7. Information ..................................................................................................... 58
8. Appendix A SF_Capture ................................................................................ 59
9. Appendix B Updating Firmware .................................................................... 60
10. Appendix C Touchpad Alignment .............................................................. 60
1. Introduction
1.1 Applications
1.1.1 Intended Use
®
The Stat Fax
system designed to read and calculate the results of assays, which are read in 96 well
microplates or strip trays.
4200 is a compact, microprocessor-controlled, multi-purpose photometer
The Stat Fax
®
4200 may be used for in-vitro diagnostics. It is a user-programmable open
system with selectable plate formatting, alphanumeric test naming, automatic interpretation
options, duplicate well options, curve plotting and editing, flags and error messages from
which the user can create a menu of pre-programmed assays for instant recall.
Stat Fax
®
4200 is intended to be used as a stand-alone instrument without a PC but it has
a connection in the event a user wishes to download results. For a PC-controlled
microplate reader ask about the ChroMate
TM
Model 4300.
It is a general purpose instrument intended to be used by trained laboratory professionals
who are capable of selecting the appropriate features and options for each specific clinical
application. Contact your company’s instrument service provider to arrange for training if
the information provided in this manual is not sufficient for your applications.
FOR IN-VITRO DIAGNOSTIC USE
POUR L'USAGE DIAGNOSTIQUE IN VITRO
1.1.2 Summary of the Instrument
This instrument reads monochromatically (optional) or bichromatically (optional) and has
two models:
• Four-filter model (405, 450, 492, and 630 nm)
• Six-filter VIS model (405, 450, 492, 545, 600, and 630 nm)
Substitute filters in the range of 405 nm - 750 nm are available on special order models.
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 1
Summary of the Instrument (continued)
The Stat Fax® 4200 accepts all standard microplates. Trays containing microstrips may also
be read whether they are 12 wells in length, 8 wells or even partial strips.
Pre-programmed Modes - The basic calculations are permanently stored in
memory and include several single and multi-point equations. Provisions are
made in certain cases for reading specimens in duplicate and/or using the
mean reading in calculations. Each calculation mode is described in detail in
Section 2.5-Modes of Operation. The following calculation modes are
offered: Factor, Single Standard, Point to Point, Regression, Cubic Spline,
and Cut Off.
User-programmable Memory
®
The Stat Fax
store assays. Standard curves are also stored in this memory. Test protocols remain
stored until either changed or deleted by the user.
1.1.3 Principles of Operation
The Stat Fax
maintenance design which reads 8 wells simultaneously. A 96 well plate can be read and
printed in the absorbance mode in approximately twenty seconds.
Light from a lamp installed in a sealed light-tight compartment is filtered using a rotating
filter wheel. The filtered light is then divided into 8 channels using a fiber optic bundle.
4200Microplate Reader’s software allows the operator to create, edit and
®
4200 Microplate Reader’s optical system provides an economical and low
The plate carrier precisely positions each row of 8 wells into the optical path for reading.
Light passing through the 8 wells is directed to 8 photo-detectors.
The filter wheel moves to a second position to read the 8 wells at a second wavelength.
Using bichromatic differential absorbance values corrects for optical imperfections in the
plastic wells and removes the effects of meniscus and turbidity.
The photo-detectors generate electronic signals which are amplified and interpreted by the
instrument circuitry.
2 Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K
Refer to Manual
Manuel
These terms may appear on the product: Les marques sur le produit:
DANGER
Indicates an injury immediately accessible as you read this marking
WARNING
CAUTION
BIOHAZARD
BIOHAZARDS are biological agents that can cause disease in
agents.
1.2 Warning Markings Inscriptions d’avertissement
1.2.1 Safety Symbols Le Symboles de Sûreté
Symbols that may appear on the product:
Les symboles de sûreté peuvent apparaitre sur le produit:
WARNING
AVERTISSEMENT
Risk of Shock
Risque de Choc
Protective Ground
La Terre Electrique
(Earth) Terminal
Prise de Terre
1.2.2 Safety Terms Terminologie de Sûreté
These terms may appear in this manual: Les marques dans l’opérateur manuel:
DANGER Le “de marque:
DANGER”
AVERTISSEMENT! Le “de
marque: WARNING”
L’ATTENTION “Le de marque:
CAUTION”
Indique le risque immédiat de dommages (assessible tandis que vous
lisez la marque)
WARNING statements identify conditions or practices that could
result in injury or loss of life. WARNING indicates an injury hazard not
immediately accessible as you read this marking.
Ces rapports identifient les conditions ou les pratiques qui pourraient
avoir comme conséquence les dommages ou les pertes humaines.
CAUTION statements identify conditions or practices that could result
in damage to this product or other property.
Ces rapports identifient les conditions ou les pratiques qui pourraient
avoir comme conséquence les dommages a ce produit ou a toute
autre propriété.
CAUTION
L’ATTENTION
Se Rapportent a
BIOHAZARD
BIOHAZARD
Risk of Infection
Risque d’infection
humans. Lab workers handling potentially infectious materials must
use universal precautions to reduce the risk of exposure to these
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 3
1.3 Safety Precautions
below.
Refer any questions to your instrument service provider.
may void the warranty.
instrument.
Instructions
manual, or the protection provided by the instrument may be impaired.
Supply
for the country of use.
making connections to the instrument.
be free of vibrations.
clearances: 10cm around perimeter of unit, 10cm on top, 1.27cm bottom (1/2”).
Covers
Circuitry
present.
Avoid Excessive Dust
Do not operate in an area with excessive dust.
Suspected Failures
by a qualified service person.
within their acceptable limits, disregard test results.
To assure operator safety and prolong the life of your instrument, carefully follow all instructions outlined
Read Instructions Take time to read this manual carefully before using this instrument. Review
the following safety precautions to avoid injury and prevent damage to this
instrument or any products connected to it. To avoid potential hazards, use this
instrument only as specified. For best results, familiarize yourself with the
instrument and its capabilities before attempting any clinical diagnostic tests.
Servicing There are no user-serviceable parts inside the instrument. Refer servicing to
qualified service personnel. Use only factory-authorized parts. Failure to do so
Wear Protective
Apparel
Follow Operating
Use Proper Power
Observe All Terminal
Ratings
Install as Directed This instrument should be installed on a sturdy, level surface capable of safely
Provide Proper
Ventilation
Do Not Operate
Without Protective
Avoid Exposed
Do Not Operate With
Do Not Operate in
Wet/Damp Conditions
Do Not Operate In An
Explosive Atmosphere
Operating Precautions Be sure to run a sufficient number of controls in each assay. If controls are not
Many diagnostic assays utilize materials that are potential biohazards.
WARNING: Always wear protective apparel and eye protection while using this
WARNING: Do not use this instrument in a manner not specified by the
WARNING: Use only the power supply specified for this product and certified
WARNING: To avoid fire or shock hazard, observe all ratings and markings on
the instrument. Consult this manual for further ratings information before
supporting the instrument’s weight 12 lbs (5.4kg). The mounting surface should
Refer to the installation instructions for details on installing the product so it has
proper ventilation. The instrument should be surrounded by the following
WARNING: Do not operate this instrument with covers and panels removed.
WARNING: Do not touch exposed connections and components when power is
WARNING: If you suspect there is damage to this instrument, have it inspected
WARNING: Do not operate instrument in wet/damp conditions.
WARNING: Do not operate instrument in an explosive atmosphere.
Keep Instrument
Surfaces Clean and
Dry
CAUTION: Solvents such as acetone or paint thinner will damage the
instrument.
√Do not use solvents to clean the unit. Avoid abrasive cleaners; the display
overlay is liquid-resistant, but easily scratched.
√Clean the exterior of the instrument with a soft cloth using plain water. If
needed, a mild all-purpose or nonabrasive cleaner may be used.
√Use as a disinfectant a 10% solution of chlorine bleach (5.25% Sodium
Hypochlorite) or 70% isopropyl alcohol.
√ Take special care not to spill liquid inside the instrument
4 Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K
CAUTION! L’ATTENTION!
CAUTION! L’ATTENTION!
utilisé et disposer des matières infectieuses.
WARNING: If any materials are overturned during operation,
immediately set the power switch to OFF. This material should be
treated as potentially biohazardous. Appropriate cleanup and disposal
of biohazardous waste should be used.
Avertissement! Lors du fonctionnement, si on renverse des matériaux,
coupez immédiatement le courant. Placez le commutateur électrique a
AU LOIN(0). Traitez le matériel comme biohazardous, utilisant
approprie nettoient et des méthodes de disposition.
1.3.1 Disposal
Dispose of according to local regulations.
Before the instrument is removed from the laboratory for disposal or servicing, it must be
decontaminated.
Decontamination should be performed by a well-trained authorized person, observing all
necessary safety precautions. Instruments to be returned must be accompanied by a
decontamination certificate completed by the responsible laboratory manager. If a
decontamination certificate is not supplied, the returning laboratory will be responsible for
charges resulting from non-acceptance of the instrument by the servicing center or from
any authority’s intervention.
Caution: Treat all components during use and disposal as you would any
biohazardous material.
L’Attention: Utiliser et disposer des matériaux de la même manière que vous
Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K 5
1.4 Specifications
Photometric
Linear Measurement Range: ..................................... 0.0 to 4.0 Absorbance Units (A)
Interface: .................................................................... USB mouse; USB port used for thumb
drive interface for firmware updates
(mouse not included; thumb drive not
included)
6 Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K
Software
NOTE: Although it may be safe to operate in these conditions, it may not be
suitable for the performance of your tests. Check with your reagent supplier.
Speed: ........................................................................ Reads absorbance of 96 wells in about
10 seconds
Calculation Modes: ..................................................... Factor, Single Standard, Point-to-Point,
Regression, Cubic Spline, Cut Off Mode
Test Storage: .............................................................. Stores 120 tests; stores all parameters
including wavelengths, calculations, unit
codes, linear and normal ranges,
standard values, test names, and
previous standard curve.
Other
Enclosure: .................................................................. Flame-retardant ABS/PVC alloy material
cover with aluminum base
Dimensions: ............................................................... Approximately 13.5”Wx17”Dx8-1/4”H
Recommended Environmental Conditions
Indoor use
Mains supply voltage: Fluctuations not to exceed + 10% of
nominal voltage
Recommended Operating Temperature .................... 18-35°C
Recommended Operating Humidity: .......................... less than 80%
Storage Temperature: ................................................ 10 to 50°C
Accessories sold separately: Thermal paper (contact your dealer)
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 7
1.5 Installation
frequent, an Uninterruptible Power Supply (UPS) is recommended.
power cord to the power supply.
1.5.1 General
Carefully unpack the instrument, removing it from its plastic bag. Report any damage to
your freight carrier at once. Retain the original packing material for future use in the event
that the instrument is shipped to another location or returned for service. Items packed with
the instrument include:
1.5.2 Installation/Preparation
Instrument Placement
and Use
Place the instrument on a flat working surface capable of safely
supporting the weight of the instrument, approximately 5.4 kg (12lbs).
A clearance of at least 8cm (3”) around the instrument is required to
assure optimal ventilation.
Assure Clean Power
Availability
The circuit used should be substantially free of large voltage
transients (Kilovolt amp loads) such as large pumps, large
centrifuges, refrigerators and freezers, air conditioners, large
autoclaves, ovens, and dryers. The instrument may fail to operate
normally if the power supply is interrupted. If this occurs, turn the
instrument off for a moment. When the instrument is turned back on,
it will resume normal operation, but data that was not stored in
nonvolatile memory will be lost. If power fluctuations or loss are
Power Switch Position The ON/OFF switch is on the back of the instrument. The power
switch should be in the OFF position (down) before connecting the
8 Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K
Power Cord
A
Printer paper
Requirements
1.5.3 Load Paper
With the power switch in the OFF [down] position, insert the DC
connector attached to the end of the power supply module cable to
the instrument. Insert the mating end of the AC power cord to the
inlet of the power supply module, and plug the other end of the AC
power cord into an AC outlet. Use only the power cord and supply
module specified for this product and certified for the country of
use.
For 110-120 V used in the United States, use a UL listed cord set
consisting of an 18 AWG, Type SPT-1 two conductor cord
maximum 3 meters (10 feet) in length, rated 7 A, 125 V, with a
polarized parallel blade type attachment plug.
For 220-240 V used inside the United States, use a UL listed cord as
above, except rated 250V.
For other locations, use the power cord certified for the country of
use.
compartment
release button
Figure 1.5.3-1 Paper compartment lid
Locate the roll of thermal printer paper P/N
150003.
Press down on the printer paper compartment
release button (A) and the paper compartment lid
will open.
Place the paper roll in the well such that the
leading edge of the paper feeds upward from the
front of the printer as shown.
Figure 1.5.3-2 Printer paper installed
Pull up at least 1 inch of paper and then
press the compartment cover down until it
snaps closed. Turn the power switch OFF
and back ON. The wake up routine will
follow. The printer will print several lines.
Wait until it has stopped. If there is no
printing, the internal printer is disabled
(reference Section 2.3.2 Settings).
NOTE: The printer manufacturer strongly
recommends the use of thermal paper
part number 150003 to reduce lint and
extend printer life. Contact your dealer for
replacement rolls.
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 9
1.5.4 Touch Screen Description
B A C
The touch screen responds to touch pressure which causes electrical contact between the
conductive and resistive layers. The touch screen provides the following advantages:
• High touch resolution
• Pressure sensitive, works with any stylus
• Not affected by dirt, dust, water or light
• Durable technology
1.6 Parts and Controls
The following terms are used in this manual to describe parts and controls of the Stat Fax
microplate reader. More details on the operation of each feature are provided in Section 2-Operating Procedures.
A) Door open
B) Printer and printer
paper compartment
C) Display
®
4200
Figure 1.6-1 Front of instrument
A) Power supply module
B) Power switch
C) Power supply connection
D) USB Port
E) USB ports: mouse and thumb drive
F) Stylus
Figure 1.6-2 Connections
10 Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K
1.7 Checkout Procedure
Visually confirm the following items:
Power switch is set to OFF.
described.
Loading …
AWARENESS
Follow this procedure to verify that the instrument is ready for use.
Checkout Procedure
√ Power supply is connected to the unit and the power plug is inserted into an AC
outlet.
√
The instrument is now ready for power-up. Confirm that the instrument responds as
√ Set the power switch to the ON position.
The printer will print out the
instrument model, serial
number, laboratory name (if
defined under the Settings
menu Laboratory Name option;
the default is no laboratory
name), firmware version,
revision level, date and time
(Figure 1.7-2).
Click Tray Out and make sure
door opens and microplate tray
extends out.
Click Tray In and make sure
door closes.
Before running any assays,
read at least one plate in
absorbance mode to verify
plate carrier operation
Figure 1.7-1 Power up Screen Display
Stat Fax 4200 Ver: Eng-01.26.00
Laboratory Name (if defined)
Serial Number: 9999
TECHNOLOGY
INC
Stat Fax
4200
If the instrument still produces
results other than those
described here, refer to Section
4. Troubleshooting, or contact
your dealer for assistance.
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 11
12-20-2018 16:34:40
Figure 1.7-2 Representation of the power up
print out
2. Operating Procedures
1. How will absorbance readings be converted into a final result?
wavelengths)
materials of known concentrations.
4. What is the calibrator value or values?
to mean materials of unknown concentrations.
row number, and well number for each.
comparison? If so, what cutoffs or ranges are to be used for each?
the positive range will be required.
If so, results less than which cutoff should be labeled as negative?
10. Will only a partial plate be read?
2.1 Operating Precautions
• Avoid lifting, leaning or turning the instrument over when a plate is in place.
• Be sure to run a sufficient number of controls in each assay. If controls are not within
their acceptable limits or if incomplete, disregard test results.
2.2 General Selections
For every test, the instrument will require a mode selection and filter combination. Thereafter,
only those questions that pertain to the mode selected will be asked. In the Absorbance Mode, for
example, there are no further selections required.
Review the following questions before beginning a test (refer to the reagent kit insert for required
information):
2. Is a Blank required? Which filters are optimal? (primary and differential
3. How many calibrators will be used? NOTE: For the purpose of this manual, the
terms calibrator and standard are used interchangeably to designate reference
5. Will calibrators and /or specimens be read in duplicate, or singly? NOTE: For the
purpose of this manual, the terms specimen and sample are used interchangeably
6. Will locations of one or more controls be marked? If so, indicate plate number,
7. Will acceptance cutoffs or ranges for controls be entered for automatic
8. Will a cutoff value be used to label positive samples? If so, the value that begins
9. Will a cutoff value be used to label negative samples or define an equivocal zone?
12 Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K
2.3 Main Display Menu Options
The Main display menu options are:
• Run Test explained in Section 2.3.1
• Settings explained in Section 2.3.2
• Manage Tests explained in Section 2.3.3
• Utilities explained in Section 2.3.4
• Tray In – the plate carrier transports the plate into the and the door closes
• Tray Out – the door opens and the plate carrier extends out for the ease of
inserting a test plate on the carrier
Lamp status indicator Time Reprint button
Figure 2.3-1 Main display menu screen
The status line displays the Lamp status (OFF or ON) and the time. The printer button (PRN) has
two options: 1. to advance the paper, 2. to reprint the last 64 lines of data. NOTE: PRN does not
print graphs, only data.
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 13
2.3.1 Run Test
Arrows
Use the ^ UP and V DOWN arrow keys to highlight
screen
Select
The highlighted test is executed once the Select
key is pressed
By #
Allows the user to enter a specific test number to
Cancel
Returns to the main power on display screen
1 – TSH
2 – T3
3 – T3 Uptake
4 – T4
5 – T4 Total
6 – FSH
7 – hCG
8 – LH
9 – Testosterone
By #
Cancel
Select
v
<<
>>
The Run Test feature allows the operator to recall user tests that have been stored in the
instrument’s memory (see example in Figure 2.3.1-1). The Stat Fax
®
4200 stores up to
120 complete test setups in nonvolatile memory, making it easy for the user to recall
complete test configurations.
Select Run Test from the Main display. Tests programmed and saved on the instrument
will display sorted by their test number. Each of the test parameters, including the mode,
wavelengths, standards, units, and the ranges are all stored for reuse.
Blanks and standards (including entire standard curves) that have been read are also
saved. When the test is recalled, the user has the option of using the previous saved
curve or reading a new one.
Figure 2.3.1-1 Run Test - Example of stored test menu
Feature Function
selection; use the >> side arrows to advance to
the next screen; use << arrows for previous
recall.
14 Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K
2.3.2 Settings
Adjust Date and Time
Printer Setup
Lamp Control
Plate Assignment
Sound Settings
Laboratory Name
External Output
Plate Print Format
Printer Setup
Enables/disables the internal printer. Graphs can be printed when
Arrow Keys: Utilize the
^
v
Select
Done
Select the Settings option on the main display screen and the Unit Settings will display:
The functions of the Unit Settings features are:
Feature Function
Unit Settings
arrow keys to scroll up and
down the list of options or
press the option directly
from the list; the option will
highlight
Select: The high-lighted
option is executed once the
Select key is pressed
Figure 2.3.2-1 Unit Settings
Adjust Date and
Time
Provides access to the Set Time option Hours, Minutes, Seconds
as well as Month, Day, Year
the internal printer is set to OFF. Allows the contrast to be
adjusted light to dark. It is advisable to print at contrast level set at
3. Higher t levels may degrade after several lengthy printouts.
Text height options are small, medium and large.
Lamp Control Lamp idle feature will automatically turn off the lamp; the default
setting is 600 seconds (10 minutes). Lamp Warm up range is
between 45 and 240 seconds (45 seconds is the default setting).
Plate Assignment Offers plate format options such as 8 way or 12 way; duplicates
side by side.
Sound Settings Provides access to and status of touch screen sound settings (ON
or OFF), and sound volume (1=low to 8=high). Sound options are:
start up sound, key press & release and other sounds (i.e. sound
when the Quit button is selected).
Laboratory Name Allows a laboratory named to be entered; laboratory name will
print out (reference Checkout Procedure in Section 1.8).
External Output Allows external output to be turned ON or OFF. Reference the
External Output Section 2.3.2.1 for setup instructions.
Plate Print Format User can select either a 1 to 12 or H to A printout of plate data.
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 15
2.3.2.1 External Output
To enable output to be sent to a personal computer, a user must use a software
application such as SF_Capture to transmit and store data from the Stat Fax
®
4200
and have the following in place:
•Power on the instrument and set the External Output to “ON” by: (a) selecting
Settings from the main display, (b) highlight External Output on the Unit Settings
display, (c) press the Select button, (d) on the External Output Setup display
select the ON button and ensure that the Status field displays ON, (e) select the
Save button, and then, (f) select the Done button.
•Connect the type B end of the USB cable to the type B port of the
instrument.
•With the personal computer (PC) powered on, connect the type A end of the USB
cable to a USB port on the PC.
Type A
Figure 1 USB Type A port and connector
Type B
Figure 2 USB Type B port and connector
2.3.2.2 USB Mouse Connection
To use a USB mouse with the Stat Fax
®
4200, be sure that the mouse is connected
to the USB port A on the instrument before the instrument is powered on. If the
mouse is plugged in after the instrument is on, it will not be operational.
16 Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K
Create TestMulti-Test
Edit TestClone Test
Delete Test
Print Test ListExit
Create Test
2.3.3 Manage Tests
Detailed descriptions of the Modes of Operation may be found in Section 2.5.
Select Manage Tests on the Main display screen and the following options will display:
Figure 2.3.3-1 Manage Tests options
The functions of the Manage Tests options are:
Feature Function
Allows user to:
• Name the test
• Select the Mode (Absorbance, Factor, Single Standard, Point to
Point, Regression, Cubic Spline, Cut Off)
• Select the Primary Filter
• Select the Differential Filter
• Enable/Disable Blank (Yes or No) Note: Default is ‘yes’ for
Single Standard Mode.
• Set number of blank replicates
• Interpretation Criteria (Positive and Negative interpretation, or
Normal and Valid Ranges)
• Select Units
• Input Decimals
• Enable % Absorbance (Yes or No)
• Input number of Standards
• Input number of Sample Replicates
• Select Axes
• Controls management (Enable; name; number of replicates;
input low/high ranges; input action to take (i.e. warn, continue
or end test); record Lot Number; input expiration date; location).
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 17
Print Test List
Prints the list of stored tests (maximum 120 tests)
Feature Function
Create Test
(Continued)
• Set Blank Minimum and or Maximum
• Set Normal Range Low and or High
• Define Valid Range Low and or High
Edit Test Used to change test programming. Editing a test will erase any
stored blank or standard values for that test
Delete Test Allows selection and deletion of user programmed and stored
tests
Multi Test User can select a number of different tests to be read in the same
plate. (2, 3, or 6 in an 8-way formatted plate or 2 or 4 in a 12-way
formatted plate)
Clone Test Duplicates an existing test.
Exit Returns to the Main power on display screen
2.3.3.1 Multi-Test
With the Multi-Test option, any stored assay regardless of mode or wavelength may
be used. The read direction (8 or 12-way) will be the instrument was set to read in the
Settings menu. With the 8-way format, two, three or six different test modes may be
run in the same plate; with the12-way format up to four tests may be run.
If any test is selected to run in the multi-test plate that has a stored curve, the user
will be prompted to accept, print or discard the curve.
The instrument will prompt the user to select the next test. To read less than a full
plate, select “Cancel” after choosing the last test to be run.
Print Plate will enable the location and absorbance reading of each well to be printed,
as shown on the plate map.
In order to print the test information and results, after each test is run, select
“Results”.
If all tests in the multi-test plate have the same filter configuration, then the plate is
read only one time. The user can select “Print Plate” and “Results” for each test.
If different filters are used for any other test(s), the plate will automatically be read
again.
There is no option to select number of wells in Multi-Test.
18 Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K
Multi-Test (Continued)
Example 1
1) Two tests with different wavelengths have been selected to run.
2) Select “Read Plate” and the first test will run. When that test is done, use the
“Results” button to printout out data. (Print Plate can be used to verify well and
well absorbance reading location.)
3) The next test will be recalled and run automatically.
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 19
Multi-Test (Continued)
Example 2,
1) Three tests – 2, 1 and 3 - were selected to run;
2) Reading of the Standards in the second test, TSH, results in an invalid curve.
3) The next test in the plate will not run. Multi-test mode is ended.
20 Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K
2.3.4 Utilities
Change Access Level
Show Version Details
Erase ALL Tests
Recall Calibration
Print Unit Settings
Filter Voltages
Check Wheel Speed
Technical Support for pass code.
number, build date, current time.
Arrow Keys: Utilize the arrow
^
v
Select
Done
Select Utilities from the Main display screen and Utility Menu options used for diagnostic
purposes will display. The arrow keys control movement up and down the list of options as
well as using the edge of the list to scroll up and down. To select an option from the list,
press the Select button. Press the Done button when finished.
Utility Menu
keys to scroll up and down
the list of options or press the
option directly from the list;
the option will be highlighted.
Select: The high-lighted
option is executed once the
Select key is pressed.
Done: Returns to the Main
display screen
Figure 2.3.4-1 Utility Menu options
The functions of the Utility Menu options are:
Feature Function
Change Access Level Display prompts to ‘Enter Pass Code’. Contact
Show Version Details Displays current firmware version, model, serial
Erase All Tests Display will prompt “Please Confirm. Erase All Tests?”
press OK to clear all tests or Cancel.
Recall Calibration Allows user to restore the original factory settings,
including calibration.
Print Unit Settings Prints all information about an individual instrument’s
calibration settings and other information.
Filter Voltages Displays real time filter wheel voltages; provides
option to print them out. Acceptable range is from 2 to
10 v.
Check Wheel Speed Displays results as the wheel speed is tested.
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 21
2.4 General Operation
Place the plate into the
2.4.1 Bichromatic Differential Operation
The option to operate this instrument using differential absorbance readings is available for
every mode. The absorbance readings at the differential wavelength are subtracted from
the absorbance readings at the operating (primary) wavelength. Use of the bichromatic
differential absorbance values corrects for optical imperfections in the plastic wells and
removes the effects of meniscus and turbidity.
Whenever possible, differential reading is recommended because precision is significantly
improved.
In order to preserve sensitivity, it is important not to choose a differential wavelength where
the chromophore being assayed exhibits substantial absorbance. To test your
chromophore, read a darkly colored solution in the Absorbance Mode at the operating
wavelength with no differential filter, and again at the operating wavelength with the
differential filter selection. If the two absorbance readings are within 10% of each other,
then bichromatic differential reading is beneficial. If the difference between the absorbance
readings with and without a differential wavelength is greater than 25%, then the
chromophore is absorbing at or near the differential wavelength and bichromatic reading at
this wavelength is probably not desirable.
If no bichromatic wavelength is selected, exercise every measure to enhance repeatability.
2.4.2 Loading and Formatting a Plate
®
The footprint for most 96 well plates and strip trays are similar and will fit the Stat Fax
4200. Test the fit of an empty plate before beginning an assay. Place the plate into the
instrument with well A-1 at the rear right corner so that row 1 is going into the reader first.
As you press the first row back and down you will feel slight tension on the plate stretching
the carrier so that the front fits in. The plate requires a snug fit.
instrument with well A-1 at the
right rear so that row 1 is
going into the reader first.
When using a single Blank(s)
alw
ays is located in A1.
22 Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K
Loading & Formatting a Plate (Continued)
TEST DEFINITION Control 1
EnabledYES
# Replicates1
Name^
Range Lowv
Range High
ActionEdit
Done
Location
Control 1
Warn, Continue
When using a strip tray, make sure wells are pushed down into tray so that they will not
cause the plate to jam on entry. Use care that well tabs do not extend over other wells.
Do not place the tabbed ends of strips in row 1; they should be in row 12. Be sure to place
the strips in the order in which Blanks, Calibrators and Samples are to be read.
When using a Blank, it should always be located in Well A1. If replicate blanks are used,
they will be placed on the plate map as the next well from A1, to B1 if reading 8 way strips,
or below if reading 12 way strips.
For best results, do not fill wells completely; 200-250µl depending on well total volume is
the maximum fill recommended.
Use caution when attaching labels to plates so they do not jam in reader or interfere with
read path.
The microplate may be formatted for 8 way or 12 way placement of Blanks, Controls and
Samples. Selecting 8 way or 12 way is done on the Settings screen, Plate Alignment
option.
2.4.2.1 Locating Controls
Unlike Blank(s), Controls can be located in any location of the strip carrier.
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 23
The default location is the well or wells adjacent to the Blank(s). To use a different
location, while enabling the use of Controls, select the Location field (Figure 2.4.2.1-
1).
Figure 2.4.2.1-1
H G F E D C B A
NC1
2
3
4
PC PC5 Location
6A5
7
8 Select
9
10
11 Defaul t
12 Cancel
H G
F
E
D C
B A
NC PC PC1
2
3
4
5 Location
6
7
8 Select
9
10
11
Def ault
12
Cance l
H
G F
E
D C
B A
PC
PC1
2
3
4
NC5
Location
6A5
7
8 Select
9
10
11 Default
12 Cancel
NOTE: If the
A new screen will open up showing the current plate map (Figure 2.4.2.1-2). Controls
can be placed anywhere on the map by manually selecting a well location. The well
location selected will then be displayed in the box “Location”.
In the example below, a cutoff mode assay is being run and strips are to be read in
the 8 way read direction. Duplicate Positive Controls (PC) and a single Negative
Controls (NC) are being used in the assay (Figure 2.4.2.1-2).
The Negative control is to be read in well A5. This is done by selecting that box on
the screen. The new location is presented in the Location box. .
24 Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K
Figure 2.4.2.1-2 Figure 2.4.2.1-3
When using replicates, only the location of one Control is to be selected. The
replicates will follow according to the Plate Alignment chosen in Settings.
Controls or their
Replicates overlap
due to the plate
setup, an error flag
will display to warn
the operator:
“Enabled control
location already
identified!”
Figure 2.4.2.1-4
H
G
F
E
D
C
B
A
P1
S6S5S4S3S2
S1
B2B1
1
21C3C3C2C2C1
C12
109876543
3
4
5
6
7
8
9
10
11
12
QUIT
# SamplesRead Plate
2.4.3 Reading a Plate
Select and confirm a stored test, or create a new test. The display will show the plate,
indicating where Blanks, Standards and Controls are located.
The example below (Figure 2.4.3.-1) shows the layout of a Point to Point mode assay that
has duplicate Blanks, 5 Standards, using all three Controls read in duplicate.
If fewer than a full plate of wells is to be read, select the ‘#Samples’ button. Once the
plate has been read (via Read Plate), absorbance readings will be on the display for each
well.
There is an option of ‘Print Plate’ to print out the plate as seen on the display. This is
especially useful when running Absorbance mode.
Select the Results button and the test results will be printed. The message “Run another
plate of samples” will appear. The user can choose to read more or to Quit.
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 25
Figure 2.4.3-1
2.4.4 Unit of Measurement Codes (Unit Code)
Exit
Pos/Neg
Conc^
1v
1
Edit
<<SaveRunPrint
TEST DEFINITION Final page
Interpretation Mode
Units
Decimals
# Sample Replicates
Conc
To access the list of Unit Codes, select Manage Tests, select Create Test select any mode
except Absorbance, and advance to the Test Definition Final Page. Select the Units field
and the list will display. Seven units of measurement designations are stored for labeling
the concentration column plus a blank customizable selection.
Figure 2.4.4-1 Test Definition Final page
IU/mL
e
g/dL
mmol/L
umol/L
ppm
ppb
Customize
Figure 2.4.4-2 Units of Measure
Create a customized label for the concentration column by selecting the option
‘Customize’ from the list of units available. A keyboard will display, enter the desired
label name and press Enter. The custom Unit label will display in the Unit field.
26 Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K
2.4.5 Offset Absorbance
NameExit
Mode
^
Primary Filter405v
Differential Filter630
Blank NOEdit
Offset Absorbance
PrintSaveRun
TEST DEFINITION
Absorbance
If Differential Filter is set to "None", then the Offset Absorbance field is displayed.
Test Definition – Offset Absorbance field displayed
The user should determine the value of the Offset Absorbance. This value will be
subtracted from the absorbance readings to correct for the meniscus effect for the test that
is being run. Select the Offset Absorbance field and enter the user-determined offset
absorbance value. Press the Save button to save the settings.
Offset Absorbance feature is only needed when both of these conditions occur:
Monochromatic reading
No blank well
It is a mathematical blank to correct for the differential between air and liquid with a
meniscus.
To determine the user-entered value, pipette a volume of wash buffer into a well that is
equal to the total volume in test wells at the read time.
For example, if the well has 100µl of Substrate Part A per 100µl of Substrate Part B, plus
50µl of stop solution, the total volume in the test wells at the read time is 250µl. Therefore
read the absorbance of 250µl of colorless wash buffer to determine the correct value for
the Offset Absorbance. Typical values are from .250 to .400A.
2.4.6 Lamp Saver and Lamp Warmup
After a period of inactivity, there is a lamp time-out feature that will automatically turn off
the lamp. The default setting is 600 seconds (10 minutes). This can be adjusted using the
Lamp Control feature located in Settings. The range is 30 to 3600 seconds.
Lamp Warmup has a default setting of 45 seconds. This can also be adjusted from the
Lamp Control feature. The range is 45 to 240 seconds
.
Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K 27
2.5 Modes of Operation
Mode Function
Absorbance See Section 2.5.1 - Absorbance mode reads and prints the
monochromatic or bichromatic differential absorbance at the userselected wavelengths. Blanking is optional. Most Assays require a mode
other than Absorbance Mode. In this mode, no calculations are made only absorbance values are reported.
Factor See Section 2.5.2 – In Factor Mode, the endpoint absorbance readings
will be multiplied by a user-entered factor to calculate a result.
Single Standard See Section 2.5.3 – Single Standard mode reads a calibrator, and then
calculates concentrations based on a single-point standard curve
passing through the point (0,0). A blank is required to determine the
(0,0) point. A factor (equal to the concentration of calibrator ÷ the
absorbance of calibrator) is generated in this mode, and then multiplied
by subsequent absorbance readings to determine concentrations.
Point to Point See Section 2.5.4 – The plate reader accepts a number of calibrators
and calculates concentrations based on the point-to-point calibration
curve. The resulting calibrator curve is a series of lines connecting the
calibrator points, which may be entered in ascending or descending
order of absorbance. The direction of slope between the first and
second calibrators determines the direction of the curve. If the direction
of the curve changes direction, the curve will be flagged as being
“invalid” and no interpretations will be printed.
Unknown samples are calculated as follows:
The unknown sample’s absorbance is read and compared to the
absorbance of the calibrator.
An unknown Sample, with absorbance higher than the calibrator with
the highest absorbance value, is calculated using a line that passes
through the two Calibrator points with the highest absorbance value. An
unknown Sample with absorbance lower than the lowest Calibrator
absorbance is calculated from the line that passes through the two
Calibrators with the lowest absorbance values.
28 Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K
AbsorbanceCancel
Factor
Single Standard^
Point to Point
Regressionv
Cubic Spline
Cut OffSelect
SELECT MODE
CAUTION!
is applied to prevent curve overshoot.
options and Quality Control options.
Regression See Section 2.5.4- In Regression mode, the instrument accepts a
number of Calibrators and calculates concentration values based on a
best-fit curve (linear regression).
Cubic Spline See Section 2.5.4– Cubic Spline Mode accepts a minimum of 3 and a
maximum of 8 calibrators and calculates concentrations based on the
Cubic Spline (constrained) calibration curve.
Calibrator materials of known concentrations are used to calibrate the
instrument so that concentrations of unknown Samples are calculated
from the generated curve. The resulting calibrator curve is a smooth
curve connecting the Calibrator points, which may be entered in
ascending or descending order of absorbance. A constraining algorithm
Cut Off See Section 2.5.5 – Cut Off Mode provides Cut Off Value (COV) formula
Mode options are located under Manage Tests on the Main display screen. Select Create
Test and the Test Definition screen will display; select the Mode field and the list of Mode
options will display.
Each of the seven (7) modes follow the same basic pattern of options, with other variables
available for selection dependent upon the mode being used.
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 29
Available Modes
CAUTION! Verify mode selection matches with the test kit package
insert or manufacturer’s instructions before reporting test results! Always
include normal and abnormal controls.
2.5.1 Absorbance Mode
NameExit
Mode
^
Primary Filter405v
Differential Filter630
Blank NOEdit
PrintSaveRun
TEST DEFINITION
Absorbance
405Cancel
450^
492v
630Select
SELECT FILTERS
NoneCancel
405^
450
492v
630Select
SELECT FILTERS
NOTE: Every user should begin by learning to use the Absorbance Mode.
Absorbance mode will read and print sample absorbance values at user selected
wavelengths. All of the other modes that follow function the same way but use the
measured absorbance to calculate final results. Press Manage Tests, press Create Test,
and the Test Definition window will display.
Initial Test Definition display
2.5.1.1 Filter Selection
To edit the Primary or Differential Filter wavelength fields, use the arrow keys to advance
to the desired field, and then press the Edit key. The Select Filters screen will display the
available filter options; make a selection and press the Select button.
Primary Filter displayDifferential Filter display
If Differential Filter is set to "None", then the Offset Absorbance field is displayed.
30 Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K
2.5.1.2 Blank
TEST DEFINITION Blank
^
Yesv
1
Edit
Done
# Replicates
Enabled
AbsorbanceCancel
Factor
Single Standard^
Point to Point
Regressionv
Cubic Spline
Cut OffSelect
SELECT MODE
The Test Definition Blank screen will display when the Blank field is selected. This screen
allows for the Blank setting to be enabled (Yes or No); and for inputting the number of
Replicates. Press the Done button when editing is completed.
Test Definition Blank display
NOTE: All of the other modes function the same way but use the measured absorbance
to calculate final results.
2.5.2 Factor Mode
Factor Mode will read and print sample concentrations at user selected wavelengths. A
previously determined factor is entered by the operator, and the measured absorbance is
then multiplied by the factor to obtain concentration.
To operate in Factor Mode, press Manage Tests and then press Create Test. A Test
Definition window will display. Press the Mode field and the Select Mode screen will
display. Use the arrow keys to advance through the list to the desired mode and press
Select or use the stylus to highlight Factor.
The Factor Mode Test Definition first page will display the available options: Name, Mode
(Factor), Primary Filter, Differential Filter, and Blank. Enter each as done in Absorbance
Mode Section 2.5.1.
Select Mode screen
Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K 31
Factor Mode (Continued)
Exit
Factor
^
Control 1
Control 2v
Control 3Edit
<<SaveRun>>
Disabled
Disabled
Disabled
TEST DEFINITION 2nd page
10
EnabledNO
# Replicates1
Name^
Range Lowv
Range High
ActionEdit
Lot NumberDone
Expiration [mm.yyyy]
Location
TEST DEFINITION Control 1
Control 1
Warn, Co ntinue
The Test Definition 2nd page will display and allow editing of Factor and Controls. Use the
arrow keys to advance through the options, press the Edit key to change the values
displayed on the screen.
NOTE: The Factor comes from the assay package insert or from a previous assay in the
Single Standard Mode.
Factor Mode Test Definition – 2nd Page
Selecting any of the Control fields will open a Test Definition Control screen with options to
enable the Control (Yes or No), input the number of Replicates, Name the Control, input a
Low and a High Range, set the Action to Take (Warn, Continue or End Test), input a Lot
Number, input an Expiration Date, and input a Location. Press the Done button when
finished with editing fields.
32 Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K
Test Definition Control screen
Factor Mode (Continued)
Exit
Pos/Neg
Conc^
1v
1
Edit
<<SaveRunPrint
TEST DEFINITION Final page
Interpretation Mode
Units
Decimals
# Sample Replicates
Interpretation^
v
NOEdit
Done
Neg <
Reversed
TEST DEFINITION Interpretation
Pos/Neg
Pos >=
Interpretation^
Normal Range Lowv
Normal Range High
Valid Range L owEdit
Valid Range HighDone
Normal
TEST DEFINITION Interpretation
NOTE: Double
The Factor Mode Test Definition Final Page will display and allow editing of Interpretation
Mode, Units, Decimals, and number of Sample Replicates. Selecting the Units field on the
Factor Mode Test Definition – Final Page will display a list of seven measurement
designations including a customizable selection, which are stored for labeling the
concentration column (see Section 2.4.3 Unit of Measurement Codes). Use the arrow keys
to advance through the options and press the Edit key to change the values of the
selections on the screen. Press Save to save the selections.
Factor Mode Test Definition - Final Page
Selecting the Interpretation Mode field will open the Test Definition Interpretation screen for
editing the Interpretation controls. Use the arrow keys to advance through the options,
press the Edit key to change the value of the selection.
check that the
interpretation
information you
Factor Mode Test Definition –Pos/Neg Interpretation
Editing the Interpretation Pos/Neg field will display the Interpretation Normal option for
inputting Normal Range Low, Normal Range High, Valid Range Low, or Valid Range High.
Press the Done button when editing is completed.
Factor Mode Test Definition –Normal Interpretation
Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K 33
2.5.3 Single Standard Mode
Absorbance
Cancel
Factor
Single Standard
^
Point to Point
Regressionv
Cubic Spline
Cut OffSelect
SELECT MODE
NameExit
Mode
^
Primary Filter
405v
Differential Filter630
Blank YES
Edit
PrintSave
>>
TEST DEFINITION
Single Standard
Exit
Standard C onc
# Std Replicates1^
Control 1
Control 2v
Control 3Edit
<<SaveRun>>
TEST DEFINITION 2nd page
10
Disabled
Disabled
Disabled
To operate in Single Standard Mode, select Manage Tests and then select Create Test.
The Test Definition screen will display. Select the mode field and the Select Mode screen
will display the list of modes. Use the arrow keys to scroll through the mode selections to
the desired selection and press Select, or use the stylus to highlight Single Standard.
Select Mode screen
The Single Standard Mode Test Definition first page will display the available options:
Name, Mode (Single Standard), Primary Filter, Differential Filter, and Blank. These
operate the same as in the Absorbance Mode except that the Blank is required, therefore
‘YES’ will automatically appear in the Blank enabled field.
The Single Standard Mode Test Definition 2nd page will display and allow editing of
Standard Concentration, Controls, and number of Standard Replicates. Use the arrow
keys to advance through the options and press the Edit key to change the values of the
selections on the screen.
34 Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K
Single Standard Mode Test Definition – 1
st
Page
Single Standard Mode Test Definition –2nd Page
Single Standard Mode (Continued)
EnabledNO
# Replicates1
Name^
Range Lowv
Range High
ActionEdit
Lot NumberDone
Expiration [mm.yyyy]
Location
TEST DEFINITION Control 1
Control 1
Warn, Co ntinue
Exit
Pos/Neg
Conc^
1v
1
Edit
<<SaveRunPrint
TEST DEFINITION Final page
Interpretation Mode
Units
Decimals
# Sample Replicates
Selecting any of the Control fields will open a Test Definition Control screen with options to
enable the Control (Yes or No), input the number of Replicates, Name the Control, input a
Low and a High Range, set the Action to Take (Warn, Continue or End Test), input a Lot
Number, input an Expiration Date, and input a Location. Press the Done button when
finished with editing fields.
When Controls are read in duplicate, the mean of the duplicates can be calculated and that
one value used, or, each Control can be required to fall within the ranges individually. The
instrument will prompt ‘Check Mean?’ once duplicate Controls are selected.
When setting the Controls, if there are Replicates, the Location will be the starting location.
If the Controls or their Replicates overlap due to the plate setup, an error flag will display
while running the test to warn the operator.
The Single Standard Mode Test Definition Final page will display and allow editing of
Interpretation Mode, Units, Decimals, and number of Sample Replicates. Use the arrow
keys to advance through the options; press the Edit button to change the values of the
selections on the screen. Press Save to save the selections.
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 35
Test Definition Control screen
Single Standard Mode Test Definition - Final Page
Single Standard Mode (Continued)
Interpretation
^
v
NOEdit
Done
TEST DEFINITION Interpretation
Pos/Neg
Pos >=
Neg <
Reversed
NOTE: Double check
MULTI-POINT
Selecting the Interpretation Mode field will open the Test Definition Interpretation screen for
editing Interpretation controls. Use the arrow keys to advance through the options, press
the Edit key to change the value of the selection. Press the Done button when editing is
complete.
that the interpretation
information entered is
correct.
Single Standard Mode Test Definition - Interpretation
2.5.4 Multi-Point Modes (Point to Point, Regression and Cubic Spline)
®
The Stat Fax
and Cubic Spline.
In Point to Point and Cubic Spline modes, the standard concentrations must be read in
either increasing or decreasing order.
4200 offers three varieties of multi-point modes: Point-to-Point, Regression
All of the Test Definition screens, use of Controls, and items related to Interpretations work
the same for each type of multi-point. Use the arrow keys to scroll through the mode
selections to the desired selection and press Select or use the stylus to highlight
selections.
MODES
Select Mode screen
36 Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K
Multi-Point Modes (Point to Point, Regression and Cubic Spline)
NameExit
Mode
^
Primary Filter405v
Differential Filter630
Blank NOEdit
Offset Absorbance0.000
PrintSaveRun
TEST DEFINITION
Point to Point
% AbsorbanceNOExit
Standards
5
# Std Replicates1^
Axes
Control 1 Disabledv
Control 2
Disabled
Control 3 Disabled
Edit
<<SaveRun>>
Y=Abs X=Conc
TEST DEFINITION 2nd page
(Continued)
The Test Definition first page displays the available options.
Test Definition – 1
If Differential Filter is set to "None", then the Offset Absorbance field is displayed
The Test Definition 2nd page includes use of % Absorbance, number of Standards,
number of Standard replicates, selection of Axes, and the use of Controls.
st
page with Offset Absorbance field displayed
Note: Point-to-Point and Regression require at least two Standards, Cubic Spline mode
requires at least three.
Test Definition –2nd Page
Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K 37
The % Absorbance option is used with multipoint uptake assays.
% Absorbance assigns a value of 100% to the first Standard, which must be entered from
darkest to lightest.
An additional calculation (%A/A0) is calculated, which is the absorbance of the sample
divided by the absorbance of the first calibrator, or “percent of first calibrator”. The first
calibrator is considered 100%, and all subsequent samples are calculated as percent of
calibrator.
The %Abs value is displayed in the Interpretation field of the report. This mode does not
support ln, logit or log scales for the absorbance axis. Upon recall of a % Absorbance
mode test, the user will be asked whether to reread the highest value Standard.
Select the Standards field and the Standards Test Definition screen will open.
Standards Test Definition screen
38 Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K
2.5.4.2 – Editing a Curve
An “X” will appear in the
When Duplicate Standards are used, the curve may be edited by deleting one of any pair
of standards. For Point to Point and Regression Modes, the minimum is two; for Cubic
Spline Mode, the minimum is three.
Once the curve has been read, select the ‘Edit’ option before the ‘Accept’ or ‘Print’
options. Using the stylus, select the duplicate standard to be deleted
.
location on the display
that is to be deleted.
The curve will be
recalculated and the
options to Accept, Edit,
Print or Quit will be
available.
Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K 39
Example of editing a curve with Duplicate Standards
As shown in the example below, Duplicate Standards are labeled as ‘Dupl I’ and ‘Dupl II’.
2.5.4.3 – Select Axes
SELECT AXES
Cancel
Y=Ln(1000*Abs) X=Conc^
v
Y=Ln(1000*Abs) X=Ln(Conc)Select
Y=Logit(Abs) X=Log(Conc)
Y=Abs X=Ln(Conc)
Y=Abs X=Conc
concentration data (x) are linear.
Axes selections include linear-linear, ln (= natural log)-linear, linear-ln, or ln-ln
calculations. A logit-log calculation is also available. Absorbance, or ln of (1000 *
absorbance), is always on the “Y” axis. Concentration, or ln of concentration, is always on
the “X” axis.
Calculation Explanation
“Y= ABS, X=CONC” Both the absorbance data (y) and the
“Y=Ln (1000*ABS), X=Conc” The natural log of the absorbance is plotted
against the concentration. The absorbance
values are multiplied by 1000 before taking the
logs.
“Y=ABS, X=Ln(Conc)” Absorbance is plotted against the natural log of
the concentration.
“Y= Ln (1000*ABS), X=Ln(Conc)” The natural log of the absorbance is plotted
against the natural log of the concentration.
“Y= Logit(Abs), X= Log(Conc)” Select to calculate unknowns using the
Refer carefully to the Reagent Kit product insert for correct selection.
NOTE: Remember log of 0 is undefined so do not set concentration to 0 (zero) on a
logarithmic scale.
Point-to-Point and Regression modes require at least two standards, Cubic Spline
requires at least three standards.
40 Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K
Select Axes (Continued)
SELECT AXES
Cancel
Y=Ln(1000*Abs) X=Conc
^
v
Y=Ln(1000*Abs) X=Ln(Conc)Select
Y=Logit(Abs) X=Log(Conc)
Y=Abs X=Ln(Conc)
Y=Abs X=Conc
Log / logit
The Log / Logit option is available under the Point to Point Mode, Regression Mode,
and the Cubic Spline Mode. Concentrations must be greater than zero in this axes
mode. The program automatically increases the number of Standards by one.
option –
the program
automatically
increases the
number of
Standards by
one.
When using a non-linear axis:
If absorbance of a Sample is > than the highest calibrator's absorbance, “ ****** “ is
reported for the Concentration
• Standards must be read in descending absorbance order
• If no Standard programmed has a 0.0 concentration assigned to it, an additional
Standard is forced, and the concentration for that Standard is replaced with a
series of stars [ **** ]
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 41
2.5.4.4 – Controls
EnabledNO
# Replicates1
Name^
Range Lowv
Range High
ActionEdit
Lot NumberDone
Expiration [mm.yyyy]
Location
TEST DEFINITION Control 1
Control 1
Warn, Co ntinue
Exit
Pos/Neg
Conc^
1v
1
Edit
<<SaveRunPrint
Decimals
# Sample Replicates
TEST DEFINITION Final page
Interpretation Mode
Units
Interpretation Pos/Neg^
v
NOEdit
Done
TEST DEFINITION Interpretation
Pos >=
Neg <
Reversed
NOTE: Double check
entered is correct.
Selecting any of the Control fields will open a Test Definition Control screen with
options to enable the Control (Yes or No), input the number of Replicates, Name the
Control, input a Low and a High Range, set the Action to Take (Warn, Continue or
End Test), input a Lot Number, and input an Expiration Date. Press the Done button
when finished with editing fields.
When Controls are read in duplicate, the mean of the duplicates can be calculated
and that one value used, or, each Control can be required to fall within the ranges
individually. The instrument will prompt ‘Check Mean?’ once duplicate Controls are
selected.
Test Definition Control screen
The Test Definition Final page will display Interpretation Mode, Units, Decimals, and
number of Sample Replicates
.
Test Definition – Final Page
Selecting the Interpretation Mode field will open the Test Definition Interpretation screen
for editing the Interpretation controls
.
that the interpretation
information you
42 Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K
Test Definition –Pos/Neg Interpretation
Controls (Continued)
Interpretation^
v
Normal Range High
Valid Range Lo wEdit
Valid Range HighDone
TEST DEFINITION Interpretation
Normal
AbsorbanceCancel
Factor
Single Standard^
Point to Point
Regressionv
Cubic Spline
Cut OffSelect
SELECT MODE
Editing the Interpretation Pos/Neg field will display the Interpretation Normal option
for inputting Normal Range Low, Normal Range High, Valid Range Low, or Valid
Range High.
Press the Done button when editing is completed.
Test Definition –Normal Interpretation
2.5.5 Cut Off Mode
Many qualitative assays interpret results based on a pre-determined, or cut off, absorbance
such that a sample absorbance greater than the cut off absorbance is interpreted as
positive (or negative) and samples with absorbance less than the cut off be interpreted as
negative (or positive). Companies that manufacture these assays use a variety of means of
obtaining (reading or calculating) the essential cut off point. Stat Fax
universal equation that can be adapted to fit most commercially available kits by inserting
user-entered factors ‘X’ and/or ‘Y’ as well as a coefficient ‘F’. The equation is (X)(mNC) +
(Y)(mPC)+F = COV (Cut Off Value).
In the Cut Off Mode, ‘X ‘is multiplied by the mean of the negative controls. ‘Y’ is multiplied
by the mean of the positive controls, and each of these two numbers is added to ‘F’ to
arrive at the Cut Off Value (COV). The ‘X’, ‘Y’ and ‘F’ factors can be positive or negative
numbers, one or zero, or decimals such as 0.5 to fit the test requirements provided in the
Kit package insert or as instructed by the Kit manufacturer. In this mode be careful to enter
all QC parameters completely and correctly. They are usually provided as Absorbance
limits.
To operate in Cut Off Mode, select Manage Tests and then select Create Test. The Test
Definition screen will display. Select the Mode field and the Select Mode screen will display
the list of modes. Use the arrow keys to scroll through the mode selections to the desired
selection and press Select or use the stylus to highlight Cut Off.
®
4200 employs a
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 43
Cut Off Mode (Continued)
NameExit
Mode
^
Primary Filter405v
Differential Filter630
Blank NOEdit
PrintSave>>
TEST DEFINITION
Cut Off
COV=
Exit
X=1
Y=1^
F=0v
Neg Ctrl Enabled
Pos Ctrl EnabledEdit
QC Criteria
<<SaveRun>>
TEST DEFINITION Cutoff
X*mNC+Y*mPC+F
COV, mNC, mPC
The Cut Off Mode Test Definition first page displays the available options: Name, Mode
(Cut Off), Primary Filter, Differential Filter, and Blank. Enter wavelength and blank as
shown in Section 2.5.1 Absorbance Mode.
Sample of Cut Off Mode Test Definition 1
The Test Definition Cutoff 2nd page allows editing of Cut Off Value (COV), values of X, Y,
and F; Negative Control; Positive Control; and QC Criteria.
st
page
44 Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K
Sample of Cut Off Mode Test Definition 2nd page
Cut Off Mode (Continued)
EnabledYES
# Replicates1
NameNeg Ctrl^
Range Low0.200v
Range High0.400
ActionEnd TestEdit
Lot Number1222Done
Expiration [mm.yyyy]12.2013
TEST DEFINITION Neg Ctrl
When Cutoff Controls are read in multiples, the mean of the replicates
Controls are selected.
mean
Selecting any of the Control fields will open a Test Definition Control screen with options to
enable the Control (Yes or No), input the number of Replicates (1 to 5), Name the Control,
input a Low and a High Range, set the Action to Take (Warn, Continue or End Test), input
a Lot Number, input an Expiration Date, and input a Location. Press the Done button when
finished editing fields.
Test Definition Control Screen
can be calculated and that value used, or each Control can be required
to fall within the ranges set. “Check Mean” will display once replicate
Example - Cut Off Mode output showing mean
Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K 45
Cut Off Mode (Continued)
to determine the COV.
See explanation
Select COV Formula
X*mNC+Y*mPC+FExit
^
v
Select
X*mNC+F
y*mPC+F
X*mCC+F
Set the Controls by selecting the appropriate control field.
Note: When setting the Controls, if there are Replicates, the Location will be the starting
location. If the Controls or their Replicates overlap due to the plate setup, an error flag will
display while running the test to warn the operator. (Reference Section 2.4.2.1)
Selecting the ‘COV =’ field will open the Select COV Formula screen. In addition to the
universal equation, Stat Fax
Where COV is the Cutoff Absorbance Value, mNC is the mean of the Negative Controls;
mPC is the mean of the Positive Controls. X, Y, and F are user-entered coefficients that
can have any positive or negative numerical value including zero and 1.
of choices in
the table below.
Highlight the desired formula from the list and press the Select button.
®
4200 offers three more optional formulas.
COV Formula options
COV Formula Explanation
X*mean(NC) + Y*mean(PC) + F Universal equation, used to calculate the cutoff
value based upon readings or positive and/or
negative controls.
• In this equation, NC and PC are used to determine
the COV (cutoff value).
• Enter the variables provided in the package insert.
The software calculates the mean of the positives
and the mean of the negatives.
X * mean(NC) + F
• Use this equation if only a negative control is used
46 Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K
Cut Off Mode (Continued)
and ranges.
QC criteria.
COV Formula Explanation
Y * mean(PC) +F
X*mean(CC) + F
• Use this equation if only a positive control is used
to determine the COV.
• For interpretation, you may choose either the
Regular Cut Off Mode (positive >= cutoff, negative
< cutoff), or the Reverse Cut Off Mode. With the
Reverse Cut Off option, the Samples with values
lower than the cut off are labeled as “positive”. If
you choose this option, be careful to follow the <
and > signs in the prompts for entering cut offs
• COV = cut off value for positive or negative
interpretation
• X = a variable provided in the Assay Kit package
insert. NOTE: X will equal 1 if no value of variable
provided.
• mCC = the calculated mean of the absorbance
values of the cutoff controls when test is run
• Blanking is optional, dependent upon the Assay
Kit
• F = factor added to the mCC. If needed, this value
is provided in the Assay Kit package insert. If none
provided, use zero for the value of F.
• Positive and/or negative controls may be used for
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 47
Cut Off Mode (Continued)
Exit
Pos/Neg
Abs/COV^
3v
1
Edit
<<SaveRunPrint
# Sample Replicates
TEST DEFINITION Final page
Interpretation Mode
Units
Decimals
(mPC / mNC):
(mPC – mNC): Enter
Note: ‘Reversed’ field is ‘YES’
Positive and High Absorbances
are Negative.
Backup to be used in case
too low.
TEST DEFINITION Cutoff Quality
^
Minimum COVv
Maximum COV
Edit
(mPC/mNC)>=Done
(mPC-mNC)>=
Reversed
Selecting the QC Criteria field opens the Test Definition Cutoff Quality screen:
Enter the lowest
acceptable ratio for
the mean of the
positive controls to
the mean of the
negative controls
the lowest
acceptable difference
for the mean
absorbance of the
positive controls
minus the mean
absorbance of the
negative controls.
controls read are too high or
when Low Absorbances are
Quality Criteria screen
The Blank is not subtracted in the Reverse Cutoff option. Pay close attention to “<” less
than and “>” greater than symbols when entering control limits in this case.
The Cut Off Mode Test Definition final page displays Interpretation Mode, Units, Decimals,
and number of Sample Replicates.
Cut Off Mode Test Definition - Final page
48 Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K
Cut Off Mode (Continued)
^
Pos >=
1.000 *COVv
Neg <1.000 *COVEdit
ReversedNODone
TEST DEFINITION Interpretation
NOTE: Double check
interpretation
By selecting Pos/Neg, Positive and Negative Interpretations can be entered. It is not
necessary to enter both Positive and Negative. These can also be set to be evaluated in
the reverse, where Positive is LESS than or equal to (<=) the value entered.
that the COV
calculations and
Cut Off Mode Test Definition – Interpretation
Press the Done button when finished editing.
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 49
3. Cleaning and Maintenance
3.1 Cleaning
3.1.1 Exterior
CAUTION: Solvents such as acetone or thinner will damage the instrument! Use only water
and recommended cleaners! Avoid abrasive cleaners. The display area is liquid-resistant,
but is easily scratched.
The exterior of the instrument may be cleaned with a soft cloth using plain water. If
needed, a mild all-purpose (nonabrasive) cleaner may be used. A 10% solution of
chlorine bleach (5.25% Sodium Hypochlorite) or 70% isopropyl alcohol may be used as a
disinfectant. Take special care not to spill any liquid into the read well.
3.2 Maintenance
3.2.1 Calibration and Linearity
Each instrument is calibrated during manufacturing using standards that are traceable to the
National Institute for Standards and Testing (NIST), and is tested to verify its linearity to 3A.
This preset calibration is very stable. Absolute calibration can be verified with the use of
NIST filters, or by periodic comparison to a reference instrument that is known to be
calibrated to NIST filters.
Absolute calibration can be verified using a specifically designed product called DRI-DYE
Check Strips, available from your instrument supplier.
The best way to assure quality instrument performance is to include a sufficient number of
controls in each assay to cover the entire operational range.
In vertical photometry, the fill volume and degree of meniscus determine the pathlength; and
absorbance is proportional to pathlength. Since lab test results are typically based upon
standards rather than upon absolute absorbances, the linearity of the instrument is the more
critical indicator of instrument performance.
A reduction in linearity may be indicative of filter deterioration. In this event, filter
replacement is required for continued reliable operation. A monthly verification of instrument
linearity is generally required to comply with regulatory agencies.
DRI-DYE
Alternately, the degree of linearity can be checked by reading a 1:2 serial dilution of a
material having a peak absorbance at or near one of the wavelengths, and observing a 1:2
ratio in the absorbance readings.
If you prepare dilutions to verify linearity, uniform pipetting (pathlength) and minimal transfer
error are important, since the tolerance limits which you will establish must take into
consideration the sources of error that are not due to the instrument
®
Check Strips can also be used to verify linearity.
.
®
50 Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K
Calibration and Linearity (Continued)
Place the instrument in the original packaging material.
functional tests be performed as if setting up the instrument for the first time.
Pipetting errors can be identified by comparing duplicate readings. The instrument should
give the expected value ± (1% of the expected value + .01A).
For example, if the 1/4 dilution reads .520 A, then you can expect your 1/2 dilution sample to
read twice as much or 1.04 A ± [(1% of 1.04) +0.01 A], which is ± .02.
A good working instrument would then give a result between 1.02 A and 1.06 A. You will
know that your filters are deteriorating when darkly colored solutions consistently read
lower than expected.
Since the procedures for verifying the calibration and linearity of a vertical photometer are
quite cumbersome and not conducive to stringent criteria, the best way to assure quality
instrument performance is with DRI-DYE® Check Strips.
3.2.2 Storage
The instrument may be stored under the following recommended environmental conditions:
• Temperature: 10 to 50°C
• Humidity: Less than 80% relative humidity, non-condensing.
Store the instrument using original packaging if possible.
Perform the following steps before storing:
Set the power switch to OFF and remove the power cord.
When returning the instrument to service from storage, it is recommended that
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 51
4. Troubleshooting
while calculating the concentration.
the maximum readable absorbance for the device.
negative absorbances!”
absorbance.
Edited!”
flag set to prevent it from being edited.
4.1 Messages and Flags
Messages are displayed to alert the operator when certain limits are approached. After displaying
the warning the instrument will continue to function normally.
Flags are displayed, for example, when an invalid range is entered. They are intended to help the
user correct the problem. Some error messages appear as dialog boxes that require action in
order to clear; others appear on the display screen and appear on the printout.
Message/Flag Explanation
>10**6” Appears in Concentration Column when Concentration is
greater than 10**6
“******” 1. A sample has exceeded the absorbance of the reference
well in a log-logit test. Curve Axes is Y=Logit(Abs),
X=Log(Conc) and Absorbance >reference(zero) absorbance
value.
2. Will print if a curve with a log axis tries to log a zero, a
negative number, or an invalid curve has been read.
3. Prints in Concentration column whenever an error occurs
“>3” Prints in ABS column if a read absorbance is greater than
“%Absorbance Mode
should not be used with
“Assay Cannot Be
“Cannot match test
wavelength with filters
currently installed on
device!”
“Do not select the same
filter for primary and
differential”
This message will display if the user is in %Absorbance
Mode and a standard is read which has a negative
User has tried to edit an OEM preprogrammed test that has
Pop up message that appears when running a test; occurs if
the device cannot find filters for the wavelengths set in the
test definition.
Appears on Page 1 of Test Definition when the same
wavelengths are entered for the Primary and the Differential
filters.
52 Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K
Messages andFlags (continued)
and < value NEG cutoff.)
Regression; minimum 4 for Cubic Spline).
Message/Flag Explanation
“EQUIV” This is reported when a Sample is less than the POS
cutoff and greater than the NEG cutoff concentration.
(If it is reversed, then the sample is > value POS cutoff
“Factory Settings Restored”
OR
“User Settings Restored “
“Invalid Curve!” This message indicates that readings for standards
“Invalid Curve – Log Error” An error in the creation of a curve or readings of a
”log(<=0)” Prints in Interpretation column whenever a curve axis
“LOW”, “HIGH” Prints in Interpretation column whenever a value is
“NEG” Negative – prints if a Sample is <NEG value (or if
“OUT” Blank is outside min/max range or sample is outside
“POS” Sample is >= POS value (or if reversed, sample is
One of these messages can be triggered when the
operator chooses to restore parameters or could be
seen at startup if the unit is forced to recover from bad
settings.
cannot be used to obtain a valid curve for the
calculation mode being used. If duplicate standards are
used, the user has the option to edit the curve if there
are enough points (minimum 3 for Point-to-Point and
negative absorbance will result in this message,
followed by Test Ended.
mode with a log tries to log a zero or negative number.
lower or higher than the range of values entered to
create the test.
reversed, sample is >NEG value).
low/high valid range values. Also appears in Cubic
Spline mode when the sample is outside of the range
of valid standards.
<=POS value.
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 53
Messages and Flags (continued)
increasing concentration.
Message/Flag Explanation
“Pos must be greater than or
equal to Neg!”
“Range Error” This indicates there is a limit to the range of numbers
“>Ref” Prints when a sample exceeds the absorbance of the
Test Data Corrupted!
OR
Test Definition Corrupted!
“Unable to assign control!” This message refers to control location in the test
“When reversed, Pos must be
less than Neg!”“POS”
“Wrong date format
(mm.yyyy)!”
This message appears during test definition when the
user is defining positive/negative interpretations. A
value was entered for both the positive and negative,
but the positive value is less than the negative value
entered.
that can be entered
reference well in a log-logit test.. This is usually due to
the calibrator concentrations being entered incorrectly
or the calibrators being pipetted into the strip
incorrectly. The first, or "Zero" calibrator must have the
highest absorbance (darker) and the remaining
calibrators must be of descending absorbance and
Test (assay) Data (including calibration data) does not
match what was stored.
OR
Test Definition does not match the data stored.
Once a test shows that it is corrupted, there is no way
to edit the test. It must be reprogrammed.
definition. If there are Replicates, then the location will
be the starting location. If the Controls or their
Replicates overlap due to user error (plate setup direction or side by side duplicates can affect this) the
user is warned and the test is ended.
Pop up message that appears during test definition
when the user is defining Positive/Negative
interpretations; a value was entered for both the
positive and negative and the reverse was selected,
but the positive value is greater than the negative value
entered. Sample is >= POS value (or if reversed,
sample is <=POS value.
This message indicates the expiration date for a control
is not entered in the correct format
54 Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K
4.2 Error Messages
Open the Utility menu and select Print Unit Settings.
power-off, wait, power-on"
Contact Technical Support
a plate to see
Indicates a problem was encountered and the
This indicates the instrument is not able to store assay
Contact Technical Support
Contact Technical Support
Error Messages are displayed when the instrument fails to operate correctly. They are intended to
help the user locate the problem. Some error messages require action in order to clear. Most
require help from technical support.
Error Message Explanation
Error 102
Provide this information to Technical Support.
Error 109 “Air Values Too
Dark”
“201, Read/Plate: Aborted,
Plate Movement Error”
"i2c Bus error, 0x%02x.
Unit may need service. Try
“Calibration Checksum
Faulty”
Although anything that blocks the path of the light to the
photometer can cause this, a burned out lamp is usually
the cause.
If an error occurs during plate movement or during running
a test, message will display and terminate the test.
Contact Technical Support for assistance.
If restarting the instrument does not clear this message,
contact technical support.
“Check unit Calibration”
Or
“Check Calibration”
“Default Settings Loaded”
“Error Saving Assay Data!”
“Factory Data Not Saved”
Factory Settings Invalid
Go to the "Utilities" menu
Run "Recall Calibration"
Message will be posted “Reload Factory Settings?”. Select
YES. “Factory Settings Restored” will be printed.
Select “done to return to the main menu”
Turn the instrument off, then back on and read
the message has been eliminated
If the message still appears, contact Technical Support.
instrument’s settings have been returned to the same
status as when instrument was shipped from the factory.
information into memory. If less than 120 assays have
already been stored, contact technical support.
OR
User Settings Invalid
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 55
Error Messages (continued)
Technical Support for a passcode.
Updating internal firmware
If the system finds an older version of FTDI firmware,
Error Message
FTDI initialization error
===========================
Hardware Init Error (#10)
5. References
Description
This message will be seen if the FTDI startup
initialization fails, tries again, but then fails a second
time.
=======================================
If the FTDI initialization fails again, the user will see
this message. If turning the instrument off and then
back on does not resolve the error, there is a
Troubleshooting button which requires contacting
this message will be displayed and the update will
happen automatically. Do not turn off the unit until the
update is complete
1. Engineering data supplied by Awareness Technology, Inc. Palm City, Florida (1987 to
current date)
2. Data on DRI-DYE
Florida (1989 to current date)
®
Check Strips, provided by Awareness Technology, Inc. Palm City,
56 Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K
6. Optional Accessories
6.1 Thermal Printer Paper
Reorder replacement Thermal Printer Paper Roll P/N 15003. NOTE: The printer manufacturer
strongly recommends the use of thermal paper part number 150003 to ensure the best print
quality and the effective operation of the print head. Contact your dealer for replacement rolls
6.2 Dri-Dye® Check Strips
•Dri-Dye® Check Strips offer an easy method of verifying proper calibration and linearity
on a routine basis. The charts provided allow simple interpretation of results. If a strip or
plate reader can repeatedly obtain acceptable results with this kit, then the instrument
demonstrates linearity, calibration, filter integrity, repeatability, and lack of stray light.
®
•Proper use of the Dri-Dye
compliance with most quality control and licensing agency requirements for instrument
performance verification.
®
•Dri-Dye
Check Strips allow the instrument owner to perform testing that is very similar
to the Q.A./Q.C. testing initially done by the instrument manufacturer. Dye absorbance
values are determined by a reference instrument that is routinely calibrated with materials
traceable to the National Institute of Standards and Technology (NIST, formerly NBS).
Dyes are then precisely dispensed into the wells, dried, and packaged.
•Using pre-dispensed dyes reduces the degree of pipetting precision required to
reconstitute the check set strips. Since the readers look through the samples from top to
bottom, volume is proportional to path-length, and therefore, to absorbance. If a well is
reconstituted with too much water, the decrease in concentration will be corrected by the
increase in path-length. Simply stated, the system is self-correcting for random pipetting
errors up to 10%.
Check Strips, along with good record keeping, will constitute
.
®
•DRI- DYE
Check Strips can be ordered from your dealer.
The charts provided with these kits are designed to replace the normal calculations with easy
to visualize, built-in acceptance ranges. Kits are available for the following wavelengths:
®
o DRI-DYE
o DRI-DYE
o DRI-DYE
Check Strips-405 - for 405nm
®
Check Strips-450 - for 450nm
®
Check Strips-492 - for 492nm
Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K 57
7. Information
Awareness Technology, Inc.
In the unlikely event that a problem occurs with the instrument, please contact your instrument
supplier.
Factory Information:
Telephone: USA 772-283-6540
Fax: USA 772-283-8020
E-mail: support@awaretech.com
Mailing Address:
1935 SW Martin Highway
Palm City FL 34990 USA
®
Important: When contacting us, please have the Model and Serial Number of the Stat Fax
question. Have a description of the problem with as much detail as possible. Save any relevant jobs
or logs and send or e-mail us the information.
4200 in
Model:__________________
Serial Number:_____________
WARNING: Instruments to be returned must be accompanied by a
decontamination certificate completed by the responsible laboratory manager. If a
decontamination certificate is not supplied, the returning laboratory will be
responsible for charges resulting from non-acceptance of the instrument by the
servicing center or from any authority’s intervention.
58 Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K
8. Appendix A SF_Capture
SF_Capture is a software program for use with this reader. Contact your dealer for a copy of the
program. To install drivers, refer to Section 2.3.2.1 External Output.
Install the program by double clicking on the setup.exe file found in the SF_Capture folder. The
installation process is the same as other software – follow the prompts until it is done.
Power on the instrument, connect the serial port to USB cable from the reader to the PC. Verify
that External Export has been switched to ON from the instrument's Settings menu
NOTE: An installation of drivers will need to be done the first time SF_Capture is installed
on a PC.
Once the cable connection is made between the reader and the PC, Windows® will pop up a
message box stating that the VNC1-A As Slave is not installed and that drivers will need to be
found.
Correct response is to let Windows
First, it will search for and install USB Serial Converter driver(s).
The cables can be connected and the instrument's power up can be done in any sequence. If the
instrument is OFF, but SF_Capture is open and the cable is connected to the instrument, select
ReConnect.
Select Settings. This will open a dialog box where you select the instrument type and the COM
port connected to on the PC side
®
look for the appropriate driver this one time.
Data is exported to Excel, to Notepad and to a Log file each time QUIT/Test Ended occurs.
The default location for these files is C:\4200. This can be changed from the default by using the
Browse option for the Device Root Folder.
Stat Fax® 4200Operator’s Manual Doc. 4200 Rev. K 59
9. Appendix B Updating Firmware
Adjustment
Touch Pad Align
To update firmware, follow the instructions below. Contact your dealer or technical support if you
have any questions.
1. Erase/delete anything that is on the USB pen drive (or any USB drive used).
2. Unzip the attached file and install the folder romimage.42x on the USB drive.
3. Turn off the reader and insert the USB drive with only the one romimage.42x folder
on it into the port.
4. Place a finger on the display and then power up the reader.
5. After the instrument begins to beep you can remove your finger from the screen.
6. The update process will take about 3-5 minutes, depending upon the instrument’s
bootloader version. To determine the version installed, go to Utilities – Show Version
Details.
.
10. Appendix C Touchpad Alignment
The ‘TouchPad Align’ feature is enabled via pass code from your Administrator. This feature
provides the ability to adjust the touchpad ranges by using the stylus or the mouse.
Selection of each of the eight squares should produce numbers on the display. If an area outside
of the acceptable range is pressed with the stylus or mouse, a sound will be produced. Each of
the pairs (top, left, bottom and right) must have a difference between them of not more than 3
points, such as on the left, 17 and 20 are good, whereas 17 and 24 are not. The SAVE button will
not appear until acceptable points have been selected. It is necessary to select SAVE in order to
store the new values.
Please carefully use a stylus to touch each of
the 8 small squares on the edges to help
adjust the touchpad ranges. Touch as closely
square
to the outside
for best results. If the points are not close
enough (3 pointsapart), some pairs may be
reset.
edge of the screen as possible
60 Stat Fax® 4200 Operator’s Manual Doc. 4200 Rev. K
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.