The information contained herein is the property of Avery Berkel Limited and is supplied without liability for errors and omissions. No part may be reproduced or used except as authorised by contract or other written permission. The copyright and the foregoing restriction on reproduction and use extend to all media in which the information may be embodied.
Warnings
For your protection, all mains (110V or 230V) equipment used where damp or wet conditions may occur, must be supplied from a correctly fused source and protected by an approved RCD to BS7071 or BS7288 or IEC1008-2-2 BS EN 61008:1995. IF IN DOUBT SEEK ADVICE FROM A QUALIFIED ELECTRICIAN.
To avoid the possibility of electric shock or damage to the machine, always isolate from the mains power supply before carrying out any routine maintenance.
Cleaning the scale
Harsh abrasives, solvents, scouring cleaners and alkaline cleaning solutions, such as washing soda, should not be used especially on the display windows. Under no circumstances should you attempt to wipe the inside of the machine.
The outside of the machine may be wiped down with a clean cloth moistened with water containing a small amount of washing up liquid.
Installation
FX110 only
Remove parts 1- 4 before installing the machine
1
Make sure that the work surface is firm, steady and free from vibration and
draughts.
2
Hold the scale by the base when picking up. Do not pick up the scale by the
weigh plate or cross.
3
Adjust the feet if the scale is not level.
FX110
Installation
FX120
To avoid the risk of RSI (Repetitive Strain Injury) it is important to ensure that the machine is placed on a surface which is ergonomically satisfactory to the user. In cases where usage is prolonged EN614-1 recommends frequent breaks for the user.
F 110 connectionsX
Power
Programming
FX 120 connections
1 Remove weigh plate.
Display
RS 232 Serial
2
Connect power cable.
To connect remote display.
3
OR
Replace the weighplate. Connect to power point.
4
Fitting the FX 120 scale mounted display
Make sure the scale is not connected to the power supply and remove
1
weigh plate.
2
Turn over machine and place on a safe, level surface.
3
Turn the machine round so the required mounting slot is next to the edge of
the counter.
Slot 1
Slot 3 - required mounting slot
4
Remove the screw in the mounting slot.
Slot 2
Place the mounting bracket into the slot.5
6
Place washer over screw hole and fix with the large replacement screw.
7
Refer to the drawings on the following pages and:
!
Plug the display cable into the base. Feed the slack cable round the base.
!
Secure the cable clips.
!
8
Place machine upright.
9 Replace the weigh plate. Connect to power point.
Caution:
Do not lift the scale by the column display or the weighplate.
Cable route: single display
Cable from slot 3
Cable from slot 2
4
8
7
3
3
4
5
6
2
2
9
1
1
Cable from slot 1
1
2
3
4
Cable route: dual display
Cable from slot 3
Cable from slot 1
4
8
7
3
5
6
2
9
1
1
2
3
4
Cable route: dual display
Cable from slot 2
Cable from slot 1
1
2
3
4
1
2
3
4
Cable route: dual display
Cable from slot 2
Cable from slot 3
5
4
1
2
3
4
8
7
3
5
6
2
9
1
Securing the cable clips
FX 120 battery operation
Note: Battery operation is not available on the FX110
Plug the battery pack into the scale (use 6 x 1.5v 2‘D’ size cells) and place the pack into the recess. Replace the cover and weighplate.
NiCad
Caution:
flat or damaged batteries immediately. Please note: Rechargable NiCad batteries are not suitable for this application due to their inherent short circuit thermal characteristics which can adversely affect the performance of the scale.
When you see the symbol flashing on the display (or Battery LED flashing), then you must either replace the batteries (or recharge the rechargeable battery).
Using a rechargeable battery pack
into the scale and place the pack into the recess. Replace the cover and weighplate.
Recharging the battery pack
located on the base of the scale, then switch on using the mains ON/OFF switch.
To avoid damage to the scale caused by leaking batteries, remove
- Plug the rechargeable battery pack
- Plug the mains cable into the socket
Note:
You can still use the scale while the battery is being charged.
Error messages
If a persistent error message appears or the scale locks up,
F30
disconnect then reconnect the scale to the power supply. If the scale remains inoperative, contact your authorised service agent.
Temporary
error
Error codes:
F10 F11 F30 F35 F36
Contact your service agent if you see any other F messeges.
E 120 E 121 E 122 E 123 E 124 E 130 E 131
Battery failure. Power supply voltage is too high. Memory checksum invalid. Prohibited function. Prohibited scale capacity.
Tare weight unsteady Tare too heavy. Tare already in operation. Cannot use EPOS tare. Mismatch between EPOS and scale tare increments. Unit price not accepted. Scale has not returned to zero between EPOS transactions.
Weight
unsteady
.
Balance
failed
Under
range
Over
range
Using the scale
Testing the display
Or
LCD
Displaying service information
LCD
LED
Resetting the machine
Press and hold
LCD
Selecting sleep mode
LED
Or
LED
Press and hold
Waking up the scale
Depending on configuration...
Either
Or
T
Zeroing the scale
Make sure that the scale is at zero before weighing any goods.
Or
LCD
Weighing goods
Place the goods onto the scale.
Using tares
T
Cumulative tare
LED
To cancel a tare
T
T
T
Repeat until all the goods have been weighed.
Management Mode
Entering management mode
Or
LCD
Navigating management mode
Branch
C
12.95
Go to the next branch (press and hold) Go to next sub-branch
Go to the previous branch (press and hold) Go to previous sub-branch
Sets the value to 0 enabling you to display a new value Return to Branch 00 (press and hold)
Displays the value. Press again to re-display the branch / sub-branch
After clearing a value to be changed each key press increments the value displayed. Pressing and holding the key increments the value by 10.
Note:
go to the next branch to accept the value.
Sub-branch Value
1
After changing a value, you must
LED
For Example:
Branch ( )
12
Sub-branch ( )
Branch ( )
13
Sub-branch ( ) Sub-branch ( )
C
98
00 01
Exit management mode
Press and hold
Example: turn the keyboard bleeper off and change the backlight time to 5 seconds
0.00
Press and hold
.
Press and hold
19.00
20.00
19.01
C
1
C
4
0
29.00
Press and hold
Press and hold
0
1
Switching the bleeper alarm on/off
Go to branch 19, Sub-branch 00
19.00
0=Off 1=On
1
Switching the keyboard bleeper on/off
Go to branch 19, Sub-branch 01
19.01 1
0=Off 1=On
Target bleep
Branch 19, Sub-branch 02 - Not applicable for this machine.
Error bleeper
Go to branch 19, Sub-branch 03
19.03 1
0=Off 1=On
Changing the bleeper volume
Go to branch 19, Sub-branch 04
19.04 1
0 = Quiet 1 = Loud
Changing the backlight time
Go to branch 20, Sub-branch 00
20.00 1
0 = always off 1 = 5 seconds 2 = 1 minute 3 = 5 minutes 4 = always on
Changing the sleep time
Go to branch 20, Sub-branch 01
20.01 1
0 = none 1 = 1 minute 2 = 5 minutes 3 = 30 minutes
Changing the duration of a ‘Long' press (press and hold)
Go to branch 29, Sub-branch 00
29.00
200
Enter a value in units of 10 milliseconds (50 -255). For example : 200 = 2 seconds 150 = 1½ seconds 100 = 1 second
Enabling key timeout
Go to branch 74, Sub-branch 01
74.01 1
0 = Disabled 1 = Enabled
Changing the timeout period
Go to branch 74, Sub-branch 02
74.02 20
Enter a value in units of 10 milliseconds (1-255). For example : 200 = 2 seconds 150 = 1½ seconds 100 = 1 second
Declaration of Conformity
The non-automatic weighing instrument
Manufacturer Avery Berkel Type Number of EC type approval certificate UK 2625
Manufacturer Avery Berkel Type FX 120 Number of EC type approval certificate T5532
Corresponds to the requirements of the following EC directives:
FX 110
Non-Automatic Weighing Instruments Directive EMC Directive Low Voltage Directive The applicable harmonised standards are:
90/384/EEC* 89/336/EEC 73/23/EEC EN 45 501* EN 50 081-1 EN 60 950
EMC Compliance
This is a class A product. In a domestic environment, this product may cause radio interference in whch case the user may be required to take adequate measures.
*This declaration is only valid if the non-automatic weighing instrument was verified by the manufacturer or with a certificate of conformity issued by a notified authority.
A copy of the original signed declaration for this instrument is available from: Foundry Lane, Smethwick, West Midlands, England. B66 2LP
Email: sales@averyberkel.com
Internet: http://www.averyberkel.com