ATO SKY2000 User Manual

www.ato.com
Welcome to visit ATO.com!
Tel: 800-585-1519 Email: sales@ato.com Webwww.ato.com
User Manual - SKY2000
Thanks for purchasing our product. Please read the user manual carefully before using.
Keys Operation:
Turn on/off: long press for 3s to turn on/off the device, short press to confirm or save data.
Menu: in detection mode, press this button to enter into parameter setting mode, and in parameter setting mode, press this button to move the cursor.
Plus: in parameter setting mode, short press it to select items upward or add value. Long press for 3 seconds to turn on/off alarm function quickly.
Minus: in parameter setting mode, short press it to select items downward or decrease value. Long press for 3 seconds to turn on/off air pump quickly.
Exit: in parameter setting mode, short press it to go back to previous menu. Long press to shift the gas concentration unit quickly.
(Note: The long press function only work in detection mode. When you change or reset any data, please remember to click ON/OFF button to confirm and save the settings. Remember to turn on the pump when start the detection. Please turn off the pump when there is no need to detect, which will be helpful for the service life of the pump)
Power On / Off
Put the detector in clean air, and press the ON/OFF for three seconds, the detector will be turned on and the pump will start to work. The screen displays the following interface successively: Brand and Logo, main parameter interface, initialing countdown. The countdown would be 60 or 120 seconds, which is to make sure that the sensors are fully activated. After countdown, the device enter normal detection interface. Take the detector out of the detected environment and wait for the data to get down to zero, then press the ON/OFF to turn off the detector
Zero Calibration
Zero calibration must be done in clean air. Better to do once after alarming happens. Press MENU to enter parameter setting mode, select the "calibrate zero", and press ON/OFF to enter the
mode. Observe the value of “real time concentration”. If the reading is not zero and the drifting is too big when the reading is stable, then it is necessary to conduct zero point calibration.
Press ON/OFF, the real-time concentration will go back to zero, and press ESC to go back to detection mode, gas concentration values also back to zero.
Note: Zero calibration of oxygen detector, nitrogen detector and carbon dioxide detector (target gases are component of the air) can not be done in the air. Only the target point calibration (value of target point concentration is the standard concentration values in the clean air) can be done.
www.ato.com
Welcome to visit ATO.com!
Tel: 800-585-1519 Email: sales@ato.com Webwww.ato.com
Process of Over-range incorrect operation
Users should avoid using gases which exceed the measure range to impact the sensor, because it will affect the lifespan and sensitivity of the sensor, even, "poison" the sensor. If there is any over-range incorrect operation makes the detector displaying an concentration at a large reading, the remedy is to take the device out of the environment immediately and put it in clean air for over half an hour, and then observe the reading, if it keeps going down, then wait until the reading back to zero before powering it off, and do the zero point calibration next time before using it. If the reading maintained at full scale, user should send the device back to manufacturer or agent for repair or replacing the sensor.
— Air Inlet
Air Outlet
air chamber of sensor ④ LCD ⑤ PLUS ⑥ MINUS ⑦ MENU ⑧— ESC — ON/OFF
-----------------------------------------------------------------------------------------------------------------------------
Note
*If detector is used for the gas cylinder detection, considering the high pressure in the cylinder, please use a pressure regulator, and keep it as the below data: Flow rate: 800mL~1L/minute Pressure: 0.1MPa or 1Bar *Please put the detector in clean air for about 5 minutes and waiting for the data falling down to normal before turning it off after measurement.