Atkins 9319C User Manual

Page 1
60°C
25°C
0°C
BATTERY
(RED - LOW)
ON
OFF
9319C TYPE K THERMOCOUPLE PROVER
60°C
25°C
0°C
BATTERY
(RED - LOW)
ON
OFF
Specications:
Accuracy: ± 0.14°C Three Output Test Points: 0°C, 25°C and 60°C Ambient Conditions: 10°C to 32°C, 90%RH (non-condensing) Battery Type: CR2032 3 Volt Lithium Battery Life: 100 Hours minimum
Low battery indicator (LED)
certain areas:
Hold the prover like this
Operating Instructions:
NOTE: The prover must be allowed 30 minutes to acclimate
prior to use.
1. Set the ON / OFF slide switch to ON position.
2. The battery indicator light will blink once to indicate the battery life:
Green LED: The battery is good
Red LED: Replace the battery (See Battery
Replacement Procedure)
3. Allow the prover to warm-up for a period of 10 seconds.
4. Disconnect the probe from thermocouple instrument and connect the
9319C Prover.
5. Set the output slide switch to the temperature to be validated. The instrument should read the selected temperature.
6. Consult the thermocouple instrument instruction sheet if
recalibration is required.
7. Set the ON / OFF slide switch to OFF position
Battery Replacement Procedure:
1. Use a #1 Phillips screwdriver to remove the two housing screws.
2. Lift the circuit board from the housing.
3. Before removing the spent battery, take note of the battery orientation
in the holder. It is very important that the new battery be placed in the proper ± orientation.
4. Remove the spent battery and dispose of responsibly.
5. Place the new battery into the battery holder.
6. Replace the circuit board, cover, and screws.
NOTE: The 9319C does not have automatic shut-off, so be sure to turn off the prover after each use to preserve battery life.
Not like this
LIMITED ONE YEAR WARRANTYLIMITED ONE YEAR WARRANTY
33 Reeds Gap Road • Middleeld, CT 06455-0450 U.S.A.
Cooper-Atkins Corporation
860-347-2256 • www.cooper-atkins.com
27-206