Copyright, 1997, AVL Scientific Corporation. All rights reserved. Unless otherwise noted, no part of this
publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into
any language in any form without the written permission of AVL Scientific Corporation.
For information contact:
AVL Scientific CorporationAVL MEDICAL INSTRUMENTS AGAVL LIST GmbH
50 Mansell CourtStettemerstrasse 28Kleiststrasse 48
P.O. Box 337CH-8207 SchaffhausenA-8020 Graz
Roswell, Georgia USA 30077SwitzerlandAustria
1-800-526-227241-848-800-88543-316-787-0
Printed in USA
US0285 REV K
ii
Important Information!
This Operator´s Manual contains important warnings and safety information to be observed by the
user.
This instrument is only intended for one area of application which is described in the instructions. The most important prerequisites for application, operation and safety, are explained to
ensure smooth operation. No warranty or liability claims will be covered if the instrument is
applied in areas other than those described or if the necessary prerequisites and safety
measures are not observed.
The instrument is only to be operated by qualified personnel capable of observing these
prerequisites.
Only accessories and supplies either delivered by or approved by AVL are to be used with the
instrument.
Due to the operating principle of the instrument, analytical accuracy not only depends on
correct operation and function, but also upon a variety of external influences beyond the
manufacturers control. Therefore, the test results from this instrument must be carefully
examined by an expert, before further measures are taken based on the analytical results.
Instrument adjustment and maintenance with removed covers and connected power mains are
only to be performed by a qualified technician who is aware of the dangers involved.
Instrument repairs are only to be performed by the manufacturer or qualified service personnel.
SymbolExplanation
Attention symbol - Refer to the Operator's Manual or Service
Manual for further instructions. This symbol is located on the
inside of the instrument.
Type B instrument symbol - An instrument of the B type falls under
safety categories I, II, or III, or has an internal power supply providing
the required insulation against discharge current and reliable ground
connections.
Important Information!
iii
Operating Safety Information
•This instrument falls under Safety Category I.
•This instrument is a Class B instrument.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) this device may not cause harmful interference’s, and (2) this device must accept
any interference received, including interference that may cause undesired operation.
Warning: Changes or modifications to this unit not expressly approved by the party
responsible for compliance could void the user’s authority to operate the equipment.
Note: This equipment has been tested and found to comply with the limits for a Class B
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference in a residential installation.
This equipment generates, uses, and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause harmful interference to
radio communication. However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does not cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and on, the
user is encouraged to try to correct the interference by one or more of the following
measures:
•Reorient or relocate the receiving antenna
•Increase the separation between the equipment and receiver
•Connect the equipment into an outlet on a circuit different from that to which
the receiver is connected.
•Consult the dealer or an experienced radio TV technician for help
Caution:
•The instrument is designed as a conventional device (closed, not waterproof type).
•Do not operate the instrument in an explosive environment or in the vicinity of
explosive anesthetic mixtures containing oxygen or nitrous oxide.
•This instrument is suitable for continuous operation.
•The power plug is to be plugged into a ground socket only. When using an extension
cord, make sure that it is of the proper size and is properly grounded.
•Any breakage of the ground lead inside or outside the instrument or a loose ground
connection can cause a hazardous condition when operating the instrument.
Intentional disconnection of the grounding is not permitted.
•When replacing the fuses, make sure that they are of the same type and rating as the
original fuses. Never use repaired fuses or short-circuit the fuse holders.
9.2 Example Data String Information............................................ 108
9.3 Data Link Information............................................................ 108
vi
1Introduction
The Service Manual for the 9120, 30, 40, 80, and 81 Electrolyte Analyzers
contains the technical information needed to ensure easy fault identification.
This manual is intended to be complementary to the Operator's Manual where
detailed instructions for operation, maintenance and troubleshooting are
provided.
As for all clinical instrumentation, a thorough understanding of the principles
of operation is prerequisite to attempting service of this product. Training
along with experience will enhance the use of this manual.
Service and repair of 9120, 30, 40, 80, and 81 analyzers should be performed
only by qualified repair technicians. Care should be taken when removing the
covers as hazardous voltages are exposed. Use only accepted electronic test
procedures and static protection when replacing and handling all electronic
parts.
This manual is divided into 9 chapters to facilitate location of technical information. Chapter 2 provides specifications and operating parameter information. Chapter 3 includes flow charts for all system functions and detailed
operation of system test procedures. In Chapter 4, all mechanical, fluidic and
electronic assemblies are described. Chapter 5 outlines routine maintenance
and troubleshooting procedures. Chapter 6 includes electronic adjustments
and Chapter 7 provides the system block diagram and all circuit schematic
diagrams and wiring interconnection information. In Chapter 8, part identification, location and description is provided by the illustrated parts list.
Finally, in Chapter 9 interface specifications are provided for the RS232
output.
Safety categoryI
Device typeB (according to ÖVE-MG/EN 60601-1,
IEC 601-1)
Mode of operationcontinuous operation
Protection classificationIP 20
Explosion protectionthe device is not designed for operation in
explosive environments.
ApprovalsCSA, CE, FCC Class B
CLIA Complexity Category: Moderate
Test System Code: 04739
FDA 510(k) number: K972673
Analyte Codes:Na+:5805
K+:4910
Cl-:1018
iCa++: 1004
Li+:3712
Data subject to change without notice. Technical information is
supplied for general informational purposes only.
4
3 Software Operation
3.1 Software Operation - 9120, 9130, 9140
Software operation of the 9120, 9130, 9140 analyzers can be accessed by YES/
NO selection on the analyzer front panel. The following flow chart diagrams are
provided to identify operating sequences of the 9120, 9130, 9140 Analyzers:
1. Overall Program Operating Flow
2. Main Menu
3. Measurement Sequence
4. Print Functions Menu
5. QC/Standard/Urine Sample Menu
6. Daily Maintenance Menu
7. Operator Functions Menu
8. Program Instrument Menu
9. Service Functions Menu
10. Calibration Sequence
3.2 Software Operation - 9180
1. Overall Program Operating Flow
2. Main Menu
3. Measurement Sequence
4. Print Functions Menu
5. QC/Std/Dialysate/Urine Sample Menu
6. Daily Maintenance Menu
7. Operator Functions Menu
8. Program Instrument Menu
9. Service Functions Menu
10. Calibration Sequence
11. Power-up Sequence
5
3.3 Software Operation - 9181
1. Overall Program Operating Flow
2. Main Menu
3. Measurement Sequence
4. Print Functions Menu
5. QC/Std/Dialysate/Urine Sample Menu
6. Daily Maintenance Menu
7. Operator Functions Menu
8. Program Instrument Menu
9. Service Functions Menu
10. Calibration Sequence
11. Power-up Sequence
The Service Functions Menu is also provided with a more detailed description
of each test parameter for use in testing instrument subassemblies.
6
3.1 Software Operation - 9120, 30, 40
1 Overall Program Operating Flow - 9120, 30, 40
NoNoNoNoNoNoNoNo
.
CALIB R AT ION?
SERVICE
PROGRAM
OP E R AT OR
DAILY
CALI B R AT E ..
No
Test
YesYesYes
Enter Code:
Check/Change
D aily
E lectrodes?
No
AAA
No
SnapPak?
No
Cleaning?
FUNCTIONS?
INS T R U ME N T ?
FUNCTIONS?
MAIN T E N ANCE ?
No
Test
Sample Sensor?
No
Program QC
Level 1 Ranges?
No
Go to
S t andby M ode?
No
D aily
Conditioning?
No
Test
S am ple D oor ?
No
Program QC
Level 2 Ranges?
No
Set
Time/Date?
No
Maintenance?
R emain in Daily
Yes
No
Test
Pump?
No
Program QC
Level 3 Ranges?
No
Reset
Sample Number?
CALI B R AT E
No
Test
Valves?
No
Program
Normal Ranges?
No
Operation?
Take Out of
Test Interface?
Program
Rem ain in
Jump Pins 2 & 3
Corr. Factors?
Operator F unc.?
Yes
No
No
No
No
Test
Am plifier?
No
Program
P ri nter S etup?
Enter
Rem ain in
S ervice Code?
Program Func.?
Yes
No
No
R emain in
No
Service Func.?
Yes
YesYesYes
QC/S T ANDAR D /
URINE SAMPLE?
Yes
FUNCTIONS?
R E ADY 12:45PRINT
No
Sample?
QC Level 1
No
P rint las t
Sample Report?
No
Sample?
QC Level 2
No
Print
Cal. R eport?
No
Sample?
QC Level 3
No
and S tatis tics ?
Print QC Values
No
Sample?
Standard
No
Functions?
Rem ain in P r int
Yes
No
Urine
Sample?
No
Remain in QC/
Std/Urine Sample?
Yes
7
2 Main Menu - 9120, 30, 40
MAIN MENU
READY 12:45
NO
PRINT
FUNCTIONS?
NO
QC/STANDARD/
URINE S AMPLE?
NO
DAILY
MAINTENANCE?
NO
OPERATOR
FUNCTIONS?
NO
LEGEND
Display prompt
Action performed by analyzer
2-second display prompt
Decision-making by analyzer
PROGRAM
INSTRUMENT?
NO
SERVICE
FUNCTIONS?
NO
NO
CALIBRATION?
8
3 Measurement Sequence - 9120, 30, 40
READY 12:45
OPEN
FLAP
Please
Introduce Sample
CHECK SS
VOLTAGE
> air
voltage
YES
NO
NO
< air
voltage
PRINT
FUNCTIONS?
Open Sample Door
Introduce Sample
DOOR OPENED
WITHIN 2s?
YES
Remove Sample
and Wipe Probe
Please Close
Sample Door
Serum Sample
in Process 49
Na K Cl
151 5.1 114
YES/NO
1 min.
Remove Sample
and Wipe Probe
2s2s
Please Close
Sample Door
FLAP
CLOSED
Blood Sample
in Process 49
Na K Cl
151 5.1 114
CLOSED
FLAP
YES/NO
1 min.
9
4 Print Functions Menu - 9120, 30, 40
YES
PRINT
FUNCTIONS?
YES
Print Last
Sample Report?
NO
Print
Cal. Report?
NO
Print QC Values
and Statistics?
NO
Remain in Print
Functions?
NO
YES
YES
YES
NO
QC/STANDARD/
URINE SAMPLE?
PRINTING
...
READY 12:45PRINTING
...
PRINTING
...
READY 12:45
10
5 QC/Standard/Urine Sample Menu - 9120, 30, 40
QC/STANDARD/
URINE SAMPLE?
YES
QC Level 1
Sample?
2 min.
2 min.
2 min.
NO
QC Level 2
Sample?
NO
QC Level 3
Sample?
NO
NO
YESOPENSAMPLE
YESOPENSAMPLE
YES
DAILY
MAINTENANCE?
Open Sample Door
Introduce Sample
Open Sample Door
Introduce Sample
Open Sample Door
Introduce Sample
2s
2s
2s
2s
OPEN
2s
2s
Introduce
Sample
VALUES
REJECTED!
VALUES
ACCEPTED!
Introduce
Sample
VALUES
REJECTED!
VALUES
ACCEPTED!
Introduce
Sample
VALUES
REJECTED!
VALUES
ACCEPTED!
IN
NO
YES
IN
NO
YES
SAMPLE
IN
NO
YES
Wipe Probe/Close
Sample Door
2s
Store
Values?
2 min
READY 12:45
Wipe Probe/Close
Sample Door
2s
Store
Values?
2 min
READY 12:45
Wipe Probe/Close
Sample Door
2s
Store
Values?
2 min
READY 12:45
CLOSED
2s
CLOSED
2s
CLOSED
2s
QC Level 1
in Process 49
Na K Cl
99.65.34 88.2
QC Level 2
in Process 49
Na K Cl
99.6 5.34 88.2
QC Level 3
in Process 49
Na K Cl
99.6 5.34 88.2
Standard
Sample?
2 min.
Sample?
2 min.
Remain in QC/
Std/Urine Sample?
YES
Urine
YESOPENSAMPLE
Open Sample Door
Introduce Sample
NO
YESOPENSAMPLE
Open Sample Door
Introduce Sample
NO
READY 12:45
NO/
2 min.
5s
5s
Introduce
Sample
Introduce
Sample
IN
IN
Wipe Probe/Close
Sample Door
Wipe Probe/Close
Sample Door
CLOSED
CLOSED
Standard Sample
in Process 49
Urine Sample
in Process 49
Na K Cl
158.8 5.82 113.2
1 min.
NO/
YES
Na K Cl
238 18.8 198
1 min.
NO/
YES
11
6 Daily Maintenance Menu - 9120, 30, 40
DAILY
MAINTENANCE?
Daily
Cleaning?
Daily
Conditioning?
Remain in Daily
Maintenance?
YES
CLEANING AND/OR
COND. PERFORMED?
YES
NO
NO/1min.
NO
YES
Open Sample Door
OPEN
OPEN
Introduce Sample
YES
Open Sample Door
Introduce Sample
YES
OPERATOR
FUNCTIONS?
30 SECOND
STABILIZATION
OPENCLOSEYES
OPENCLOSEYES
Wipe Probe/Close
Sample Door
Cleaning Solution
not detected!
Wipe Probe/Close
Sample Door
Condition. Solution
not detected!
CALIBRATE
Cleaning Solution
detected?
NO
Condition. Solution
detected?
NO
READY 12:45
CLEAN
CONDITION
12
NO
Calibration
due?
NO
YES
7 Operator Functions Menu - 9120, 30, 40
OPERATOR
FUNCTIONS?
YES
Check/Change
SnapPak?
NO
Go to
Standby Mode
NO
Set
Time/Date?
NO
NO
YES
YES
NO
PROGRAM
INSTRUMENT?
37%
remaining
***STANDBY!***
YES -> REA D Y
YES/NO
Leave
Standby Mode?
YES
NOYES
PRESS YES
WITHIN 2s
Date: 01- FEB-9 4
Time: 12:00 ok ?
NO
Date: 01- FEB-9 4
Time: 12:00 ok
YES
YES
Change
SnapPak?
NO
IS
CALIBRATION
DUE?
YES
Daily
Cleaning?
YES
READY 12:45
YES
New Snap Pa k
installed?
NO
NO
NONO
Daily
Conditioning?
YES
CONDITIONCLEAN
30 SECON D
STABILIZATION
CALIBRATE
YES
Fluid Counter
reset to 100%
Maint . due?
NO
New Pack
Prime ...
CALIBRATE
READ Y 12:45
YES
Remain in Daily
Maintenance?
1min/NO
IS
CALIBRATION
DUE?
YES
YESYES
NO
Sample Number?
Operator Func?
NO
Take out of
Operation?
Remain in
YES
NO
NO
YES
Are you sure?Reset
Shutdown Plug /
Water installed?
YES
ShutdownComplete
Turn Pow er Off
YES
YES/
NO
YES
YES
SET SAMPL E
NUMBER TO 0
Transp Housing
Ref installed?
Do You Wish to
DELETE all Data?
YESYES
Remove Water
and Pres s YES
YES
Pump Windings
Relieved?
YES
All Electrodes
& Plug Rem oved?
YES
4 Relief Clam ps
Inserted?
13
8 Program Instrument Menu - 9120, 30, 40
PROGRAM
INSTRUMENT?
YES
Enter Code:
AAA
CORRECT
Program QC
Level 1 Ranges?
NO
Program QC
Level 2 Ranges?
NO
Program QC
Level 3 Ranges?
NO
Program
Normal Ranges?
NO
Program
Corr Factors
NO
NO
IN-
CORRECT
YES
YES
YES
YESYESYESYES
Current Lot: 0000
Cl low = 050
Cl high = 200
"see Level
"see Level
Na low = 135.0
Na high = 150.0
Factor (default)?
YES
Cl (b) = +00.0
Cl (m) = 1.000
YES
CODE ERROR!
RETRY?
Change Lot?
NO
1"
1"
Reset Corr
YES
NO
YES
NOYES
SERVICE
FUNCTIONS?
Print old Values
and Statistics?
K low = 1.5
K high = 15.0
K low = 3.50
K high = 5.50
Input/Verify
Corr Factors?
K (b) = +00.0
K (m) = 1.000
NO
NOYES
NO
YES
YESYES
New Lot!
Delete old Data?
Na low = 040
Na high = 205
Cl low = 95.0
Cl high = 108.0
Remain in
Corr Factors?
NO
Na (b) = +00.0
Na (m) = 1.000
YESYESYESYES
YES
Enter last 4 Dig
of Lot#: 0000
YES
Lot Number
ok?
NO
Program Printer
NO
Program Func?
YES
NO
READY 12:45
14
Setup?
Remain in
YES
Patient Report
off?
YES
NO
Print two
Patient Reports?
YES/
NO
Automatic
Cal Report?
YES/
NO
9 Service Functions Menu - 9120, 30, 40
SERVICE
FUNCTIONS?
YES
Test
Electrodes?
NO
Test
Sample Sensor?
NO
Test
Sample Door?
NO
Test Pump?
NO
Test Valves?
NO
Test Interface?
Jump Pins 2&3
NO
Test
Amplifier?
NO
Enter
Service Code?
NO
NO
YESYES
YES
YES
YES
YES
YES
YESYESYES
CALIBRATION?
Test
Standard A?
NONO
Test
Standard B?
NONO
Test
External Sample?
YES
Remain in
Test Electrodes?
NO
DRY: 80-120: 102
YES=Pump/NO=Exit
Sample Door: C
C:Closed O:Open
Test Pump!
Very slow speed
Test Valve A:
YES/NO=ON/OFF
-- Test --
-- Interface --
Na=-1030 K=-1200
Cl = -1093
Enter Code:
AAA ok?
NO
YESYES
YES
YES
NO
2s2s2s
NONONO
NO
Std. A: Na =0000
K =0000 Cl =0000
Std. B: Na =0000
K =0000 Cl =0000
Open Door and
Introduce Sample
Ext.: Na =0000
K =0000 Cl =0000
NO
External: 102
Pump: NO=OFF
Test Pump!
Slow speed
Test Valve B:
YES/NO=ON/OFF
Interface
ok!
Ground = 2
-2500mV = -2496
Is Code
valid?
NO
Code
Error!
CLOSED
OPEN
Introduce Sample
Wipe Probe and
Close Sample Door
wash...
Test Pump!
Medium speed
Test Valve V:
YES/NO=ON/OFF
YES
Remain in
Test Interface?
NO
Code
Accepted!
Test Pump!
Fast speed
2s
Test Valve R:
YES/NO=ON/OFF
NO
YES
Enter Other
Service Code?
NO
YES
Remain in
Service Func?
NO
Calibration?
Calibrate...
READY 12:45
15
10 Calibration Sequence - 9120, 30, 40
CALIBRATION?
YES
Calibration in
Process 124
Clear Sample
Path
Calibrate
Sample Sensor
Sample Sensor
Voltage ok?
YES
Prime
Standard B
Standard B
detected?
YES
Prime
Standard A
Standard A
detected?
NO
SAMPLE SENSOR
NO
**STANDARD B**
NOT DETECTED
NO
**STANDARD A**
NOT DETECTED
***CHECK***
A
Asp. Std. A into
Meas. Chamber
Standard A
detected?
YES
Obtain Electrode
Voltages Std. A
3 Std. A Values
obtained?
YES
Are Values
reproducible?
YES
Aspir. Std. B into
Meas. Chamber
Standard B
detected?
YES
Obtain Electrode
Voltages Std. B
NO
**STANDARD A**
NO
NO
NO
**STANDARD B**
NOT DETECTED
NO
6 Aspir. of
Std.A performed?
YES
***CHECK***
ELECTRODES
NOT DETECTED
16
YES
Pressurize Ref.
Asp. Std. A into
Meas. Chamber
Reference
Housing ok?
YES
A
NO
***CHECK***
REF HOUSING
Perform
Wash Cycle
Aspir. Std. A
into Meas.
Chamber
Standard A
detected?
Calculation of
Std.A-Std.B
(slope)
Slope of all
Electrodes in
Range?
YESYES
READY 12:45READY 12:45
NO
**STANDARD A**
NOT DETECTED
YES
NONO
At least 1 slope
XX NOT CAL'D
in Range?
***CHECK***
ELECTRODES
3.2 Software Operation - 9180
1 Overall Program Operating Flow - 9180
NoNoNoNoNoNoNoNo
CALIBRATION?
SERVICE
PROGRAM
OPERATOR
DAILY
CALIBRATE...
No
Test
FUNCTIONS?
INSTRUMENT?
FUNCTIONS?
MAINTENANCE?
Electrodes?
AAA
Enter Code:
SnapPak?
Check/Change
Cleaning?
Perform Daily
No
No
No
YesYesYes
No
Test
Sample Sensor?
No
Program QC
Level 1 Ranges?
No
Go to
Standby Mode?
No
Perform Daily
Conditioning?
No
Test
SnapPak Sensor?
No
Program QC
Level 2 Ranges?
No
Set
Time/Date?
No
Maintenance?
Remain in Daily
No
Test
Language Switch?
No
Program QC
Level 3 Ranges?
No
Configuration?
Select Parameter
CALIBRATE...
Test
Program
Reset
No
Sample Door?
No
Normal Ranges?
No
Sample Number?
Test
Program
Take Out of
No
Pump?
No
Corr. Factors?
No
Operation?
No
Test
Valves?
No
Corr Facto r s?
Program Bicarb.
No
Remain in
Operator Func.?
Yes
No
Test Interface?
Jump Pins 2 & 3
No
Corr Facto r s?
Program Acetate
No
Test
Amplifier?
No
Corr Facto r s?
Program Urine
Enter
Program
Service Code?
Printer Setup?
No
No
Program
No
Remain in
Service Func.?
Yes
No
Interface?
No
Remain in
Program Func.?
Yes
Yes
YesYesYes
/URINE SAMPLE?
QC/STD/DIALYSATE
Yes
FUNCTIONS?
Na K Cl READYPRINT
No
Sample?
QC Level 1
No
Print last
Sample Report?
No
Sample?
QC Level 2
No
Print
Cal. Report?
No
Sample?
QC Level 3
No
and Statistics?
Print QC Values
No
Sample?
Standard
No
Functions?
Remain in Print
Yes
No
Sample?
Bicarbonate
No
Acetate
Sample?
No
Urine
Sample?
No
Remain in QC/
Std/Urine Sample?
Yes
17
2 Main Menu - 9180
MAIN MENU
NA K CL
READY
NO
PRINT
FUNCTIONS?
NO
QC/STD/DIALYSATE
/URINE SAMPLE?
NO
DAILY
MAINTENANCE?
NO
OPERATOR
FUNCTIONS?
LEGEND
Display prompt
Action performed by analyzer
2-second display prompt
Decision-making by analyzer
18
NO
PROGRAM
INSTRUMENT?
NO
SERVICE
FUNCTIONS?
NO
NO
CALIBRATION?
3 Measurement Sequence - 9180
NA K CL READY
OPEN
FLAP
Please
Introduce Sample
CHECK SS
VOLTAGE
> air voltage
NO
YES
NO
< air voltage
PRINT
FUNCTIONS?
Open Sample Door
Introduce Sample
DOOR OPENED
WITHIN 2s?
YES
Remove Sample and
Wipe Probe
Please Close
Sample Door
Serum Sample
in Process 49
Na K Cl
151 5.1 114
YES/NO
1 min.
Remove Sample and
Wipe Probe
2s2s
Please Close
Sample Door
FLAP
CLOSED
Blood Sample
in Process 49
Na K Cl
151 5.1 114
FLAP
CLOSED
YES/NO
1 min.
19
4 Print Functions Menu - 9180
NO
YES
PRINT
FUNCTIONS?
YES
Print Last
Sample Report?
NO
Print
Cal. Report?
NO
Print QC Values
and Statistics?
NO
Remain in Print
Functions?
YES
YES
YES
NO
QC/STD/DIALYSATE
/URINE SAMPLE?
PRINTING...
PRINTING...
PRINTING...
NA K CL READY
NA K CL READY
20
5 QC/Standard/Dialysate/Urine Sample Menu - 9180
QC/STD/DIALYSATE
/URINE SAMPLE?
YES
QC Level 1
Sample?
2 min.
2 min.
2 min.
2 min.
2 min.
2 min.
2 min.
NO
QC Level 2
Sample?
NO
QC Level 3
Sample?
NO
Standard
Sample?
NO
Bicarbonate
Sample?
NO
Acetate
Sample?
NO
Urine
Sample?
NO
NO
YESOPENSAMPLE
OPEN
YESOPENSAMPLE
OPEN
YESOPENSAMPLE
OPEN
YESOPENSAMPLE
OPEN
YESOPENSAMPLE
OPEN
YESOPENSAMPLE
OPEN
YESOPENSAMPLE
OPEN
DAILY
MAINTENANCE?
Open Sample Door
Introduce Sample
Open Sample Door
Introduce Sample
Open Sample Door
Introduce Sample
5s
Open Sample Door
Introduce Sample
5s
Open Sample Door
Introduce Sample
5s
Open Sample Door
Introduce Sample
5s
Open Sample Door
Introduce Sample
Introduce SampleWipe Probe/Close
IN
2s
2s
2s
2s
2s
2s
VALUES
REJECTED!
VALUES
ACCEPTED!
Introduce SampleWipe Probe/Close
VALUES
REJECTED!
VALUES
ACCEPTED!
Introduce SampleWipe Probe/Close
VALUES
REJECTED!
VALUES
ACCEPTED!
Introduce SampleWipe Probe/Close
Introduce SampleWipe Probe/Close
Introduce SampleWipe Probe/Close
Introduce SampleWipe Probe/Close
NO
YES
IN
NO
YES
IN
NO
YES
IN
IN
IN
IN
Sample Door
2s
Store
Values?
2 min
NA K CA READY
Sample Door
2s
Store
Values?
2 min
NA K CA READY
Sample Door
2s
Store
Values?
2 min
NA K CA READY
Sample Door
Sample Door
Sample Door
Sample Door
CLOSED
2s
CLOSED
2s
CLOSED
2s
CLOSED
CLOSED
CLOSED
CLOSED
QC Level 1
in Process 49
Na K Ca
99.6 5.34 1.250
QC Level 2
in Process 49
Na K Ca
99.6 5.34 1.250
QC Level 3
in Process 49
Na K Ca
99.6 5.34 1.250
Standard Sample
in Process 49
Bicarb. Sample
in Process 49
Acetate Sample
in Process 49
Urine Sample
in Process 49
Na K Ca
158.8 5.82 1.250
1 min. NO/YES
Na K Ca
158.8 5.82 1.250
1 min. NO/YES
Na K Ca
158.8 5.82 1.250
1 min.
NO/YES
Na K Cl
238 18.8 198
1 min. NO/YES
Remain in QC/
Std/Urine Sample?
YES
NO/2 min.
NA K CA READY
21
6 Daily Maintenance Menu - 9180
DAILY
MAINTENANCE?
Perform Daily
Cleaning?
Perform Daily
Conditioning?
Remain in Daily
YES
Maintenance?
CLEANING AND/OR
COND. PERFORMED?
YES
NO
NO/1min.
NO
YES
OPEN
YES
OPEN
YES
OPERATOR
FUNCTIONS?
Open Sample Door
Introduce Sample
Open Sample Door
Introduce Sample
30 SECOND
STABILIZATION
OPENCLOSEYES
OPENCLOSEYES
Wipe Probe/Close
Sample Door
Cleaning Solution
not detected!
Wipe Probe/Close
Sample Door
Condition. Solution
not detected!
CALIBRATE
Cleaning Solution
detected?
NO
Condition. Solution
detected?
NO
NA K CA READY
CLEAN
CONDITION
22
NO
Calibration
due?
NO
YES
7 Operator Functions Menu - 9180
*****STATUS*****
**NO SnapPak**
Check/Change
Select Parameter
Sample Number?
Operator Func?
NO
OPERATOR
FUNCTIONS?
YES
SnapPak?
NO
Go to
Standby Mode
NO
Set
Time/Date?
NO
Configuration?
NO
NO
Take out of
Operation?
NO
Remain in
YES
NO
YES
YES
YESNONONO
YES
YES
PROGRAM
INSTRUMENT?
37%
remaining
***STANDBY!***
YES -> READY
YES/NO
Leave
Standby Mode?
YES
NOYES
PRESS YES
WITHIN 2s
Date: 01-FEB-94
Time: 12:00 ok ?
Date: 01-FEB-94
Time: 12:00 ok
NO
YES
Sel. Parameter:
[Na] [K] [ ] ok?
Are you sure?Reset
Shutdown Plug and
Water installed?
Shutdown
Complete
Turn Power Off
YESYESYES
YES
NO
YES/
NO
YES
YES
YES
Change
SnapPak?
NO
IS
CALIBRATION
DUE?
YES
Daily
Cleaning?
YES
NA K READY
Sel. Parameter:
[Na] [K] [Cl] ok?
SET SAMPLE
NUMBER TO 0
Transp Housing
Ref installed?
Do You Wish to
DELETE all Data?
YES
NO
NONO
NO
YESYES
Pack
detected?
GLOBAL*
New SnapPak
installed?
NO
Daily
Conditioning?
CONDITIONCLEAN
30 SECOND
STABILIZATION
CALIBRATE
Sel. Parameter:
[Na] [K] [Ca] ok?
Sel. Parameter:
([Na]) [ ] [Li] ok?
Remove Water
and Press YES
Pump Windings
Relieved?
NO
YES
YES
YES
YESYES
*Background check in all menus
except "Take out of Operation?"
YES
Fluid Counter
reset to 100%
Maint. due?
CALIBRATE
Na K READY
Remain in Daily
Maintenance?
CALIBRATION
YES
Parameter
NO
Electrodes
Sel. Parameter:
[Na] [K] [Li] ok?
YES
NO
Sel. Parameter:
[Na] [ ] [Li] ok?
All Electrodes
& Plug Removed?
5 Relief Clamps
Inserted?
New Pack
Prime ...
IS
DUE?
New
added?
in place?
YESYES
NO
YES
1min/NO
NO
YES
YES
YES
NO
YES
23
8 Program Instrument Menu - 9180
PROGRAM
INSTRUMENT?
YES
Enter Code:
AAA
CORRECT
Program QC
Level 1 Ranges?
NO
Program QC
Level 2 Ranges?
NO
Program QC
Level 3 Ranges?
NO
Program
Normal Ranges?
NO
Program
Corr Factors
NO
Program Bicarb.
Corr Factors
NO
Program Acetate
Corr Factors
NO
Program Urine
Corr Factors
NO
NO
IN-
CORRECT
YES
NO
YES
YES
YES
YES
SERVICE
FUNCTIONS?
YES
CODE ERROR!
RETRY?
Current Lot: 0000
Change Lot?
NO
Additional
Parameters?
YES
"see Level 1"
"see Level 1"
Na low = 136
Na high = 145 ok?
Reset Corr
Factor (default)?
YES
Reset Bicarb.
Factor (default)?
YES
Reset Acetate
Factor (default)?
YES
Reset Urine
Factor (default)?
YESYES
NO
YESNOYES
Print old Values
and Statistics?
K low = 1.5
K high = 15.0 ok?
Cl low = 050
Cl high = 200 ok?
YESYES
K low = 03.5
K high = 05.1 ok?
NOYES
Input/Verify
Corr Factors?
NO
Input/Verify
Bicarb. Factors?
NOYES
Input/Verify
Acetate Factors?
NOYES
Input/Verify
Urine Factors?
NO
New Lot!
Delete old Data?
Na low = 040
Na high = 205 ok?
Ca low = 0.2
Ca high = 5.0 ok?
YES
NOYES
YES
NO
YES
NOYES
YES
NOYES
Remain in
Corr Factors?
NO
YES
Remain in
Bicarb Factors?
NO
YES
Remain in
Acetate Factors?
NO
YES
Remain in
Urine Factors?
NO
Enter last 4 Dig
of Lot#: 0000
YESYESYES
Enter last 4 Dig
of Lot#: 0000 ok?
Li low = 0.1
Li high = 6.0 ok?
Na (b) = +00.0
Na (m) = 1.000
Na (b) = +00.0
Na (m) = 1.000
Na (b) = +00.0
Na (m) = 1.000
Na (b) = +00.0
Na (m) = 1.000
YES
NO
YESYESYES
K (b) = +00.0
K (m) = 1.000
YES
K (b) = +00.0
K (m) = 1.000
K (b) = +00.0
K (m) = 1.000
K (b) = +00.0
K (m) = 1.000
YES
YES
YES
YES
YES
24
Program Printer
Setup?
NO
Program
Interface?
NO
Remain in
Program Func?
YES
YES
NO
Patient Report
off?
YES
Activate
Data Link?
YES/NO
NA K READY
NO
Print two
Patient Reports?
YES/
NO
Automatic
Cal Report?
YES/
NO
Enter Comment
Line?
NOYES
YES
Enter Comment:
_
Loading...
+ 84 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.