Atmos Cam, Strobo LED, S 61 Servant vision Mounting Instructions

English
Mounting Instructions
ATMOS Cam
ATMOS Strobo LED
in ATMOS S 61 Servant vision
2012-08 Index: 01
Mounting Instructions
ATMOS MedizinTechnik GmbH & Co. KG
Ludwig-Kegel-Straße 16 79853 Lenzkirch/Germany
Tel. +49 7653 689-222 Fax +49 7653 689-292
M531.2010.B M531.2020.B M531.2055.B M531.2065.B M531.2070.B M531.2080.B M531.2090.B M531.2050.B
service@atmosmed.de www.atmosmed.de
1.0 General Notes
These assembly instructions explain the retrofi tting of the options in the ATMOS S 61 Servant vision from construction year
07/2009:
531.2010.0 ATMOS Cam without image storage (retrofi t kit 531.2010.N)
531.2020.0 ATMOS Cam 21 DV (retrofi t kit 531.2020.N)
531.2080.0 ATMOS Cam 21 DV DATA (retrofi t kit 531.2080.N)
531.2065.0 ATMOS Cam 31 (retrofi t kit 531.2065.N)
531.2055.0 ATMOS Cam 31 DATA (retrofi t kit 531.2055.N)
531.2070.0 ATMOS Cam 31 DV (retrofi t kit 531.2070.N)
531.2090.0 ATMOS Cam 31 DV DATA (retrofi t kit 531.2050.N)
531.2050.0 ATMOS Strobo 21 LED (retrofi t kit 531.2050.N)
On the older versions of the ATMOS S 61 Servant vision the rear lead frame on the camera side may need to
be adjusted.
1.1 Gültige Gebrauchsanweisung
The valid operating instructions for the ATMOS S 61 Servant (REF 53x.xxxx.B) are constituent part of these service instructions. Precise observation of the instructions
regarding fi rst-time operation, handling and maintenance
is a precondition for carrying out service work.
1.2 Information on manufacturer
Further information, accessories and spare parts
are available from:
ATMOS
Medizintechnik GmbH & Co. KG Ludwig-Kegel-Str. 16 79853 Lenzkirch Germany
Phone: + 49 7653 689-222
1.3 Exclusion of liability
No warranty claims will be accepted for any damage caused by the use of non-original accessories and consumables.
ATMOS only assumes responsibility for the unit, as far as safety, dependability and functionality are concerned, if assembly, readjustment, modifi cations, extensions or
repairs are carried out by ATMOS or by persons authorized by ATMOS.
The user‘s safety and the trouble-free operation of the de­vice are only guaranteed if original ATMOS components are used. In the case that non-original accessories and consumables
are use, ATMOS cannot guarantee safe operation and safe
function of the components from the ATMOS S 61 Servant.
Fax: + 49 7653 689-392 (Vertrieb Inland / Domestic Sales) + 49 7653 689-391 (Vertrieb Ausland / Export Sales) + 49 7653 689-292 (Kundenservice / Client Service)
E-Mail: service@atmosmed.de Internet: http://www.atmosmed.de
Reproduction and distribution of these instructions
– even in part – only with the written consent from
ATMOS Medizintechnik GmbH & Co. KG.
2
Particularly important notes
1.0 General Notes
1.4 For your safety
The ATMOS S 61 Servant ENT workstation is produced
according to IEC 601/ EN 60601 and listed in the following classes:
VDE Class of protection 1
Class IIa (EEC 93/42).
The device may only be connected to a properly installed
shockproof socket. Correct confi guration for installation to country-specifi c connections:
green/yellow: protective conductor (PE) blue: neutral conductor (N) black resp. brown: phase (L)
Do not constrict the air supply on the back of the unit!
The ATMOS S 61 Servant may be used in supervised ope-
ration by qualifi ed personnel only which has been authori-
sed by ATMOS and which has been trained for operating the appliance (MPBetreibV §2 Abs. 1 and 2).
The mains voltage specifi ed on the type plate must corre-
spond with the data of the power supply system.
Prior to opening, switch off the unit and separate it from
the line voltage.
The operation via an isolating transformer is only permitted
when an insulation monitoring or another type of monitoring function for electrical protection is provided.
The ambient conditions specifi ed in section „Technical
specifi cations“ must be strictly observed!
Switch off main switch after fi nishing work in practice and
close water supply, if present.
Attention when working with endoscopes near the light
sources. Do not look directly into the cone of light! At a contingent light blackout slowly remove the endoscope out of the working area!
The ATMOS S 61 Servant may be operated only in rooms
used for medical purposes, but not in areas subject to
explosion hazards.
The ATMOS S 61 Servant fully complies with the electro-
magnetic immunity requirements of standard IEC 601-1-2/
EN 60601-1-2 „Electromagnetic compatibility - Medical Electrical Equipment“.
The ATMOS S 61 Servant may not be operated with units
not complying with the requirements of standard EN
60601-1 „Medical Electrical Equipment“ and EN 60601-
1-2 „Electromagnetic compatibility (Medical Electrical
Equipment)“.
ATMOS cannot be hold liable personal injury and damage
to property, if
no original ATMOS parts are used,
the instructions for use in the operating instructions or
in these service instructions are not being observed,
assembly, new settings, alterations, extensions and
repairs have been carried out by personnel not autho­rised by ATMOS.
These service instructions correspond with the construction
of the unit and with the current status of safety-related standards at the time of printing. Proprietary rights are
existing for all described circuits, processes, names, soft-
ware programs and units.
During function tests at the open device all measures
must be taken which are required for protection against an electric shock.
Attention! Danger of injury at rotating pumps respectively
motors.
Attention! When working at the light module, there is a
danger of eye injury (dazzlement or burn) caused by the
extreme brightness. In addition there is the danger of skin burn due to the extremely high temperature generated by the bulb.
During service work at the ATMOS S 61 Servant - espe-
cially at the suction system - hygiene regulations must strictly be observed.
After each maintenance or service work a function test
must be performed according to the check list as well as a safety-related inspection acc. to DIN EN 0751-1.
ATMOS recommends an inspection according to
the testing instructions and a safety-related in­spection once a year.
3
1.0 General Notes
1.5 Hygiene measures during repair work at ATMOS treatment units
Disposable gloves must be worn when working with contaminated materials.
The single-use gloves must be changed prior to touch other things respectively parts which
are not contaminated.
During work it is most important not to hurt yourself - DANGER OF INFECTION!
Dismantled, contaminated parts which were replaced require waterproof and airtight pa-
cking and must then be disposed of professionally.
After you fi nished your work you must thoroughly clean and disinfect the tools you used.
Please observe the contact time prescribed by the manufacturer!
Furthermore please thoroughly clean and disinfect the working area with surface disinfec-
tant. Please observe the contact time prescribed by the manufacturer!
At the end you must thoroughly clean and disinfect your hands.
For service technicians we recommend all required active immunisations.
4
2.0 Termination 531.0030.0
If both options, camera and Strobo LED are used simultaneously the ATMOS S 61 Servant vision must be equip-
ped with a termination for the power supply cord (531.0030.0) instead of the demountable mains cable!
008.0704.1 Bayonet cap
530.1147.0 Mains connection bracket
008.0877.0 Cable gland
530.1147.0 Mains connection bra­cket
530.1148.0 Insulating board
531.0168.0
cable S61 fi xed connection
Fig. 1 Fixed mains connection 531.0030.0
008.0026.1 Clamp 1P (3x)
531.0169.0 Cable fuse holder
008.0704.0 Fuse holder
Fig. 2
008.0877.0 Cable gland
531.0168.0
Cable S 61fi xed connection
5
Loading...
+ 11 hidden pages