Atmos i View 21 COLPO, i View 31 COLPO Operating Instructions Manual

English
ATMOS® i View COLPO
Operating Instructions
GA1GB.420101.0
2018-04 Index 24
Table of contents
1.0 Introduction .........................................................3
1.1 Notes on operating instructions ............................3
1.2 Intended use .........................................................4
1.3 Function ................................................................5
1.4 Explanation of pictures and symbols ....................6
1.5 Scope of supply ................................................... 7
1.6 Transport and storage ...........................................7
2.0 For your safety ....................................................8
3.0 Setting up and starting up .................................9
3.1 Overview .............................................................. 9
3.2 Setting up ............................................................10
3.2.1 Connection to the mains supply ..........................10
3.2.2 Microscope overview ..........................................10
3.2.3 Operating elements at the colposcope ...............11
3.2.4 Rear view of the control device of the ATMOS
3.2.5 Rear view of the control device of the ATMOS
(not with an integrated HD camera) .................... 11
3.2.6 Rear view of the control device of the ATMOS View 31 COLPO with an integrated HD camera .12
3.3 Integration possibilities........................................12
3.4 Starting up...........................................................13
3.5 Operation requirements ......................................13
3.6 Starting up at a glance ........................................14
®
i View 21 COLPO ................................. 11
®
i View 31 COLPO
®
i
4.0 Operation ...........................................................15
4.1 Colposcope suspension ......................................15
4.2 Mechanical arm...................................................15
4.3 Hand grips...........................................................15
4.3.1 T-hand grip ..........................................................15
4.3.2 Lateral double hand grip .....................................15
4.4 Adjusting the interocular distance .......................16
4.5 Adjusting the eye pieces .....................................16
4.6 Exchanging the lenses ........................................17
4.7 Exchanging the lenses with manual ne
focusing...............................................................17
4.8 Exchanging the VarioFocus lens .........................17
4.9 Adjustment of the 5 fold magnication changer ..17
4.10 Focussing............................................................18
4.10.1 Fine focussing .....................................................18
4.11 Exchanging the binocular tube............................18
4.12 Pivoting green lter .............................................18
4.13 Pivoting H.A.S.I. lter ..........................................19
4.14 Shadowless illumination......................................19
4.15 Colposcope zoom and object eld size...............19
4.16 Measuring scale ..................................................19
4.17 Image and video recording .................................20
4.17.1 Adjusting the light mode of the integrated HD
camera ................................................................20
4.18 Endoscope adapter .............................................21
4.19 HD adapter..........................................................21
5.0 Cleaning and care .............................................22
5.1 Basic instructions ................................................22
5.2 Cleaning the mechanical microscope surface ....22
5.3 Cleaning of lenses / eyepieces ...........................22
5.3.1 Cleaning optical surfaces ....................................22
5.3.2 Optical surface of the endoscope port ................22
5.3.3 Fogging of optical surfaces .................................23
5.4 Recommended surface disinfectants ..................23
5.5 Hygiene plan .......................................................23
Further information, accessories, consumables and spare parts are available from:
ATMOS
MedizinTechnik GmbH & Co. KG
Ludwig-Kegel-Straße 16
79853 Lenzkirch
Germany
Phone +49 7653 689-0
Fax: +49 7653 689-190
+49 7653 689-292 (Service Centre)
atmos@atmosmed.de
www.atmosmed.de
6.0 Maintenance and Service .................................24
6.1 Basic instructions ................................................24
6.2 Sending in the device..........................................24
6.3 Exchange of spare parts .....................................24
7.0 Troubleshooting ................................................25
8.0 Accessories and Options .................................26
9.0 Technical data ...................................................27
10.0 Disposal .............................................................28
11.0 Notes on EMC ....................................................29
1.0 Introduction
1.1 Notes on Operating Instructions
These operating instructions contain important notes on how to operate the ATMOS® i View COLPO safely, correctly and effectively. Their reading helps to avoid risks, and also to reduce repair costs and down-times. This increases, amongst other things, the reliability and service-life of the colposcope.
These operating instructions serve not only for new operating personnel to be instructed in its use, but also for use as a reference manual. Reprints (also in extracts) only with permission in written form by ATMOS.
These operating instructions must always be kept available near the colposcope.
Care and safety inspections in conjunction with professional execution provide for operational safety and readiness for use of your ATMOS
®
i View COLPO and are therefore a must besides regular cleaning.
Repair work and safety inspections may be carried out only by expert personnel authorised by ATMOS. By applying only original spare parts you will have the guarantee that operational safety, readiness for work and the value of your ATMOS
• The product ATMOS
®
i View COLPO will be preserved.
®
i View COLPO bears CE marking CE according to the EC Directive of the council
for medical products 93/42/EEC and meets the basic requirements of Appendix I of the directive.
®
• The product ATMOS
i View COLPO complies with all applicable requirements of the Directive 2011/65/
EC restricting the use of certain hazardous substances in electrical and electronic equipment (“RoHS”).
• The declaration of conformity and our general standard terms and conditions can be obtained on our website at www.atmosmed.com.
The quality management system applied at ATMOS has been certied according to international standards EN ISO 13485.
• Prior to start-up please peruse chapter 2.0 „For your safety“, in order to be prepared for any possible dangerous situations.
These operating instructions are valid for the following devices:
®
ATMOS
i View 21 COLPO .........................REF 605.0000.0
Colposcope with an integrated, fanless, high transmission, high performance LED light in the colposcope head
®
ATMOS
i View 31 COLPO .........................REF 606.0000.0
Colposcope with an integrated, fanless, high transmission, high performance LED light in the colposcope head
Please keep this document for future consultation!
1.0 Introduction
1.2 Intended use
Name: ATMOS® i View 21 COLPO
®
ATMOS
Main functions: Visualization of natural body orices (genital area) for examinatory purposes
i View 31 COLPO
Medical indications / application:
Specication of the main
function:
User prole: Examinations and surgical interventions which are carried out using the
Patient groups: No restrictions
Application organ: Natural orices (portio and vulva)
Application time: Short-term, under normal conditions for permanent use over a maximum up to
Application site: Application sites are hospitals, doctor’s ofces, ORs at gynaecologists. The
Standard gynaecological examinations. Visual control of the genital area.
The application organ is illuminated for examination purposes and can be visualized on a monitor if desired by an integrated, fanless high transmission,
high performance LED light source, 5 fold magnication changer, integrated camera module, pivoting green lter and an automatic light control via tilt
sensor; the light output is min. 120 kLux (200 mm)with a colour temperature of
5.500 K ± 10%.
colposcope may only be performed by doctors with appropriate training.
Only qualied personal with a proper hygiene training may prepare the
colposcope for surgical interventions.
Installation and maintenance may only be carried out by service technicians who were trained and authorised by the manufacturer.
30 days
examination with the colposcope may only be executed by medically trained persons.
The colposcope may only be used in closed rooms, on rm ground or mounted
to the ATMOS
In ORs an appropriate protective sleeve must be used.
®
Chair 41 Gyne .
Contraindications: • No application in ophthalmology.
The product is: active
Sterility: The colposcope is not a sterile product.
Single-use product / reprocessing:
The colposcope is intended for multiple use. The device and parts of the accessories are reusable. For information on reprocessing, cleaning and disinfection please see the operating instructions.
1.0 Introduction
1.3 Function
The ATMOS® i View COLPO is a complete colposcope system, consisting of optics and lighting. It produces outstanding pictures for examination purposes with the use of latest LED technology and patent registered optics. The interaction between the integrated
fanless, high transmission, high performance LED, the apochromatic optics and the precisely tting options offer best working
quality.
The ergonomically assorted buttons, two selectable hand grips and the integrated control panel provides the user with highest level of ergonomic comfort and suitability for daily use as well as an outstanding and intuitive handling. Via the control panel the individual options of the ATMOS / stopping of possible video sequences, the operator is capable of manually switching the LED light source on and off despite the activated automatic light control. Due to the variety of options which the ATMOS
position to congure a colposcope to suit his requirements. The following functions can be chosen optionally:
4 lenses with different focal distances (200, 250, 300 and 400 mm) with or without ne focussing or a VarioFocus 200-350 mm
(easy exchange of the lenses due to the respective thread on the colposcope head)
5 fold magnication changer. Exact adjustments due to turning disks on both sides.
• Binocular tubes, simple adaption due to the Zeiss dove tail.
Pivoting colour lter
• Measuring scale
• Shadowless illumination
®
i View COLPO can be activated. Besides the triggering of the camera (freeze frame) and starting
®
i View COLPO has to offer, the user is in a
Due to the LED light source and the integrable camera solution (SD integrated respectively as HD or endoscope adapter for the connection of an external camera) the ATMOS
In connection with the mechanical support arm and the numerous connection possibilities to units and stands the ATMOS
®
i View COLPO is a guarantor for best image quality.
®
i View
COLPO offers countless system possibilities, which can be individually adapted to suit the users environment!
®
These operating instructions describe all functions with a maximum conguration of the ATMOS
i View 31 COLPO.
1.0 Introduction
1.4 Explanation of pictures and symbols
Short cuts / symbols contained in these operating instructions
Follow the arrows
General information
Move, plug... in this direction
Please press where dot indicates
Please read, important information
Graphic symbols contained in these operating instructions
Warning, special diligent notice
Numeration
Check
!
Symbols ATMOS
SN Serial number
®
i View
Manufacturing date Manufacturer
Observe operating instructions! Observe operating instructions!
REF
click
) Important information
Order number
Turn, shift... in this direction
Replace
Engage, check
correct t
Weight adjustment for the carrier arm Professional disposal
Alternating current Fuse
2
Control panel buttons
Single use product - not for reuse. Exchange after use.
Do not look directly into the light source of the ATMOS
Light on / off (independent of automatic light control)
Video recording (start / stop)
®
i View COLPO.
This product complies with the relevant requirements of EU Directives
Sitting is prohibited.
Freeze frame
With an integrated HD camera: Adjusting the light mode of the camera
1.0 Introduction
Foot Switch
Out 1
Out 2 Out 3
Video Out
Foot Switch
Only ATMOS® i View 21 COLPO
Output of the power supply for the electron­ics in the colposcope
Fuse
®
Only ATMOS
i View 31 COLPO
Fuse Potential equalisation acc. to IEC 604175021
Colposcope Foot switch
Record function Freeze
Potential equalisation acc. to IEC 604175021
not in use
USB port
S-video-output (not with an integrated HD camera)
Video In Intern 1
Out 1
Video Out
Out 2 Out 3
Output signals of the tilt sensor in the carrier arm system
Video In Extern 2
Input video signal internal / external (Only with an integrated HD camera)
Output video signal (Only with an integrated HD camera)
1.5 Scope of supply
• Prior to dispatch, the ATMOS® i View COLPO was subjected to an extensive functional test and was carefully packed. Nevertheless, please compare the contents of the shipment on completeness immediately upon receipt (see delivery note).
Colposcope
Operating Instructions
Colposcope arm
Protective cover
1.6 Transport and storage
• After the transport of the ATMOS® i View in
temperatures below 0°C or prior to rst start up
it should be kept at room temperature for at least six hours. If the ATMOS acclimatized it may not be used as damages to the electronic components may be the result.
Only transport the device in a shipping carton, which is
padded and offers sufcient protection.
If damage occurs during transport:
• Document and report the transport damage.
• Send the device to ATMOS (Chapter „6.2 Sending in the device“ on page 24).
®
i View COLPO is not
Ambient conditions:
• Transport / storage:
-10...+50 °C;
30...95 % humidity without condensing air pressure 500...1060 hPa
• Operation: +10...+35 °C;
30...95 % humidity without condensing air pressure 700...1060 hPa
2.0 For your safety
!
For your safety
• To safely disconnect the unit from the mains, the power cord must be removed from the IEC connector of the control device!
®
• The ATMOS with IEC 60601-1-1/EN60601-1 and it is a device with protection class I. In order to avoid the risk of electrical shock, this unit may only be connected to a mains supply with properly installed earth conductor.
• Power cables, accessories and access cables need to be checked for defects prior to setting up the ATMOS COLPO. Damaged cables must be replaced immediately.
• The ATMOS
qualied personnel.
• The ATMOS in an explosion-hazardous environment. Explosion-
hazardous areas may be caused by the use of ammable
anaesthetics, skin cleansing products and skin disinfectants.
If uids penetrated the ATMOS be sent in and may only be used after the check up of an ATMOS authorised person.
• After the transport of the ATMOS
below 0°C or prior to rst start up it should be kept at room
temperature for at least six hours. If the ATMOS COLPO is not acclimatised it may not be used.
• Do not plug in electric connections (plug, socket) under the use of force. If this is not possible check whether the
plug ts the socket. If you should ascertain a defect in the
connection you should have it repaired by our service.
• Never look straight into the sun with lenses or eye lenses.
• Always make sure that you do not blind patients with the light source! Watch out that patients do not look directly into the light source! Never look directly into the light source! > Damage to the eyes due to the strong glare.
• Please pay attention to the period tests in chapter 6 „Service and maintenance“ on page 24.
• Prior to every use the colposcope suspension (all joints included) need to be checked for safe connections.
• Take care that the patient does not touch the device or have any contact with it.
• Please observe the EMC Directives. Failure to follow this guideline can result in a hazard.
• Make sure that the unit is positioned so that all the controls and the on/off switch are always accessible.
• Dispose of wrappings accordingly.
• Before connecting the ATMOS be checked whether the requested mains voltage of the ATMOS mains power supply.
• Only proper and undamaged plugs and extension cables may be used.
i View COLPO is a device designed in line
®
i View
®
i View COLPO may only be operated by
®
i View COLPO is not designed to be used
®
i View COLPO it needs to
®
i View in temperatures
®
i View
®
i View COLPO it needs to
®
i View COLPO matches the mains voltage of the
• To disconnect the ATMOS
mains supply, rst remove the plug from the wall outlet.
Disconnect the connection line on the ATMOS
®
i View COLPO from the
®
i View COLPO afterwards only. Never touch plug or line with wet hands.
• Please observe the ambient conditions stated in the technical data (chapter 9.0).
®
• The ATMOS
i View COLPO meets the immunity to interference requirements of IEC 60601-1-2 / EN 60601­1-2 „Electromagnetic Compatibility – Medical Electrical Devices“.
• ATMOS is not liable for personal injury and damage to property if
- no original ATMOS parts are being used,
- the advice for use in these operating instructions is not being observed,
- assembly, new settings, alterations, extensions and repairs have been carried out by personnel not authorised by ATMOS.
• Unplug the device immediately if you observe fumes, sparks or weird noises.
• With every light source a warming of tissue due to absorption may occur. Please make sure to reduce duration of use to a minimum, to switch off the light source when not in use and to check heat development if necessary.
®
• The ATMOS
i View COLPO may be operated only in rooms used for medical purposes, but not in areas subject to explosion hazards and in oxygen rich environments.
• Take into consideration, when setting up the colposcope, that the elastic force of the arm – without colposcope head – is exceedingly strong. Operate the break of the height adjustment carefully.
• Risk of injury! Take care not to roll the mobile stand over your feet when moving the stand.
• Please note that only PCs and monitors with IEC 606010­1/EN 60601-1 approval may be connected to the video outlet of the ATMOS
®
i View COLPO supply module!
• During operation, the user is obliged to regularly check the microscope for proper function. In the unlikely event of failure of the microscope, the user must take precautions to continue the treatment of the patient with suitable methods.
3.0 Setting up and starting up
3.1 Overview
ATMOS® i View 21 COLPO ATMOS® i View 31 COLPO
Examination colposcope with an inte-
Description
Integrated high performance white light LED
Automatic light control
Optimised stereo effect
Measuring scale Optional Optional
Integrated operating panel
Colour lter Optional
Integrated camera -
HD adapter for an external camera
Endoscope adapter - Optional
Mains voltage 100–240 V 100–240 V
grated, fanless, high transmission, high performance LED light in the colposcope head
Optional Optional
- Optional
Examination colposcope with an inte­grated, fanless, high transmission, high performance LED light in the colposcope head
Optional green lter or Optional H.A.S.I. lter
Optional SD camera or
Optional HD camera
min. 120 kLux (200 mm)
Light output
Operating life of the LED
Colour temperature 5.500 K ± 10 % 5.500 K ± 10 %
Scope of delivery
min. 80 kLux (250mm)
min. 55 kLux (300 mm)
min. 30 kLux (400 mm)
50 000 hours 50 000 hours
Dust cover, operating instructions
min. 120 kLux (200 mm)
min. 80 kLux (250mm)
min. 55 kLux (300 mm)
min. 30 kLux (400 mm)
Dust cover, operating instructions
3.0 Setting up and starting up
3.2 Assembly
Please make sure that the static conditions stated by ATMOS MedizinTechnik are met (for details see the separately enclosed
!
document „Static requirements for installing the ATMOS i View“). The fullment of these requirements must be conrmed by an
authorized expert.
Mains voltage and fuse: Mains voltage: 100-240 V; 50/ 60 Hz, fuse: 2 x T 3.15 A
Please note that only PCs and monitors with IEC 606010-1/EN 60601-1 approval may be connected to the video outlet of the ATMOS
3.2.1 Connection to the mains supply
Potential equalisation:
The ATMOS potential equalisation rail in the room if need be. Hereby user/patient safety can be increased especially in the case of a defective earth conductor. For connecting the device's potential equalisation plug with the potential equalisation rail of the room, use the potential equalisation cord with REF 530.0030.0.
3.2.2 Microscope overview
®
i View COLPO supply module!
®
i View COLPO supply module has a rear connection for potential equalisation which can be connected to the
Wide eld eye
piece lenses with binocular tube
Rotating disk 5
fold magnication
changer
Lens
Lateral double hand grip (optional)
T-hand grip (optional)
Pivoting colour lter
(optional)
Measuring scale (optional)
Brightness control
HD adapter for SONY digital camera (optional)
Control panel (optional)
10