CONTROLS AND FUNCTIONS CONT.
6. Max/Min Hold Button is used to enable the maximum or minimum
value to be displayed and updated during measurement. Press once,
minimum value will be displayed and updated. Press again, maximum
value will be displayed and updated. Press the third time and the meter
will return to normal measurement mode. The Zero function will be
disabled if MAX/MIN is enabled.
7. Zero/Relative Button sets the current reading to zero and is
used as a zero reference value for all subsequent measurements. This
function is also used to remove offset values caused by the residual
magnetism remaining in the core after a DC current measurement.
The Zero/Relative function will be disabled if the MAX/MIN button is
pressed.
8. LCD is a 3 3/4 digit Liquid Crystal Display with a maximum indication
of 3999. The functions included are units, symbols, bar graph, sign,
decimal points, low battery symbols, MAX/MIN symbols, and zero
symbol.
9. Battery Symbol appears if the battery voltage drops below the
minimum required voltage. Refer to page 3 for battery replacement.
10. Zero/Relative Symbol means a reference value has been
subtracted from the actual reading. The reading shown is an offset
value. Press and hold the zero button for 2 seconds to disable this
function.
11. Data Hold Symbol appears on the LCD when the hold button is
pressed.
12. Bar graph has forty segments. It displays segments proportional to
the actual reading. Each segment represents one count.
13. MAX or MIN is displayed on the LCD when the MAX/MIN button is
pressed.
14. Continuity Symbol appears on the LCD if the ohm and continuity
function is selected.
15. Units are displayed on the LCD when a function is selected. The
corresponding symbols are: V, A, W, and Hz.
16. VWHz Input Terminal (Red) is used as input for voltage, ohm/
continuity or frequency measurements.
17. COM Terminal (Black) is used as a common reference input.
18. Hand Strap allows you to put your hand through to avoid accidental
drop of the clamp meter.
STARTER DRAW TEST
STARTER
SETUP
1. Check with manufacture's proper procedure to disable ignition if
needed. Improper methods could cause damage to vehicles that
are computer controlled.
2. Determine the manufactures Starter Draw specications. If not
available use the following chart as a guide.
3. Select DC 200 Amp and zero the DM-40 by pressing the yellow
button.
4. Press the MAX/MIN button twice and make sure the display
reads HOLD MAX.
5. Clamp the DM-40 around the positive conductor leading to the
starter (See illustration above). Make sure the small arrow inside
the clamp is pointing in the direction of the current towards the
starter. Also make sure the DM-40 is clear of any moving belts,
fan blades or anything that might damage the meter.
6. Start or crank the engine for 3 to 5 seconds.
7. Read the DM-40 and compare with manufactures specications.
8. High Amp reading indicates a faulty starter or engine timing
that would put an extra load upon the starter. Possible starter
problems are shorted windings, bent armature, broken housing or
bad bearings.