DR. HARVEV
UNIVERSITY
DEPARTMENT
9500 GIU^AN DRIVE
J,
KARTEN,
OF
CALIFORNIA,
OF
NEUROSCIENCES, 0608
M.D.
SAN
DIEGO
LA JOLLA, CA 92093-0608 February 1990
AXOCLAMP-2A MICROELECTRODE CLAMP
THEORY AND OPERATION
Written for Axon Instruments, Inc.
by Alan Finkel, Ph.D.
Copyright 1988, 1990 Axon Instruments, Inc.
No part of this manual may be reproduced, stored in a retrieval system, or transmitted, in any
form or by any means, electronic, mechanical, photocopying, microfilming, recording, or
otherwise, without written permission from Axon Instruments, Inc.
QUESTIONS? Call (415) 571-9400
Part Number 2500-000 REV B PriMcd in
U.S.A.
OUT i(r..;.,) • • St^f 4<i-.^^
Ill
COPYRIGHT
THE CIRCUITS AND INFORMATION IN THIS MANUAL ARE COPYRIGHTED AND
MUST NOT BE REPRODUCED IN ANY FORM WHATSOEVER WITHOUT WRITTEN
PERMISSION FROM AXON INSTRUMENTS, INC.
VERIFICATION
THIS INSTRUMENT IS EXTENSIVELY TESTED AND THOROUGHLY CALIBRATED
BEFORE LEAVING THE FACTORY. NEVERTHELESS, RESEARCHERS SHOULD
INDEPENDENTLY VERIFY THE BASIC ACCURACY OF THE CONTROLS USING
RESISTOR/CAPACITOR MODELS OF THEIR ELECTRODES AND CELL MEMBRANES.
DISCLAIMER
THIS EQUIPMENT IS NOT INTENDED TO BE USED AND SHOULD NOT BE USED IN
HUMAN EXPERIMENTATION OR APPLIED TO HUMANS IN ANY WAY.
AXOCLAMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
IV
Illustrations of the rear-panel view of the
AX0CLAMP-2A are shown on the fold-out page at the
rear of the manual.
AXOCLAMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
9500 GILMAN DR^VE '"'''^''' °^°^
LA JOLLA, CA 92093-0608
TABLE OF CONTENTS
Page
INTRODUCTION 1
FEATURES 3
FEATURES ..3
GLOSSARY 9
QUICK GUIDE TO OPERATIONS 11
DETAILED GUIDE TO OPERATIONS 15
ANTI-ALIAS FILTER 15
BATH PROBE i, 16
Bath Potential Measurement 16
Grounding 16
BLANKING 16
BRIDGE MODE 17
Description 17
Suggested Use 17
Intracellular Balancing 18
BUZZ 20
Remote Buzz 20
CALIBRATION SIGNAL 21
CAPACITANCE NEUTRALIZATION AND INPUT CAPACITANCE 21
Primary 21
Secondary 21
AXOCLAMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
High-Order Lowpass Filters For Low-Noise Recordings 51
Rise Time Of High-Order Filters 51
Note On Ultimate Rise Time ;.... 51
OUTPUT IMPEDANCE AND PROTECTION 51
PANEL METERS 51
V„(mV) 51
V2(mV) 52
I(nA) 52
PHASE • 52
POWER-SUPPLY GLITCHES 53
POWER SUPPLY VOLTAGE SELECTION & FUSE CHANGING 54
Supply Voltage 54
Changing The Fuse 54
REMOTE 55
RMP BALANCE 57
AXOCLAMP-2A THEORY & OPERATION,
COPYRIGHT
FEBRUARY
1990,
AXON
INSTRUMENTS.
INC.
Vlll
Page
SERIES RESISTANCE 57
Origin 57
Problem 57
Solutions 57
What is Uie True Membrane Potential Time Course? 58
SEVC MODE - CONTINUOUS.... 60
Important Note - Anti-Alias Filter 60
Suggested Use 60
cSEVC Compared WiUi Whole-Cell Patch Clamp 62
SEVC MODE - DISCONTINUOUS .....64
Description 64
Suggested Use 67
Important Note 70
Which SEVC to use witii a Suction Electrode 70
Minimum Sampling Rate and Maximum Gain 74
Clamp Error 74
Gain.... 74
SPACE CLAMP 75
TEN-TURN POTENTIOMETERS 75
TEVC MODE ; 75
Description 75
Suggested Use 76
Extremely Important Note - Coupling Capacitance 76
Saturation During The Capacitance Transient 79
Choosing the Microelectrode Resistances 79
TRIGGERED CLAMPING 79
TROUBLE SHOOTING 80
UNITY-GAIN RECORDING - THIRD POINT 80
AXOCLAMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
Page
IX
VIRTUAL-GROUND CURRENT MEASUREMENT
10.V„
AND I^
SPECIFICATIONS
REFERENCES
WARRANTY
RMA FORM
B-1
C-1
POLICY STATEMENT
SERVICE
D-l
COMMENT FORM
OUTPUTS
81
82
A-1
D-l
E-1
80
FRONT AND REAR PANEL -
f^ir^yf'
iZ^^f ^Iffff^^P
AXOCLAMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY
1990,
AXON INSTRUMENTS,
INC.
CThis page ia intentionally left blank)
AXOCLAMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
iNTRODUcrroN Page 1
INTRODUCTION
!
The AXOCLAMP-2A Microelectrode Clamp can be used as a dual channel microelectrode probe, or as a
microelectrode voltage clamp.
Voltage clar; ping is a powerful technique for the control of membrane potential and for the investigation of
processes alfecting membrane conductance. Voltage clamping has traditionally been performed using two
intracellular microelectrodes and the AXOCLAMP-2A can be used for this purpose.
The AXOCLAMP-2A can also be used for discontinuous single-electrode voltage clamping (dSEVC) and for
continuous single-electrode voltage clamping (cSEVC). A single-electrode voltage clamp (SEVC) is more
convenient to use than a two-electrode voltage clamp (TEVC) in very small cells and cells which cannot be
visualized. A particular advantage of a dSEVC is that the voltage drop due to current flow through the
series component of cell membrane resistance (Rg) is not clamped. In addition, for both types of SEVC
instabilities due to coupling capacitance and coupling resistance between two microelectrodes do not arise.
The disadvantages of a dSEVC compared with a TEVC are that the response speed is slower, the maximum
achievable gain is lower, and the noise in the current and voltage records is greater. The design of the
AX0CLAMP-2A reduces these disadvantages towards their theoretical minimums, thereby allowing singleelectrode voltage clamping to be performed in the many situations where conventional voltage clamping is
not suitable.
A cSEVC i.s as low in noise as a TEVC but has a severe disadvantage in that the voltage drop across the
microelectrode is clamped unless compensation is made. Since the required compensation is never
perfect, tha rSF,V(;^ y^in nnly he. psfid wh^-ij thg e|ectrode resistance is very small compared with the cell
input resistance. These favorable conditions can often be achieved by the whole-cell patch technique.
Because of the AXOCLAMP-2A's advanced design, it itself does not limit the achievable performance.
Instead, the dominant factor affecting SEVC performance is the microelectrode. Users of the
AXOCLAMP-2A in eidier of the SEVC modes should be quick to question, then adjust, the microelectrode
and its placement.
TTie AXOCLAMP-2A is a sophisticated instrument. Even experienced researchers are advised to read this
manual thoroughly and to familiarize themselves widi the instrument using model electrodes (i.e. resistors)
and cells (e.g. parallel RC) before attempting experiments with real microelectrodes and cells.
We will be pleased to answer any questions regarding the theory and use of the AX0CLAMP-2A. Any
conmients and suggestions on the use and design ofthe AX0CLAMP-2A will be much appreciated.
We would be most grateful for reprints of papers describing work performed with the AX0CLAMP-2A.
Keeping abreast of research performed helps us to design our instruments to be of maximum usefulness to
you who use them.
Axon Instruments, Inc.
AX0CLAMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
Page 2 iNTRODUcnoN
(This page is intentionally left blank)
AXOCLAMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
FEATURES
Page 3
FEATURES
The AXOCLAMP-2A is a complete microelectrode current and voltage clamp for intracdiular investigations. It
combines state-of-the-art single-electrode voltage clamping, two-electrode voltage clamping, and two complete
bridge amplifiers into one instrument. Precision command voltages, meters, filters, offsets and many other
features are built in to give you unprecedented flexibility.
4 discontinuous single-electrode voltage clamping
4 continuous single-electrode voltage clanqiing
4 two-electrode voltages clamping
4 discontinuous current clamping
4 two complete bridge amplifiers
4 high-speed headstages
4 low-noise low-hum operation
4 push-button selection of operating mode
4 computer selection of operating mode
4 two digital meters for voltage display
4 digital counter for display of sample rate
4 3-input digital meter for current display
4 separate current-measurement circuits for
each microelectrode
4 virtual-ground current measurement
VOLTAGE CLAMPING Voltage clamp with one or two microelectrodes — your choice is dictated
by the needs of your investigation; the AXOCLAMP-2A does both.
Discontinuous Single-Electrode Voltage Clamping (dSEVC) is based on
the technique of sampling the membrane potential while zero current flows
and then retaining this sampled value while current is injected into the
cell. This procedure is rapidly repeated to produce a smooth response.
Continuous Single-Electrode Voltage Clamping uses a low resistance
electrode to continuously record membrane potential and inject current.
The error caused by voltage drop across the electrode resistance can be
partially reduced by series resistance compensation. With Two-Electrode
Voltage Clamping (TEVC) one microelectrode is used to continuously
record membrane potential while the other is used to inject current. ~^
4 bath potential measurement and compensation
4 intemally generated precision command voltages
4 automatic clamping at resting membrane potential
4 offset compensation
4 rejection of stimulus artifacts
4 output bandwidth selection
4 calbration signal on outputs
4 electrode buzz
4 electrode clear
4 hands-free operation of buzz and clear
4 anti-alias filter
4 phase control
4 sampling clock synchronization
4 model cell
Gain of the voltage-clamp amplifier is quickly set on a smooth-acting
nonlinear control. The phase response of the amplifier is altered from
lead to lag by a Phase Shift potentiometer with a Center Frequency
switch to select the range.
A unique variable Anti-Alias Filter helps reduce noise towards the
theoretical minimum during dSEVC by slowing the response of the
sampling circuit to suit the sample rate and the microelectrode response.
The Sample Rate can be continuously altered from a low value of 500 Hz
to a high of
in noise and response times occurring when faster sampling rates are used.
AXOCLAMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
SO
kHz. This enables you to take advantage of the decrease
Page 4
FEATURES
The sample clocks of two AXOCLAMP-2A's can be synchronized in a
'Master-Slave' configuration. This is useful in experiments in which two
cells in the same preparations are independently voltage clamped using
dSEVC. Linking the two clocks prevents the generation of spurious
signals which would otherwise appear at harmonics of the difference in the
two clocks firequencies.
Output compliance in TEVC mode is ±30 V. This reduces the chance of
saturation while the membrane capacitance is charging after a step change
in voltage. To further minimize the chance of saturation during TEVC a
relay-switched headstage (HS-4) is available to automatically bypass the
current-sensing resistor inside the headstage. The HS-4 headstage must
therefore be used in conjunction with a virtual-ground current monitor
(VG-2).
The HS-4 headstage is recommended only when large, ultra-fast
voltage steps in big cells must be established.
Another unique control is a Resting Membrane Potential (RMP) Balance
Indicator which enables you to preset the clamp offset so that when you
switch into voltage-clamp mode the cell membrane will automatically be
clamped at its resting value, irrespective of the clamp gain.
A remarkable "BLANK" facility can be used to force the voltage clamp
system to ignore stimulus artifacts that would otherwise be picked up by
the voltage-recording circuit and result in large current artifacts which
could damage the cell under clamp.
A "Monitor" output enables the input to the sampling circuit to be
observed. It is essential to observe this signal during dSEVC to ensure
that the microelectrode voltage due to current passing has time to
adequately decay at the end of each cycle. An oscilloscope trigger signal
at the sample rate is provided for use with the Monitor signal.
The AX0CLAMP-2A allows very fast discontinuous single-electrode
voltage clamping. In a test cell (see specifications) the 10% to 90% rise
time is only 100 /ts. In a real setup the response speed is limited by the
microelectrode characteristics, but membrane potential rise times (without
overshoot) of less than 1 ms have been regularly achieved in a variety of
cell types. Two-electrode voltage clamping is much faster.
CURRENT CLAMPING Two controls for each microelectrode are devoted to clearing blocked
microelectrode tips and assisting cell penetration. One is a "Clear"
switch which can be used to force large hyperpolarizing or depolarizing
currents through the microelectrode. The other is a "BUZZ" switch
which causes the mocroelectrode voltage to oscillate. Depending on the
microelectrode and the preparation, one of these two methods will often
succeed in lowering the resistance of blocked microelectrode tips. When
used while the tip of the microelectrode is pressing against the membrane,
Buzz and Clear may also cause the microelectrode to penetrate the cell.
AXOCLAMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
FEATURES
Page 5
HEADSTAGES
Unity-voltage-gain HS-2 headstages are available in several current gains.
These cover the range of cell input impedances from less than 1 MO to
greater than 1 GO. Ultrahigh-input impedance versions are also available
for ion-sensitive electrodes.
High speed and low noise are achieved by using bootstrapped power
supplies for the input circuit of each headstage. These bootstrapped
power supplies are derived from special high-voltage circuits so that the
headstages will not be saturated by the large voltages that may occur
during the passage of ciurent through high-resistance microelectrodes.
Capacitance Neutralization is also derived from high-voltage circuits so
that fast responses are not degraded during large input signals.
Current in each microelectrode is continuously measured during both
voltage clamp and current clamp. This measurement does not include
currents from sources other than the microelectrode (e.g. hum,
ionophoresis, the other microelectrode) and indicates zero if the
microelectrode blocks.
Headstages have a gold-plated 2 mm (0.08") input socket to directly
accept standard microelectrode holders. 2 mm plugs are supplied with the
headstages to connect wire leads, if used.
COMMAND GENERATORS
In any mode, level and step commands can be generated intemally.
Level Commands (one for voltage clamp and one for each microelectrode
for a total of 3) are set on precision ten-tum potentiometers. The Step
Command is set on a 3'/i-digit thumbwheel switch and can be directed to
either one of the microelectrodes or to the voltage clamp. An indicator
light for each microelectrode illuminates during current commands.
Extemal command sources can be used simultaneously with the intemal
command sources.
OUTPUTS Two dedicated Digital Voltmeters continuously display the
microelectrode voltages while a third displays the currmt in the selected
microelectrode or in a virtual-ground circuit, if used. Front-panel
controls for each microelectrode and the virtual ground set the scaling of
the current meter to suit the gain of your headstage.
A Digital Counter lets you know precisely what sampling rate you are
using during single-electrode voltage clamp or discontinuous current
clamp.
Offset Controls are provided for each microelectrode, and a variable
Lowpass Filter is provided for the microelectrode used in single-electrode
voltage clamping. As well, an intemally generated Calibration Signal
can be superimposed onto each of the outputs. Hence, the output signals
in many cases can be wholly conditioned within the AXOCLAMP-2A to
suit your recording apparatus.
AXOCLAMP;2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
Page 6
FEATURES
Six outputs are conveniently located at the front panel for connectmg to
your oscilloscope. These outputs are repeated at the rear panel, where
the other outputs, the inputs and the headstage connectors are also
located.
REMOTE CONTROL
Hands-free operation of Buzz is possible using the footswitches supplied
with every AXOCLAMP-2A. Selection of the operating mode can be made
remotely for computer sequencing of experiments.
All AXOCLAMP-2AS have a Buzz oscillator to assist in cell penetration.
The duration of the Buzz oscillation is normally equal to the time that the
front-panel switch is pressed. Practically, the shortest duration that this
switch can be pressed is about 100 ms. For small cells, 100 ms Buzz
oscillation sometimes damages the cells immediately after penetration.
The Remote Buzz Duration Control supplied with the AX0CLAMP-2A is
a hand held control that contains a trigger switch to buzz either electrode,
and a duration control for setting the Buzz duration in the range 1-50 ms.
An appropriate duration can be fotmd for most cells that is sufficiently
long to allow penetration of the membrane but short enough that the cell is
not damaged after penetration.
MODEL CELL Every AXOCLAMP-2A is supplied with a CLAMP-1 model cell. This
model cell plugs directly into the input sockets of the headstages. A
switch allows the
—
two 50 MQ electrodes to ground, or (b) CELL mode — two electrodes
CLAMP-1
model cell to be configured as (a) BATH mode
connected to a 50 MO // 500 pF cell.
The CLAMP-1 model cell can be used to test and practice using bridge
current clamp, discontinuous current clamp, single-electrode voltage
clamp and two-electrode voltage clamp. It is a useful tool to use while
leaming the operation of the AX0CLAMP-2A and subsequently to verify
the correct operation of the AXOCLAMP-2A and the recording pathway.
GENERAL A third HS-2 headstage can be used extracellulariy to record bath
potential. The bath potential is then subtracted from the potentials
recorded by the two intracellular microelectrodes to compensate for shifts
in bath potential due to changing of solutions or temperature.
A VG-2 Virtual-Ground headstage may be used to measure total bath
current. Generally, the built-in current monitors are more useful since
they yield the microelectrode currents separately without any interfering
currents (e.g. from ionophoresis). Since both microelectrode amplifiers
are complete, one microelectrode can be used for ionophoresis while the
AXOCLAMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
FEATURES Page?
other is used intracellularly. Internally generated hum due to the built-in
power supply has been prevented by using a specially constmcted lowradiation transformer, by placing the supply well away from the rest of
the circuitry, and by using intemal shielding. The incoming power is
filtered to remove radio-frequency interference (RFI).
QUALITY The excellence of the components and constmction will be obvious to you
from the high quality of the cabinet and controls. Precision ten-tura
potentiometers and reliable switches abound. But the high qualify is more
than "skin deep' gold plated connectors ar^ used throughout, ultralowdrift operational amplifiers are used in all critical positions, I.C.s are
socketed for easy maintenance, and the circuit designs and operation have
been well tested in laboratories throughout the world. All this adds up to
low-noise, low-drift, reliable and accurate operation. And the
excellence does not stop with the hardware. We also provide a detailed
operator's manual that serves as a handbook of procedures for
microelectrode users. A separate service manual is also supplied.
FURTHER INFORMATION
AND ORDERING
The AXOCLAMP specification sheet contains complete technical details
and ordering information. Please call the factory for answers to any
questions you may have.
AXOCLAMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
Pages
FEATURES
Crhis page is intentionally left blank)
AXOCLAMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
GLOSSARY
Page 9
GLOSSARY
AXOCLAMP and AX0CLAMP-2A are used interchangeably.
Cia Total input capacitance of the headstage due mainly to the microelectrode and any
connecting cable
Cm Input capacitance of cell
cSEVC Continuous single-electrode voltage clamp
DCC Discontinuous current clamp
dSEVC Discontinuous single-electrode voltage clamp
fg Sampling rate; rate for switching from current passing to voltage recording in DCC
and dSEVC modes
G The average gain during dSEVC
GT
The instantaneous gain of the controlled current source during dSEVC
H Headstage current gain
11 Continuous current flow in microelectrode 1
12 Current flow in microelectrode 2
Im Membrane current flow
Lag High-frequency cut
Lead High-frequency boost
MEI Microelectrode 1
ME2 Microelectrode 2
Re Electrode resistance
Rg Resistance in series with membrane
RMP Resting membrane potential
Rm.Rin Input resistance of cell membrane
SEVC Single-electrode voltage clamp
TEVC Two-electrode voltage clamp
Vl Continuous voltage recorded by microelectrode 1
Vj Voltage recorded by microelectrode 2
VC Voltage Clamp
VG Virtual-ground output attenuation
Vm Membrane potential recorded by microelectrode 1
Vmon Voltage at the input of the sample-and-hold amplifier (SHI)
AX0CLAMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
Page 10 GLOSSARY
Crtiis page is intentionally left blank)
AXOCLAMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
QUICK
GUIDE TO OPERATIONS Page 11
QUICK GUIDE TO OPERATIONS
The controls and operation of die AXOCLAMP-2A are very briefly described in this section. Detailed
explanations are given in the alphabetically organized Section E of this manual.
Dl.
HEADSTAGES
(1) HS-2 Series
HS-2 series headstages are standard. Two supplied with AX0CLAMP-2A.
All HS-2 headstages record voltage at unity gain.
Z'
i .^. K [J> CC ^^^-M- j
Available in several headstage current gains (H). Front-panel controls read direcdy in indicated units
when H = xl. All H values are powers of 10. Small H values used widi high-resistance cells and
electrodes. Large H values used to pass large curretits.
H= xlO, xl, xO.l, xO.Ol for recording and clamping. H = 0.0001 for ion-sensitive electrodes.
Headstages normally supplied in L version (low-noise, low capacitance-neutralization range). M version
can be supplied to compensate large capacitance of grounded shield.
Red connector: Microelectrode input
Gold Connector: Driven shield; case
Yellow connector: Ground output
(2) HS-4 Series
Optional for current-passing electrode (ME2) in two-electrode voltage clamp. (Requires VG-2 for current
measurement.) Bypasses internal current-setting resistor during two-electrode voltage clamp so output
voltage appliad directly to electrode.
Supplied in L or M versions only.
When AX0CLAMP-2A is not in two-electrode voltage clamp mode HS-4 operates same as HS-2.
(3) VG-2 Series
Optional virtual-ground headstage measures total badi current. Not required for normal operation.
Required in two-electrode voltage clamp if HS-4 headstage used. Virtual Ground output attenuation (VG)
specifies the sensitivity. Smaller VG is more sensitive; used for low currents.
AXOCLAMP-2A
THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
Page 12
D2.
MODE GROUP
QUICK GUEIE TO OPERATIONS
Illuminated pushbuttons reconfigure AXOCLAMP-2A for different operating modes.
BRIDGE: Two conventional microelectrode amplifiers.
DCC:
Discontinuous current clamp on microelectrode 1.
SEVC: Single-electrode voltage clamp on microelectrode 1.
Discontinuous SEVC (dSEVC) uses time-sharing technique (electrode switches repetitively
from voltage recording to current-passing).
Continuous SEVC (cSEVC) is .analogous to whole-cell patch clamp (electrode
simultaneously does voltage recording and current passing).
TEVC: Two-electrode voltage clamp. Microelectrode 1 does voltage recording. Microelectrode
^2 does current passing. -——- - . - •
Cont./Discont.: Switch and lamps operate only in SEVC mode.
Deliberate overutilization of capacitance neutralization. Oscillation helps
cell penetration. Footswitches supplied as standard accessories.
Bridge:
Compensates electrode voltage drop during current passing. Resistance
(scaled by H) read on ten-tum dial. Range automatically reduced tenfold
during cSEVC.
Input Offset:
Adds ±500 mV DC to electrode voltage at early stage,
electrode voltage while extracellular.
DC Current Command:
For injection of j^nsjant current. Magnitude set on ten-tum dial.
Polarity set on switch. LED indicates when current injection activated.
Clear:
Passes large hyperpolarizing and depolarizing current to clear blocked
electrodes or help cell impalement.
Voltmeter:
Indicates membrane potential (Vm) in mV.
Use to zero
D4.
MICROELECTRODE 2 (ME2) GROUP
An independent intracellular/extracellular electrometer similar to MEI. Differences are:
Potential is labelled V2.
Output offset adds ±500 mV to electrode voltage in output stage.
AXOCLAMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
QUICK GUIDE TO OPERATIONS
D5.
VOLTAGE-CLAMP GROUP
Page 13
Gain:
Sets open-loop gain during voltage clamp. In SEVC modes output is
current source. Therefore gain is nA/mV. In TEVC mode ou^ut is
voltage source. Therefore gain is V/V.
Holding Position:
RMP Balance Lamps:
Sets holding potential during voltage clamp. Range ±200 mV.
Null during Bridge or DCC so that when activated, voltage clamp will be
at resting membrane potential.
Phase shift:
Modifies frequency response of voltage-clamp amplifier. Compensates for
nonideal phase shifts of membrane. Potentiometer adds phase advance
Oead) or phase delay Gag). Switch selects range.
Anti-Alias Filter:
D6.
STEP-COMMAND GROUP
Used in DCC or dSEVC modes to reduce noise of electrodes that have fast
and slow setding characteristics.
Uses D/A converter to generate precision command voltage.
Destination Switch:
Selects voltage clamp or either microelectrode as target for command.
Commands are mV or nA respectively.
Thumbwheel Switch:
Sets magnitude widi 0.05% resolution.
Ext./Cont./Off Switch:
Cont. position activates step command. Ext. position thumbwheel switch
is off unless logic level HIGH applied to rear-panel Step Activate input.
Off position overrides logic input.
Indication:
When destination is a microelectrode and step command is activated, lamp
in microelectrode DC Current Command Section illuminates.
D7.
RATE GROUP
Counter indicates sampling rate (cycling rate) in DCC and dSEVC modes.
Potentiometer adjusts rate from 500 Hz to 50 kHz.
AXOCLAMP-2A THEORY St OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
current outputs.
Step Activate input: Logic HIGH activates Step Command.
Blank Activate input: Logic HIGH activates Blank. During Blank,
Thus stimulus artifacts
Ext.
VC
Command input: Voltage
Ext.
MEI
Command input: Voltage
Ext. ME2 Command input: Voltage
R» Comp. input: Used
to
normally required.
VBATH
IN
input: Bath potential recorded
connected
D9.
REMOTE
Allows certain functions
to be
remotely activated
on
this input converted into voltage-clamp command.
on
this input converted into
on
this input converted into ME2 current command.
compensate voltage drop across membrane
to
this input.
are
See
service manual
by
by
computer
rejected.
MEI
current command.
for
suggested circuit.
other equipment subtracted from
or
switches. These
Clear.
DIG. CLOCK LINK-UP
Allows sampling clocks from
talk when
two
AXOCLAMP-2As
two
AX0CLAMP-2AS
in
dSEVC mode used
to be
synchronized. This eliminates electrode cross-
to
clamp
two
cells
LU-1 link-up cable.
Vm
prevented from updating.
R,
during TEVC.
Not
Vi and V2 if
are
Mode, Buzz
in
same preparation. Requires
and
AXOCI-AMP-2A
THEORY & OPERATION, COPYRIGHT FEBRUARY
1990,
AXON INSTRUMENTS,
INC.
DETAILED GUIDE TO OPERATIONS Page 15
DETAILED GUIDE TO OPERATIONS
ANTI-ALIAS FILTER
A property of all digital sampling systems is that noise in the input signal at frequencies greater than 0.5 of
die sample rate (fg) is folded down to appear as extra noise in the bandwidth from zero to 0.5 of fg (see
section on noise). This phenomenon is known as aliasing.
Aliasing can be overcome by filtering die input signal before sampling, thereby reducing die highfrequency noise content. However, this filtering procedure degrades the dynamic response of the input
signal and when used with an ideal microelectrode worsens the clamp performance.
The voltage across a real microelectrode often has a two-phase decay after the end of a current pulse,
either because of redistribution of ions in the tip, or because of the distributed nature of the capacitance
through the wall of the microelectrode (see Fig. 1). The final stages of the decay may often be so slow
that additional delay introduced by a filter us^ to prevent aliasing (an Anti-Alias Filter) causes
insignificant worsening of the dynamic response. The Anti-Alias Filter can be used by the experimenter
to trade off the noise recorded in DCC and dSEVC modes against the dynamic response. That is,
increasing the Anti-Alias Filter setting decreases the noise but can lead to instability in dSEVC and can
make it more difficult in DCC to balance the response to a current step.
The Anti-Alias Filter also has an effect in the continuous modes. It acts as a lowpass filter on the voltage
recorded by MEI. Thus the effects during TEVC and cSEVC are the same as those due to a slow voltagerecording microelectrode. Using the Anti-Alias Filter in these modes is not recommended.
Rotating the Anti-Alias Filter control clockwise logarithmically increases die amount of filtering. In the
fully counterclockwise position the filter time constant is 0.2 ;xs and the discontinuous clamp responses are
unaffected. In the fiiUy clockwise position the filter time constant is 100 /iS. There is a maximal
reduction in noise but the maximum sampling rate which can be achieved is severely limited (to about 1
kHz or less).
FIGURE 1 - TWO-PHASE MICROELECTRODE DECAY
AXOCI-AMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
Page 16 DETAILED GUIDE TO OPERATIONS
BATH PROBE
Bath Potential Measurement
In certain experimental circumstances it is desirable to make all voltage measurements relative to a
reference point in the bathing solution radier than relative to ground. (These conditions may include
precision measurements during changes of temperature or ion content of the saline, or cases of restricted
access from the extracellular space to the grounding point.)
All measurements are normally made relative to the system ground. However, if an HS-2 headstage is
plugged into the rear-panel Bath Probe connector, measurements by both MEI and ME2 are automatically
made relative to the potential recorded by this headstage. For optimum voltage-clamp performance, the
bandwidth of die bath potential is limited to 300 Hz before it is subtracted from the potentials recorded by
MEI and ME2 (see Finkel & Gage, 1985). The bath microelectrode cannot be used for current passing.
The fiill-bandwiddi voltage recorded by die badi microelectrode is available at the VBATH OUT connector.
If there is no HS-2 headstage plugged into the Bath Probe connector, a reference potential from an external
amplifier can be subtracted by connecting a reference source to the VBATH IN connector.
Grounding
It is quite uncommon to measure the bath potential. Irrespective of whether or not the bath potential is
measured, the preparation bath should be grounded by direcdy connecting it to the yellow ground
connector on the back ofthe MEI headstage (or to a virtual-ground headstage if used).
BLANKING
A common problem when using stimulating electrodes is that some of the stimulus is direcdy coupled into
the recording microelectrode. This can saturate subsequent high-gain amplifiers and die coupling
capacitors of AC circuits. The saturation effects may take tens or hundreds of milliseconds to subside.
The best way to minimize or even eliminate this artifact is at the source, by using small stimuli, isolated
stimiilators, placing an grounded shield between the stimulating electrodes and the microelectrodes, etc.
Often, though, it is not possible to reduce the artifact to manageable levels.
The AXOCLAMP-2A can circumvent the effects of the stimulus artifact by Blanking. At the moment the
logic level of the Blank Activate input goes HIGH the value of
Vm
is sampled and saved. For the duration
ofthe HIGH signal, this saved value is used instead ofthe actual potential.
In voltage-clamp modes the voltage-clamp current during the Blanking period will be held at the level
which existed at the start of the period. A small deviation from the command potential may develop
during the Blanking period as a result of comparing the command to the sampled value of
the instantaneous value of
Vm.
This deviation will only be seen when the Blanking period ends. Usually
Vm
instead of
this deviation is preferable to the situation that can occur if Blanking is not^used. If Blanking is not used
the artifact pick^ up by MEI is treated by the voltage-clamp circuit as an attempt by the cell to change its
potential. Therefore, the voltage-clamp circuit causes a current to be passed into the cell to clamp this
presumed membrane potential change. If
the
stimulus artifact is large, the consequent current artifact can
be large enough to damage the cell.
AXOCLAMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRITMENTS, INC.
DETAILED GUIDE TO OPERATIONS Page 17
The width of the Blanking period should be no longer than the minimum width required to cover the
period of the stimulus artifact. It is important not to Blank for longer than necessary since during
Blanking no updating of
minimize the artifact at the source.
BRIDGE MODE
Description
In Bridge mode the microelectrode voltages are monitored continuously, and continuous currents can be
injected down MEI or ME2.
Associated with the current flow (I) in a microelectrode is a voltage drop across the microelectrode which
depends on the product of the current and the microelectrode resistance (Re)- This unwanted IR voltage
drop adds to the recorded potential. The Bridge Balance control can be used to balance out this voltage
drop so that only membrane potential is recorded. The term "Bridge" refers to the original Wheatstone
Bridge circuit used to balance the IR voltage drop and is retained by convention even though die circuitry
has been r^laced by operational amplifier techniques.
Vm
is allowed. Even when Blanking is used, attempts should still be made to
The particular setting required to balance the Bridge is aJmeasure^Qhe-microelectrodTiresistance. j =• /\
In cSEVC mode the Bridge potentiometer compensates electrode IR voltage drop at one-tenth sensitivity.
Suggested Use
Set die Destination switch to ME 1/2 and externally trigger die Step Command generator so that ^uls^ of
current are repetitively injected into MEl/2. (Altematively, derive the command for injecting current
pulses by connecting a signal source to the Ext. ME 1/2 Command input.) Start with the Bridge Balance
control set to zero. Advance the dial until the fast voltage steps seen at die start and finish of the current
step are just eliminated. The Bridge is correcdv balanc^. The residual transient at the start and finish
of the current step is due to the finite response speed of'the ihicroelectrode. No attempt is made to
balance this transient since it covers a very brief period only and it is a useful indication of the frequency
response of the microelectrode. The transient can be minimized by correcdy setting the Capacitance
Neutralization.
The Bridge balancing procedure is illustrated in Fig. 2. The trace in A was recorded in a_model cell when
the Bridge Balance control was correctly set. In response to a positive current pulse the membrane
potential began to charge up. Before the membrane potential reached its final value die current pulse was
terminated and the membrane potential exponentially decayed to its
Hie traces in B were recorded at a sweep speed which was fast compared with the membrane time
constant, hence the membrane responses look like straight lines. The top trace shows the voltage
recorded when no Bridge Balance was used. The response was dominated by the IR voltage drop across
the electrode. In the middle trace the Bridge Balance was optimum and in the bottom trace it was slighdy
overused.
final
value.
7
When the Bridge is correctly balanced the resistance of the microelectrode can be read directly from the
dial. The sensitivity is 10 -^ H
AXOCLAMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
MO
per turn.
Page 18
DETAILED GUIDE TO OPERATIONS
The Bridge Balance controls operate on die 10.Vm output and on the
V2
output. On the 10.Vm output the
Bridge Balance control satiirates when the IR voltage drop exceeds ±600 mV referred to the input.
Intracellular Balancing
The traces in Fig 2. were all recorded with the electrode inside the cell. Since the electrode response and
the oscilloscope swe^ speed were fast compared with the membrane time constant (as in Fig. 2B), the
correct Bridge Balance setting was easy to see, even through die electrode was inside the cell.
It is sometimes usefiil to inject a brief small current pulse at the start of each oscilloscope sweep in order to
continually check the Bridge Balance setting during the course of an experiment.
Figure 2
Illustration of Bridge balancing technique. All traces were recorded from the 10.Vm output. The model
cell was 10 M0//1 nF.
R*
was 10 MO.
Recording bandwiddi: 30 kHz.
Vertical calibration: 20 mV referred to Vm-
A. Response.to.+5.nA.10.nis current pulse,. Bridge correctly balanced. Trace is membrane
response only"!
Cal. bar: 20 ms.
B.
Response to -t-5 nA 1 ms pulse.
Cal. bar:
1
ms.
Top trace: No Bridge balance used.' Fast voltagesteps at start and finish of the current
pulse are the electrode IR voltage drop.
Middle
trace:
Bridge correcdy balanced. Trace is membrane response only. Transient
electrode response remains.
Bottom trace: Bridge balance overused. Negative going step is introduced by the Bridge
Balance circuit.
AXOCLAMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
DETAILED OinDE TO OPERATIONS
Page 19
B
FIGURE 2 - BRIDGE BALANCING PROCEDURE
AXOCLAMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
Page 20
DETAILED GUIDE TO OPERATIONS
BUZZ
When the Buzz switch or the footswitch is depressed, the amount of Capacitance Neutralization is
increased. If the Capacitance Neutralization control is within a few tums of optimuni, this extra
compensation causes the headstage to go into high-frequency oscillation. If this is done while the tip of
the microelectrode is pressing against the cell membrane the oscillation will often help the microelectrode
impale the cell. The exact mechanism is unknown, but it may involve attraction between the charge at
the tip of
the
microelectrode and bound charges on the inside of
the
membrane.
To use the FS-3 footswitches, plug them into the 4 mm jacks on the back panel. The red jack labelled
"+5 V" is shared by the two footswitches. There is one violet jack for each of the two footswitches.
Precise control of the duration of Buzz can be achieved by connecting a pulse generator to pin 15 of the
Remote connector (see Remote Section). For some small cells a long duration Buzz can be deadly. In
this case it may be helpful to use an external pulse generator connected to pin 15 of the Remote connector
to gate the Buzz oscillation so that it is on for just a few milliseconds. The hand-held Remote Buzz
generator (see next page) is designed to allow you to conveniently generate Buzz durations between 1 and
50 ms.
It is difficult to interpret the operation of Buzz by watching die 10.Vm trace. This is because the xlO gain
and lowpass filter on the 10.Vm output strongly affect the amount of headstage oscillation seen. As a
quick guide, if the 10.Vm trace is unaffected dien Buzz did not succeed (so increase die Capacitance
Neutralization setting). If
the
10.Vm trace jumps then Buzz was successful.
The Buzz oscillation can be clearly observed on the Vi Cont. output.
If a grounded shield adds a lot of capacitance to ME2 the Capacitance Neutralization range may be
insufficient when an HS-2L headstage is used. In this case it will be necessary to use an HS-2M headstage
(see Headstage Section).
Ranote Buzz
Installation: Plug the Buzz control into the rear-panel ' remote' connector of the Axoclamp.
If you want to use some of the pins on the rear-panel remote connector to remotely select
the operating mode or activate the Clear currents, you will have to remove the cover from
the plug on the Remote Buzz unit and solder your inputs to the appropriate spare pins on
this plug.
Use:
Set the desired Buzz duration on the Duration control of die Remote Buzz unit. Press the
button corresponding to the electrode you want to buzz. Note that the Duration control is
shared by the two electrodes.
For Buzz durations greater dian 50 ms, use the buttons on the front panel of the
Axoclamp. Neither the buttons on the front panel of the Axoclamp nor the footswitches
use the duration set on the Remote Buzz unit.
AXOCLAMP-2A THEORY & OPERATION, COPYRIGHT FEBRUARY 1990, AXON INSTRUMENTS, INC.
Loading...
+ 76 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.