No part of this manual may be reproduced, stored in a retrieval system, or transmitted, in
any form or by any means, electronic, mechanical, photocopying, microfilming, recording,
or
othenwise,
without written permission from Axon
Instruments,
Tnc.
QUESTIONS? Call (415) 340-9988
COPYRIGHT
THE CIRCUITS AND INFORMATION IN THIS MANUAL ARE COPYRIGHTED AND MUST
NOT BE REPRODUCED IN ANY FORM WHATSOEVER WITHOUT WRITTEN PERMISSION
FROM AXON INSTRUMENTS. INC.
VERIFICATION
THIS INSTRUMENT IS EXTENSIVELY TESTED AND THOROUGHLY CALIBRATED
BEFORE LEAVING THE FACTORY. NEVERTHELESS. RESEARCHERS SHOULD
INDEPENDENTLY VERIFY THE BASIC ACCURACY OF THE CONTROLS USING
RESISTOR/CAPACITOR MODELS OF THEIR ELECTRODES AND CELL MEMBRANES.
DISCLAIMER
THIS EQUIPMENT IS NOT INTENDED TO BE USED AND SHOULD NOT BE USED IN
HUMAN EXPERIMENTATION OR APPLIED TO HUMANS IN ANY WAY.
iv
V
CONTENTS
Page
INTRODUCTION .....;..........A-1
FEATURES AND SPECIFICATIONS ...B-1
GLOSSARY C-1
QUICK
DETAILED
GUIDE TO
GUIDE TO
OPERATIONS D-1
OPERATIONS E-1
Audio Monitor E-2
Bath Probe : E-3
Blanking E-4
Buzz E-5
Calibration Signal .E-6
Capacitance Neutralization and Input Capacitance E-7
Primary • Secondary
Clear E-9
Command Generators E-10
Step command generator • DC current command * Pulse current command •
Extemal command inputs * Mixing commands
Current Measurement E-13
C^Comp
E-14
Enhancement * Neutralization • Transient coupling • DC coupling
Grounding and Hum E-17
Headstages E-19
The meaning of H • Which headstage to use • Capacitance neutralization range •
Headstage connectors * Tip potentials • Interchangeability • Cleaning *
Input leakage current and how to trim it to zero • Warning • DC removal *
Input resistance
Holders E-28
Vll
Page
in Use/Standby E-31
lonophoresis E-32
Ion Sensitive Electrodes - Special Considerations E-33
Voltage • Current • Round-off error and zero error
Power Supply Glitches E-42
Power Supply Voltage Selection and Fuse Changing E-43
Supply voltage • Changing the fuse
Resistance Compensation E-45
Description • Suggested use * Intracellular balancing
Ten-Turn Potentiometers E-48
Test Current ....E-49
Cell resistance
Troubleshooting E-50
References E-51
GENERAL INFORMATION F-1
Warranty F-1
RMA form F-3
Policy statement F-5
Service F-5
Comment form • F-7
Vlll
A-1
INTRODUCTION
The AX0PR0BE-1A multipurpose microelectrode amplifier contains two microelectrode
amplifiers (ME1 and ME2). These amplifiers may be independently used for Intracellular
and extracellular voltage recording with simultaneous current passing, or differentially for
recording with ion-sensitive electrodes. Many built-in features make the AX0PR0BE-1A
extremely convenient to use.
To learn how to make the most of these features, we advise first-time users of the
AX0PR0BE-1A to read this manual thoroughly and to familiarize themselves with the
instrument using resistor/capacitor models of their electrodes and cells.
We will be pleased to answer any questions regarding the theory and use of the
AX0PR0BE-1A. Any comments and suggestions on the use and design of the
AXOPROBE-1A
We would be most grateful for reprints of papers describing work performed with the
AXOPROBE-1 A. Keeping abreast of research performed helps us to design our
instruments to be of maximum usefulness to you who use them.
will
be much appreciated.
Axon Instruments, Inc.
Axon Instruments.
^
Inc
ff'
AXOPROBE-1
Multipurpose Microelectrode Amplifier
CURRENT DISPLAY UtCROELECTROOE I (ME1
t^
STEP COMMAND
9
^H \^4<ofT Imttunienii ^m^^ ^*^^^
AXOPROBE 1A
MULUPURPOSE UICRD
CLICTHOOF AMPttflfn
i
••:.*
•.**>•
A
MICROELECTRODE 2 |ME7|
0
7 0
••;:,*
^^•^'
0
viy ^wi^^ ^*^^ mu
® ifel
AUOK) UONITOn
f»>
.«.
Fl INPUT rilNPUT
^
^^i
'1 no FREO |H/) -3 ita FR£0 |Hf)
m?
d
Q
DUAL INTRACELLULAR AMPLIFrER
ION-SENSITIVE ELECTRODE AMPLIFIER
EXTRACELLULAR AMPLIFIER
lONOPHORESIS & CURRENT PASSING
The
AXOPROBE-1 A is a dual-channel microelectrode amplifier designed
and differential recording make
current passing through high-resistance electrodes
reliable
to enhance flexibility. Programmable "Buzz"
and
easy
to
operate. Command generators, digital voltmeters, lowpass filters
AXOPROBE-1
full range of microelectrode resistances, from hundreds
of kn to tens of Cft. To accomodate this range, a variety
of interchangeable headstages are available. These all
record at unity voltage gain but have different currentpassing gains (H).
Headstages also come with a choice of capacitance
neutralization ranges. The range depends on the size of
the capacitor used for neutralization. The high-frequency
noise is also proportional to this capacitor. Usually the
headstages are supplied with the smallest capacitor (L
version) but other choices are available.
The actual current in each microelectrode is mea-
sured.
That is, the measurement falls to zero if the micro-
electrode blocks even though a current command is set
up.
During current passing, up to ±30 V can be applied
across the microelectrode to enable control of current in
high-resistance electrodes.
For ion-selective electrode measurements either one
or both headstages can be ultrahigh input impedance
types.
A is designed to be used with the
VOLTAGE RECORDING
and CURRENT PASSING
Voltage recording and current passing are always performed simultaneously. Passive voltage recording is
ply achieved by switching off the current commands.
MICROELECTRODE 2 (ME2)
RESISTANCE
COMPENSATION
(lOvH MU/TURNI
CAPITANCE
NEUTRALIZATION
sim-
During current passing the voltage drop across the microelectrode can be eliminated from the electrode voltage
by setting the Resistance Compensation control (also
known as Bridge Balance control). At this setting the
microelectrode resistance can be read from the
ternatively, the microelectrode resistance can be read on
the digital panel meter by using the Test switch.
Capacitance Neutralization uses a 10-turn control for
maximum sensitivity.
To clear blocked tips, a Clear switch forces large hyperpolarizing and depolarizing currents through the electrode.
In some preparations, flicking the Clear switch
helps the microelectrode impale the
When using ultrahigh-impedance electrodes (e.g. ionselective electrodes), there can potentially be a DC error
introduced by leakage through the capacitor used for
capacitance neutralization. This is totally eliminated in
the AXOPROBE-1 A by an original circuit that removes
the DC voltage from across this capacitor, thus preventing
any leakage.
A common problem when using stimulating electrodes
is that some of the stimulus is directly coupled into the
recording microelectrode. The AXOPROBE-1 A has special circuits to blank the stimulus artifact.
cell.
dial.
Al-
COMMAND GENERATORS
Several current commands can be generated internally.
Each amplifier has a DC Current Command and a Pulse
Current Command. In addition, a Step Command can
be directed to either amplifier. The Step Command
generator uses a thumbwheel switch to set the output of
a digital-to-analog converter. Thus a high degree of precision and repeatability is achieved. Timing is set externally, indicator LEDs light whenever an internal command is activated.
All commands add linearly. Externai command
sources can be used simultaneously with the internal
command sources.
STEP COMMAND
o
DC CURRENT
- COMMAND
(10H nA TURN)
.t-OM
PULSE CURRENT
COMMAND
(100H nA/TURN)
W •^ii^
TEST
CURRENT
OFF
«
CONT
9 ME1
DESTINATION
'^
(xH nA)
ME2 Q
DIFFERENTIAL MEASUREMENT
For ion-sensitive measurements
and some current determinations
in cylindrical cells it is necessary
to measure
V^-Vj.
This signal is
provided as an output and can
^C^COMP_^^^
/ IN ENHANCE \
OUT NEUT
also be displayed on one of the
digital panel meters. A C, Compensation control enhances the
cross capacitance (C^) between
the two electrodes thereby en-
OFF
MAX
abling their responses to common-mode potentials to be
exactly matched.
For double-barrel current-passing experiments the C^
Compensation control can be used to neutralize the cross
capacitance. The coupling resistance can also be neutrallized.
DISPLAYS, OUTPUTS
and LOWPASS FILTERS
Two dedicated digital panel meters continuously display the microelectrode voltages and a third displays the
current in the selected microelectrode. The decimal point
of the current meter automatically shifts to comply with
the various headstage current gains (H).
The Vg outputs are the raw electrode voltages. The xlO
outputs contain the Offset voltage and Resistance Compensation.
coupled outputs useful for extracellular measurements.
The xlOO outputs are ultralow-noise AC-
— FILTERED OUTPUTS —
40 dB/DECADE LOWPASS
Fl INPUT
10(V,-V2
F2 INPUT
VBATH
Internally or externally generated calibration voltages
can be added to all of the outputs except Vg.
Two second-order lowpass filters have twelve select-
able -3 dB frequencies. The filters have been designed
for low noise and zero overshoot. The input to each filter
can be switched to one of six signals. Thus all signals
can be monitored on the two front-panel connectors
without shuffling cables or cluttering up the equipment
rack. The unfiltered and filtered signals are separately
available at output connectors on the rear panel.
BUZZ
The most crucial stage of an
intracellular experiment is obtaining a good impalement. A commonly used method is to press the
microelectrode tip against the
membrane and then oscillate
(BUZZ) the microelectrode
tage.
For unknown reasons (vibra-
vol-
tion of the tip and electrostatic
attraction to bound charges on the
inside membrane surface have
been postulated) this causes the
tip to penetrate. Until now, deliberately over utilized capacitance neutralization has been
used to establish the oscillation.
This produces widely variable results because the oscillation
parameters are uncontrolled. In
the AXOPROBE-1 A a revolution- i MIN
ary approach is used. The experi- \^^
menter is given complete control
of the three essential Buzz
parameters — frequency, duration and amplitude — so
that requirements for easy impalement can be optimized.
Buzz can be activated by a front-panel switch or by the
footswitches provided.
/— BUZZ
/^PARAMETERS
FREQUENCY
fkHz
MAX
(10H
mV/nA)
-3 dB FREO. (Hz)
100 200
50 \ / 500
BYPASS
w
ACTIVE
Fl OUT
:®i
(10H
mV/nA)
-3 dB FREQ. (Hz)
100 200
50 \ / SOO
10K
BYPASS
ACTIVE
F2 0UT
AUDIO MONITOR
So the experimenter can watch the preparation without
interruption, an audio monitor indicates the voltage
change following a successful cell impalement. The moment that the electrode tip first touches the solution can
also be clearly heard. The selected input (V, or Vj) determines the pitch of the monitor.
AUDIO MONITOR
VOLUME
^fl^
MIN
PHONE
MAX
SOURCE
OFF
GENERAL
A third electrode can be used extracellularly to record
the bath potential. To compensate for potential shifts
caused by changing the bath solution or temperature,
the bath potential is subtracted from the potentials recorded by the two main electrodes.
A specially constructed low-radiation transformer
eliminates the source of line-frequency noise (hum). The
incoming line voltage is filtered to remove radio-fre-
quency interference (RFI).
FURTHER INFORMATION
and ORDERING
The Specifications Sheet contains complete technical
details and ordering information. Please call the factory
for answers to any questions you may have.
Strong emphasis has been placed on quality. Precision
ten-turn potentiometers, reliable switches and
plated connectors are used throughout. Ultralow-drift
operational amplifiers are used in all critical positions
and ICs are socketted for easy maintenance. Detailed
operator's and service manuals are provided.
gold-
Axon Instruments, Inc.
1437 Rollins Road
Burlingame, CA 94010
U.S.A.
Phone (415) 340-9988
Telex: 6771237
B-1
Axon Instruments, Inc.
-^
AXOPROBE-1
HEADSTAGES
Types: HS-2 series headstages are standard.
Voltage Gain: All headstages record voltage at a unity
Headstage Current Gain (H): Select on basis of cell input resistance
(Ri„) and maximum current capacity dmax)H = x
0.0001
H = x 0.01 for Rin greater than approx. 300 Mfl.
H = x 0.1 for R,„ approx. 30-300 Mfl.
H = X 1 for Ri„ approx. 3-30 Mil.
H = x 10 for Ri„ approx. 0.3-3 Mn.
These ranges suggested for optimum performance. Considerable
overlap Is allowable.
Extracellular Electrodes: Operate with any H value (but check Ip,,,
and Input Leakage Current).
lonophoresis: Typically uses H = x 1.
'"Current Setting and Measuring Resistance (RJ: Located inside
headstage. R^, determines H. R^ is nof the input resistance.
'''input Resistance: Inversely proportional to H.
'•"'max: Maximum current that can be passed with input grounded via
indicated
Compliance
high-resistance electrodes.
'*'Noise:
Values measured at V^ output with input grounded via ReSingle-pole lowpass filter used to set measurement circuit -3 dB
bandwidth.
of Vj is non-oyershooting and so -3 dB bandwidth of v^ is equal
to measurement circuit bandwidth.
Hum (Power Line Noise): Less than 20 )xV peak-to-peak, grounded
input, input-referred.
for ion-sensitive electrodes with current passing.
sum
of cell resistance
Voltage: ± 30
(Rin) and
V.
Max voltage that can be applied across
Capacitance Neutralization adjusted so step response
A Multipurpose Microelectrode Amplifier
Note:
Numbered Hems are detailed in the TaUe.
electrode resistance
SPECIFICATIONS
gain.
(R^).
November 1986
'"10-90%
'"Capacitance Neutralization Range: Ten-turn potentiometer. L ver-
'"Case/Shield: Case and shield connector connected to Capacitance
"'input Leakage Current die,),) vs. Temperature. Temperature depen-
PREAMPS (Two Channels)
Settling Time
(fia_9o):
Two values
shown;
f,o_,o,i,
for voltage
step applied to Input via R^; fro_9o.i for current step into same R^.
Capacitance Neutralization adjusted for fastest non-overshooting
response.
Input Capacitance: Largely eliminated from step response considera-
tions by bootstrapped power supplies and Capacitance Neutralization.
See Settling Time specifications.
sion headstages have smallest range and lowest noise. M version
headstages have larger range but more noise.
Capacitance Neutralization Leakage Current: Prevented by removal
of DC potentials from neutralization capacitor
and
shield.
Removal
has 1 s or 10 s time constant.
Neutralization or to unity-gain buffered electrode potential (V,.).
dance measured near 25°C. At fixed temperature input leakage
Fl Inputs: Switch selected. 10V,,
F2 Inputs: Switch selected. 10V,,
Bypass Switch: In ACTIVE position signals are filtered. In BYPASS
INTERNAL COMMANDS
Note:
DC Current Command: One for each preamp. ±100H nA max.
Pulse Current Command: One for each preamp. ± 1000H nA max.
Step Command: Shared by preamps. Destination switch determines
EXTERNAL COMMANDS
Sensitivities: 20H nAA'
Input Impedance: 100 kft
Max. Input Vbltage: ±30 V
CALIBRATION SIGNAL
Intemal:
Extemal:
Accuracy: 1% typical.
Audio Monitor: Pitch proportional to V, or Vj. Internal speaker by-
BATH POTENTIAL SUBTRACTION
Signal recorded by bath headstage or by external amplifier Is sub-
Approx. 5-500 ms. Logarithmic potentiometer.
potentiometer.
Front-panel debounced switch, footswitches, or logic
HIGH level on rear-panel connector.
and 1^: /Acfoa/electrode currents. 10-^H mV/nA.
point accordingly. Maximum meter reading is less than headstage
current limit
be measured on I, and Ij outputs. Display selections are I, andlj.
V, meter also displays
tance between electrodes. ENHANCE position adds ME2 voltage
via a capacitor into ME2 headstage input, thus matching electrode
responses to common-mode signals. Used for Ion-sensitive electrode recording. NEUTRALIZE position subtracts ME2 voltage from
ME2 headstage input. Used for double-barrel current passing.
2K, 5K, 10K Hertz. Continuous rotation.
position signals are wideband.
Commands from all sources sum linearly.
Ten-turn potentiometers.
Ten-turn potentiometers. Activated by HIGH control signal on
PULSE GATE input or by front-panel switch.
which preamp command goes to. ± 199.9H nA max. Set on thumbwheel switch. Activated by HIGH control signal on STEP GATE
input or by front-panel switch.
by front-panel switch. Input-referred values: 10 mV on x 10
outputs, 1 mV on x 100 outputs, 10H nA on I outputs.
2 mVA' on xlO outputs, 0.2 mVA/ on xlOO outputs, 2H nAA'on
I outputs. 100 kft Input impedance.
passed when earphone plugged in.
tracted from xio outputs. Subtraction band-limited to 10 kHz. If
bath potential not measured system automatically reverts fo using
0 V as reference potential. Standard headstages work as bath
headstages if plugged into bath headstage connector.
{l,^J-
Currents exceeding current meter range can
Vi-Vj.
lOVj.
Matched to 0.01%
lOVj,
10(V,-Vj), V^,, 100V,, I,.
lOVj,
VBATH-
'^e2' lOOVj, Ij.
Activated by HIGH control signal on CAL. GATE injjut or
Proportional to applied voltage. Input-referred values:
GROUNDING
Signal ground is isolated fnjm chassis and power ground.
CONTROL INPUTS
Above 3 V accepted as logic HIGH. Below 2 V accepted as logic
LOW. Inputs protected to ±15 V.
PAIRING BRACKET (BR-1)
BR-1 bracket (optional extra) for mounting two headstages
HEADSTAGE DIMENSIONS
Case is 2.25 x 1.14 x 0.87" (57.2 x 29.0 x 22.1 mm). Mounting rod
is 4 (102)
or VB (6.3, 7.9 or 9.5). Specify required mounting rod diameter
with order. Cable length is 10 feet (3 m).
HEADSTAGE CONNECTORS
Sockets for microelectrode input, shield drive and ground output
are 0.08" (2 mm) diameter. Input socket is Teflon insulated.
CABINET DIMENSIONS
7 (177)
19"
SUPPLY REQUIREMENTS
Line VolUge: 100-125 VAC or 200-250 VAC. User selectable by an
internal switch.
Line Frequency: 50-60 Hz.
Power: 20 W.
Fuse:
Line Filter: RFI filter is included.
Line
ACCESSORIES PROVIDED
Operator's Manual
Service Manual
2 mm plugs for use with headstage
Low-capacitance test resistor for each headstage
Spare fuse
Footswitches to operate Buzz of both electrodes.
ORDERING INFORMATION
When ordering please specify:
1.
2.
3. Diameter (D) of headstage mouriting rods.
Unless you specify otherwise, the AXOPROBE-1 A will be supplied
with one HS-2 H = x0.1L and one HS-2 H = xlL headstage, each
with D = Vif. (7.9 mm). Domestic and international sales are direct
from the factory.
10%
AXOPROBE-1 As by a single group. For non-simultaneous
purchases, 10% discount applies to second and subsequent
AXOPROBE-1 As purchased by a single group within 12 months.
Discount must be requested when placing order.
WARRANTY
12 months parts and labor from date of receipt.
SERVICE
Service is available at the factory. A detailed service manual is
supplied with each AXOPROBE-1 A.
For further information call us. A factory expert will be pleased to
answer your technical and ordering inquiries.
long.
Available mounting rod diameters (D) are Vi, Vio
high,
19 (483) wide, 12.5 (317) deep. Mounts in standard
rack. Handles included.
0.5 A slow. 5 x 20 mm.
cord:
Shielded line cord is provided.
Current gain (H) and type of two headstages provided.
Current gain (H) and type of any extra headstages.
discount applies to simultaneous purchase of two or more
as
a pair.
AXON INSTRUMENTS, INC.
1429 Rollins Road Burlingame, CA 94010
U.S.A.
Phone: (415) 340-9988 / Telex: 677-1237
c-1
CAL.
CONT.
Cx COMP
EXT.
F1/F2
H
'l/'2
ME1/2
GLOSSARY
Calibration.
Continuous.
Cross capacitance compensation.
External.
Lowpass
Headstage current gain.
Current in microelectrode 1/2.
Microelectrode 1/2.
OF
filter.
FRONT PANEL ABBREVIATIONS
VBATH
Vel/Ve2
V1/V2
10V1/10V2
10(Vi-V2)
IOOV1/IOOV2
Bath electrode potential. Unity gain.
Raw electrode potential. Unity gain.
Unity-gain electrode potential. Derived by dividing
10Vi/10V2byten.
electrode potential. Includes Calibration. AC coupled.
IOV1
and IOV2.
D-1
QUICK GUIDE TO OPERATIONS
The controls and operation of the AXOPROBE-1 A are very briefly described in this section. Detailed
explanations are given in the alphabetically organized section E of this manual.
D1.
REST POSITIONS OF CONTROLS
Current meter display: l.|
Step Command: 000.0
OFF
ME1
Panel meters: V^, IN USE
Capacitance Neutralization: Counterclockwise
Resistance Compensation: Counterclockwise (zero)
Offset: Mid position (approx. 5.0)
DC Current Command: OFF
Counterclockwise (zero)
Pulse Current Command: OFF
Counterclockwise (zero)
C^ Compensation: Counterclockwise, IN
Output Calibration: OFF
Buzz frequency: Mid position
Buzz duration: Mid position
Buzz amplitude: Mid position
Audio Monitor OFF
Counterclockwise
Filters: lOV^.IOVg
10 kHz
Active
D-2
D2.
HEADSTAGES
HS-2 series headstages are standard. Two supplied with
All HS-2 headstages record voltage at unity
Available in several headstage current gains (H). Front-panel controls read directly in indicated units when
H =
x1.
All H values are powers of 10. Small H values for high-resistance cells and electrodes. Large H
values for passing large currents.
H = x10,
H = xO.OOOl for ion-sensitive electrodes.
Headstages normally supplied in L version (low-noise, low capacitance-neutralization range).
M version can be supplied to compensate large capacitances.
Red connector Microelectrode input
Gold connector Driven shield; case
Yellow connector Ground output
Compensates electrode voltage drop during current passing. Resistance (scaled by H) read on ten-
turn
dial.
If cross-capacitance compensation (0,^ COMP) is "on" and in the NEUT (neutralize)
position,
Test Current:
Applies constant current to electrode. Electrode response in mV corresponds to resistance.
Resistance Compensation setting must be zero.
then this control compensates electrode coupling resistance, scaled by H2 (H of ME2).
Offset:
Compensates ±500 mV electrode tip potential. Works an xlO output only. Use to zero electrode
voltage while extracellular.
DC Current Command:
D-3
For injection of constant current. Magnitude set on ten-turn
indicates when current injection activtrted.
Pulse Current Command:
For injection of pulsed or constant current. Magnitude set on ten-turn
CONT. position for continuous activation. EXT. GATE position for gating by logic HIGH level applied
to rear-panel PULSE GATE input LED indicates activation.
Clear
Passes large hyperpolarizing and depolarizing current to clear blocked electrodes or to impale
Display:
Digital panel meter displays either
Buzz:
Use for cell impalement. Connects oscillating voltage to microelectrode. Parameters of oscillation
set in Buzz Parameters section. Activate by front-panel pushbutton switch, or by external switch or
V.|
or
V.| — V2.
dial.
Polarity set on switch. LED
dial.
Polarity set on switch.
cell.
logic device to connect 5 V TO BUZZ input on rear panel.
D4.
MICROELECTRODE 2 (ME2)
Independent intracellular/extracellular electrometer similar to ME1. One difference is Standby switch
replacing Display switch. Standby position used when ME2 not required (disables digital panel meter and
capacitance neutralization). Second difference is that when cross compensation is used it is applied from
ME2 to
ME1,
not vice versa.
D-4
D5.
BUZZ PARAMETERS:
For optimization of headstage oscillation used to assist impalement. Frequency, Duration and Amplitude of
oscillation independently set.
D6.
FILTERED OUTPUTS
Two independent two-pole filters. One of six output signals can be selected for each filter. Twelve —3 dB
cutoff frequencies can be selected. Bypass/Active switch determines whether signal is filtered (active
position) or widet}and (filter bypassed).
D7.
C^ COMP
IN:
Enhances capacitance coupling of V2 into V.|. Speeds ME1 response to common signals so that
common signals are recorded identically by ME1 and ME2. Useful for ion-sensitive electrode
recording.
OUT: Neutralizes resistance and capacitance coupling of V2 into V.|. Useful for current passing through
one ban-el of a double-ban-el electrode.
D8.
OUTPUT CAL
Adds a 100 mV signal to current outputs, xlO and xlOO voltage outputs. Input-referred values are 10H nA,
10 mV and 1 mV respectively. CONT. position for continuous activation. EXT. GATE position for gating by
logic HIGH level applied to rear-panel CAL GATE input. Additional calibration signals separately generated
by applying a signal to the EXT. CAL SIGNAL input on the rear panel.
D9.
CURRENT DISPLAY
Meter displays average current. Decimal point automatically placed to suit H. Display is in nA for H = xl,
xO.l.xO.OI;
pAforH =
x0.0001;
^forH = x10. Display
11
or I2.
D-5
DIG.
STEP COMMAND
D/A converter generates precision command voltages. Destination switch selects either MEI or ME2 to be
target for command. Thumbwheel switch sets magnKude with 0.05% resolution. Magnitudes are scaled by
H. CONT position for continuous activation. EXT. GATE position for activation by logic HIGH level applied
to rear-panel STEP GATE input.
D11.
AUDIO MONITOR
Pitch of audible tone depends on potential of selected Input (V.| or V2). Potentiometer sets volume.
Earphone can be plugged into phone jack. If
D12.
BATH POTENTIAL
Use a low-resistance microelectrode with an HS-2 headstage to record potential of bath solution (Vgy^ji^).
This potential is sutjtracted from 10V.| and lOVj outputs. If not required, do not plug headstage into rearpanel connector. Vg^^-p^ recorded t)y outside equipment can be subtracted by connecting to EXT.
connector
D13.
INPUTS AND OUTPUTS
Located on rear-panel, but Fl OUT and F2 OUT repeated on front-panel.
F1,F2 outputs:
Filtered outputs.
so,
speaker is disabled.
Vp^^ji.
VQ
output
Raw electrode potential.
10V output:
Modified electrode potential. xlO
potential.
gain.
Includes Offset Resistance Compensation, CAL and bath
D-6
10(V^-V2) output:
Difference signal.
100V output:
AC-coupled (1 Hz) electrode potential x100
VBATH
°"*P"t=
Potential recorded by bath electrode.
I output
Electrode current.
Logic Levels:
Over 3 V accepted as logic HIGH. Below 1 V accepted as logic LOW. ±15 V safe operating range.
Step Gate input:
Logic HIGH activates Step Command.
Pulse Gate input
Logic HIGH activates Pulse Current.
gain.
Includes CAL
100 mV CAL Gate input:
Logic HIGH activates 100 mV OUTPUT CAL
EXT. CAL Signal input:
Voltage on this input converted into proportional calibration signal.
EXT. ME Command input
Voltage on this input converted into proportional current
EXT. Vgy^ji^ input:
Bath potential recorded by outside equipment subtracted from 10V outputs.
V.| Blank Gate Input:
Logic HIGH causes
Vg.|
to be sampled and
held.
Used for stimulus artifact rejection.
Buzz ME input:
Logic HIGH on this input activates Buzz.
+5V output:
Used with footswitches to generate logic HIGH level for Buzz. Protected by 150 ohm series resistor.
D-7
E-1
DETAILED GUIDE TO OPERATION
The controls and operation of the AXOPROBE-1 A are described In this section. The
topics are arranged in alphabetical order.
E-2
AUDIO MONITOR
The audio monitor is a voltage controlled oscillator (VCO) that drives a small speaker A switch is used to
select
V.|
or V2 as the control voltage (V^). As V^ varies, a so too does the pitch of the audio tone.
When
VQ
= 0 the frequency is about 2.25 kHz. This frequency drops by approximately 3 octaves as V^
decreases to-100 mV.
The Audio Monitor enables changes in the electrode potential to be recognized without having to look at the
oscilloscope or panel meters. Thus one can detect a successful cell impalement while still looking through
the microscope. There is also an abrupt change in tone when a new electrode first touches the solution.
Thus the electrode can be lowered towards the preparation very rapidly and stopped as soon as the tone
change indicates contact with the solution.
The volume control on minimum makes the tone inaudible. The volume control can be left in its usual
position and the Audio Monitor switched off by using the center position of the input-selector switch.
An earphone can be plugged into the phone jack provided. This disables the speaker
E-3
BATH PROBE
In certain experimental circumstances it is desirable to make all voltage measurements relative to a
reference point in the bathing solution rather than relative to ground. (These conditions may include
precision measurements during changes of temperature or Ion content of the saline, or cases of restricted
access from the extracellular space to the grounding point.)
All measurements are normally made relative to the system ground. However, if a unity-gain headstage is
plugged into the rear-panel Bath Headstage connector, measurements by both MEI and ME2 are made
relative to the potential recorded by this headstage. The bandwidth of the bath potential is limited to 10 kHz
t}efore it is subtracted from the potentials recorded by ME1 and ME2. The bath microelectrode cannot be
used for current passing.
If there is no unity-gain headstage plugged into the Bath Headstage connector, a reference potential from
an extemal amplifier can be subtracted by connecting It to the EXT. Vg^-p^ connector
Loading...
+ 54 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.