AtriCure cryoICE BOX ACM1, cryoICE BOX ACM2 User Manual

cryoICE BOX
Version 6
USER’S MANUAL
European Representative: Herbert Köntges Köntges SPRL Avenue Hellevelt 35 B-1180 Brussels Belgium Tel: +32 (0) 2 375 51 63 FAX: +32 (0) 2 375 89 06 email: herbert.kontges@skynet.be
i
Manufactured by: AtriCure, Inc. 7555 Innovation Way, Mason, Ohio 45040 USA
1.866.349.2342
P001323 Rev A
Table of Contents
FOREWORD .................................................................................................................................... IV
CAUTION ......................................................................................................................................... IV
IMPORTANT ................................................................................................................................... IV
INDICATIONS FOR USE .............................................................................................................. IV
DISCLAIMER .................................................................................................................................. IV
PATENT INFORMATION .............................................................................................................. V
WARNINGS AND CAUTIONS ....................................................................................................... V
WARNINGS ...........................................................................................................................v
CAUTIONS .......................................................................................................................... vi
SAFETY CSA MARK INFORMATION ............................................................................. ix
1. SYSTEM OVERVIEW ............................................................................................................... 1
The AtriCure cryoICE BOX ...................................................................................................1
AtriCure cryoICE BOX Front and Rear Panels – Illustrations and Nomenclature ................1
Operating Modes .....................................................................................................................3
READY Mode .................................................................................................................... 3
FREEZE Mode.................................................................................................................... 3
DEFROST Mode ................................................................................................................ 3
FAULT Condition ............................................................................................................... 3
2. TECHNICAL SPECIFICATIONS ........................................................................................... 4
Mechanical Specifications ......................................................................................................4
Electrical Specifications..........................................................................................................4
Mains Fuses ............................................................................................................................4
Footswitch Specifications .......................................................................................................4
Equipment Type / Classification .............................................................................................4
3. ATRICURE CRYOICE BOX, DETACHABLES, AND ACCESSORIES ........................... 5
AtriCure cryoICE BOX Set-Up and Preparation ....................................................................6
N2O Cylinder Installation .......................................................................................................6
Exhaust Tubing .......................................................................................................................8
Heater Band Installation .........................................................................................................8
Turning On the AtriCure cryoICE BOX .................................................................................9
Resetting the N2O Gas Gauge ...............................................................................................10
System Check........................................................................................................................11
4. DEVICE USE ............................................................................................................................. 11
Install AtriCure cryoICE Probe ............................................................................................11
Set Ablation Time .................................................................................................................13
Start Ablation ........................................................................................................................13
ii
5. SPECIAL CASES ...................................................................................................................... 14
Abort FREEZE......................................................................................................................14
Change Ablation Time during Ablation ...............................................................................14
Emergency Stop ....................................................................................................................14
Set Default Ablation Time ....................................................................................................14
Operate Without Temperature Reading ................................................................................14
6. SYSTEM DISASSEMBLY AFTER USE ............................................................................... 15
Disconnecting the AtriCure cryoICE Probe .........................................................................15
N2O Cylinder Removal .........................................................................................................15
7. PREVENTIVE MAINTENANCE AND CLEANING OF THE ATRICURE CRYOICE
BOX ............................................................................................................................................. 16
Cleaning and Disinfecting Instructions .................................................................................16
Preventive Maintenance ........................................................................................................17
AtriCure Address / Toll Free Telephone Number ................................................................18
Corporate Website ................................................................................................................18
Customer Service/ Product Inquiries ....................................................................................18
Replacement of AC Line Fuses ............................................................................................18
Tank Hose Assembly without canisters – Standard ..............................................................20
Tank Hose Assembly with canisters – Alternate (Replacement of Desiccant Filter) ...........21
Disposal.................................................................................................................................22
8. TROUBLESHOOTING ........................................................................................................... 23
AtriCure cryoICE BOX Error Codes ....................................................................................27
9. ELECTROMAGNETIC COMPATABILITY TABLES ..................................................... 28
Electromagnetic Emissions ...................................................................................................28
Electromagnetic Immunity – Enclosure Port ........................................................................28
Electromagnetic Immunity – Input A.C. Power Port ............................................................29
Electromagnetic Immunity – Input D.C. Power Port – Not Applicable ...............................30
Electromagnetic Immunity – Patient Coupling Port .............................................................30
WARRANTIES ................................................................................................................................. 31
iii

FOREWORD

This manual and the equipment it describes are for use only by qualified medical professionals trained in the particular technique and surgical procedure to be performed. The AtriCure cryoICE BOX also referred to as the AtriCure Cryo Module (ACM) consisting of two model units: ACM1 and ACM2.

CAUTION

Federal (USA) law restricts this device to sale by or on the order of a physician.
Please read all information carefully. Failure to properly follow the instructions may lead to serious surgical consequences including patient and caregiver harm.

IMPORTANT

This user manual is designed to provide instructions for use of the AtriCure cryoICE BOX (A000896-3 & A000897-3 assembly/A000898-3 & A000899-3 packaged assembly) with the AtriCure cryoICE probes and AtriCure Detachable and Accessory Devices (See page 22 of this manual for specific part numbers information.) This user manual is not a reference to surgical technique.

INDICATIONS FOR USE

The AtriCure cryoICE BOX is a non-sterile, reusable device which delivers cryogenic energy, namely nitrous oxide, to AtriCure’s cryo-ablation probes.
The AtriCure cryoICE BOX unit is an electro-mechanical cryogenic surgical unit that delivers a cryogenic Nitrous Oxide (N2O) energy source to a cryoICE probe to create lines of ablation through tissue. The AtriCure cryoICE BOX is part of a system which includes the N2O gas cylinder, N2O gas line hose, N2O exhaust hose, cylinder heater band, an optional footswitch, and single-use cryoICE probe. The system provides controlled lesion forming temperatures below -40°C, with typical operating ranges between -50°C to -70°C.
The AtriCure cryoICE BOX is designed to operate only with AtriCure designed and developed cryoICE probes. The AtriCure cryoICE probe will be referred in this User’s Manual as the “cryoICE probe”.
This User’s Manual provides a description of the AtriCure cryoICE BOX, its controls, displays, indicators, and a sequence for its operation with the cryoICE probe. This User’s Manual also supplies other information of importance to the user. For information about the cryoICE probes, please refer to the cryoICE probe Instructions for Use.
Do not operate the AtriCure cryoICE BOX before thoroughly reading this manual.

DISCLAIMER

Users assume responsibility for approving the acceptable condition of this product before it is used, and for ensuring that the product is only used in the manner described in these instructions
iv
for use. Under no circumstances will AtriCure, Inc. be responsible for any incidental, special or consequential loss, damage, or expense, which is the result of the deliberate misuse of this product, including any loss, damage, or expense which is related to personal injury or damage to property.

PATENT INFORMATION

May be covered by one or more patents.

WARNINGS AND CAUTIONS

The safe and effective use of the cryo device and equipment is highly dependent upon factors under the control of the operator. There is no substitute for a properly trained operating room staff. It is important that the operating instructions supplied with the AtriCure cryoICE BOX unit be read, understood, and followed before use.

WARNINGS

• Note: Do not operate the cryoICE BOX unit before thoroughly reading this manual.
Note: Do not use cryo surgical equipment unless properly trained in the specific.
procedure being undertaken. This manual and the equipment it describes are for use only by qualified medical professionals trained in the particular technique and surgical procedure to be performed.
Fire Hazard: Do not use extension cords.
Trip Hazard: Standard care should be used to reduce the risk of tripping on the
Footswitch cable, as well as the N2O exhaust hose.
• No modification of this equipment is allowed.
The voltage selector is factory set and should not be changed by the user. The voltage
setting and the fuse rating must be appropriate as identified to prevent cryoICE BOX malfunction and potential instrument damage.
Electric Shock Hazard: Connect the cryoICE BOX power cord to a properly grounded
receptacle. Do not use power plug adapters.
Electric Shock Hazard: Do not connect wet accessories to the generator.
Electric Shock Hazard: Ensure that the cryoICE probe is correctly connected to the
cryoICE BOX and that no thermocouple wires are exposed from the cable, connector, or the cryoICE probe.
Use of accessories, transducers and cables other than those specified or provided by AtriCure could result in increased electromagnetic emissions or decreased electromagnetic immunity of the cryoICE BOX and result in improper operation
Use of the cryoICE BOX adjacent to or stacked with other equipment should be avoided because it could result in improper operation
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of
v
the cryoICE BOX, including cables specified by the AtriCure. Otherwise, degradation of the performance of this equipment could result.
Note: The Emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not offer adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or reorienting the equipment.

CAUTIONS

Use only with the cryoICE probes intended for use with the cryoICE BOX.
Do not transition into FREEZE mode until the cryoICE probe is properly positioned at the
ablation site.
The system status indicators and displays are important safety features. Do not obstruct either the ablation or the system status indicators.
Do not remove the cryoICE BOX cover as there is a potential for electrical shock. Refer to authorized personnel for service.
The Power Cord of the cryoICE BOX must be connected to a properly grounded receptacle. Extension cords and/or adapter plugs must not be used.
Do not contact cryoICE probes with a RF device.
Compressed Air Hazard: Do not operate N2O cylinders with a pressure greater than 1000
PSIG (6900 kPa).
Nitrous Oxide connections should only be unplugged when the cryoICE BOX is in the READY mode and properly vented.
vi
Meanings of Symbols on AtriCure cryoICE BOX
Power Off
Caution
Alternating Current
Equipotential Terminal
Type CF Applied Part (cryoICE Probe)
READY
FREEZE
DEFROST
N2O Gas Gauge
Timer
Timer Increase Button
Timer Decrease Button
cryoICE Probe Temperature
Thermocouple/Probe
Cylinder Valve On/Off
N2O Gas Gauge Reset
Gas Exhaust
vii
Maintenance Needed
Fault Condition
Cylinder Heater Band
Footswitch
Maximum Pressure
Gas Inlet
Gas Outlet
Non-Sterile
Manufacturer
Catalog Number
Serial Number
Prescription Only
Follow Instructions for Use
Humidity and Temperature Storage Limits
Waste Electrical and Electronic Equipment (WEEE)
viii
Classification in accordance with EN 60601-1

SAFETY CSA MARK INFORMATION

CLASS 8750 01 - MEDICAL ELECTRICAL EQUIPMENT/SYSTEMS CLASS 8750 81 - MEDICAL ELECTRICAL EQUIPMENT/SYSTEMS - Certified to US
Standards Cryogenic Ablation Device, Model AtriCure Cryo Module, ACM1 & ACM2, cord connected/ appliance coupler / transportable, rated: 115/230Vac, 4/2A, 50/60 Hz
1. Type of protection against electric shock: Class I
2. Degree of protection against electric shock: Type CF
3. Degree of protection against ingress of water: IPX0
4. Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide
5. Mode of operation: Continuous Environmental Conditions: Normal: 10-40°C, 15-90% rH, 980-1050mb
ix
1. Activation Button
9. cryoICE Probe Gas Outlet Port
2. N2O Gas Gauge Indicator Display
10. cryoICE Probe Gas Inlet Port
3. N2O Gas Gauge Indicator Display Reset
11. Ablation Status Indicator
4. Ablation Timer Display
12. Cylinder Heater Band Indicator
5. Ablation Timer Decrement
13. Maintenance Needed Indicator
6. Ablation Timer Increment
14. System Fault Indicator
7. cryoICE Probe Temperature
15. Thermocouple Open Indicator
8. Future Probe Connection
16. cryoICE Probe Thermocouple Ports

1. SYSTEM OVERVIEW

The AtriCure cryoICE BOX

This section provides a detailed description of the cryoICE BOX including its function and operating features.
The cryoICE BOX unit is an electro-mechanical cryogenic surgical unit that delivers a Nitrous Oxide (N2O) cryogenic energy source to a cryoICE probe to create lines of ablation through tissue. The cryoICE BOX is part of a system which includes the N2O cylinder, N2O gas line hose, N2O exhaust hose, cylinder heater band, an optional footswitch, and single-use cryoICE probe. The system provides controlled lesion forming temperature that is below -40°C, with typical operating ranges between -50°C to -70°C.
Along with the Activation Button on the front panel of the cryoICE BOX, an optional Footswitch can also be used to activate and terminate the cryo ablation cycle.
The cryoICE BOX is designed to operate only with AtriCure cryoICE probes. Refer to the cryoICE probe Instruction for Use for complete description and indications for use of
these devices.

AtriCure cryoICE BOX Front and Rear Panels – Illustrations and Nomenclature

Illustrations of the cryoICE BOX front panel (Figure 1) and rear panel (Figure 2) are shown below.
Figure 1: AtriCure cryoICE BOX Front Panel
1
17. N2O Exhaust Port
23. Power Plug Receptacle
18. N2O Manual Exhaust Knob
24. Power Switch
19. N2O Inlet Port
25. Mains Fuse Location
20. N2O Exhaust Switch
26. cryoICE BOX Voltage Rating Label
21. Activation Footswitch Connection Port
27. Equipotential Terminal
22. Heater Band Cord Receptacle
28. RS232 Data Connection
Figure 2: AtriCure cryoICE BOX Rear Panel (Domestic & International)
2

Operating Modes

The cryoICE BOX operates in one of three modes: READY, FREEZE, and DEFROST. These modes are identified by the system status indicator LEDs and the ablation status indicator LEDs located on the front of the cryoICE BOX unit.

READY Mode

This mode is entered automatically upon successful execution of Power-on-self ­test when the unit is first turned on or following DEFROST Mode upon the cryoICE probe reaching approximately 10°C and automatically venting. This indicates that the system is ready for the next cryo ablation run.

FREEZE Mode

This Mode is entered from the READY Mode when the user initiates the cryo ablation cycle by pressing and releasing the Activation Switch or the Footswitch. In this mode, the N2O gas is allowed to cycle through the cryoICE probe causing a temperature drop to take place at the cryoICE probe.

DEFROST Mode

This Mode is entered automatically from FREEZE Mode upon expiration of the ablation timer, or manually by the operator when the Activation Switch or the Footswitch is actuated while in the FREEZE Mode. In this mode, the cryoICE probe temperature is actively forced towards the ambient temperature.
Once the cryoICE probe temperature is above approximately 10°C, the cryoICE BOX unit will transition back to the READY Mode.
Note: cryoICE BOX does allow early transition out from the DEFROST Mode into
either the READY Mode or the FREEZE Mode by pressing the Activation Button.
Note: cryoICE probe temperature may drop temporarily upon transition from
DEFROST to READY state.

FAULT Condition

This is entered upon detection of any unrecoverable error condition during any Mode. The system is inoperable in this Mode until the unit is first power cycled, and only if the fault condition no longer exists or has been remedied.
3
Size:
17.5 in (44.5 cm) - (W) × 27.0 in (68.6 cm) - (D) × 4.5 in (11.4 cm) - (H) maximum
Weight:
45 lb. (20.4 kg) absolute maximum
Temperature
Humidity
Atmospheric
Operational
+10°C to +40°C
15% to 90%
980mb to 1050mb
Storage
-28°C to +37°C
15% to 90%
980mb to 1050mb
Transit
-28°C to +37°C
15% to 90%
980mb to 1050mb

2. TECHNICAL SPECIFICATIONS

Mechanical Specifications

Environmental Specifications
pressure
Temperature
+50°F to +104°F
-20°F to +100°F
-20°F to +100°F
relative humidity
relative humidity
relative humidity

Electrical Specifications

AtriCure cryoICE BOX, model ACM1 – 115 (100-120)VAC, 4A, 50/60 Hz
AtriCure cryoICE BOX, model ACM2 – 230 (220-240)VAC, 2A, 50/60 Hz

Mains Fuses

AtriCure cryoICE BOX, model ACM1 – 115 (100-120)VAC, 4A, 50/60 Hz
Replace fuses as marked: 4.0A/250V, T-lag, 5 × 20 mm, UL Recognized, IEC Approved
AtriCure cryoICE BOX, model ACM2 – 230 (220-240)VAC, 2A, 50/60 Hz
Replace fuses as marked: 2.0A/250V, T-lag, 5 × 20 mm, UL Recognized, IEC Approved

Footswitch Specifications

Moisture protection rating: IP68

Equipment Type / Classification

Class 1 Equipment
4
A
E
D
C
B

3. AtriCure cryoICE BOX, DETACHABLES, AND ACCESSORIES

As shown in Figure 3, the system is comprised of the following:
A: AtriCure cryoICE BOX Cylinder Heater Band (CMH15 or CMH22) - Detachable
B: AtriCure cryoICE BOX Tank Hose Assembly, without Canister Set, Standard - Detachable
C: AtriCure cryoICE BOX N2O Exhaust Hose - Detachable
D: AtriCure cryoICE BOX Tank Hose Assembly, with Filter Canister Set (Optional) -
Detachable
E: AtriCure cryoICE BOX
F: AtriCure cryoICE BOX Footswitch (Optional – not shown) - Accessory
G: AtriCure cryoICE BOX Power Cord (not shown) - Detachable
H: AtriCure cryoICE probe with integral tube set (not shown) - Type CF Applied Part
I: AtriCure cryoICE BOX Heater Band Extension Spring (not shown) – Detachable
Figure 3: AtriCure cryoICE BOX, Detachables, and Accessories
5

AtriCure cryoICE BOX Set-Up and Preparation

This section will outline the preliminary set-up for the cryoICE BOX, including cylinder installation, heater band installation, turning on the cryoICE BOX, and resetting the cylinder gauge on the cryoICE BOX user interface. Note: The cryoICE BOX should be set up at least 15-minutes prior to the procedure to allow time for the heater to warm the N2O cylinder to operating temperature.

N2O Cylinder Installation

Use only nitrous oxide gas with a water content not exceeding 3 ppm. Automotive grade nitrous oxide should not be used due to the inclusion of hydrogen sulfide.
The cryoICE BOX is designed to use 20-pound (9-kg.) cylinders.
Always install a completely full cylinder so the cylinder volume can be indicated
correctly.
To install a new N2O cylinder, first find the N2O gas line receptacle on the rear panel and connect this end into the corresponding end of the N2O gas line. Insert and push in the connector until you hear it “click” in place and the connection is fully seated and secured from unlatching as seen below in Figure 4.
Figure 4: N2O Inlet Connection
Next, match the opposite black knob end of the N2O gas line with the threaded connection port of a new N2O gas cylinder.
Screw the cryoICE BOX gas line into place by hand tightening the knob as shown in Figure 5. Over tightening this fitting with a wrench may cause damage, allowing N2O gas to leak.
To open gas cylinder valve, slowly turn the knob on the top of the cylinder counter- clockwise as seen in Figure 6.
6
Figure 5: Attach Black Knob to Threaded Connection
Figure 6: Turn Valve Counter-Clockwise to Open
Listen for leaks. If a leak is detected, tighten the black knob with a wrench if needed.
If the Low-Pressure indicator, as seen in Figure 7, illuminate’s amber this indicates that the cryoICE BOX is not detecting proper pressure. Check to ensure that the gas cylinder valve is open fully and that the cylinder you have attached is not empty.
7
Figure 7: Low Pressure Indicator

Exhaust Tubing

Note: Ensure the Exhaust tubing (hose) is firmly attached to the cryoICE BOX N2O
exhaust port, see Figure 2 item 17.
Be sure to route the N2O vent tubing to a safe area prior to use.
If a scavenger system is used, it must be able to accommodate a continuous flow of
60-liters per minute.

Heater Band Installation

Ensure the cryoICE BOX is properly connected to an N2O gas cylinder.
Place heater band with the cord facing upward.
Secure all tensioning spring retainers around the gas cylinder, starting with the very
bottom and very top retainers and then proceed to secure the middle retainers as shown in Figure 8.
The Heater band must be positioned less than 2-inches (5-cm) from bottom of the cylinder to ensure that the N2O is heated efficiently.
Plug heater band cord into the appropriate indicated receptacle located on the rear panel of the cryoICE BOX unit as shown in Figure 9.
Verify that the Cylinder Heater Band Icon on the front of the unit is not illuminated.
8
Figure 8: Secure All Tensioning Spring Retainers
Figure 9: Plug Heater Band Cord into Receptacle

Turning On the AtriCure cryoICE BOX

Plug in the cryoICE BOX unit into an approved hospital outlet.
Turn-On the cryoICE BOX unit with the switch located on the back as seen in Figure 10.
The power switch is used to connect mains power (Turn-On) or disconnect mains power (Turn-Off) to the cryoICE BOX unit.
After powering up, the Activation Button on the front of the cryoICE BOX interface will be illuminated. If no light is observed, check for proper power cord connection and switch position.
9
Figure 10: Turn-On AtriCure cryoICE BOX with Switch

Resetting the N2O Gas Gauge

Only reset the gauge when a new full cylinder has been installed.
Ensure cryoICE BOX is powered on.
Ensure the unit is in READY mode.
Find the gas cylinder display on the front of the cryoICE BOX and note the RESET
button to the right of this display, see Figure 11.
Press and hold the RESET button for one second.
Note: Once the N2O gas gauge is reset, the display can take up to several minutes to
display the remaining volume in the tank.
The gauge can only be reset to full after a system power cycle or following a cylinder swap out. If RESET button is pressed following usage the gauge will reset to the estimated cylinder volume.
Figure 11: N2O Gas Gauge RESET Button
10
Meaning of gas gauge indicators seen in Figure 12
Figure 12: N2O Gauge Indicators
3-Segments On = Approximately 20-40 minutes remaining 2-Segments On = Approximately 15-20 minutes remaining 1-Segment On = Approximately 5-10 minutes remaining 1-Segment Flashing = Approximately 5 minutes or less remaining – CHANGE TANK

System Check

Verify neither the Maintenance Needed or System Fault icons are illuminated.

4. DEVICE USE

Install AtriCure cryoICE Probe

1. Ensure cryoICE BOX is properly connected to a N2O gas cylinder.
2. The cryoICE probe may be connected before the cryoICE BOX has been turned on, while the cryoICE BOX is being turned on, or when the cryoICE BOX unit is on and in READY mode.
3. Insert the corresponding connections on the pneumatic connectors as shown below in Figure 13. The sliding ring will need to be manually pushed-in on the orange connector.
Figure 13: Color Coded Pneumatic Connectors
11
Figure 15
4. Ensure each pneumatic connection is fully seated by listening for an audible “click” as each connector engages its receptacle. Gently tug on each tube to ensure proper engagement with connector.
5. Insert the corresponding red and black colored connections into the thermocouple connectors, see Figure 15.
Figure 14
6. The cryoICE probe icon, seen above in Figure 14, will extinguish if the cryoICE probe is functioning properly and the approximate room temperature will be displayed on the temperature display (typically 10 to 25° C). An example of this is shown in Figure 16.
Figure 16: Probe Temperature Display
7. A test run is advised to ensure the cryoICE probe and system is working properly prior to the case.
8. Pneumatic connectors should only be unplugged when the cryoICE BOX is in the READY mode.
12

Set Ablation Time

1. The time of ablation is displayed in the middle of the interface of the cryoICE BOX and is indicated by a clock underneath the display. The display shows the time of ablation in seconds, see Figure 17.
Figure 17: Ablation Time Display
2. To change the duration of the ablation, press either of the up or down arrows to the right of the time display. The display will change in increments of ten seconds. The timer will reset to the default setting after a single cycle has been run.

Start Ablation

1. Ensure cryoICE BOX is powered on and the cryoICE probe and N2O is connected properly.
2. Check that desired ablation time is displayed, change if needed.
3. Press and release the Activation Button at the left of the device to begin the ablation.
4. The temperature display on the front panel displays the cryoICE probe temperature. A double-beep will indicate that the therapeutic temperature has been reached (typically -40°C), and the ablation timer will begin to count down. A short beep will sound every 30 seconds. A series of beeps will indicate the last 5-seconds of the Ablation cycle.
5. At the conclusion of the Ablation cycle, the cryoICE BOX will automatically transition into the DEFROST mode. The DEFROST indicator will illuminate indicating probe warming until it has reached the transition temperature which ends DEFROST, then the unit will automatically transition into READY and vent the probe. During the DEFROST cycle, a triple-beep will alert the user that the temperature of the probe has transitioned above 0°C degrees.
13

5. SPECIAL CASES

Abort FREEZE

To stop ablation during a FREEZE cycle, press and release the Activation Button during the ablation. The system will then transition into DEFROST mode.

Change Ablation Time during Ablation

To change the current ablation time, the up and down arrows can be used to add or decrease time in 10 second increments.

Emergency Stop

To stop ablation and depressurize the cryoICE probe during a FREEZE or DEFROST push the Activation Button to vent cryoICE probe until the cryoICE BOX system has sequenced into READY mode.
The unit can also be stopped by turning off power in the back of the unit or unplugging it from the AC power outlet. The flow of N2O will stop, however gas will be trapped within the cryoICE probe and the cryoICE BOX. This gas will be vented the next time the cryoICE BOX is powered on.

Set Default Ablation Time

1. Ensure cryoICE BOX is powered on.
2. Press and hold both up and down arrows simultaneously for one second to initiate the mode that allows a change to the default ablation time.
3. The time display will flash and the default time can now be changed by using the up or down arrows. The time will change in increments of 10 seconds. The time cannot be set lower than 20 seconds, nor higher than 270 seconds.
4. To save the set default time, the display will stop flashing after 5 seconds and the new default will be set.

Operate Without Temperature Reading

If the cryoICE BOX does not display a temperature and the cryoICE probe is properly plugged in (red and black connectors) the cryoICE probe should not be used. If the Activation Button is pressed with this condition, the cryoICE BOX will flash and beep for 5- seconds. If the Activation Button is pressed again within 5-seconds, the cryoICE BOX will sequence into FREEZE mode and the counter will start the countdown immediately. This should only be done at the discretion of a physician as there will not be temperature feedback.
14

6. SYSTEM DISASSEMBLY AFTER USE

Check to see that the service icon is not illuminated. If so, notify AtriCure service to correct the problem.

Disconnecting the AtriCure cryoICE Probe

1. The cryoICE probe can only be removed in the READY mode.
2. Remove the cryoICE probes pneumatic connections by pushing in the sliding ring on the receptacle while pulling out the cryoICE probe side of the connector.
3. Remove the black and red connections for the thermocouples.

N2O Cylinder Removal

1. Turn-Off the N2O cylinder by turning the knob clockwise.
2. Purge the N2O from the system by pressing and holding the N2O Exhaust Switch in the back of the unit. Watch the pressure gauge on the cylinder to see that all the pressure has been released. If the cryoICE BOX is powered off, pull and hold the N2O Manual Exhaust Knob until the pressure is relieved.
3. Disconnect the gas cylinder inlet fitting on the back of the cryoICE BOX by sliding the collar back.
4. Disconnect the hose from the N2O cylinder by unscrewing the black knob.
5. Turn-Off power and unplug the cryoICE BOX.
15

7. PREVENTIVE MAINTENANCE AND CLEANING OF THE AtriCure cryoICE BOX

Cleaning and Disinfecting Instructions

Note: Do not spray or pour liquids directly on the unit. Note: The unit and/or accessories cannot be sterilized. WARNING: Ensure Isopropyl Alcohol (IPA) is completely dry before operating the unit. CAUTION: Avoid caustic or abrasive cleaners
Guidelines
The following guidelines are recommended for cleaning the unit. It is the user’s responsibility to qualify any deviations from these processing methods.
1. Disconnect the unit or cart from the outlet before cleaning.
2. If the unit and/or accessories are contaminated with blood or other body fluids, they shall be cleaned before the contamination can dry (within two hours of contamination).
3. The outer surfaces of the unit and/or accessories shall be cleaned with 70% -90% Isopropyl alcohol (IPA) wipes for a minimum of two minutes. Do not allow fluids to enter the chassis.
4. Pay attention to all areas where fluids or soil may gather, such as under/ around the handles or any tight crevices/ grooves.
5. Dry the unit and/or accessories with a dry, white lint-free cloth.
6. Conduct a final confirmation of the cleaning process by visually inspecting the white cloth for remaining soil.
7. If soil remains on the white cloth, repeat steps 3 through 6.
8. After cleaning is complete, turn the unit on to perform Power On Self-Test (POST). If
any errors are received, contact AtriCure to begin return process.
16

Preventive Maintenance

AtriCure service representatives or the hospital biomed personnel shall conduct annual preventative maintenance procedures to ensure all cryoICE BOX components are functioning as defined within this manual. Pay particular attention to operational and safety features, including but not limited to:
Electrical power cords for fraying, damage, and proper grounding
• AC power switch
Any front panel display damage including switches, numeric displays and indicator lights.
cryoICE probe electronic interface connector damage, cracking or inability to insert
and latch cryoICE probe connector.
cryoICE probe pneumatic interface connector damage or inability to insert and latch cryoICE probe pneumatic connector.
• Carrying handle damage or inability to fold.
Rubber feet damage, cracking or inability for the cryoICE BOX to remain stable on a
flat surface.
Rubber alignment cup damage, cracking or inability for the ASB/ASU to remain stable atop cryoICE BOX and within the alignment cup.
Listen for leaks when pressurized.
Other medical equipment that may be used simultaneously with the cryoICE BOX
should also be inspected for damage. Specifically, check for insulation damage of electrical cables and associated connectors.
17
AtriCure cryoICE
ACM1
T 4A H 250V
Schurter
0001.2510
ACM2
T 2A H 250V
Schurter
0001.2507
The cryoICE BOX does not have any customer serviceable parts aside from mains fuses and gas line desiccant filter for cryoICE BOX units so equipped. For servicing issues, contact AtriCure, Inc. at:

AtriCure Address / Toll Free Telephone Number

AtriCure, Inc. 7555 Innovation Way, Mason, Ohio 45040 USA
1.866.349.2342

Corporate Website

www.atricure.com

Customer Service/ Product Inquiries

Telephone: 513-755-4100 866-349-2342 Toll Free Fax: 513-755-4567

Replacement of AC Line Fuses

Tools and Parts
• Needle Nose Pliers
Fuses
BOX Model
The cryoICE BOX unit has been pre-set at the factory to a nominal voltage of 115V (ACM1) or 230V (ACM2). The Rating Label below the Power Entry Module on the back panel of the cryoICE BOX indicates the selected Input Voltage for this unit. This setting should only be adjusted by the manufacturer or by an authorized AtriCure technical service representative. Note: cryoICE BOX unit should be powered off and unplugged before continuing with the fuse replacement procedure.
Fuse Type Manufacturer Part Number
Procedure to Replace AC Mains Fuses
1. Determine the fuse type by looking at the cryoICE BOX Model Number or the cryoICE
BOX Rating Label.
18
2. Using the needle nose pliers, carefully extract the fuse box from the power entry
module by squeezing down on the fuse box tabs in the slots as shown in Figure 18.
Figure 18: Fuse Box Tabs
3. Replace the (2) two fuses located in the fuse box. Make sure the fuses are aligned
properly.
Figure 19: Guide Tab Location
4. Align the fuse cartridge so the guide tab is towards the power entry side.
Return the fuse box to the power entry module and push in firmly.
5.
Confirm operational status by plugging in the cryoICE BOX and turning power on.
6.
Ensure that the self-test is completed without errors.
19
A001053
Packaged, ACM Accessories- Domestic
A001054
Packaged, ACM Accessories- International
A001056
Packaged, Gas Line Hose Module Assembly- Domestic
A001055
Packaged, Tank Hose Assembly- International
Component “C”
Tip Washer
AtriCure
F021837

Tank Hose Assembly without canisters – Standard

New AtriCure cryoICE BOX Installation
Existing AtriCure cryoICE BOX Upgrade
Figure 20: N2O Cylinder Interface Figure 21: AtriCure cryoICE BOX Tank Hose
Assembly without canisters
Replacement Part
20
Item
Supplied By
Part Number
Filter Cartridge
AtriCure
F021720
Filter O-ring
AtriCure
F010924
Tip Washer
AtriCure
F021837
O-Ring Lubricant
AtriCure
C002502
D

Tank Hose Assembly with canisters – Alternate (Replacement of Desiccant Filter)

This section only applies to cryoICE Box Systems equipped with the Tank Hose Assembly which contains the canister set.
Tank Hose Assembly with Canisters Replacement Parts
Figure 22: Gas Line Components
Desiccant Filter Cartridge (A) Note: Replace desiccant filter cartridge every time the N2O tank is replaced. Filter Housing (B)
Tip Washer (C)
Filter O-Ring (D)
Note: Replace filter O-Ring with replacement of the desiccant filter cartridge.
Procedure
1. Prior to changing the Desiccant Filter Cartridge, assure that the cryoICE probe is disconnected from the patient and the cryoICE Box is turned off.
2. Unscrew the filter cartridge housing by rotating it counter-clockwise. Refer to Figure 23 below.
Figure 23: Filter Housing Removal
21
Item
Supplied
USA
International
ACM Footswitch
AtriCure
A000708
A000708
Tank Hose Assembly with canisters
AtriCure
A000837
A000838
Tank Hose Assembly without canisters
AtriCure
A001056
A001055
Heater Band Extension Springs (Qty. 6)
AtriCure
A000836
A000836
N2O Exhaust Hose (50ft./15.2m.)
AtriCure
C002051
C002051
Cylinder Heater Band (CMH15)
AtriCure
A000728
A000728
Watlow
120150509 or
120150509 or
Cylinder Heater Band (CMH22)
AtriCure
A000727
A000727
Watlow
120220507 or
120220507 or
AC Power Cord
AtriCure
C000262
C002090
3. Remove the desiccant filter cartridge by rotating it counter-clockwise using hand force only. Refer to Figure 24 below.
Figure 24: Desiccant Filter Cartridge Removal
4. Remove the old black O-ring from the top of the filter housing fixture.
5. Slide the new O-ring onto the filter housing fixture, making sure that it is fully seated in the groove at the top.
6. Apply a thin film of O-Ring Lubricant around the new O-Ring.
7. Replace the desiccant filter cartridge with the new cartridge.
8. Replace the filter housing by screwing on clockwise using hand force only.
9. Remove the old Tip Washer and replace it with the new washer.
Other Detachable and Accessory Devices
International
Domestic
By
Part Number
SK025877-DWG7
SK025877­DWG10
125 VAC, 10A. (10ft./3.3m.)
Part Number
SK025877-DWG7
SK025877-DWG10
250 VAC, 10A. (11.5ft./3.5m.)

Disposal

The cryoICE BOX does not contain hazardous substances. Follow local governing ordinances and recycling plans regarding disposal or recycling of device components. The used cryoICE probe is considered bio-hazardous. Follow facility procedures for disposal.
22
Problem
Possible Cause
Action
Front displays
• No power.
• Check power switch on back of
cryoICE BOX electrical failure.
• Call AtriCure Service.
Cylinder Heater
• Heater not plugged in.
• Check connection on back of
• N2O cylinder valve closed.
• Ensure N2O valve is open.
• Empty N2O cylinder.
• Replace N2O cylinder.
• Extremely cold N2O cylinder.
• Allow 15 minutes to warm up.
• Heater not attached to N2O
• Attach Heater Band to cylinder.
• Heater malfunctioning.
• Call AtriCure Service.
Temperature
cryoICE probe not plugged in.
• Ensure cryoICE probe
• Malfunctioning cryoICE probe.
• Replace cryoICE probe.
cryoICE BOX malfunctioning.
• Call AtriCure Service.
cryoICE BOX has
cryoICE probe not plugged in.
• Plug in cryoICE probe.
• N2O cylinder empty.
• Replace N2O Cylinder.
• N2O cylinder valve closed.
• Open cylinder valve.
• Inlet Gas Connection not secure.
• Ensure Inlet Gas Connection is

8. TROUBLESHOOTING

not lit.
Band Icon Illuminated.
cryoICE BOX.
• Check plug connection on back of cryoICE BOX.
• Check AC plug in wall socket.
• Ensure power is available at wall socket.
unit.
cylinder.
Not Displayed.
power but will not go into FREEZE mode.
thermocouple leads are firmly seated within their receptacles.
completely seated.
23
Problem
Possible Cause
Action
cryoICE probe
• Heater Band not properly
• Check heater installation and
• N2O cylinder low or out of gas.
• Replace N2O cylinder.
Temperature
cryoICE probe plugged in
• Ensure cryoICE probe black and
• Malfunctioning cryoICE probe.
• Replace cryoICE probe.
cryoICE BOX malfunctioning.
• Call AtriCure Service.
Bottom segment
• N2O cylinder empty.
• Replace with full cylinder.
• N2O cylinder cold.
• Make sure heater blanket is
• Indicator not reset when
• Press Reset when cylinder is
N2O Gas Gauge
• N2O cylinder pressure is below
• Make sure heater blanket is
• N2O cylinder empty.
• Replace with full cylinder.
not getting cold enough.
Display indicates incorrect values.
installed.
• Exhaust filter is clogged.
incorrectly.
• Electromagnetic interference
heater icon.
• Exhaust connector (orange) is frosting/ freezing ice (liquid condensate is not uncommon) call AtriCure Service.
red plugs are in correct receptacles.
• Relocate or Reorient cryoICE BOX
on N2O icon flasing.
flashing.
cylinder was replaced.
650psi.
installed and working. Allow time for the cylinder to warm up if it is cold.
replaced.
installed and working. Allow time for the cylinder to warm up if it is cold.
24
Problem
Possible Cause
Action
Amber Low
• N2O cylinder not turned on.
• Ensure the N2O cylinder is fully
Difficulty
• Trapped N2O within the system.
• Power-On the cryoICE BOX
Wrench Icon
• Heater band over temperature
• Unplug heater band if clicking
Pressure Indicator on N2O icon flashing.
connecting a cryoICE probe to the cryoICE BOX.
• Quick connector out of sequence, sleeve on blue connector is forward.
• Quick connector O-ring dried out and/or swelling.
turned on.
which clears trapped gas exerting pressure on the connector.
• Push the sleeve toward the
cryoICE BOX
until it locks back.
(usually clicks)
• Lubricate the connector inside with silicon-based O-ring. lubrication such as AtriCure Part No. C002502.
flashing and clicking heard inside cryoICE BOX, may also include display flashing.
• Heater band over temperature
due to empty N2O Cylinder.
due to loose fit on N2O cylinder.
stops and/ or display flashing stops, check if tank is warm to the touch – If so, tank is likely empty, replace tank with full tank. Power-off, then Power-on
cryoICE BOX to reset wrench
Icon.
• Heater band is to be tight and positioned at bottom of tank, cord at top edge. If problem is not corrected by above two actions, return cryoICE BOX and heater band to AtriCure.
25
Problem
Possible Cause
Action
cryoICE probe
• The cryoICE BOX and probe
• If probe does not reach desired
• The N2O quality is not adequate
• Medical grade nitrous oxide,
• N2O cylinder contains a siphon
• Verify the N2O cylinder does not
getting colder than -75°C and not defrosting.
system are flooded with liquid N2O.
defrost temperature, apply warm sterile saline to the tissue and probe area as necessary.
• Replace the Tank Hose Assembly which has canister set with Tank Hose Assembly without canister set.
A001056 – Domestic Tank Hose Assembly without canisters A001055 – International Tank Hose Assembly without canisters
Power-On cryoICE BOX within a
few minutes of cryoICE probe use to minimize N2O condensing in system.
to be used as a refrigerant.
tube or a dip tube.
26
3ppm water maximum, is preferred for use with AtriCure cryogenic Devices.
contain a siphon tube or dip tube. Cylinder valve body should be blank (no mark of: S, DT, or D.)

AtriCure cryoICE BOX Error Codes

If an error condition should occur, the Maintenance Needed Indicator or the System Fault Indicator will illuminate. The probe Temperature display on the front panel will temporarily display one of the following error codes during the power-up sequence. Contact AtriCure Service if one of these conditions occurs.
Error ID Error Likely Cause
001 No 24 VDC Fuse (F2)
002 Cylinder Over Temperature Heater blanket
003 probe Overpressure Pressure regulator
004 Unwanted probe Pressure Leaky inlet valve
005 No 230 VAC Fuse (F1)
008 Cylinder Over Pressure/Temperature Overheated Cylinder
PPP Power On Self-Test Error
Activation Button/Footswitch Pressed during power up
27
Harmonic distortion
See IEC 61000-3-2 b) (Class A)
a) Professional healthcare facility environment.
Guidance and manufacturer’s declaration – Enclosure Port Immunity
The AtriCure cryoICE BOX is intended for use in the electromagnetic environment specified below. The customer or the user of the AtriCure cryoICE BOX unit should assure that it is used in such an environment.
Immunity Test Levels
Professional healthcare facility
environment
± 8 kV contact ± 2kV, ± 4kV, ± 8kV, ± 15 kV air
3 V/m f)
80% AM at 1kHz c)
Refer to Table 9 in IEC 60601-1-
equipment
Rated power frequency magnetic fields
30 A/m g) 50 Hz or 60 Hz
a) The interface between the PATIENT physiological signal simulation, if used, and the AtriCure cryoICE BOX shall
for the minimum expected distance.

9. ELECTROMAGNETIC COMPATABILITY TABLES

Electromagnetic Emissions

Guidance and manufacturer’s declaration – Electromagnetic Emissions
The AtriCure cryoICE BOX is intended for use in the electromagnetic environment specified below. The customer or the user of the AtriCure cryoICE BOX
Phenomenon Professional healthcare facility environment a)
Conducted and radiated RF EMISSIONS CISPR 11 (Group 1, Class A)
Voltage fluctuations and flicker IEC 61000-3-3 b)
b) This test is not applicable in this environment unless the AtriCure cryoICE BOX used there will be connected to the PUBLIC MAINS NETWORK and the power input Is otherwise within the scope of the Basic EMC standard.

Electromagnetic Immunity – Enclosure Port

unit should assure that it is used in such an environment.
Phenomenon Basic EMC standard or test method
ELECTROSTATIC DISCHARGE IEC 61000-4-2
Radiated RF EM fields a) IEC 61000-4-3
Proximity fields from RF wireless communications equipment
d) e)
be located within 0.1 m of the vertical plane or the uniform field area in one orientation of the AtriCure cryoICE BOX. b) ME EQUIPMENT and ME SYSTEMS that intentionally receive RF electromagnetic energy for its operation shall be tested at the frequency of reception. Testing may be performed at other modulation frequencies identified by the RISK MANAGEMENT PROCESS. This test assesses the BASIC SAFETY and ESSENTIAL PERFORMANCE of an intentional receiver when an ambient signal is in the passband. It is understood that the receiver might not achieve normal reception during the test. c) Testing may be performed at other modulation frequencies identified by the RISK MANAGEMENT PROCESS. d) Applies only to ME EQUIPMENT and ME SYSTEMS with magnetically sensitive components or circuitry. e) During the test, the AtriCure cryoICE BOX may be powered at any NOMINAL input voltage, but with the same frequency as the test signal. f) Before modulation is applied. g) This test level assumes a minimum distance between the AtriCure cryoICE BOX and sources of power frequency magnetic field of at least 15 cm. If the RISK ANALYSIS shows that the AtriCure cryoICE BOX will be used closer than 15 cm to sources of power frequency magnetic field, the IMMUNITY TEST LEVEL shall be adjusted as appropriate
IEC 61000-4-3
IEC 61000-4-8
80 MHz – 2.7 GHz b)
2:2014 – Test specification for Enclosure Port Immunity to RF wireless communication
28
Guidance and manufacturer’s declaration – Input A.C Power Port Immunity
The AtriCure cryoICE BOX is intended for use in the electromagnetic environment specified below. The customer or the user of the AtriCure cryoICE BOX unit should assure that it is used in such an environment.
Immunity Test Levels
Professional healthcare facility
environment
Electrical fast transients / bursts
a)
± 2 kV 100 kHz repetition frequency
Surges
a) b) j) o)
Line-to-line
± 0.5 kV, ± 1 kV, ± 2 kV
3 V/m m)
80% AM at 1kHz e)
0% UT; 0.5 cycle g)
270° and 315° q)
0% UT: 1 cycle
Single phase: at 0°
Voltage interruptions
f) i) o) r)
IEC 61000-4-11
0% UT: 250/300 cycle h)
a) The test may be performed at any one power input voltage within the AtriCure cryoICE BOX’s RATED voltage
q) At some phase angles, applying this test to ME EQUIPMENT with transformer mains power Input might cause

Electromagnetic Immunity – Input A.C. Power Port

Phenomenon Basic EMC standard or test method
l) o)
IEC 61000-4-4
IEC 61000-4-5 ± 0.5 kV, ± 1 kV
a) b) j) k) o)
Surges Line-to-ground
Conducted disturbances Induced by RF fields
c) d) o)
IEC 61000-4-5
IEC 61000-4-6
0.15 MHz - 80 MHz 6 V/m
0.15 MHz and 80 MHz
At 0°, 45°, 90°, 135°, 180°, 225°,
Voltage dips
f) p) r)
IEC 61000-4-11
And 70% U
range. If the AtriCure cryoICE BOX is tested at one power input voltage, It Is not necessary to re-test al additional voltages. b) All AtriCure cryoICE BOX cables are attached during the test. c) Calibration for current injection clamps shall be performed in a 150 Ω system. d) If the frequency stepping skips over an ISM or amateur band, as applicable, an additional test frequency shall be used In the ISM or amateur radio band. This applies to each ISM and amateur radio band within the specified frequency range. e) Testing may be performed at other modulation frequencies identified by the RISK MANAGEMENT PROCESS. f) ME EQUIPMENT and ME SYSTEMS with a D.C. power input intended for use with A.C.-to-D.C. converters shall be tested using a converter that meets the specifications of the MANUFACTURER of the ME EQUIPMENT or ME SYSTEM. The IMMUNITY TEST LEVELS are applied to the A.C. power input of the converter. g) Applicable only to ME EQUIPMENT and ME SYSTEMS connected to single-phase A.C. mains. h) E.g. 10/12 means 10 periods at 50 Hz or 12 periods at 60 Hz. i) ME EQUIPMENT and ME SYSTEMS with RATED Input current greater than 16 A/phase shall be interrupted once for 250/300 cycles at any angle and at all phases at the same time (if applicable). ME EQUIPMENT and ME SYSTEMS with battery backup shall resume line power operation after the test. For ME EQUIPMENT and ME SYSTEMS with RATED input current not exceeding 16 A, all phases shall be interrupted simultaneously. j) ME EQUIPMENT and ME SYSTEMS that do not have a surge protection device in the primary power circuit may be tested only at ± 2 kV line{s) to earth and ± 1 kV line(s) to line(s). k) Not applicable to CLASS 11 ME EQUIPMENT and ME SYSTEMS. I) Direct coupling shall be used. m) R.M.S., before modulation Is applied. n) The ISM (Industrial, scientific and medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to
6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz The amateur radio bands between 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz, 3.5 MHz to 4.0 MHz, 5.3 MHz to 5.4 MHz, 7 MHz to 7.3 MHz, 10.1 MHz to 10.15 MHz, 14 MHz to 14.2 MHz, 18.07 MHz to 18.17 MHz,
21.0 MHz lo 21.4 MHz, 24.89 MHz to 24.99 MHz, 28.0 MHz to 29.7 MHz and 50.0 MHz to 54.0 MHz. o) Applicable to ME EQUIPMENT and ME SYSTEMS with RATED input current less than or equal to 16 A/phase and ME EQUIPMENT and ME SYSTEMS with RATED Input current greater than 16 A/phase. p) Applicable to ME EQUIPMENT and ME SYSTEMS with RATED Input current less than or equal to 16 A/ phase.
m)
in ISM bands between
: 25/30 cycles h)
T
n)
29
an overcurrent protection device to open. This can occur due to magnetic flux saturation of the transformer
RATED input voltage within the range.
Guidance and manufacturer’s declaration – Patient Coupling Port Immunity
The AtriCure cryoICE BOX is intended for use in the electromagnetic environment specified below. The customer or the user of the AtriCure cryoICE BOX unit should assure that it is used in such an environment.
Immunity Test Levels
Professional healthcare facility
environment
± 8 kV contact
± 2kV, ± 4kV, ± 8kV, ± 15 kV air
3 V b)
80% AM at 1 kHz
a) The following apply:
PERFORMANCE.
core after the voltage dip. If this occurs, the AtriCure cryoICE BOX shall provide BASIC SAFETY during and after the test. r) For ME EQUIPMENT and ME SYSTEMS that have multiple voltage settings or auto ranging voltage capability, the test shell be performed at the minimum and maximum RATED input voltage. ME EQUIPMENT and ME SYSTEMS with a RATED input voltage range of less than 25 % of the highest RATED input voltage shall be tested at one

Electromagnetic Immunity – Input D.C. Power Port – Not Applicable

Electromagnetic Immunity – Patient Coupling Port

Phenomenon Basic EMC standard or test method
ELECTROSTATIC DISCHARGE c) IEC 61000-4-2
Conducted disturbances induced by RF fields
- All PATIENT-COUPLED cables shalt be tested, either Individually or bundled
- PATIENT-COUPLED cables shall be tested using a current clamp unless a current clamp is not suitable. In cases were a current clamp is not suitable, an EM clamp shall be used.
- No intentional decoupling device shalt be used between the injection point and the PATIENT COUPLING POINT in any case.
- Testing may be performed at other modulation frequencies identified by the RISK MANAGEMENT PROCESS.
- Tubes that are intentionally filled with conductive liquids end intended to be connected to a PATIENT shalt be considered to be PATIENT-COUPLED cables.
- If the frequency stepping skips over an ISM or amateur radio band, as applicable, an additional test frequency shall be used In the ISM or amateur radio band. This applies to each ISM and amateur radio band within the specified frequency range.
- The ISM (Industrial, scientific and medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to
6.795 MHz: 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz. The amateur radio bands between 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz, 3.5 MHz to 4.0 MHz,
5.3 MHz to 5.4 MHz, 7 MHz to 7.3 MHz, 10.1 MHz to 10.15 MHz, 14 MHz to 14.2 MHz, 18.07 MHz to
18.17 MHz, 21.0 MHz lo 21.4 MHz, 24.89 MHz to 24.99 MHz, 28.0 MHz to 29.7 MHz and 50.0 MHz to
54.0 MHz. b) R.M.S., before modulation Is applied c) Discharges shall be applied with no connection to an artificial hand and no connection to PATIENT simulation. PATIENT simulation may be connected after the test as needed in order to verify BASIC SAFETY and ESSENTIAL
a)
IEC 61000-4-6
0.15 MHz - 80 MHz
b)
6V
in ISM bands between
0.15 MHz and 80 MHz
30

Warranties

Limitation on Liability
This warranty and the rights and obligations hereunder shall be construed under and governed by the laws of the State of Ohio, U.S.A.
AtriCure, Inc. warrants this product to be free from defects in material and workmanship under normal use and preventive maintenance for the respective warranty period shown below. AtriCure’s obligation under this warranty is limited to the repair or replacement, at its option, of any product, or part thereof, which has been returned to AtriCure, Inc. or its Distributor within the applicable time period shown below and which examination disclosed, to AtriCure’s satisfaction, to be defective. This warranty does not apply to any product, or part thereof, that has been: (1) adversely affected due to use with devices manufactured or distributed by parties not authorized by AtriCure, Inc. (2) repaired or altered outside AtriCure’s factory in a way so as to, in AtriCure’s judgment, affect its stability or reliability, (3) subjected to improper use, negligence or accident, or (4) used other than in accordance with the design and use parameters, instructions and guidelines for the product or with functional, operational or environmental standards for similar products generally accepted in the industry. AtriCure has no control over the operation,
inspection, maintenance or use of its products after sale, lease or transfer, and has no control of the selection of Customer’s patients.
AtriCure’s products are warranted for the following periods after shipment to the original purchaser:
AtriCure cryoICE BOX Unit ............................................................................................. One (1) Year
AtriCure Cylinder Heater Band ..................................................................................... One (1) Year
AtriCure Gas Line Hose Assembly……………………………………………… ................................ One (1) Year
Grounded Electrical Cord .............................................................................................. One (1) Year
AtriCure Cryo Footswitch .............................................................................................. One (1) Year
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING THE WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, AND OF ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF ATRICURE, INC. AND IS A PURCHASER’S EXCLUSIVE REMEDY. IN NO EVENT SHALL ATRICURE, INC. BE LIABLE FOR SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES INCLUDING, WITHOUT LIMITATION, DAMAGES RESULTING FROM LOSS OF USE, PROFITS, BUSINESS OR GOODWILL.
AtriCure, Inc. neither assumes nor authorizes any other person to assume for it any other liability in connection with the sale or use of any of AtriCure Inc. products. There are no warranties that extend beyond the terms presented unless an extended warranty is purchased before the original warranty expires. No agent, employee or
representative of AtriCure has any authority to change any of the foregoing or assume or bind AtriCure to any additional liability or responsibility. AtriCure, Inc. reserves the right to make changes to products built and/or sold
by them at any time without incurring any obligation to make the same or similar changes on products previously built and/or sold by them.
31
Loading...