Table of Contents ..................................................................................................................................... i
Specifications.............................................................................................................................................. ii
Service Menu Structure ..........................................................................................................................2
Moving Through Menus ...............................................................................................................2
Accessing the Menu ......................................................................................................................3
Exiting the Menu............................................................................................................................3
Using the Service Menu ..........................................................................................................................4
About Menu...............................................................................................................................................4
Audit Menu ................................................................................................................................................5
Test Menu ..................................................................................................................................................6
Setup Menu ................................................................................................................................................ 8
Top Level of 110 Menu ................................................................................................................9
Service Menu .................................................................................................................................... 96-98
Tips on setting up and using the Harmonizer® filtering ...............................14
Tips on setting up and using the Harmonizer® filtering ...............................26
i
Page 3
WI-127 Specifications*
Power requirements:
115 Volts AC, +10% to -15% @ 0.3Amp maximum
230 Volts AC, +10% to -15% @ 0.15 AMP maximum
50/60 Hz
Excitation: 10 Volts DC
Supports up to twelve 350-ohm weight sensors
Operational keys:
Five yellow standard keys: Zero, Tare, Print, Units, Select
Three function keys: F1, F2, F3
Numeric keys: 0-9
Operational annunciators:
Gross, Tare, Net, Print, Zero, Motion
Under, Accept, Over, Cutoff, ID,
Three units of measure
Display: Eight digit, seven segment, 0. 8-inch high LED
Display rate: Selectable (1, 2, 5, 10)
Analog to digital conversion rate: 60 times per second
Unit of measure:
Three, independently programmable:
Pounds, kilograms, grams, ounces, ton, tonne, custom, Off
Standard outputs:
Three outputs, open collector design
Relay power supply, 24 VDC at 150mA
Bi-directional serial port (RS-232 or RS-422/485 or
20mA current loop)
Serial Command Inputs
Programmable serial response to ASCII character input
Self diagnostics:
Display, keys, inputs, outputs, serial port,
A to D converter, loadcell output display, voltages
Circuitry protection: RFI, EMI, and ESD protection
Options:
Two additional serial ports
BCD parallel
10 cutoffs
Analog output
0-5, 0-10 volts
1-5, 4-20, 10-50 mA
Operating temperature:
-40 to 140° F (-40 to 60° C)
100% relative humidity including washdown
Capacity selections:
999,999 with decimal located from zero to five places
Incremental selections:
Multiples and sub-multiples of 1, 2, 5
Programmable selections:
Zero range, motion detection,
automatic zero tracking, five-point linearization.
Time and date / RAM:
Battery backed up real time clock and RAM are standard
Internal resolution: 6,291,456 counts per mV/V per sec.
HarmonizerTM digital filtering:
Fully programmable to ignore noise and vibration
Standard inputs:
Seven configurable logic level inputs for functions such as
PB tare, print, zero, units, select, gross and net.
Enclosure: NEMA 4X stainless steel enclosure
Dimensions:
12" W x 8" H x 4" D (without mounting bracket)
12.3" W x 11.0" H x 5.3" D (with mounting bracket)
Weight: 12.5 lb, 5.7 kg
Agencies:
NTEP Class III/IIIL:10,000d, COC #96-140.A1
Consumer and Corporate Affairs, Canada, #AM-5167
UL/CUL/CSA
FCC Class A
ii
* Specifications subject to change without notice.
Page 4
Introduction
When installing the WI-127, the
power socket-outlet must be nearby
and easily accessible.
This equipment uses double pole/
neutral fusing
This manual covers the service issues for the WI-127 indicator. The manual is
divided into the following sections:
• Introduction
• Error Messages
• Service Menu Structure
• Using the Service Menu
• About Menu
• Audit Menu
• Test Menu
• Setup Menu
• Reset Menu
• Calibrating the WI-127
• Customizing the Serial Output
• WI-127 Service Menu
• Technical Illustrations
Main sections of this manual are set apart by the large black bar as seen above.
Subsections are labeled in the left column of each page. Notes, cautions and
warnings are also listed in the left hand column.
Error Messages
If you find inaccuracies in this manual or have suggestions on how to improve it,
please call 507-238-4461 or 800-368-2039 and ask for a technical writer.
The following are displays you may see if problems occur or if invalid operations
are attempted with your WI-127:
DisplayDescription
O. LoAd
L.C. Error
Overrange weight.
Underrange weight.
Recovering from lockup or out of range condition.
A-D converter is not functioning.
A-D converter subjected to an input signal beyond ±5.00000 mV/V
The unit cannot perform a function. Displayed only while key is held
down.
Corrupted data in the reset menus. See the Reset Menu/Master Clear
section. (* = RESET, SETUP, or CAL)
Displayed while a key is pressed when attempting to modify a sealed
selection without edit privileges.
1
Page 5
Auto. 0
Displayed while waiting for a stable, valid weight to use as a zero
reference on power-up.
Displayed when input voltage to excitation regulator drops below
10.5 VDC. Will clear when input voltage rises above 11.5 VDC.
Service Menu Structure
Moving Through Menus
You configure, calibrate and do testing of the WI-127 using a menu structure
which you move through using directional keys on the front panel. The directional
keys are shown below:
1 Busy
Displayed when the ready/busy handshake has exceeded its time out
limit. Default is 2 seconds. This can also apply to optional 2nd and
3rd serial ports.
Figure 1
Directional keys
Press this key to exit a menu parameter without saving any changes. Use to move
“up” in the menus.
Press this key to end digit entry, accept a change made, or select an item from a
function list. Use to move “down” in the menus.
Press this key to access menus. Use to move “right” in the menus.
Press this key to backspace (deletes the last digit or punctuation mark entered)
while in numeric entry. Use to move “left” in the menus.
2
Page 6
Accessing the
Service Menu
Caution
Do not break any seal on the
indicator unless absolutely necessary.
This may cause the need for the
indicator to be recertified and
resealed.
Entering the service menu disables all outputs and inputs, disables/stops all serial
output, B.C.D. output and analog output.
To enter the service menu structure, key in the default password (127) then press
and hold the ESCAPE key for two seconds. If you do not know the password
you may remove the nylon plug on the back of the WI-127 and press the SEAL
switch inside. If you do not want to make any changes in the service menu but
want to view the items, enter the menu without keying in a password.
If your password has been changed since leaving the factory, enter your current
password instead of the default.
The display should show About. This is the first item in the menu structure. The
following section explains the menu items.
Exiting the
Service Menu
If you do not press SELECT with
Save displayed, none of your
configuration changes will be saved.
You may exit the service menu and return to weight display mode at any time by
pressing SELECT.
If you made changes to the menu, the indicator will display SAVE (asking you if you
wish to save your changes) before returning directly to weight display mode.
To exit and save menu changes:
1.With SAVE displayed, press
SELECT. . .The indicator will return to weight
display mode and your changes are
saved.
To exit without saving changes:
1a. With SAVE displayed, press
MENU. . .no SAVE is displayed.
OR
1b. With SAVE displayed, press
ESCAPE. . .Indicator returns to weight display
mode.
2.Press SELECT. . .Indicator returns to weight display
mode without saving any changes.
3
Page 7
Using the Service Menu
Figures 2 through 19 show the service menu structure in the WI-127. Following
each figure are explanations for each of the service menu items.
The complete service menu structure
can be viewed on the last fold out
pages in the back of this manual.
Figure 2
First menu level
To exit back to normal weighing
mode, press the SELECT key and
save changes as needed by pressing
ENTER with SAVE? displayed.
About Menu
About MenuInformation about the software
Audit MenuAudit counters for calibration and configuration
Test MenuFor testing the hardware of the indicator.
Setup MenuFor setup of the indicator as a 127 or a 110 clone
Figure 3
About menu
These are the items listed in the About menu:
127The unit designation.
12345Parent part number of the software.
-1234The dash portion of the software part number.
Rev. AThe revision level of the software.
4
Page 8
Audit menu
Figure 4
Audit menu
These are the items listed in the Audit menu:
Cal. 0000This is an example of how the calibration audit trail counter will
appear. The actual value will be between 0000 and 9999. It is
nonresettable and may not be edited.
Cfg. 0000This is an example of how the configuration audit trail counter
will appear. The actual value will be between 0000 and 9999. It is
nonresettable and may not be edited. The counter is
incremented each time a metrological item is modified in the
setup menus and saved and so may be used as a control audit.
5
Page 9
Test Menu
Figure 5 shows the
The description of each menu item follows Figure 5.
Test menu. Use this menu to check functions of the WI-127.
Figure 5
Test menu
These are the items listed in the
DisplayPress to start and stop a continuous display test.
With the test stopped press the or key to
move backward or forward one step at a time. Press
again to resume the automatic test or press to stop
the test and return to diSPLAy.
ButtonsThis provides a test of the keypad. The name of the key that is
pressed appears on the display. If no key is pressed, nOnE is
displayed. Press to return to buttonS.
A to DPress this key to see the current A to D value. The displayed
resolution is 1 part in 20,000 per mV/V. This test exists so that
the offset and gain of the electronics can be checked. The offset
is initially set to the nominal offset of the electronics, but you can
press the ZERO key to establish the actual offset, allowing the
gain to be evaluated.
6
Test menu from left to right:
Page 10
To exit back to normal weighing
mode, press the SELECT key and
save changes as needed by pressing
ENTER with SAVE? displayed.
LoadcellWith LOAdCELL displayed, press the
live display of the current counts coming from the loadcell.
400400 is the example shown in Figure 5. Press the UNITS
key to toggle between the counts display mode and the mV/V
display, which appears in this format: 2.00200. This stands for
2.002 mV/V. The decimal indicates you are looking at mV/V and
not current counts.
SerialThis item accesses the internal port serial tests. With SEriAL
displayed, press the key. Port 1 is displayed. This is
always the internal serial port. When the optional serial boards
are installed in the same stack, port two is always on the bottom.
When installed in different stacks (side by side), port two is
always closest to the power supply. Press the
key to select the port you wish to test.
Ready/
BusyWith the port you want to test displayed, press the
key. rEAdy or bUSy is displayed telling you if the hardware
input line is ready or busy. This is useful in tracking down serial
output problems.
Loop/
No LoopPress the
Connect the transmit line to the receive line at some point in
the cabling. The WI-127 checks if it receives the same characters
that it transmits. If it can, LOOP is displayed. If it cannot, noLOOP is displayed. This is useful in isolating serial output
problems to the WI-127, cable, or connected device by looping
back at the corresponding points.
OutputsThis test allows you to check the operation of the onboard and
optional outputs. The onboard outputs are copies of the first
three outputs on the optional I/O board when it is installed.
key to see the Loop - No Loop test.
key to see a
or
With motors connected to external
circuits, this is a good way to test
your hardware.
☞
Sequence This is the first item in the Outputs submenu. Press the
key to test the outputs. Each output is turned on and
off sequentially. The display will show Out. nn. The nn being the
number of the output being tested. The outputs will sequence
every half second. Press the or key to
end the test and return to the SEqUEnCE display.
Out 1-3 or
(Out 1-8)
(Out 9-16) This is the second item in the Outputs submenu. This allows you
to enable or disable any of the outputs 1-3 (1-16 if the optional
boards are installed). Press the key to see the
display of the outputs status.
The screen has 0s and 1s displayed in this format: 00.010000. In
this example, output #4 is active. The zeros and ones represent
the status of each output. A 1 means it is activated and a 0
means it is deactivated. The left digit is output #1 or #9.
To change the status of an output, press the or
key to move the decimal point to the right of the
output you want to change. Press the key to toggle
the output from one status to the other.
Press the key to exit the test. The outputs remain
as selected until you exit the Test menu. Returning to the
normal weigh mode returns the control of the outputs to the
cutoff values.
7
Page 11
This allows you to test your remote
push button switches (ie. remote
zero, print, etc.)
☞
InputsThis test allows you to check the operation of the onboard and
optional inputs. Press the key to access the submenu.
StandardThis submenu item lets you check the status of the onboard
inputs. In this example, 1000100, inputs #1 and #5 are active.
The inputs are ordered 1-8 from left to right. A 1 means
activated and a 0 means deactivated.
As you view the inputs, #8 is actually a flag that is dependent on
the states of inputs 6 and 7. Input 6 resets flag #8 to a false 0.
Input 7 sets flag #8 to a true 1. Input 8 does not terminate at a
connector.
OptionThis menu item is available only if the option board is installed. It
works the same way as the Standard example above.
VoltagesThe submenu under VoltAgES lets you see the power supply
voltages. The voltage readings are updated 2 times per second.
13 voltsThis test displays the unregulated weight sensor excitation
power supply voltage. If the voltage drops below 10.5 the display
will show Lo. Volt. The error condition will not clear until
voltage reaches 11.5.
- 5 voltsThis test displays the -5 volt excitation voltage.
10 voltsThis test displays the unregulated 5 volt logic supply voltage.
Setup Menu
24 voltsThis test display shows the relay supply voltage. If this voltage
drops too far it may not be possible to activate certain relays.
Nominal level for this power supply voltage is 22.8.
Setup is the next top level menu item. See Figure 6.
Figure 6
Setup menu
The two top level menu items under Setup are 110 and 127.
110This selection allows access to the menus necessary to make the
127This selection allows all setup menus to be accessed.
8
WI-127 behave as a clone of the WI-110 indicator.
Page 12
Top Level of 110 Menu
To exit back to normal weighing
mode, press the SELECT key and
save changes as needed by pressing
ENTER with SAVE? displayed.
The top level menu items of the 110 menu are shown in Figure 7.
Figure 7
110 Menu
AdjustUse this submenu for calibration of the scale.
ScaleUse this submenu for configuring units, capacity, divisions, zero,
stability, A.Z.T., update rate, averaging and filtering.
OptionsUse this submenu to configure the buttons on the front panel.
SerialUse this submenu to setup the baud rate, data bits, parity and
stops of the serial ports.
SealUse this submenu to setup a custom password and to set the
sealing choices for the unit.
110 Adjust Menu
Be sure the correct capacity and
division sizes have been selected
before calibrating.
Press the ZERO key to delete a
displayed calibration point.
Below is the 110 Adjust submenu. If the factory calibration has become corrupt,
the word AdJUSt will flash on and off. You may use the indicator under this
condition by calibrating with real weights, not by entering previously recorded
count values. See note at left.
Figure 8
110 Adjust menu
9
Page 13
To exit back to normal weighing
mode, press the SELECT key and
save changes as needed by pressing
ENTER with SAVE? displayed.
Reset is very helpful if you get
confused and would like to put
working values back in the WI-127
calibration menu.
PointsSee the section Calibrating the WI-127 for indepth instructions.
In this submenu you may view, edit, insert or delete calibration
points in a list. The WI-127 allows calibration using up to five
calibration points.
The WI-127 comes from the factory with two calibration points:
0 and 5000 lbs. (These weight values may differ depending on
your unit of measure.)
These points may be calibrated in two ways: 1) by standard
weight calibration or 2) by entering count values. Counts are
calibrated to a 1 mV/V signal from the factory.
A 0When you access the Points submenu the first item is A 0. This
stands for Apply 0 weight. This is the zero calibration point.
A 5000The next calibration point in the default list is the calibration
point for full capacity. Your indicator is factory calibrated to
5000 lbs (1 mV/V). You may keep this value or key in a new
value for full capacity.
DeadloadLets you view the count value for the current deadload. Press
the UNITS key to change the display to mV/V.
DisplayShows the weight display without having to exit the menus. An
example of what is shown—d 5000. The d indicates a live display
of weight and reminds you that the unit is in calibration. The
value you see depends on the selected unit of measure, the
calibration and the weight applied. The UNITS key works in this
mode.
ResetUse this item to reset the calibration points to factory default
values. If you choose yES the values are set to 0 lbs at 0 mV/V
and 5000 lbs at 1 mV/V.
10
Page 14
110 Scale Menu
The next item in the 110 menu structure is Scale. In this group of submenus you
set scale related parameters. Figure 9 shows the Scale menu.
Figure 9
110 Scale menu
UnitsThis submenu lets you select what unit of measure will be
assigned to the three annunciators on the front panel.
Unit 1The unit of measure you choose for this item will be assigned to
the annunciator labeled “lb” on the front panel. You can choose
from the following list of units of measure:
Pounds = pounds
1000g = kilograms
g = grams
Oz = ounces
Tons = tons
Tonnes = tonnes (metric tons)
None = no unit of measure is assigned.
Unit 2The unit of measure you choose for this item will be assigned to
the annunciator labeled “kg” on the front panel. You can choose
a unit of measure from the same list as above.
Unit 3The unit of measure you choose for this item will be assigned to
the annunciator labeled “oz” on the front panel. You can choose
a unit of measure from the same list as above.
11
Page 15
Don’t forget to re-calibrate your
system after you change capacity.
CapacityThe next scale menu item is capacity. Use this item to view or
edit the capacity of the unit in any unit of measure configured
under Units. With CAPACity displayed, press the
key. The current capacity is displayed. Use the numeric keypad
to key in a new value and press to accept it and
return to the CAPACity display.
You can view the capacity in other units of measure by pressing
the UNITS key.
If you change capacity or division
size in any unit of measure, this
automatically changes all the other
enabled units of measure as well.
If you want leading zeroes in your
printouts or broadcast and autosend
information, key in leading zeroes
when setting division size.
To exit back to normal weighing
mode, press the SELECT key and
save changes as needed by pressing
ENTER with SAVE? displayed.
DivisionThis selection allows you to view and edit the division size of the
enabled units of measure. You can enter any division size. The
indicator will use the closest division size for each enabled unit of
measure (1, 2 or 5 divisions).
You can view the division in other units of measure by pressing
the UNITS key. The number is stored in the resolution you
enter but is displayed in the closest valid division size.
ZeroUse this menu item to set pushbutton zero related options.
Zero range is specified as a percent of capacity referenced from
the deadload. There are two items in the submenu: -Range and
Range.
-RangeUse this to set the negative range (below reference zero) within
which the unit may be zeroed. 100% is the default value.
RangeUse this to set the positive range (above reference zero) within
which the unit may be zeroed. 100% is the default value.
StableUse this menu item to set the motion detection parameters.
There are two items in the submenu: Range and Display.
RangeSpecifies the number of ±divisions for the motion window. Use
the numeric keypad to enter appropriate value. Most common
settings are 0.5, 1, or 3 divisions.
A.Z.T.This stands for Automatic Zero Tracking. Range is the only item
UpdateUse this to set the display update rate from these choices:
12
DisplayChoose ON to if you want the display on while the indicator
senses scale motion. Choose OFF to blank the display while
there is scale motion.
in this submenu.
RangeUse this item to set the range within which the indicator will
automatically adjust the zero balance towards zero. Use the
numeric keypad to enter appropriate value. Most common
settings are 0.5, 1, or 3 divisions.
1One update per second.
2Two updates per second
5Five updates per second
10Ten updates per second.
Page 16
AverageThe next menu item is Average. This can be entered in one of
two methods: x disp. or x a-ds. Press the UNITS key to switch
back and forth between the two choices.
The suggested method of setting the average is by picking a value
for x disp. Doing this insures that a multiple of the display rate is
always being averaged. This results in a steadier weight display.
Use x a-ds if you need an exact number of A-D conversions for
your particular situation.
Changing the update rate changes
the x disp. or x a-ds value based on
the new update rate.
Harmonizer® filter is used only if the
average value will not filter out the
vibration by itself.
Values change with unit of measure.
x disp.1 disp. is the default display when you access this item. x is the
number of display interval(s) over which the data is internally
averaged prior to being displayed. The number of A-Ds averaged
is based on the display update rate you set under the Update
menu item. Default is 1 disp.
x a-dsx is the number of a-d conversions to average for each display.
The A-D weight conversion happens 60 times per second in this
indicator. Average is the number of conversions you want to
average for the weight that is displayed. Default is 12 a-ds when
Update is at default of 5 and x disp. is at default value of 1.
FilterUse this menu item to configure the Harmonizer® filter settings.
Constant and Threshld are the two items you can configure in
this submenu.
ConstantThis number represents the amount of filtering. Choose a setting
between 1 and 10. Choose 1 for the least amount of filtering but
the fastest response. Choose 10 for the most filtering but the
slowest response. Choose Off to disable the Harmonizer
functions and default to the lowest filtering.
ThreshldThis is the window, in the current unit of measure, within which
weight changes are altered according to the constant. 10.0 is the
default value shown in Figure 9. You should set the threshold
value between 130 and 150% of the weight swings that need to
be suppressed.
13
Page 17
Tips on setting up and using the Harmonizer® filtering
To find the best settings for your filter needs, follow the steps listed below.
1. What to Do: Determine the amount of positive and negative force
exerted by the vibration on the scale.
How to Do It: Set Threshld to 0.0, Constant to OFF, and Average to
1.0 A-Ds. Return to weigh mode and observe the weight swings. Record
the difference between the highest and lowest displayed weight values.
Add 30 to 50% to this value. This is a good starting value for the
Threshld setting. Do not set your indicator to this value until told to in
step 7.
2. Setting the Average to higher values increases the filtering effect.
What to Do: Set Threshld to 0.0, Constant to OFF and Average to
15.0 A-Ds. Check the stability of the scale.
How to Do It: Save changes and exit to normal weight mode. Observe
the Center of Zero light. If it is on all the time your scale is stable within
¼ division. If the Center of Zero light blinks more filtering is required.
Go to step 3.
3. Repeat step 2 but increase the Average by 15.0 A-Ds. Keep repeating
steps 2 and 3 until the scale is stable or you’ve tried the entire range of
Average (300 A-Ds). If the scale is still not stable go to step 4.
4. Setting the Constant to higher values increases the filtering effect.
What to Do: Set Threshld to 0.0, Constant to 1.0 and Average to 300
A-Ds. Check the stability of the scale.
How to Do It: Save changes and exit to normal weight mode. Observe
the Center of Zero light. If it is on all the time your scale is stable within
¼ division. If the Center of Zero light blinks more filtering is required.
Go to step 5.
5. Repeat step 4 but increase the Constant by 1.0. Keep repeating steps 4
and 5 until the scale is stable or you’ve tried the entire range of Constant (10). If the scale is still not stable, decrease your display update rate
and start over at step 1 using the new, slower display rate.
6. After the Constant value is established you may wish to lower the
Average value to improve display response time.
7. After a final value for Constant and Average has been set, enter the
Threshld value established in step 1. If this value is too small your scale
will act as if the filtering is off or not working. Increase the Threshld
value until your scale stabilizes.
14
If the Threshld value is too high your scale will react slowly to weight
changes.
When Harmonizer is properly adjusted the scale will be stable at zero
and will rapidly display a stable test weight value.
Page 18
110 Options Menu
The third item in the 110 menu structure is
Menu. Use this menu to enable/disable and configure the buttons on the front
panel.
Options. Figure 10 shows the Options
Figure 10
110 Options menu
SelectEnables or disabled the SELECT key.
UnitsEnables or disables the UNITS key.
TareThis parameter enables or disables the pushbutton tare and
numeric tare entry.
PB TareOptions under this item are Yes (allows pushbutton tare for all
nonnegative values), Non-Zero (enables pushbutton tare for all
nonnegative values excluding zero), and No (disables pushbutton
tare).
n TareEnables and disables numeric tare entry.
15
Page 19
Print
PB PrintThis item determines which layouts are sent to which ports for
pushbutton print. Press and Group is displayed.
From this item you may either disable the pushbutton print
option, or select a group to print. To select a group, press
again. Using the keypad, enter the number of the
group you wish to print. To disable, press , then
. The four default print groups are as follows:
To exit back to normal weighing
mode, press the SELECT key and
save changes as needed by pressing
ENTER with SAVE? displayed.
110 Serial Menu
1 = Displayed weight
2 = Gross, Tare, Net (default)
3 = Gross and Net
4 = Net and Tare.
ZeroEnables or disables the ZERO key.
UserEnables or disables the User’s Menu.
F1Enables or disables the F1 key. The default setting for this key
accesses the tare registers.
F2Enables or disables the F2 key. The default setting for this key
accesses the identification number entry.
F3Enables or disables the F3 key. The default setting for this key
accesses the cutoff values.
The fourth item in the 110 menu structure is Serial. This submenu sets up the
baud rate, data bits, parity, and stops of the serial port(s). Figure 11 below outlines
the Serial Menu.
All are sent out
of Port #1 only.
16
Figure 11
110 Serial menu
Page 20
Ports 2 and 3 present if optional
hardware is installed.
A flashing value indicates an
invalid combination.
Port 1, 2 & 3
Port 1 is the onboard serial port. Any additional boards installed
are configured under Ports 2 & 3. Ports 2 & 3 show up in the
Serial Menu only if additional boards are installed.
BusyEnables or disables the ready/busy input. If the ready/busy input
is enabled, you may enter a timeout period. This value determines how long a port can be busy before the indicator displays
a port busy message.
DataChoose between 7 (default) & 8 data bits.
ParityChoices are:
No -Specifies that no parity bit is to be included.
Even - Specifies that a parity bit which insures an even number
of logic one bits is transmitted. (default) (7, 1or 2, even)
Odd - Specified that a parity bit which insures an odd number
of logic one bits is transmitted. (7, 1 or 2, odd)
Clear - Specified that a logic zero bit is always transmitted after
the data bits (space parity). (8, 1, none)
Set -Specifies that a logic one bit is always transmitted after
the data bits (mark parity). (7, 2, none)
StopsSelect the number of stop bits. Choices are 1 or 2.
17
Page 21
110 Seal Menu
The last item in the 110 Setup menu is Seal. Use this submenu to set up a custom
password and to set the sealing choices for the unit. Items in the
(Figure 6) can be protected from unrecorded changes.
Two internal counters record changes to items in the Setup menu. View these
counters under AUDIT in the Service menu (Figure 4). These counters cannot be
reset and thus can be used by auditors or inspectors to check if changes have been
made. One counter is for scale calibration items and the other for configuration
items. The level of protection is set in the Seal menu. The Seal menu is shown in
Figure 12.
Setup menu
To exit back to normal weighing
mode, press the SELECT key and
save changes as needed by pressing
ENTER with SAVE? displayed.
Figure 12
110 Seal Menu
Code No.This item allows you to enter a personalized security code
number. Digits are not shown on the display as you key them in,
so the display prompts you to enter the code number twice.
All & Phys.The two parameters in the SEAL menu are All and Phys. Below
are explanations of the choices you can make.
ALL set to YES
Any time you access the setup menu and change any item in Table 1, the appropri-
ate counter increments one count. Changing multiple items on one visit to the
menu increments the counter only one count. It's the number of visits with
changes that are counted, not the number of changes per visit.
ALL set to NO
With this setting the calibration internal counter will increment when you access
the setup menu and change any calibration item in Table 1. The internal configuration counter will increment only if you change one of the configuration items in
bold print from Table 1.
18
Page 22
Phys. set to Yes
If Phys. is set to Yes, you must remove the physical seal (rear sealing plug) of the
WI-127 to access an internal switch. When you press this switch you have full
editing privileges and the display shows the first item in the Service menu, About,
without the need to enter the password.
If you enter the Service menu using the password and not the internal switch, you
can change only the configuration items in Table 1 that are in normal, not bold,
print.
PHYS set to NO
If PHYS. is set to NO, you can still
press the internal switch and have
instant setup menu access and
editing privileges.
If PHYS. is set to NO, correct password entry is the only way to have editing
privileges of all the items in Table 1 without breaking the physical seal. See note to
left.
If the password is not entered correctly, the setup menu items can be viewed but
not edited.
Calibration ItemsConfiguration Items
Any item in the Adjust menuAny item in the Scale menu
Any item in the Options menu
Any item in the Seal menu
Any item in the Serial menu
Any item in the Analog menu
Any item in the B.C.D. Out menu
Table 1
Calibration and Configuration list
This is the end of the 110 section. The next section covers the 127 section of the
Setup menu.
19
Page 23
Top Level of 127 Menu
To exit back to normal weighing
mode, press the SELECT key and
save changes as needed by pressing
ENTER with SAVE? displayed.
The top level menu items of the 127 menu are shown in Figure 13.
Figure 13
127 Menu
AdjustUse this submenu for calibration of the scale.
ScaleUse this submenu for configuring units, capacity, divisions, zero,
stability, A.Z.T., update rate, averaging, filtering, and overload.
OptionsUse this submenu to configure the buttons on the front panel.
SerialUse this submenu to setup the baud rate, data bits, parity and
stops of the serial ports.
B.C.D. OutUse this submenu to configure the output of an optional B.C.D.
board. This menu will not appear if the optional board is not
installed.
AnalogUse this to configure the analog output option board.
OutputsUse this to configure the WI-127’s outputs.
InputsUse this to configure the WI-127’s hardwired inputs (ie. PB
Tare).
SealUse this submenu to setup a custom password and to set the
sealing choices for the unit.
TransferUse this submenu to setup the WI-127 to receive or send print
formats from the WI-127 downloader software package in your
PC.
20
Page 24
127 Adjust Menu
Be sure the correct capacity and
division sizes have been selected
before calibrating.
A step-by-step description of
calibrating the WI-127 and using
the Adjust menu can be found in the
section titled Calibrating the WI-
127.
Below is the Adjust submenu for the 127. If the factory calibration has become
corrupt, the word AdJUSt will flash on and off. To correct this you must send it
to the factory. You may use the indicator under this condition by calibrating with
real weights, not by entering previously recorded count values. See note at left.
Figure 14
127 Adjust menu
PointsSee the section Calibrating the WI-127 for indepth instructions.
Press the ZERO key to delete a
displayed calibration point.
Reset is very helpful if you get
confused and would like to put
working values back in the WI-127
calibration menu.
12/31/9910:39 AM
lbmV/V
--------- -------00.00000
50002.00000
Serial No.
--------
In this submenu you may view, edit, insert or delete calibration
points in a list. The WI-127 allows calibration using up to five
calibration points.
The WI-127 comes from the factory with two calibration points:
0 and 5000 lbs. (These weight values may differ depending on
your unit of measure.)
These points may be calibrated in two ways: 1) by standard
weight calibration or 2) by entering count values. Counts are
calibrated to a 1 mV/V signal from the factory.
A 0When you access the Points submenu the first item is A 0. This
stands for Apply 0 weight. This is the zero calibration point.
A 5000The next calibration point in the default list is the calibration
point for full capacity. Your indicator is factory calibrated to
5000 lbs (1 mV/V). You may keep this value or key in a new
value for full capacity.
DeadloadLets you view the count value for the current deadload. Press
the UNITS key to change the display to mV/V.
DisplayShows the weight display without having to exit the menus. An
example of what is shown—d 5000. The d indicates a live display
of weight and reminds you that the unit is in calibration. The
value you see depends on the selected unit of measure, the
calibration and the weight applied. The UNITS key works in this
mode.
ResetUse this item to reset the calibration points to factory default
values. If you choose yES the values are set to 0 lbs at 0 mV/V
and 5000 lbs at 1 mV/V.
PrintUse this item to print the calibration information shown in the
left column from port 1 by choosing yes. These numbers can be
valuable if you have an indicator failure.
21
Page 25
127 Scale Menu
The next item in the 127 menu structure is Scale. In this group of submenus you
set scale related parameters. Figure 15 shows the Scale menu.
To exit back to normal weighing
mode, press the SELECT key and
save changes as needed by pressing
ENTER with SAVE? displayed.
22
Figure 15
127 Scale menu
UnitsThis submenu lets you select what unit of measure will be
assigned to the three annunciators on the front panel.
Unit 1The unit of measure you choose for this item will be assigned to
the annunciator labeled “lb” on the front panel. You can choose
from the following list of units of measure:
Pounds = pounds
1000g = kilograms
g = grams
Oz = ounces
Tons = tons
Tonnes = tonnes (metric tons)
Other = This allows you to enter the information to create a
custom unit of measure. You need to enter the ratio
values for the custom unit based upon another unit of
measure and a string of text to use in serial output.
Under Basis choose the unit and amount upon which
the custom unit is based. Under Equals enter the
equivalent custom unit value.
Page 26
Refer to the ASCII character chart
located in the section Customizingthe Serial Output.
Example 1: To create a ‘stone’ unit of measure, the
Basis would be 14 pounds and Equals should be 1
stone.
Example 2: To create an ounce unit of measure, the
Basis would be 1 pound and Equals should be 16
ounces.
Example 3: To create a kg unit of measure, the Basis
would be 10 pounds and Equals should be 4.5359237
kilograms. Ten pounds is used so all eight digits can be
entered as the Equals value.
Under Label you create the serial label for the custom
unit of measure. With the first ASCII value displayed
you can use the keys to do the following:
ENTER key-accepts the displayed value and
returns to Label display.
Custom label can be a maximum of
16 characters in length.
MENU key -accepts the displayed value and moves
to the next ASCII character value.
ESCAPE key - exits the display without making a
change.
+/- key -Inserts a new ASCII value before the
currently displayed ASCII value. Key in
the ASCII value and continue by
pressing another key.
ZERO key -deletes the currently displayed ASCII
value.
Continue editing, inserting, and deleting until you are
done. Press to exit the label entry.
None = no unit of measure is assigned.
Unit 2The unit of measure you choose for this item will be assigned to
the annunciator labeled “kg” on the front panel. You can choose
a unit of measure from the same list as above.
Unit 3The unit of measure you choose for this item will be assigned to
the annunciator labeled “oz” on the front panel. You can choose
a unit of measure from the same list as above.
Don’t forget to re-calibrate your
system after you change capacity.
CapacityThe next scale menu item is capacity. Use this item to view or
edit the capacity of the unit in any unit of measure configured
under Units. With CAPACity displayed, press the
key. The current capacity is displayed. Use the numeric keypad
to key in a new value and press to accept it and
return to the CAPACity display.
You can view the capacity in other units of measure by pressing
the UNITS key. The number is stored in the resolution you
enter but is displayed in the division size stored in Division.
23
Page 27
If you change capacity or division
size in any unit of measure, this
automatically changes all the other
enabled units of measure as well.
DivisionThis selection allows you to view and edit the division size of the
enabled units of measure. You can enter any division size. The
indicator will use the closest division size for each enabled unit of
measure.
If you want leading zeroes in your
printouts or broadcast and autosend
information, key in leading zeroes
when setting division size.
This feature keeps the indicator
from functioning at power up in an
unstable environment.
You can view the division in other units of measure by pressing
the UNITS key. The number is stored in the resolution you
enter but is displayed in the closest valid division size. Any
additional resolution is used in calculating division size in the
other units of measure.
ZeroUse this menu item to set zero related options. Zero range is
specified as a percent of capacity referenced from the deadload.
There are three items in the submenu: -Range, Range and
Start.
-RangeUse this to set the negative range (below reference zero) within
which the unit may be zeroed. 100% is the default value.
RangeUse this to set the positive range (above reference zero) within
which the unit may be zeroed. 100% is the default value.
StartUse this parameter to determine whether or not the indicator
must reach a stable reading within the above range before it will
exit the start-up sequence, automatically zero the scale and begin
weighing. While trying to acquire a stable zero the unit displays
Auto. 0. Select No if you want no start-up zero restrictions.
Choose Yes if you want the start-up zero restriction.
To exit back to normal weighing
mode, press the SELECT key and
save changes as needed by pressing
ENTER with SAVE? displayed.
StableUse this menu item to set the motion detection parameters.
There are three items in the submenu: Range, Delay and
Display.
RangeUse this to specify the number of ±divisions for the motion
window. Default is 1.0 division.
DelayUse this to specify the number of seconds during which the
weight must be within range (described above) before a nomotion condition is displayed. Default value is 0.4 seconds.
DisplayChoose ON to if you want the display on while the indicator
senses scale motion. Choose OFF to blank the display while
there is scale motion. Default is ON.
A.Z.T.This stands for Automatic Zero Tracking. There are two items in
this submenu; Range and Net.
RangeA.Z.T. adjusts the zero balance towards zero at the rate of .1
divisions per second after being within the range given for at
least 2 seconds.
NetYou may also enable or disable automatic zero tracking of net
weight.
UpdateUse this to set the display update rate from these choices:
24
1One update per second.
2Two updates per second
5Five updates per second
10Ten updates per second.
Page 28
Changing the update rate changes
the x disp. or x a-ds value based on
the new update rate.
AverageThe next menu item is Average. This can be entered in one of
two methods: x disp. or x a-ds. Press the UNITS key to switch
back and forth between the two choices.
The suggested method of setting the average is by picking a value
for x disp. Doing this insures that a multiple of the display rate is
always being averaged. This results in a steadier weight display.
Use x a-ds if you need an exact number of A-D conversions for
your particular situation.
x disp.1 disp. is the default display when you access this item. x is the
number of display interval(s) over which the data is internally
averaged prior to being displayed. The number of A-Ds averaged
is based on the display update rate you set under the Update
menu item. Default is 1 disp.
x a-dsx is the number of a-d conversions to average for each display.
The A-D weight conversion happens 60 times per second in this
indicator. Average is the number of conversions you want to
average for the weight that is displayed. Default is 12 a-ds when
Update is at default of 5 and x disp. is at default value of 1.
Harmonizer® filter is used only if the
average value will not filter out the
vibration by itself. See Tips on
setting up and using the
Harmonizer filtering on the next
page.
Values change with unit of measure.
When using overload or over
capacity alone, set the other menu
item to its maximum value to disable
any conflicts.
FilterUse this menu item to configure the Harmonizer filter settings.
Constant and Threshld are the two items you can configure in
this submenu.
ConstantThis number represents the amount of filtering. Choose a setting
between 1 and 10. Choose 1 for the least amount of filtering but
the fastest response. Choose 10 for the most filtering but the
slowest response. Choose Off to disable the Harmonizer
functions and default to the lowest filtering.
ThreshldThis is the window, in the current unit of measure, within which
weight changes are altered according to the constant. 10.0 is the
default value shown in Figure 15. You should set the threshold
value between 130 and 150% of the total weight oscillations that
need to be suppressed. If weight changes are +20 lbs and -10 lbs,
set this value to 130 to 150% of 30 lbs.
OverThis submenu lets you setup the overload and capacity setting
which cause the O. load error message.
OverloadThis value is expressed as a percent of capacity referenced from
the deadload determined by calibration. Overload is restricted to
200% or lower. 105% is the default value.
Over Cap. This value expressed as a number of divisions over the capacity
referenced from the zero value. A value of 9 satisfies UK
requirements. Configurable from 0 to 999999 divisions. Default
is 9 divisions.
25
Page 29
Tips on setting up and using the Harmonizer® filtering
To find the best settings for your filter needs, follow the steps listed below.
1. What to Do: Determine the amount of positive and negative force
exerted by the vibration on the scale.
How to Do It: Set Threshld to 0.0, Constant to OFF, and Average to
1.0 A-Ds. Return to weigh mode and observe the weight swings. Record
the difference between the highest and lowest displayed weight values.
Add 30 to 50% to this value. This is a good starting value for the
Threshld setting. Do not set your indicator to this value until told to in
step 7.
2. Setting the Average to higher values increases the filtering effect.
What to Do: Set Threshld to 0.0, Constant to OFF and Average to
15.0 A-Ds. Check the stability of the scale.
How to Do It: Save changes and exit to normal weight mode. Observe
the Center of Zero light. If it is on all the time your scale is stable within
¼ division. If the Center of Zero light blinks more filtering is required.
Go to step 3.
3. Repeat step 2 but increase the Average by 15.0 A-Ds. Keep repeating
steps 2 and 3 until the scale is stable or you’ve tried the entire range of
Average (300 A-Ds). If the scale is still not stable go to step 4.
4. Setting the Constant to higher values increases the filtering effect.
What to Do: Set Threshld to 0.0, Constant to 1.0 and Average to 300
A-Ds. Check the stability of the scale.
How to Do It: Save changes and exit to normal weight mode. Observe
the Center of Zero light. If it is on all the time your scale is stable within
¼ division. If the Center of Zero light blinks more filtering is required.
Go to step 5.
5. Repeat step 4 but increase the Constant by 1.0. Keep repeating steps 4
and 5 until the scale is stable or you’ve tried the entire range of Constant (10). If the scale is still not stable, decrease your display update rate
and start over at step 1 using the new, slower display rate.
6. After the Constant value is established you may wish to lower the
Average value to improve display response time.
7. After a final value for Constant and Average has been set, enter the
Threshld value established in step 1. If this value is too small your scale
will act as if the filtering is off or not working. Increase the Threshld
value until your scale stabilizes.
26
If the Threshld value is too high your scale will react slowly to weight
changes.
When Harmonizer is properly adjusted the scale will be stable at zero
and will rapidly display a stable test weight value.
Page 30
127 Options Menu
The third item in the 127 menu structure is Options. Figure 16 below shows the
Options Menu. Use this menu to configure the keys on the front panel as well as
define print layouts.
Figure 16
127 Options Menu
27
Page 31
Buttons
SelectEnables or disabled the SELECT key.
UnitsEnables or disables the UNITS key.
TareThis parameter enables or disables the pushbutton tare and
numeric tare entry.
PB Tare Options under this item are Yes (allows pushbutton
tare for all nonnegative values), Non-Zero (enables
pushbutton tare for all nonnegative values excluding
zero), and No (disables pushbutton tare).
n TareEnables and disables numeric tare entry.
Print
PB Print This item determines which layouts are sent to
which ports for pushbutton print. Press
and Group is displayed. From this item you may
either disable the pushbutton print option, or select
a group to print. To select a group, press
again. Using the keypad, enter the number of the
group you wish to print. The four default print
groups are as follows:
1 = Displayed weight
2 = Gross, Tare, Net
3 = Gross and Net
All are sent out
of Port #1 only.
4 = Net and Tare.
n PrintThis item enables or disables group print entry. If
enabled, you may print any group from the front
panel. To use this feature during normal weighing
operations, key in the number of the group you wish
to print, then press PRINT.
ZeroEnables or disables the ZERO key.
UserEnables or disables access to the User’s menu.
F1Enables or disables the F1 key. The default setting for this key
accesses the tare registers.
F2Enables or disables the F2 key. The default setting for this key
accesses the identification number entry.
F3Enables or disables the F3 key. The default setting for this key
accesses the cutoff values.
DisplayThe Display submenu allows you to customize the order in
If Net is removed from Select mode,
the SELECT key cannot enable net
mode, but taring a nonzero value
will put the indicator temporarily into
the net mode.
28
which the WI-127 cycles through its weight display modes and
units of measure.
SelectCustomize the list of weight display modes here. The default list
is: Gross!Tare!Net. Repeatedly pressing SELECT while in
the weight display mode will cycle through the list in that order.
You may customize the list by inserting and/or deleting items in
the list. The list can contain a maximum of four items.
Page 32
To exit back to normal weighing
mode, press the SELECT key and
save changes as needed by pressing
ENTER with SAVE? displayed.
To add an item: Determine where in the list you wish to insert
an item. Display the item which will immediately follow the one
you are inserting. Press +/-. The ACCEPT annunciator illuminates. Cycle through the options by pressing the and
keys until the item you wish to insert is displayed, then
press
. That item is now included in the list.
To delete an item: With that item displayed, press ZERO. That
item is deleted from the list.
UnitsCustomize the list of units of measure here. The default list is:
Unit 1!Unit 2!Unit 3. Repeatedly pressing UNITS while
displaying a weight in the weight display mode will cycle through
the list in that order. You may customize the list by inserting
and/or deleting items in the list. The list can contain a maximum
of four items.
To add an item: Determine where in the list you wish to insert
an item. Display the item which will immediately follow the one
you are inserting. Press +/-. The ACCEPT annunciator illuminates. Cycle through the options by pressing the
and
keys until the item you wish to insert is displayed, then
press . That item is now included in the list.
To delete an item: With that item displayed, press ZERO. That
item is deleted from the list.
DefineRefer to the section “Customizing your Serial Output”
for detailed instructions about this submenu.
The Define submenu allows you to customize printouts, build
ASCII strings, create groups and enable continuous send.
ASCIIThe WI-127 can store up to sixteen ASCII strings, each contain-
ing up to 32 individual ASCII characters. The WI-127 contains
one default ASCII string. You may customize this string as well as
build fifteen additional ones.
LayoutsThe WI-127 has sixteen available print layouts (of which eight
have default settings) which may include a combination of sixteen
items. Build your custom layouts by choosing among nineteen
print items, including ASCII strings, weight values, time and date,
and other layouts.
GroupsHere is where you specify which ports will print which layouts. A
total of nine groups is available. Within each group, up to three
ports can output layouts. Port 1 is standard; ports 2 & 3 are
offered only if additional serial boards are installed. Each port is
then assigned a layout to print.
Broadcast Enables or disables continuous send. To enable continuous send,
a group number must be entered under GrouP. This number
defines the group to be activated at each display update.
Autosend Enables or disables autosend. To enable auto send, a group
number must be entered under GrouP. This number defines the
group to be printed each time weight stabilizes above 1%
capacity. The weight must fall below 1% capacity for the indicator
to initiate another print.
29
Page 33
Except for Reply the actions at
right can only be executed if the
character is received while the
indicator is in the Display mode.
Protocol A protocol is defined as the set of input characters assigned to specific
actions for each port. Under this item you choose a port, choose an
input number, assign an ASCII character (decimal value) to the action,
and choose the associated response or action from the following list
(see note at left):
Print -Causes a Print to occur. You can specify which group to
output.
Reply -Causes a Print to occur. You can specify which group to
output. This runs as a background fuction without regard
for motion. Reply can be executed when the indicator is
in Display mode and also in the User Menu.
Zero -Performs a Zero function.
None -Assigns no function.
Units -Switches indicator to the next available unit of measure.
Unit 1 -Switches indicator to Unit 1, if available.
Unit 2 -Switches indicator to Unit 2, if available.
Unit 3 -Switches indicator to Unit 3, if available.
Select -Switches indicator to next available display mode.
Gross -Switches display to the gross mode.
Net -Switches display to the net mode.
Tare -Switches display to the tare mode.
P.B.Tare - Causes a pushbutton tare operation.
Example:This is how you would set up the indicator if you want to zero the
scale through a serial port from a computer or PLC:
Select a port (1, 2 or 3) under Protocol.
Connect a cable between this port and your computer or PLC.
Configure a serial Input to perform an action. A maximum of 16
serial inputs may be configured.
Under the selected input, under Ch (Channel), choose an ASCII
character, for example Z, represented by decimal value 90. See Table
5 in the WI-127 Service Manual for a complete list of ASCII characters.
30
Go over from Ch to Action. Under Action drop down and select
Zero from the menu choices.
Now when you send a Z from your PC or PLC to the serial port,
this will cause the display to zero out, the same as if you pressed the
front panel ZERO key.
Page 34
127 Serial Menu
The fourth item in the 127 menu structure is Serial. Figure 17 below shows the
Serial Menu. Use this menu to configure serial ports and communications protocols.
Figure 17
127 Serial Menu
Ports 2 and 3 present if optional
hardware is installed.
By default the WI-127 ports accept
an enquire (ENQ) (05) character. If
an ENQ character is received, the
group selected for the pushbutton
print will be output. This is configurable under Define, Protocol
for each serial port separately.
A flashing value indicates an
invalid combination.
Port 1, 2 & 3Port 1 is the onboard serial port. Any additional boards installed
are configured under Ports 2 & 3. Ports 2 & 3 show up in the
Serial Menu only if additional boards are installed.
BusyEnables or disables the ready/busy input. If the ready/busy input
is enabled, you may enter a timeout period. This value determines how long a port can be busy before the indicator displays
a port busy message.
DataChoose between 7 (default) & 8 data bits.
ParityChoices are:
No -Specifies that no parity bit is to be included.
Even - Specifies that a parity bit which insures an even number
of logic one bits is transmitted. (default) (7, 1or 2, even)
Odd - Specified that a parity bit which insures an odd number
of logic one bits is transmitted. (7, 1 or 2, odd)
Clear - Specified that a logic zero bit is always transmitted after
the data bits (space parity). (8, 1, none)
Set -Specifies that a logic one bit is always transmitted after
the data bits (mark parity). (7, 2, none)
StopsSelect the number of stop bits. Choices are 1 (default) or 2.
31
Page 35
127 BCD Out Menu
To exit back to normal weighing
mode, press the SELECT key and
save changes as needed by pressing
ENTER with SAVE? displayed.
The next item in the 127 menu structure is B.C.D. Out. Figure 19 below shows
this menu. This submenu configures what the BCD output board will print. It is
offered only if the BCD board is installed. Choices are: off, displayed weight, gross
weight, and net weight.
If you choose displayed weight, BCD will output gross weight or net weight
depending on what you are using at the time. If you access the user menu, the
BCD output will be whatever was last displayed. If you use the SELECT key to
access the tare display mode, BCD output will be Gross if tare = 0, Net if tare ≠ 0.
Figure 18
127 B.C.D. Out menu
127 Analog Menu
The Analog submenu is next. This submenu configures the analog output option
board. Choices are: off, displayed weight, gross weight, and net weight.
If you choose displayed weight, the analog board will output gross weight or net
weight depending on what you are using at the time. If you access the user menu,
the analog output will be whatever was last displayed. If you use the SELECT key
to access the tare display mode, analog output will be Gross if tare = 0, Net if tare ≠ 0.
Figure 19
127 Analog menu
OutputThis item specifies which weight the analog output will follow.
FullWhen selected, the indicator will display the last value used or
32
Choices are: off, displayed weight, gross weight, or net weight.
the default value. You enter the full capacity of the analog output
which may be less than or greater than the capacity of the scale.
For example, the capacity of the indicator might be 5000 lb, but
it may be desirable to have 3000 lbs as the full capacity of the
Page 36
analog output. In any case, the analog output has nominal under
and over range limits of 20%. After entering this value, press
UNITS to select the unit of measure for the value. (This
selection is not offered if off is selected under Output.)
AdjustThese choices—Zero and Full—allow the zero and the span of
the analog outputs to be adjusted without actually putting
weights on and off the scale.
Selecting Zero lets you adjust the zero of the analog output for a
zero weight reading. Adjust the zero by entering a percent value
or by pressing and to make incremental
adjustments. The number on the display gives a visual representation of the zero setting with 00.000 being the nominal value
The zero adjustment has a +/- 10% range, -10.000 to +10.000 on
the display.
Selecting Full lets you adjust the span of the analog output for
the full capacity weight reading. Adjust the span by entering a
percent value or by pressing
incremental adjustments. Holding these keys will cause the values
to change faster and faster. The number on the display gives a
visual representation of the zero setting with 100.000 being the
nominal value The zero adjustment has a +/- 10% range, 90.000
to 110.000 on the display. Weight does not have to be on the
scale to perform this task.
and to make
127 Outputs Menu
* denotes default settings
This submenu, shown in Figure 20, is where you configure the WI-127’s outputs.
UpdateChoose an update display rate for the cutoffs.
Display*This selection updates the cutoffs at the indicator’s configured
display update rate:
1, 2, 5 or 10 times per second.
FastThis selection updates the cutoffs at 20 times per second. The
indicator will continue to update at its configured display update
rate.
OnboardChoose the configuration of the onboard outputs.
Cutoff*This selection configures the onboard outputs as cutoffs.
BoundsThis selection configures the onboard outputs as over, under and
accept outputs.
Figure 20
Outputs menu
33
Page 37
127 Inputs Menu
Use this submenu, shown in Figure 20a, to assign actions to the onboard hardware
inputs. The action for each input is selectable from this list: Print; Reply; Zero;
None; Units; Unit 1; Unit 2; Unit 3; Select; Gross; Net; Tare; P.B. Tare. See the
explanations for each of these at the bottom of the page.
Figure 20a
127 Inputs Menu
See the table below for the default assignments. Use this submenu, shown in
Figure 20a, to assign actions to the onboard hardware inputs. See the table below
for the default assignments.
Except for Reply the actions at right
can only be executed if the character
is received while the indicator is in
the Display mode.
Print -Causes a Print to occur. You can specify which group to output.
Reply -Causes a Print to occur. You can specify which group to output. This
runs as a background fuction without regard for motion. Reply can be
executed when the indicator is in Display mode and also in the User
Menu.
Zero -Performs a Zero function.
None -Assigns no function.
Units -Switches indicator to the next available unit of measure.
Unit 1 -Switches indicator to Unit 1, if available.
Unit 2 -Switches indicator to Unit 2, if available.
Unit 3 -Switches indicator to Unit 3, if available.
Select -Switches indicator to next available display mode.
Gross -Switches display to the gross mode.
Net -Switches display to the net mode.
Tare -Switches display to the tare mode.
P.B.Tare -Causes a pushbutton tare operation.
34
Page 38
127 Seal Menu
The next item in the 127
password and to set the sealing choices for the unit. The Seal Menu is shown in
Figure 21.
Setup menu is Seal. Use this submenu to set up a custom
Figure 21
127 Seal Menu
To exit back to normal weighing
mode, press the SELECT key and
save changes as needed by pressing
ENTER with SAVE? displayed.
Items in the
Two internal counters record changes to items in the Setup menu. View these
counters under AUDIT in the Service menu (Figure 4). These counters cannot be
reset and thus can be used by auditors or inspectors to check if changes have been
made. One counter is for scale calibration items and the other for configuration
items. The level of protection is set in the Seal menu. The Seal menu is shown in
Figure 22.
Setup menu (Figure 6) can be protected from unrecorded changes.
Figure 22
110 Seal Menu
Code No.This item allows you to enter a personalized security code
number. Digits are not shown on the display as you key them in,
so the display prompts you to enter the code number twice.
All & Phys.The two parameters in the SEAL menu are All and Phys. Below
are explanations of the choices you can make.
35
Page 39
All set to Yes
Any time you access the setup menu and change any item in Table 2, the appropri-
ate counter increments one count. Changing multiple items on one visit to the
menu increments the counter only one count. It's the number of visits with
changes that are counted, not the number of changes per visit.
All set to No
With this setting the calibration internal counter will increment when you access
the setup menu and change any calibration item in Table 2. The internal configuration counter will increment only if you change one of the configuration items in
bold print from Table 2.
Phys. set to Yes
If Phys. is set to Yes, you must remove the physical seal (rear sealing plug) of the
WI-127 to access an internal switch. When you press this switch you have full
editing privileges and the display shows the first item in the Service menu, About,
without the need to enter the password.
If you enter the Service menu using the password and not the internal switch, you
can change only the configuration items in Table 2 that are in normal, not bold,
print.
If Phys. is set to No, you can still
press the internal switch and have
instant setup menu access and
editing privileges.
Phys. set to No
If Phys. is set to no correct password entry is the only way to have editing
privileges of all the items in Table 2 without breaking the physical seal. See note to
left.
If the password is not entered correctly, the setup menu items can be viewed but
not edited.
Calibration ItemsConfiguration Items
Any item in the Adjust menuAny item in the Scale menu
Any item in the Options menu
Any item in the Seal menu
Any item in the Serial menu
Any item in the Analog menu
Any item in the B.C.D. Out menu
Table 2
Calibration and Configuration list
127 Transfer Menu
36
Use this submenu to setup the WI-127 to receive or send print formats from the
WI-127 downloader software package in your PC. See Figure 21.
This is the end of the 127 section.
Page 40
Reset Menu / Master Clear
User Master Clear as a last
resort before sending in the
unit for repair.
You must press the Seal switch if the
unit is physically sealed.
The reset menu may not contain all
three items shown in Figure 23. If an
item is at default and not corrupted
it will not appear in the menu.
Corrupted items will flash. Items not
at default and not corrupted will
appear but will be solid (not
flashing). These items are not
required to be reset.
If the SEAL PHYS. selection is
corrupted, the unit assumes that the
selection is YES.
The reset menu shown in Figure 23 appears in two cases.
1.If you do a Master Clear (powering up the unit with both the MENU and
F2 keys pressed).
2.If setup, calibration, or data becomes corrupted.
In case 1, you will need to enter the password the same way as explained in the
Service menu. After correctly entering, the reset menu will be displayed.
1. With Reset displayed, press the
key. . .The first menu item will be displayed.
See note at left.
2. Press the key. . .No is displayed.
3. Choose Yes* to reset to default
values or No to leave the values as
they are. Toggle between the
choices with the
key. When the choice
you want is displayed, press the
key. . .If you reset the item, the display will
or
* If the unit is physically sealed, you
must press the internal switch to
select YES.
show the next item. When you reset
an item to defaults it disappears from
the menu.
You can press the ESCAPE key to
exit the Reset Menu when you don’t
want to reset any more items.
A software change requires a reset.
You must press the seal switch to
force the reset to occur properly.
4. Repeat steps 2 and 3 for each item
in the menu. . .When all items are either accepted or
reset the indicator reboots automatically.
Figure 23
Reset menu
In case 2, the display bypasses the password and goes right to Reset. Repeat steps
1 through 4 above.
37
Page 41
Calibrating the WI-127
The WI-127 allows calibration using up to five calibration points. These points can
be any weight value in any unit of measure. Standard calibration generally uses two
calibration points; for linearization, more than two may be used.
The WI-127 comes from the factory with two calibration points: 0 and 5000 lbs.
(These weight values may differ depending on your unit of measure.) To perform
Any changes you make within the
calibration menu will be immediately
implemented, so take care when
recalibrating your system.
While in the Calibration menu, you
may print the calibration data out of
port #1 by selecting YES. See
sample below.
12/31/9910:39 AM
lbmV/V
--------- -------00.00000
50002.00000
Serial No.
--------
linearization, you can insert more calibration points (up to a total of five points).
These points appear in a list and may be inserted and deleted. The unit will order
the points based on increasing count values.
These points may be calibrated in two ways: 1) by standard weight calibration or
2) by entering count values. Counts are calibrated to a 1 mV/V signal from the
factory. This allows you to view the deadload, calibration point counts, and
loadcell test in our standard count value (200,000 counts per mV/V) or press
UNITS to view and edit the real mV/V. There are benefits to this feature:
•You can enter these values into a new indicator hooked up to the same
scale. This is useful if the old indicator needs servicing and a quick
turnaround is needed.
•Another benefit is the ability to enter the profile of a weight sensor
without having to calibrate the indicator conventionally. This profile is
used mostly in batching bars and force measurement devices, which have
a standard or known output and are more difficult to calibrate in the
field.
Entering the
Calibration menu
Be sure the correct capacity and
division sizes have been selected
before calibrating.
To exit back to normal weighing
mode, press the SELECT key and
save changes as needed by pressing
ENTER with SAVE? displayed.
To calibrate your indicator you must enter the calibration menu. See note at left
before following these instructions:
1.From weight display mode, key
in the security code (default code
is 1 2 7). . .The code number is displayed.
2a. Press and hold for
two seconds. . .About is displayed.
OR
2b. Press the SEAL switch inside the
WI-127. . .About is displayed.
3.Press . . .SEtUP is displayed.
4.Press . . .110 or 127 is displayed.
5.Press . . .AdJuSt is displayed.
6.Press . . .PointS is displayed. You are now in the
calibration menu.
38
Page 42
Weight Calibration
To calibrate your scale using live weight calibration, follow these steps:
1.With PointS displayed,
press . . .A 0 is displayed. This is the zero
calibration point.
To view or edit the weight in another
configured unit of measure, press
the UNITS key at any time during
calibration.
If the scale is too unstable, BUSY
will be displayed for several seconds.
Correct the filtering and recalibrate.
2.To calibrate your scale’s zero
point, press
3.Remove all weight from the scale
and press
4.Press . . .A 0 is displayed.
5.Press . . .A 5000 is displayed. This is the full
6.You may use 5000 lbs to calibrate
this point, or you may change the
value for this calibration point.
To change this calibration point,
key in the new value now and
press or to use the
current value, press . . .CAL. is displayed.
. . .CAL. is displayed.
. . .bUSy is displayed for at least ½ second
while the unit obtains a stable value,
then d 0 is displayed. Note: Pressing
while bUSy is displayed
will abort the calibration and the
indicator will return to the previous
display.
capacity calibration point. Full capacity
is factory calibrated at 1 mV/V input..
After your system is fully calibrated,
write down and save the COUNT
values for each calibration point. If
your indicator ever needs replacing
you can key these values into your
new indicator and be assured the
calibration will be correct.
To exit back to normal weighing
mode, press the SELECT key to
save changes.
7.Put the appropriate weight
calibration value on the scale and
press . . .bUSy is displayed for at least ½ second
while the unit obtains a stable value,
then d XXXX is displayed.
Note: Pressing while
bUSy is displayed will abort the
calibration and the indicator will return
to the previous display.
8.Press . . .A XXXX is displayed. You have now
calibrated the two standard calibration
points using live weight.
9.Press SELECT to save your
calibration.
39
Page 43
Count Calibration
To view or edit the weight in another
configured unit of measure, press
the UNITS key at any time during
calibration.
To calibrate your scale using count calibration, follow these steps:
1.With PointS displayed,
press
2.Press
3.Press . . .CountS is displayed.
. . .A 0 is displayed. This is the zero
calibration point.
. . .CAL. is displayed.
4.Press
5.Key in the count value for the
zero calibration point. . .Value is displayed.
6.Press
7.Press . . .A 0 is displayed.
8.Press . . .A 5000 is displayed. This is the full
9.a You may leave this point at
5000 lbs, or you may change the
value for this calibration point.
To change this calibration point,
key in the new value now and
press . . .CAL. is displayed.
. . .Current count value is displayed. Press
UNITS to view and edit the count in
mV/V.
. . .Value is accepted and CountS is
displayed.
capacity calibration point. Full capacity
is factory calibrated at 1 mV/V input..
After your system is fully calibrated,
write down and save the COUNT
values for each calibration point. If
your indicator ever needs replacing
you can key these values into your
new indicator and be assured the
calibration will be correct.
To exit back to normal weighing
mode, press the SELECT key to
save changes.
40
9.b To leave the point at 5000
lbs, press . . .CAL. is displayed.
10. Press . . .CountS is displayed.
11. Press . . .Current count value is displayed. Press
UNITS to view and edit the count in
mV/V.
12. Key in the count value for the
A XXXXX calibration point. . .Value is displayed.
13. Press . . .Value is accepted and CountS is
displayed.
14. Press . . .A XXXX is displayed. You have now
calibrated the standard two calibration
points using count calibration.
15. Press SELECT to save your
calibration.
Page 44
Adding Calibration
Points
You have the option of adding one, two or three additional calibration points for
linearization. You may add these points at the same time you are calibrating the
zero load and full capacity points. Points do not have to be inserted in the correct
order. The WI-127 will automatically order the points based on count values.
To add linearization points:
1.With A XXXX displayed,
press +/- . . .A __ is displayed.
2.Key in the calibration value for the
new point. . .A XXXX is displayed.
3.Press . . .CAL. is displayed.
To perform a live weight calibration:
4a. 1. Put the correct weight on the
scale and press
. . .bUSy is displayed for at least ½ second
while the unit obtains a stable value,
then d XXXX is displayed.
To exit back to normal weighing
mode, press the SELECT key to
save changes.
Deleting Calibration
Points
2. Press
To perform a count calibration:
4b. 1. Press . . .CountS is displayed. Press UNITS to
2. Key in the correct count and
press . . .CountS is displayed.
3. Press . . .A XXXX is displayed.
5.To add more calibration points,
repeat steps 1-4 above.
6.Press SELECT to save changes.
There are two methods of deleting calibration points. NOTE: You may not have
less than two calibration points.
. . .A XXXX is displayed.
view and edit the count in mV/V.
Method A
1.With the point you wish to delete
displayed. . .A XXXX
2.Press ZERO. . .Point is deleted.
Method B:
1.With the point you wish to delete
displayed. . .A XXXXX
41
Page 45
2.Press
. . .CAL. is displayed.
3.Press . . .CountS is displayed.
4.Press again. . .dELEtE is displayed.
To exit back to normal weighing
mode, press the SELECT key to
save changes.
5.Press
6.Press
. . .no is displayed.
. . .yES is displayed.
7.Press . . .The point is deleted and A XXXX
(next calibration point) is displayed.
8.Press SELECT to save changes.
42
Page 46
Customizing the Serial Output
Predefined
Print Layouts
The WI-127 has sixteen available print layouts. Nine have default settings. Examples of the nine default layouts are shown below in printout form. Their actual
layout codes are shown on the following pages.
Pressing the PRINT key on a new indicator will automatically print Layout #6.
Layout 1Prints the weight label, gross weight, and unit of measure.
Layout 2Prints the tare register number, weight label, tare weight, and
unit of measure.
Layout 3Prints the tare register number, weight label, net weight, and
unit of measure.
Layout 4Prints the weight label, displayed weight, and unit of measure.
Layout 5Prints Layout 4 (weight label, displayed weight, and unit of
measure) plus a form feed.
Layout 6This is the default printout when the PRINT key is pressed.
Prints Layouts 1, 2 & 3 (tare register numbers, weight labels,
gross, tare and net weights, and units of measure) plus a form
feed.
and tare weights, units of measure) plus a form feed.
Layout 9Prints weight label, displayed weight, unit of measure, time and
date.
43
Page 47
Use the key to move
down the list of printable items in
Table 3. Use the key move
up the list.
Just as in other Weigh-Tronix indicators (WI-125, WI-150, etc.) the layouts within
the WI-127 may be customized. If the nine default layouts do not fit your specific
applications, or if you wish to include, for example, custom wording, you may
easily customize a print layout. This is done within the “Options” menu of the
service menu. See the section Step by Step Instructions / Entering the Define Submenu
located just before Figure 24. But please read the following section in order to
understand what items are available for customization.
The WI-127’s layouts can be a maximum of sixteen items long. Any combination
of the following nineteen items (up to sixteen maximum) may be used to form a
layout. These printable items, along with their descriptions, are listed below.
Table 3: Layout Print Items (Blocks)
Print ItemsDescription
Status byte = 8 bits
Transmitted as a single character.
The bits appear as follows: 0011LEBM
L has a weighted value of 8
E has a weighted value of 4
B has a weighted value of 2
M has a weighted value of 1
where L is set to logic 1 when a Low
voltage condition exists; logic 0
otherwise. E is set to 1 when an a-d
Error condition exists; 0 otherwise. B
is 1 when the weight is beyond
displayable range (over- or undercapacity); 0 otherwise. And M is set
to logic 1 when an in-motion
condition exists; set to logic 0 when
the weight is stable. The upper four
bits are set to 0011 to cause the
value to be printed as a digit or
symbol in row 3 of the ASCII
character set.
These are the most common characters you will see on a terminal:
“0” = Stable
“1” = In-motion
“2” = Range error
“4” = A-D error
“8” = Low voltage
END and DELAY are not
printable items.
GROSSGross weight is transmitted.
NETNet weight is transmitted.
DISPLAYCurrent displayed live weight is transmitted (gross or net),
depending on the current display mode.
TARECurrent tare (general or numbered register) is transmitted.
T. REGCurrent tare register number is transmitted (a space is
transmitted for the general tare register).
SELECTEDCurrently selected item from the select list is transmitted
(gross, tare or net).
IDIdentification number is transmitted.
HOURCurrent time is transmitted.
DAYCurrent date is transmitted.
STATUSCurrent status is transmitted. See note at left.
* LABELWeight label is transmitted. Five choices are available:
• gross-outputs the gross label
• net-outputs the net label
• display-outputs gross or net depending on the displayed live
weight
• tare-outputs the tare label
• selected-outputs the label gross, tare or net following
SELECTED item
UNITSUnit of measure label is transmitted
* LAYOUTA predefined layout may be included within another layout.
For example, Layout 6 includes Layouts 1, 2, & 3.
Note: A “layout error” will occur if a layout uses its own
layout within itself or if a “loop” of layouts is used (for
example, Layout 1 cannot use Layout 2 if Layout 2 includes
Layout 1)
* DELAYPauses serial port output by this many seconds (0-255)
ENDInsert to delete any printable items that follow
* ASCIIA configurable ASCII string can be inserted and transmitted
* Ch.A character (ASCII value) can be entered to be transmitted
* SPACESA configurable number of spaces is transmitted
* CRSA configurable number of carriage returns is transmitted
* LFSA configurable number of line feeds is transmitted
FFA form feed is transmitted
44
* denotes items that require detailed information. For example: if “LABEL” is
inserted within a layout, you must specify the label to be printed; if “SP” is
inserted, you must specify the number of spaces to be transmitted.
Page 48
Listed below are the predefined print layouts. You may customize these, or create
up to seven new layouts.
Layout 1
Layout 2
Layout 3
Layout 4
Anyplace you see in a
layout at right, you will actually see
on your indicator.
Layout 5
Layout 6
Layout 7
Layout 8
Layout 9
45
Page 49
ASCII Strings
As mentioned in the previous section, the layouts may include a number of items,
one of which is ASCII strings. These ASCII strings allow you to customize your
printouts with custom wording.
ASCII is an acronym for American Standard Code for Information Interchange.
ASCII codes are simply numbers (code values) a computer can translate into
letters, numbers and actions.
The WI-127 can store up to sixteen ASCII strings, each containing up to 32
individual characters. These strings are numbered 1-16. Below is a worksheet to
help define several ASCII strings. Write in your custom wording in the white
boxes, then convert the characters to ASCII code values using the table on the
next page. Write these values in the gray boxes. If a letter or action is repeated
several times, place a decimal point after the code value, then insert the number of
times it is to be repeated. For example, 13.3 means three successive carriage
returns.
The WI-127 contains one default ASCII string. String #1 will spell out “WEIGHTRONIX WI-127” when added to a customized layout. Feel free to fill in the
worksheet below to keep track of your custom defined ASCII strings.
Complete instructions for programming these ASCII strings into the WI-127 can
be found in the section Step by Step Instructions preceding Figure 24.
46
Page 50
Table 5 ASCII Control Code Values
NOTENOTE
NOTE: To repeat a control code a number of times, enter the control code #, a decimal, then the number of times you
NOTENOTE
want it repeated. Spaces, letters, or carriage returns can easily be repeated this way.
47
Page 51
Step by Step
Instructions
To customize layouts and ASCII strings you must enter the Define submenu of the
Service Menu, then follow the flowchart pictured below.
Follow these instructions:
Entering the Define
submenu
1.From weight display mode, key
in the security code (default code
is 1 2 7). . .The code number is displayed.
2a. Press and hold for
two seconds. . .About is displayed.
OR
2b. Press the SEAL switch inside the
WI-127. . .About is displayed.
3.Press . . .SEtUP is displayed.
4.Press . . .110 or 127 is displayed.
If 110 is displayed, press
. . .127 is displayed.
5.Press . . .AdJuSt is displayed.
6.Press . . .SCALE is displayed.
7.Press . . .OPtionS is displayed.
8.Press . . .buttonS is displayed.
9.Press . . .diSPLAy is displayed.
10. Press . . .dEFINE is displayed. You are now in
the Define submenu shown below.
48
Figure 24
Define menu
Page 52
Creating ASCII strings
It is within the ASCII submenu that you create/customize the ASCII strings to be
used in your layouts.
Label
The default labels are: “G” for gross,
“N” for net, “T” for tare, and ID for
ID. You can customize the labels
and use up to 16 characters.
This item allows you to edit the gross, net, tare, and ID labels.
1.With LAbEL displayed,
press . . .GroSS is displayed.
2.Press again. . .71 (the ASCII value for the letter “G”)
is displayed.
3a. Press if you don’t wish
to make a change or your are done.GroSS is displayed.
or
3b. To change this label, press the
+/- key. . .A — (dash) is displayed.
Key in your next ASCII value(s)
and press . . .ENd is displayed.
4.Press . . .NEt is displayed.
5.Repeat steps 1-4 for Net,
Tare, and ID.
49
Page 53
Strings
This item allows the creation of up to sixteen different ASCII strings. On a new
indicator, there is one default ASCII string. You may edit the existing string as well
as create up to fifteen more.
Viewing strings
Creating new strings
1.With StringS displayed,
press
. . .ASCII 01 (default string) is displayed.
2.Press . . .The first ASCII value in the string is
displayed.
3.Press and to
scroll through the entire string of
ASCII values.
4.When you are finished viewing
the values, press . . .ASCII 01 is displayed.
5.Press . . .On a new indicator, End is displayed.
This means that there is only one ASCII
string (ASCII 01) defined so far. You
may create up to fifteen more.
1.With StringS displayed,
press
. . .ASCII 01 (string #1) is displayed.
2.Press until . . .End is displayed.
To delete the displayed ASCII value
press the ZERO key.
3.Press +/- . . .The next ASCII string in the list will be
added. On a new indicator, this will be
ASCII 02.
4.With ASCII 02 displayed, press
. . .End is displayed. This means that there
are no ASCII values entered for this
string.
5.Press +/-. . .— is displayed.
6.Key in your first ASCII value,
then press . . .The value is stored and End is
displayed.
7.Continue repeating steps 5 & 6
until you have entered all your
ASCII values. If you make an error,
refer to the next section:
Editing strings.
8.To view your newly entered ASCII
values, press
repeatedly. . .The indicator scrolls through all the
values.
9.Repeat steps 2-7 as new ASCII
strings are needed.
50
Page 54
Editing strings
There are two keys to remember when inserting or deleting ASCII values in a
string: +/- will insert and ZERO will delete. We will use the following example to
illustrate how to edit strings: remove and then replace the hyphen from the
default string ASCII 01: WEIGH-TRONIX WI-127.
1.With StringS displayed,
press . . .ASCII 01 is displayed.
2.Press . . .87, the first ASCII value in the string, is
displayed.
3.Press or to
scroll through the entire string of
ASCII values. Stop when the ASCII
value you want to delete (in this
case, 45-hyphen) is displayed. . .45 is displayed.
4.With 45 displayed, press ZERO. . .The hyphen character is deleted and
the next value in the list--84--is
displayed.
5.To insert a character, scroll
through the ASCII string until the
value that the character will
precede is displayed. . .84 is displayed.
Creating Layouts
Viewing layouts
6.Press +/- . . .
7.Key in your ASCII value,
then press . . .The character is inserted in the string
After your labels and strings are defined, it is within the Layouts submenu that you
create/customize the items which will be transmitted to a printer.
The WI-127 allows the creation of up to sixteen different layouts. A new indicator
contains nine default layouts. You may edit any or all of these nine, or create seven
additional, new layouts.
1.With LAYOutS displayed,
press . . .LAY. 01 (the first default layout) is
2.Press . . .The first item in the layout is displayed.
3.Press or to
scroll through all the items in the
entire layout.
is displayed.
—
and the string name (ASCII 01) is
displayed.
displayed.
4.When you are finished viewing
the items, press . . .LAY. 01 is displayed.
51
Page 55
5.Press
. . .LAY. 02 is displayed. On a new indica-
6.Repeat steps 1-5 above to view
all the layouts.
tor, the nine default layouts are
present. You may create up to seven
more.
Creating new layouts
1.With LAYOutS displayed,
press
. . .LAY. 01 (string #1) is displayed.
2.Press until . . .End is displayed.
3.Press +/- . . .The next layout in the list will be
added. On a new indicator, this will be
LAY. 10.
4.With the new layout
displayed, press . . .End is displayed. This means that there
are no items entered for this layout.
5.Press +/-. . .GroSS is displayed and the ACCEPT
annunciator illuminates indicating that
you are selecting items.
6.Scroll through the list of available
layout items (blocks) (see Table 3)
by pressing the key.
When the item you wish to
insert in the layout is displayed,
press . . .The ACCEPT annunciator goes off and
that item is displayed. Some items (the
ones with asterisks in Table 3) require
that additional information be entered
after selecting the item. For example, if
you choose LF, you must next enter a
value for the number of line feeds you
want, then press once
again.
7.Continue repeating steps 5 & 6
8.To view your newly entered layout,
52
until you have entered all your
layout items. If you make an error,
refer to the next section: EditingLayouts.
press repeatedly. . .The indicator scrolls through all the
layout items.
Page 56
Editing layouts
There are two keys to remember when inserting or deleting items in a layout:
+/- will insert and ZERO will delete. We will use the following example to
illustrate how to edit layouts: remove and then replace the line feed from default
layout 01.
1.With LAYoutS displayed,
press
. . .LAY. 01 is displayed.
To delete any displayed block, press
the ZERO key.
2.Press
. . .1 SP, the first item in the layout is
displayed.
3.Press to scroll through
the layout. Stop when the item you
want to delete--1 LF (one line
feed)--is displayed. . .1 LF is displayed.
4.With 1 LF displayed, pressZERO. . .The item is deleted and the next item--
in this case, End--is displayed.
5.To insert an item in a layout, you
must display the item it will
precede. In this example, we will
insert a line feed in front of End. . .End is displayed.
6.Press +/- . . .GroSS is displayed and the ACCEPT
annunciator illuminates indicating that
you are selecting items.
7.Scroll through the list in Table #3
of available items by pressing the
key until LFS is
displayed. . .LFS is displayed.
8.Press . . .0 is displayed.
9.Key in the number of line feeds
you want, 1 for this example, and
press . . .1 LF is displayed.
10. Press to return to. . .LAY. 01
53
Page 57
Creating Groups
A group is the combination of your layout and the assigned serial port through
which the data is sent. The Groups submenu is where you specify which ports will
print which layouts.
A total of nine groups is available. Within each group, up to three ports can
output layouts. Port one is standard; ports 2 & 3 are offered only if additionalserial boards are installed. Each port is then assigned a layout to print.
The default for the PRINT key is
Group 2.
You may print other groups by
keying in the correct group number,
then pressing PRINT. Note: This
option must be enabled in the
service menu first.
The default group settings are:
Group 1:Port 1 prints Layout 5
Group 2:Port 1 prints Layout 6
Group 3:Port 1 prints Layout 7
Group 4:Port 1 prints Layout 8
Here are two examples of how the groups might be set up using three serial
ports:
Group 1:Port 1 prints Layout 1Group 2:Port 1 prints Layout 5
Port 2 prints Layout 2Port 2 prints nothing
Port 3 prints Layout 3Port 3 prints Layout 2
Follow the instructions below to set up your groups.
1.With GrouPS displayed,
press . . .GrouP 1 is displayed.
2.Press to scroll through
the list of available groups.
To insert a new group, with End
displayed, press +/-. A total of nine
groups is available. When the
group you wish to define is
displayed, press . . .Port 1 is displayed.
With more than one serial port
installed, the group# must assign a
layout# or the choice NONE for each
serial port for every group used.
54
3.Press . . .LAYOut is displayed.
4.a. If you do not want to print from
this port, press . . .NONE is displayed.
b.If you do want to print from this
port, press , then key
in the correct layout number. . .XX (layout number) is displayed.
5.Press . . .Port 1 is displayed. (If you are only
setting up this one port, go to step 8.)
6.You may now select a layout for
Port 2 (if available). With Port 1
displayed, press . . .Port 2 is displayed.
7.Repeat steps 3-5 to set up Port 2.
Press to set up Port 3
(if available). . .Port 3 is displayed.
8.Repeat steps 3-5 to set up Port 3.
Press . . .Inhibit is displayed.
Page 58
9.Press . . .YES is displayed.
10. a. If you wish to inhibit the printout
when motion is detected,
press
b.If you do not wish to inhibit the
printout when motion is
detected, press , then
press . . .No is selected and Inhibit is displayed.
11. Press . . .GrouP 1 is displayed.
12. Press . . .GrouP 2 is displayed.
13. Press . . .Port 1 is displayed.
14. Repeat steps 3-13 above to set up
all the remaining groups.
. . .Yes is selected and Inhibit is displayed.
Broadcast
If you want leading zeroes in your
printouts or broadcast and autosend
information, key in leading zeroes
when setting division size.
Autosend
The 1% trigger threshold cannot be
altered.
This item enables or disables continuous send. If you enable this feature, a group
number must be defined.
1.With broAdcSt displayed,
press . . .no is displayed. If you do not wish to
enable continuous send, STOP HERE.
2.To enable, press . . .GrouP is displayed.
3.Press . . .1 is displayed.
4.Key in the group number you
want continuously sent, then
press . . .The new group is selected and
broAdcSt is displayed.
This item enables or disables autosend. Autosend is automatic print after motion
ceases with weight on the scale above 1% of capacity. To retrigger autosend the
scale must go below 1% stable weight and then above 1% stable weight. If you
enable this feature, a group number must be defined.
1.With AutoSEnd displayed,
press . . .no is displayed. If you do not wish to
enable autosend, STOP HERE.
2.To enable, press . . .GrouP is displayed.
3.Press . . .1 is displayed.
4.Key in the group number you
want sent, then press . . . The new group is selected and
AutoSEnd is displayed.
55
Page 59
Using Inputs and Outputs on the WI-127
Standard Inputs
(defaults shown)
The inputs are configurable. See 127
Inputs Menu section.
For Inbound/Outbound or In-Motion
defaults see the appropriate section
in the back of this manual.
Standard Outputs
InputConnectionAction
#1TB15-2Remote Zero Key
#2TB15-3Remote Print Key
#3TB15-4Go to Net Mode
#4TB15-5Go to Gross Mode
#5TB15-6Remote (pushbutton) Tare Key
#6TB15-7Remote Select Key
#7TB15-8Remote Unit Key
TB15-9Ground
To activate an input, wire them through a normally open switch to ground on
TB15.
To cause the related action to occur, activate the input. If the action is not able to
be performed (i.e. because of motion), the indicator will keep trying while the
input is activated. Individually, an input must be deactivated before it can perform
the action again.
OutputConnectionCutoffBounds
TB14-1+22.8 VDC
TB14-2Positive Relay Voltage
#1TB14-3Cutoff 1Follows Under Annunciator
#2TB14-4Cutoff 2Follows Accept Annunciator
#3TB14-5Cutoff 3Follows Over Annunciator
Follow the instructions in this section to disassemble the WI-127.
1.Unplug the WI-127 from the power source.
2.Remove the back of the WI-127 by removing the sixteen acorn nuts and
pulling the back cover from the case. See Figure 25.
Figure 25
Back view of the WI-27
3.If you need to remove the mother board disconnect all the wires and the
ribbon cable leading to the mother board and remove the ten hold-down
screws shown in Figure 26.
Figure 26
Hold-down screws on the motherboard
4.To reassemble the unit reverse the above procedure.
57
Page 61
Installing Option Cards
1.Follow the disassembly instructions on previous page.
2.Option cards plug into J19 or J20 on the mother board. Remove the four
3.Install the standoffs in these holes and plug in the option card.
When serial boards are installed in
the same stack, port two is always
on the bottom. When installed in
different stacks (side by side) port
two is always closest to the power
supply.
4.Attach the option card to the standoffs with the screws.
hold-down screws shown in Figure 27 for J19 or J20.
58
Figure 27
Main board
5.Reassemble the unit.
Page 62
WI-127 Inbound/Outbound Software
This section of the WI-127 Service manual covers the optional inbound/outbound
software. This software has the following additions made to the service menus.
The complete service menus are included at the end of this section. Below are
descriptions of each additional item.
G.T.N. (After Define in
the Options menu)
These parameters configure the items that affect the inbound/outbound functionality of the indicator.
InboundSelects the print group for the inbound transaction.
If Group is selected, you may then key in the group number to print.
If None is selected, a group will not be printed. Group 2 is the
default.
Outbound Selects the print group for the outbound transaction.
If Group is selected, you may then key in the group number to print.
If None is selected, a group will not be printed. Group 3 is the
default.
ReportSelects the print group for the channel report.
If Group is selected, you may then key in the group number to print.
If None is selected, a group will not be printed. Group 5 is the
default.
EditAllows editing of the inbound/outbound function.
If Sealed is selected, you may not edit the inbound/outbound data.
If Unsealed is selected, editing is allowed. Sealed is the deafult.
Tares ( After G.T.N. in
the Options menu)
This parameter configures the items that affect the tare register functionality of
the indicator.
ReportSelects the print group for the tare register part of the indicator.
If Group is selected, you may then key in the group number to print.
If None is selected, a group will not be printed. Group 6 is the
default.
59
Page 63
WI-127 Inbound/Outbound Serial Communications
Serial communication in the WI-127 involves layouts, ASCII strings, and groups.
Their relationship is this: The ASCII strings allow entry of alpha-numeric charac-
ters that can be assigned to a layout; the layouts contain the information sent out
each port; and the groups set up each port and what is transmitted from it.
The WI-127 Inbound/Outbound indicator has a total of 32 available print layouts.
Layouts
These layouts replace the sixteen found in the standard WI-127.
Twenty-nine of these layouts are predefined. You may use them as they are, or you
may customize them to fit your particular application.
Six of these layouts are used as defaults. Below you will find examples of these
regularly used default printouts. Note that the layouts use other layouts within
themselves (see the italics to the right of each element of the printouts). This
allows for a high degree of customization. You will find all the predefined layouts
mapped out on the following pages.
Layout #1: Displayed Weight
Layout #25: Inbound ticket
Layout # 26: Outbound Ticket
Layout 2
Layout 3
Layout 4
Layout 5
Layout 6
Layout 7
Layout 3
Layout 8
Layout 4
Layout 5
Layout 9
Layout 10
60
Layout 11
Page 64
Layout # 28: Inbound/Outbound Report
Layout # 27: Stored Tare Transaction
Layout 12
Layout 17
Layout 21
Layout 21
Layout 21
Layout 21
Layout 21
Layout 18
Layout 22
Layout # 29: Stored Tare Report
Layout 12
Layout 13
Layout 14
Layout 15
Layout 16
Layout 12
Layout 19
Layout 23
Layout 23
Layout 23
Layout 23
Layout 23
Layout 20
Layout 24
Layout 24
Layout 24
Layout 24
Layout 24
61
Page 65
Predefined layouts
Below are the actual layout codes for the 29 predefined layouts. They are numbered in the same order in which they are available in the indicator.
Layout 1
Layout 2
Layout 3
Layout 4
Layout 5
Layout 6
Layout 7
Layout 8
Layout 9
Layout 10
Layout 11
Layout 12
Layout 13
Layout 14
62
Page 66
Layout 15
Layout 16
Layout 17
Layout 18
Layout 19
Layout 20
Layout 21
Layout 22
Layout 23
Layout 24
Layout 25
Layout 26
Layout 27
Layout 28
Layout 29
63
Page 67
Layout items
Print layouts are constructed using various elements including ASCII strings/
characters, print items, and other layouts. Following are 16 print items which have
been added to the WI-127 Inbound/Outbound indicator. The standard WI-127 has
19 items. Adding these 16, you will have a total of 35 items with which to build
print layouts.These items are found in the “Layouts” section of the Options SubMenu of the Service Menu.
As you scroll through the list of items, you will find the following new items:
Print ItemsDescription
ChannelCurrently selected inbound/outbound channel is transmitted.
Hour inTime of the inbound/outbound weighment for the currently selected
channel is transmitted.
Day inDate of the inbound/outbound weighment for the currently
selected channel is transmitted.
InboundInbound weight for the currently selected channel is transmitted.
Ch GrossInbound/outbound gross weight for the currently compleyed
transaction is transmitted.
Ch TareInbound/outbound tare weight for the currently completed
transaction is transmitted.
Ch NetInbound/outbound net weight for the currently completed
transaction is transmitted.
Ch TotalInbound/outbound net accumulator value for the currently
selected channel is transmitted.
Ch CountInbound/outbound transaction counter value for the currently
selected channel is transmitted.
RegisterCurrently selected tare register number is transmitted.
Rg TypeSource of the tare weight for the currently selected tare register
is transmitted. The two default values are “Pb” (pushbutton) and
“KP” (keypad).
Rg TotalNet accumulator value for the currently selected tare register is
transmitted.
Rg CountTransaction counter value for the currently selected tare
register is transmitted.
nn By Ch“Layout nn” is used once for every inbound/outbound channel in
existence, in ascending order. This supports the ability to
generate inbound/outbound database reports.
nn By In“Layout nn” is used once for every inbound/outbound channel
that has an inbound weight, in ascending order. This supports
the ability to generate inbound/outbound database reports.
nn By Rg“Layout nn” is used once for every tare register in existence, in
ascending order. This supports the ability to generate tare
database reports.
64
Page 68
ASCII strings
As mentioned in the previous section, layouts may include a number of elements,
on of which is ASCII strings. These ASCII strings allow you to customize your
printouts with custom wording.
The WI-127 Inbound/Outbound can store up to sixteen ASCII strings, each
containing up to 32 individuals characters. These strings are numbered 1-16.
The WI-127 Inbound/Outbound contains 15 default ASCII strings. Below is a table
containing the 15 default strings. Use these, or develop your own custom messages.
65
Page 69
Printing Groups
In order to print, the layouts in the WI-127 are assigned to groups. These groups
are in turn assigned to specific serial ports. A group may contain several layouts or
only one layout.
Each group can have up to three serial ports. Each port can have a different layout
assigned to it. Port #1 is always the onboard seril port; ports #2 and #3 are addon option cards.
The default group settings in the WI-127 are as follows:
Each group can be printed separately by using the “N print” feature under “Print”
in the set-up menu. The operator presses the number of the group before
pressing the PRINT key. The information assigned to that group number will be
printed. Pressing 1 and PRINT will print group #1, pressing 2 and PRINT will
print group #2, etc.
66
Page 70
Reset Menu
An additional parameter is added to the Inbound/Outbound Reset Menu.
DatabaseThis parameter will flash if the database is corrupted. This
parameter is not visible if there are no entered records.
NoWill not reset database to default values.
YesWill reset database to default values.
User Menu Modifications
User Menu
These items are offered in the user menu.
G.T.N. - Accesses inbound/outbound function.
Tares - Accesses multiple tare registers.
Hour - View/edit the time.
Day - View/edit the date.
Id - View/edit the ID number.
Bounds - Accesses the checkweighing parameters.
This section of the WI-127 Service manual covers the optional in-motion software
for Conveyor/Monorail scales. This software has the following additions made to
the service menus. The complete service menus are included at the end of this
section. Below are descriptions of each additional item.
Conveyor (After Define
in the Options menu)
Tare ( After Conveyor in
the Options menu)
NoChoose this if you want the indicator to operate as normal. The inputs
used for in-motion weighing are ignored. The cutoffs, analog output,
B.C.D. output, and displayed weight, all work as normal.
YesChoose this to use the indicator for in-motion weighing. The indicator
uses the built in reset/set inputs to start and stop averaging weight
readings. An in-motion weighment is the average of all these weight
readings. Every time an in-motion weighment is calculated, the indicator
does the following:
Updates the Gross and Net variables with the new weighment
Updates the Cutoff outputs, Checkweighing outputs, B.C.D. output and
the Analog output for the new weight
Triggers an auto-print, if enabled
Displays the new weighment for 7 seconds or until the next weighment
begins (dashes) or an error occurs.
It is always possible to enter tare to any resolution but this selection determines
how gross, tare, and net are displayed, printed, and calculated.
0.1 d* The net weight is calculated internally to 10 times the resolution, but
displayed and printed to the nearest division. Tare values are printed and
displayed at 0.1 division.
Default Setup
Modifications
1 dAllows no extra resolution on the tare value. The net weight is calculated
as on a standard indicator.
Scale
Capacity
100 lb
Division
.02 lb
AZT
3 divisions
Update
1 Update/Second
75
Page 79
Filter/Threshold
Filter constant = 3
100 lb(The filter threshold should be set at the configured capacity as a
minumum.)
Options
Define
Autosend
Group 1 (Displayed Weight)
Buttons
Print/P.B. Print
Group 1 (Displayed Weight)
Autoprint
While in the in-motion mode, autoprint changes functionality to automatically
print each time an in-motion weighment is made through the in-motion logic. The
default group for Autoprint is 1. (Displayed weight) While in in-motion mode, the
autoprint functions in the background.
Cutoffs/Checkweigh
While in the in-motion mode, the cutoffs and over/under/accept outputs are only
updated as each in-motion weighment is made. The outputs follow the displayed
in-motion weighments (gross or net) If an erroneous in-motion weighment is
made, the outputs drop out until the next valid weighment is made.
Analog output/B.C.D. output
The weight output cards follow the in-motion weighments while in the in-motion
mode. As usual configuration determines whether to follow the gross, net, or
displayed weighments.
Dynamic Weighment Serial Output
The Gross and Net weights in the layout menus follow the in-motion weighments
while in the in-motion modes. Each time the indicator calculates a new average,
the gross and net weights are calculated based on the new weighment. The gross
and net change from the last In-motion weighment value when the indicator
returns to live weight display. Dashes are printed if ‘no’ is selected for ‘inhibited’
for that group and dashes are on the display to indicate in-motion sampling.
User Menu
These items are offered in the user menu.
76
Id - View/edit the ID number.
Hour - View/edit the time.
Day - View/edit the date.
Page 80
F1
Tare Registers - View/Edit the 10 tare registers.
F2
Over - View/edit the over register.
Under - View/edit the under register.
Target - View/edit the target register.
F3
Cutoffs - View/edit the cutoff registers.
New Layout Items
Label
Averaged
Outputs an ASCII string defined to represent the status of the Selected
weight, Averaged or Live. The default for an averaged weight is “AVERAGED”. The default for a static weighment is a null string. Configuration
for the two ASCII strings appear in the ASCII/Labels menu under the
headings “Averaged” and “Live”.
Layouts and Groups
No changes are made to the default layouts and groups.
Inputs-Onboard
Inputs are redefined.
InputIn-Motion Assignments*
IRemote Zero Key
2Remote Print Key
3Remote Select Key
In these display modes, the only function that is offered from the front panel is the
entrance into the Service menus. After entering the service menus, the indicator
stops the in-motion functionality.
77
Page 81
If the indicator is in a user menu or editing a register, these display modes do not
appear on the display. In this case the in-motion operations appear to occur in
the background.
- - - - - - - -
This indicates that the indicator is sampling the current item on the
scale. As soon as the sampling is completed, the weight is calculated
and displayed.
Annunciator 5 (ID)
This annunciator is used to indicate that the display is showing the
current in-motion weighment. This display appears for seven seconds
after each in-motion weighment is calculated. This mode can be
interrupted by a new weighment (dashes).
Weighing
Errors
Sampling occurs from leading edge to trailing edge of the reset/set input. The
leading edge must correspond to the weight being entirely on the live rail, and the
trailing edge must correspond to the weight just ready to begin leaving the live rail.
1) Six seconds maximum are allowed from arrival to exit. This supports a 12
inch package at a minimum speed of 30 ft/min on a 4 ft conveyor scale. If 6
seconds elapse, an error message is displayed, and auto-print is not initiated.
The error message persists until a new weighment is begun, but no longer
than 7 seconds (unless retriggered by a subsequent weighment.) After 7
seconds, the display reverts to live weight.
2) At least two samples are required to be accumulated to consider a weighment
to be valid. This corresponds to 1/30 th of a second (or 1/10 ft at 180 ft/min).
This requirement is violated when boxes are too long or a conveyor is moving
too fast. If less than 2 samples are accumulated, the weighment is considered
to be in error and is handled in the same manner as (1) above.
3) Weighments less than the upper limit of the AZT window are considered to
be in error and are handled in the same manner as (1) above.
WI-127 to Conveyor and Monorail Wiring Information
The wiring information to connect a WI-127 indicator to a CVC/CVCS conveyor
scale or in-motion monorail scale is as follows:
The WI-127 would need to have the special in-motion software installed to be
compatible with the CVC operation. The recommended filter settings for the inmotion operation is a Filter time constant of 3 and a threshold equal to full scale
capacity. The in motion system will only weigh and transmit data properly if there
is only one package on the scale at a time. With packages identical in size, this
means front edge to front edge spacing of at least the length of the conveyor.
With varying length package applications there are two criteria which must be
met:
1) With a package on the conveyor scale, the next one coming on the scale
must not break the entrance photocell beam prior to the one on the scale
breaking the exit beam.
2) With a package on the conveyor scale, the exiting package must clear the exit
beam prior to the package entering the scale clearing the entrance beam.
If either of these rules are violated erroneous weights may be transmitted or
weight values not transmitted at all.
This simple formula will help you figure the minimum package spacing needed for
unequal package sizes.
The diagram below shows a formula for figuring throughput of identically sized
packages.
All serial data istransmitted out of TB16, 17, and18 at the
same time. To configurethe serial port for bi-directional
communication you must select which format (RS-232, etc.)
will receive data from the external device. See serial port
chart for correct jumpersettings.
EXCITATION CHART
P6 Jumper
Position
Typ.Exc
Voltage
typical 3-410.0
2-35.0
1-26.0
1-2, 3-4
16.0
NOTE:
When using barriers, thetwo pin jumper is not
connected and should be stored on pin-1. The
barrier should drop theexcitation voltage down
to the range of (2.0to 3.8 volts) . If only one load
cell is being used, the voltage could be greater than
3.8 volts. In thiscase, add a load resistor toincrease
the current through the barrier to drop the voltage.
All serial data istransmitted out of TB34, 35, and36 at the
same time. To configurethe serial port for bi-directional
communication you must select which format (RS-232, etc.)
will receive data from the external device. See serial port
chart for correct jumper position.
Page 94
WI-127 INDICATOR
(115/230VAC)
OPTION BOARDS
CUTOFF & INPUT
AND
BCD PARALLEL
91
Page 95
WI-127 INDICATOR
(115/230VAC)
KEYPAD OVERLAY P/N49058-0016
(STANDARD)
KEYPAD OVERLAY P/N49058-0024
(GTN)
AND
MATRIX
92
Page 96
WI-127 INDICATOR
(115/230VAC)
OPTION BOARDPROCEDURE
AND
INSTALLATION
93
ITEM
NO.DESCRIPTIONW-T P/NQTY
1Nut, Kep, #6-321025-00114 4/board
2---- NO PART ---3Standoff, m/f, 6-32 x.56L15437-5000 4/board
4Screw, Sem, #6-32 x .25L1006-02600 4/board
5Analog Output Power Cable Assy49375-00531
6Analog Output pc. BoardAssy49148-0018----
<or>50097-0017---Serial I/O pc Board Assy49152-0011---Cutoff & Input pc BoardAssy49156-0017---BCD Parallel Output pcBoard Assy 49160-0011----
Page 97
WI-127 INDICATOR
(115/230VAC)
WALL MOUNT OUTLINE DRAWING
94
Page 98
WI-127 INDICATOR
(115/230VAC)
PANEL MOUNT OUTLINE DRAWING
95
Page 99
Page 100
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.