AtriCure ACM1, ACM2 User Manual

Page 1
AtriCure Cryo Module
(ACM)
Model ACM1 – 115 VAC Model ACM2 – 230 VAC
AtriCure Incorporated 7555 Innovation Way Mason, Ohio 45040 USA
Customer Service: 1-866-349-2342 (toll free) 1-513-755-4100 (phone)
P000669 Rev. J
Page 2
European Representative: Herbert Köntges Köntges SPRL Avenue Hellevelt 35 B-1180 Brussels Belgium Tel: +32 2 375 51 63 FAX: +32 2 375 89 06 email: herbert.kontges@skynet.be
Page 3
Table of Contents
1. GETTING STARTED .................................................................................................................... 4
1.1. System Description ......................................................................................................... 5
1.2. Unpacking ......................................................................................................................... 6
1.3. Warnings and Precautions ............................................................................................. 6
1.4. EMC Guidance and Manufacturer’s Declaration ........................................................ 7
1.5. Responsibility of the Manufacturer ............................................................................. 11
2. THE ATRICURE CRYO MODULE (ACM) ............................................................................. 12
2.1. Device Description ......................................................................................................... 12
2.2. ACM Front Panel – Il lustration and Nomenclatur e ................................................... 13
2.3. ACM Rear Panel – Il lustration and Nomenclatur e ................................................... 20
3. INSTALLING THE ACM ............................................................................................................ 23
3.1. Transporting of ACM ..................................................................................................... 23
3.2. Preparing the ACM For Use ......................................................................................... 23
3.3. Installing Power Cord .................................................................................................... 23
3.4. Installing the Footswitch ............................................................................................... 23
3.5. Installing Cylinder Heater Band ................................................................................... 24
3.6. Installing N2O Gas Line Hose ...................................................................................... 25
3.7. Installing N2O Exhaust Hose........................................................................................ 27
3.8. Connecting and Disconnecting the Handpiece ......................................................... 28
4. INSTRUCTIONS FOR USE ....................................................................................................... 30
4.1. Powering Up the ACM .................................................................................................. 30
4.2. Operating Modes ........................................................................................................... 31
4.3. Delivering Cryo Energy ................................................................................................. 33
4.4. Parameter Entry Mode / Adjusting the System Default Parameter Values .......... 37
5. TROUBLESHOOTING ............................................................................................................... 38
5.1. No ACM Power / Display Function ............................................................................. 38
5.2. ACM Fault Codes .......................................................................................................... 38
5.3. ACM Error Codes .......................................................................................................... 39
5.4. Handpiece Error Codes ................................................................................................ 41
6. SYMBOLS USED ........................................................................................................................ 42
7. TECHNICAL SPECIFICATIONS .............................................................................................. 45
7.1. Mechanical Specific ations ............................................................................................ 45
7.2. Environmental Specifications ....................................................................................... 45
7.3. Electrical Specifications ................................................................................................ 45
7.4. Mains Fuses ................................................................................................................... 45
7.5. Footswitch Specifications ............................................................................................. 45
7.6. Equipment Type / Classification .................................................................................. 45
8. PREVENTIVE MAINTENANCE AND CLEANING OF ACM .............................................. 46
8.1. Preventive Maintenance ............................................................................................... 46
8.2. Cleaning and Disinfecting ............................................................................................. 47
2
Page 4
9. SERVICING OF ACM UNIT ...................................................................................................... 48
9.1. Replacement of AC Mains Fuses ............................................................................... 48
9.2. Inspection and Replaceme nt of In-Line N2O Filter ................................................... 49
9.3. Replacement of Gas Line Desiccant Filter and Tip Washer ................................... 51
9.4. Other Replacement Components ............................................................................... 53
10. ACCESSORIES ........................................................................................................................... 54
10.1. CMH15, Cryo Module Cylinder Heater Band ............................................................ 54
10.2. CMH22, Cryo Module Cylinder Heater Band ............................................................ 54
10.3. CMF1, Cryo Module Footswitch .................................................................................. 54
11. DISPOSAL .................................................................................................................................... 54
3
Page 5
1. Getting Started
This manual a nd the equip ment it describes are for use onl y by qualified medical professionals trained in the particular technique and surgical procedure to be performed.
Caution: F eder al (USA) law res tr icts thi s dev ic e to s ale by or o n the or der of a
physician.
Please rea d al l informati on carefully. F ailure to properly fol low the ins tructions may lead to serious s ur g i cal consequ ences.
Important: This manual is designed to provide instructions for use of the AtriCure Cryo Module (ACM) with the AtriCure Cryo Handpiece (CRYO1 and cryoICE) and AtriCure Accessory Devices (CMF1, CMH15, CMH22). This manual is not a reference to surgical technique.
The AtriCure Cryo Module (ACM) unit is an electro-mechanical and pneumatic cryogenic surgical system that d elivers a cryogenic ener gy source, namely Nit rous Oxide, t o c reate lines of a blation thro ugh cardiac tissue. The ACM is part of a sys t e m which includes the nitrous oxi de gas , N2O gas line h ose, N2O exhaust hos e, cy l inder hea ter ban d, a footswitch, and single-use cryo ha ndpiece with integrated electronics cabl e / g as hos es. The syste m pr ov i d es controlled lesion forming temperature that is below -40°C, with typical operating ranges between -50°C to -70°C. The system is also equipped with the ability t o tu ne or set the terminal handpiece temperature.
The ACM is designed to operate only with AtriCure designed and developed handpieces. Along with the Ablation Activation Switch on the front panel of the ACM, a footswitch ca n also be used to activate the ablation cycle. Refer to the handpiece Instruction for Use for a complete description of the indications and uses of these devices.
For the user’s conv e nienc e, t he AtriC ur e Cryo Module will be referred to in this User’s Manual as the “ ACM”. The AtriCure Cryo Handpiece will b e r e f erred in this User’s Manual as the “Hand pi ece” .
This User’s Manual provides a description of the ACM, its controls, displays, indicators, and a sequence for its operation with the Handpiece. This User’s Manual also supplies other information of importance to the user. This manual is intended as a User’s Manual only. Do not operate the ACM before thoroughly reading this manual.
4
Page 6
A E D C B
1.1. System Description
As shown i n F i g ure 1, the s ystem is c o mpri sed of the following:
A: AtriCure Cr yo Module C ylinder Heat er B and (CMH15 or C M H 2 2)
B: AtriCure Cryo Footswitch (CMF-1)
C: ACM N2O Exhaust Hos e
D: ACM N2O Gas Line Hose Module
E: AtriCure Cryo Module (ACM)
ACM Power Cor d ( n o t shown)
AtriCure Cr yo Handpiec e with integ ral c a ble (not shown)
Accessory d evices are de s cr i b e d in Section 10.
Figure 1 – ACM, Footswitch, N2O Gas Line Hos e Module, N2O
5
Exhaust Hose, and Cylinder Heater Ba nd
Page 7
1.2. Unpacking
1.2.1. ACM Unit
Lift the ACM unit from the box and remove the protective wrapping and side caps.
Note: The ACM unit is packed upside down for transit purpose. Note: Take precaution s when lifting the unit as it weig hs approx 45 lbs (20 kg) .
1.2.2. Accessories
ACM accessor ies are ship ped sep ara tely fro m th e ACM uni t. Th e acc ess ory box contains th e foll ow ing items:
ACM Power Cord
N2O Gas Line Hose
N2O Exhaust Hose
ACM Cylinder Heater Band
ACM Footswitch
It is recomme nde d that th e ori gin al s hippi ng box and pr ote ctiv e w ra pping be sav e d for future stori ng a nd/ or tr a nspor ti ng o f th e devi ce.
1.3. Warnings and Precautions
The safe and effective use of the cryo d ev ice an d equi p men t is highly dependent
upon factors und er th e co ntr ol o f the op era tor . Ther e is no su bsti tut e f or a pr op erly trained operating room staff. It is important that t h e op erating instructions supplied with the ACM unit be read, understood, and foll ow ed b efor e use.
1.3.1. WARNINGS
Do not operat e the ACM unit before thoroug hl y rea ding thi s man ual .
Do not use cryo surgical equipment unless pro perly tr ai ned i n the spe ci fic
procedure b eing und ert aken . Thi s man ual and the equip men t it desc ri bes are for use only by quali fi ed m edi cal pro fessi o nal s trai ne d in t he par ti cul ar te c hniq ue and surgical proc e dur e to be per for m ed.
Fire Hazard: Do not use extension cords.
Trip Hazard: Stan dar d care sho ul d be use d t o red uce th e ri sk o f tri ppi ng on the
Footswitc h cable, as well as t he N2O exhaust hose.
No modificati on of t his eq ui pme nt is al low ed. The voltage selec tor is fact ory set and sh oul d no t be chang ed by the us er. The
voltage setting and the fuse r ati ng m ust be ap propr i ate as id enti fi ed to prev ent ACM malfunction and potential instrument damage.
Electric S ho c k Hazard: Connect the ACM Power C or d to a properl y grounded receptacle. Do not use power plug adapters.
Electric S ho c k Hazard: Do not conn ec t wet accessories to the generator.
6
Page 8
Electric S ho c k Hazard: Ensure that th e Handpiece is correc tl y connected t o the ACM and that no thermocouple w ires are expo se d fro m the cabl e, conne ct or, or the Handpiece.
1.3.2. PRECAUTIONS
Use only with the Atri Cure Handpi ec es intended for use with t h e ACM.
Do not activat e th e ACM unit until the Handpiece is properly positioned at the
ablation site.
The system status indicators and di spl ays are importa nt safety features. Do n ot
obstruct either the ablation or the s ystem stat u s indicators. Do not remove the ACM cover , ex cep t for th e in-line gas filter ac ces s por t, as
there is a p otential for electrical shock. Re fer to authoriz e d per sonnel for s er vice.
Use only the Footswitch provided with the ACM.
The Power Cord of the ACM must be connected to a properly grounded
receptacle . Exte nsi o n cord s an d/or ada pter pl ugs m ust not be use d.
Do not contac t Atri C ure Ha ndpi ec es w ith a RF dev ic e.
Compressed Air Hazard: Do not operate N2O tanks wit h a pr essure grea ter than
1000 PSIG (690 0 kP a).
1.4. EMC Guidance and M an ufacturer’s Declarat ion
1.4.1. Electromag n et ic Requirements
1.4.1.1. The AtriCure Cryo Module (ACM) has been tested and found to comply
with the limits for m e di c al devices in IEC 60601-1-2:2007. These limits are designed to pr ovi de r eas ona ble prot ecti on against harmful interference in a typical medic al ins tall at i on.
1.4.1.2. The AC M can radi at e radio frequency ener g y and, if not installed and used
in accorda nce with the instructions, may caus e harmful in terference t o other devi ces in the vicinity.
1.4.1.3. Port abl e an d mobile RF communications equipment can also affect ACM
performanc e an d car e sh oul d be tak e n to mini miz e suc h inter fer e nce . However, t h er e i s no guarantee that int erference will not occur in a particular installation.
1.4.1.4. If th e ACM does c a use harmful i nter fer e nc e to ot her dev ices , w hic h can be
determined by turning t he A CM off and on, the us er is encouraged to try to correct the int erf ere nce by on e or mor e of t he follow i ng m easur es :
Reorient or rel oc at e the rec eiv ing devi ce .
Increase the separation between the ACM and the other devices.
Connect the ACM into an outlet on a ci r cuit different from t h at to which
the other devi ce( s) are conn ect ed.
Contact t he A tr i C ure servic e representative for help.
7
Page 9
Emissions test
Compliance
Electromagnetic environment – guidance
The ACM unit uses RF energy only for its internal
electronic equipment.
RF emissions
Class A
The ACM unit is suitable for use in all establishments
Voltage fluctuations/
1.4.2. Electromag net ic Em is sion s
Guidance and manufacturer’s declaration – electromagnetic emissions
The AtriCure Cryo Module (ACM) is intended for use in the electromagnetic environment specified below. The customer or the user of the ACM unit should assure that it is used in such an environment.
RF emissions CISPR 11 Group 1
CISPR 11 Harmonic emissi on s IEC
61000-3-2
flicker emissions IEC 61000-3-3
Class A
Complies
function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby
other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domesti c purposes.
8
Page 10
± 6 kV contact
± 6 kV contact
Floors should be wood, concrete,
least 30 %.
Electrical fast
± 2 kV for power
± 2 kV for power
lines
Mains power quality should be that ± 1 kV differential
mode
Mains power quality should be that
Voltage dips, short
<5 % UT
for 5 s
<5 % UT
for 5 s
Mains power quality should be that
Power frequency
3 A/m
3 A/m
Power frequency magnetic fields
environment.
1.4.3. Electromag net ic I mmu nity
Guidance and manufacturer’s declaration – electromagnetic immunity
The AtriCure Cryo Module (ACM) is intended for use in the electromagnetic environment specified below. The customer or the user of the ACM unit should assure that it is used in such an environment.
IMMUNITY test IEC 60601 test level Compliance level
Electromagnetic environment –
guidance
Electrostatic discharge (ESD) IEC 61000-4-2
transient/burst IEC 61000-4-4
Surge IEC 61000-4-5
interruptions and voltage variations on power supply input lines IEC 61000-4-11
± 8 kV air
supply lines ± 1 kV for
input/output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
(>95 % dip in UT) for 0,5 cycle
40 % U
T
(60 % dip in UT) for 5 cycles
70 % U
T
(30 % dip in UT) for 25 cycles
<5 % U
T
(>95 % dip in UT)
± 8 kV air
supply Lines
± 1 kV for input/output
mode ± 2 kV common
(>95 % dip in UT) for 0,5 cycle
40 % U
T
(60 % dip in UT) for 5 cycles
70 % U
T
(30 % dip in UT) for 25 cycles
<5 % U
T
(>95 % dip in UT)
or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at
of a typical commercial or hospital environment.
of a typical commercial or hospital environment.
of a typical commercial or hospital environment. If the user of the ACM unit requires continued operation during power mains interruptions, it is recommended that the ACM unit be powered from an uninterruptible power supply or a battery.
(50/60 Hz) magnetic field IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
9
should be at levels characteristic of a typical location in a typical commercial or hospital
Page 11
Portable and mobile RF communications
d = 1.2 √P 80 MHz to 800 MHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
absorption and reflection from structures, objects, and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
1.4.4. EMC Guidan ce and Manuf act urer’s Declaration
Guidance and manufacturer’s declaration – electromagnetic immunity
The AtriCure Cryo Module (ACM) is intended for use in the electromagnetic environment specified below. The customer or the user of the ACM should assure that it is used in such an environment.
IMMUNITY test
Conducted RF IEC 61000-4-6
IEC 60601 TEST LEVEL
3 Vrms 150 kHz to 80 MHz
Compliance level
3 Vrms
Electromagnetic environment – guidance
equipment should be used no closer to any part of the ACM, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation di st ance d = 1.2 √P
Radiated RF IEC 61000-4-3
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the ACM is used exceeds the applicable RF compliance level above, the ACM should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the ACM.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
3 V/m 80 MHz to 2.5 GHz
3 V/m
d = 2.3 √P 800 MHz to 2.5 GHz where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
b
a
10
Page 12
The AtriCure Cryo Module (ACM) is intended for use in an electromagnetic environment in which radiated
the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended separation distance
absorption and reflection from structures, objects, and people.
1.4.5. Recommended Separation Distance
Recommended separation distances between portable and mobile RF communications equipment and the AtriCure Cryo Module
RF disturbances are controlled. The customer or the user of the ACM can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and t he ACM as recommended below, according to the maximum output power of
Rated maximum
output power of
transmitter W
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d = 1.2 P
80 MHz to 800 MHz
d = 1.2 P
800 MHz to 2.5 GHz
d = 2.3 P
1.5. Responsibility of the Manufacturer
AtriCure is responsibl e for safety , reliability, and per formance of the equip me n t only if:
Installati on pr oce dur es in this m anual are followed.
Persons aut horiz e d by A triC ur e car ry out mo di fic ati ons or re pai rs.
The electri cal install ation of the relevant r o o m complies w i th local codes a n d
regulatory requi r eme nts such as IEC and BS I.
The equipme nt is us ed in acc ord anc e w ith the Atr iC ure Us er’ s Manual.
11
Page 13
2. The AtriCure Cryo Module (ACM)
This section prov id es a de tail e d desc rip ti on of t he ACM including its function and operating f eatur es .
2.1. Device Description
The AtriCure Cryo Module (ACM) unit is an electro-mechanical and pneu mati c
cryogenic surgical system that delivers a cr yogenic sour ce, Nitrous O xide, to create lines of ablation through cardiac t issue.
The ACM is part of a system which includes the nitrous oxide gas and its cylinder,
N2O gas line hose, N2O exhaust hos e, cy l inder heat er b an d, a footswitch, system power cord, and single-use cryo handpiece with integrated electronics cable / gas hoses.
The system prov id es a co ntr oll ed l esi on for mi n g temperature that is below -40°C,
with typical operating ra ng es between -50°C to -70°C for the tr eat ment o f cardi ac arrhythmias , cr eat ing a n infl am mator y r es pons e (cry o nec rosi s) that bloc ks the cardiac el ec t r ical conduc ti o n p athway.
The system is also equipped with the ability to tune or set the terminal handpiece
temperature.
Along with the Ablation Activation Switch on the front panel of the ACM, the ACM
Footswitc h can al so be us ed as the i npu t dev ic e to ac tiv at e and t erminate the cryo ablation cyc le.
The presence of the ACM Footsw itch i s not a r equi re me nt for the A CM unit to
operate and is mad e avai la bl e as a se con dary mea ns of int er faci ng with th e ACM unit, and to provide control of the ablation cycle to the surgeon.
The ACM is desig ne d to op er at e only with AtriCure designed and developed
handpieces.
Refer to the Handpiece Instruc ti on for Us e for co mple te des cri pt io n of t he
indications an d us es of t hes e dev ice s.
12
Page 14
1. Ablation Activation Switch
7. Ablation Timer Increase Button
1
5
7
12
11
6
3
8
2
4
10
9
1
2
3
2.2. ACM Front Panel – Illustration and Nomencl ature
An illustration of the ACM front pa n el i s shown in Figure 2, below.
Figure 2 – ACM Fro nt Panel
2. Ablation Status Indicator 8. Handpiece Probe Te mper ature
3. N
O Gas Gauge Display 9. Handpiece Thermocouple Port
2
4. System Sta tu s Indicator 10. Handpiece Electronic Interface Port
5. Ablation Timer Decrease Button 11. Handpiece Gas Connection Port
6. Ablation Timer Display 12. Handpiece Probe Temperature Cont rol
2.2.1. Ablation Status Indicators
There are three indicators on the front panel of the ACM: READY, FREEZE, and DEFROST. These indicators are descr ibed below.
13
Figure 3 – AC M Ablatio n Status Indic a tor s
Page 15
Indicator Symbol Description
READY – Green LED
This status indicates the ACM is in
1
stand-by mode and is ready to start the cryo ablation cycle.
2
3
FREEZE – Blue LED This status indicates the ACM is in cryo
freeze state and the N2O gas is allowed to flow through the Handpiece causing drop in probe temperature.
DEFROST – Amber LED Constant:
This status indicates th e ACM is in defrost state, N2O gas flow is ceased, Handpiece is bac k pr es suri zed , a nd the Handpiece probe is warming back up to the ambient temperature.
Blinking: Indicates t he ACM is now ready to transitio n to the nex t st ate. Upon activation o f eit her th e Abl ati on Act iv ati on Switch or the Foot sw itc h, the ACM unit will transition into the next state depending on the ACM model as described bel ow .
International – ACM
Transitions to Ready state only
US – ACM
Press and rel eas e – Freeze stat e
Press and h old – Ready state
14
Page 16
1
2
3
4
6
5
2.2.2. System Status Indicators
There are six system status indicators on the front panel of the ACM: SYSTEM POWER, SYSTEM FAULT, MAINTENANCE NEEDED, LOW CYLINDER PRESSURE, CYLINDER HEATER BAND ACTIVE, and FOOTSWITCH ACTIVE. These indicators are described below.
Figure 4 – ACM Status Indicators
15
Page 17
Indicator Symbol Description
SYSTEM POWER – Green LED
1
2
3
This status indicates system logic power for the ACM is available.
SYSTEM FAULT – Red LED This status indicates ACM has
encountered a system fault condition and has halted all operation. The ACM unit will not operate until the fault condi ti on has been addressed.
MAINTENANCE NEEDED – Amber LED This status indicates th e ACM syst em is s till
in normal, fu ncti o nal, and saf e oper ati ng mode. How ever, maintenance is being requested to address such items as replacing in-line N2O gas filter to opti mize system perfor m anc e.
16
Page 18
Indicator Symbol Description
4
LOW CYLINDER PRESSURE – Amber LED
This status indicates the N2O cylinder gas pressure into the ACM unit is below (initial system power up) or is dropping below (during usage) the optimal operating level.
Two LED indication modes are available. Initial Power Up: Constant:
Indicates the cylinder pressure is still below the optimal operating level.
Blinking:
Indicates cylinder valve is closed, or if the valve is open, the cylinder pressure is well below the operating level or is empty of N2O gas.
In both of these cases, the cylinder heater band will be active to raise the cylinder pressure to its recommended operating level as long as the cylinder valve is in the open position.
The cylinder heater band will not be active if the cylinder valve is in the CLOSED position.
The indicator LED will extinguish when the cylinder pressure has reached its optimal operating level.
During Usage: Constant:
Indicates the cylinder pressure has dropped below the optimal oper a ti ng level, and the cylinder heater band will be activated. The indicator LED will extinguish when the cylinder pressure has risen back to the optimal operating level.
Blinking:
Indicates the cylinder pressure is well below the recommended operating level, and the cylinder has less than 10 minutes of functional gas left.
17
Page 19
HANDPIECE THERMOCOUPLE
HANDPIECE ELECTRONIC
HANDPIECE PNEUMATIC Receptacle
Indicator Symbol Description
5
6
2.2.3. Front Panel Functions
Item Description
CYLINDER HEATER BAND ACTIVE –
Amber LED This status indicates ACM unit has
energized the N2O cylinder heater band and it is active. The indicator LED will extinguish when the cylinder heater band is turned off by the ACM unit (when optimal cylinder pressure is reached).
FOOTSWITCH ACTIVE – Blue LED This status indicates the ACM footswitch
pedal is depressed and is in activate state. The indicator LED will extinguish when the footswitch pedal is released.
Receptacle This 2-pin receptacle accepts the AtriCure
Handpiece thermocouple connectors. This connection is patient-isolated.
INTERFACE Receptacle This 14-pin receptac le is reserv e d for
future releas e o f Atri C ur e Hand piec e . This connection is patient-isolated.
This 2 port pneumatic receptacle provides the N2O interfac e with the AtriCure Handpiece. The ports are sized differ en tly to prevent c ross-coupling of the connections.
18
Page 20
HANDPIECE PROBE TEMPERATURE
Item Description
ADJUST Knob
This control knob prov i des t he op er ator with the a bil ity to adj us t the terminal temperature of the ha ndpiece pro b e. Turning the kn ob in th e coun ter-clockwise rotation low er s th e prob e te mper at ur e, while turn ing the knob in clockwise rotati on raises the probe temperature. In normal operation the knob wi ll be adjusted to provide max imu m neg at iv e temp er atur e.
2.2.4. Front Panel Displays
2.2.4.1. N2O Gas Gauge
This displ ay indicates, i n minutes, the approximate amount of useable N2O gas remaining in the cylinder. As the gas is being c onsumed, this value will adjust based on the actual gas flow tim e and the cylinder pressure. As a resul t, the displ ayed value may decrease quicker than anticipated.
This time value should no t be tak e n as a n abs olu te me as urement, but only as a r e f erence in gaug ing the amount of gas remaining in the cylinder.
The display value wil l start to blink when there is appr oximately 1 0 minutes or les s of a ntic i pate d us eabl e g as le ft in th e cy li nder.
NOTE: Replacement of the N2O cylinder at this po int i s high ly
recommended.
NOTE: Along with the blinking gas gauge , th e Lo w Cylinder
Pressure in dicator LED may also be blinking in dicative of empty or near empty N2O cylinder.
Typical usag e ti me for a new cyl in der is between 30 to 40 minutes depending on the ty pe and si z e of the cyl in der w hen used with the ACM cylinder he ater ban d. The usage time will be less when the cylinder heater band is not us ed .
The gas gauge value can be reset upon replacement of the N2O cylinder by pressing down on th e Ti m er Increase and Timer Decrease key simult aneously an d r eleasing.
Secondary function of the gas gauge is to display the N2O cylinder gas pressure dur ing the abl ati on cy cle . Thi s prov id es th e user wi th additional r e al -time info r m ation regarding the gas status w i t hi n the cylinder during system operation in psi x10.
19
Page 21
1. N2O Exhaust Port
5. Equipotential Terminal
1 5 32 4 6 8 7
2.2.4.2. Ablation Timer
This display indi cat es t he desir ed abl ati on dur ati on in uni ts o f sec ond s.
The default dur at ion time is 120 s econds per ablation run.
This timer v alue can be adjusted pr ior to, or d uring the ablation cycle,
by using either the Ti mer Increase or Timer Decrease Button.
Each key press of th e Ti mer Increase Button or Timer Decrease Button will modi fy the target ablation ti m e value by 1 0 s ec onds, either addi ng or subtracting from its present value.
At the termination o f the freeze c y c l e ( Defrost mod e) , the displ ay will show total time in freeze state. A full 120 second run will resul t in the display show ing 120 as an ex ampl e. An abor ted freez e cycle will show the number of seconds the system operated in the Freeze mode.
2.2.4.3. Handpiece Probe T e m perature
This displ ay indicates the current handpiece probe temperature in Celsius, and is cont inuously updated each second.
2.3. ACM Rear Panel – Illustration and Nomenclature
An illustration of the ACM rear pan el i s shown in Figure 5, bel ow .
Figure 5 – ACM Re ar P anel
2. System Pressure Relief Knob 6. Cylinder Heater Band Receptacle
3. N
O Inlet Port 7. AC Power Entry Module
2
4. Footswitch Receptacle 8. Input Voltage S el ec to r S witch
Note: Input Vol tag e Level Switch is factory set a nd i s covered by unit mod el label. The switch and t he label should not be tampered with or changed ex cep t b y aut hori z ed AtriCure personnel.
20
Page 22
2.3.1. Rear Panel Functions
Item Description
Power Entry Module –
This module co nt ains bot h th e ON/OFF switch and the w all receptacle mains fuses.
Fuse Box – The Fuse Box conta ins fus es selec t ed for the input voltage. S ee T echnical Specificati ons in S ecti o n 7 of this manu al .
Cylinder Heater Band Receptacle – This receptacle accepts the ACM Cylinder Heater Band connector. It provides power for the heating element embedded in the heater band, as well as transmitting back to the ACM unit the N2O cylinder wall temperature.
Footswitch Receptacle – This receptacle accepts the ACM Footswitch connector. The single momentary actuation p edal pr ovides for the activation and termination of the cryo ablation cycle.
N2O Inlet Receptacle – This receptacle accepts the N2O gas line hose with the quick connect / disconnect interface used to transfer gas from the N2O gas cylinder to the ACM unit to be utilized by the Handpiece.
21
Page 23
protective earth connection.
Item Description
N2O Exhaust Receptacle –
This ½ inch (1.25cm) receptacle accepts the 50 ft (15 m) flexible tubing to direc t the expelled N2O gas in a manner appropriate with local and hospital regulation.
System Pressure Relief Knob – This knob, when pulled, will depressurize and expel trapped N2O gas within the ACM unit.
Be sure the N2O cylinder valve is CLOSED prior to performing this task.
Equipotential Terminal –
This terminal provides a means of securely linking the earth ground of the AtriCure ACM unit to other grounded equipment.
This terminal is accessible during any position of normal use and mini mi zes risk of accidental disconnecti on. Th e conductor is detached without use of a tool.
This terminal is not to be used for a
22
Page 24
3. Installing the ACM
Inspect the ACM system for any signs of physical damage to the front and re ar panel, N2O gas couplers, connector receptacles, chassis or cover.
NOTE: If any physical damag e is f oun d , DO NOT USE THE UNIT. Contact
AtriCure for a replacement .
All return s must be app r oved by AtriCure.
3.1. Transporting of ACM
The use of handles is highly recommended in lifting, carrying, or moving of the ACM unit. The ACM wei g hs approximately 45 l bs ( 20 kg) so c are should be taken when transporting the ACM unit.
3.2. Preparing the ACM For Use
3.2.1. The ACM unit may be placed on a cart or on any sturdy t a bl e or platform. It is recommended that carts have cond uctive wheel s. Refer to hospital p r oce dures or local codes f or det ail ed in for mati on.
3.2.2. The ACM unit may be pl a c e d together or stacked w it h o ther AtriCure capital equipment s uch as the ASU , A SB, and t he ORLab. Howev er, car e m ust be tak en when stacki ng or pl ac ing the AC M uni t adj ac en t to ot her medi cal equipment and unit performance should be obs er ved to veri fy normal operation in th e configuration in whic h it will be used.
3.2.3. Provide at le ast f our t o si x inc hes of space around the sides of the ACM for convection cooling. U n der continuous use for extended peri ods of time, i t is normal for t he top pane l t o be warm to the touch.
3.3. Installing Power Cord
The ACM unit is ship ped with an approved hospital grad e power cord. Insert the pow er c ord i nto the pow er en try mod ule of the AC M u nit as shown in Fig ure 7, and then into a grounded hospi tal AC power receptacle.
NOTE: ACM power cord is shipped wi th the accessories and found in the acc ess or y
package.
NOTE: Do not use extension c ords or thr ee-prong to two-prong adapters . T he Pow er
Cord assem bl y should be p er iodically c he c k ed for damaged insulatio n or connectors.
3.4. Installing the F ootswitch
3.4.1. Inspect the Footswitch
Inspect t he Footswitc h for any signs o f p hysical damage to the cable and connector. If physical damage is found or the Footswitch does not perform within specification, contact AtriCure for replacement.
3.4.2. Connecting and Disconnecting the Fo otswitch
With the connector al i g n me nt arrow in the 12 o’cl ock position , push the Fo ots witch Connector into the Footswitch Receptacle on the rear panel of the ACM as shown
23
Page 25
in Figure 7. To disconnect, firmly grasp the connector body and pull out from the Footswitc h Receptacle .
NOTE: Typically, you will connect the Footswitch to the ACM prior to powering up the
unit. However, the Footswitch may be connected to the ACM when the unit is powered up and is in READY operating mode ( s ee Secti on 4 reg ar ding the READY mode).
3.4.3. Preparing the Footswitch for Us e
The Footsw i tc h should be pl aced on a fl at floor. I t i s recommended that the area near the Foot sw itch b e ke pt dry to red uc e the ri sk o f slip pag e. Appropriat e precautio n s s hould be tak en to ensure t hat the cabl e connecting the Footswitch to the ACM does not creat e a hazard in the operating room.
3.5. Installing Cylinder Heater Band
3.5.1. Inspect th e C y li nder Heat er Band
Inspect the Cylinder Heater Band for any s i g ns of physic al da m ag e to the ca ble and connector . If phy si c al da mage is foun d or th e Cylinder Heater Band does not perform w i thi n specificati o n, contact AtriCure for replacement.
3.5.2. Wrapping the Cylinder Heater B and
Place the Cyl ind er He at er Ba nd aro un d the bottom of the N2O cylinder, with the cord at the top of the Heater Band, and s e c ure all 6 tensioning s pr ing retainers as shown in Fi g ure 6. Sta r t with the top a nd bottom spr ing retain er s . Be sur e he at er band is flat agai ns t the cy li nder (n o w rink les , no pu cker s ).
NOTE: Cylinder Heater Band should be positioned approximately 2.0 inch (5.0 cm)
from the cylinder bottom for optimal performance.
NOTE: If the nitrous oxide cylinder is too large or too small for the supplied ACM
Heater Band to fit securely, contact AtriCure Customer Supp or t for replacement.
Figure 6 – Cyli nd er Heater Band Installation
24
Page 26
3.5.3. Connecting and Disconnecting the Cylinder Heater Band
With the c onnector alig nment arrow i n t he 12 o’clock position, pu s h t he Cylinder Heater Ba nd Connector into the Cylinder Heater Band R ec e p t acle on the r ear panel of the ACM as shown in Figure 7. T o disconnect , fi rmly gr asp the connector body, push dow n on the l ocki ng latc h, an d p ull out fr om th e Cy lin der Hea ter Band Receptacle.
NOTE: Cylinder Heater Band must be connected to the ACM prior to powering up the
unit. However, the ACM unit can operate without the Cylinder Heater Band in place but the system performance will not be optimal.
NOTE: It is highly recommended that the ACM unit and the cylinder heater band be
activated at least 15 minutes prior to fi rst use to all ow for t he cy lind er pr ess ur e to reach opti m al o perating level.
Figure 7 – Connecting the Footswitch, Cylinder Heater Band, and Power Cord to ACM
3.6. Installing N2O Gas Line Hose
3.6.1. Inspect th e N2O Gas Line Hose Module
Inspect the N2O Gas Line Hose for any signs of phys ic al da mag e to th e hose and pneumatic connectors on both ends, including pressure gauge, press ure rel i ef vessel, and the d esic ca nt filt er as sem bly . If physical damage is found do not use the N2O Gas Line Hose and contact AtriCure for replacement.
NOTE: Refer to the ACM Tank Hose Subassembly Cart Attachment Instruction
(P000862) for details on securing the N2O Gas Line Hose M o dul e to the AtriCure System Cart.
3.6.2. Connecting and Disconnecting the N2O Gas Line Hose with ACM unit
25
Page 27
Align the quick-connec t gas co nnec t or from the gas hose line with the ACM
N2O inlet port.
Insert and push in the conn ect or unt il you h ear it “cli ck ” in pl ac e and the
connection is fully seated and secured from unlatc hi ng.
To disconnec t th e N2O gas line connector from the ACM, slide back on the
N2O inlet port connector sleeve and remove. See Figure 8.
NOTE: The N2O cylinder valve should be in the CLOSED position during
connection and discon nec ti o n fro m the ACM uni t.
Figure 8 – Aligning the N2O gas line quick connect to the ACM N2O inlet port
3.6.3. Connecting and Disconnecting the N2O Gas Line Hose with Gas
Cylinder
Align the CGA326 hand knob (or equival e nt cylinder coupler) from the gas
hose line with the N2O gas cylinder’s threaded connection port , and screw i n place by turning the hand knob in clockwise rotation.
26
Page 28
Hand tighten the connection and verify there is no leak when cylinder valve is
opened.
If a leak is de tect e d, tig ht en the hand knob conne cti on f urt her . A wren ch m ay
be needed to tighten the connection.
If the leak cannot be eliminated, either the N2O gas line hose or the N2O gas
cylinder may need to be repl ac ed.
To disconnec t fr om the gas cy l in der , tur n the hand k nob in a counter-clockwise
rotation un til the connection is disengaged.
NOTE: Keep gas cylinder or gas line c anisters stan di ng vertically, in an up-right position
NOTE: Only me di c al g rade N2O gas and an N2O gas cy l inder should be used in the gas line connection.
3.7. Installing N2O Exhaust Hose
Align the 50 ft (15m) exhaust hose with the N2O Exhaust por t of the AC M an d push
in place.
Run the exhaus t hos e t o an ap pro pri at e vent in g mecha nis m tha t mee t s the hos pi tal
and local r egulation.
NOTE: The venting mechanism needs to vent at least 60L per minute.
To remove the exhaust hose fr om th e ACM uni t , si mply pull bac k on th e tu bing unti l it
releases. See Figure 9.
NOTE: It is not necessary or required for the N2O Exhaust Hose to be removed from
the ACM unit at the end of the procedure. The hose may be left in place.
27
Page 29
Figure 9 – Connecting the N2O gas line hose and N2O exhaust hose to the ACM
3.8. Connecting and Disconnecting the Handpiece
Connect the Handpiece directly to the ACM.
Insert the Handpiece pneumatic interface connectors into the pneumatic
receptacles on the front panel of the ACM, ensuring that each pneumatic inlet and exhaust connection is fully seated and secured.
Insert the thermocouple connectors into their respective color coded connection
ports on the front panel of the ACM. See Figure 10.
NOTE: Typically, you will connect the Handpiece to the ACM when the ACM has
been powered up and is in READY operating mode (see Section 4.2 regarding the READY mode). However, the Handpiece may be connected when powered up, or prior to powering up the ACM.
NOTE: Once the Handpiece has been connected, it cannot be disconnected from the
ACM by pulling on the cable. To disconnect the Handpiece, slide back on the pneumatic connector sleeve and remove it from the ACM pneumatic receptacle.
NOTE: Connection and disconnection of the Handpiece with the ACM unit should
ONLY take place with the ACM unit in the READY state, N2O cylinder valve in the CLOSED position, and ACM unit fully depressurized by pulling on the System Pressure Relie f Knob i f the unit has b een pr ess ur i z ed pri or .
NOTE: Refer to the Handpiece instruction sheet for more detailed information about
connecting the Handpiece to the ACM in a sterile environment.
28
Page 30
Figure 10 – Connecting the Handpiece to the ACM
29
Page 31
4. Instructions For Use
4.1. Powering Up the ACM
1. Ensure that the ACM has been plugged into a proper hospital pow er receptacle.
NOTE: Do not use extension c ords or thr ee-prong to two-prong adapt ers where
applicable. The power cord ass embly sho ul d be per i odic al ly check ed f or damaged insulation or connectors .
2. Apply power t o the ACM unit by flipping the ON/OFF power swi tch located on the
power entry module at the rear panel. The GREEN LED on th e pow er switch should energize i f p ower is being applied to the system.
3. The ACM unit will go through a system Power On Self Test (POST). If POST
passes, the system transitions to the READY mode. If any por tio n of the POS T fails, the system transitions to the FAULT mode.
NOTE: Refer to Section 4.2 ., bel ow , for a full d escr i pti on of t he READY and FAULT
modes, as w el l as all the other op er a t ing modes.
4. It is highly recommended that the ACM system be prepare d a nd po w ered up w ell i n
advance of st art of the proc e dur e to verify its o perational readiness a s well as preheating of N2O cylinder g as pressure to raise it to its optimal performance level.
NOTE: N2O cylinder val ve m us t be op en f or th e sy stem to de tect its pres enc e an d to
start the preh ea ti ng cy cle.
NOTE: Heating of the N2O cylinder is auto mat e d and cont rol le d by t he ACM uni t
only as needed and when appropriate. Operator intervention is no t required.
30
Page 32
4.2. Operating Modes
The ACM operates in one of five modes: READY, FREEZE, DEFROST, ERROR and FAULT modes. These mod es are i de n tified by the syste m s t a tu s indicator LEDs and the ablation s ta tus indicat or LEDs located on th e front of the ACM unit. See Figure 11, below.
4.2.1. READY Mode – This mode is en ter ed au to matic al ly upon successful
execution of Power On Self Test when the uni t i s firs t turned on, or following DEFROST Mode u pon venting of the Handpiece by depressi ng the Ablatio n Act iv ati on Sw it ch or t he Footswitch. This indicates that the system is rea dy f or th e next c ry o abl ati on r un.
4.2.2. FREEZE Mode – This Mod e i s en ter ed from the READY Mode when the
user initi ates the cry o ablation cy c le by pressin g and relea s ing the Ablat ion Activation Switch or the Footswitch. In this mode the N2O gas is allowed to cycle thr oug h the handpiece causing a temperature drop to take place at the handpiec e pr obe.
4.2.3. DEFROST Mode – This Mo de is enter ed automatically from FREEZE
Mode upon expi ra ti on of th e abl ati on ti mer , or man uall y by the op era t or when the Ablat ion Acti v ati on Swi tc h or th e Footswitch is actuated while in the FREEZE Mode. In this mode th e h an dpiece probe temperatur e i s actively forc e d towards the ambient temperature.
Once the h a nd pi e c e temperature has reached 10°C or higher, the DEFROST status LED will start to blink to indi cate to the us er the ACM unit is ready to transition back to the READY Mode or retur n t o t he FREEZE Mode to start the next ablation cycle.
International units:
The ACM unit returns t o the READY M ode by press i ng a nd releasing t h e Ablation Activation Switch or the Footswitch. The ACM unit will automatical ly vent and depr ess ur iz e the ha ndp iec e in preparati on for the next freeze cy cl e.
US units:
The ACM unit can be returned to the FREEZE Mode by pre ssi ng and releasing the Ablation Acti va tio n Switch or depressing and rel easi ng the Footswitch, which causes the handpiece to vent and start the freeze cy cle again.
The ACM unit can be r eturned to t he R EADY Mode by pressing t h e Ablation Acti va tio n Sw itc h or depressing the Footswitch and holding it in place for 2 sec onds or l ong er. The ACM unit will automatically vent and depressurize the han dpiec e. Onc e d epr ess uriz ed , th e Abla tio n Ac tiv ati on Switch or the Footswitch can be releas ed .
NOTE: ACM does allow e ar l y transition o ut from the DEFROST Mode into
either th e READY Mode or the FREEZE Mode upon user req ues t and upon handp iec e t emper at ure reaching above 0°C.
31
Page 33
NOTE: Handpiece probe temperature may drop temporarily during
depressurization phase (transition from Defrost to Ready state).
4.2.4. ERROR Mode – This Mode is ent ere d up on d etec ti on o f any recov er abl e
error conditi on s duri ng any Mod e exc ludi ng t h e FAUL T Mod e. The ACM unit will display the correspondi ng error me ss a g e when in READ Y Mode, and may halt operation temporarily upon detection of the error conditio n. These are recoverable errors and do not impact continued use of the system. The System Maintenance LED will be active to alert the user that an error condition has occurred.
To reset t he Maintenance LED, hold down bot h th e Tim er Increase and Timer Decrease button and power cycle the unit. The system will undergo power-on self test for system functionality confirmation and return to the operating state if appropriate.
4.2.5. FAULT Mode – This Mode is entered upon detection of any unrecoverable
error condi tion duri ng a ny Mod e. T he sy stem i s in opera ble i n thi s Mo de un til the unit is first power cycled, and only if the fault condition no longer exists or has been remedied.
32
Page 34
4.3. Delivering Cryo Energy
4.3.1. Connect the Handpiece, Pneumatics, Cylinder Heater Band, and
Footswitch
Connect the Handpiece, N2O Gas Line Hose, N2O Exhaust Hos e, C ylinder H e at er Band, and Footswitch as described in Sections 3.4. through 3.8., and note no errors are i n di cated by t he AC M unit. The ACM should indicate that the system is in the READY mo de, a nd the di spl ay s how ing abl ati o n par am eter s. S ee Figure
11. If a Handpiece or the Handpiece thermocouple is not connected, the ACM unit will
indicate suc h by displ a y ing an error mess ag e i n the pr ob e tem per atur e di spl ay window wi th “E-H”. See Figures 12A and 12B.
Figure 11 – ACM Indicating READY Mode
Figure 12A – ACM Indicating Missing Handpiece Thermocouple
33
Page 35
Figure 12B – ACM Indicating Missing Handpiece
4.3.2. Position the Handpiece
To positio n the Handpi ece, follow the Instructions for Use provided with t he Handpiece.
4.3.3. Delivery of Cryo Energy
1. To start the cryo ablation cycle, simply press and release the Ablation
Activation Switch or the Footswitch. The ACM unit’s Ablation Status Indicator LED will transition from Ready to Freeze.
2. N2O starts to flow through the ACM pneumatics and into the cryo ablation
handpiece, and the probe temperature starts to drop.
3. Upon probe temperature dropping below the set threshold (value is
programmable with default value of -40°C) the ablation timer starts its decrement from a preset value (this value is also programmable with default value of 120 s).
NOTE: The Freeze Mode can be manually terminated at any moment by
pressing and releasing the Ablation Activ ati on Switch or the Footswitch, causing the ACM unit to transition into Defrost Mode.
NOTE: The Ablation Timer value can be manually adjusted from its start
value either prior to the start of the ablation cycle or during the freeze cycle using the Timer Increase and Timer Decrease key pad. Each key press increases or decreases the timer value by 10 seconds.
NOTE: The factory default values of 120 seconds for ablation duration and
-40°C for the probe temperature threshold value can be altered by the operator, and set as a site default values, follow i ng the parameter entry procedure as defined in Section 4.4.
4. When the ablation timer reaches zero, the ACM unit automatically
transitions into Defrost Mode. As it transitions into the Defrost Mode, the Ablation Timer display will now show the actual total ablation run time. The ACM unit’s Ablation Status Indicator LED will transition from Freeze to Defrost.
34
Page 36
5. At this point, the ACM unit can be made to transition into one of two modes
depending on the ACM model: Freeze Mode or Ready Mode.
International – ACMs
Defrost to Ready:
Press and release the Ablation Activation Switch or the Footswitch. The ACM unit vents and depressurizes the handpiece. The ACM Ablation Status Indicator LED will transition from Defrost back
to Ready. As it transitions back to the Ready Mode, the Ablation Timer display also resets to the target ablation time value.
Once the Handpiece has been depressurized, it is ready for the next ablation run, which may take place immediately thereafter.
US – ACMs
Defrost to Freeze:
Press and release the Ablation Activation Switch or the Footswitch. The ACM Ablation Status Indicator LED will transition from Defrost back
to Freeze and the ablation cycle starts again. As it transitions back to the Freeze Mode, the Ablation Timer display also resets to the target ablation time value.
Defrost to Ready:
Press and hold the Ablation Ac ti v at ion Switch or the Footswitch for three seconds, until the ACM unit begins to vent and depressurize the handpiece. Total depressurization requires 30 seconds to complete.
The ACM Ablation Status Indicator LED will transition from Defrost back to Ready. As it transitions back to the Ready Mode, the Ablation Timer display also resets to the target ablation time value.
Once the Handpiece has been depressurized, it is ready for the next ablation run, which may take place immediately thereafter.
NOTE: The transition from Defrost to either Freeze Mode or Ready Mode
will not occur until the probe temperature is warmer than 0°C . Actuation of the Ablation Activation Switch or the Footswitch when the probe temperature i s colder than 0°C is simply ignored by the ACM unit.
NOTE: The Defrost status LED will start to blink indicating to the operator
when the ACM unit is ready to transition back to either the Freeze Mode or the Ready Mode.
NOTE: Handpiece probe temperature may drop temporarily during
depressurization phase (transition from Defrost to either Freeze Mode or Ready Mode).
NOTE: Total Handpiece depressurization requires 30 seconds to complete.
35
Page 37
4.3.4. Powering down the ACM
Upon completion of the procedure, the ACM uni t can be powered down.
NOTE: Verify that the ACM unit is in the Ready Mode prior to powering down
the unit. If the unit is not in the Ready Mode, cycle the ACM by continuously
pressing the Ablation Activation Switch or the Footswitch until the unit returns to the Ready Mode.
CAUTION: Do not remove cryo handpiece from the ACM unit if the ACM has
been powered down prior to the system being in the Ready Mode as N2O gas may be trapped within the cryo handpiece interface and the handpiece couplers may still be under pressure.
To remove any trapped N2O gas, turn the ACM unit back ON. Depress the Ablation Activation Switch or the Footswitch to initiate an ablation. After two seconds, stop the ablation by depressing and releasing the Ablation Activation switch or the Footswitch. Depress and hold the Ablation Activation Switch of the Footswitch until depressurization is initiated, then release the Ablation Activation Switch or the Footswitch. Wait 30 seconds for complete depressurization, and then turn OFF the unit utilizing the ON/OFF switch.
When the ACM enters the Ready Mode, the ACM may now be powered down, and the cryo handpiece can removed from the ACM safely and disposed of properly.
NOTE: When un plugging the handpiece, remove the large diameter tube first.
If a rush of gas is heard or felt, pause before unplugging the small diameter tube until no gas is noticed.
1. Close the N2O gas cylinder valve.
2. Depressurize the ACM by pulling on the Pressure Relief Valve Knob located
in the rear of the ACM. This will remove excess N2O that may be trapped within the ACM pneumatic system.
3. Unplug the power cord from the hospital mains outlet, and position on the
Cart.
4. Disconnect the cryo handpiece from the ACM unit and dispose of properly.
5. Position the footswitch from the ACM on the Cart.
6. Return ACM Cart back to proper storage.
36
Page 38
4.4. Parameter Entry Mode / Adjusting t he S ystem Default Param et e r Values
1. Power up the ACM unit. The system will go through its Power On Self Test (POST)
routine.
2. Upon completion of POST, press and hold down the Ablation Activation Switch for
approximately three seconds and release.
3. The User Display will start to blink indicating entry into the parameter entry mode.
4. There are two parameters that can be modified by the operator:
Ablation Timer / Duration: Freeze duration value in seconds
Probe Temperature Threshold: Temperature value at which the ablation timer
starts its countdown sequence
5. Ablation Timer / Duration Adjustment
Using the Timer Increase and the Timer Decrease key pad adjust the timer value until the desired time value is shown. The displayed value corresponds to the desired length of freeze time in units of seconds. Valid range is from 010 to 300 seconds in 10 second increments.
Once the desired time value is displayed, press the Ablat ion Activation Switch to accept the value.
The system will now start blinking the display of the probe temperature threshold parameter.
6. Probe Temperature Threshold
This value is used by the system to start the ablation timer countdown when the probe temperature has dropped below the set threshold value.
Using the Timer Increase and the Timer Decrease key pad adjust the probe temperature threshold value until the desired temperature value is shown. The displayed value corresponds to the desired probe temperature threshold value in units of degrees Celsius. Valid range is from -10°C down to -70°C in 5°C increments.
Once the desired temperature value is displayed, press the Ablat ion Activation Switch to accept the value.
7. The entry of the parameter value is now complete, and the system exits the parameter entry mode and returns to the operating mode with the new set of parameters.
8. Factory default values are:
Ablation Timer / Duration = 120 seconds
Probe Temperature Thr eshol d = -40°C
37
Page 39
Possible Cause
Solution
DISPLAY
MESSAGE
DESCRIPTION
SOLUTION
Cold junction temperature out of range
System Controller Board
Service for support
5. Troubleshooting
Use the follow ing sec tio ns to hel p tro ubles h oot po ssi bl e pr obl ems with t he AC M. To clear a faul t or error code from the ACM:
1. Power OFF the ACM
2. Depress and hold bot h th e increase (+) and decrease (-) buttons
3. Power ON the ACM. An “E” will b e di splayed in th e N2O gauge display
4. Release the increase(+) and decrease (-) buttons simultaneously
5. ACM will cycle thr o ugh its startup test an d c o nt i nu e into Ready Mode
6. If the Fault/Error Code returns, repeat ste ps 1-5, ensuring complete d epr essi o n of
the buttons
5.1. No ACM Power / Display Function
If there is no ACM power or display func ti on, attempt to correct this problem using the following checklist.
ACM not turned on Turn power on ACM not plugged in Confirm electrical connections and then turn power on Blown fuse Replace fuse, confirm electrical connections and turn power on. Internal ACM failure Contact AtriCure Customer Service
5.2. ACM Fault Codes
If a fault condi tio n shoul d occ ur , the N2O Gas Gaug e display on the front panel will display the following fault code. Contact AtriCure Custo mer Service for support.
NOTE: The ACM will be unusable until the P-Error is resolved .
P1
P4
System Controller Board – Fuse F2 blown
Contact AtriCure Customer Service for support
requires service. Contact AtriCure Customer
38
Page 40
DISPLAY
MESSAGE
DESCRIPTION
SOLUTION
System Controller Board Error – Fuse
Contact AtriCure Customer Service
System Controller Board Error – Triac
Contact AtriCure Customer Service
Handpiece pressure regulator out of
Contact AtriCure Customer Service
Internal pressure difference greater than
Wait for N2O Cylinder to cool down,
N2O Cylinder Temperature High
Reposition Heater Band, or
5.3. ACM Error Codes
If an error condition should occur, the N2O Gas Gaug e display on the front panel will display th e following error code while the A CM unit is in the READ Y Mode. The System Maintenance Needed indicator LED will also be lit. Use the table below to attempt to resolve th e following application er rors.
NOTE: The ACM will remain usable while displaying an E-Error
E1
E2
E3
E4
E6
blown
inoperative
range
check value
Cylinder over heated (pressure gage over 900psi)
In-line N
N2O Gas Cylinder Empty, or
Heater Band is not attached to
tank, or
Heater Band secured improperly
o Bottom of Heater Band is not
o Some spring retainers are not
O gas filter dirty
2
approximately 2 inches from the bottom of the cylinder
secured
for support
for support
for support
or Replace cylinder with cool cylinder,
or Pull Red Knob to vent pressure, or Visually inspect In-line filter element
for debris (see section 9), or Contact AtriCure Customer Service
for support
Secure all 6 tensioning spring retainers, or
Wait for N or
Replace with new N or
Contact AtriCure Customer Service for support
O Cylinder to cool down,
2
O gas cylinder,
2
39
Page 41
DISPLAY
MESSAGE
DESCRIPTION
SOLUTION
Trapped N2O gas in the Inlet Line
For ACMs WITH an “L” on the
For ACMs WITHOUT an “L” label
No cylinder heater band current
Verify Cylinder Heater Band is
E7
-System
Maintenance
Light
illuminated (no
error code displayed)
Trying to enter Freeze Mode before connecting the handpiece), or
Powering down ACM before fully depressurizing the handpiece, or
Leaking Inlet Valve
Note: User may be unable to connect handpiece to the ACM.
Cylinder heater band
Heater Band is not secured to tank (i.e. lying on floor, table, etc.), or
Heating element malfunction, or
Thermal cutoff switch was activated
(heater band reached 65°C – cylinder will feel very warm to the touch adjacent to the heater band)
Note: May notice hot rubber smell Caution: cylinder heater band may be
hot to the touch.
top of the unit:
Turn off ACM for a minimum of 4 seconds and turn the ACM back on, or
If problem persists, contact AtriCure Customer Service for support.
on the top of the unit:
Close the N2O cylinder. Connect the Handpiece Thermocouple Connections. Enter into Freeze Mode. Pull the red System Pressure Relief knob in the back of the unit until all gas has been released. Depress Button to initiate Defrost.Depress Button and hold 3­seconds for Ready state Connect probe and re-open the N cylinder, or
If problem persists, contact AtriCure Customer Service for support.
wrapped around the N2O gas cylinder, and all latches engaged, or
Wait for Cylinder to cool down and allow thermal cutoff switch to reset, or
Replace Cylinder Heater Band if defective heating element is determined, or
Contact AtriCure Customer Service for support
O
2
40
Page 42
DISPLAY
MESSAGE
DESCRIPTION
SOLUTION
READY Mode Only
Connect Handpiece
READY Mode Only
Replace Handpiece.
READY Mode Only
Replace Handpiece.
FREEZE and DEFROST Mode Only
Replace Handpiece.
FREEZE and DEFROST Mode Only
Replace Handpiece.
5.4. Handpiece Error Codes
If Handpiece related error condition should occur, the Probe Temperature display on the front panel will displ ay the following error code. Use the table below to attempt to r e s ol v e the followi ng ap plic ati o n err ors .
E-H
E-L
HI
LO
Handpiece Not Connected.
Defective Thermocouple.
Defective Thermocouple.
Probe temperature has risen above the operating limit.
Probe temperature has dropped below operating limit.
Thermocouple Connection
If Problem Persists, Contact AtriCure Customer Service.
If Problem Repeats, Contact AtriCure Customer Service.
If Problem Repeats, Contact AtriCure Customer Service.
If Problem Repeats, Contact AtriCure Customer Service.
41
Page 43
6. Symbols Used
Attention: consult accompanying documents
Dangerous Voltage
Power OFF Power ON Alternating Current
Fuses
Equipotential Termi nal
Type CF Applied Part
Start
READY
FREEZE
DEFROST
Gas Gauge (Time Remaining)
Timer
42
Page 44
Timer Decrease Button
Timer Increase Button
Probe Temperature
Connector
Gas Inlet
Gas Outlet
Temperature Adjust
Gas Exhaust
Power
Maintenance Needed
Low Gas Pressure
43
Page 45
Cylinder Heater Band
Footswitch
Maximum Pressure
Follow instructions for use
44
Page 46
7. Technical Specifications
7.1. Mechanical Specifications
Size: 17.5 in (W) × 27.0 in (D) × 4.5 in ( H) maximum
44.5 cm (W) × 68.6 cm (D) × 11.4 cm (H) maximum
Weight: 50 lb. (23 kg) absolute maximum.
7.2. Environmental Specifications
Operational temperature: +10°C to +40°C
Storage and Transit temperature: –35°C to +54°C
Humidity: 15 to 90% relati ve h umi dit y
7.3. Elect ri cal Sp ec ifi cat io ns
ACM1: 100-120VAC, (115 VAC nominal), 50/60 Hz
ACM2: 220-240VAC, (230 VAC nominal), 50/60 Hz
7.4. Mains Fuses
100 -120VAC, 50 / 60 Hz,: Replace fus es as mark ed:
4.0A/250V, T-lag, 5 × 20 mm, UL Recog niz ed , IEC A ppr ov ed
220-240VAC, 50 / 60 Hz,: Replace fuses as marke d :
2.0A/250V , T-lag, 5 × 2 0 mm, UL Rec og niz ed, IEC A ppr oved
7.5. Footswitch Specifications
Moisture protection rating: IPX8
7.6. Equipment Ty pe / Cla ssifi cation
Class 1 Equipment
45
Page 47
8. Preventive Main tenance and Clean ing of ACM
8.1. Preventive Maint enance
Perform annu al pr eve nt ativ e mai nt ena nc e proc edur es to ens ure all ACM components are functioni ng as de fin ed withi n thi s ma nual . Pay par ti cul ar atten tio n to oper at io nal and safety features, including but not limited to:
Electrical pow er cor ds for fr ay ing, da mag e, an d pro per gro undi ng
AC power swi tch
Indicator da mag e (P ow er On, Fault, Maintenance Needed, Low Cylinder Pressure,
Cylinder Heater Band A c ti ve, Footsw it ch A c t ive, Ready, Freeze, Defrost)
N2O Gas Gauge dis pl ay damage or loss of information
Ablation Ti mer di spl ay damag e or los s of in for mati on
Temperatur e display da mage or loss of information
Handpiece elec tro nic i nter fac e co nnec t or da ma ge, cr ack ing or i nabi li ty to inser t an d
latch Hand piece connector
Handpiece p ne u m ati c interfac e connector damage or i n abi l ity to ins er t and latch Handpiece pneumatic connector
Carrying handle damage or ina bili ty to fol d
Rubber feet damag e, crac ki ng or ina bili ty f or t he AC M to remai n st abl e on a fla t
surface.
Rubber align men t cu p dam age , cr acki ng or in abili ty for the AS B/ ASU to re mai n stabl e atop ACM a n d within th e al ignment cup.
Footswitch cord fr aying or damage
Footswitch connector damage cracking or inability to insert and latch footswitch plug
Footswitch pedal damage; check acti v ati on by pres sing and rele asi ng the pe dal
Other medical equipme nt that may be used sim ul t an eously with t h e ACM should als o be inspected for da mag e. Sp eci fic all y , chec k for ins ula tio n d amage o f electrical cables and associated connectors .
Visually ins pec t th e foo t sw itch for fl uid s or ot he r infec ti ous haz ar ds. Clean as nec ess ary using the i n s tructions in Section 8.2.
The ACM does not hav e any customer serviceabl e part s aside from mains fuses, gas line desiccant filter, and Inline gas filter element. For servi c ing issues, contact AtriC ure, Inc. at:
7555 Innovation Way Mason, Ohio 45 040 USA Telephone: 513-755-4100 866-349-2342 Toll Free
46
Page 48
8.2. Cleaning and Disinfecting
NOTE: Do not spra y or po ur liq ui ds dir ec tly on t he unit . NOTE: The unit and/or accessories cannot be sterilized. WARNING: Ensure Isopropyl Alcohol (IPA) is completely dry before operating the unit. CAUTION: Avoid caustic or abrasive cleaners
Guidelines The following guideline s a re rec ommended f or cleaning the unit. It is the user ’ s respon sib ility to qualify
any deviations from these processing methods.
1. Disconnect the unit or ca rt from the o ut let be fore cleaning.
2. If the unit and/or ac ces sor ies are co ntam ina te d w ith blo od or ot he r body fluids, they shall be cleaned before the contamination ca n dry (within two hours of con tami nation).
3. The outer surfaces of the unit and/or accessories shall be cleaned with 70% -90% I sopropyl alcohol (IPA) wipes for a minimu m of two min utes. Do n ot allow f lui ds to en ter the c ha ssis.
4. Pay attention to all are a s whe re f lu ids or soi l may g a ther, such a s under / a ro und t he ha nd les or any tight crevices/ grooves.
5. Dry the unit and/or ac c essorie s wi th a dry, white lint-free cloth.
6. Conduct a final confir ma tion of the c lea ning pr o ce ss by visually inspecting the white cloth for remaining soil.
7. If soil remains o n the w hite c lot h, re pea t step s 3 t hroug h 6.
8. After cleaning is com ple te, tur n t he un it on to per f orm Pow er O n Self-Test (POST). If any errors are received, contact Atri Cur e to beg in return process.
47
Page 49
ACM Model
Fuse Type
Manufacturer
Part Number
AtriCure
P/N
ACM1
T 4A L 250V
Schurter
0034.5049
C002262
ACM2
T 2A L 250V
Schurter
0034.5046
C002261
9. Servicin g of ACM Unit
The ACM unit does not have any cust omer serv i ceable parts aside from fuses and gas filter. The manufacturer does not require or recommend routine calibration / output verification.
As stated in Section 4, the ACM performs its own maintenance check every time it is powered ON. T he che c k is cal le d th e Pow er On Sel f Tes t (PO S T). No calibrat ion is needed for the system and the unit will generate an error if further maintenance is required.
Following identifies servicing that can be performed on the ACM unit based on Section 5, Troublesho ot ing guide.
9.1. Repla cement of A C Mains Fuses
9.1.1. Tools and Parts
Needle Nose Pliers
Fuses:
9.1.2. Procedure
The ACM unit has be en pr e-s e t at t he fac tor y to eith er 1 10V – 120V (ACM1) or 220V – 240V (ACM2). The Rating La bel below the Power E n tr y Module on the back panel of the ACM indica tes t he sel ect e d Inp ut Volt age for thi s uni t. This setting sho uld only be adj ust ed by the m anu fac tur er or by an au t horiz e d Atri Cur e technical service re pr es entative.
NOTE: ACM unit sh oul d be po w ered o ff a nd un plug ge d befor e c onti n uing wit h
the fuse replacement procedure.
1. Determine the fuse type by looking at the ACM Model Nu mb er or th e ACM Rati ng
Label.
2. Using needle nose pliers, car eful ly extra ct the f use box fr om th e pow er entry
module by sque ez ing d ow n on t he fus e box tabs in t he sl ots as sh ow n.
48
Page 50
In-Line N2O
Filter
AtriCure
P/N
Type
Manufacturer
Part Number
Filter
C002458
TF Series
Swagelok
SS-4F-K4-60
Filter Bonnet
Nut
C002500
TF Series
Swagelok
SS-4TF-K5
Filter Gasket
C002501
TF Series
Swagelok
SS-4TF-K2
1
3. Replace the ( 2) two fuses located in t h e fuse box. Ma ke sure the fuses are lined
up straight.
4. Line up the fuse cartridge so the guide tab is tow ar ds the power entry side as
shown.
5. Slide the fuse box back into the power entr y mod ule and seat it firmly.
6. Confirm operati on al st a tus by plug ging in the ACM and p owering it u p. Check to
see that t he self-test is completed without er rors.
9.2. Inspection and R ep lace ment of I n-Line N2O Filter
9.2.1. Tools and Parts
3/32 inch Allen Wrench
1.0 inch hex Socket
Socket Wrench
Filter:
9.2.2. Procedure
NOTE: ACM unit sh oul d be totally depressuri z ed, in the ready state , then
powered off and unplugged before continuing with the In-Line N2O Filter Inspectio n and/ or replacement procedure.
1. Turn over the A C M unit so that the botto m s ide is up.
2. Remove the 4 s crews holding the filter service cover in i ts place usi ng the 3/ 3 2
inch Allen wr ench .
49
Page 51
1 2 3
3. Using the 1 i nch (25 mm) hex Socket, loo se n the fil ter bonnet nut (Note: s n a p r ing
does not n e ed to be removed).
4. Visually inspec t th e out er s ur fac e of th e fil ter el eme nt for an acc u mul ati on of
debris. If mor e tha n hal f o f the outer sur fa ce is c oat ed w ith d ebris pr ocee d to step 5 to replace the filt er el e ment , ot her w ise insert the Fil ter El eme nt in to th e ho usi ng, verify the filter gas k et is in position and t hr e a d the Filter B onnet Nut ti g ht onto the housing proc eed to st ep 7.
5. Remove the old filter element and filt er gask et and replace with new. Breakdown
of each compone n t i s shown below for reference.
50
Page 52
1. Filter Bonnet Nut
2. Filter Element
3. Filter Gasket
Item
Supplied By
Part Number
Filter Cartridge
AtriCure
F021720
Filter O-ring
AtriCure
F010924
O-ring lubricant
AtriCure
C002502
US Tank Hose Assembly
AtriCure
A000837
International Tank Hose Assembly
AtriCure
A000838 Tip Washer
AtriCure
F021837
1. Desiccant Filter Cartridge
2. Filter Housing
3. Tip Washer
1 2 3
6. Insert the filt er el em ent c ent ering it i n the filter housing. Insert the filter gasket into
the housing. Install the Filter Bonnet Nut finger tight and tighten to approximately
45.8 ft-lbs (62.2 N-m).
7. Replace the filt er servi c e cov er to the ACM bottom. Tighten the four cover
retaining screws without stripping them.
8. Confirm oper ational status by plugging in the ACM and pow er ing it up. Check to
see that t he self-test is completed without er rors.
9.3. Repla cement of Ga s L in e Desiccant F ilter and Tip Washer
9.3.1. Parts
Replacement parts:
9.3.2. Procedure – Desiccant Filt er
1. Unscrew the filt er car tr i dge h ous ing by ro tati ng it counter-clockwise. Refer to the
figure bel ow.
51
Page 53
Note: There are two fil ter car tr idg e housi ng . Make sur e s erv ici ng of th e fil te r
cartridge is per for med onl y on the uni t nex t to th e pre ss ure gaug e as shown in t he figure below.
2. Remove the desiccant filter cartridge by rotating it counter-clockwise. Refer to the
figure bel ow.
3. Replace the desic ca nt filter car tr idg e with a new one.
4. Replace the filt er housi ng .
9.3.3. Procedure – Tip Washer (US Mode l)
1. Unscrew the w orn out Tip Washer by rotating i t counter-clockw ise and disc ard.
Refer to the figure below.
52
Page 54
Item
Supplied By
Part Number
ACM Footswitch
AtriCure
A000708
N2O Gas Line Hose Assembly (2.5ft /
0.75m)
AtriCure
A000837 N2O Exhaust Hose
AtriCure
C002051
Cylinder Heater Band (CMH15)
AtriCure
A000728
Cylinder Heater Band (CMH22)
AtriCure
A000727
AC Power Cord – 10 ft (USA)
AtriCure
C000262
Item
Supplied By
Part Number
ACM Footswitch
AtriCure
A000708
N2O Gas Line Hose Assembly (2.5ft /
0.75m)
AtriCure
A000838 N2O Exhaust Hose
AtriCure
C002051
Cylinder Heater Band (CMH15)
AtriCure
A000728
Cylinder Heater Band (CMH22)
AtriCure
A000727
AC Power Cord – 3.5 m (Europe)
AtriCure
C002090
2. Replace with new tip was her.
9.4. Other Replacement Components
9.4.1. USA Only
NOTE: Standard cylinder size for the AtriCure Cart is 20lbs with a CGA326 valve.
9.4.2. International
NOTE: Nitrous Oxide cylinders and valves are country specific.
9.4.3. Electromag n et ic Compatib i l ity
Use of access orie s ot her than ide nti fie d in Sec tio n 9.5.1 and 9.5.2 may result in increased EM ISS ION S or decr eas ed IMMUN I TY of the AtriCure Cryo Module.
53
Page 55
10. Accessories
10.1. CMH15, Cryo Module Cylinder Heater Band
Cylinder heater band with dimensions of 12 inch x 15 inch (30.5 cm x 38.1 cm).
10.2. CMH22, Cryo Module Cylinder Heater Band
Cylinder heater band with dimensions of 12 inch x 22 inch (30.5 cm x 55.9 cm). Spring Extension Kit is available for larger diameter cylinders – contact
AtriCure customer service
10.3. CMF1, Cryo Module Footswitch
Momentary contact, pedal switch with 15 feet signal cord which can be used in place of the Ablation Activation Switch at the ACM front panel.
11. Disposal
Follow loc al g overning ordinances a n d recyclin g pl a ns regarding di sposal or rec y c l ing of device components.
54
Page 56
Warranties
Limitation on Liability
This warranty and the rights and obligations hereunder shall be construed under and governed by the laws of the State of Ohio, U.S.A.
AtriCure, Inc. warrants this product to be f ree from defects in material and workm anship under norm al use and preventive maintenance for the respective warranty period shown below. AtriCure’s obligation under this warranty is limited to the repair or replacem ent, at its option, of any product, or part thereof, wh ich has been retur ned to AtriCure, Inc. or its Distributor within the applicable time period shown below and which examination disclosed, to AtriCure’s satisfaction, to be defective. This warranty does not apply to any product, or part thereof, that has been: (1) adversely affec ted due t o us e with devices manufactured or dis tribut ed by parties not authorized b y Atri Cure, Inc. (2) repaired or alt ered outsid e AtriCure ’s fac tory in a wa y so as to, in A triCur e’s judgm ent, affec t its stabilit y or reliability, (3) subjected to improper use, negligence or accident, or (4) used other than in accordance with the design and use parameters, instructions and guidelines for the product or with functional, operational or environmental sta ndards for sim ilar prod ucts g enera lly ac cepte d in t he industr y. AtriCure has no control over
the operation, inspection, maintenance or use of its products after sale, lease or transfer, and has no control of the selection of Customer’s patients.
AtriCure’s products are warranted for the following periods after shipment to the original purchaser:
AtriCure Cryo Module Unit ................................................................................ One (1) Year
AtriCure Cylinder Heater Band ......................................................................... One (1) Year
AtriCure Gas Line Hose Assembly……………………………………………… .. One (1) Year
AtriCure Cryo Footswitch .................................................................................. One (1) Year
Grounded Electrical Cord .................................................................................. One (1) Year
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING THE WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, AND OF ALL OTHER OBLIGATIONS OR LIABILI TIES ON T HE PART O F ATRICUR E, INC. AND IS A PURCH ASER’S EX CLUSIVE REMEDY. IN NO EVENT SHALL ATRICURE, INC. BE LIABLE FOR SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES INCLUDING, WITHOUT LIMITATION, DAMAGES RESULTING FROM LOSS OF USE, PROFITS, BUSINESS OR GOODWILL.
AtriCure, Inc. neith er assum es nor authorizes a ny other pers on to assum e for it any other liability in con nection with the sale or use of any of AtriCure Inc. pr oducts. There are no warranties that extend beyond the term s presented unless an extended warranty is purchased before the original warranty expires. No agent, employee
or representative of AtriCure has any author ity to change any of the foregoing or assume or bind Atri Cure to any additional liability or responsibility. AtriCure, Inc. reserves the right to make changes to products built
and/or sold by them at any time without incurring any obligation to make the same or similar changes on products previously built and/or sold b y them .
55
Loading...