for the display setting n.
Up to 16 characters can be entered.
5-7
UNIT_SELECTION_n
*2
You can select whether the parameter in display setting n will be displayed using the
unit associated with it or another desired unit.
Select Auto (0) in order to display the parameter with the associated unit as assigned
for the displayed value, and select Custom (1) in order to display the parameter with
another desired unit.
Please refer to the parameter list, which includes the display parameter, and section in
Table 5-4, “LUI Display Unit String.”
CUSTOM_UNIT_n
*2
Enter the text you want to display as the unit.
This parameter is displayed only when Custom is selected in UNIT_SELECTION_n.
Up to 16 characters can be entered for CUSTOM_UNIT_n.
*1 If changing the display time, change the parameter setting.
*2 If changing the display detail from the factory default PV value of the Flow Transducer Block to
other parameter, change these parameters.
Regular Display
With the factory default setting, the PV value of Flow Transducer Block is displayed.
With the factory default setting, it gets displayed periodically according to the following
sequence.
Seq. No.123
Main Disp.
Aux. Disp
DisplayTime
FTB:PV valueFTB:PV valueFTB:PV value
PRIMARY_VALUE(Specified Unit) (Status)
555
For example, if you wish to change the display detail to Out value of AI Function Block,
1. Set AI_FB_01 in BLOCK_TAG_SELECTION_1. (AI_FB_01 of Block Tag is
the factory default value. If it has been changed since factory shipment, set this
modified value)
2. Verify that BLOCK_TYPE_SELECTION_1 is now AI Function Block(0x0101).
3. Select 8:OUT for PARAM_SELECTION_1
4. For example, change DISPLAY_TAG_1, which is the tag to be displayed, to OUT.
For other displayable parameters, please refer to Table 5-3.
Note: If you change the Block Tag of the block that you are trying to set the display for, the
setting will not get properly implemented.
If you had changed the Block Tag, turn the device power off and then on again, or
write “4{Restart Processor}” to Resource Block’s RESTART section and then restart
your device.
5-8
Several Parameter Display
Display Transducer Block can display a maximum of 4 parameters sequentially and
periodically. This section will explain the setting of displaying two parameters as an
example.
This explains how to set it to display the PV value of Flow Transducer Block and OUT
value of AI Function Block periodically.
- The factory default setting sets it so that it only displays the PV value of Flow
Transducer Block. In addition, if adding OUT value of AI Function Block as the second
parameter, enable bit1:Parameter 2 of the DISPLAY_PARAM_SELECTION.
DISPLAY_PARAM_SELECTION will change from the factory default 0x01 to 0x03.
- Next you will set BLOCK_TAG_SELECTION_2 to AI_FB_01. Verify that BLOCK_
TYPE_SELECTION_2 is now set to 0x0101 AI Function Block (AI_FB_01 of Block
Tag is the factory default value. If it has been changed since factory shipment, set this
modified value).
- Select 8:OUT for PARAM_SELECTION_2.
- Set the DISPLAY_TAG_2, a display tag, to OUT for example.
By conducting the above setting, it gets displayed periodically according to the following
sequence.
Seq. No.123456
Main Disp.
Aux. Disp
DisplayTime
FTB:PV
value
PRIMARY_
VALUE
555555
FTB:PV
value
(Specified
Unit)
FTB:PV
value
(Status)OUT
AI:OUT
value
AI:OUT
value
(Specified
Unit)
AI:OUT
value
(Status)
Set the following to display the third parameter along with the above.
DISPLAY_PARAM_SELECTION: 0x07 (enable bit2:Parameter 3 as well)
BLOCK_TAG_SELECTION_3: set the Block Tag to be displayed
PARAM_SELECTION_3: select the parameter to be displayed
DISPLAY_TAG_3: set the Tag to be displayed
Set the following to display the fourth parameter as well.
DISPLAY_PARAM_SELECTION: 0x0f (enable bit3:Parameter 4 as well)
BLOCK_TAG_SELECTION_4: set the Block Tag to be displayed
PARAM_SELECTION_4: select the parameter to be displayed
DISPLAY_TAG_4: set the Tag to be displayed
Please refer to Appendix B for detail on Display Transducer Block parameters.
Note: If you change the Block Tag of the block that you are trying to set the display for, the
setting will not get properly implemented.
If you had changed the Block Tag, turn the device power off and then on again, or
write “4{Restart Processor}” to Resource Block’s RESTART section and then restart
your device.
5-9
Status Display
Please refer to Table 5-5 for the status displayed by the string part.
Alarm Display
When an alarm is activated, the alarm display strings of Table 5-6 will get periodically
displayed.
Irregular Display
During OOS and main board communication error, the regular display will be switched to
irregular display.
OOS Display
When Display Transducer Block is OOS (Out Of Service), it will display the following.
Main Disp.
Aux. Disp
(upper)
(lower)
DISPLAY OFF
DISP TB O/S
Main Board Communication Error Display
When a main board communication error occurs inside the transmitter, the following
gets displayed.
Main Disp.
Aux. Disp
(upper)
(lower)
HW ERROR
COMM ERROR
Table 5-3 Display Transducer Block Displayable Parameter List
BlockProfileNumberParameterIndexRangeIndex
FlowTB0x0113PRIMARY_VALUE15PRIMARY_VALUE_RANGE 16
SECONDARY_VALUE18
DiagTB0x8018SCALE_LEVEL15SCALE_LEVEL_RANGE16
AI0x0101OUT8OUT_SCALE11
PID0x0108OUT9OUT_SCALE11
IN15PV_SCALE10
CAS_IN18PV_SCALE10
BKCAL_IN27OUT_SCALE11
BKCAL_OUT31PV_SCALE10
RCAS_IN32PV_SCALE10
ROUT_IN33OUT_SCALE11
RCAS_OUT35PV_SCALE10
ROUT_OUT36OUT_SCALE11
TRK_VAL39TRK_SCLAE37
FF_VAL40FF_SCALE41
AR0x0127OUT8OUT_RANGE11
IN14PV_SCALSE10
IN_LO15PV_SCALSE10
IN116PV_SCALSE10
IN217PV_SCALSE10
IN318PV_SCALSE10
TOTALIZER0x0144OUT9OUT_RANGE12
PRESET_IN20XD_RANGE11
SECONDARY_VALUE_
RANGE
19
5-10
Table 5-4 LUI Display Unit String List
Unit CodeUnit String Displayed in LUIDescription
1000KKelvin
1001degCdegree Celsius
1002degFdegree Fahrenheit
1003degRdegree Rankine
1034m3cubic meter
1036cm3cubic centimeter
1038Lliter
1048galUS gallon
1049ImpGalImperial gallon
1051bblbarrel
1088kgkilogram
1089ggram
1092tmetric ton
1094lbpound (mass)
1130Papascal
1131GPagigapascal
1132Mpamegapascal
1133kPakilopascal
1134mPamillipascal
1135uPamicropascal
1136hPahectopascal
1137barbar
1138mbarmillibar
1139torrtorr
1140atmatmospheres
1141psipounds per square inch
1142psiapounds per square inch absolute
1143psigpounds per square inch gauge
1144gcm2gram per square centimeter
1145kgcm2kilogram per square centimeter
1146inH2Oinches of water
1147inH2O_4Cinches of water at 4°C
1148inH2O_68Finches of water at 68°F
1149mmH2Omillimeters of water
1150mmH2O_4Cmillimeters of water at 4°C
1151mmH2O_68Fmillimeters of water at 68°F
1152ftH2Ofeet of water
1153ftH2O_4Cfeet of water at 4°C
1154ftH2O_68Ffeet of water at 68°F
1155inHginches of mercury
1156inHg_0Cinches of mercury at 0°C
1157mmHgmillimeters of mercury
1158mmHg_0Cmillimeters of mercury at 0°C
1318g/sgram per second
1319g/mingram per minute
1320g/hgram per hour
1321g/dgram per day
1322kg/skilogram per second
1323kg/minkilogram per minute
1324kg/hkilogram per hour
1325kg/dkilogram per day
1326t/smetric ton per second
1327t/minmetric ton per minute
1328t/hmetric ton per hour
1329t/dmetric ton per day
1330lb/spound per second
5-11
Unit CodeUnit String Displayed in LUIDescription
1331lb/minpound per minute
1332lb/hpound per hour
1333lb/dpound per day
1334STon/sshort ton per second
1335STon/minshort ton per minute
1336STon/hshort ton per hour
1337STon/dshort ton per day
1338LTon/slong ton per second
1339LTon/minlong ton per minute
1340LTon/hlong ton per hour
1341LTon/dlong ton per day
1342%percent
1347m3/ scubic meter per second
1348m3/mincubic meter per minute
1349m3/ hcubic meter per hour
1350m3/ dcubic meter per day
1351L/sliter per second
1352L/minliter per minute
1353L/hliter per hour
1354L/dliter per day
1355ML/dmegaliter per day
1356CFScubic feet per second
1357CFMcubic feet per minute
1358CFHcubic feet per hour
1359ft3/dcubic feet per day
1360SCFMstandard cubic feet per minute
1361SCFHstandard cubic feet per hour
1362gal/sUS gallon per second
1363GPMUS gallon per minute
1364gal/hUS gallon per hour
1365gal/dUS gallon per day
1366Mgal/dmega US gallon per day
1367ImpGal/sImperial gallon per second
1368ImpGal/minImperial gallon per minute
1369ImpGal/hImperial gallon per hour
1370ImpGal/dImperial gallon per day
1371bbl/sbarrel per second
1372bbl/minbarrel per minute
1373bbl/hbarrel per hour
1374bbl/dbarrel per day
1449mgal/smilli US gallon per second
1450kgal/skilo US gallon per second
1451Mgal/smega US gallon per second
1453mgal/minmilli US gallon per minute
1454kgal/minkilo US gallon per minute
1455Mgal/minmega US gallon per minute
1457mgal/hmilli US gallon per hour
1458kgal/hkilo US gallon per hour
1459Mgal/hmega US gallon per hour
1461mgal/dmilli US gallon per day
1462kgal/dkilo US gallon per day
1463Mgal/dmega US gallon per day
1464mImpGal/smilli imperial gallon per second
1465kImpGal/skilo imperial gallon per second
1466MImpGal/smega imperial gallon per second
1468mImpGal/minmilli imperial gallon per day
1469kImpGal/minkilo imperial gallon per day
5-12
Unit CodeUnit String Displayed in LUIDescription
1470MImpGal/minmega imperial gallon per day
1472mImpGal/hmilli imperial gallon per hour
1473kImpGal/hkilo imperial gallon per hour
1474MImpGal/hmega imperial gallon per hour
1476mImpGal/dmilli imperial gallon per day
1477kImpGal/dkilo imperial gallon per day
1478MImpGal/dmega imperial gallon per day
1482Mbbl/smegabarrel per second
1486Mbbl/minmegabarrel per minute
1490Mbbl/hmegabarrel per hour
1494Mbbl/dmegabarrel per day
1496mm3/scubic millimeter per second
1497km3/scubic kilometer per second
1498Mm3/scubic megameter per second
1500mm3/mincubic millimeter per minute
1501km3/mincubic kilometer per minute
1502Mm3/mincubic megameter per minute
1504mm3/hcubic millimeter per hour
1505km3/hcubic kilometer per hour
1506Mm3/hcubic megameter per hour
1508mm3/dcubic millimeter per day
1509km3/dcubic kilometer per day
1510Mm3/dcubic megameter per day
1511cm3/scubic centimeter per second
1512cm3/mincubic centimeter per minute
1513cm3/hcubic centimeter per hour
1514cm3/dcubic centimeter per day
1518kL/minkiloliter per minute
1519kL/hkiloliter per hour
1520kL/dkiloliter per day
1522Nm3/sNormal cubic meter per second
1523Nm3/minNormal cubic meter per minute
1524Nm3/hNormal cubic meter per hour
1525Nm3/dNormal cubic meter per day
1527Sm3/sStandard cubic meter per second
1528Sm3/minStandard cubic meter per minute
1529Sm3/hStandard cubic meter per hour
1530Sm3/dStandard cubic meter per day
1532NL/sNormal liter per second
1533NL/minNormal liter per minute
1534NL/hNormal liter per hour
1535NL/dNormal liter per day
1537SL/sStandard liter per second
1538SL/minStandard liter per minute
1539SL/hStandard liter per hour
1540SL/dStandard liter per day
1589mL/minmilliliters per minute
1617ML/hmegaliter per hour
1618ML/minmegaliter per minute
1619kL/skiloliter per second
1620kft3/ dcubic kilofeet per day
1621kCFHcubic kilofeet per hour
1622kCFMcubic kilofeet per minute
1623kCFScubic kilofeet per second
1624mft3/ dcubic millifeet per day
1625mCFHcubic millifeet per hour
1626mCFMcubic millifeet per minute
5-13
Unit CodeUnit String Displayed in LUIDescription
1627mCFScubic millifeet per second
1648kgalkilogallon
1649kImpGalkilo-imperial gallon
1653Mft3/ dcubic Megafeet per day
1654Mm3/ dcubic Megameters per day
Table 5-5 Display Transducer Block Display Status List
QualitySubstatusDisplay StringStatus Detail
0 : Bad0Bad-Non_specNon-specific
1Bad-ConfigErrorConfiguration Error
2Bad-NotConnectedNot Connected
3Bad-DeviceFailurDevice Failure
4Bad-SensorFailurSensor Failure
5Bad-NoCommWitLNo Comm, with LUV
6Bad-NoCommWitNNo Comm, no LUV
7Bad-OutOfServiceOut of Service
8Bad-TrnsducInMATransducer in MAN
1 : Uncertain0Uncertn-Non_specNon-specific
1Uncertn-LUVLast Usable Value
2Uncertn-SubsttutSubstitute / Manual Entry
3Uncertn-InitValuInitial Value
4Uncertn-S-CvNotASensor Conversion not Accurate
5Uncertn-EgUnRaVEngineering Unit Range Violation
6Uncertn-SubNomaSub-normal
7Uncertn-TrdInMATransducer in MAN
2 : GOOD(NC) 0GOOD-NC-Non_spNon-specific
1GOOD-NC-ActBkAActive Block Alarm
2GOOD-NC-ActAdAActive Advisory Alarm
3GOOD-NC-ActCrAActive Critical Alarm
4GOOD-NC-UakBkUnack Block Alarm
5GOOD-NC-UakAdUnack Advisory Alarm
6GOOD-NC-UakCrUnack Critical Alarm
8GOOD-NC-IFSInitiate Fault State(IFS)
3 : GOOD(C)0GOOD-C-Non_specNon-specific
1GOOD-C-InitAckInitialization Acknowledge
2GOOD-C-InitReqInitialization Request
3GOOD-C-NotInvitNot Invited
4GOOD-C-NotSelctNot Selected
6GOOD-C-LoclOvrrLocal Override
7GOOD-C-FaltStAcFault State Active
8GOOD-C-IFSInitiate Fault State(IFS)
5-14
Table 5-6 Display Transducer Block Display Alarm List
FD_xxx_ACTIVE BitDisplay StringDescription
0CheckCheck Function Bit
1Not used
2Not used
3Not used
4Gain CalibrationGain Calibration
5Zero CalibrationZero Calibration
6Fixed EX CurrentFixed EX Current
7Touch-Key ActiveTouch-Key Active
8Not used
9Not used
10Not used
11Not used
12Not used
13Zero Calib. FailZero Calib. Failure
14Scale DetectedScale Detected
15Not CalibratedNot Calibrated
16Not usedNot used
17Not usedNot used
18Not usedNot used
19Not usedNot used
20Not usedNot used
21Not usedNot used
22Empty DetectedEmpty Detected
23Flow Rate OverFlow Rate Over
24Not usedNot used
25Not usedNot used
26Not usedNot used
27Not usedNot used
28MainBoardCommErrMainBoardCommunicationError
29Extube Coil OpenExtube Coil Open
30Main Board FailMain Board Failure
31FFopt Board FailFF Option Board Failure
5-15
Diag Transducer Block
S
Overview of function
Diag Transducer Block calculates values for diagnosing the integrity of the flow rate signal
based on the sensor output from the electromagnetic flowmeter.
The calculated diagnostic value is output to AI Function Block.
The block judges the diagnostic value by comparing it against the threshold value. The
result is output in the device status summary of Resource Block as an alarm.
Function Block diagram
The configuration of Diag Transducer Block is presented in the figure below.
Diag_TB
ensor value
SCALE_FACTOR
AI_FB
DEVICE_STATUS_SUM1
RB_FB
Figure5-2 Function Block diagram
Parameters
Major parameters of Diag Transducer Block are described as follows.
Refer to the attached Parameter List for more descriptions of parameters.
Output
SCALE_LEVEL
Displays the value for diagnosing the integrity of the flow rate signal.
Setting
SCALE_LEVEL_RANGE
Indicates setting values for the upper and lower range limits, unit, and the position of
the decimal point for SCALE_LEVEL (value for diagnosing integrity of the flow rate
signal).
The unit for the diagnostic value is UNIT_LESS (1615). The unit must be identical to
the unit assigned for XD_SCALE of the block being used in order to connect with AI
Function Block.
SCALE_LEVEL_MOVING_AVERAGE_MODE
Select whether to calculate a moving average for the diagnostic value.
SCALE_LEVEL_MOVING_AVERAGE_TIME
Assign a moving average time for the diagnostic value.
SCALE_LEVEL_THRESHOLD
Specify a threshold value for diagnosing the integrity of the flow rate signal.
If the diagnostic value exceeds the threshold value, an alarm is generated as an output.
SCALE_LEVEL_HYSTERESIS
Specify a hysteresis for diagnosing the integrity of the flow rate signal.
Assign a value smaller than the threshold value for diagnosing the integrity of the flow
rate signal.
5-16
6. Maintenance and Troubleshooting
Overview
This chapter describes maintenance and checkup procedures for the electromagnetic
flowmeter and reference information for troubleshooting.
6.1 Troubleshooting
Types of trouble
Introduction
There are three types of trouble that could happen when the electromagnetic flowmeter is
started up for operation.
Trouble caused by the deviation of actual conditions of use from the specifications of
the electromagnetic flowmeter
Trouble caused by mistakes in settings or operation
Trouble caused by failures of the electromagnetic flowmeter
Trouble experienced during the operation of the electromagnetic flowmeter is recognized
either as a “major trouble” or a “minor trouble” by the self-diagnosis functions of the
transmitter, before it is displayed or addressed.
If any trouble occurs, refer to the troubleshooting guide described here to take appropriate
corrective action.
Major trouble
A case of major trouble refers to a state or failure that significantly impairs the operation
of the electromagnetic flowmeter and will lead to the damage of the flowmeter if
left unaddressed. If a case of major trouble takes place during the operation of the
electromagnetic flowmeter, an error message appears in the display panel of the
transmitter’s main unit.
Minor trouble
A case of minor trouble refers to a state or failure that does not significantly impair the
operation of the electromagnetic flowmeter. If the transmitter self-diagnoses that a trouble
event that occurred during operation is a case of minor trouble, the electromagnetic
flowmeter continues to output the instantaneous flow rate value.
6-1
Trouble at the start of operation
Troubleshooting
Address a problem occurred at the start of operation in accordance with the table below.
If the instruction in the table does not help eliminate the trouble, the electromagnetic
flowmeter may be out of order. Please contact us by referring to the end of this instruction
manual.
TroubleCheckpoint and action
Nothing is displayed in the
LUI even when the power is
turned on.
No output signal is
transmitted even when the
power is turned on.
· Check the specifications of the power supply for the
transmitter.
· Make sure that the ambient temperature is not below
-25 °C.
Make sure that the LUI display is turnThis completes the
startup of the electromagnetic flowmeter.
6-2
Trouble during operation
Troubleshooting
Address a problem occurred during operation by taking the following steps.
1. Check if the same trouble is described in the table. If yes, address the problem as
instructed in the table.
2. If the trouble cannot be resolved after following the above step, the electromagnetic
flowmeter may be out of order.
Please contact us by referring to the end of this instruction manual.
TroubleCheckpoint and action
The output value fluctuates
substantially when
compared to the anticipated
fluctuation range of flow
rate.
The output value remains at
0%.
Display Transducer Block Does Not Switch To Auto (LUI display says “DISPLAY_O/S.”)
TroubleCheckpoint and action
Display TB Does Not
Switch To Auto (LUI
display says “DISPLAY
O/S”)
Note: When writing is not possible for BLOCK_TAG_SELECTION_n (n=1 to 4), or
PARAMETER_SELECTION_n (n=1 to 4), switch the settings of RB FEATURE_SEL
Bit_12 (Deferral of Inter-Parameter Write Checks) to ON (enabled).
· Make sure that the electromagnetic flowmeter is properly
grounded.
· Make sure that the damping constant is properly
configured.
· Clean the electrodes.
· Check if the piping is empty.
(i.e., the empty detection function is activated)
· Check if the signal line is properly connected.
· Check if the upstream or downstream valves are open.
· Check if BLOCK_TYPE_SELECTION_n (n=1 to 4) is
configured. If not, BLOCK_TYPE_SELECTION_n will
be 0. In that case, assign the BLOCK_TAG(block tag
name) of the parameter you want to display to BLOCK_
TAG_SELECTION_n (n=1 to 4).
· Check if PARAMETER_SELECTION_n (n=1 to 4) is set
to the parameter you want to display.
· Check if DISPLAY_PARAM_SELECTION is set to the
parameters you want to display.
Example: If DISPLAY_PARAM_SELECTION=Parameter
1 and Parameter 2
BLOCK_TAG_SELECTION_1, PARAM_SELECTION_1,
BLOCK_TAG_SELECTION_2, and PARAM_
SELECTION_2 are appropriately configured.
6-3
Error messages and corrective actions
This section describes errors that may be experienced in each block and corrective actions.
Swiftly take proper measures or corrective actions.
Errors experienced with Resource Block
Error displayError descriptionCorrective action
Option Board Fail
Main Board Fail
Extube Coil Open
MainBoardCommErr
Flow Rate Over-Range
Empty Detected
Not Calibrated
Scale Detected
Zero Calib. Fail
Local User I/F Active
Fixed Excitation
Current Mode
Zero Calibration on
Gain Calibration on
ROM error or EEPROM error in
the option board
NVM error, AD converter error,
ROM error, or RAM error in the
main board
Coil disconnection1. Check the connection.
Communication error between
boards
Flow velocity exceeds 10 m/s.1. Reduce the flow velocity to
The empty pipe detector is
activated.
Calibration has not been made.1. Reboot.
The value for diagnosing the
integrity of the flow rate signal
exceeds the threshold.
Failure in zero adjustmentCheck if the measurement
The LUI is enabled.Return to the measuring mode.
Excitation out mode EXX,
excitation out mode EXY, or
excitation current off
Zero adjustment is being
performed.
Transmitter gain is adjusted at
0 m/s, 2.5 m/s, and 10.0 m/s.
1. Reboot.
2. Replace the main P/C.
1. Reboot.
2. Replace the main P/C.
2. Measure the coil’s resistance.
3. Reboot.
1. Reboot.
2. Replace the main P/C.
3. Replace the Fieldbus option
P/C.
10 m/s or below.
2. Replace the flowtube with
another one with larger
diameter.
1. Check if the piping is filled
with water.
2. Clean the electrodes.
2. Replace the main P/C, then
contact us.
1. It is time to perform
maintenance work. Clean
inside the flowtube.
piping is empty or if the flow
velocity exceeds 0.2 m/s
Bring the excitation current to
the original state by selecting “0”
(None) for EX_OUTPUT_CMD
in Flow Transducer Block.
Wait until zero adjustment is
complete.
Wait until gain adjustment is
complete.
6-4
Errors experienced with Flow Transducer Block
Error displayCorrective action
1. Check the connection.
SENS EX FAILURE
SENS NVM FAILURE
SENS ADC FAILURE
SENS ROM FAILURE
SENS RAM FAILURE
IO FAILURE
AUTO ZERO FAILURE
NOT CALIBRATED
FLOWTUBE_CONFIG
ERROR
PVR CONFIG ERROR
SVR CONFIG ERROR
FLSW_1 CONFIG ERROR
FLSW_2 CONFIG ERROR
2. Measure the coil’s resistance.
3. Reboot.
1. Reboot.
2. Replace the main P/C.
1. Reboot.
2. Replace the main P/C.
1. Reboot.
2. Replace the main P/C.
1. Reboot.
2. Replace the main P/C.
1. Reboot.
2. Replace the main P/C.
3. Replace the Fieldbus option P/C.
Check if the measurement piping is empty or if the flow velocity
exceeds 0.2 m/s.
1. Reboot.
2. Replace the main P/C.
Check the diameter and flowtube, and then enter the correct data.
PVR exceeds the flow rate equivalent to 10 m/s.
Check the settings of PVR, diameter, flowtube type, and dummy.
SVR exceeds the flow rate equivalent to 10 m/s.
Check the settings of SVR, diameter, flowtube type, dummy,
and density.
flsw_1_threshold exceeds the value of PVR when
flsw_1_source is PV and flsw_1_threshold exceeds the value
of SVR when flsw_1_source is SV. Check the settings of
flsw_1_threshold.
flsw_2_threshold exceeds the value of PVR when
flsw_2_source is PV and flsw_2_threshold exceeds the value
of SVR when flsw_2_source is SV. Check the settings of
flsw_2_threshold.
Errors experienced with Diag Transducer Block
Error displayCorrective action
scale_level_hysteresis exceeds
PARA_ERROR
SENSOR_ERROR
scale_level_threshold. Check the settings of
scale_level_threshold.
A sensor error has occurred.
Check BLOCK_ERR_DESC in Flow Transducer Block.
6-5
6.2 Input of simulated signals from calibrator
Introduction
A special calibrator has been designed for the electromagnetic flowmeter. This special
calibrator can generate the same kind of signal as the flow rate signal produced by
the flowtube. This simulated signal can be used for checking relevant functions of the
transmitter.
Getting started
Prepare the following devices and components.
A special calibrator (MGZ13) and a special cable
A digital multimeter
250- resistance
Checking method
Check the relevant functions in accordance with F1X 1000 Flow Meter Calibrator User's
Manual
(CM2-F1X100-2001).
6-6
6.3 Measurement of excitation current
0.0000
0.0000
0.0000
0.0000
0.0000
0.0000
0.0000
Introduction
The value and direction of the excitation current that flows in the coil in the flowtube can
be checked.
The current value is checked by connecting an ammeter in series to the excitation cables
of the flowtube and transmitter.
Measuring method using LUI
StepProcedureScreen
Follow the procedure for entering MAINTENANCE MODE
to display the screen for checking the excitation current.
1
In this state, the excitation current flows from X to Y.
Confirm that the current value becomes 160 mA.
EX CHECK
>
EXX1 6 0 . 0
Touch the
2
Touch the
from Y to X. This changes the polarity to the opposite of the
3
key once to move the cursor underneath E.
key once to cause the excitation current to flow
state in Step 2.
Touch the
4
current.
Finally, touch the
5
underneath >.
Use the
excitation returns to the current signal of the rectangular
6
key once to stop the flow of the excitation
key once to move the cursor
or key to transit to a different screen. The
wave.
Next, display the screen for setting the excitation current
7
value. In this state, the excitation current flows from X to Y.
Use the
8
that was measured in Step 1.
Use the
excitation returns to the current signal of the rectangular
9
key or key to set the excitation current value
or key to transit to a different screen. The
wave.
EX CHECK
>
EXX1 6 0 . 0
EX CHECK
>
EXY1 6 0 . 0
EX CHECK
>
EXY1 6 0 . 0
EX CHECK
>
OFF1 6 0 . 0
6-7
Method using Fieldbus communication
Accessing the option from the parameter list
1. Set TARGET of MODE_BLK in FLOW Transducer Block to O/S (Out of Service).
2. Set “1” (EXX) to EX_OUTPUT_CMD in FLOW Transducer Block.
The excitation current is fixed to EXX.
Measure the excitation current and write down the excitation current value.
3. Set “0” (None) to EX_OUTPUT_CMD in FLOW Transducer Block.
The excitation current is returned to the original state.
4. Read the value of EX_CURRENT_VALUE in FLOW Transducer Block.
If this value is different from the value written down in Step 2, write the value written
down in Step 2 into EX_CURRENT_VALUE.
5. For flow measurement, set TARGET of Transducer Block in FLOW Transducer Block to
AUTO.
Accessing the option from the menu
Note: Use METHOD in the menu to check or adjust the excitation current.
METHOD may not operate with some hosts.
In such cases, check or adjust the excitation current by accessing the option through
the parameter list.
Device menu for communicators
Check or adjust the excitation current by selecting Device Maintenance Excitation
Current Calibration (METHOD).
Block menu for communicators or Block menu for PCs
Check or adjust the excitation current by selecting FLOW Transducer Block Block
Maintenance Excitation Current Calibration (METHOD).
6-8
6.4 Gain adjustment
Measuring method using LUI
StepProcedureScreen
Enter CALIBRATION MODE and display the screen shown
1
on the right.
Set the cursor at the position of READY and
press the key to start gain calibration at the
zero point (0 m/s).
2
When the screen shown on the right is displayed, the gain
calibration at the zero point (0 m/s) has been completed.
Perform gain calibration at 2.5 m/s next.
Display the screen shown on the right.
3
Use the calibrator to input 2.5 m/s.
0.0000
0.0000
Gain calibration starts (the
main display blinks).
A few seconds later
Gain calibration is complete.
0.0000
2.5000
2.5000
Set the cursor at the position of READY and
press the key to start gain calibration at 2.5
m/s.
4
When the screen shown on the right is displayed, the gain
calibration at 2.5 m/s has been completed.
Perform gain calibration at 10 m/s in the same way.
Gain calibration starts (the
main display blinks).
A few seconds later
Gain calibration is complete.
2.5000
6-9
Method using Fieldbus communication
Accessing the option from the parameter list
1. Set TARGET of MODE_BLK in FLOW Transducer Block to O/S (Out of Service).
2. Use the calibrator to input 0.0 m/s.
3. Set “1” (0.0 m/s) to GAIN_CALIBRATION_CMD in FLOW Transducer Block.
Zero adjustment starts.
To cancel the zero adjustment, set “4” (Canceled) to GAIN_CALIBRATION_CMD in
FLOW Transducer Block.
4. Execution results of automatic zero adjustment can be checked at
GAIN_CALIBRATION_STATUS in FLOW Transducer Block.
While “1” (Executing) is displayed during execution of zero adjustment, “2” (Success) is
displayed when zero adjustment is completed successfully.
5. Use the calibrator to input 2.5 m/s.
6. Set “2” (2.5 m/s) to GAIN_CALIBRATION_CMD in FLOW Transducer Block.
Gain adjustment of 2.5 m/s starts.
To cancel the 2.5 m/s gain adjustment, set “4” (Canceled) to
GAIN_CALIBRATION_CMD in FLOW Transducer Block.
7. Execution results of 2.5 m/s gain adjustment can be checked at
GAIN_CALIBRATION_STATUS in FLOW Transducer Block.
While “1” (Executing) is displayed during execution of 2.5 m/s gain adjustment, “2”
Success} is displayed when 2.5 m/s gain adjustment is completed successfully.
8. Use the calibrator to input 10.0 m/s.
9. Set “3” (10.0 m/s) to GAIN_CALIBRATION_CMD in FLOW Transducer Block. Gain
adjustment of 10.0 m/s starts.
To cancel the 10.0 m/s gain adjustment, set “4” (Canceled) to
GAIN_CALIBRATION_CMD in FLOW Transducer Block.
10. Execution results of 10.0 m/s gain adjustment can be checked at
GAIN_CALIBRATION_STATUS in FLOW Transducer Block.
While “1” (Executing) is displayed during execution of 10.0 m/s gain adjustment, “2”
(Success) is displayed when 10.0 m/s gain adjustment is completed successfully.
11. For flow measurement, set TARGET of Transducer Block in FLOW Transducer Block
to AUTO.
Accessing the option from the menu
Note: Use METHOD in the menu to perform gain adjustment.
METHOD may not operate with some hosts.
In such cases, perform gain adjustment by accessing the option through the
parameter list.
Device menu for communicators
Perform gain adjustment by selecting Device Maintenance Gain Adjustment
(METHOD).
Block menu for communicators or Block menu for PCs
Perform gain adjustment by selecting FLOW Transducer Block Block Maintenance
Gain Adjustment (METHOD).
6-10
Appendix A. View List
Resource Block
Index Parameter MnemonicVIEW_1VIEW_2VIEW_3VIEW_4VIEW_4_2
1ST_REV22222
2TAG_DESC
3STRATEGY2
4ALERT_KEY1
5MODE_BLK44
6BLOCK_ERR22
7RS_STATE11
8TEST_RW
9DD_RESOURCE
10MANUFAC_ID4
11DEV_TYPE2
12DEV_REV1
13DD_REV1
14GRANT_DENY2
15HARD_TYPES2
16RESTART
17FEATURES2
18FEATURE_SEL2
19CYCLE_TYPE2
20CYCLE_SEL2
21MIN_CYCLE_T4
22MEMORY_SIZE2
23NV_CYCLE_T4
24FREE_SPACE4
25FREE_TIME44
26SHED_RCAS4
27SHED_ROUT4
28FAULT_STATE11
29SET_FSTATE
30CLR_FSTATE
31MAX_NOTIFY1
32LIM_NOTIFY1
33CONFIRM_TIME4
34WRITE_LOCK1
35UPDATE_EVT
36BLOCK_ALM
37ALARM_SUM88
38ACK_OPTION2
39WRITE_PRI1
40WRITE_ALM
41ITK_VER2
42FD_VER 2
43FD_FAIL_ACTIVE44
FD_OFFSPEC_
44
ACTIVE
FD_MAINT_
45
ACTIVE
FD_CHECK_
46
ACTIVE
47FD_FAIL_MAP4
44
44
44
A-1
Index Parameter MnemonicVIEW_1VIEW_2VIEW_3VIEW_4VIEW_4_2
Index Parameter Mnemonic VIEW_1VIEW_2 VIEW_3 VIEW_4
1ST_REV22222222
2TAG_DESC
3STRATEGY2
4ALERT_KEY1
5MODE_BLK44
6BLOCK_ERR22
7UPDATE_EVT
8BLOCK_ALM
TRANSDUCER_
9
DIRECTORY
TRANSDUCER_
10
TYPE
TRANSDUCER_
11
TYPE_VER
12XD_ERROR11
COLLECTION_
13
DIRECTORY
BLOCK_ERR_
14
DESC_1
DISPLAY_
15
PARAM_SEL
DISPLAY_INFO_
16
SEL
17DISPLAY_CYCLE1111
BLOCK_TYPE_
18
SEL_1
BLOCK_TAG_
19
SEL_1
PARAM_
20
SELECTION_1
21DISPLAY_TAG_132
UNIT_
22
SELECTION_1
23CUSTOM_UNIT_132
24EXPONENT_SEL_11
BLOCK_TYPE_
25
SEL_2
BLOCK_TAG_
26
SEL_2
PARAM_
27
SELECTION_2
28DISPLAY_TAG_232
UNIT_
29
SELECTION_2
30CUSTOM_UNIT_232
31EXPONENT_SEL_21
BLOCK_TYPE_
32
SEL_3
BLOCK_TAG_
33
SEL_3
34PARAM_SEL32
35DISPLAY_TAG332
2222
44
2
2
2
VIEW_4_2 VIEW_4_3 VIEW_4_4 VIEW_4_5
2
1111
1111
32
2
1
32
2
1
32
A-6
Index Parameter Mnemonic VIEW_1VIEW_2 VIEW_3 VIEW_4
36UNIT_SEL31
37DISPLAY_UNIT332
38EXP_MODE31
BLOCK_TYPE_
39
SEL_4
BLOCK_TAG_
40
SEL_4
41PARAM_SEL42
42DISPLAY_TAG432
43UNIT_SEL41
44DISPLAY_UNIT432
45EXP_MODE41
46CLEAR_HISTORY
OPERATION_
47
HISTORY1
OPERATION_
48
HISTORY2
OPERATION_
49
HISTORY3
OPERATION_
50
HISTORY4
OPERATION_
51
HISTORY5
OPERATION_
52
HISTORY6
OPERATION_
53
HISTORY7
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
2
VIEW_4_2 VIEW_4_3 VIEW_4_4 VIEW_4_5
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
32
A-7
Index Parameter Mnemonic VIEW_1VIEW_2 VIEW_3 VIEW_4
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
54
HISTORY8
OPERATION_
56
HISTORY10
1542399105105105105
VIEW_4_2 VIEW_4_3 VIEW_4_4 VIEW_4_5
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
A-8
DIAG Transducer Block
Index Parameter MnemonicVIEW_1VIEW_2VIEW_3VIEW_4
1ST_REV22
2TAG_DESC
3STRATEGY2
4ALERT_KEY1
5MODE_BLK44
6BLOCK_ERR22
7UPDATE_EVT
8BLOCK_ALM
TRANSDUCER_
9
DIRECTORY
TRANSDUCER_
10
TYPE
TRANSDUCER_
11
TYPE_VER
12XD_ERROR11
COLLECTION_
13
DIRECTORY
BLOCK_ERR_
14
DESC_1
15SCALE_FACTOR55
SCALE_FACTOR_
16
RANGE
SCALE_FACTOR_
17
MOVING_
AVERAGE_MODE
SCALE_FACTOR_
18
MOVING_
AVERAGE_TIME
SCALE_FACTOR_
19
THRESHOLD
SCALE_FACTOR_
20
HYSTERESIS
21FB_CYCLE_MIN
22FB_CYCLE_MAX
23FB_EXEC_MIN
24FB_EXEC_MAX
2222
44
11
20152019
2
1
1
4
4
A-9
A-10
Appendix B. Parameter List
Resource Block (Base INDEX)
INDEX Parameter nameDescription
Indicates the number of changes of the
Static parameter that belongs to the
1ST_REV
2TAG_DESC
3STRATEGY
4ALERT_KEY
5MODE_BLK
6BLOCK_ERR
7RS_STATEIndicates the operating state of the device.-D-R1
8TEST_RW
9DD_RESOURCE (Unused)-S-R32spaces
10MANUFAC_ID
11DEV_TYPE
12DEV_REV
13DD_REV
Resource block. It is incremented by one
(0x0001) when a change is made for a
parameter whose access attribute is “S-.”
A tag name for the Resource block set
by the user. It can be referenced by the
host devices and thus does not affect the
execution of Function block operation at
all.
An arbitrary group number for the
Resource block. It does not affect Function
block operation.
The identification number of the relevant
in-plant device. It does not affect Function
block operation.
Indicates the mode parameter group of
the Resource block. The configuration is
as follows:
· Target: Parameter for setting mode
from host devices.
· Actual: Indicates the current value of the
mode.
· Permitted: Indicates a mode value used
for the Function block.
· Normal: Indicates a mode value which
should be in a steady state.
Indicates the error state regarding the
Resource block.
This is a parameter for compatibility
testing of communication software. Users
are not supposed to use this parameter.
An identification number specific to each
manufacturer registered in the Fieldbus
Association.
An identification number that indicates a
device model defined by the manufacturer.
A revision number defined by the
manufacturer.
Revision number of a DD file to be applied
to this device.
Sub-parameter
name
-S-R20 ≤ X ≤ 655350
-S-R/W32spaces
-S-R/W20 ≤ X ≤ 655350
-S-R/W11 ≤ X ≤ 2550
TargetN-R/W1
ActualD-R1
PermittedS-R/W1
NormalS-R/W1
-D-R2
-D-R/W112
-S-R40x0DFC960x0DFC96
-S-R20 ≤ X ≤ 0xFFFF0x1601
-S-R10 ≤ X ≤ 0xFF0x01
-S-R10 ≤ X ≤ 0xFF0x01
Access
attribute
Size
(Byte)
bit3: Auto
bit7: OOS
bit3: Auto
bit7: OOS
bit3: Auto
bit7: OOS
bit3: Auto
bit7: OOS
0: Other
1: Block Configuration Error
2: Link Configuration Error
3: Simulate Active
5: Device Fault State Set
6: Device Needs Maintenance
Soon
9: Memory Failure
10: Lost Static Data11:Lost NV
Data
13: Device Needs Maintenance
Now
14: Power -up
15: Out-of-Service
0: Undefined
1: Start/Restart
2: Initialization
3: Online Linking
4: Online
5: Standby
6: Failure
Range
*1 : Refer to Display Transducer Block INDEX18 BLOCK_TYPE_SELECTION_1
*2 : Refer to Table 5-3
Factory default
setting
0x08
bit3:Auto
-
0x88
bit3: Auto
bit7: OOS
0x08
bit3:Auto
-
-
B-1
INDEX Parameter nameDescription
This parameter is used to permit/deny
14GRANT_DENY
15HARD_TYPES
16RESTART
17FEATURES
18FEATURE_SELMakes an option setting for device use.-S-R/W2
19CYCLE_TYPE
20CYCLE_SELSets the Function block execution method.-S-R/W2bit0: Scheduled0
21MIN_CYCLE_T
22MEMORY_SIZE
23NV_CYCLE_T
24FREE_SPACE
25FREE_TIME
access from MMI or other host devices to
parameters in this block.
Indicates the type of hardware in which
this Resource block exists.
The device is restarted manually. A restart
type can be selected from several types in
the specification.
Sets an option that can be selected for
FEATURE_SEL in the option settings for
device use.
Indicates the current operating state
based on the setting of CYCLE_SEL in the
Function block execution method.
Indicates the minimum period in which
the Function block can be executed.
Indicates the memory capacity that can be
used as a guideline for adding Function
blocks. (Unused)
Indicates the minimum time needed for
writing N-type parameters in non-volatile
memory. (Unused)
Indicates the available memory as a
guideline for adding configurations.
Indicates the loaded condition showing
how much idle time is available for
Function block execution time. (Unused)
Sub-parameter
name
GrantD-R/W1
DenyD-R/W1
-S-R2bit0: Scalar Input
-D-R/W1
-S-R2
-S-R2bit0: Scheduled
-S-R440004000
-S-R200
-S-R4345600000(3h)345600000(3hr)
-D-R40 ≤ X ≤ 100-
-D-R40 ≤ X ≤ 100-
Access
attribute
Size
(Byte)
bit0: Program
bit1: Tune
bit2: Alarm
bit3: Local
bit4: Operate
bit5: Service
bit6: Diagnostic
bit0: Program Denied
bit1: Tune Denied
bit2: Alarm Denied
bit3: Local Denied
bit4: Operate Denied
bit5: Service Denied
bit6: Diagnostics Denied
1: Run
2: Restart resource
3: Restart with defaults
4: Restart processor
11: Restores Factory default
blocks
12: Resets transducer block
Factory calibration
bit0: Unicode strings
bit1: Reports supported
bit2: Fault State supported
bit3: Soft Write lock supportedx
bit10: Multi-bit Alarm(Bit-
Alarm) Support
bit12: Deferral of Inter-
Paramter Write Checks
bit0: Unicode strings
bit1: Reports supported
bit2: Fault State supported
bit3: Soft Write lock supportedx
bit10: Multi-bit Alarm (Bit-
Alarm) Support
bit12: Deferral of Inter-
Paramter Write Checks
Range
Factory default
setting
-
-
0x01bit0: Scalar
Input
-
0x140F
bit0: Unicode
strings
bit1: Reports
supported
bit2: Fault State
supported
bit3: Soft Write
lock supportedx
bit10:Multi-bit
Alarm(Bit-Alarm)
Support
bit12:Deferral of
Inter-Paramter
Write Checks
0x100A
bit1: Reports
supported
bit3: Soft Write
lock supportedx
bit12:Deferral of
Inter-Paramter
Write Checks
0x0001
bit0: Scheduled
B-2
INDEX Parameter nameDescription
Sets the timeout time for writing the value
for a setting value change (SPC) from the
host arithmetic unit connected with the
RCAS_IN parameter when MODE of the
26SHED_RCAS
27SHED_ROUT
28FAULT_STATEIndicates the fail-safe condition.-N-R1
29SET_FSTATEStarts the fail-safe state.-D-R/W1
30CLR_FSTATECancels the fail-safe state.-D-R/W1
31MAX_NOTIFY
32LIM_NOTIFY
33CONFIRM_TIME
34WRITE_LOCK
35UPDATE_EVT
Function block is RCAS. If the setting
value is not written within this set time,
the Function block automatically transits
to the mode which has been set to the
SHED_OPT parameter in the Function
block.
Sets the timeout time for writing the value
for an output value change (DDC) from
the host arithmetic unit connected with
the ROUT_IN parameter when MODE of
the Function block is ROUT. If the setting
value is not written within this set time,
the Function block automatically transits
to the mode which has been set to the
SHED_OPT parameter in the Function
block.
The maximum number of alerts to be
retained.
Limit of the number of alerts. Being set by
the user, the number of alerts to be notified
to the host can be limited to prevent
overflow of the host.
Parameter for setting the wait time for
confirmation of an alert.
Prohibits setting values from being written
from outside.
Parameter for generating an alert when
fixed data (data whose access attribute is
“S-”) of the Resource block is changed. The
configuration is as follows:
· Unacknowledged: Determined state
· Update_State: Change state
· Time_stamp: Change time
· Static_Revision: Revision after change
· Relative_Index: Changed parameter
identification number
Sub-parameter
name
-S-R/W4
-S-R/W4
-S-R133
-S-R/W10 ≤ X ≤ 33
-S-R/W4
-S-R/W1
Unacknowledged
Update StateD-R1
Time StampD-R8Static Revision D-R20 ≤ X ≤ 65535-
Relative Index D-R20 ≤ X ≤ 65535-
Access
attribute
D-R/W1
Size
(Byte)
0 ≤ X ≤ 0x
FFFFFFFF
0 ≤ X ≤ 0x
FFFFFFFF
1: Clear
2: Active
1: Off
2: Set
1: Off
2: Set
0 ≤ X ≤ 0x
FFFFFFFF
1: Unlocked
2: Locked
0: Undefined
1: Acknowledged
2: Unacknowledged
0: Undefined
1: Update reported
2: Update not reported
Range
Factory default
setting
640000(20sec)
640000(20sec)
1: Clear
-
-
640000(20sec)
1: Unlocked
-
-
36BLOCK_ALM
37ALARM_SUM
Parameter for generating an alert when
fixed data (data whose access attribute is
“S-”) of the Resource block is changed. The
configuration is as follows:
· Unacknowledged: Determined state
· Update_State: Change state
· Time_stamp: Change time
· Static_Revision: Revision after change
· Relative_Index: Changed parameter
identification number
Parameter that indicates the state of
BLOCK_ALM of the Resource block
comprehensively. The configuration is as
follows:
· Current: Current generation state
· Unacknowledged: Alarm checking state
· Unreported: Reporting state to the host
devices
· Disabled: Alarm detection inhibit state
0: Undefined
Unacknowledged
Alarm StateD-R1
Time StampD-R8SubcodeD-R2Valu eD-R1CurrentD-R2bit1: High high alarmUnacknowledged
UnreportedD-R2bit6: Deviation low alarm-
DisabledS-R/W2bit7: Block alarm0
D-R/W1
D-R2bit5: Deviation high alarm-
1: Acknowledged
2: Unacknowledged
0: Undefined
1: Clear - reported
2: Clear - not reported
3: Active - reported4
: Active - not reported
B-3
-
-
INDEX Parameter nameDescription
Permits or inhibits automatic checking
for generation of BLOCK_ALM
38ACK_OPTION
39WRITE_PRI
40WRITE_ALM
41ITK_VER
42FD_VER
FD_FAIL_
43
ACTIVE
FD_OFFSPEC_
44
ACTIVE
FD_MAINT_
45
ACTIVE
FD_CHECK_
46
ACTIVE
47FD_FAIL_MAP
of the Resource block. Automatic
checking means acknowledgment in
communication which is regarded as
equivalent to the operation by an operator.
Sets the priority of WRITE_ALM. Aside
from controlling priority, setting this
parameter also makes it possible to
disable informing alarms and eliminate
acknowledgement.
Generates an alarm when WRITE_LOCK
is cancelled.
Indicates the version of the FF
authentication test (interoperability test)
conducted on the device.
A value consistent with the major version
of the Field Diagnostics specification of the
device.
Parameter that reflects activation of the
error state classified into this category
(FAIL). Because this parameter is a bit
string parameter, it indicates more than
one state.
Parameter that reflects activation of the
error state classified into this category
(OFFSPEC). Because this parameter is a
bit string parameter, it indicates more than
one state.
Parameter that reflects activation of the
error state classified into this category
(MAINTENANCE). Because this
parameter is a bit string parameter, it
indicates more than one state.
Parameter that reflects activation of the
error state classified into this category
(CHECK). Because this parameter is a bit
string parameter, it indicates more than
one state.
Parameter that selects a condition
classified into this alarm category (FAIL).
The same condition may become active for
more than one category of the four alarm
categories.
Sub-parameter
name
-S-R/W2
-S-R/W10 ≤ X ≤ 150
Unacknowledged
Alarm StateD-R1
Time StampD-R8SubcodeD-R2Valu eD-R1-
-S-R26
-S-R21
-D-R4
-D-R4*1-
-D-R4*1-
-D-R4*1-
-S-R/W4*1
Access
attribute
D-R/W1
Size
(Byte)
0: Auto Ack Disabled
1: Auto Ack Enabled
0: Undefined
1: Acknowledged
2: Unacknowledged
0: Undefined
1: Clear - reported
2: Clear - not reported
3: Active - reported
4: Active - not reported
*1
bit31: Fieldbus Option Board
Failure
bit30: Main Board Failure
bit29: Flowtube Coil Open
Circuit
bit28: Main Board
Communications Error
bit23: Flow Rate Over-Range
bit22: Empty Detected
bit15: Not Calibrated
bit14: Scale Detected
bit13: Auto Zero Calibration
Failure
bit7: Local User I/F Active
bit6: Fixed Excitation Current
Mode
bit5: Performing Auto Zero
Calibration
bit4: Performing Gain
Calibration
Range
Factory default
setting
0: Auto Ack
Disabled
0xF0000000
bit31/bit30/bit29/
bit28
-
-
-
B-4
INDEX Parameter nameDescription
Parameter that selects a condition
FD_OFFSPEC_
48
MAP
49FD_MAINT_MAP
FD_CHECK_
50
MAP
51FD_FAIL_MASK
FD_OFFSPEC_
52
MASK
FD_MAINT_
53
MASK
FD_CHECK_
54
MASK
55FD_FAIL_ALM
FD_OFFSPEC_
56
ALM
classified into this alarm category
(OFFSPEC). The same condition may
become active for more than one category
of the four alarm categories.
Parameter that selects a condition
classified into this alarm category
(MAINTENANCE). The same condition
may become active for more than one
category of the four alarm categories.
Parameter that selects a condition
classified into this alarm category
(CHECK). The same condition may
become active for more than one category
of the four alarm categories.
Parameter for the user to set multiple states
of this category (FAIL), which should not
be notified to the host through an alarm
parameter. Setting the bit to 1 inhibits
the notification, while setting the bit to 0
permits the notification.
Parameter for the user to set multiple
states of this category (OFFSPEC), which
should not be notified to the host through
an alarm parameter. Setting the bit to 1
inhibits the notification, while setting the
bit to 0 permits the notification.
Parameter for the user to set multiple states
of this category (MAINTENANCE), which
should not be notified to the host through
an alarm parameter. Setting the bit to 1
inhibits the notification, while setting the
bit to 0 permits the notification.
Parameter for the user to set multiple
states of this category (CHECK), which
should not be notified to the host through
an alarm parameter. Setting the bit to 1
inhibits the notification, while setting the
bit to 0 permits the notification.
Parameter for notifying the state change
of unmasked items of this alarm category
(FAIL) to the host system.
Parameter for notifying the state change
of unmasked items of this alarm category
(OFFSPEC) to the host system.
Sub-parameter
name
-S-R/W4*1
-S-R/W4*1
-S-R/W4*1
-S-R/W4*10
-S-R/W4*10
-S-R/W4*10
-S-R/W4*10
unacknowledged
alarmstateD-R1
timestampD-R8subcodeD-R4valueD-R1-
unacknowledged
alarmstateD-R1
timestampD-R8subcodeD-R4valueD-R1-
Access
attribute
D-R/W1
D-R/W1
Size
(Byte)
0: Undefined
1: Acknowledged
2: Unacknowledged
0: Undefined
1: Clear - reported
2: Clear - not reported
3: Active - reported
4: Active - not reported
0: Undefined
1: Acknowledged
2: Unacknowledged
0: Undefined
1: Clear - reported
2: Clear - not reported
3: Active - reported
4: Active - not reported
Range
Factory default
setting
0x00C00000
bit23/bit22
0x0000E000
bit15/bit14/bit13
0x000000F
0bit7/bit6/bit5/
bit4
-
-
-
-
B-5
INDEX Parameter nameDescription
Parameter for notifying the state change
57FD_MAINT_ALM
58FD_CHECK_ALM
59FD_FAIL_PRI
FD_OFFSPEC_
60
PRI
61FD_MAINT_PRI
62FD_CHECK_PRI
63FD_SIMULATE
FD_
64
RECOMMEN_
ACT
CAPABILITY_
65
LEV
HARDWARE_
66
REV
67SOFTWARE_REV
68SIM_ACTIVE_SW
of unmasked items of this alarm category
(MAINTENANCE) to the host system.
Parameter for notifying the state change
of unmasked items of this alarm category
(CHECK) to the host system.
Parameter that sets the priority of this
alarm category (FAIL).
Parameter that sets the priority of this
alarm category (OFFSPEC).
Parameter that sets the priority of this
alarm category (MAINTENANCE).
Parameter that sets the priority of this
alarm category (CHECK).
Parameter that can change the state
manually when simulation is enabled.
When simulation is disabled, both the
simulated diagnostic value and the
diagnostic value indicate the actual
state. While FD_SIMULTE is enabled,
the recommended action indicates that
simulation is active.
Indicates the most important condition
that has been detected by the device.
Indicates the capability level of the device.-S-R1
Indicates the hardware revision of the
device.
Indicates the software revision of the
device.
Selects between enabled and disabled for
the simulation functionality.Select Set
Simulate Active to enable the simulation
functionality.
Sub-parameter
name
unacknowledged
alarmstateD-R1
timestampD-R8subcodeD-R4valueD-R1-
unacknowledged
alarmstateD-R1
timestampD-R8subcodeD-R4valueD-R1-
-S-R/W10 ≤ X ≤ 150
-S-R/W10 ≤ X ≤ 150
-S-R/W10 ≤ X ≤ 150
-S-R/W10 ≤ X ≤ 150
diag_sim_value D-R/W4*1diag_valueD-R/W4*1-
enableD-R/W1
-D-R2
-S-R32spacesspaces
-S-R32
-D-R/W2
Access
attribute
D-R/W1
D-R/W1
Size
(Byte)
0: Undefined
1: Acknowledged
2: Unacknowledged
0: Undefined
1: Clear - reported
2: Clear - not reported
3: Active - reported
4: Active - not reported
0: Undefined
1: Acknowledged
2: Unacknowledged
0: Undefined
1: Clear - reported
2: Clear - not reported
3: Active - reported
4: Active - not reported
0: Not Initialized
1: Simulation Disabled
2: Simulation Active
0: Uninitialized
1: No Action Required
2: Replace H/W
3: Check Fluid
4: Recalibrate
5: Remove Scale
6: Calibrating - Please Wait
7: Other
0: capability level not supported
1: Standard Model
2: Scale Diagnostic Model
0: Disabled
1: Active
Range
Factory default
setting
1: Standard
Modelor
2: Scale
Diagnostic Model
-
-
-
-
-
-
-
B-6
FLOW Transducer Block
INDEX Parameter nameDescription
Indicates the number of changes of the
1ST_REV
2TAG_DESC
3STRATEGY
4ALERT_KEY
5MODE_BLK
6BLOCK_ERR
7UPDATE_EVT
Static parameter that belongs to FLOW_
TB. It is incremented by one (0x0001)
when a change is made for a parameter
whose access attribute is “S-.”
The tag name of FLOW_TB set by the user.
It can be referenced by the host devices
and thus does not affect the execution of
Function block operation at all.
An arbitrary group number for FLOW_
TB. It does not affect Function block
operation.
The identification number of the relevant
in-plant device. It does not affect Function
block operation.
Indicates the mode parameter group
of FLOW_TB. The configuration is as
follows:
· Target: Parameter for setting mode from
host devices.
· Actual: Indicates the current value of the
mode.
· Permitted: Indicates the mode value used
for the Function block.
· Normal: Indicates the mode value which
should be in a steady state.
Indicates the error state regarding
FLOW_TB.
Parameter for generating an alert when
fixed data (data whose access attribute is
“S-” or “N-”) of FLOW_TB is changed.
The configuration is as follows:
· Unacknowledged: Determined state
· Update_State: Change state
· Time_stamp: Change time
· Static_Revision: Revision after change
· Relative_Index: Changed parameter
identification number
Indicates an abnormal state specific to the
device.
Definition information of the parameter
group for efficient access of the host
devices to parameters having the same
attribute.
Type of the PRIMARY_VALUE value
(volume flow rate value)
PRIMARY_VALUE value (volume flow
rate value) and its status
Range upper limit value and lower limit
value, unit, and setting value of the
decimal point position of the PRIMARY_
VALUE value (volume flow rate value).
Type of the SECONDARY_VALUE value
(mass flow rate value)
SECONDARY_VALUE value (mass flow
rate value) and its status
Range upper limit value and lower
limit value, unit, and setting value
of the decimal point position of the
SECONDARY_VALUE value (mass flow
rate value).
Option setting for determining the status
of the output value.
Indicates the upper limit value and
lower limit value, unit, and setting
value of the decimal point position of
the measurable flow velocity.Setting
MeasurementSENSOR_RANGE is a
parameter that indicates the measurable
range of the sensor. This parameter cannot
be set.
Serial number of the sensor module
(Unused for MGG)
Indicates the method of the flowmeter
adjustment that was last carried out.
Indicates the location of the flowmeter
adjustment that was last carried out.
The time of the flowmeter adjustment that
was last carried out can be set.
The person who carried out the flowmeter
adjustment last can be set.
Sub-parameter
name
[0]S-R40
StatusD-R1-
Valu eD-R4-
EU at 100%S-R/W4EU at 0%S-R/W400
Units IndexS-R/W2refer to table 1-
Decimal Point S-R/W1-128 to +1272
StatusD-R1-
Valu eD-R4-
EU at 100%S-R/W4EU at 0%S-R/W400
Units IndexS-R/W2refer to table 2-
Sets the diameter of the flowtube.S-R/W1refer to table 3-
Selects a flowtube type from MGG, KID,
and NNK.
S-R/W1
0: KID
2: NHK
3: MGG
Sets the flowtube constant that is specific
32
FLOWTUBE_
FAC TO R
to the flowtube.The flowtube constant has
been determined based on the actual flow
calibration at factory and is stamped on
S-R/W4100.0 ≤ X ≤ 999.9-
the name plate of the device.
0: 0
1: 1
2: 2
3: 3
4: 4
5: 5
6: 6
33
DUMMY_
NUMBER
Sets the number of dummy flowtubes to
install.This parameter needs to be set only
when an NNK flowtube is used.
S-R/W1
7: 7
8: 8
9: 9
EMPTY_PIPE_
34
DETECTOR
DENSITY_
35
CONSTANT
DENSITY_
36
CONSTANT_
Checks whether the empty detection
function is set.
D-R1
0: OFF
1: ON
Sets the density of measurement fluid.S-R/W40.1 ≤ X ≤ 9.9999-
Sets the density unit of measurement fluid.S-R/W21100: g/cm
3
1100: g/cm
UNITS
37
FLOW_
DIRECTION
Sets the flow direction of measurement
fluid.
S-R/W1
0: Forward
1: Reverse
0: Forward
Sets the correction coefficient.A value
38COEFFICIENT
which is multiplied by the correction
S-R/W40.1 ≤ X ≤ 9.99991.0
coefficient is output.
DAMPING_
39
CONSTANT
MOVING_
40
AVERAGE_
MODE
MOVING_
41
AVERAGE_TIME
PRIMARY_
42
VALUE_SPIKE_
CUT_MODE
PRIMARY_
43
VALUE_SPIKE_
CUT_TIME
PRIMARY_
44
VALUE_SPIKE_
CUT_LEVEL
Sets the damping constant to cut small
fluctuation components in the measured
S-R/W40.1 ≤ X ≤ 199.93.0
instantaneous flow rate value.
Selects between ON and OFF for the
moving average function.When ON is
selected, the moving average value of the
S-R/W1
0: OFF
1: ON
0: OFF
set moving average time is output.
Sets the moving average time.S-R/W41.0 ≤ X ≤ 30.01.0
Selects the spike cut mode from OFF,
AUTO, and MANUAL.
S-R/W1
0: OFF
1: Auto
2: Manual
0: OFF
Sets the spike cut time.This parameter
should be set only when the MANUAL
S-R/W40.0 ≤ X ≤ 99.91.0
spike cut mode is selected.
Sets the spike cut level.This parameter
should be set only when the MANUAL
S-R/W41.0 ≤ X ≤ 99.91.0
spike cut mode is selected.
setting
-
-
-
-
3
B-9
INDEX Parameter nameDescription
SECONDARY_
45
VALUE_SPIKE_
CUT_MODE
SECONDARY_
46
VALUE_SPIKE_
CUT_TIME
SECONDARY_
47
VALUE_SPIKE_
CUT_LEVEL
PRIMARY_
48
VALUE_LOW_
FLOW_CUTOFF
SECONDARY_
49
VALUE_LOW_
FLOW_CUTOFF
50AC_FREQUENCY
51EX_FREQUENCY Sets the frequency of excitation current.S-R/W1
EX_OUTPUT_
52
CMD
EX_OUTPUT_
53
STATUS
EX_CURRENT_
54
VALUE
AUTO_ZERO_
55
CALIBRATION_
CMD
AUTO_ZERO_
56
CALIBRATION_
STATUS
MANUAL_ZERO_
57
CALIBRATION_
CMD
Selects the spike cut mode from OFF,
AUTO, and MANUAL.
Sets the spike cut time.This parameter
should be set only when the MANUAL
spike cut mode is selected.
Sets the spike cut level.This parameter
should be set only when the MANUAL
spike cut mode is selected.
Sets the low flow cut function.This setting
can be made between 0% and 10% of the
setting range. When 0% is set, the low flow
cut function does not work.If the flow rate
is lower than the set value, the flow rate
output becomes zero and the display is also
fixed at zero.
Sets the low flow cut function.This setting
can be made between 0% and 10% of the
setting range. When 0% is set, the low flow
cut function does not work.If the flow rate
is lower than the set value, the flow rate
output becomes zero and the display is also
fixed at zero.
Indicates the frequency of the AC power
supply being used.
Outputs the excitation current at a fixed
value.
Displays the output state of the excitation
current.
A value used to adjust the excitation
current.
Executes the automatic zero adjustment.D-R/W1
Displays the result of the automatic
zero adjustment.If the automatic zero
adjustment cannot be executed, check
whether the measurement piping is empty
or whether the flow velocity is over 0.2 m/
s.
AC_FREAQUENCY is 60: For
60 Hz
0: 7.5Hz
1: 15.0Hz
2: 30.0Hz
0: None
1: EXX
2: EXY
3: EX Current Off
4: Normal
0: Normal
1: EXX
2: EXY
3: EX Current Off
73FLOWTUBE_SN Displays the serial number of the flowtube.S-R/W32-
Executes the gain adjustment of the
transmitter.
Displays the result of the transmitter gain
adjustment.
Displays the flow switch output and
outputs the state from DI Function Block.
Selects between Primary Value and
Secondary Value for the output to which
Flow Switch 1 is applied.
Selects between High alarm and Low
alarm for the alarm output of Flow Switch
1.
Sets the threshold value for the flow
switch.If Flow Switch 1 Mode is “High,”
Flow Switch 1 becomes On when the
threshold value exceeds this setting value.If
Flow Switch 1 Mode is “Low,” Flow Switch
1 becomes On when the threshold value
becomes below this setting value.
Sets hysteresis until Flow Switch 1 becomes
OFF again after it has become ON.
Displays the flow switch output and
outputs the state from DI Function Block.
Selects between Primary Value and
Secondary Value for the output to which
Flow Switch 2 is applied.
Selects between High alarm and Low
alarm for the alarm output of Flow Switch
2.
Sets the threshold value for the flow
switch.If Flow Switch 2 Mode is “High,”
Flow Switch 2 becomes On when the
threshold value exceeds this setting value.If
Flow Switch 2 Mode is “Low,” Flow Switch
2 becomes On when the threshold value
becomes below this setting value.
Sets hysteresis until Flow Switch 2 becomes
OFF again after it has become ON.
Displays the transmitter model number.S-R/W32-
Displays the serial number of transmitter.S-R/W32-
Static parameter that belongs to DISPLAY_
TB. It is incremented by one (0x0001)
when a change is made for a parameter
whose access attribute is “S-.”
The tag name of DISPLAY_TB set by
the user. It can be referenced by the host
devices and thus does not affect the
execution of Function block operation at
all.
An arbitrary group number for DISPLAY_
TB. It does not affect Function block
operation.
An identification number of the relevant
in-plant device. It does not affect Function
block operation.
Indicates the mode parameter group of
DISPLAY_TB. The configuration is as
follows:
· Target: Parameter for setting mode from
host devices.
· Actual: Indicates the current value of the
mode.
· Permitted: Indicates the mode value used
for the Function block.
· Normal: Indicates the mode value which
should be in a steady state.
Indicates the error state regarding
DISPLAY_TB.
Parameter for generating an alert when
fixed data (data whose access attribute is
“S-” or “N-”) of DISPLAY_TB is changed.
The configuration is as follows:
· Unacknowledged: Determined state
· Update_State: Change state
· Time_stamp: Change time
· Static_Revision: Revision after change
· Relative_Index: Changed parameter
identification number
Parameter that indicates the configuration
regarding DISPLAY_TB and the abnormal
state in execution. The configuration is as
follows:
· Unacknowledged: Generation determined
state
· Alarm_State: Alarm generation state
· Time_stamp: Alarm generation/recovery
time
· Subcode: Alarm content sub-code
· Value: Alarm value
The header information of DISPLAY_TB.
Users do not directly use this parameter.
Indicates the type of DISPLAY_TB.-S-R20xffff
Indicates the version of DISPLAY_TB.-N-R2-
Indicates an abnormal state specific to the
device.
Definition information of the parameter
group for efficient access of the host
devices to parameters having the same
attribute.
Sub-parameter
name
-S-R20 ≤ X ≤ 655350
-S-R/W32spaces
-S-R/W20
-S-R/W11 ≤ X ≤ 2550
TargetN-R/W1O/S, AUTO
ActualD-R1-
PermittedS-R/W1
NormalS-R/W1
-
Unacknowledged
Update StateD-R1
Time StampD-R8-
Static Revision D-R2-
Relative Index
Unacknowledged
Alarm StateD-R1
Time StampD-R8SubcodeD-R2-
Valu eD-R1-
-S-R20
-D-R119: Configration Error-
-S-R40
Access
attribute
D-R2
D-R/W1
D-R2-
D-R/W1
Size
(Byte)
bit0: Other
bit1: Block Configuration Error
bit15: Out-of-Service
(0: Undefined)
1: Acknowledged
2: Unacknowledged
0: Undefined
1: Update reported
2: Update not reported
- reported2=Clear not reported3=Active reported4=Active - not
reported
Range
Factory default
setting
bit3:AUTO
0x08
bit3:AUTO
bit7:O/S
0x88
bit3:AUTO
0x08
-
-
-
-
-
B-14
INDEX Parameter nameDescription
BLOCK_ERR_
14
DESC_1
DISPLAY_
15
PARAM_
SELECTION
DISPLAY_INFO_
16
SELECTION
17DISPLAY_CYCLE Selects a display update cycle.-S-R/W11 ≤ X ≤ 105
BLOCK_TYPE_
18
SELECTION_1
BLOCK_TAG_
19
SELECTION_1
PARAM_
20
SELECTION_1
21DISPLAY_TAG_1
UNIT_
22
SELECTION_1
CUSTOM_
23
UNIT_1
EXPONENT_
24
SELECTION_1
BLOCK_TYPE_
25
SELECTION_2
BLOCK_TAG_
26
SELECTION_2
PARAM_
27
SELECTION_2
28DISPLAY_TAG_2
UNIT_
29
SELECTION_2
CUSTOM_
30
UNIT_2
EXPONENT_
31
SELECTION_2
BLOCK_TYPE_
32
SELECTION_3
Indicates the details of BLOCK_ERR.-D-R4
Select a parameter to be displayed from
four display formats.
Selects at least one of TAG, status, and unit
of the parameter for display.
Displays the block type of the block
selected by BLOCK_TAG_SEL_1.
Inputs Block TAG of the parameter to be
displayed on Screen 1.
Selects a parameter to be displayed on
Screen 1.
Enter the name (TAG) of a parameter to
display on Screen 1.
Select a unit of the parameter to be
displayed on Screen 1.
Freely set a unit of the parameter to be
displayed on Screen 1.
Select an exponent of the parameter to be
displayed on Screen 1.
Displays the block type of the block
selected by BLOCK_TAG_SEL_2.
Inputs Block TAG of the parameter to be
displayed on Screen 2.
Selects a parameter to be displayed on
Screen 2.
Enter the name (TAG) of a parameter to
display on Screen 2.
Select a unit of the parameter to be
displayed on Screen 2.
Freely set a unit of the parameter to be
displayed on Screen 2.
Select an exponent of the parameter to be
displayed on Screen 2.
Displays the block type of the block
selected by BLOCK_TAG_SEL_3.
Sub-parameter
name
-S-R/W1
-S-R/W1
-D-R2
-S-R/W32"FLOW_TB"
-S-R/W1*2-
-S-R/W321 character ≤ X ≤ 32 characters-
-S-R/W1
-S-R/W321 character ≤ X ≤ 32 characters spaces
-S-R/W1
-D-R2*10
-S-R/W32spaces
-S-R/W1*20
-S-R/W321 character ≤ X ≤ 32 characters spaces
-S-R/W1
-S-R/W321 character ≤ X ≤ 32 characters spaces
-S-R/W1
-D-R2*10
Access
attribute
Size
(Byte)
bit0: Parameter 1 Configuration
bit1: Parameter 2 Configuration
bit2: Parameter 3 Configuration
bit3: Parameter 4 Configuration
bit4: Parameter/Information
bit0: Parameter 1
bit1: Parameter 2
bit2: Parameter 3
bit3 :Parameter 4
bit0: Tag
bit1: Unit
bit2: Status
*1
0x0000: 0x0101: Analog Input (AI)
0x0108: Proportional-Integral-
Inputs Block TAG of the parameter to be
displayed on Screen 3.
Selects a parameter to be displayed on
Screen 3.
Enter the name (TAG) of a parameter to
display on Screen 3.
Select a unit of the parameter to be
displayed on Screen 3.
Freely set a unit of the parameter to be
displayed on Screen 3.
Select an exponent of the parameter to be
displayed on Screen 3.
Displays the block type of the block
selected by BLOCK_TAG_SEL_4.
Inputs Block TAG of the parameter to be
displayed on Screen 4.
Selects a parameter to be displayed on
Screen 4.
Enter the name (TAG) of a parameter to
display on Screen 4.
Select a unit of the parameter to be
displayed on Screen 4.
Freely set a unit of the parameter to be
displayed on Screen 4.
Select an exponent of the parameter to be
displayed on Screen 4.
Deletes the operation history sent from the
digital operator panel.
Saves up to 10 data items of the time at
which LUI input mode was changed and
the mode after the change.
Saves up to 10 data items of the time at
which LUI input mode was changed and
the mode after the change.
Saves up to 10 data items of the time at
which LUI input mode was changed and
the mode after the change.
Saves up to 10 data items of the time at
which LUI input mode was changed and
the mode after the change.
Saves up to 10 data items of the time at
which LUI input mode was changed and
the mode after the change.
Saves up to 10 data items of the time at
which LUI input mode was changed and
the mode after the change.
Saves up to 10 data items of the time at
which LUI input mode was changed and
the mode after the change.
Sub-parameter
name
-S-R/W32spaces
-S-R/W1*20
-S-R/W321 character ≤ X ≤ 32 characters spaces
-S-R/W1
-S-R/W321 character ≤ X ≤ 32 characters spaces
-S-R/W1
-D-R2*10
-S-R/W32spaces
-S-R/W1*20
-S-R/W321 character ≤ X ≤ 32 characters spaces
-S-R/W1
-S-R/W321 character ≤ X ≤ 32 characters spaces
-S-R/W1
-S-R/W
DateN-R80
Valu eN-R1
DateN-R80
Valu eN-R1
DateN-R80
Valu eN-R1
DateN-R80
Valu eN-R1
DateN-R80
Valu eN-R1
DateN-R80
Valu eN-R1
DateN-R80
Valu eN-R1
Access
attribute
Size
(Byte)
0: Auto
1: Custom
0: None
1: 1
2: 2
3: 3
4: 4
5: 5
6: 6
0: Auto
1: Custom
0: None
1: 1
2: 2
3: 3
4: 4
5: 5
6: 6
0{None}
1{Erase}
0x00: Local User I/F Inactive
0x80: Local User I/F Active
0x00: Local User I/F Inactive
0x80: Local User I/F Active
0x00: Local User I/F Inactive
0x80: Local User I/F Active
0x00: Local User I/F Inactive
0x80: Local User I/F Active
0x00: Local User I/F Inactive
0x80: Local User I/F Active
0x00: Local User I/F Inactive
0x80: Local User I/F Active
0x00: Local User I/F Inactive
0x80: Local User I/F Active
Saves up to 10 data items of the time at
which LUI input mode was changed and
the mode after the change.
Saves up to 10 data items of the time at
which LUI input mode was changed and
the mode after the change.
Saves up to 10 data items of the time at
which LUI input mode was changed and
the mode after the change.
Display Transducer Block
INDEX Parameter nameDescription
Indicates the number of changes of the
1ST_REV
2TAG_DESC
3STRATEGY
4ALERT_KEY
Static parameter that belongs to DIAG_TB.
It is incremented by one (0x0001) when
a change is made for a parameter whose
access attribute is “S-.”
The tag name of DIAG_TB set by the user.
It can be referenced by the host devices
and thus does not affect the execution of
Function block operation at all.
An arbitrary group number for DIAG_TB.
It does not affect Function block operation.
An identification number of the relevant
in-plant device. It does not affect Function
block operation.
DateN-R8
Valu eN-R1
DateN-R80
Valu eN-R1
DateN-R80
Valu eN-R1
Sub-parameter
name
-S-R20–655350
-S-R/W32spaces
-S-R/W20–655350
-S-R/W11–2550
Access
attribute
0
0x00: Local User I/F Inactive
0x80: Local User I/F Active
0x00: Local User I/F Inactive
0x80: Local User I/F Active
0x00: Local User I/F Inactive
0x80: Local User I/F Active
Size
(Byte)
Range
0
0
0
Factory default
setting
5MODE_BLK
6BLOCK_ERR
7UPDATE_EVT
Indicates the mode parameter group of
Diag Transducer Block. The configuration
is as follows:
· Target: Parameter for setting mode from
host devices.
· Actual: Indicates the current value of the
mode.
· Permitted: Indicates the mode value used
for the Function block.
· Normal: Indicates the mode value which
should be in a steady state.
Indicates the error state regarding DIAG_
TB.
Parameter for generating an alert when
fixed data (data whose access attribute is
“S-”) of Diag Transducer Block is changed.
The configuration is as follows:
· Unacknowledged: Determined state
· Update_State: Change state
· Time_stamp: Change time
· Static_Revision: Revision after change
· Relative_Index: Changed parameter
identification number
TargetN-R/W1
ActualD-R1
PermittedS-R/W1
NormalS-R/W1bit3: Auto0x08bit3: Auto
-D-R2
Unacknowledged
Update StateD-R1
Time StampD-R8-
Static Revision D-R2-
Relative Index D-R2-
D-R/W1
bit3: Auto
bit7: O/S
bit3: Auto
bit4: Man
bit7: O/S
bit3: Auto
bit7: O/S
bit1: Block Configuration Error
bit7: Input Failure/Process has
BAD status
bit15: Out-of-Service
(0: Undefined)
1: Acknowledged
2: Unacknowledged
0: Undefined
1: Update reported
2: Update not reported
0x80bit7: O/S
0x80bit7: O/S
0x88bit3:
Autobit7: O/S
-
-
-
B-17
INDEX Parameter nameDescription
Parameter for generating an alert when
fixed data (data whose access attribute is
“S-”) of Diag Transducer Block is changed.
The configuration is as follows:
8BLOCK_ALM
TRANSDUCER_
9
DIRECTORY
TRANSDUCER_
10
TYPE
TRANSDUCER_
11
TYPE_VER
12XD_ERROR
COLLECTION_
13
DIRECTORY
BLOCK_ERR_
14
DESC_1
15SCALE_LEVELIndicates the scale diagnostic value.
SCALE_LEVEL_
16
RANGE
SCALE__LEVEL_
MOVING_
17
AVERAGE_
MODE
SCALE__LEVEL_
18
MOVING_
AVERAGE_TIME
SCALE__LEVEL_
19
THRESHOLD
SCALE__LEVEL_
20
HYSTERESIS
· Unacknowledged: Determined state
· Update_State: Change state
· Time_stamp: Change time
· Static_Revision: Revision after change
· Relative_Index: Changed parameter
identification number
The header information of DIAG_TB.
Users do not directly use this parameter.
Indicates the type of DIAG_TB.S-R20xffff: Other0xffff
Indicates the version of DIAG_TB.N-R20x01010x0101
Indicates an abnormal state specific to the
device.
Definition information of the parameter
group for efficient access of the host
devices to parameters having the same
attribute.
Indicates the details of BLOCK_ERR.D-R4
Range of the scale diagnostic value.
Specifies whether to perform a moving
average of the scale diagnostic value.
Specifies the moving average time for the
scale diagnostic value.
Specifies the threshold value for scale
diagnosis.
Specifies hysteresis for scale diagnosis.S-R/W40.0 ≤ X ≤ 99.00
Sub-parameter
name
Unacknowledged
Alarm StateD-R1
Time StampD-R8-
SubcodeD-R2-
Valu eD-R1-
StatusD-R1-
Valu eD-R4-
EU at 100%N-R499.099
EU at 0%N-R40.00
Units IndexN-R21615: Unitless1615
Decimal Point N-R1
Access
attribute
D-R/W1
S-R200
D-R1
S-R400
S-R/W1
S-R/W10 ≤ X ≤ 300
S-R/W40.0 ≤ X ≤ 99.099
Size
(Byte)
(0: Undefined)
1: Acknowledged
2: Unacknowledged
0: Undefined
1: Clear - reported
2: Clear - not reported
3: Active - reported
4: Active - not reported
Secondary Value.StatusParameter Flow_TB
Secondary Value.ValueParameter Flow_TB
Mass Flow TrendChartFlow_TB
BlockWINDOWAll
Basic SetupDefault setting that should be
implemented before the device
starts operation upon delivery.
General users do not need to
touch categories other than this
category in the initialization
Primary Value
Range.Units Index
Primary Value
Range.EU at 100%
Mass FlowGROUP Flow_TB
Secondary Value
Range.Units Index
Secondary Value
Range.EU_100
Density ConstantParameter Flow_TB
Damping C onstantParameter Flow_TB
Auto Zero CalibrationMethod Flow_TB
ConfigurationDetailed settingsPAG EAll
Flowmeter
Configuration
Volumetric FlowGROUP Flow_TB
Primary Value TypeParameter Flow_TB
Primary Value Range.
EU at 100%
Primary Value Range.
EU at 0%
Primary Value Range.
Units Index
Primary Value Range.
Decimal
Mass FlowGROUP Flow_TB
Secondary Value TypeParameter Flow_TB
Secondary Value
Range.EU at 100%
Secondary Value
Range.EU at 0%
Secondary Value
Range.Units Index
Secondary Value
Range.Decimal
DensityGROUP Flow_TB
Density ConstantParameter Flow_TB
Damping
Constant
Flow DirectionParameter Flow_TB
CoefficientParameter Flow_TB
Flowtube Model
Number
Flowtube Serial
Number
Flowtube SizeParameter Flow_TB
Flowtube TypeParameter Flow_TB
Flowtube FactorParameter Flow_TB
Excitation
Frequency
Dummy NumberParameter Flow_TB
MaintenanceUsed by the user service
Auto Zero CalibrationMethod Flow_TB
Manual Zero
Calibration
Gain CalibrationMethodFlow_TB
Excitation Current
Calibration
Restores Factory
default blocks
Resets transducer
block Factory
calibration
Calibration DetailsGROUP Flow_TB
Sensor Calibration
method
Sensor Calibration
Location
Sensor Calibration
Date
Sensor Calibration
Who