1.0 Introduction
These operating instructions contain important notes on how to operate the ATMOS Chair 41 Gyne
safely, correctly and eectively. Their reading helps to avoid risks, and also to reduce repair costs
and down-time. That increases, amongst other things, the reliability and service-life of the device.
These operating instructions serve not only for new operating personnel to be instructed in its use,
but also for use as a reference manual. Full or partial reproduction is subject to prior written
approval from ATMOS.
These operating instructions must always be kept available near the device.
Care and safety inspections in conjunction with professional execution provide for operational safety
and readiness for use of your ATMOS Chair 41 Gyne and are therefore a must besides regular
cleaning.
Repair work and safety inspections may be carried out only by expert personnel authorised by
ATMOS. By applying only original spare parts you will have the guarantee that operational safety,
readiness for work and the value of your ATMOS Chair 41 Gyne will be preserved.
●The product ATMOS Chair 41 Gyne bears CE marking CE according to the EC
Directive of the council for medical products 93/42/EEC and meets the basic requirements of
annex I of this directive.
●The product ATMOS Chair 41 Gyne complies with all applicable requirements of the Directive
2011/65/EC restricting the use of certain hazardous substances in electrical and electronic
equipment (“RoHS”).
●The declaration of conformity and our general standard terms and conditions can be obtained
on our website at www.atmosmed.com.
●The quality management system applied at ATMOS has been certied according to international
standards EN ISO 9001 and EN ISO 13485.
●Prior to start-up please peruse chapter 1.2 „For your safety“, in order to be prepared for any
possible dangerous situations.
Information provided in these operating instructions
These operating instructions describe the ATMOS Chair 41 Gyne purposes and provide information on how it works, its functions, maintenance,
trouble shooting and relative solutions.
Marks and symbols
The following marks and symbols are being used in this handbook:
Note:This points out, that the text contains some important information.
Attention:This message may appear before certain procedures. If it is ignored, the equipment could be damaged.
Warning:This message may appear before certain procedures. If it is ignored, the operator could be injured and
the equipment damaged.
General information
The ATMOS Chair 41 Gyne does not electromagnetically or otherwise aect other equipment nor is the chair itself aected by these phenomena,
caused by other equipment, as it is manufactured in compliance with the Directive, and meets the requirements of the harmonised standard
EN 60601-1:1996-03.
4
Page 5
1.0 Introduction
1.1 Intended use
Name: ATMOS Chair 41 Gyne
Main functions: Examinations and therapy in the gynaecological sector
Med. indications/ application: Direct contact with the skin exclusively with the prescribed / recommended paper
SpecicationofthePositioning and bedding of the patient for gynaecological
main function:examination and treatment like pap smear test,colposcopy, diagnostic hysteroscopy, conus biopsy, intravaginal and abdominal echography
Application organ: Female sexual / reproductive organs
Application time: Temporary
Application site: In gynaecologist clinics and practices
Contraindications: Not to be used outside medical areas: Device must not be
used in areas with a potential explosive hazard in
rooms specied for medical treatment.
The product is: active
Sterility: Not necessary
Single use product / No single use product
reprocessing:
1.2 For your safety
- Max. patient weight: 150 kg
- Max. lifting capacity: 200 kg
- Prior to cleaning, the mains switch must be switched o.
- The ATMOS Chair 41 Gyne may be operated only in rooms
used for medical purposes, but not in areas subject to
explosion hazards and in oxygen rich environments.
5
Page 6
1.0 Introduction
This chapter describes the features of the gynaecological chair ATMOS Chair 41 Gyne. The main equipment components
and relative accessories are also illustrated.
Warning:
This equipment has been conceived for gynaecology purposes only. Neither the chair nor its components
It must not be used for purposes other than gynaecological ones (especially not in operating theatres).
Warning:
When moving the chair make sure that the patient has her hands at the chair and that third persons and things are
in adequate distance from the chair.
The ATMOS Chair 41 Gyne has the following features:
• The chair has a driven backrest adjustment for the Trendelenburg position and a lift mechanism.
1.3 MAIN CHARACTERISTICS
The ATMOS Chair 41 Gyne oers the following features:
• perfectly symmetrical chair;
• 3 user speciable position memory systems;
• auto return system to “zero” position;
• control pedals reextractable from the base;
• Trendelenburg position;
• liquid collection basin, plastic or stainless steel (optional);
• leg section extension built in the chair (optional);
• leg rests, adjustable (optional);
• seamless chair upholstery;
• handles; (optional)
• magnetic head-rest cushion (optional);
• paper roll holder integrated in the chair back;
• removable foot rest for the doctor (optional);
• step for helping the patient get on the chair;
• double electrical socket in the base (230 V) Attention: max. load is 460 W! (optional)
• low voltage controls (5 V).
Warning:
During the Trendelenburg movement the patient must grasp the handles and have a support for the shoulders.
Warning:
During an anaesthesia, an assistant must attend for safety reasons.
Warning:
When several devices are connected to one socket, the allowed load and leakage currents must be observed.
Note:
In order to ensure the proper functioning of the position memory system, it is highly recommended to reset the chair
to “zero” position by pressing key “0”, after every medical visit.
6
Page 7
1.0 Introduction
1.4 Type label
The type label is situated on the rear part of the chair and indicates the model, the serial number, the REF and other
information (see Fig. 1.1).
1.5 Technical information
Fig. 1.1
Fig. 1.2
7
Page 8
2.0 Scope of supply
For removing the packing, all screws must be loosened and the strap must be removed.
The following components must be supplied inside the packing of the bottom part:
• 1 bottom part of the ATMOS Chair 41 Gyne
• 1 operating instructions
• 1 bottle of detergent
• 1 power supply cable
• 1 guarantee certicate
The following must be supplied inside the packing of the upper part:
• 1 upper part of the ATMOS Chair 41 Gyne
8
Page 9
3.0 Setting up and starting up
Loosen the two screws that secure the chair to the wooden pallet to free it. A second person is required to help to move the chair
from the pallet. Use the transport handles to move the chair. Use the lateral handles at the seat to position the upper part onto
the lower part.
1) Place the upper part of the chair on top of the lower part, matching the xing holes of the seat support with the xing holes
of the counter seat, as shown in the gure.
2) Fix the upper part and the lower part by means of the screws provided.
3) Connect the orange 4 poles crimping terminal of the seat support (Fig. 3.1 A) with that of the counter seat (Fig. 3.1 B) for
the mains power supply of the backrest and the respective micro switch.
Fig 3.1
9
Page 10
3.0 Setting up and starting up
Once the two parts of the chair have been assembled and placed in the desired position, securing it, if necessary to the
oor, exploiting the holes in the base (see Fig. 3.2), proceed as follows:
1. Remove the handles used to move the chair.
2. Make sure that the place in which the chair is installed is provided with a suitable earth system.
3. Make sure that the voltage of the electrical power supply system of the place in which the chair is installed corresponds
to that indicated on the type label.
4. Make sure that the fuses on the back of the chair are tted rmly as they could have loosened during transport.
5. Plug the power supply cable into the socket in the chair base and then into the mains socket.
6. Turn the chair on by pressing the main ON/OFF switch.
The chair is now ready to be used.
If it should fail to work, see chapter 7.0 Trouble shooting.
Thanks to the integrated control electronics, the chair has many functions, which are easy to use. All the chair adjustments and
functions are co-ordinated, processed and stored by an integrated microprocessor. The ATMOS Chair 41 Gyne can easily be
controlled when using the foot pedal, with it the gynaecologist has his hands free for the patient.
Warning:
Make sure that the chair has been correctly installed before putting the information provided in this section of the
operating instructions into practice (see chapter 3).
4.1 Controls
The chair has the following control elements:
Left-hand foot pedal (Fig. 4.1):
• Up arrow “↑” - upward movement of chair
• Down arrow “↓” - downward movement of chair
• Right-hand arrow “→” - Trendelenburg position -10˚
• Left-hand arrow “←” - Trendelenburg position +25˚
RH FOOT PEDAL CONTROL (Fig. 4.1)
• Up arrow “↑” - upward movement of chair
• Down arrow “↓” - downward movement of chair
• Right-hand arrow “→” - downward tilt of chair back
• Left-hand arrow “←” - upward tilt of chair back
• “0” - Automatic return to “zero” position.
• “I” - Program 1 (position memory)
• “II” - Program 2 (position memory)
• “III” - Program 3 (position memory)
Keep pressed the key for activating the relevant movement. A short key press activates the position memory.
The electronics requires a short delay of approx. 0.5 sec to dierentiate between "short key press" and "keep depressed".
Therefore, not eort but patience is required.
Left Right
Fig. 4.1
11
Page 12
4.0 Operation
Warning:
When moving the chair make sure that the patient has her hands at the chair and that third persons and things are
in adequate distance from the chair.
4.2 Programming
Proceed as follows to program (keys “I”,”II” and “III” of the right-hand foot pedal):
1. Bring the chair into basic position by pressing the "0" key shortly.
2. Bring the chair into the requested position.
3. Keep depressed the programming switch M, which is on the “cover”, on the left of the lifting mechanism and lightly
press key “I”, “II” or ”III” depending on the requested storage place.
4. An acoustic signal (beep) conrms successful saving.
4.3 Leg rest adjustment
Proceed as follows to move the leg rest:
Hold the patient’s leg with one hand, loosen handle 1 and
adjust the height of the leg rest. When the required position is
reached, tighten handle 1 again.
To adjust the positionable leg rest 2, loosen handle 3, move the
leg rest to the required position and retighten handle 3.
4.4 Leg rest (REF 503.0552.0)
For extracting the leg rest:
Catch the leg rest in centre between seat and secretion bowl.
Now pull out the leg rest until it is on the same level like the
seat.
2
1
3
Fig 4.2
When pulling the leg rest in:
ATTENTION! Do not touch the guiding rail of the leg rest
(see Fig. 4.3). There is a risk of injury!
Grasp the centre of the leg support. With the other hand
grasp the metal bar under the leg support and hold it tight.
Now slide the support back until it declines. As soon as the
leg support declines, take your hand from the metal bar and
slide the support backwards until it snaps in.
12
Fig 4.3
Page 13
5.0 Cleaning and care
In case the chair is left inoperative for a longer period, it is advisable to cover it with a sheet, to switch it o and disconnect it
from the mains supply.
5.1 Cleaning
For a longer service life, cleaning with a wet cloth is recommended at regular intervals.
Upholstery
For removing stains on the upholstery, only the supplied detergent may be used. For cleaning a soft cloth must be used to
prevent supercial cracks and to ensure the exibility and softness of the material.
Substances such as blood need to be removed immediately to prevent stains on the cover material of the upholstery.
Note:
Please reorder the detergent once it is used up.
Parts made of polyurethane
The parts made of polyurethane must be cleaned using a cloth soaked in soapy water.
Painted parts
The painted parts must be cleaned using a cloth soaked in soapy water.
Attention:
Please do not use substances, detergents and strong abrasives for cleaning, and even not for removing tenacious
stains.
Furthermore please take care when using colour-intensive substances, like, for example, Betaisodona, as this may
lead to discolourations and unattractive stains.
5.2 Paper roll replacement
Proceed as follows to replace the paper roll (see Fig. 5.1):
1. Bring the chair back into horizontal position.
2. Remove the chair back upholstery from the chair back.
3. Block the upholstery using the appropriate locking device integrated in the chair back.
4. Extract the tube support from the chair back.
5. Insert the tube support in the new paper roll.
6. Position the roll in the chair back, running the end of paper upwards.
7. Close the upholstery, making sure that the paper is fed out from the chair back.
Warning:
The manufacturer cannot be hold liable for any damages that may occur from the non-observance of the regulations
stated in these operating instructions.
The function of the roll paper feed can be guaranteed only when using the original paper roll (REF 503.0557.0).
13
Page 14
5.0 Cleaning and care
Fig 5.1
5.3 Recommended surface disinfectants
The surfaces of the Chair 41 Gyne can be cleaned with disinfectants with the following active ingredients:
• quaternary ammoniumchlorides
5.3.1 Recommended disinfectant for cushion
Disinfectant Contents (in 100 g) Manufacturer
Green & Clean SK Alkyldimethylbenzylammoniumchloride < 1 g Metasys, Rum (Ausrtia)
Maintenance, repairs and period tests may only be carried out by persons who have the appropriate technical knowledge
and are familiar with the product. To carry out these measures the person must have the necessary test devices and
original spare parts.
ATMOS recommends: work should be carried out by an authorized ATMOS service partner. This ensures that repairs and
testing are carried out professionally, original spare parts are used and warranty claims remain unaected.
At least every 24 months a repeat test of the electrical safety should be performed according to IEC 62353.
ATMOS recommends an inspection according to the manufacturer's specications.
6.1 Exchange of upholstery
Seat upholstery
This upholstery is xed with nylon snap fasteners and can therefore easily be removed in pulling it o. If required, adjust
the lower and the upper snap fastener for a perfect t of the upholstery.
Back upholstery
This upholstery is xed with magnets and nylon snap fasteners. For removing the upholstery, loosen the xation screws
of the upholstery support, which is xed at the hinges near the seat.
15
Page 16
7.0 Trouble-shooting
ProblemCheckTrouble shooting
The chair
cannot be
moved
ProblemCheckTrouble shooting
Foot control
pedal does
not work
properly
Mains cable is connected? Connect the mains cable to the mains supply.
Main switch is on? Switch on the mains switch.
Is the green control lamp illuminated? Check the fuses, 2 pieces 3,15 A,
at the power input.
Is the green LED on the board "PWR" Check the connector on the board. Fuse on the board
switched on and is a beep audible T 315 mA and T 12,5 A.
during switch on? Check the fuse F 12,5 A between transformer and board.
Voltage at the plug CN9 Fuse between transformer and board F 12,5 A or
(24–28 V AC on the board exchange transformer.
near the small transformer)
Observe operating instructions. Press the control pedal straight and not skewed.
In order not to twist the pedal, observe the delay regarding
the position memory.
Despite perfect operation, the foot control pedal Exchange the board in the foot pedal and check perfect pass
does not always work properly. of the cable, the way the cover ts perfectly.
ProblemCheckTrouble shooting
Chair only
moves
upwards and
then it stops
ProblemCheckTrouble shooting
Chair can
be moved
but position
storage is
not possible
ProblemCheckTrouble shooting
Trendelen-
burg motor
does not
work
ProblemCheckTrouble shooting
Deviation
from the
chair
position
stored
Does the relay on the board work properly? If not, exchange the board.
Measure the voltage at the motor (24 – 28 V AC). If voltage is connected and the motor does not work,
exchange the motor.
Check the connector on the board. Rectify the fault (exchange the connector or the board).
Check the button for position memory. If required, exchange the switch.
Does the relay on the board work properly? If not, exchange the board.
Measure the voltage at the motor (24 – 28 V AC). If voltage is connected and the motor does not work,
exchange the motor.
Correct adjustment of requested position? Bring the chair into "0" position, then adjust the requested
position and store it, as described in the operating
instructions.
ProblemCheckTrouble shooting
Motor
makes
noises
Loud cracking and sometimes the Probably the gear mechanism is defective, therefore
motor does not work. exchange the motor.
16
Page 17
8.0 Accessories, further options and consumables
Accessories
Headrest magnetic pillow
Colour: Pastel blue
Colour: Saffron yellow
Colour: Salmon
Colour: Leaf green
Colour: Light grey
Colour: Silk grey
Colour: White
Colour: Sapphire blue
Leg support all-cushion 1 pairr
Colour: Pastel blue
Colour: Saffron yellow
Colour: Salmon
Colour: Leaf green
Colour: Light grey
Colour: Silk grey
Colour: White
Colour: Sapphire blue
Leg support stainless steel
1 pair, with cushion, foot support plate stainless steel
Colour: Pastel blue
Colour: Saffron yellow
Colour: Salmon
Colour: Leaf green
Colour: Light grey
Colour: Silk grey
Colour: White
Colour: Sapphire blue
Leg holders „Goepel“
1 pair, without cushion, simultaneous adjustment of
height, inclination and distance
Cushion for leg holders „Goepel“
Colour: Pastel blue
Colour: Saffron yellow
Colour: Salmon
Colour: Leaf green
Colour: Light grey
Colour: Silk grey
Colour: White
Colour: Sapphire blue
Handles, 1 pair, high-grade steel
REF
503.0556.0
503.0556.1
503.0556.2
503.0556.3
503.0556.4
503.0556.5
503.0556.6
503.0556.7
503.0558.0
503.0558.1
503.0558.2
503.0558.3
503.0558.4
503.0558.5
503.0558.6
503.0558.7
503.0554.0
503.0554.1
503.0554.2
503.0554.3
503.0554.4
503.0554.5
503.0554.6
503.0554.7
503.0551.0
503.0565.0
503.0565.1
503.0565.2
503.0565.3
503.0565.4
503.0565.5
503.0565.6
503.0565.7
503.0573.0
17
Page 18
8.0 Accessories, further options and consumables
Accessories
Leg rest
extendable, width 440 mm (complete length 1970 mm)
REF
503.0552.0
Secretion Basin, Plastic
To be pulled out, tiltable storage
Secretion basin stainless steel for the attachment
kit (REF 503.0571.0) 325 x 265 mm, height 65 mm,
additional price
Docotor´s foot support
Examination stool
height adjustable, safety chassis on 5 double castors
Colour: Pastel blue
Colour: Saffron yellow
Colour: Salmon
Colour: Leaf green
Colour: Light grey
Colour: Silk grey
Colour: White
Colour: Sapphire blue
503.0571.0
503.0555.0
503.0559.0
503.0570.0
503.0570.1
503.0570.2
503.0570.3
503.0570.4
503.0570.5
503.0570.6
503.0570.7
18
Colposcope holder
mountable left or right, for colposcope für ATMOS i View
Kaps support with tube diameter from 30 - 38 mm;
Consumables for the ATMOS Chair 41 Gyne
Paper role
503.0553.0
503.0557.0
Page 19
9.0 Technicalspecications
Nominal voltage230 V ± 10%
Nominal frequency50-60 Hz
Fuses2 × T 3.15 A 250 V
Control voltage5 V
Motor voltage24 V
Conditions
Transport and Storage
max. seat height (mm)900
min. seat height (mm)570
Width (mm)677
Length (mm)1970
Trendelenburg (degree)-10 + 25
Tilt of back rest (degree)0 ÷ 70
Net weight (kg)200
Dimensions packing (mm)800 x 1360 x 845
max. lift (kg)150
max. lifting weight200 kg
Protection class (EN 60601-1)I
Degree of protectiontype B
Classication according to appendix IX
EC Directive 93/42/EEC
Applied standardsEN 60601-1: 1996-03
Type of application discontinuous (18 sec ON - 340 sec OFF)
UMDNS-Code 10-960
GMDN-Code 38447
+10...+35°C
20...80 % atmospheric pressure from
500...1060 hPa
800 x 1360 x 365
I
19
Page 20
10.0 Disposal
●The ATMOS Chair 41 Gyne is not comprised of any hazardous materials.
●The materials of the housing can be recycled completely.
●Prior to disposal, device and accessories must be decontaminated.
●The materials are to be separated carefully.
●Pay attention to country-specic regulations for disposal (e. g. waste incineration).
Disposal within the EC
The product described above is a high-quality medical product with a long service life. After its life cycle it must be disposed of
professional. According to the EC directives (WEEE and RoHS) the device may not be disposed of in domestic waste. Please
observe existing national laws and rules for disposal of old devices.
Disposal within the Federal Republic of Germany
In the Federal Republic of Germany the law for electrical devices (ElektroG) rules the disposal of electrical devices. In order
to guarantee a proper disposal of your old device, please either pass on your old device to your specialised dealer or send it
directly to ATMOS MedizinTechnik for a professional disposal.
Disposal of the packing
All the materials used to pack the chair respect the environment and are 100% recyclable:
- untreated wooden pallet,
- carton,
- polyethylene bag,
- bubble-pack polyethylene.
The recovery and recycling of the materials used to pack the chair help to save on raw materials and to reduce waste volumes.
Hand the packing over to a disposal company that is authorised to recover recyclable materials.
Disposal of the chair
Once the chair has reached its expected lifetime and before it is nally disposed of, it must be made unusable by unplugging it
from the mains socket and cutting the power supply cable.
The chair disposal does not involve any hazard whatsoever.
For the correct safeguard of the environment, hand the unserviceable chair over to a disposal company that is authorised to
recover recyclable and non-recyclable materials.
• the chair has been installed, adjusted and repaired by personnel authorised by the manufacturer;
• the electrical system of the place in which the chair is installed complies with current standards;
• the equipment is used in compliance with the operating instructions.
20
Page 21
11.0 Notes on EMC
Guidelines and Manufacturer´s Declaration - Emissions
The ATMOS Chair 41 Gyne is intended for use in the electromagnetic environment specied below. The customer or user of the
ATMOS Chair 41 Gyne should ensure that it is used in such an environment.
The ATMOS Chair 41 Gyne uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The ATMOS Chair 41 Gyne is suitable for use in all establishments,
including domestic, and those directly connected to the public lowvoltage power supply network that supplies buildings used for domestic
purposes.
Guidelines and Manufacturer´s Declaration - Immunity for ATMOS Chair 41 Gyne
The ATMOS Chair 41 Gyne is intended for use in the electromagnetic environment specied below. The customer or user of the
ATMOS Chair 41 Gyne should ensure that it is used in such an environment.
Immunity Test
ESD
IEC 61000-4-2
EFT
IEC 61000-4-4
IEC 60601Test Level
± 6 kV Contact
± 8 kV Air
± 2 kV Mains
Compliance Level
± 4 kV Contact
± 8 kV Air
± 1 kV MainsMains power quality should be that of a typical
Electromagnetic Environment Guidance
Floors should be wood, concrete, or ceramics
tile. If oors are synthetic, the relative humidity
should be at least 30%.
commercial or hospital environment.
± 1 kV I/Os
Surges
IEC 61000-4-5
± 1 kV
symmetric
± 1 kV
symmetric
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage Dips / Dropout
IEC 61000-4-11
Power Frequency 50/60
Hz Magnetic eld
IEC 61000-4-8
NOTE U
is the mains alternating current prior to application of the test levels.
T
± 1 kV for input and
not applicable
output power
< 5 % U
(> 95 % Dip of the UT)
for 0.5 Cycle
T
< 5 % UT
(> 95 % Dip of the UT)
for 0.5 Cycle
Mains power quality should be that of a typical
commercial or hospital environment. If the
user of the ATMOS Chair 41 Gyne demands
continued function even in case of interruptions
40 % U
T
(60% Dip of the UT)
for 5 Cycles
70% U
T
(30 % Dip of the UT)
for 25 Cycles
< 5 % UT
(>95 % Dip of the UT)
for 5 s
40 % U
T
(60% Dip of the UT)
for 5 Cycles
70% U
T
(30 % Dip of the UT)
for 25 Cycles
< 5 % UT
(>95 % Dip of the UT)
for 5 s
of the energy supply, it is recommended to
supply the ATMOS Chair 41 Gyne from an
uninterruptible current supply or a battery.
3 A/mnot applicablePower frequency magnetic elds should be that
of a typical commercial or hospital environment.
21
Page 22
11.0 Notes on EMC
Guidelines and manufacturer's declaration – Electromagnetic noise immunity –
for devices and system which are not life-sustaining
The ATMOS Chair 41 Gyne is intended for use in the electromagnetic environment specied below. The customer or
user of the ATMOS Chair 41 Gyne should ensure that it is used in such an environment.
Immunity Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601Test Level
= 3 Ve
V
1
150 kHz bis 80 MHz
= 3 V/m
E
1
80 MHz bis 2,5 GHz
Compliance Level
3 V
3 V/m
Electromagnetic Environment Guidance
Portable and mobile
communications equipment
should be separated from the
ATMOS Chair 41 Gyne incl.
the cables by no less than the
distances calculated/listed below.
Recommended distances
d = (3,5 / V
d = (3,5 / E
d = (7 / E
) * √(P)
1
) * √(P)
1
) * √(P)
1
where „P“ is the max. power
in watts (W) and D is the
recommended separation
distance in meters (m).
Field strengths from xed
transmitters, as determined
by an electromagnetic site (a)
survey, should be less than the
compliance level (b).
Interference may occur in the
vicinity of equipment containing
following symbol.
NOTE 1 With 80 MHz and 800 MHz the higher frequency range applies.
NOTE 2 These guidelines are not applicable in any case. The propagation of electromagnetic sizes is
inuenced by absorptions and reections of buildings, objects and people.
22
Page 23
11.0 Notes on EMC
a) The eld strength of stationary transmitters, such as base stations of cellular phones and mobile terrain radioequipment, amateur radio transmitters, cbm broadcast and TV stations cannot be predestined exactly.
To determine the electromagnetic environment in regard to stationary transmitters, a study of the location is to
be considered. If the measured eld strength at the location where the ATMOS Chair 41 Gyne is used exceeds the above compliance level, the ATMOS Chair 41 Gyne is to be observed to verify the intended use. If abnormal performance characteristics are noted, additional measures might be necessary, e.g. a changed
arrangement or another location for the device.
b) Within the frequency range of 150 kHz to 80 MHz the eld strength is to be below 3 V/m.
Recommended separations between portable and mobile RF communications
equipment and the ATMOS Chair 41 Gyne
The ATMOS Chair 41 Gyne is intended for use in electromagnetic environment in which radiated disturbances are controlled. The
customer or user of the ATMOS Chair 41 Gyne can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF Communications equipment and the ATMOS Chair 41 Gyne as recommended below, according
to the maximum output power of the communications equipment.
Nominal output of the
transmitter
in Watts (W)
0,010,1160,1160,233
0,10,3670,3670,737
1,01,161,162,33
103,673,677,37
10011,6711.6723,33
For transmitters for which the maximum nominal output is not indicated in the above table, the recommended separation distance
d in meters (m) can be determined using the equation belonging to the respective column whereas P is the maximum nominal
output of the transmitter in watts (W) acc. to manufacturer´s specication.
NOTE 1 With 80 MHz and 800 MHz the higher frequency range applies.
NOTE 2 These guidelines are not applicable in any case. The propagation of electromagnetic sizes is inuenced by absorptions
and reections of buildings, objects and people.
Separation distance, depending on transmit-frequency m
150 kHz - 80 MHz
d = (3,5/V
) * √(P)
1
80 MHz - 800 MHz
d = (3,5/E1) * √(P)
800 MHz - 2,5 GHz
d = (7/E1) * √(P)
23
Page 24
ATMOS MedizinTechnik GmbH & Co. KG
Ludwig-Kegel-Str. 16
79853 Lenzkirch / Germany
Phone: +49 7653 689 -0
atmos@atmosmed.de
www.atmosmed.com