Atmos i View 21 COLPO, i View 31 COLPO Operating Instructions Manual

English
ATMOS® i View COLPO
Operating Instructions
GA1GB.420101.0
2018-04 Index 24
Table of contents
1.0 Introduction .........................................................3
1.1 Notes on operating instructions ............................3
1.2 Intended use .........................................................4
1.3 Function ................................................................5
1.4 Explanation of pictures and symbols ....................6
1.5 Scope of supply ................................................... 7
1.6 Transport and storage ...........................................7
2.0 For your safety ....................................................8
3.0 Setting up and starting up .................................9
3.1 Overview .............................................................. 9
3.2 Setting up ............................................................10
3.2.1 Connection to the mains supply ..........................10
3.2.2 Microscope overview ..........................................10
3.2.3 Operating elements at the colposcope ...............11
3.2.4 Rear view of the control device of the ATMOS
3.2.5 Rear view of the control device of the ATMOS
(not with an integrated HD camera) .................... 11
3.2.6 Rear view of the control device of the ATMOS View 31 COLPO with an integrated HD camera .12
3.3 Integration possibilities........................................12
3.4 Starting up...........................................................13
3.5 Operation requirements ......................................13
3.6 Starting up at a glance ........................................14
®
i View 21 COLPO ................................. 11
®
i View 31 COLPO
®
i
4.0 Operation ...........................................................15
4.1 Colposcope suspension ......................................15
4.2 Mechanical arm...................................................15
4.3 Hand grips...........................................................15
4.3.1 T-hand grip ..........................................................15
4.3.2 Lateral double hand grip .....................................15
4.4 Adjusting the interocular distance .......................16
4.5 Adjusting the eye pieces .....................................16
4.6 Exchanging the lenses ........................................17
4.7 Exchanging the lenses with manual ne
focusing...............................................................17
4.8 Exchanging the VarioFocus lens .........................17
4.9 Adjustment of the 5 fold magnication changer ..17
4.10 Focussing............................................................18
4.10.1 Fine focussing .....................................................18
4.11 Exchanging the binocular tube............................18
4.12 Pivoting green lter .............................................18
4.13 Pivoting H.A.S.I. lter ..........................................19
4.14 Shadowless illumination......................................19
4.15 Colposcope zoom and object eld size...............19
4.16 Measuring scale ..................................................19
4.17 Image and video recording .................................20
4.17.1 Adjusting the light mode of the integrated HD
camera ................................................................20
4.18 Endoscope adapter .............................................21
4.19 HD adapter..........................................................21
5.0 Cleaning and care .............................................22
5.1 Basic instructions ................................................22
5.2 Cleaning the mechanical microscope surface ....22
5.3 Cleaning of lenses / eyepieces ...........................22
5.3.1 Cleaning optical surfaces ....................................22
5.3.2 Optical surface of the endoscope port ................22
5.3.3 Fogging of optical surfaces .................................23
5.4 Recommended surface disinfectants ..................23
5.5 Hygiene plan .......................................................23
Further information, accessories, consumables and spare parts are available from:
ATMOS
MedizinTechnik GmbH & Co. KG
Ludwig-Kegel-Straße 16
79853 Lenzkirch
Germany
Phone +49 7653 689-0
Fax: +49 7653 689-190
+49 7653 689-292 (Service Centre)
atmos@atmosmed.de
www.atmosmed.de
6.0 Maintenance and Service .................................24
6.1 Basic instructions ................................................24
6.2 Sending in the device..........................................24
6.3 Exchange of spare parts .....................................24
7.0 Troubleshooting ................................................25
8.0 Accessories and Options .................................26
9.0 Technical data ...................................................27
10.0 Disposal .............................................................28
11.0 Notes on EMC ....................................................29
1.0 Introduction
1.1 Notes on Operating Instructions
These operating instructions contain important notes on how to operate the ATMOS® i View COLPO safely, correctly and effectively. Their reading helps to avoid risks, and also to reduce repair costs and down-times. This increases, amongst other things, the reliability and service-life of the colposcope.
These operating instructions serve not only for new operating personnel to be instructed in its use, but also for use as a reference manual. Reprints (also in extracts) only with permission in written form by ATMOS.
These operating instructions must always be kept available near the colposcope.
Care and safety inspections in conjunction with professional execution provide for operational safety and readiness for use of your ATMOS
®
i View COLPO and are therefore a must besides regular cleaning.
Repair work and safety inspections may be carried out only by expert personnel authorised by ATMOS. By applying only original spare parts you will have the guarantee that operational safety, readiness for work and the value of your ATMOS
• The product ATMOS
®
i View COLPO will be preserved.
®
i View COLPO bears CE marking CE according to the EC Directive of the council
for medical products 93/42/EEC and meets the basic requirements of Appendix I of the directive.
®
• The product ATMOS
i View COLPO complies with all applicable requirements of the Directive 2011/65/
EC restricting the use of certain hazardous substances in electrical and electronic equipment (“RoHS”).
• The declaration of conformity and our general standard terms and conditions can be obtained on our website at www.atmosmed.com.
The quality management system applied at ATMOS has been certied according to international standards EN ISO 13485.
• Prior to start-up please peruse chapter 2.0 „For your safety“, in order to be prepared for any possible dangerous situations.
These operating instructions are valid for the following devices:
®
ATMOS
i View 21 COLPO .........................REF 605.0000.0
Colposcope with an integrated, fanless, high transmission, high performance LED light in the colposcope head
®
ATMOS
i View 31 COLPO .........................REF 606.0000.0
Colposcope with an integrated, fanless, high transmission, high performance LED light in the colposcope head
Please keep this document for future consultation!
1.0 Introduction
1.2 Intended use
Name: ATMOS® i View 21 COLPO
®
ATMOS
Main functions: Visualization of natural body orices (genital area) for examinatory purposes
i View 31 COLPO
Medical indications / application:
Specication of the main
function:
User prole: Examinations and surgical interventions which are carried out using the
Patient groups: No restrictions
Application organ: Natural orices (portio and vulva)
Application time: Short-term, under normal conditions for permanent use over a maximum up to
Application site: Application sites are hospitals, doctor’s ofces, ORs at gynaecologists. The
Standard gynaecological examinations. Visual control of the genital area.
The application organ is illuminated for examination purposes and can be visualized on a monitor if desired by an integrated, fanless high transmission,
high performance LED light source, 5 fold magnication changer, integrated camera module, pivoting green lter and an automatic light control via tilt
sensor; the light output is min. 120 kLux (200 mm)with a colour temperature of
5.500 K ± 10%.
colposcope may only be performed by doctors with appropriate training.
Only qualied personal with a proper hygiene training may prepare the
colposcope for surgical interventions.
Installation and maintenance may only be carried out by service technicians who were trained and authorised by the manufacturer.
30 days
examination with the colposcope may only be executed by medically trained persons.
The colposcope may only be used in closed rooms, on rm ground or mounted
to the ATMOS
In ORs an appropriate protective sleeve must be used.
®
Chair 41 Gyne .
Contraindications: • No application in ophthalmology.
The product is: active
Sterility: The colposcope is not a sterile product.
Single-use product / reprocessing:
The colposcope is intended for multiple use. The device and parts of the accessories are reusable. For information on reprocessing, cleaning and disinfection please see the operating instructions.
1.0 Introduction
1.3 Function
The ATMOS® i View COLPO is a complete colposcope system, consisting of optics and lighting. It produces outstanding pictures for examination purposes with the use of latest LED technology and patent registered optics. The interaction between the integrated
fanless, high transmission, high performance LED, the apochromatic optics and the precisely tting options offer best working
quality.
The ergonomically assorted buttons, two selectable hand grips and the integrated control panel provides the user with highest level of ergonomic comfort and suitability for daily use as well as an outstanding and intuitive handling. Via the control panel the individual options of the ATMOS / stopping of possible video sequences, the operator is capable of manually switching the LED light source on and off despite the activated automatic light control. Due to the variety of options which the ATMOS
position to congure a colposcope to suit his requirements. The following functions can be chosen optionally:
4 lenses with different focal distances (200, 250, 300 and 400 mm) with or without ne focussing or a VarioFocus 200-350 mm
(easy exchange of the lenses due to the respective thread on the colposcope head)
5 fold magnication changer. Exact adjustments due to turning disks on both sides.
• Binocular tubes, simple adaption due to the Zeiss dove tail.
Pivoting colour lter
• Measuring scale
• Shadowless illumination
®
i View COLPO can be activated. Besides the triggering of the camera (freeze frame) and starting
®
i View COLPO has to offer, the user is in a
Due to the LED light source and the integrable camera solution (SD integrated respectively as HD or endoscope adapter for the connection of an external camera) the ATMOS
In connection with the mechanical support arm and the numerous connection possibilities to units and stands the ATMOS
®
i View COLPO is a guarantor for best image quality.
®
i View
COLPO offers countless system possibilities, which can be individually adapted to suit the users environment!
®
These operating instructions describe all functions with a maximum conguration of the ATMOS
i View 31 COLPO.
1.0 Introduction
1.4 Explanation of pictures and symbols
Short cuts / symbols contained in these operating instructions
Follow the arrows
General information
Move, plug... in this direction
Please press where dot indicates
Please read, important information
Graphic symbols contained in these operating instructions
Warning, special diligent notice
Numeration
Check
!
Symbols ATMOS
SN Serial number
®
i View
Manufacturing date Manufacturer
Observe operating instructions! Observe operating instructions!
REF
click
) Important information
Order number
Turn, shift... in this direction
Replace
Engage, check
correct t
Weight adjustment for the carrier arm Professional disposal
Alternating current Fuse
2
Control panel buttons
Single use product - not for reuse. Exchange after use.
Do not look directly into the light source of the ATMOS
Light on / off (independent of automatic light control)
Video recording (start / stop)
®
i View COLPO.
This product complies with the relevant requirements of EU Directives
Sitting is prohibited.
Freeze frame
With an integrated HD camera: Adjusting the light mode of the camera
1.0 Introduction
Foot Switch
Out 1
Out 2 Out 3
Video Out
Foot Switch
Only ATMOS® i View 21 COLPO
Output of the power supply for the electron­ics in the colposcope
Fuse
®
Only ATMOS
i View 31 COLPO
Fuse Potential equalisation acc. to IEC 604175021
Colposcope Foot switch
Record function Freeze
Potential equalisation acc. to IEC 604175021
not in use
USB port
S-video-output (not with an integrated HD camera)
Video In Intern 1
Out 1
Video Out
Out 2 Out 3
Output signals of the tilt sensor in the carrier arm system
Video In Extern 2
Input video signal internal / external (Only with an integrated HD camera)
Output video signal (Only with an integrated HD camera)
1.5 Scope of supply
• Prior to dispatch, the ATMOS® i View COLPO was subjected to an extensive functional test and was carefully packed. Nevertheless, please compare the contents of the shipment on completeness immediately upon receipt (see delivery note).
Colposcope
Operating Instructions
Colposcope arm
Protective cover
1.6 Transport and storage
• After the transport of the ATMOS® i View in
temperatures below 0°C or prior to rst start up
it should be kept at room temperature for at least six hours. If the ATMOS acclimatized it may not be used as damages to the electronic components may be the result.
Only transport the device in a shipping carton, which is
padded and offers sufcient protection.
If damage occurs during transport:
• Document and report the transport damage.
• Send the device to ATMOS (Chapter „6.2 Sending in the device“ on page 24).
®
i View COLPO is not
Ambient conditions:
• Transport / storage:
-10...+50 °C;
30...95 % humidity without condensing air pressure 500...1060 hPa
• Operation: +10...+35 °C;
30...95 % humidity without condensing air pressure 700...1060 hPa
2.0 For your safety
!
For your safety
• To safely disconnect the unit from the mains, the power cord must be removed from the IEC connector of the control device!
®
• The ATMOS with IEC 60601-1-1/EN60601-1 and it is a device with protection class I. In order to avoid the risk of electrical shock, this unit may only be connected to a mains supply with properly installed earth conductor.
• Power cables, accessories and access cables need to be checked for defects prior to setting up the ATMOS COLPO. Damaged cables must be replaced immediately.
• The ATMOS
qualied personnel.
• The ATMOS in an explosion-hazardous environment. Explosion-
hazardous areas may be caused by the use of ammable
anaesthetics, skin cleansing products and skin disinfectants.
If uids penetrated the ATMOS be sent in and may only be used after the check up of an ATMOS authorised person.
• After the transport of the ATMOS
below 0°C or prior to rst start up it should be kept at room
temperature for at least six hours. If the ATMOS COLPO is not acclimatised it may not be used.
• Do not plug in electric connections (plug, socket) under the use of force. If this is not possible check whether the
plug ts the socket. If you should ascertain a defect in the
connection you should have it repaired by our service.
• Never look straight into the sun with lenses or eye lenses.
• Always make sure that you do not blind patients with the light source! Watch out that patients do not look directly into the light source! Never look directly into the light source! > Damage to the eyes due to the strong glare.
• Please pay attention to the period tests in chapter 6 „Service and maintenance“ on page 24.
• Prior to every use the colposcope suspension (all joints included) need to be checked for safe connections.
• Take care that the patient does not touch the device or have any contact with it.
• Please observe the EMC Directives. Failure to follow this guideline can result in a hazard.
• Make sure that the unit is positioned so that all the controls and the on/off switch are always accessible.
• Dispose of wrappings accordingly.
• Before connecting the ATMOS be checked whether the requested mains voltage of the ATMOS mains power supply.
• Only proper and undamaged plugs and extension cables may be used.
i View COLPO is a device designed in line
®
i View
®
i View COLPO may only be operated by
®
i View COLPO is not designed to be used
®
i View COLPO it needs to
®
i View in temperatures
®
i View
®
i View COLPO it needs to
®
i View COLPO matches the mains voltage of the
• To disconnect the ATMOS
mains supply, rst remove the plug from the wall outlet.
Disconnect the connection line on the ATMOS
®
i View COLPO from the
®
i View COLPO afterwards only. Never touch plug or line with wet hands.
• Please observe the ambient conditions stated in the technical data (chapter 9.0).
®
• The ATMOS
i View COLPO meets the immunity to interference requirements of IEC 60601-1-2 / EN 60601­1-2 „Electromagnetic Compatibility – Medical Electrical Devices“.
• ATMOS is not liable for personal injury and damage to property if
- no original ATMOS parts are being used,
- the advice for use in these operating instructions is not being observed,
- assembly, new settings, alterations, extensions and repairs have been carried out by personnel not authorised by ATMOS.
• Unplug the device immediately if you observe fumes, sparks or weird noises.
• With every light source a warming of tissue due to absorption may occur. Please make sure to reduce duration of use to a minimum, to switch off the light source when not in use and to check heat development if necessary.
®
• The ATMOS
i View COLPO may be operated only in rooms used for medical purposes, but not in areas subject to explosion hazards and in oxygen rich environments.
• Take into consideration, when setting up the colposcope, that the elastic force of the arm – without colposcope head – is exceedingly strong. Operate the break of the height adjustment carefully.
• Risk of injury! Take care not to roll the mobile stand over your feet when moving the stand.
• Please note that only PCs and monitors with IEC 606010­1/EN 60601-1 approval may be connected to the video outlet of the ATMOS
®
i View COLPO supply module!
• During operation, the user is obliged to regularly check the microscope for proper function. In the unlikely event of failure of the microscope, the user must take precautions to continue the treatment of the patient with suitable methods.
3.0 Setting up and starting up
3.1 Overview
ATMOS® i View 21 COLPO ATMOS® i View 31 COLPO
Examination colposcope with an inte-
Description
Integrated high performance white light LED
Automatic light control
Optimised stereo effect
Measuring scale Optional Optional
Integrated operating panel
Colour lter Optional
Integrated camera -
HD adapter for an external camera
Endoscope adapter - Optional
Mains voltage 100–240 V 100–240 V
grated, fanless, high transmission, high performance LED light in the colposcope head
Optional Optional
- Optional
Examination colposcope with an inte­grated, fanless, high transmission, high performance LED light in the colposcope head
Optional green lter or Optional H.A.S.I. lter
Optional SD camera or
Optional HD camera
min. 120 kLux (200 mm)
Light output
Operating life of the LED
Colour temperature 5.500 K ± 10 % 5.500 K ± 10 %
Scope of delivery
min. 80 kLux (250mm)
min. 55 kLux (300 mm)
min. 30 kLux (400 mm)
50 000 hours 50 000 hours
Dust cover, operating instructions
min. 120 kLux (200 mm)
min. 80 kLux (250mm)
min. 55 kLux (300 mm)
min. 30 kLux (400 mm)
Dust cover, operating instructions
3.0 Setting up and starting up
3.2 Assembly
Please make sure that the static conditions stated by ATMOS MedizinTechnik are met (for details see the separately enclosed
!
document „Static requirements for installing the ATMOS i View“). The fullment of these requirements must be conrmed by an
authorized expert.
Mains voltage and fuse: Mains voltage: 100-240 V; 50/ 60 Hz, fuse: 2 x T 3.15 A
Please note that only PCs and monitors with IEC 606010-1/EN 60601-1 approval may be connected to the video outlet of the ATMOS
3.2.1 Connection to the mains supply
Potential equalisation:
The ATMOS potential equalisation rail in the room if need be. Hereby user/patient safety can be increased especially in the case of a defective earth conductor. For connecting the device's potential equalisation plug with the potential equalisation rail of the room, use the potential equalisation cord with REF 530.0030.0.
3.2.2 Microscope overview
®
i View COLPO supply module!
®
i View COLPO supply module has a rear connection for potential equalisation which can be connected to the
Wide eld eye
piece lenses with binocular tube
Rotating disk 5
fold magnication
changer
Lens
Lateral double hand grip (optional)
T-hand grip (optional)
Pivoting colour lter
(optional)
Measuring scale (optional)
Brightness control
HD adapter for SONY digital camera (optional)
Control panel (optional)
10
3.0 Setting up and starting up
3.2.3 Operating elements at the colposcope
LED
STROBO
AFD
MODE
Light on / off (independent of
Freeze frame
Video recording (start / stop)
With an integrated HD camera: Adjusting the light mode of the camera
automatic light control)
3.2.4 Rear view of the control device of the ATMOS® i View 21 COLPO
Output of the power supply for the electronics in the colposcope
Connection for potential equalization performance acc. to IEC 60417-5021
IEC power plug with fuse inlay for the connection to the mains power supply
3.2.5 Rear view of the control device of the ATMOS
Output signal of the „Video record function“
USB port for the transfer of the key status of the „Freeze frame “- and „Video recording -functions“.
Output signals of the tilt sensor in the carrier arm system
Output signal of the „Freeze frame function“
S-Video-output of the integrated SD camera
not in use
®
i View 31 COLPO (not with an integrated HD camera)
Connection for potential equalization performance acc. to IEC 60417-5021
Connection for the supply to the colposcope electronics and control line
Connection to the foot switch for switching between the light channels
IEC power plug with fuse inlay for the connection to the mains power supply
11
3.0 Setting up and starting up
3.2.6 Rear view of the control device of the ATMOS® i View 31 COLPO with an integrated HD camera
HD video input. Can only be used for the HD camera module
Connection for the supply to the colposcope electronics and control line
Video In
Intern 1
Microscope
not in use
HD video input from an external HD video source
Video In Extern 2
Strobo
Sensor
Output signals of the tilt sensor in the carrier arm system
HD video outlet from the video source internal 1 or external 2
Video Out
Out 1
Freeze
Record
Output signal of the „Video record function“
Out 2 Out 3
PC/USB
Output signal of the „Freeze frame function“
Connection to the foot switch for switching between the light channels
Foot Switch
100 - 240V~
2 x T 3,15A
PC connection (optional)
IEC power plug with fuse inlay for the connection to the mains power supply
Connection for potential equalization performance acc. to IEC 60417-5021
3.3 Integration options
Please note the assembly instructions for the integration options.
®
ATMOS
For the integration into ATMOS
Please note that a save combination of the ATMOS guaranteed with the ATMOS
Mobile stand
When moving the roller stand please make sure that the colposcope arm is in a retracted position and the screws are tightened.
Risk of injury! Take care not to roll the mobile stand over your feet when moving the stand.
!
When the device is placed in the working position the brakes must be locked.
Chair 41 Gyne
®
Chair 41 Gyne.
®
Chair 41 Gyne.
®
i View COLPO can only be
12
3.0 Setting up and starting up
3.4 Starting up
• Remove the colposcope from the packaging. Check whether the mains current on the type label corresponds to the mains power supply.
Peruse safety information in part 2.0 prior to starting up the device for the rst time.
• Check the scope of delivery.
• After the transport of the colposcope in temperatures below 0°C it must be kept at room temperature for at least six hours. If the colposcope is not acclimatized it must not be used.
• Take into consideration, when setting up the colposcope, that the elastic force of the arm – without colposcope head – is exceedingly strong. Operate the break of the height adjustment carefully.
®
• To activate the ATMOS
3.5 Operating requirements
Please note that the following requirements must be adhered to for further application after the installation of the device:
• All joints and connection parts which are responsible for the safety of the device are securely fastened.
• All electronic connections (cables, plugs, power cables etc.) are in good order and condition.
The specied mains voltage and frequency on the colposcope corresponds to the supply network.
• The colposcope is connected to a safety connection socket with the provided mains cable.
• Attention, never point or direct the beam into the patients eyes. Do not look directly into the light source.
!
• With every light source a warming of tissue due to radiation and absorption could occur. This could result in damage to the biological tissue. Please keep the luminosity and duration of use to a minimum. Switch off the light source when not in use and check the heat development if necessary.
i View COLPO please use the power switch on the front side of the control device.
13
3.0 Setting up and starting up
3.6 Starting up at a glance
Adjust colposcope to initial position on the colposcope suspension by use of the xing wheel.
Adjust colposcope horizontally and vertically.
Adjust all the clamps on the carrier and oat arm to secure the movability of the arm in compliance with the
requirements.
Swing in colposcope into the working space.
Adjust the interocular distance by pressing or pulling the lens tubes together or apart. The interocular distance is
perfectly adjusted when you look through and a circular picture is perceived!
Adjusting the eyepieces.
Persons without glasses Persons with glasses
Eyepieces remain in initial
position (eyepieces are pulled
out). Dioptre scale adjusted to
zero
People with defective vision
and glasses
Keep glasses on, push
eyepieces in direction of the
lens tube until they engage
audibly. Adjust dioptre scale
to zero.
People with defective
vision without glasses
(refraction values known)
Remove glasses and adjust
dioptre scale to matching
number (eyepieces are
pulled out).
People with defective
vision without glasses
(refraction values
unknown)
Remove glasses and adjust
both eyepieces to +5 dpt.
Remove the lens tube from
the colposcope head and focus on an object* in the
distance. The object still
looks blurred. Turn the
dioptre ring of the rst
eyepiece slowly in clockwise
direction until the object
is sharp. Keep your other
eye closed while adjusting
the eyepiece. Repeat this
procedure for a couple
of times to determine an
average value. Adjust
the second eyepiece by
the same procedure and
reattach the lens tubes to the
colposcope head with the
connective screw (eyepieces
are pulled out).
* Never use the sun as an
object!
14
Set the 5-fold magnication changer unit to maximum zoom (2.0). Approach the object with the colposcope
(according to the chosen focal distance) until the image is sharp. If the zoom level is changed the grade of
sharpness is retained.
Brightness can be adjusted by the rotary knob on the bottom of the device if necessary.
4.0 Operation
Rotating
knob
4.1 Colposcope suspension
By means of a corresponding suspension the colposcope head is connected laterally to the colposcope arm. The complete range of cables run through the suspension ­therefore no disturbing cables are visible from the outside (with the exception of the connection to the HD adapter and direct connection to a monitor). Due to a rotating knob, which is situated on the side of the suspension, the colposcope can be adjusted vertically to suit the individual requirements of the user.
To x the colposcope head turn the rotating knob towards
you in a clockwise direction.
To loosen the colposcope head turn the rotating knob towards you counter-clockwise.
Attention: Check the secure connection of the colposcope to the suspension prior to every use!
4.2 Mechanical arm
The mechanical colposcope arm can be adjusted via four set screws according to the individual requirements of the user. Choose the strength of the clamping so that the free movement of the arm suits your requirements. Turn the
set screw in clockwise direction to x the arm. To loosen
the arm turn counter-clockwise. To align the arm please observe the assembly instructions for the integration possibilities.
Attention: Prior to use ensure that the brakes of the support arm are set correctly.
Automatic light switching: Once the arm is in the lower position the LED light switches off automatically.
Adjusting screws
4.3 Hand grips
When purchasing the ATMOS® i View COLPO you may choose between two versions of handles.
4.3.1 T-hand grip
(see gure)
4.3.2 Lateral double hand grip
The position of the lateral double hand grip can be gradually adjusted by simultaneously pulling and turning the handle.
15
4.0 Operation
4.4 Adjusting the interocular distance
The interocular distance is adjustable between 50-75 mm.
• Swivel the colposcope into the work space.
• Look through the eye lenses and push or pull the eye lens tubes together or apart with both hands.
The interocular distance is perfectly adjusted when you
look through with both eyes and a circular eld is visible!
4.5 Adjusting the eye pieces
Persons without glasses:
• Eyepieces remain in initial position. Initial position = The eye steerings of the eyepieces are pulled out.
• Make sure that the zero of the dioptre scale complies with the index on the eyepieces.
Persons with glasses:
• People with defective vision and glasses keep their glasses on and push the eyepieces in direction of the lens tube until they engage audibly. Adjust dioptre scale to zero.
• People with defective vision (with known refraction values) should take their glasses off and adjust the dioptre scale on the eyepieces to the matching number (the eye steerings of the eyepieces are pulled out). The process of focussing is performed as described in Chapter 4.9.
• People with defective vision without glasses adjust both eyepieces to +5 dpt. Remove the binocular tube and the eyepiece from the microscope head and focus on a distant object*. The object still looks blurred. Slowly
turn the dioptre ring of the rst eyepiece in clockwise
direction until the object is sharp. The other eye must remain closed. Repeat this procedure for a couple of times in order to determine on an average value. Adjust the second eyepiece by the same procedure. Reattach the lens tubes to the microscope head with the connective screw. The process of focussing is performed as described in Chapter 4.10.
* Never use the sun as an object!
16
4.0 Operation
4.6 Exchanging the lenses
The designated thread on the colposcope head allows
for easy exchange and xation of the different lenses.
Due to the integrated screw mount lenses can be
loosened by turning it to the left hand side and xated by
turning it to the right.
Grub screw
Lens
Setting dial
4.7 Exchanging the lenses with
manual ne focussing
Mount lens as described above and secure it with the intermediate screwed ring.
4.8 Exchanging the VarioFocus lens
To loosen the VarioFocus lens from the colposcope head, turn it to the left. To tighten the VarioFocus lens on the colposcope head, turn it to the right onto the thread.
Position the setting dial
The dial can be positioned on either side of the VarioFocus lens.
Attention! During the process rmly hold the VarioFocus
lens just in case it may loosen itself from the colposcope head and fall off.
Loosen the three grub screws on the lens. Continue to hold the lens and turn the setting dial in the desired position. Tighten the three grub screws.
4.9 Adjusting the 5-fold magnication
changer
The 5-fold magnication changer from ATMOS enables
free range zoom between 0.5x up to 2.0x.
• Select the desired zoom factor by selecting one of the lateral rotating knobs.
• Pay attention that the chosen zoom factor engages audibly with the groove.
• Freely adjustable zoom factors: 2.0 - 1.4 - 1.0 - 0.7 -
0.5.
The magnication which points in the direction of the
eyepieces is the current magnication.
17
4.0 Operation
Fine focussing
4.10 Focussing
Adjust the zoom to maximum (2.0) on the magnication unit.
• Approach the object with the colposcope until the image is sharp.
• If the zoom level is changed the pre-adjusted degree of sharpness is still maintained.
4.10.1 Fine-focussing
The optional ne focussing allows for sensitive and
precise focussing in a 17 mm range. Fine focussing is necessary in order to focus accurately while zooming in.
• Replace the mounted lens with the appropriate lens
for ne focussing (simple mounting due to the screw
mount at the colposcope head. Secure with the intermediate screwed ring).
• Conduct the focussing as described above.
• Adjust focus by use of the lateral adjusting disk.
Binocular straight lens tube
Binocular angled lens tube 45°
Undo screw
Undo screw
4.11 Exchanging the binocular tube
The tubes focal distance of 160 mm allows for a comfortable and fatigue-proof observation of the object with both eyes. Working is made easier due to the exceptionally large exit pupil and an increased stereo base of 24 mm.
Please hold the lens tube with one hand while loosening the screw. Otherwise the lens tube could drop.
• Loosen the screw on top of the lens tube and remove the tubes from the colposcope head.
• Make sure that the gudgeons and grooves of the dove
tail xation engage and the tubes lie at.
• Tighten the screw again.
Check for a secure t.
4.12 Pivoting green lter
The pivoting colour lter enhances contrast of the
microscopic picture for better visibility of vessel structures.
• Turn the function knob by 90° in clockwise direction to
swing in the lter.
• By turning the knob by 90° in an anti-clockwise
direction the lter is removed from the optical beam
path of the colposcope.
18
Pivoting green lter
4.0 Operation
4.13 Pivoting H.A.S.I. lter
The pivoting H.A.S.I. lter gives a more contrasting and
clearer view of the mucosa.
• Turn the function knob by 90° in clockwise direction to
swing in the lter.
Pivoting H.A.S.I.
lter
4.14 Shadowless illumination
The option shadowless illumination prevents instruments from causing shadows in the eld of view. This option cannot be retrotted.
• For the shadowless illumination no operating steps are required.
4.15 Colposcope zoom and object eld size
• By turning the knob by 90° in an anti-clockwise
direction the lter is removed from the optical beam
path of the colposcope.
Lens f in mm
equals the
approximate
working
distance
200 6.4 / 31 9 / 22 12.8 / 16 18 / 11 25.6 / 8 16 x
250 5.1 / 39 7.2 / 28 10.2 / 20 14.3 / 14 20.5 / 10 16 x
300 4.3 / 47 6 / 33 8.5 / 23 12 / 17 17 / 12 16 x
400 3.2 / 62 4.5 / 44 6.4 / 32 9 / 22 12.8 / 16 16 x
* Read off at factor 1 when using the colposcope zoom without the zoom unit.
0.5 0.7 1* 1.4 2.0
4.16 Measuring scale
Measuring scale Figure not true to scale
Factor display on the magnication unit Eyepieces with
Total zoom / visual eld Ø in mm
20 mm
10 mm
0.75 mm 2 mm
lens tubes
f = 160 mm
5 mm
7 mm 10 mm
This gure is for guidance
only and may not be used for measuring absolute quantities.
Via a small turning knob beneath the lens a true to scale dimension scale can be faded into the eld of the illumination light path. This documentation display enables the measurement of objects regardless of the selected magnication. The scale will be
displayed in both the 3D picture and on all camera pictures and if required it can be faded out at any time.
• To fade-in the scale turn the knob by 45° in a clockwise direction.
• Via a 45° rotation in anti-clockwise direction the scale can be faded out from the path of illumination.
The following measures have to be observed: - Distance 2 mm, - Line width 0.75 mm
Please note that these specications are only correct for the following combination: Measuring scale for 300 mm lenses, 300 mm lenses with or without ne focusing or wide angle eyepieces 16 x.
19
4.0 Operation
4.17 Image and video recording
LED
STROBO
AFD
MODE
Integrated camera: If desired an SD camera or a HD camera can be integrated in the ATMOS
External video sources: External video sources can be controlled via the panel buttons if they are connected to the jack plugs “Freeze” and “Record”.
Control panel buttons:
Save image.
Start / stop the recording of a video frequency
Adjusting the light mode of the integrated HD camera.
The data are transmitted to a connected PC (USB interface). The ATMOSoft software can process the data.
Only with an integrated HD camera
You can change between the integrated HD camera and external video source by switching the LED light on or off. As soon as the LED light goes off the integrated camera is switched off and the data from the external video source is displayed (Video Out 1 - 3).
Also observe this within the automatic light switching.
®
i View 31 COLPO.
4.17.1 Adjusting the light mode of the integrated HD camera
By pressing the MODE button once the current light mode of the integrated HD camera is displayed on the monitor. By pressing the MODE button again the light mode can be changed.
Light mode Display on the monitor
Standard LED light remains unchanged.
When the power is switched on, the default setting is automatically selected.
Center LED light will be displayed with less
reections.
Suitable for recordings through a speculum.
Warm LED light appears warmer.
20
4.0 Operation
4.18 Endoscope adapter
The standardized endoscope adapter allows for an easy connection to an external ATMOS Cam or other external endoscope or digital camera (third party products). The ATMOS Cam can be easily and swiftly attached to the endoscope adapter by means of a special clip seal. Other endoscope cameras which provide a standardised connection interface can also be adapted without any trouble. To attach an external digital camera a special adapter (which is suitable for the respective digital camera) is required.
4.19 HD adapter
Due to the especially developed HD adapter it is possible to connect a SONY digital camera with e-mount bayonet to your ATMOS archive HD resolution pictures.
At dispatch the HD adapter is covered with a cover cap. This cap is to protect against contamination and has to be re­attached at any time e.g. if the camera is removed or when the adapter is unused.
®
i View Colpo. This camera enables you to take and
Please make sure that externally connected cameras do not exceed a weight of 300 g.
21
5.0 Cleaning and care
5.1 General information on cleaning and disinfection
Prior to cleaning
Medical devices like the ATMOS® i View COLPO must be fail safe at all times. Therefore we recommend prior to every use:
if
necessary
) The described action relating to cleaning and disinfection
resp. sterilisation do not substitute the relevant instructions which must be adhered to prior to operation!
• All disinfectants used for the disinfection of the ATMOS View COLPO must be approved.
®
i
) Always observe the concentration specications and
instructions by the respective manufacturer!
5.2 Cleaning the mechanical colposcope surface
All mechanical surfaces of the ATMOS® i View COLPO must be wiped and disinfected after each application. Do not use aggressive or abrasive cleansing agents.
Residues can be removed with a mixture made from equal parts of ethyl alcohol and distilled water to which a drop of standard washing-up liquid is added.
®
If uids penetrated the ATMOS
!
authorised person.
Disconnect the plug from the mains current prior to cleaning and disinfecting the colposcope surface.
!
For a sterile covering of the device the single use sterilization drapes may be used. The sterilization drapes may only be used once.
Afx the cover loosely so that there is enough room left for the colposcope support and the unit. The drapes must be especially
loose around the hand grips as the physician must be able to use the operating elements though the cover.
i View COLPO it needs to be sent in and may only be used after the check up of an ATMOS
5.3 Cleaning of lenses / eyepieces
5.3.1 Cleaning optical surfaces
The multilayer T* coating of optical components (e.g. eyepieces, lenses) results in optimum image quality.
Image quality could be reduced even by the slightest contamination of the optics or by ngerprints. In order to protect the internal
optics from dust, the instrument should never be left without a safety cover, HD adapter, lens, binocular tube or eyepiece installed when it is not in use.
After use the colposcope should be covered in order to protect it from dust. Always store lenses, eyepieces and accessories which are not being used in clean, dust-free cases.
The external surfaces of optical components should only be cleaned when required.
• Dust which has accumulated on the optical surfaces can be blown off or removed with a soft, clean brush.
5.3.2 Optical surface of the endoscope connection
The endoscope connection is protected against contamination and humidity by an end glass cover. For cleaning and care of this glass plate, proceed in the same way as with the other optical surfaces of the ATMOS following the instructions for cleaning optical surfaces.
On delivery the endoscope connection is protected with a cover against contamination and humidity.
®
i View COLPO. This can be done by
22
5.0 Cleaning and care
5.3.3 Fogging of optical surfaces
To prevent the eyepiece optics from fogging, we recommend using an anti-fogging agent.
Note:
Anti-fogging agents provided by eyecare professionals for use with eyeglass lenses are also suitable for ATMOS optics.
• Please observe the instructions supplied with each anti-fogging agent.
Anti-fogging agents do not only ensure fog-free optics they also clean and protect them against dirt, grease, dust, uff and ngerprints.
5.4 Recommended surface disinfectants
When using disinfectants containing aldehyde and amine at the same object colour changes may occur.
Do not use
- Disinfectants which contain organic or inorganic acids or bases as they could cause corrosion damage.
- Disinfectants which contain chloramines or phenol derivatives as they could cause stress cracks in the material which is used for the housing.
Suitable for
Disinfectant
Green & Clean SK x x
®
Bacillol
Kohrsolin
30 Foam x
®
FF
(Application concentrate)
®
Kohrsolin
extra
(Application concentrate)
®
Mikrobac
forte
(Application concentrate)
®
Mikrozid
SaniCloth
Sensitive Wipes x x
®
Active x x
Microscope Handle Control unit Other mechanical
surface treatment
x x x
x x x
x x x
Optical surfaces
5.5 Hygiene Plan
WHAT HOW WHEN Details
C D S
Housing X X X Manual wiping and disinfection
Lens / Optics X X X Manual wiping and disinfection
Operation parts* X X X Manual wiping and disinfection
Protective covers (disposables)
Hand grips X X X Manual wiping and disinfection
C= Cleaning, D= Disinfection, S= Sterilization
* Operation parts
Knobs to adjust (colour lter, 5 fold magnication changer, operator panel, adjusting screws on the arm)
after each
application
daily
X
weekly
monthly
Single-use product -> not for reprocessing, change after use
23
6.0 Maintenance and Service
6.1 General advice
• Prior to every use a visual inspection of the colposcope and
colposcope connection line must be performed. Damaged
cables must be replaced immediately!
• Maintenance, repairs and period tests may not be carried
out while the product is used on the patient.
• Maintenance, repairs and period tests may only be carried out by persons who have the appropriate technical knowledge and are familiar with the product. To carry out these measures the person must have the necessary test devices and original spare parts. ATMOS recommends: Work should be carried out by an authorized ATMOS service partner. This ensures that repairs and testing are carried out professionally, original spare parts are used and warranty claims remain unaffected.
• At least every 24 months a repeat test of the electrical safety should be performed according to IEC 62353. ATMOS recommends an inspection according to the
manufacturer‘s specications.
• ATMOS neither guarantees for fault-free operation nor for personal injuries and damage to property if
- no original ATMOS parts are being used,
- the advice for use in these operating instructions is not being observed,
- assembly, new settings, alterations, extensions and repairs have not been executed by ATMOS authorised personnel.
• There are no warranty claims whatsoever on defects or malfunctions which arise from the use of third party accessories or consumables.
The instructions and regulations for the respective eld of application should be observed.
6.2 Sending in the device
• Remove and properly dispose of consumables.
• Clean and disinfect the product and accessories according to the operating instructions.
• Place used accessories with the product.
• Fill in the form QD 434 „Delivery complaint / return shipment“ and the respective decontamination certicate.
) This form is enclosed to each delivery and can be found at
www.atmosmed.com.
• The device must be well padded and packed in suitable packaging.
• Place the form QD 434 „Delivery complaint / return shipment“ and the respective decontamination certicate in an envelope.
Afx the envelope to the outside of the package.
• Send the product to ATMOS or to your dealer.
6.3 Exchange of spare parts
1
Brake star grip with copper REF 538.2013.0
2
Brake star grip with POM REF 538.2015.0
Colposcope arm until 2014-12
1
Fuse T 3.15 A/H 250 V REF 008.0751.0
Prior to exchanging the main fuse the system must be disconnected from the power supply. For this it is necessary to unplug the power cord from the power outlet.
Fuse exchange
Colposcope arm since 2014-12
2
2
2
click
24
+
...
+
...
7.0 Troubleshooting
Description Possible causes Remedy
ATMOS
®
i View COLPO cannot
be switched on
®
ATMOS
ATMOS
i View COLPO is hot
®
i View COLPO is
overheated
No function whatsoever
5-fold magnication changer is
defective
Arm follows
Too little or no light at all
Mains cable is not connected
Defective fuse
Connect mains cable
exchange fuse
Please ensure sufcient air ventilation.
Switch off and let cool down for 2-3 hours
Please contact the ATMOS service.
®
ATMOS
i View is switched off Switch on the ATMOS® i View COLPO at the
connection box.
Contact the ATMOS service.
Tie bar is not vertically adjusted
®
The ATMOS
i View COLPO was moved into „parking position“ and thereby the light was switched off.
Adjust tie bar
Pull ATMOS position.
Malfunction of the LED light source
Contact the ATMOS service
Extreme decline in the LED light source.
Light source is dimmed down too low. Increase brightness of the light source.
®
i View COLPO into working
25
8.0 Options and Accessories
Lens
REF
Lens 200 mm 538.1000.0
Lens 250 mm 538.1100.0
Lens 300 mm 538.1200.0
Lens 400 mm 538.1300.0
Lens 200 mm with manual ne focussing (17 mm) 539.1700.0 Lens 250 mm with manual ne focussing (17mm) 539.1800.0 Lens 300 mm with manual ne focussing (17 mm) 539.1900.0 Lens 400 mm with manual ne focussing (17 mm) 539.2000.0
VarioFocus lens (200-350 mm) 538.4000.0
Lens tube
REF
Binocular straight tube 16-times, f = 160 mm 605.2000.0
45° adaption for binocular tubes 606.1106.0
®
Cable (only for the ATMOS
i View 31)
REF
Video cable S-video, 5 m (not with an integrated HD camera) 008.0882.0
Cable HDMI type A/C, L = 5 m (only with an integrated HD camera)
Cable HDMI extension, L = 5 m (only with an integrated HD camera)
USB cable A/B, L = 5 m (only with an integrated SD camera) 008.0910.0
538.1902.0
008.0909.0
26
9.0 Technical data
Voltage 100-240 V~ ± 10 %; 50/60 Hz
Power consumption max. 45 VA
Fuses 2 x T 3.15 A / 250 V
Operation time Continuous operation
Light intensity
F 200 min. 120 kLux
F 250 min. 80 kLux
F 300 min. 55 kLux
F 400 min. 30 kLux
Colour temperature 5000 K ± 500 K
Cooling Fanless / passive
Protective earth conductor resistance
Earth leakage current
Enclosure leakage current
Patient leakage current
Ambient conditions transport / storage
Temperature -10...+50°C
Humidity without condensation 30...95 %
Air pressure 500...1060 hPa
Ambient conditions operation
Temperature +10...+35°C
Humidity without condensation 30...95 %
Air pressure 700...1060 hPa Maximum operational altitude ≤ 3000 m
Contamination level 2
Overvoltage category II
Weight 3.65 kg - 5.6 kg
Period tests Repeat test of the electrical safety every 24 months.
Safety class (EN 60601-1) l
Degree of protection No application part available
Safety type IP X0
Classication according to Appendix IX EC Directive
93/42/EEC
CE marking CE
GMDN code 10960
UMDNS code 10-960
ID No. (REF) 605.0000.0, 606.0000.0
max. 0.1 Ω
max. 0.5 mA
max. 0.1 mA
max. 0.1 mA
Recommended: inspection according to the manufacturer‘s
specications.
Class l (according to regulation no. 12)
Issue of the Technical Data: 20.12.2017
27
10.0 Disposal
• The ATMOS® i View COLPO does not contain any hazardous materials.
• The housing is recyclable.
• Pay attention to a careful separation of the different materials.
• Please observe national disposal regulations (e.g. waste incineration).
Disposal within the EC
The colposcope described above is a high-quality medical product with a long service life. After its life cycle it must be disposed of professionally. According to the EC Directives (WEEE and RoHS) the colposcope may not be disposed of in domestic waste. Please observe existing national laws and rules for disposal of old devices in the respective country.
Disposal within the Federal Republic of Germany
In the Federal Republic of Germany the law for electrical devices (ElektroG) regulates the disposal of electrical devices. It must be assumed that these devices could be contaminated. Therefore, according to the regulations of the EAR (Stiftung Elektro-Altgeräte Register) is this type of device excluded from the ElektroG regulations. In order to guarantee a proper disposal of your old device, please either pass on your old device to your specialised dealer or send it directly to ATMOS MedizinTechnik for a professional disposal.
Before disposal respectively before transport, the colposcope surface must be disinfected.
28
11.0 Notes on EMC
• Medical electrical equipment is subject to special precautions with regard to EMC and must be installed acc. to following EMC notes.
Portable and mobile HF communication facilities can inuence medical electrical equipment.
• The use of other accessories, other converters and cables than stated may lead to an increased emission or a reduced interference immunity of the equipment or system.
11.1 Guidelines and Manufacturer's Declaration - Emissions
The ATMOS® i View COLPO is intended for use in the electromagnetic environment specied below. The customer or user of the ATMOS
®
i View COLPO should ensure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
®
RF Emissions acc.to CISPR 11 Group 1 The ATMOS
function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
HF transmission according to CISPR 11
Harmonic emissions according to IEC 61000-3-2
Voltage uctuations/icker according
to IEC 61000-3-3
Class B
Class A
Corresponds
The ATMOS establishments, including domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
i View COLPO uses RF energy only for its internal
®
i View COLPO is suitable for use in all
11.2 Guidelines and Manufacturer's Declaration - Immunity
The ATMOS® i View COLPO is intended for use in the electromagnetic environment specied below. The customer or user of the ATMOS
®
i View COLPO should ensure that it is used in such an environment.
Immunity Test IEC 60601- Test Level Compliance Level Electromagnetic Environment - Guidance
ESD IEC 61000-4-2 ± 6 kV Contact
± 8 kV Air
Fast electrical transient/ burst IEC 61000-4-4
Surges IEC 61000-4-5 1 kV
Magnetic eld at power
frequency 50/60 Hz acc. to IEC 61000-4-8
± 2 kV Mains
± 1 kV I/Os
Differential
2 kV
Common
3 A/m applicable
± 6 kV Contact
± 8 kV Air
± 2 kV Mains
not applicable
± 1 kV I/Os
1 kV
Differential
2 kV
Common
3 A/m
Floors should be wood, concrete, or ceramics
tile. If oors are synthetic, the relative humidity
should be at least 30 %.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Power frequency magnetic elds should be that
of a typical commercial or hospital environment.
29
11.0 Notes on EMC
Immunity Test IEC 60601- Test Level Compliance Level Electromagnetic Environment - Guidance
Voltage Dips / Dropout IEC 61000-4-11
NOTE U
is the mains alternating current prior to application of the test levels.
T
< 5 % U (> 95 % Dip of the U for 0,5 Cycles
40 % U (60% Dip of the U for 5 Cycles)
70% U (30 % Dip of the U for 25 Cycles)
< 5 % U (> 95 % Dip of the U for 5 s
T
T
T
T
T
T
< 5 % U
)
(> 95 % Dip of the U
T
for 0,5 Cycles
40 % U (60 % Dip of the U for 5 Cycles
70 % U (30 % Dip of the U for 25 Cycles
< 5 % U
)
(>95 % Dip of the U
T
for 5 s
11.3 Guidelines and Manufacturer´s Declaration - Immunity
T
T
T
T
T
)
T
)
T
)
T
Mains power quality should be that of a typical
)
commercial or hospital environment. If the user of the ATMOS
®
i View COLPO requires continued function during interruptions of the energy supply, it is recommended to supply the ATMOS
®
i View COLPO from an uninterruptible
power supply or a battery.
The ATMOS® i View COLPO is intended for use in the electromagnetic environment specied below. The customer or user of the ATMOS
®
i View COLPO should ensure that it is used in such an environment.
Immunity Test IEC 60601- Test
Compliance Level Electromagnetic Environment - Guidance
Level
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
V1 = 3 V
eff
150 kHz to 80 MHz
E1 = 3 V/m
80 MHz to 2.5 GHz
3 V
3 V/m
Portable and mobile communications equipment should be separated from the ATMOS
®
i View COLPO incl. the cables by no less than the distances calculated/listed below.
Recommended distances:
d = [ 3.5 / V1] √P d = [ 3.5 / E1 ] √P from 80 MHz to 800 MHz d = [ 7.0 / E1 ] √P from 800 MHz to 2500 MHz
where „P“ is the max. power in watts (W) and D is the recommended separation distance in meters (m).
Field strengths from xed transmitters, as determined by
an electromagnetic site (a) survey, should be less than the compliance level (b).
Interference may occur in the vicinity of equipment containing following symbol:
30
11.0 Notes on EMC
NOTE 1
With 80 MHz and 800 MHz the higher frequency range applies.
NOTE 2
These guidelines may not be applicable in all cases. The emanation of electromagnetic waves is affected by absorption and
reection of buildings, objects and people.
a
The eld strength of stationary transmitters, such as base stations of cellular phones and mobile terrain radio equipment,
amateur radio transmitters, cbm broadcast and TV stations cannot be predestined exactly. To determine the electromagnetic
environment in regard to stationary transmitters, a study of the location is to be considered. If the measured eld strength at the
location where the ATMOS observed to verify the intended use. If abnormal performance characteristics are noted, additional measures might be necessary, e. g. a changed arrangement or another location for the ATMOS
b
Within the frequency range of 150 kHz to 80 MHz the eld strength should be below 3 V/m.
11.4 Recommended safety distance between portable and mobile RF Communications equipment and the ATMOS® i View COLPO
®
i View COLPO is used exceeds the above compliance level, the ATMOS® i View COLPO is to be
®
i View COLPO.
The ATMOS customer or user of the ATMOS between portable and mobile RF Communications equipment and the ATMOS
®
i View COLPO is intended for use in electromagnetic environment in which radiated disturbances are controlled. The
®
i View COLPO can help prevent electromagnetic interference by maintaining a minimum distance
®
i View COLPO as recommended below, according
to the maximum output power of the communications equipment.
Safety distance, depending on transmit-frequency m
Nominal output of the
transmitter
150 kHz to 80 MHz
d = [ 3.5 / 3] √P
80 MHz to 800 MHz
d = [ 3.5 / 3] √P
800 MHz to 2.5 GHz
d = [ 7.0 / 3] √P
W
0.01 0.12 0.12 0.233
0.1 0.37 0.37 0.74
1 1.16 1.16 2.33
10 3.69 3.69 7.38
100 11.66 11.66 23.33
For transmitters for which the maximum nominal output is not indicated in the above table, the recommended safety distance d in meters (m) can be determined using the equation belonging to the respective column whereas
P is the maximum nominal output of the transmitter in watts (W) acc. to manufacturer´s specication.
NOTE 1
With 80 MHz and 800 MHz the higher frequency range applies.
NOTE 2
These guidelines may not be applicable in all cases. The emanation of electromagnetic waves is affected by
absorption and reection of buildings, objects and people.
31
ATMOS MedizinTechnik GmbH & Co. KG
Ludwig-Kegel-Straße 16
79853 Lenzkirch / Germany
Phone: +49 7653 689-0
atmos@atmosmed.de
www.atmosmed.com
Loading...