All rights reserved. All trademarks and registered trademarks are the property of their respective owners.
This product is covered by one or more of the following U.S. Patents: 7,909,783, 8,246,556, 9,283,139, and 9,439,828 and various pending
U.S. patent applications. Purchase of this package does not act to grant a use license under these patents in any jurisdiction outside of
the United States. Additional patents pending.
Information and views expressed in this document, including URL and other internet web site references, may change without notice.
This manual describes the Footbeat circulation enhancement and recovery system (“Footbeat”) and provides instructions for its use.
Please read the entire manual carefully before use. This manual contains information, contraindications and warnings that must be
observed to ensure safe operation and maintain Footbeat in a safe condition.
The Footbeat System
START RECOVERING. Footbeat is the natural, non-invasive solution for enhancing circulation throughout the body when you are seated,
immobile or lying down. Footbeat’s patented micro-mobile compression technology, speeds recovery and improves quality of life
by compressing the veins of the foot and naturally pumping blood up the legs. This unique action duplicates the circulatory benets
received from being active or walking.
Technical Description
Footbeat provides therapeutic massage through a cyclic compression to the arch of the foot when the user is in a non-weight bearing
position. The Footbeat insole contains a pad that presses into the arch of the foot approximately every 35 seconds. The pad holds for 2
seconds and then fully retracts. When a user walks or runs, the Footbeat system senses this activity and pauses. Upon return to a seated
or recumbent position, the Footbeat system will reactivate the compression cycle.
What’s In The Box
1. Quick Start Guide
2. Insoles with engines
3. Socks (2 pairs)
4. Tote bag
5. Mocs
6. Instructions For Use (IFU)
7. Remote
8. Charging cable
1 QUICK START GUIDE
6 IFU
3 SOCKS
7 REMOTE
2 INSOLES
5 MOCS
Intended Use
Footbeat provides a therapeutic foot massage that provides relief from aches and pains.
NOTE – The instructions are for information only and should not be used for diagnosis, prevention or treatment of medical conditions.
Footbeat is not a prescriptive therapy for any disease or illness. Footbeat is non-invasive and not pharmacological. There are no known
side eects for Footbeat. Please consult your healthcare provider regarding the duration and frequency that Footbeat should be used to
achieve the maximum benets from Footbeat.
4 TOTE BAG
8 CHARGING
CABLE
IMPORTANT
Mocs should not be worn outdoors.
Contraindications
Unless directed by a physician, DO NOT use Footbeat if you have:
a. Recently had a blood clot or are under the care of a physician for Deep Vein Thrombosis (DVT)
b. Pulmonary edema
c. Pulmonary embolism (PE)
d. Foot ulceration, wounds or lesion at or near the plantar venous plexus on the bottom of the foot
Warnings
1. DO NOT use in conjunction with compression stockings or compression sleeves worn around the feet, ankles or legs.
2. DO NOT wear or use while operating any type of vehicle.
3. DO NOT use with ice or cold therapy devices.
4. Failure to observe the following procedures may result in battery leakage, heat, bursting, and serious personal injury.
a. Do not remove any batteries from product.
b. Never dispose of batteries in re.
c. Keep batteries away from combustible materials.
d. Do not incinerate or heat batteries.
e. Do not disassemble, alter, or reconstruct product.
f. Batteries contain a strong colorless electrolyte, which may cause permanent loss of eyesight and injury when in direct contact
with eyes or skin. Immediately ush with clean water when in contact with leaked electrolyte and consult a physician.
To insure proper safety and optimum Footbeat operation, observe the following warnings:
1. There are no serviceable parts. To avoid the risk of electrical shock, do not attempt to open or modify the insoles, charger or remote.
2. To avoid electrical shock, product malfunction or damage, never operate with mechanical damage, or if the unit is otherwise not fully
operational.
3. To avoid the risk of electrical shock, unplug the device when in use.
4. Never operate with any AC adapter other than that provided by AVEX LLC. Use of other adapters may result in electrical shock and will
void the warranty.
Precautions
1. Anyone under the age of 18 years old should consult a healthcare provider before using Footbeat.
2. Use the device on the foot only and per instructions.
3. Not intended for use during athletic activities.
4. Because Footbeat is in sleep mode when walking or running and won’t compress the foot, it should not be worn for extensive
walking. Do not run or jump while wearing Footbeat.
5. Other conditions are presented in this IFU in appropriate locations, identied with the caution symbol.
IMPORTANT
If you have any questions about any existing health conditions, consult your
Health Care Provider prior to using Footbeat.
Symbols and Explanations
32F
14F
140F
Electrical Caution
Caution
Consult Instructions For Use
Keep Dry
Dispose of in accordance with
local and national regulations
140F
14F
Temperature Limitation
Humidity Limitation
104F
Temperature Limitation
Manufacturer
Date of Manufacture
SN
Serial Number
Refer to the warnings to avoid electrical shock.
A hazardous situation, which if not avoided, could result in injury.
Read Instructions For Use for care, storage, operation and warnings.
Follow all recommended washing directions to avoid damage.
Follow all local and national regulations when disposing of
parts or all of product.
To avoid damage, store within 14º F to 140º F (-10º C to 60º C).
Maximum 90% RH (no condensation).
To avoid damage, operate within 32º F to 104º F (0º C to 40º C).
Name of manufacturer.
Date the product was manufactured.
Assigned serial number of device(s)
REF
Catalog Reference Number
Identies the specic product.
How to properly t Footbeat
1. The insole pad should be positioned to press into the center of the arch.
2. Adjust the Moc to ensure a rm but not tight t.
IMPORTANT
A loose or too tight tting Moc will result in sub-optimal results. Always wear
socks while using Footbeat.
How to use and secure the Moc
Footbeat comes with a pair of performance Mocs specically designed to provide the optimum comfort and performance from
Footbeat. The Mocs feature an expanded insole cavity, comfortable toe box and a custom strap that provides closure and support. The
adjustable strap enables generous expansion to accommodate swelling. With the strap tightened, the t should be comfortably rm,
not tight, and never painful. The primary purpose of the Mocs is to secure Footbeat in place so that as the insole pad extends into the
arch. Footbeat should not push away from the bottom of the foot. The Mocs should be used for minimal indoor walking, while seated
or lying down.
Charge Insoles
1
Turn on with Remote
4
Insert Insoles into Mocs
2
Relax & Recover
5
Secure Moc to Foot
3
Operation
1. INSERT INSOLE into Moc.
2. TURN ON: Push and hold the button on the remote for one second. A green LED light will appear on the remote. Units will activate.
3. TURN OFF: Push and hold the button on the remote for one second. A red LED light will appear on the remote.
4. LOW BATTERY CONDITION: A red ashing LED light will appear on the remote.
5. Footbeat is engaged when you are seated or reclining. Footbeat automatically pauses as you stand or walk. Footbeat resumes as you
return to a seated or reclining position.
6. PAIRING: The Footbeat remote works exclusively with the insoles provided to you. However, the remote may un-pair with your insoles
from time to time. This can happen due to changing the remote battery or due to a loss of power to the remote. To pair them
up again, hold the remote within 6” inches of BOTH insoles, and press the button for one second. The GREEN LED will ash on the
remote as with normal TURN ON operation and the insoles will begin to function. If the GREEN LED on the remote continues to ash,
move the remote closer to BOTH insoles. Wait up to 1 minute for the Footbeat activate.
7. CHARGING:
IMPORTANT
Only use the supplied charger for charging. For international use, make sure you
use the proper outlet converter.
a. Step 1 – Plug the supplied charger into a compatible electrical outlet.
b. Step 2 – Align the wand end of the charger to the charger port on the insole. Once the wand is within two inches of the
charging port, the magnet will pull the wand into place.
c. Step 3 – Once the charging indicator light located on the charging port of the insole has turned green the insole is charged. Pull
the wand away from the charging port.
STEP 1STEP 2STEP 3
IMPORTANT
The rechargeable battery inside the Footbeat is not removable. Attempting to
NOTE – Achieving a full charge will take approximately 1 hour, which will provide 4+ hours of continual use.
NOTE – Two insoles can be charged at the same time. Due to variations in batteries, the insoles may charge at dierent rates.
NOTE – Turn o insoles before applying to the charging unit. They can be turned o using the remote.
NOTE – To turn units on or o, the remote must be within 6 feet of the insoles. If more than 6 feet away, the Footbeat may not turn on
or o properly.
self-service will void the warranty and may cause injury.
Caring For Your Footbeat
NOTE – This system does not require any user maintenance. There are no user repairable components. If a problem occurs, please
contact Footbeat Support. Opening the case on the control unit, with the exception of replacing the battery, or attempting to selfservice any part of the insole will void the warranty and may cause injury.
Routine Cleaning
1. Insole
a. A damp cloth with mild detergent may be used to wipe the outside of the insoles. DO NOT use a strong cleaning solution that
contains bleach. DO NOT put under running water or immerse in water.
b. To keep the charging port and pad clear of debris, sand and dirt, wipe with a wet cloth.
2. MOC
a. Machine wash warm. Tumble dry low. Do not bleach. Do not dry clean. Avoid the use of fabric softeners.
IMPORTANT
REMOVE THE INSOLE BEFORE WASHING THE MOC. The insole is not waterproof.
Washing the insole will cause the system to malfunction, may cause injury and
will void the warranty.
Routine Inspection
Periodically inspect for damage and wear. Check electrical cables and connectors to ensure they are intact and that no conductive wires
are exposed. If damage preventing the safe use of Footbeat is evident, please contact Footbeat Support.
Troubleshooting
If your Footbeat stops working, use your remote to turn the power o, then turn the power on to reset.
If reset does not work, then attempt to charge the battery. See charging instructions. Next attempt to turn the power on.
If charging the unit does not work, try replacing your remote battery (CR2032)
If Footbeat is still not working, contact Footbeat Support at 1-844-388-AVEX or email support@footbeat.com.