(For scale base, 3000 divisions)
±0.005% of capacity, maximum (For base, ±0.01%)
±0.005% of capacity, maximum (For base, ±0.01%)
0.005% of capacity, maximum (For base, 0.01%)
All calibration and programming is done through the front panel with data
stored in nonvolatile memory.
1, 2, 5, or 10* times per second
1*, 2, 5, or 10 display intervals
±1%, ±2%, ±5%, ±10%, ±20%, ±50%, ±100%* of Capacity
d = 1 displayed division
±0.25 d, ±0.5 d, ±0.6 d, ±1 d*, ±2 d, ±3 d, ±5 d
d = 1 displayed division
The scale displays weights up to capacity plus 9 divisions, referenced from
the zero value determined by zero setting point, or 105% of capacity,
referenced from the deadload.
* = default
QC-3265 Checkweigher User’s Manual
Page 5
Under Range Capacity
The scale displays weights in the negative direction using the same restrictions as for over-range, but further limited by the number of display digits
available.
Temperature Range
Humidity
Scale Capacity and Division
-10 to 40°C (14 to 104°F)
-30 to 60°C (-22 to 140°F) with reduced accuracy
Up to 100% relative humidity.
PoundsOuncesKilogramsGrams
6 x .002100 x .053 x .0013000 x 1
6 x .001100 x .023 x .00053000 x .5
*12 x .005200 x 0.16 x .0026000 x 2
12 x .002200 x 0.056 x .0016000 x 1
30 x .01480 x 0.215 x .00515000 x 5
30 x .005480 x 0.115 x .00215000 x 2
60 x .02960 x 0.530 x .0130000 x 10
60 x .01960 x 0.230 x .00530000 x 5
100 x .051600 x 0.545 x .0245000 x 20
100 x .021600 x 0.245 x .0145000 x 10
200 x .13200 x 190 x .0590000 x 50
200 x .053200 x .590 x .0290000 x 20
* = default
Options
• LCD & LED versions
• 230 VAC 50/60 Hz power
• Tower mounted toggle switch for setting target weight and zeroing
• Battery and charger (available with LCD version only)
• RS-232 or RS-485 interface board
• Short and extended towers
• Remote head
• NTEP load cells
QC-3265 Checkweigher User’s Manual
5
Page 6
This page left intentionally blank.
6
QC-3265 Checkweigher User’s Manual
Page 7
Introduction
This manual covers software
48307-0025B and newer on
the LED model of the QC3265 and 48307-0017C and
newer on the LCD model.
These revisions contain the
LB-OZ capability which
previous versions did not
have.
Front Panel
The Quick Check QC-3265 Checkweigher is a low-cost, high-speed production checkweigher housed in stainless steel for harsh, washdown environments. This manual is divided into the following sections:
• Introduction
• Front Panel
• Operation
• User Menu
If you have any problems with your QC-3265 Checkweigher, call your local
Weigh-Tronix distributor.
Figure 1 shows the front panel. The panel consists of the following:
• a six-digit, liquid crystal or LED display
• five keys;
QUICK CHECK
TARGET
UNITS
ZERO
(Right arrow key)
• a center of zero annunciator
• four unit of measure annunciators; lb, kg, oz, g
• three checkweigher annunciators; OVER, ACCEPT, UNDER
Figure 1
QC-3265 Front Panel
QC-3265 Checkweigher User’s Manual
7
Page 8
Keys
If for some reason the QC3265 cannot perform a key
function (due to motion, range
limits, and others) the display
will show CAN'T while the key
is held down.
Following are the keys and their functions.
Press this key to toggle between checkweigh mode and
weight display mode, assuming you have an active target
weight. Use the QUICK CHECK key to return to the
weight display and checkweigh modes from anywhere in
the menus. Press this key to accept a displayed selection
and return to a display mode.
Use the up arrow-TARGET key to set the target weight.
With no weight on the scale, press this key to remove a
target weight and enter the weight display mode. With a
weight on the scale, press this key set the target weight
and to enter the Checkweigh mode from the weight
display mode. Also, press this key to move up in the
menus. You can also set the target weight by flipping the
optional, column mounted toggle switch up.
Use the left arrow-UNITS key to change the unit of
measure in weight display mode and checkweigh mode.
This key works only if more than one unit of measure is
enabled. Use this key to move to the left in the menus.
Key names are abbreviated in
many of the instructions of this
manual. Instead of saying the
up arrow-TARGET key every
time it is shortened to this
symbol ^.
^ = up arrow key
v = down arrow key
< = left arrow key
> = right arrow key
Annunciators
Use the ZERO-down arrow key to zero the scale while in
either display mode. In the A/D test mode this key sets the
offset to zero. Use this key to move down in the menu
hierarchy. With a selection displayed, press this key to
select that item and go up one menu level. You can also
zero the scale by flipping the optional, column mounted
toggle switch down.
Use the right arrow key to move to the right in the menus.
The OVER, ACCEPT, and UNDER annunciators are to the right of the
display and are shown below. The OVER light is yellow, the ACCEPT light is
green and the UNDER light is red.
Target weight and the tolerances are saved in nonvolatile
memory. They are not lost
when the unit is powered
down.
8
QC-3265 Checkweigher User’s Manual
Page 9
The unit of measure annunciators are below the display and are reproduced
The lb and oz annunciators will
be lit when you are in the LBOZ unit of measure. Two
numbers are displayed separated by a space in LB-OZ.
The left number is the number
of pounds and the right is the
ounces to the correct division
size.
below. The lit annunciator tells you the currently selected unit of measurement.
The center of zero annunciator, reproduced below, is next to the ZERO key.
When the annunciator is lit the scale is within ¼ division of zero.
Installation and Operation
Installation
There is no ON/OFF switch.
When the unit is plugged in it
will power up. If the battery
option is installed and enabled,
press any key to start.
Explanation of Modes
of Operation
Place the QC-3265 on a stable, level surface.
Use the leveling feet to level the scale. A bubble level is visible under the
scale platform shroud. Lock the feet in this position using the locknuts on the
feet.
Plug the unit into a properly grounded 115VAC outlet.
The socket-outlet must be installed near
the equipment and easily accessible.
There are two modes of operation for the QC-3265, the weight display mode
and the checkweigh mode.
• Use the weight display mode when you want to use the scale for
displaying weight. The scale powers up in weight display mode if there
is no target weight in memory.
• Use the checkweigh mode for checkweighing operations and the use
of the OVER, ACCEPT, and UNDER LEDs. If there is a target weight
in memory, the scale will power up in the checkweigh mode. Below are
explanations of the operation in each mode.
Operation as a Scale
Press the UNITS key to view a
weight in a different unit of
measure.
With the scale in weight display mode, follow these steps to perform a
weighment:
1. Press the ZERO key to zero the
scale. . .Display shows 0 weight.
2. Place the object to be weighed on
the scale. . .The weight is displayed.
QC-3265 Checkweigher User’s Manual
9
Page 10
Gross Weight
Checkweighing
To use the QC-3265 as a checkweigher for gross weight, follow the directions given below:
1. Zero the scale by pressing the
ZERO key or flip the optional
toggle switch down. . .Display shows 0 weight.
In gross mode, the indicator
will always represent negative
weight with a leading minus
sign. If the weight becomes too
large, positive or negative,
overscores or underscores will
be displayed while the condition persists.
Changing the
Acceptance Tolerance
Over and Under tolerance can
be set from the User menu or
the Setup menu (see Service
Manual) but not both. The
choice where the tolerances
appear is made in the Setup
menu. The tolerances will not
appear in the User menu if they
appear in the Setup menu and
vice versa.
2. Place a test weight on the scale
equal to the gross weight of the
package. . .The gross weight is displayed.
3. Press the TARGET key or flip
the optional toggle switch up. . .The display shows 0 weight and the
ACCEPT LED illuminates.
4. Remove the test weight and
place a package on the scale. . .The display will show the weight
deviation from the target weight and
the OVER, ACCEPT, or UNDER
LED will illuminate accordingly.
IMPORTANT NOTE:If you use the ZERO key to tare a container weight
and use the checkweigher in net weighing mode, the unit will track zero
whenever there is an empty container on the scale. If the scale is left idle
while empty, zero drift may occur.
1. Press the > key. . .OVER is displayed.
2. Press the v key. ..o 0.5 is an example of what might
be displayed. o stands for over
tolerance and the value is the
current over tolerance. The actual
value depends on the unit of measure and division size. The over LED
is lit.
3. Press the < key to decrease the
value and the > key to increase
the tolerance value. Press the v
key to accept the displayed value. OVER is displayed.
If you change the tolerance but
decide you would rather keep
the original value, press the ^
key in step 2 to return to OVER
without accepting the displayed
value.
Target weight and the tolerances are saved in nonvolatile
memory. They are not lost
when the unit is powered
down.
10
4. Press the > key. . .UNDER is displayed.
5. Press the v key. . .u 0.5 is an example of what might
6. Repeat step 3 and press the
be displayed. u stands for under
tolerance and the value is the
current under tolerance. The under
LED is lit.
QUICK CHECK key to return to
the checkweigh or display mode.
QC-3265 Checkweigher User’s Manual
Page 11
User Menu
Move to the right by pressing the > key
Move down by pressing the v key
Move left by pressing the < key
Move up by pressing the ^ key.
Without realizing it you have learned how to access the User menu. When
you press the > key from either the checkweighing (deviation display) or the
weighing (weight display) mode you access the User menu shown below.
Figure 2
User Menu
You have already learned about UNDER and OVER tolerance. The other
items in this menu are described below.
OFFThis item appears only if battery power mode is enabled in setup,
With OFF displayed, press the v key to turn off the QC-3265.
Pressing any key turns the unit on in weight display mode. Battery
power is not an option with QC-3265s equipped with an LED
weight display.
ABOUTPress the v key with ABOUT displayed and QC3265 will be
displayed. Press the > key to step through the EPROM part
number, dash number, revision letter, and then back to QC3265.
Press the ^ key to go back to the ABOUT display.
TESTPress the v key with TEST displayed to access the test menu
shown in Figure 3. Use the v, ^, <, and > keys as before to move
through the menu.
AUDITLets you see the CA (calibration) audit number and the
CF (configuration) audit number.
DISP.Press the v key twice to perform a display segment
test. Press the < or > key to change the direction of the
test. Press the ^ key twice to stop the test and return
to DISP.
QC-3265 Checkweigher User’s Manual
11
Page 12
BUTTON Lets you check the function of each key. Figure 3
shows you what will be displayed as you press each
key.
VOLTSShows you the current input voltage going into the
voltage regulator.
A to DShows the output of the A to D converter with a
sensitivity of 20,000 counts per millivolt per volt.
SERIALLets you perform READY/BUSY and LOOP/NOLOOP
tests to check the serial port. READY/BUSY not
available unless RS-232 has been selected during
setup.
SETUPThis is a password protected menu and is used for basic setup
and configuration of the unit. See the Service Manual for details.
12
Figure 3
Test Menu
QC-3265 Checkweigher User’s Manual
Page 13
Communication
Error Displays
The QC-3265 has an optional RS-232 or RS-485 communication capability.
Your unit may be customized to print according to your needs. Examples of
the default print layout are shown below:
30.01 lb G or 0.01 kg DEV
What is transmitted is the weight or deviation, one space, the unit of measure, one space, and a label for the weight (G for gross, DEV for deviation).
Your unit may be set up to send data continuously, after each weighment (if
a 30 division movement has occurred), or upon request from a computer.
The following are displays you may see if problems occur or if invalid
operations are attempted with your QC-3265:
DisplayDescription
Overrange weight.
Underrange weight.
Recovering from lock-up or out of range condition.
A-D converter is not functioning.
A-D converter subjected to an input signal beyond
±6.66667 mV/V
The unit cannot perform a function. Displayed only while
key is held down.
Corrupted data in the reset menus. See Reset Menus.
(* = RESET, SETUP, or CAL)
A low input voltage is detected. This appears when voltage
level reaches 10.5 volts and alternates with the normal
display. The unit will shut itself off at 9.2 volts. Only
available on battery powered units.
QC-3265 Checkweigher User’s Manual
13
Page 14
Low voltage on the LED version.
Displayed while a key is pressed when attempting to
modify a sealed selection without edit privileges.
Scale unstable at power up.
Serial port handshake not responding.
Plastic Knob Installation for USDA Approved Applications
Below are the instructions and illustration for installing the plastic knobs
which are necessary for USDA approved installations.
Install knobs by
• Removing the two 10-32 hexhead screws accompanying the unit.
• Replace with knobs.
• Use supplied Belleville washers between knob and mounting surface.
• Install the washer with the major diameter bearing against mounting
surface.
14
Figure 4
Plastic knob installation
QC-3265 Checkweigher User’s Manual
Page 15
Weigh-Tronix
1000 Armstrong Dr.
Fairmont, MN 56031 USA
Telephone: 507-238-4461
Facsimile: 507-238-4195
e-mail: industrial@weigh-tronix.com
www.wtxweb.com