Gas Detector with Data Storage
Thanks for purchasing our product. Please read the user manual carefully before using.
Keys Operation:
ON/OFF: long press for 3s to turn on/off the device, short press to confirm or save data.
Menu: in detection mode, press this button to enter into parameter setting mode, and in parameter
setting mode, press this button to move the cursor.
Plus: in parameter setting mode, short press it to select items upward or add value. Long press for 3 seconds to turn on/off alarm function quickly.
Minus: in parameter setting mode, short press it to select items downward or decrease value. Long press for 3 seconds to do zero calibration quickly.
Exit: in parameter setting mode, short press it to go back to previous menu. Long press to shift the gas concentration unit quickly, but it is restricted to PPM and mg/m3.
(Note: The long press function only work in detection interface. When you change or reset any data, please remember to click ON/OFF button to confirm and save the settings)
Power On
Press the ON/OFF for three seconds, the detector will be turned on and the pump will start to work. The screen displays the following interface successively: Brand and Logo, main parameter interface, initialing countdown. The countdown would be 60 seconds, which is to make sure that the sensors are fully activated. After 60 seconds, the device enter normal detection interface.
Zero Calibration
Zero calibration must be done in clean air. Zero calibration is needed when there is zero shift happens.
Put the detector in clean air for 3-5 minutes, under the detection interface, press the MINUS buttons for three seconds, the data of all channels will reset to zero automatically.
Note: Zero calibration of oxygen detector, nitrogen detector and carbon dioxide detector (target gases are component of the air), the data will be defaults to 20.9%VOL (O2), 79.1%VOL (N2), and 400ppm (CO2). Do not do the zero calibration in not clean air or when there is no zero shift happens.
Page 2
Welcome to visit ATO.com!
Tel: 800-585-1519 Email: sales@ato.com Webwww.ato.com
Process of Over-range incorrect operation
Users should avoid using gases which exceed the measure range to impact the sensor, because it will affect the lifespan and sensitivity of the sensor, even, "poison" the sensor. If there is any over-range incorrect operation makes the detector displaying an concentration at a large reading, the remedy is to take the device out of the environment immediately and put it in clean air for over half an hour, and then observe the reading, if it keeps going down, then wait until the reading back to zero before powering it off, and do the zero point calibration next time before using it. If the reading maintained at full scale, user should send the device back to manufacturer or agent for repair or replacing the sensor.
ON/OFF
PLUS MENU ESC MINUS LCDAir Chamber of Sensor Shockproof Cover Alarm Area
Operation Interface
When the detector is turned on completely, the device will enter the detecting interface. The detected gases will be showed on the screen (for example: O2, CO, H2S, EX). Time and battery power will be showed at the top right. When the alarm is on, there will be alarm symbol at the right bottom.
Alarm Status
When the detected gas concentration is lower than the low alarm value, the low alarm is auto on, the relative gas value will be highlighted, and the alarm sound will be on, light will be flickered. When the detected gas concentration is high than the high alarm value, the high alarm is auto on, the relative gas value will be highlighted, and the alarm sound will be on, light will be flickered. When there is low alarm and high alarm at the same time, both relative gas values will be highlighted, and high alarm will take the priority.
Page 3
Welcome to visit ATO.com!
Tel: 800-585-1519 Email: sales@ato.com Webwww.ato.com
Low Battery Alarm
When the battery is low, there will be sound, light low battery alarm. There will be sound and light every 5 seconds to remind the user to charge the device in time. When the battery is used up, the device will be turned off automatically.
Alarm Setting
Press Menu”, choose the Alarm, and start to set the alarm value. There are five options under the alarm setting mode. The first option is the channel, there will be 4 channels for 4 in 1 multi gas detector. Channel choose: press ON/OFF button, the cursor will be at the channel number, then press the PLUS and “MINUS” button to choose the channel, press the ON/OFF button to confirm. Alarm setting: The second option is the low alarm value, and the third option is the high alarm value. Choose the option that you want to change, then press ON/OFF to confirm, then press the PLUS and MINUS to increase or decrease the value, press the MENU to move cursor. After the setting is done, press ON/OFF button to save the settings.
Settings
Press MENU, choose SET, and start to set the detector. There are five options: Language, gray level, LCD backlight, auto off, RST reset to factory setting. .
Storage set
Press the MENU, choose HIST, it means history, then you can get in the storage interface. There are four options in the history set.
1. CYC cycle Set: it is the storage time interval, only when you set the time interval, the detector will start to record the data. The time interval can be 5 seconds, 20 seconds, 30 seconds, etc. The highest interval is 1 hour. If you dont want to record the data, please set the interval 0000.
2. View: view the history data. At the top is the record number, press the PLUS and MINUS to change the number, and you will see the date, time and the record gas concentration.
3. DEL: Delete record data. To make sure the detector work perfectly and you can check the history data quickly, we suggest you to erase the history data regularly. Change the NO to YES, then press the ON/OFF to confirm, then all the record will be deleted.
4. Alarm: You can check the alarm record here. At the top is the total record number, press the PLUS and MINUS to change the number, and you will see the date, time and the alarm value.
Calibration
Note: Calibration needs to be done by professionals and with necessary accessories and standard gas.
Page 4
Welcome to visit ATO.com!
Tel: 800-585-1519 Email: sales@ato.com Webwww.ato.com
Press MENU, choose the CALIBRATION. There are five options in the calibration interface:
1. Channel option: you need to choose the right channel/gas type before calibration, there will be 4 channels for the 4 gases multi gas detector.
2. Zero calibration option: put the detector in clean air for 3-5 min, then press ON/OFF button, wait for the value become stable, press ON/OFF again, the zero calibration is finished. You can also do the zero calibration in the detection interface, press the MINUS” buttons for three seconds, the data of all channels will reset to zero automatically.
3. Target point calibration:
Prepare the standard gas before starting the target point calibration. Choose the target point calibration, press ON/OFF button to enter. Then you will need to enter password. (If you need password, please contact the manufacturer to get it) After entering the password successfully, press ON/OFF to enter the target point calibration interface. The value at the lower left side is current concentration of the gas, the value at the lower right side is the AD value. Change the target concentration to the standard gas concentration, after the value on the left side and right side is stable, press the ON/OFF”, to finish the calibration.
The detailed process is as below:
1. Connect the detector with standard gas cylinder by hose.
2. Enter the target point calibration interface, enter password correctly,
3. Change the target point value to the standard gas value.
4. Release the standard gas to the detector in a flow rate of 700mL/min. The value of current concentration will become larger gradually, and become stable after 30seconds.
5. When the value of concentration is stable, press the ON/OFF button, the value of current concentration will be same with or almost same with the target point value. The calibration is finished.
6. Cut off the standard gas supply. Press ESC to go back to detection interface.