CONTROLS AND FUNCTIONS
7. VΩRANGE Zero/Relative Button. On voltage or ohm function,
press this button to select manual range.
On other functions, once this button is pressed, the current reading
shall be set to zero and be used as a zero reference value for all
other subsequent measurement. The function is also used to remove
offset value caused by residual magnetism remaining in the core
for the DC current measurement. The Zero/Relative function will
be disabled if the MAX/MIN button is pressed. This button is not
available in Duty, Hz, Continuity, Diode function.
8. LCD is a 3 3/4 digit Liquid Crystal Display with a maximum indication
of 3999. The functions included are units, symbols, bar graph, sign,
decimal points, low battery symbols, max/mim symbols, and zero
symbol. Press the zero button again to disable this function.
9. Battery Symbol appears if the battery voltage drops below the
minimum required voltage. Refer to page 3 for battery replacement.
10. Zero/Relative Symbol means a reference value has been
subtracted from the actual reading. The reading shown is an offset
value. Press the zero button again to disable this function.
11. Data Hold Symbol appears on the LCD when the hold button is
pressed.
12. Bar Graph has forty segments. It displays segments proportional to
the actual reading. Each segment represents one count.
13. MAX or MIN is displayed on the LCD when the MIN/MAX button is
pressed.
14. Continuity Symbol. If continuity function is selected, this symbol
shall appear on LCD.
15. Units are displayed on the LCD when a function is selected. The
corresponding symbols are: V, A, Ω, and Hz.
16. V, Ω, Hz, Capacitance, Temperature input Terminal.
This terminal is used as input for voltage, ohm/continuity, frequency,
capacitance or temperature measurements..
17. COM Terminal (Black) is used as a common reference input.
18. Hand Strap allows you to put your hand through to avoid
accidentally dropping the clamp meter.
SPECIFICATIONS (23oC 5oC)
16
5
Range Resolution Accuracy Overload Protection
40A 10mA
+
1.5% +2dgts DC 400A
0-150A 100mA
+
1.5% +2dgts DC 400A
150-200A 100mA
+
2.2% +2dgts DC 400A
DC Current
- -
- -
- -
AC Current
Range Resolution Accuracy Overload
Protection
50/60 Hz 40-1 KHz
40A 10mA
+
1.5% +2dgts+2.5% +2dgts AC 400A
0-150A 100mA
+
1.5% +2dgts+2.5% +2dgts AC 400A
150-200A 100mA
+
2.2% +2dgts+2.5% +2dgts AC 400A
- -
- -
- -
- -
- -
- -
DC Voltage: (Overload Protection: 600V)
Range Resolution Accuracy Input Impedance
400mV 0.1mV
+
1.5% +3dgts 2.3MΩ
4V 1mV
+
1.5% +3dgts 2.3MΩ
40V 10mV
+
1.5% +3dgts 2.3MΩ
400V 100mV
+
1.5% +3dgts 2.3MΩ
600V 1V
+
1.5% +3dgts 2.3MΩ
- -
- -
- -
- -
- -
Range Resolution Accuracy Overload
Protection
50/60 Hz 40-1 KHz
0.01-4V 1mV
+
1.5% +5dgts+2.0% +5dgts AC 800V
40V 10mV
+
1.5% +5dgts+2.0% +5dgts AC 800V
400V 100mV
+
1.5% +5dgts+2.0% +5dgts AC 800V
600V 1V
+
1.5% +5dgts+2.0% +5dgts AC 800V
AC Voltage: (Input Impedance 2.3MΩ)
- -
- -
- -
- -
- -
- -
- -
- -
+
-