(preset output mode) setting is available in MT4Y-☐-43 (relay (low out)+transmission (DC4-20mA) output) and
※
Only for MT4W.
To measure the current over DC5A, please select DV type because the shunt should be used.
In case of selecting frequency display, no output will be provided even if it is output support models.
(main output, sub output and RS485 communication output)
100-240VAC
12-24VDC,
100-240VAC
100-240VAC
※
1
4100-240VAC 50/60Hz
DVDC voltage
DADC current
AVAC voltage
AAAC current
YDIN W72×H36mm
WDIN W96×H48mm
49999 (4-digit)
MTMulti meter
N
Indicator
0
Relay output
1
NPN open collector output
2
PNP open collector output
※
1
3
Relay (low out)+transmission (DC4-20mA) output
※
1
4
Relay (low out)+RS485 communication output
5
BCD dynamic output
6
Low speed serial output
N
Indicator
0
Relay+transmission (DC4-20mA) output
1
Relay output
2
NPN open collector+BCD dynamic output
3
PNP open collector+BCD dynamic output
4
NPN open collector+transmission (DC4-20mA) output
5
PNP open collector+transmission (DC4-20mA) output
6
NPN open collector+low speed serial output
7
PNP open collector+low speed serial output
8
NPN open collector+RS485 communication output
9
PNP open collector+RS485 communication output
2
12-24VDC
SENSORS
CONTROLLERS
MOTION DEVICES
SOFTWARE
(J)
Temperature
Controllers
(K)
SSRs
(L)
Power
Controllers
(M)
Counters
(N)
Timers
(O)
Digital
Panel Meters
(P)
Indicators
(Q)
Converters
(R)
Digital
Display Units
(S)
Sensor
Controllers
(T)
Switching
Mode Power
Supplies
(U)
Recorders
(V)
HMIs
(W)
Panel PC
(X)
Field Network
Devices
O-25
Page 2
MT4Y/MT4W Series
-|Transparent setting guide|-
▣ Specifications
Series
Measurement input
Power supply100-240VACᜠ 50/60Hz12-24VDC
Allowable voltage range90 to 110%90 to 110%
Power consumption5VA5W
Display method7-segment LED display (red) (character height: 14.2mm)
Display accuracy
Max. allowable input 110% F.S. for each measured input range
A/D conversion methodΣΔ (Sigma Delta) ADC
Sampling cycleDC input: 50ms, AC input: 16.6ms
Max. display range-1999 to 9999 (4-digit)
Preset output
Sub output
(transmission output)
AC measurement
Frequency measurement
Hold function
Insulation resistanceOver 100MΩ (at 500VDC megger, between external terminal and case)
Dielectric strength2000VAC 50/60Hz for 1 min (between external terminal and case)
Noise immunity±2kV the square wave noise (pulse width: 1㎲) by the noise simulator
Vibration
Shock
Relay
life cycle
Environ
-ment
Insulation type
Approval
※3
Weight
※
1: AC, frequency measurement functions are only for AC measurement input type.
※
2: MT4Y-
※
3: The weight includes packaging. The weight in parenthesis is for unit only.
※
Environment resistance is rated at no freezing or condensation.
※1
※2
Mechanical0.75mm amplitude at frequency of 10 to 55Hz (for 1 min) in each X, Y, Z direction for 2 hours
Malfunction0.5mm amplitude at frequency of 10 to 55Hz (for 1 min) in each X, Y, Z direction for 10 min
Mechanical100m/s² (approx. 10G) in each X, Y, Z direction for 3 times
Malfunction300m/s² (approx. 30G) in each X, Y, Z direction for 3 times
MalfunctionMin. 20,000,000 operations
MechanicalMin. 100,000 operations (250VAC 3A load current)
Ambient temperature -10 to 50℃, storage: -20 to 60
Ambient humidity35 to 85%RH, storage: 35 to 85%RH
Connector socket specification: Contact the manufacture for the
socket and cable.
Connector socketHIF3BA-20D-2.54RHirose Electric
I/O cable (Sold separately) CO20-HP
SpecificationsManufacture
(triple NPN/PNP open collector+BCD dynamic output)
+24VDC
30mA Max.
0V
ACD0D2
BCD OUT:
NPN OPEN COLLECTOR
12-24VDC Max. 50mA
COM2DOTD3D1DB
MAIN OUT:
PNP OPEN COLLECTOR
12-24VDC Max. 50mA
POL
※
POL: When a display value is "-", the signal of "-" will be
outputted.
triple NPN/PNP open collector+transmission (DC4-20mA) output)
+24VDC
30mA Max.
911131517191357
0V
(-)
(+)
Current:
DC4-20mA LOAD 600Ω Max.
MAIN OUT:
PNP OPEN COLLECTOR
12-24VDC Max. 50mA
PCB
-L, CO20-HP -R Autonics
+24VDC
GO
COM1 HI LO
COM1 HI LO
30mA Max.
2468
1357
+24VDC
30mA Max.
GO
2468
1357
0V
0V
● MT4W- -46 / MT4W- -47
MAIN OUT:
NPN OPEN COLLECTOR
12-24VDC Max. 50mA
COM1 HILO
● MT4W- -48 / MT4W- -49
MAIN OUT:
NPN OPEN COLLECTOR
12-24VDC Max. 50mA
COM1 HILO
O-28
(triple NPN/PNP open collector+low speed serial output)
※
+24VDC
GO
2468101214161820
135791113151719
30mA Max.
0V
POL: When a display value is "-", the signal of "-" will be
outputted.
CLOCKPOL
LATCH
MAIN OUT:
PNP OPEN COLLECTOR
12-24VDC Max. 50mA
DATA
COM2
SERIAL OUT:
NPN OPEN COLLECTOR
12-24VDC Max. 50mA
(triple NPN/PNP open collector+RS485 communication output)
+24VDC
RS485
30mA Max.
GO
2468101214161820
135791113151719
B (-)
SG
SG
※
0V
A (+)
RS485
SG (Signal Ground)
: Signal ground is the terminal which is base of the
signal voltage. It is connected with terminal 7 inside
of the product.
MAIN OUT:
PNP OPEN COLLECTOR
12-24VDC Max. 50mA
GO
2468
1357
COM1 HI LO
GO
2468
1357
COM1 HI LO
+24VDC
30mA Max.
0V
+24VDC
30mA Max.
0V
Page 5
Multi Panel Meter
-|Transparent setting guide|-
▣ Dimensions
MT4Y Series
72
● Bracket
4
60
12
MT4W Series
96
● MT4Y-
6
36
9
14.5
● MT4W- - N/ 0/ 1
6610089.5
48
-4N/40~44/46
77
30.2
● Panel cut-out
Min. 40
45
● MT4Y-
6
Min. 91
● MT4W-
-45
66.5
-42 to 49
(unit: mm)
SENSORS
10.5
0
+ 0.5
31.5
+ 0.7
68
0
14.9
30.2
CONTROLLERS
MOTION DEVICES
SOFTWARE
(J)
Temperature
Controllers
(K)
SSRs
(L)
Power
Controllers
(M)
Counters
(N)
Timers
45
(O)
Digital
Panel Meters
● Bracket
4.412
73
▣ Parameter Setting
RUN
3 sec
PA 0
PA1
2 sec
PA2
3 sec
3 sec
(P)
Indicators
● Panel cut-out
Min. 116
10
+ 0.8
92
Min. 52
0
14.5
※
※
※
※
※
MODE
Press
Press
Press
When pressing
key in RUN mode and it enters PA 0 group.
MODE
key for over 3 sec in RUN mode, it displays [
MODE
key for over 5 sec in RUN mode, it displays [
MODE
key continually, it stops displaying at [
It is advanced to current display parameter releasing
MODE
Press
key for over 3 sec in any parameter groups, it returns to RUN
PA1
PA2
MODE
].
] after [
PA2
key at [
].
PA1
].
] or [
PA1
mode.
※
If any key is not entered for 60 sec in each parameter, it returns to RUN mode.
※
After returning to RUN mode, press
MODE
key within 2 sec, it returns to previous
parameter. (Refer to the below descriptions of each parameter group.)
※
PA 0 group cannot be entered when preset output mode of [
.
OFF
PA2
] group is
0
+ 0.6
45
PA2
].
(Q)
Converters
(R)
Digital
Display Units
(S)
Sensor
Controllers
(T)
Switching
Mode Power
Supplies
(U)
Recorders
(V)
HMIs
(W)
Panel PC
(X)
Field Network
Devices
O-29
Page 6
MT4Y/MT4W Series
-|Transparent setting guide|-
▣ Parameter 1 Group
RUN
3 sec
PA1
IN-R
Display of factory
default for
measurement input
Select measurement input specification by
Refer to "
-
Measurement Input".
● Measurement input chart by model
ItemRange of measurement input
MT4Y/W-DV
.
MT4Y/W-DA
MT4Y/W-AV
MT4Y/W-AA
※
Standard specification. (110P is 440VAC/110VAC P.T)
500V
100V
50V
10V
5A 2A )5A )2A 50MA 4-20MA 5MA 2MA 5A
500V
250V
110P
(※)
5A 2.5A 1A )5A )25A )1A 50mA 5
50V
5V
20V
1V
)25V
10V
2V
50ㅡMV
1V
500V
500V
A
DISP
IN-T
When
STND
INbH
STND
AVG
is
DISP
STND
Max. display
value
It shows Max. display
value of standard
specification.
Display value is fixed.
!000
INbL00
Refer to "
-
(
Select measurement method for AC.
Setting range:
(This parameter will be displayed only for AC measurement models.
In case of selecting frequency display, it will not be displayed.)
It corrects a gradient of High-limit display value against max. input.
Setting range:
(
It corrects deviation of Low-limit display value against Min. input.
Setting range:
(
Measurement Input".
mode is only for AC measurement type.)
FREQ
(average: AVG, root mean square: RMS)
AVG, RMS
When
DOT
H-SC
L-SC
: change setting value, : shift the digit)
: set a deviation value, : shift the digit)
is
DISP
SCAL
Decimal point
position
Max. display
Min. display
to
)100
%000
to 99 (Refer to " Zero adjustment ".)
-99
Select decimal point position.
Setting range
: 0,
※
position, "
decimal point position.
Set display value against
max. measurement input.
value
(
Set display value against
min. measurement input.
value
(
(%)
)0, )00, )000
When selecting the decimal point
: shift the digit
: change setting value)
: shift the digit
: change setting value)
H-SC
",
" have
L-SC
When
(It is only displayed for AC measurement type.)
DOT
is
DISP
Decimal point
position
FREQ
Set frequency measurement
range, it is decided as
the decimal point position.
(
change the
decimal point position)
0
)0
)00
)000
※
In case of selecting frequency display, no
output will be provided even if it is output
support models (main output, Sub output
and RS485 communication output).
It corrects a gradient of display
INbH
INbE
!000
10-0
value against max.input.
Setting range: 0.001 to 9.999
( : shift the digit,
: change setting value)
Set
Setting range: 10
: shift the digit,
(
: change setting value)
1 to 9999Hz
0.1 to 999.9Hz
0.10 to 99.99Hz
0.100 to 9.999Hz
-2
to 101
※
After setting each mode, press
※
If any key is untouched for 60 sec after advance to parameter, it will return to RUN.
TypeMeasurement input and range Input impedanceDisplay range [
500V
]4.33MΩ0.0 to 500.0 (xed)
100V
]4.33MΩ0.0 to 100.0 (xed)
50V
]433.48kΩ0.00 to 50.00 (xed)
10V
]433.48kΩ0.00 to 10.00 (xed)
)25V
]2.28kΩ0.0 to 250.0 (xed)
50MV
]2.28kΩ0.00 to 50.00 (xed)
)5A
]0.22Ω0.0 to 500.0 (xed)
)2A
]0.22Ω0.0 to 200.0 (xed)
50MA
]2.22Ω0.00 to 50.00 (xed)
4-20
]2.22Ω4.00 to 20.00 (xed)
5MA
]22.22Ω0.000 to 5.000 (xed)
2MA
]22.22Ω0.000 to 2.000 (xed)
500V
]5.01MΩ0.0 to 500.0 (xed)
250V
]5.01MΩ0.0 to 250.0 (xed)
110P
]1.11MΩ0.0 to 440.0 (xed)
50V
]1.11MΩ0.00 to 50.00 (xed)
20V
]200.92kΩ0.00 to 20.00 (xed)
10V
]200.92kΩ0.00 to 10.00 (xed)
2.5A
]0.02Ω0.000 to 2.500 (xed)
)5A
]0.20Ω0.0 to 500.0 (xed)
)25A
]0.20Ω0.0 to 250.0 (xed)
)1A
]1.02Ω0.0 to 100.0 (xed)
50MA
]1.02Ω0.00 to 50.00 (xed)
" is ashes with a certain measurement input, disconnect power supply and then check the cables.
DC voltage
DC current
AC voltage
AC current
When "
※
0-500V[
0-100V [
0-50V [
0-10V[
0-5V [5V]43.48kΩ0.000 to 5.000 (xed)
0-1V [1V]43.48kΩ0.000 to 1.000 (xed)
0-250mV [
0-50mV[
0-5A[5A]0.02Ω0.000 to 5.000 (xed)
0-2A[2A]0.02Ω0.000 to 2.000 (xed)
0-500mA [
0-200mA[
0-50mA[
4-20mA[
0-5mA[
0-2mA [
0-500V [
0-250V [
0-110V [
0-50V [
0-20V [
0-10V [
0-2V [2V]20.92kΩ0.000 to 2.000 (xed)
0-1V [1V]20.92kΩ0.000 to 1.000 (xed)
0-5A [5A]0.02Ω0.000 to 5.000 (xed)
0-2.5A [
0-1A[1A]0.10Ω0.000 to 1.000 (xed)
0-500mA [
0-250mA [
0-100mA [
0-50mA [
HHHH
" or
"LLLL
O-32
] Prescale display range [
STND
DOT
0
)0
)00
)000
(Display range is variable according
to decimal point position.)
※
Connect to the input terminals
whose 30% to 100% of the input
range includes the max. value of
the input range to measure.
When the max. input value
is under the 30% of the input
terminal range, display accuracy
is degraded.
When the max. input value is over
the 100%, it may result in input
terminal damage.
In case of 0 to 110V [
※
AC voltage range and using
P.T (potential transformer) for
440V/110VAC, if 110V is input,
and the unit displays 440V
automatically by preset scale
value for P.T user's convenient.
SCAL
Display range
-1999 to 9999
-199.9 to 999.9
-19.99 to 99.99
-1.999 to 9.999
110P
]
] of
Page 9
-|Transparent setting guide|-
Sold Separately
▣
Communication converter
SCM-WF48
( Wi-Fi to RS485·USB wireless
communication converter)
SCM-US48I
(USB to RS485 converter)
Display Units (DS/DA-T Series)
DS/DA-T Series
(RS485 communication input type display unit)
Multi Panel Meter
SCM-38I
(RS232C to RS485 converter)
SENSORS
CONTROLLERS
MOTION DEVICES
SOFTWARE
(J)
Temperature
Controllers
(K)
SSRs
DS16- T
※
Connect RS485 communication input type display unit (DS/DA-T Series) and RS485 communication output model of MT4Y/MT4W Series,
the display unit displays present value of the device without PC/PLC.
DS22/DA22- T
DS40/DA40- TDS60/DA60- T
▣ Functions
AC frequency measurement
[PA 1 group:
It measures input signal frequency when it is AC input.
It uses fixed decimal point [PA1 :
can be changed by setting and measured range of decimal
point position is as below chart. It is available to adjust the
upper gradient at [PA 1:
measure frequency normally, input signal, over 10% F.S. of
the measured range, should be supplied. Please select the
proper point of
Measurement range
①
Decimal point
position
Measurement
range
※
Accuracy of frequency measurement:
Below 1kHz, F.S. ±0.1rdg ±2-digit.
From 1kHz to 10kHz, F.S. ±0.3rdg ±2-digit.
INbH
②
③
: 0.100 to 9.999 [Gradient adjustment of high value]
INbE:
10
DISP
]
DOT
], measured range
INbH
]and [PA 1:
0.000 0.00 0.0 0
0.100 to
9.999Hz
-2
, 10-1, 100, 101 [Index adjustment of
0.10 to
99.99Hz
INbE
0.1 to
999.9Hz
]. In order to
1 to
9999Hz
INbH
]
Zero adjustment
[Deviation correction function of low limit
display value]
It adjusts the display value of the optional configured input
value as zero by force, zero point error can be adjusted
with 3 ways as below. When zero point adjustment with
front key and Hold terminal is finished normally, zero point
of measurement terminal is displayed and the adjusted
value at saved in
INbL
automatically.
Operation
Description
※
It sets transmission output for the display value at the
output current DC4-20mA.
It sets display value for 4mA at
and the range between
※
※
Refer to "
Input correction
value
PA 1: Direct
input correction
value method at
INbL
.
Error correction", " Error display" and
Front panel
key
Press ,
key for 3 sec
at the
RUN mode.
Input external signal
Short-circuit external
Hold terminal 11,12
[6, 7 (MT4W)] over
min. 50m.
" Parameter 2 Group" for function and error.
Transmission (DC4-20mA) output scale
[PA2 group:
When min. set interval between
set as under 10% F.S., it changed as over 10% F.S.
automatically.
Preset display value is fixed to output as 4mA at under
FS-L
and 20mA at over
Output
20mA
4mA
]
/
FS-L
FS-H
and
FS-L
FS-H
FS-H
.
Min.
setting range
F.S. 10%
FS-LFS-H
and 20mA at
should be 10%
FS-L
and
Display value
FS-H
is
(L)
Power
Controllers
(M)
Counters
(N)
Timers
(O)
Digital
Panel Meters
(P)
Indicators
(Q)
Converters
(R)
Digital
Display Units
(S)
Sensor
Controllers
(T)
Switching
Mode Power
Supplies
(U)
Recorders
(V)
HMIs
(W)
Panel PC
(X)
Field Network
Devices
O-33
Page 10
MT4Y/MT4W Series
-|Transparent setting guide|-
Initialization
It initializes as the factory default status.
If press , , keys together for 2 sec in RUN mode,
is displayed and press , , keys, (NO) will flash
INIT
every 0.5 sec. It will be initialized as the factory default
when press
key after change
MODE
NO→ YES
.
Startup compensation time
This time function limits
[PA 2 group:
STAT
]
the operation of an output until the measured input
(overvoltage or inrush current) is stable at moment
of power on. All outputs are off during startup
compensation time setting after power is applied.
Setting range: 00.0 to 99.9 (unit: sec)
Factory default: 00.0
Error display
Display Description
Flashes when measured input is exceeded the max. allowable
HHHH
input (110%)
Flashes when measured input is exceeded the min. allowable
LLLL
input (-10%)
Turns ON when display input is exceeded the max. display range (9999)
D-HH
or
setting value
H-SC
Turns ON when display input is exceeded the min. display
D-LL
range (-1999) or
Turns ON when input frequency is exceeded the max. display
F-HH
value of measurement range
Flashes when it exceeds zero range (±99)
OVER
※
Error display is released automatically when it is in the
measured and display range.
※
"
" is displayed when the input specification is DC.
LLLL
※
After flashing "
adjustment range, it returns to RUN mode.
Display scale [PA 1 group:
This function is to display setting (-1999 to 9999) of
particular High/Low-limit value in order to display High/Lowlimit value of measured input. If measured inputs are 'a' and
'b' and particular values are 'A' and 'B', it will display a=A,
b=B as below graphs.
B
Display
value
A
a b
Input value
Display
A
value
B
b
a
Input value
setting value
L-SC
" 2 times when it exceeds the zero
OVER
H-SC/L-SC
B
Display
value
Display
value
A
a
Input value
a
A
B
Input value
Display
value
b
b
Display
value
]
B
a
A
Input value
A
a
B
Input value
b
b
Gradient correction [PA 1 group:
INbH
]
This function is to correct a gradient of prescale value and
display value. (Fig.1) Display value Y can be used as α, β
times against X input value by correction function [
INbH
].
And also can be used as correction function of max.
display value (
). Adjustment range is 0.100 to 5.000
H-SC
and multiply current gradient.
E.g.) Input: DC200mV, Display: 3.000 for MT4W-DV
Display value for
αY
α time
Y
βY
Display value
0
(Fig. 1)
Select 0-1VDC[1V] for measurement input in Parameter1.
①
Standard specification in input: 0-1VDCᜡ and 1.000
②
X
β time
Input
value
therefore it has to be 15.000 (
measuring input
15.000
12.000
9.000
6.000
3.000
Display value
1.000
0.2 0.4 0.60.8 1V
(example of gradient correction)
H-SC
) for 1VDCᜡ (input)
Input
value
in order to display 3.000 for 200mVDCᜡ (input). But it
is disable due to setting range is 9.999
In this case, please check below chart.
③
Please set as
Setting
①
②
③
④
⑤
Error correction [PA 1 group:
INbH× H-SC
=15.000
H-SCL-SCINbH
Disable 0.000 1.000
7.500 0.000 2.000
5.0000.000 3.000
3.7500.000 4.000
3.0000.0005.000
Other
-
In this case, any setting
methods display the
same display value.
INbH
INbL
/
]
It corrects display value error of measured input.
: ±99 [Adjust deviation of low value]
INbL
: 5.000 to 0.100 [Correct gradient (%) of high value]
INbH
Display value= (measured value ×
INbH
) +
INbL
E.g.) When the measured range is 0 to 500V, and the
display range is 0 to 500.0. If the low display value is "
to 0V input, set -12 as
value to display ")0" by
INbL
1.2
adjusting offset of the low value. The display value to 500V
measured input varies by adjusting the offset of low value.
If this display value is "
", calculate 500.0/501.0
501.0
(desired display value/the display value), and set the 0.998
correction value as the
INbH
to display
50)0
by adjusting
gradient of high value.
※
The offset correction range of
for D-0, D
-1
digit regardless of decimal point.
is within -99 to 99
INbL
"
O-34
Display cycle delay [PA 2 group:
D1sT
]
In some applications the measured input may fluctuate
which in turn causes the display to fluctuate. By adjusting
the display cycle delay function time in the
D1sT
of
parameter 2, the operator can adjust the display time within
a range of 0.1 sec to 5 sec. For example, if the operator
sets the display cycle time to 4.0 sec, the display value
displayed will be the average input value over 4 sec and
also will show any changes if any every 4 sec.
Page 11
Multi Panel Meter
-|Transparent setting guide|-
Monitoring peak display value
[PA 0 group:
hPEK/lPEK
, PA 2 group:
It monitors max./min. value of display value based on the
current displays value and then displays the data at
lPEK
of parameter 0. Set the delay time (0 to 30 sec)
PEkT
at
of parameter 2 in order to prevent malfunction
caused by initial overcurrent or overvoltage, when
monitoring the peak value.
Delay time is 0 to 30 sec and it starts to monitor the peak
value after the set time. When pressing any one of
hPEK, lPEK
keys at
of parameter 0, the monitored data
is initialized.
※
hPEK, lPEK
monitoring delay time [
as 00 sec [
parameters is not displayed when
PEkT
00 S
].
] of parameter 2 group is set
PEkT
hPEK
Preset output operation mode
[PA 2 group:
Mode Output operationOperation
OFF
GO
lST
LO
H
HI
hST
GO
HI
LhST
GO
LO
H
HI
HhST
GO
LO
HI
GO
LlST
LO
GO
LdST
LO
※
"H" means hysteresis and able to set 1 to 99 at "
mode in PA 2 among above comparison output chart.
※
is displayed according to the setting of output
hSET
operation mode, when user sets "
not displayed.
※
Only
lST
MT4Y-☐-44 models.
H
H
H
setting is available in MT4Y-☐-43 and
OUtT
H
H
H
]
H: Hysteresis
No output
If it is equal or smaller than low
setting value, LO output will be ON.
If it is bigger than low setting value,
GO output will be ON.
If it is equal or bigger than high
setting value, HI output will be ON.
If it is equal or smaller than high
setting value, GO output will be ON.
If it is equal or smaller than low
setting value and equal or bigger
than high setting value, the output
will be ON. If it is bigger than Low
setting value and smaller than high
setting value, GO output will be ON.
If it is equal or bigger than low set
and equal or bigger than high set
value, output will be ON. If it is
smaller than low setting value and
high setting value, GO output will
H
be ON.
If it is equal or smaller than low
setting value, LO output will be ON.
If it is equal or smaller than high
setting value, HI output will be ON.
If it is bigger than low setting value
and High setting value, GO output
will be ON.
This operation is the same as
But it doesn't operate at initial low
set value, it will operate at next low
set value. If this is higher than low
set value, Go output will be ON.
",
OFF
hSET/ lSET
H
lST
"
HYS
Sub output
● RS485 communication output
]
It is able to set address (01 to 99)
It is able to transmit by selecting modulation speed
,
(transmitted number of signal per 1 sec) of serial trans
-mission. (selectable 1200, 2400, 4800, 9600bps)
● Low-speed serial output
It outputs current display value as Low-frequency (50Hz)
type.
● Transmission (DC4-20mA) output
It outputs DC4-20mA against High/Low-limit scale.
(resolution: 12000 division)
● BCD dynamic output
It outputs display value as BCD Code.
※
Only one sub-output is selectable.
(More than one sub-output is not allowed.)
Time chart of BCD dynamic output and
Serial output
● BCD dynamic output (negative logic)
Digit signal
Input Data
D3
D2
Digit
D1
D0
A
HLHLH
B
HLLHH
C
Data
LHHHH
D
HHHHH
Hdot
Display
● Serial output (negative logic)-Clock frequency:50Hz
are
Latch
Clock
Data
Input
ordering
Clock
Data
Latch
Display
Cn-1
Dn-1DnD1
12345
DP D C B A
HHLHH HHHLL LHHLH HHHHL
6 7 8 9 10
DP D C B A
t
tatw
Data
31.2ms
HLHH
tw
tat
CnC1
11 12 13 14 15
DP D C B A
tb
tb
16 17 18 19 20
DP D C B A
t: 7.8ms
ta: 0.5ms
tw: 5ms
tb: 2.3ms
t: 20ms
ta: 0.05ms
tw: 5ms
tb: 4.95ms
tw
ta
tb
1 2
DP D
SENSORS
CONTROLLERS
MOTION DEVICES
SOFTWARE
(J)
Temperature
Controllers
(K)
SSRs
(L)
Power
Controllers
(M)
Counters
(N)
Timers
(O)
Digital
Panel Meters
(P)
Indicators
(Q)
Converters
(R)
Digital
Display Units
(S)
Sensor
Controllers
(T)
Switching
Mode Power
Supplies
(U)
Recorders
(V)
HMIs
(W)
Panel PC
(X)
Field Network
Devices
O-35
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.