Atlas XENOTEST ALPHA, XENOTEST ALPHA HIGH ENERGY Operating Manual

XENOTEST ALPHA / ALPHA HIGH ENERGY
Operating manual
Operating manual XENOTEST ALPHA / ALPHA HIGH ENERGY ATLAS
Page 2
ATLAS Operating manual XENOTEST ALPHA / ALPHA HIGH ENERGY
Page 3
XENOTEST ALPHA / ALPHA HIGH ENERGY
Operating manual
Operating manual XENOTEST ALPHA / ALPHA HIGH ENERGY ATLAS
Page 4
1. Hints for this operating manual................................................................................................. 7
1.1 Sign explanation .................................................................................................................. 8
1.2 Safety instructions for the customer ................................................................................... 9
1.3 Safety instructions for the operator ................................................................................... 10
2. Features ...................................................................................................................................... 11
3. Setting up the unit .................................................................................................................... 12
3.1 Scope of delivery ............................................................................................................... 12
3.2 Installation requirements ................................................................................................... 12
3.2.1 Laboratory ......................................................................................................................... 12
3.2.2 Unit air exhaust ................................................................................................................. 12
3.2.3 Electrical supply ................................................................................................................ 12
3.2.4 Water quality ..................................................................................................................... 13
3.2.5 Water supply ..................................................................................................................... 13
3.2.6 Space requirements .......................................................................................................... 14
3.2.7 Unit dimensions ................................................................................................................. 15
4. Description of the unit .............................................................................................................. 16
4.1 Front panel of the unit ....................................................................................................... 16
4.1.1 Program control ................................................................................................................. 17
4.1.2 Test chamber ALPHA ........................................................................................................ 18
4.1.3 Test chamber HIGH ENERGY .......................................................................................... 18
4.2 Right side of the unit ......................................................................................................... 20
4.3 Upper side of the unit ........................................................................................................ 20
4.4 Left side of the unit............................................................................................................ 22
4.5 Rear panel ......................................................................................................................... 22
4.6 Ventilation ALPHA ............................................................................................................. 24
4.7 Ventilation HIGH ENERGY ............................................................................................... 24
5. Options and accessories ......................................................................................................... 26
5.1 Options .............................................................................................................................. 26
5.1.1 Rain- and humidity function .............................................................................................. 26
5.1.2 Printer ................................................................................................................................ 27
5.1.3 Mains transformers ........................................................................................................... 29
5.1.4 Interface ............................................................................................................................ 29
5.1.5 Turning mode gear ............................................................................................................ 29
5.2 Accessories ....................................................................................................................... 30
5.2.1 Filter system ...................................................................................................................... 30
5.2.2 Standard set of sample holders ........................................................................................ 35
5.2.3 Special holder for thick samples ....................................................................................... 35
5.2.4 Sample holder for blue scale ............................................................................................. 36
5.2.5 Cover plates ...................................................................................................................... 36
5.2.6 Software package XENOVIEW .........................................................................................36
5.2.7 XENOSENSIV sensor ....................................................................................................... 37
5.2.8 Adapter for XENOSENSIV sensor .................................................................................... 37
5.2.9 Calibration adapter and end plug ...................................................................................... 38
ATLAS Operating manual XENOTEST ALPHA / ALPHA HIGH ENERGY
Page 5
6. Start up ....................................................................................................................................... 39
6.1 Filter variations .................................................................................................................. 39
6.1.1 Fitting the filter to the latern .............................................................................................. 39
6.1.2 Insertion of the light exposure system .............................................................................. 40
6.1.3 Insertion of the lamp and the lantern ................................................................................ 40
6.1.4 Exchange of lamp ............................................................................................................. 40
6.2 Water supply ..................................................................................................................... 42
6.3 Insert the XENOSENSIV sensor ...................................................................................... 42
6.4 Adjust the XENOSENSIV sensor...................................................................................... 44
6.4.1 Fitting of the end plug ....................................................................................................... 44
7. Function ..................................................................................................................................... 45
7.1 Measuring and control sensor system .............................................................................. 45
7.1.1 Irradiance .......................................................................................................................... 45
7.1.2 Black standard temperature .............................................................................................. 45
7.1.3 Fan speed.......................................................................................................................... 45
7.1.4 Test chamber temperature ................................................................................................ 46
7.1.5 Test chamber humidity ...................................................................................................... 46
7.2 Rain ................................................................................................................................... 46
7.3 Program control ................................................................................................................. 47
7.4 Temperature diagrams ...................................................................................................... 47
8. Programing example ................................................................................................................. 49
8.1 Task of the program ......................................................................................................... 49
8.2 Solving the programing task.............................................................................................. 49
8.3 Program start .................................................................................................................... 53
8.4 Print out of current program parameters .......................................................................... 54
8.5 Change speed of blower during program operation.......................................................... 54
8.6 Stop a running program .................................................................................................... 54
8.7 Trouble shooting ................................................................................................................ 55
9. Setting up tests ......................................................................................................................... 58
9.1 Equip sample holders........................................................................................................ 58
9.2 Select unit parameters ...................................................................................................... 58
10. Care and cleaning ..................................................................................................................... 59
10.1 Cleaning of quartz- and glass components ...................................................................... 59
10.2 Cleaning the test chamber ................................................................................................ 59
10.3 Cleaning the water tank .................................................................................................... 59
10.4 Cleaning the air filter ......................................................................................................... 59
11. Maintenance ............................................................................................................................... 60
12. List of options and accessories .............................................................................................. 60
12.1 Options .............................................................................................................................. 60
12.2 Accessories ....................................................................................................................... 61
13. Technical data ............................................................................................................................ 62
Operating manual XENOTEST ALPHA / ALPHA HIGH ENERGY ATLAS
Page 6
Fig. 1 Space requirements .......................................................................................................... 14
Fig. 2 Unit dimensions ................................................................................................................. 15
Fig. 3 Front side of equipment .................................................................................................... 16
Fig. 4 Program control ................................................................................................................. 17
Fig. 5 Test chamber ALPHA ........................................................................................................ 19
Fig. 6 Test chamber HIGH ENERGY .......................................................................................... 19
Fig. 7 Main switch........................................................................................................................ 20
Fig. 8 Right side of the unit ......................................................................................................... 21
Fig. 9 Left side of the unit............................................................................................................ 23
Fig. 10 Ventilation scheme ALPHA ............................................................................................... 25
Fig. 11 Ventilation scheme HIGH ENERGY ................................................................................. 25
Fig. 12 Water circuits ..................................................................................................................... 26
Fig. 13 Printer ................................................................................................................................ 27
Fig. 14 Printer parameter .............................................................................................................. 28
Fig. 15 Change of the paper roll .................................................................................................... 29
Fig. 16 Mains transformers ........................................................................................................... 29
Fig. 17 XENOCHROME-filter system ALPHA and path of radiation ............................................ 31
Fig. 18 Spectral Distribution ALPHA
XENOCHROME 300 and CIE 85...................................................................................... 32
Fig. 19 Spectral Distribution ALPHA
XENOCHROME 270 and 320 ........................................................................................... 32
Fig. 20 Spectral Distribution HIGH ENERGY
XENOCHROME 300 and CIE 85...................................................................................... 33
Fig. 21 Spectral Distribution HIGH ENERGY
XENOCHROME 270 and 320 ........................................................................................... 33
Fig. 22 Absorption-filter system .................................................................................................... 34
Fig. 23 Sunlight behind window glass ...........................................................................................34
Fig. 24 Light fastness of car interior .............................................................................................. 34
Fig. 25 Spectral Distribution ALPHA
Absorption filter system .................................................................................................... 35
Fig. 26 Standard sample holder .................................................................................................... 36
Fig. 27 Special sample holder for thick samples ........................................................................... 36
Fig. 28 Sample holder für blue scale............................................................................................. 37
Fig. 29 Cover plates ...................................................................................................................... 37
Fig. 30 XENOSENSIV-sensor ....................................................................................................... 38
Fig. 31 Adapter for XENOSENSIV sensor .................................................................................... 38
Fig. 32 Calibration adapter and end plug ...................................................................................... 38
Fig. 33 Fitting the exposure system .............................................................................................. 41
Fig. 34 Fitting the XENOSENSIV sensors .................................................................................... 40
Fig. 35 Calibration adapter ............................................................................................................ 40
Fig. 36 End plug ............................................................................................................................ 42
Fig. 37 Humidity diagram ALPHA and HIGH ENERGY ................................................................ 46
Fig. 38 Temperature fields ALPHA with XENOCHROME filter system ........................................ 47
Fig. 39 Temperature fields ALPHA with absorption filter system.................................................. 48
Fig. 40 Temperature fields HIGH ENERGY with XENOCHROME filter system .......................... 48
ATLAS Operating manual XENOTEST ALPHA / ALPHA HIGH ENERGY
Page 7
Safety instructions Unit set-up Unit description Start-up
Function Setting up tests Care and cleaning Maintenance
Options und accessory
ALPHA and HIGH ENERGY features
Program example
List of options and
accessories
1. Hints for this operating manual:
The maximum unit versions XENOTEST ALPHA LM-W and XENOTEST ALPHA HIGH ENERGY LM-
W are described in this operating manual.
For simplification these unit versions are merely called ALPHA or HIGH ENERGY (HE) in this operating
manual.
Options and accessories are presented in chapter 5 in this operating manual. In case you have a unit
version without much options and accessories, please skip the corresponding sections in this operating
manual.
Mod el T W HT LM-G LM-W HE-LM-G HE-LM-W
- Equipment feature
- Basic unit Alpha
- Basic unit ALPHA HIGH ENERGY
X
-
X
-
X
-
X
-
X
-
-
X
-
X
- Non-turning mode
- Non-turning mode for use with XENOSENSIV
- Non-turning mode / turning mode Non-turning mode / turning mode for use wi th XE NOS ENSIV
-
-
X
-
-
-
X
-
-
X
-
-
-
X
-
-
-
-
-
X
-
X
-
-
-
-
-
X
- Control of lamp power in %
- Control of lamp rating in W/m²
X
-
X
-
-
X
-
X
-
X
-
X
-
X
Test chamber temperature
- controlled up to 45 °C
- controlled up to 70 °C
X
-
X
-
-
X
-
X
-
X
-
X
-
X
Measuring sensors at sample level XENOSENSIV E(300-400 nm) and BST
- - X X X X X
Stationary sensor for test chamber temperature and relative humidity
X X X X X X X
Ultrasonic humidification system X X X X X X X
Rain, pressureless - X - X X X X
Water supply with integrated waterreservoir (capacity 70 l)
X X X X X X X
Electrical versions:
- 200V, 50/60Hz
- 210V, 50/60Hz
- 220V, 50/60Hz
- 230V, 50/60Hz
- 240V, 50/60Hz
Must be indicated to customer specification !
Software versions:
- German
- Englisch
- French
Must be indicated to customer specification !
Operating manual XENOTEST ALPHA / ALPHA HIGH ENERGY ATLAS
Page 8
Safety instructions Unit set-up Unit description Start-up
Function Setting up tests Care and cleaning Maintenance
Options und accessory
ALPHA and HIGH ENERGY features
Program example
List of options and
accessories
1.1 Sign explanation
HINT!
HINT! is used when user tips and
helpful hints are given.
ATTENTION!
ATTENTION! is used when the non-compliance with i
nstructions may cause damage to the system.
DANGER!
DANGER! is used when the non-compliance with
instructions may cause danger to the user.
DANGEROUS VOLTAGE!
DANGEROUS VOLTAGE! is used when the non-compliance with instructions
may cause danger to the user because of dangerous voltage.
ATLAS Operating manual XENOTEST ALPHA / ALPHA HIGH ENERGY
Page 9
Safety instructions Unit set-up Unit description Start-up
Function Setting up tests Care and cleaning Maintenance
Options und accessory
ALPHA and HIGH ENERGY features
Program example
List of options and
accessories
1.2 Safety instructions for the customer
General:
The electrical connection of the equipment has to be made by an electric specialist.
Please note the special safety hints in the chapters "unit set-up" and "maintenance".
Instruct the operating personnel with the help of this operating manual in the operation, function and care of the unit. The unit may only be operated by instructed personnel. The operating manual should be kept close to the unit.
Inform the operating staff about accident prevention regulations regularly.
For work at and with the unit the user needs instructions in his mother language based on the operating manual (FRG: UVV VBG1 § 7 (2)).
Before starting the unit ensure that the necessary requirements for a failurefree operation of the unit are given. Observe the installation requirements in chapter 3.2 of this manual as well as valid
national laws, regulations and guiding principles.
Intended areas of application:
The ALPHA and HIGH ENERGY units are designed for research, development and test laborato­ries
The ALPHA and HIGH ENERGY units shall only be used for the exposure and weathering of materials in the intended areas of application.
Inflammable or explosive materials must not be inserted into the unit. Do not insert materials which despite their application as directed release toxical substances.
Handling and usage contrary to the intended areas of application described in this operating manual can present unforeseeable risk and should be avoided.
Maintenance and cleaning:
The lateral two-leaf doors must only be opened by Xenotest personnel, an autorised customer service or an electrical specialist.
After one year it is necessary to subject the unit - especially the safety components - to regular inspections.
Mounting, extensions, replacements and repair work may only be carried out by Xenotest or an authorised customer service engineer.
The safety of the unit is only given, when the necessary maintenance and repair work is kept within time and carried out by Xenotest or an authorised customer service engineer.
Operating manual XENOTEST ALPHA / ALPHA HIGH ENERGY ATLAS
Page 10
Safety instructions Unit set-up Unit description Start-up
Function Setting up tests Care and cleaning Maintenance
Options und accessory
ALPHA and HIGH ENERGY features
Program example
List of options and
accessories
1.3 Safety instructions for the operator
The unit must be operated by instructed and authorised personnel only.
Overheating and damage of the unit and the samples may occur, when the filters are not inser­ted. Always insert the filters!
Disconnect the unit from the mains, pull mains plug and secure against replugging, when fitting the filter system, the lamp and the lantern as well as when carrying out care and cleaning work.
Lamps, specimens, specimen holders and sensors heat up considerably during the weathering tests. Protective gloves must therefore be worn when removing them to prevent injury to your hands. This applies in particular to high temperature tests.
When removing the lamp exhaust air hose while the unit is running UV radiation may emerge and may cause irritation of the retina and the scin. Remove the lamp exhaust air hose only when the unit is switched off.
Please observe the special safety hints in the text of the chapters "unit description", "start-up", "function", "setting-up tests", "care and cleaning".
The unit complies with the safety standards of:
DIN VDE 0411 part 1 / 03.94 Measuring, Regulating, Controlling Safety requirements for electric motor-operated measuring, regulating, controlling and laboratory units.
DIN VDE 0711 part 1 / 09.91 Lamps
DIN VDE 0875 part 14 / 09.91 Limits and methods of measurement of radio disturbance characteristics of electric motor­operated and thermal appliances for household and similar purposes, electric tools and similar electric apparatus. (EN 55 014)
DIN VDE 0871 part 11/09.87 Limits of radio disturbance and measuring methods for industrial, scientific and medical high­frequency apparatus (EN 55 011).
ATLAS Operating manual XENOTEST ALPHA / ALPHA HIGH ENERGY
Page 11
Safety instructions Unit set-up Unit description Start-up
Function Setting up tests Care and cleaning Maintenance
Options und accessory
ALPHA and HIGH
ENERGY features
Program example
List of options and
accessories
2. Features
a
Pos. Feature XENOTEST ALPHA LM
1 Filter configuration Non-turning mode Warranty Turning mode Warranty
XENOCHROME 270 70 - 155 W/m² 135 W/m² 36 - 80 W/m² 70 W/m² XENOCHROME 300 50 - 110 W/m² 100 W/m² 26 - 57 W/m² 51 W/m² XENOCHROME 320 45 - 95 W/m² 80 W/m² 23 - 49 W/m² 42 W/m² Absorption filters
7 IR
45 - 90 W/m² 75 W/m² 23 - 46 W/m² 38 W/m²
Absorption filters 4 IR + 3 window glass
50 - 95 W/m² 90 W/m² 26 - 49 W/m² 46 W/m²
Absorption filters 6 IR + 1 UV
40 - 85 W/m² 70 W/m² 21 - 44 W/m² 36 W/m²
2 Rel. electrical power without
XENOSENSIV sensor
Continuously adjusted to preset value 53 - 120 %
3 Test chamber temperature** 20 - 70 °C
4 Black standard temperature ** 25 - 130 °C 25 - 100 °C 5 Relative humidity** 10% - 95% 6 Fan speed 500 - 2500 rpm
Pos. Feature XENOTEST ALPHA LM HIGH ENERGY
1 Filter configuration Non-turning mode Warranty Turning mode Warranty
XENOCHROME 270 80 - 270 W/m² 270 W/m² 42 - 140 W/m² 140 W/m² XENOCHROME 300 40 - 220 W/m² 190 W/m² 21 - 114 W/m² 99 W/m² XENOCHROME 320 40 - 180 W/m² 170 W/m² 21 - 94 W/m² 88 W/m² Absorption filters
7 IR
40 - 140 W/m² --- 21 - 72 W/m² ---
Window glass 10 WG
40 - 110 W/m² --- 21 - 57 W/m² ---
Absorption filters 4 IR + 3 window glass
40 - 140 W/m² --- 21 - 72 W/m² ---
Absorption filters 6 IR + 1 UV
40 - 140 W/m² --- 21 - 72 W/m² ---
2 Rel. electrical power without
XENOSENSIV sensor
Continuously adjusted to preset value
53 - 120 % 3 Test chamber temperature** 20 - 70° 4 Black standard temperature ** 25 - 130 °C 25 - 100 °C 5 Relative humidity** 10% - 95% 6 Fan speed 500 - 2500 rpm
Note: * Depending on the filtering, ATLAS guarantees a lamp lifetime of 1500 h for the values in bold type; unless separate values are indicated, it is applicable for the whole adjustment range. ** The values achieveable depend on the chosen filter system, the irradiance settings and the laboratory environmental conditions.
Operating manual XENOTEST ALPHA / ALPHA HIGH ENERGY ATLAS
Page 12
Safety instructions Unit set-up Unit description Start-up
Function Setting up tests Care and cleaning Maintenance
Options und accessory
ALPHA and HIGH ENERGY features
Program example
List of options and
accessories
3. Unit set-up
3.1 Scope of delivery
Supply in a suitable case. Scope of delivery of LM­W versions:
1. Control system with display and operating keyboard,
2. 1 Xenon lamp,
3. non-turning-, and turning mode gearing with connection for XENOSENSIV sensor,
4. sensor to measure test chamber temperature and rel. humidity,
5. XENOSENSIV sensor to measure the irradiance and the black standard tem­perature,
6. end plug for operation without XENOSENSIV sensor,
7. adapter to fit XENOSENSIV in test cham­ber,
8. outer cylinder made of special UV glass (Suprax),
9. water tank,
10. Thermo printer,
11. interface RS232 / RS485
12. Technical documentation:
- Operating manual
- Software documentation
- Circuit diagrams
- Spare part lists
Sensitive components such as:
1. Xenon lamp,
2. XENOSENSIV sensor,
3. Outer cylinder,
are packed separately.
Additional sample holders and optical filter sysytem are necessary for the run­ning of the unit.
3.2 Space requirements
3.2.1 Laboratory
Set up the unit in a air-conditioned and air­filtered laboratory at 18°C - 25°C and a rela­tive air humidity of max. 50%.
Supply an amount of approx. 300m3 air per hour via an air filter to the laboratory (filter type EU 4).
Keep the laboratory free of dust.
3.2.2 Exhaust air
The exhaust hoses for the lamp and the test chamber must be made of flexible heatresistant material, as the temperatures of the exhaust air may reach up to a max. of 120°C.
The max. length of the exhaust hose is limited to 3m including the option of two 90 degree bends.
In case of longer exhaust air ducts or with more than two 90 degree bends an external blower is needed.
3.2.3 Electrical supply
The electrical connection to the mains has to be made by an electrical specialist. The unit is designed for a mains supply of 230V +/- 10%. Different voltages are obtainable by fitting mains trans-formers. Connection is ensured by (1,N,PE or 2, PE), CEE (32A, 3pol.,6h).
For the electrical supply observe UVV VGB 4.
Mains conditions have to agree with the data on the name plate.
Observe VDE- and technical connection regulations of the energy supply compa­nies.
ATLAS Operating manual XENOTEST ALPHA / ALPHA HIGH ENERGY
Page 13
Safety instructions Unit set-up Unit description Start-up
Function Setting up tests Care and cleaning Maintenance
Options und accessory
ALPHA and HIGH ENERGY features
Program example
List of options and
accessories
3.2.4 Water quality
The water must be of high-grade purity to prevent deposits. The use of ion exchange cartridges alo­ne is not sufficient.
Regarding the water purity we refer to the mini­mum requirements of some important standards, such as DIN 53387 „Artifical light and weather fast­ness of plastics“:
1. Electrical conductivy: < 5 µS/cm
2. PH value: 6-8
3. content of solid bodies < 1 ppm measured as silicon dioxide.
3.2.5 Water supply
To ensure the functioning of the water circuit and the humidification system, the apparatus must be aligned by means of its height adjustable feet.
By supply line: fill the water tank (fig.9) with a pressure-resistant hose to the water inlet of the tank (inner diameter 13mm, 3bar pressure).
without supply line: fill the water tank manually.
Operating manual XENOTEST ALPHA / ALPHA HIGH ENERGY ATLAS
Page 14
Safety instructions Unit set-up Unit description Start-up
Function Setting up tests Care and cleaning Maintenance
Options und accessory
ALPHA and HIGH ENERGY features
Program example
List of options and
accessories
3.2.6 Space requirement (fig. 1)
Dimensions:
(Width x depth x height): 900mm x 780mm x 1800mm
For operating,- maintenance,- and repair work please observe the necessary space requirements.
For adequate cooling of the xenon lamp, take care that the air inlets and outlets at the top and back side of the equipment are not obstructed. Allow for the neces­sary space for operation, maintenance and repair of the equipment.
Figure 1: Space requirement
Set-up area
Service area
H
x
H
R min
=2,5
x
x
Dimensions in
m
x
H
R min
H
0,9
2,9
1,0
1,0
1
,
0
1
,
0
Hx=1,8 H
R min
=2,5
x
X = XENOTEST ALPHA LM and ALPHA HIGH ENERGY
= unit height
= minimum laboratory height
0,9
2,9
1,0
1,0
2
,
8
1
,
0
0
,
8
1
,
0
0
,
8
x
ATLAS Operating manual XENOTEST ALPHA / ALPHA HIGH ENERGY
Page 15
Safety instructions Unit set-up Unit description Start-up
Function Setting up tests Care and cleaning Maintenance
Options und accessory
ALPHA and HIGH ENERGY features
Program example
List of options and
accessories
3.2.7 Unit dimensions (Fig. 2)
Figure 2: Unit dimensions (in mm)
Operating manual XENOTEST ALPHA / ALPHA HIGH ENERGY ATLAS
Page 16
Safety instructions Unit set-up Unit description Start-up
Function Setting up tests Care and cleaning Maintenance
Options und accessory
ALPHA and HIGH ENERGY features
Program example
List of options and
accessories
4. Description of the unit
Figure 3: Front side of equipment
1 Program control 2 Door locking test chamber 3 Door locking water tank 4 Height adjustable feet
4.1 Front side of equipment (fig. 3)
2
3
1
4
ATLAS Operating manual XENOTEST ALPHA / ALPHA HIGH ENERGY
Page 17
Safety instructions Unit set-up Unit description Start-up
Function Setting up tests Care and cleaning Maintenance
Options und accessory
ALPHA and HIGH
ENERGY features
Program example
List of options and
accessories
The control panel on the front side of the apparatus controls and regulates the course of the entered weathering program. The program keyboard can
be locked during operation of the machine against unauthorized access by means of a key-operated switch.
Figure 4: Program control
4.1.1 Program control (fig. 4)
1 4-line LCD display for user guidance and
data display 2 Data key for data display 3 Print key to print out the stored program,
parameter measurements and system
status from the printer 4 Program input keyboard 5 Enter key to confirm the input values 6 Escape key 7 Key-operated switch to allow locking
of the keyboard when the system is
running 8 START key
9 Green LED light comes on when main
switch is on 10 STOP key 11 Red LED flashes when the program is
interrupted with the STOP key or when
the system is defective. The red LED is
continuously on when certain limit values
are exceeded 12 Key to interrupt the rotation of the
specimen holder carousel during a running
program resp. to move the carousel when
a program is interrupted (When door is
closed)
1
2
3
4
56
7
8
9
10
11
12
0
1
Start Stop 123
456
789
0
ESCAPE ENTER
DATA PRINT
Operating manual XENOTEST ALPHA / ALPHA HIGH ENERGY ATLAS
Page 18
Safety instructions Unit set-up Unit description Start-up
Function Setting up tests Care and cleaning Maintenance
Options und accessory
ALPHA and HIGH ENERGY features
Program example
List of options and
accessories
Open the test chamber by pressing the door lok­king. A program is interrupted when the test cham­ber door is opened while a unit is running.
Are located in the test chamber :
- Chosen outer cylinder (1),
- Xenon lamp (2),
- Filter lantern (3),
- 11 sample holders (4),
- sample holder mountings (5).
Function: The samples rotate around the lamp and filter lan­tern. In turning mode operation the samples are turned by 180° per caroussell revolution.
Ventilation: Test chamber and Xenon lamp are supplied by two separate air flows. A uniform air ventilation is achie­ved by an appropriate air flow (fig. 10 ventilation scheme).
Temperature, humidification and rain: Air is taken from the test chamber by an air ex- haust hose (6), warmed up by a heating system and humidified in the humidifier, and then sent back to the test chamber by an air supply hose (7). The maximum chamber humidity can be adjusted up to 95% r. h. depending on the preset tempera­ture.
The rain bar (8) mounted in the test chamber is used to spray ultra-pure water in cyclical intervals and without any pressure on to the specimens.
Sensors: At sample level a XENOSENSIV sensor (9) for measuring the irradiance in the range between 300-400nm and the black standard temperature in the surface temperature range up to 130°C can be fitted to the test chamber by means of an ad- apter (10) (fig. 31). The sensor is connected to the program control via a connection socket (11). If no sensor is used the end plug must be fitted to the connection socket (fig. 36). A sensor to measure the test chamber tempe- rature and rel. humidity (12) is mounted in the upper rear of the test chamber. The sensors com­bined with the program control guarantee control of constant setting values.
Contrast to ALPHA:
Are locared in the test chamber:
- arc-shaped Xenon lamp (1)
Ventilation: A uniform air ventilation of the sample area is achie­ved by an air conduction plate (2) (fig. 11 venti­lation scheme).
4.1.3 Test chamber HIGH ENERGY (fig. 6)
4.1.2 Test chamber ALPHA (fig. 5)
ATLAS Operating manual XENOTEST ALPHA / ALPHA HIGH ENERGY
Page 19
Safety instructions Unit set-up Unit description Start-up
Function Setting up tests Care and cleaning Maintenance
Options und accessory
ALPHA and HIGH
ENERGY features
Program example
List of options and
accessories
2 31
4
5
6
7
8
9
10
11
12
1 arc-shaped
Xenon lamp
2 Air conduction plate
Figure 6: Test chamber HIGH ENERGY
Figure 5: Test chamber ALPHA
1 Outer cylinder 2 Straight Xenon lamp 3 Filter lantern 4 Sample holder 5 Sample holder mounting 6 Air exhaust hose 7 Air supply hose 8 Rain bar 9 XENOSENSIV sensor 10 Adapter 11 Connection socket 12 Sensor for test chamber
temperature and rel. humidity
21
Operating manual XENOTEST ALPHA / ALPHA HIGH ENERGY ATLAS
Page 20
Safety instructions Unit set-up Unit description Start-up
Function Setting up tests Care and cleaning Maintenance
Options und accessory
ALPHA and HIGH ENERGY features
Program example
List of options and
accessories
4.2 Right-hand apparatus side ALPHA and HIGH ENERGY (fig. 8)
4.3 Apparatus top ALPHA and HIGH ENERGY (fig. 8)
Data output: Data of stored programs as well as current measu­ring values can be printed-out by a thermoprinter (2). The interface (3) on the right-hand side of the main switch provides for the data output and ex­ternal computer control.
Electrical components: The mains transformer of the humidifier (5), the
control transformer (6), a mains voltage inter- ference filter (7) and the fuse base of the con­trol voltage (8) are located below the test cham- ber (4).
The power supply unit (10) of the lamp is moun- ted below a protective wire mesh (9). The mains transformer (11) and the lamp choke (12) (not HIGH ENERGY unit) are located below the power supply unit. The heating relay (14) is seated next to the main contactor (13) and the control of the test chamber fan (15). There is a 25 amps lamp fuse (16) protection and a 10 amps
heating fuse (17) protection. The operating hour counter (18) records the operating time of the
equipment.
Ventilation: The air intake for the lamp cooling (19) passes through a filter pad in the equipment door.
The main switch (1) (fig.7) is located on the upper left of the unit. Use this switch to disconnect the unit from the mains in case of trouble.
The lateral two-leaf doors of the unit must only be opened by XENOTEST, or an authorized customer service engineer or an electrical specialist. Contact with live parts of a unit connec­ted to the mains may involve electrical shock. Before opening the lateral two-leaf doors disconnect unit from the mains. Unplug the connection socket and secure against replugging!
Figure 7: Main switch
0
1
The exhaust air flow of the lamp is evacuated at the top of the unit via an exhaust air socket
(20) to the open. (note figure 10 and 11, ventilation schemes).
Loading...
+ 44 hidden pages