These operating instructions contain important notes on how to operate the ATMOS® Chair 51 D safely,
correctly and effectively. Their reading helps to avoid risks, and also to reduce repair costs and downtimes. This increases, amongst other things, the reliability and service-life of the chair.
These operating instructions serve not only for new operating personnel to be instructed in its use, but
also for use as a reference manual. Reprints (also in extracts) only with permission in written form by
ATMOS.
These operating instructions must always be kept available near the device.
Care and safety inspections in conjunction with professional execution provide for operational safety
and readiness for use of your ATMOS
Repair work and safety inspections may be carried out only by expert personnel authorised by ATMOS.
By applying only original spare parts you will have the guarantee that operational safety, readiness for
work and the value of your ATMOS
• The product ATMOS
®
Chair 51 D bears CE marking CE according to the EC Directive of the council
®
Chair 51 D will be preserved.
for medical products 93/42/EEC and meets the basic requirements of Appendix I of the directive.
®
• The product ATMOS
Chair 51 D complies with all applicable requirements of the Directive 2011/65/
EC restricting the use of certain hazardous substances in electrical and electronic equipment
(“RoHS”).
• The declaration of conformity and our general standard terms and conditions can be obtained on our
website at www.atmosmed.com.
• The quality management system applied at ATMOS has been certied according to international
standards EN ISO 13485.
• Prior to start-up please peruse chapter 2.0 „For your safety“, in order to be prepared for any possible
dangerous situations.
1.2 Intended use
Name: ATMOS® Chair 51 D
Main function: This doctor's chair enables the optimum
positioning of the doctor with regard to height and access to
the patient
Medical indications / application: For positioning the
doctor at the patient during his treatment
Specication of the main function:
• Height adjustment of the seat by gas compression spring
from approx. 56 cm up to approx. 74 cm
• 360° rotation
• Lockable in every heights position
• Backrest adjustable in height and inclination, rotatable
• Mobile on 5 castors
Application organ: Positioning of the patient
Application time: Temporarily (up to 60 minutes)
Application site: In clinics and practices for ENT doctors
and phoniatricians. The application of the doctor’s chair
must be executed by medically trained persons only.
®
Chair 51 D and are therefore a must besides regular cleaning.
Contraindications:
• Maximum weight load: 140 kg
• May only be used as seating on horizontal surfaces
The product is: not active
Sterility: Not necessary
Single-use product / reprocessing: No single use product
1.3 Function
• With the doctor's chair ATMOS® Chair 51 D the examiner
can take optimum position to the patient.
®
• The ATMOS
height adjustment. The seat height can be varied
between 56 and 74 cm.
• The ATMOS
equipped with a backrest which is adjustable in height
and inclination. In addition it is rotatable and can
therefore be used as elbow rest.
Chair 51 D is equipped with a mechanical
®
Chair 51 D is mobile on 5 castors and
3
1.0 Introduction
1.4 Explanation of pictures and symbols
Short cuts / symbols contained in these operating instructions
Follow the arrows
Please press where dot
indicates
Please read,
important information
Graphic symbols contained in these operating instructions
!
Warning, special diligent notice
Symbols of ATMOS® Chair 51 D
General information
■
Numeration
●
Sub-numeration
→
Check
Important information
Move, plug... in this
direction
Turn, shift ... in this
direction
Replace
click
Observe operating instructions.
Engage, check
correct t
Serial number
SN
Order number
REF
Manufacturing date
This product complies with the relevant
requirements of EU Directives
4
1.0 Introduction
1.5 Scope of supply
• Prior to dispatch, the ATMOS® Chair 51 D was subjected to an extensive functional test and was carefully packed.
Nevertheless, please compare the contents of the shipment on completeness immediately upon receipt (see delivery note).
Operating
Chair
1.6 Transport and storage
Instructions
• The transport of the chair may be effected only in a
dispatch carton upholstered and offering sufcient
protection.
• Please note down and immediately report any damages
which occurred during shipping. Please make use of the
attached QD 434 delivery complaint / return shipment
form when complaining or sending back.
Ambient conditions:
Transport / storage:
• -10...+50°C;
• 30...95 % air humidity without condensation
• at an air pressure of 700...1060 hPa
5
2.0 For your safety
!
For your safety
• Warranty period for this unit: 2 years. This period is
unaffected by any repair or maintenance carried out
under the terms of warranty.
• There are no warranty claims whatsoever on defects
which arise from the use of third party accessories or
consumables.
• ATMOS is not liable for personal injury and damage to
property if
- no original ATMOS parts are being used,
- the advice for use in these operating instructions is
not being observed,
- assembly, new settings, alterations, extensions
and repairs have been carried out by personnel not
authorised by ATMOS.
®
• Before putting the ATMOS
visually check it for any signs of damage. Defect chairs
must be replaced immediately. Prior to use please check
proper function of the chair.
• Dispose of wrappings accordingly.
• Please observe the ambient conditions stated in the
technical data (chapter 7.0).
Chair 51 D into operation,
6
3.0 Setting up and starting up
5
3.1 Setting up the ATMOS® Chair 51 D
• Insert the gas compression spring including cover (2)
1
into the star base (
3
4
• Attach the seating surface (
2
spring (
).
• Fix the backrest cushion (
2
).
3
) to the gas compression
5
) to the backrest (4).
1
Fig. 1.
3.2 Height and inclination adjustment of
backrest
• Height and inclination of the backrest can be adjusted
with the screw (
1
1
, Fig 2) using a hexagonal wrench.
Fig. 2.
1
Screw for the backrest height adjustment
7
3.0 Setting up and starting up
3.3 Adjusting the position of the
backrest
Fig. 3. Adjusting the backrest position
1
• The position of the backrest (, Fig. 3) can be adjusted
in height and inclination. In addition it is rotatable and can
therefore be used as elbow rest.
Fig. 4. Adjusting the backrest position
Upwards and downwards
Fig. 5. Adjusting the backrest position
Forward and back
8
3.0 Setting up and starting up
2
Fig. 6.
right
1
centre
left
3.4 Mounting the backrest
• Loosen the screws ( and , Fig 6). Attach the backrest
to the relevant drillings and rex the screws. The backrest
is rotatable and can therefore be used as elbow rest
as well. There are 3 mounting positions possible: right,
centre, left.
3.5 Using the backrest resp. elbow rest
Fig. 7.
Fig. 8.
3.6 Adjusting the seat height
• By pulling the lever (, Fig 9) upwards, the height
adjustment is released (for upward and downward
movement). The gas operated compression spring
provides for the upward movement. The height can be
1
reduced by simultaneously pressing on the seat. When
releasing the lever the chair remains arrested in the
desired position.
Fig. 9.
9
4.0 Cleaning and care
4.1 General information on cleaning and disinfection
Prior to cleaning
®
Medical chairs such as the ATMOS
Chair 51 D must be failsafe at all times.
) The described measures relating to cleaning and
disinfection resp. sterilisation do not replace the relevant
instructions which must be adhered to prior to operation!
• For disinfection, you may use all surface and upholstery
disinfectants listed in chapter 4.3 and 4.4 “Recommended
disinfectants”.
) Always observe the concentration specications and
instructions by the respective manufacturer!
Do notuse
• Disinfectants which contain organic or inorganic acids or
bases as they could cause corrosion damage.
• Disinfectants containing chloramides, phenol derivatives
or anionic tensides, as these may cause stress cracks in
the plastic material used.
4.2 Cleaning and disinfection of the device surface and upholstery
• The surfaces of the ATMOS® Chair 51 D are resistant against all the recommended surface disinfectants stated in chapter 4.3
and 4.4. Nevertheless after a longer period of use discolouration could occur.
• Substances such as blood must be removed immediately to prevent stains.
• The device itself can be wiped off with a moist (not wet) cloth.
Do not use aggressive or abrasive cleansing agents. For the upholstery standard dry foam may be used.
!
Treatment with a commonly used care product for articial leather is recommended once a week in order to keep the upholstery
soft and smooth.
4.3 Recommended surface disinfectants
• The surfaces of the ATMOS® Chair 51 D can be cleaned / wiped with disinfectants containing the following active ingredients:
QAV (quaternary ammonium compounds)
4.4 Recommended disinfectants for the upholstery
DisinfectantIngredients(in 100 g) Manufacturer
®
ATMOS
(Application solution)
Green & Clean SK
Alkyl dimethyl benzyl ammonium chloride< 1 g Metasys, Rum (Austria)
Alkyl dimethyl ethyl benzyl ammonium chloride< 1 g
Di alkyl dimethyl ammonium chloride< 1 g
10
5.0 Maintenance
Maintenance, repairs and period tests may only be
carried out by persons who have the appropriate technical
knowledge and are familiar with the product. To carry out
these measures the person must have the necessary test
devices and original spare parts.
ATMOS recommends: Work should be carried out by an
authorized ATMOS service partner. This ensures that repairs
and testing are carried out professionally, original spare
parts are used and warranty claims remain unaffected.
Please comply with the country-specic guidelines regarding
regular testing.
There are no warranty claims whatsoever on defects
which arise from the use of third party accessories or
consumables.
6.0 Technical data
5.1 Sending in the device
• Remove and properly dispose of consumables.
• Clean and disinfect the product and accessories
according to the operating instructions.
• Place used accessories with the product.
• Fill in the form QD 434 „Delivery complaint / return
shipment“ and the respective decontamination certicate.
) This form is enclosed with each delivery and can be
found at www.atmosmed.com .
• The device must be well padded and packed in suitable
packaging.
• Place the form QD 434 „Delivery complaint / return
shipment“ and the respective decontamination certicate in an envelope.
• Afx the envelope to the outside of the package.
• Send the product to ATMOS or to your dealer.
Height adjustmentManual, gas compression spring
Seat diameter42 cm
Seat height55 - 74 cm
Maximum weight load140 kg
Ambient conditions transport / storage
• Temperature-10...+50° C
• Humidity without condensation30...95 %
• Air pressure500...1060 hPa
Dimensions
• Star baseØ 60 cm
Weight14 kg
Classication according to Appendix IX EC
Directive 93/42/EEC
CE markingCE
UMDNS code10-794
Issue of technical data: 12.03.2018
Class 1
7.0 Checking
A regular and thoroughly cleaning and disinfection of the ATMOS® Chair 51 D according to these operating instructions is taken
for granted.
11