ATI Technologies Q46-85 O & M Manual

O & M Manual
Model Q46-85
Peracetic Acid Monitor
Analytical Technology, Inc. ATI (UK) Limited 6 Iron Bridge Drive Unit 1 & 2 Gatehead Business Park Collegeville, PA 19426 Delph New Road, Delph Phone: 800-959-0299 Saddleworth OL3 5DE 610-917-0991 Phone: +44 (0)1457-873-318 Fax: 610-917-0992 Fax: + 44 (0)1457-874-468 Email: sales@analyticaltechnology.com Email: sales@atiuk.com
Table of Contents
PART 1 - INTRODUCTION ................................. 4
1.1 General ...................................................... 4
1.2 Standard System ....................................... 5
1.3 Features ..................................................... 8
1.4 Q46/85 System Specifications .................. 9
1.5 Q46H Performance Specifications ...........10
PART 2 – ANALYZER MOUNTING .................11
2.1 General .....................................................11
2.2 Wall or Pipe Mount ..................................12
2.3 Panel Mounting ........................................14
PART 3 – SENSOR/FLOWCELL MOUNTING 15
3.1 General .....................................................15
3.2 Constant-Head Flowcell ...........................15
3.3 Sealed Flowcell ........................................16
3.4 Submersion Mounting ..............................18
PART 4 – ELECTRICAL INSTALLATION ......19
4.1 General .....................................................19
4.2 Power Connection ....................................19
4.4 Direct Sensor Connection ........................23
4.5 Sensor Wiring ......................................24
4.6 Junction Box Connection .........................25
PART 5 – SENSOR ASSEMBLY .........................26
5.1 PAA Sensor Preparation ..........................26
5.2 Optional pH Sensor ..................................28
PART 6 – CONFIGURATION .............................29
6.1 User Interface ...........................................29
6.11 Keys .........................................................30
6.12 Display .....................................................30
6.2 Software ...................................................32
6.21 Software Navigation ...............................32
6.22 Measure Menu [MEASURE] ...................35
6.23 Calibration Menu [CAL].............................36
6.24 Configuration Menu [CONFIG] ..............36
6.25 Control Menu [CONTROL] ....................41
6.26 Diagnostics Menu [DIAG] .........................45
PART 7 – CALIBRATION ...................................48
7.1 PAA Calibration ......................................48
7.11 PAA Zero Cal ..........................................48
7.12 PAA Span Cal ..........................................49
7.2 Temperature Calibration ..........................50
7.3 pH Calibration..........................................51
7.31 Two-Point pH Cal ....................................51
7.32 One-Point pH Cal .....................................52
PART 8 – PID CONTROLLER DETAILS .........53
8.1 PID Description .......................................53
8.2 PID Algorithm .........................................53
8.3 Classical PID Tuning ...............................55
8.4 Manual PID Override Control..................55
8.5 Common PID Pitfalls ...............................55
PART 9 – SYSTEM MAINTENANCE ................57
9.1 General .....................................................57
9.2 Analyzer Maintenance .............................57
9.3 PAA Sensor Maintenance ........................57
9.31 Sensor Acid Cleaning ..............................58
9.4 Flow Cell Maintenance ............................58
PART 10 – TROUBLESHOOTING ....................59
10.1 General .....................................................59
10.2 External Sources of Problems ..................59
10.3 Analyzer Tests .........................................60
10.4 Display Messages ....................................60
10.5 Sensor Tests .............................................63
SPARE PARTS ......................................................65
O&M Manual Rev F (8/17)
2
Table of Figures
FIGURE 1 - TYPICAL SYSTEM DIAGRAM .......................................................................................... 5
FIGURE 2 - MONITORING SYSTEM W/BAYONET STYLE PH SNSOR .................................................... 6
FIGURE 3 - SEALED FLOWCELL ASSEMBLIES W/FLOW CONTROL ...................................................... 7
FIGURE 4 Q46 ENCLOSURE DIMENSIONS .................................................................................. 11
FIGURE 5 - WALL OR PIPE MOUNT BRACKET ................................................................................. 12
FIGURE 6 - WALL MOUNTING DIAGRAM ........................................................................................ 13
FIGURE 7 - PIPE MOUNTING DIAGRAM .......................................................................................... 13
FIGURE 8 - PANEL MOUNT CUTOUT ............................................................................................. 14
FIGURE 9 - CONSTANT HEAD FLOWCELL DETAILS ......................................................................... 15
FIGURE 10 - SEALED FLOWCELL DETAILS ..................................................................................... 16
FIGURE 11 - SEALED PH FLOWCELL DETAILS ............................................................................... 17
FIGURE 12 - SUBMERSIBLE SENSOR MOUNTING ASSEMBLY .......................................................... 18
FIGURE 13 - LINE POWER CONNECTION ....................................................................................... 20
FIGURE 14 - OPTIONAL RELAY BOARD WIRING ............................................................................. 21
FIGURE 15 - OPTIONAL RELAY BOARD WIRING ............................................................................. 22
FIGURE 16 - OPTIONAL ANALOG OUTPUT WIRING ......................................................................... 22
FIGURE 17 - SENSOR CABLE PREPARATION ................................................................................. 23
FIGURE 18 - SENSOR WIRING ...................................................................................................... 24
FIGURE 19 - JUNCTION BOX INTERCONNECT WIRING .................................................................... 25
FIGURE 20 - PAA SENSOR ASSEMBLY ......................................................................................... 26
FIGURE 21 - SUBMERSIBLE PAA SENSOR ASSEMBLY ................................................................... 27
FIGURE 22 - USER INTERFACE ..................................................................................................... 29
FIGURE 23 - SOFTWARE MAP ...................................................................................................... 34
FIGURE 24 - AUTOMATIC PH BUFFER TABLES ............................................................................... 40
FIGURE 25 - CONTROL RELAY EXAMPLE, HYSTERSIS AND TWO OPPOSITE PHASE OPTIONS ............ 43
FIGURE 26 - ALARM RELAY EXAMPLE .......................................................................................... 44
FIGURE 27 - Q46H ISA (IDEAL) EQUATION ................................................................................... 53
FIGURE 28 - Q46H DSIAPLY MESSAGES ...................................................................................... 61
FIGURE 29 - DISPLAY MESSAGES (CONTINUED) ........................................................................... 62
FIGURE 30 - PT100 RTD TABLE .................................................................................................. 63
O&M Manual Rev F (8/17)
3
Part 1 - Introduction
1.1 General
The Model Q46/85 is an on-line monitoring system designed for the continuous measurement of Peracetic Acid (PAA) concentration in solution. The full scale operating range of the system may be selected by the user for 0-20.00 ppm, 0-200.0 ppm, or 0-2000 ppm, and the sensing system will operate on water streams with temperatures ranging from 0 to 50°C. The Q46/85 PAA Measurement System is engineered for applications in which PAA is applied to water for sterilization purposes
The basic sensing element used in the PAA monitor is a polarographic membrane sensor which measures PAA directly. Water simply flows past the sensor and directly to drain, with the flow rate and pressure across the sensor controlled by a constant head flow cell assembly. The PAA measurement does not alter the sample or add any chemicals to the sample stream, so the water flow can return to the system if desired.
Q46/85 Monitors are available in two basic electronic versions, a 90-260 VAC powered monitor, or a 12-24 VDC powered monitor – both with three relays and two 4-20 mA analog outputs. Both versions allow flexible output options which include additional 4-20 mA current outputs, additional 0-30 VDC signal relays, PID control, and digital communication cards.
O&M Manual Rev F (8/17)
4
ATI Q46H/85 PAA System Part 2 – Analyzer Mounting
1.2 Standard System
The standard model Q46/85 system includes three main components, the Q46 analyzer, a constant head flow cell, and a PAA sensor. A low-volume flowcell is also available for applications where sample flowrate and pressure can be carefully controlled. Figure 1 shows a typical installation including the optional pH sensor.
For connection of the sensor to the electronics, a 25' cable is supplied. An additional 100 feet of interconnect cable may be added using #07-0100 junction box. All required spare parts are also provided with the basic system, including spare membranes, electrolyte, o-rings, and any special hardware.
O&M Manual Rev F (8/17)
Figure 1 - Typical System Diagram
5
ATI Q46H/85 PAA System Part 2 – Analyzer Mounting
Figure 2 below shows the same standard flowcell assembly and PAA sensor along with the conventional type pH sensor. A special adapter is required to hold the pH sensor in its proper location in the flowcell inlet chamber.
Figure 2 - Monitoring System w/Bayonet Style pH Snsor
O&M Manual Rev F (8/17)
6
ATI Q46H/85 PAA System Part 2 – Analyzer Mounting
Figure 3 below shows an installation using a (00-1522) sealed flowcell for the PAA sensor and a (00-
1527) sealed flowcell for the pH sensor. This type of installation requires careful flow control. We recommend the use of our (03-0371) flow control assembly when using sealed flowcells. This assembly consists of an in-line filter and a fixed-flow regulator which will maintain a constant 400 cc/min flowrate through the system. This flow will be maintained so long as inlet pressures are between 5 and 125 PSIG. The in-line filter is used mainly to protect the flow control element against larger particles that might cause plugging of the device.
Figure 3 - Sealed Flowcell Assemblies w/Flow Control
7
O&M Manual Rev F (8/17)
ATI Q46H/85 PAA System Part 2 – Analyzer Mounting
1.3 Features
Standard Q46H Analyzers have fully inputs and outputs. Analog outputs are additionally completely isolated from each other
Available in either 90-260 VAC or 12-24 VDC power supply systems. All features remain the
same in both variations.
Output Hold, Output Simulate, Output Alarm, and Output Delay Functions. All forced changes in output condition include bumpless transfer to provide gradual return to on-line signal levels and to avoid system control shocks on both analog outputs.
Three 6 amp SPDT relay outputs and two analog 4-20 mA outputs are standard. Software settings for relay control include setpoint, deadband, phase, delay, and failsafe. An optional 3­relay card for low voltage signals is available to bring the total to 6 relays.
Selectable PID controller on main analog output. PID controller can operate with instrument configured as loop-power transmitter, or as one of the two outputs on the AC powered instrument. PID includes manual operation feature, and diagnostic “stuck-controller” timer feature for relay notification of control problems.
Two analog outputs on the relay version may be configured to track PAA and temperature, PAA and PAA, or PAA and pH. Both analog outputs can be individually programmed to fail to specific values.
Digital communication option for Profibus-DP, Modbus-RTU, or Ethernet-IP. See Q46 Digital
Communications Manual for Specifications.
Diagnostic messages provide a clear description of any problem with no confusing error codes to look up. Messages are also included for diagnosing calibration problems.
Quick and easy one-point calibration method and sensor zero-cal. To provide high accuracy, all calibration methods include stability monitors that check temperature and main parameter stability before accepting data.
Security lock feature to prevent unauthorized tampering with transmitter settings. All settings can be viewed while locked, but they cannot be changed.
O&M Manual Rev F (8/17)
8
ATI Q46H/85 PAA System Part 2 – Analyzer Mounting
1.4 Q46/85 System Specifications Displayed Parameters Main input, 0.01 ppm to 2000 ppm
Sensor temperature, -10.0 to 55.0 °C (23 to 131 ºF) Sensor Current, 0.0–999.9 nA, 0.000 to 99.99 uA Loop current, 4.00 to 20.00 mA Sensor slope/offset Model number and software version PID Controller Status Optional pH Input value. 0.00 to 14.00 pH
Main Parameter Ranges Manual selection of one of the following ranges,
0.00 to 20.00 ppm
0.0 to 200.0 ppm 0 to 2000 ppm
Display 0.75” (19.1 mm) high 4-digit main display with sign
12-digit secondary display, 0.3" (7.6 mm) 5x7 dot matrix.
Integral LED back-light for visibility in the dark.
Weight Approximately 5 lbs. (2.3 Kg) with sensor & flowcell Ambient Temperature Analyzer Service, -20 to 60 °C (-4 to 140 ºF)
Sensor Service, -5 to 55°C (23 to 131 °F)
Ambient Humidity 0 to 95%, non-condensing EMI/RFI Influence Designed to EN 61326-1, CE Marked
Output Isolation 600 V galvanic isolation Filter Adjustable 0-9.9 minutes additional damping to 90% step input Temperature Input Pt100 RTD with automatic compensation
Sensor 2-electrode polarographic membrane sensor for direct
Sensor Materials PVC and PEEK Electrical Certification Ordinary Location, cCSAus (CSA and UL standards - both CSA
Sensor Cable 25 ft. (7.5 meter) cable with 6-pin plug. Max. Sensor-to-Analyzer 100 feet (30.5 meters), with junction box
Distance Flow Cell Constant head overflow, clear cast acrylic, 7-30 GPH, 15 GPH
Storage, -5 to 70 °C (-22 to 158 ºF)
measurement of PAA,
approved), pollution degree 2, installation category 2
recommended, inlet is ¼” hose barb at 1/8” MNPT, outlet is ½” hose barb at 3/8” MNPT
O&M Manual Rev F (8/17)
9
ATI Q46H/85 PAA System Part 2 – Analyzer Mounting
Power 90 - 260 VAC, 50-60 Hz, 10 VA max or
12-24 VDC, 500 mA max.
Enclosure NEMA 4X (IP-66), polycarbonate, stainless steel hardware,
weatherproof and corrosion resistant,
Mounting Options Wall or pipe mount bracket standard. Bracket suitable for either
1.5” or 2” I.D. U-Bolts for pipe mounting.
Panel mount adapter optional.
Conduit Openings Five ½” NPT openings, Adapter can be removed to provide a 1”
NPT opening in the bottom of the enclosure. Gland seals provided but not installed.
Relays, Electromechanical: Three SPDT, 6 amp @ 250 VAC, 5 amp @ 24 VDC contacts.
Software selection for setpoint, phase, delay, deadband, hi-lo alarm, and failsafe. A-B indicators on main LCD, and C indicator on lower display.
Optional Relays: Three additional non-isolated low power relays available. SPST,
1 Amp @ 30 VDC.
Analog Outputs Two 4-20 mA outputs. Output one programmable for PPM PAA
or PID. Output 2 programmable for PPM PAA or Temperature. Max load 500 Ohms for each output. Outputs ground isolated and isolated from each other. An additional 3rd analog option is available.
1.5 Q46H Performance Specifications
Accuracy 0.5% of selected range or 0.02 PPM Repeatability 0.3% of selected range or 0.01 PPM Sensitivity 0.05% of selected range Non-linearity 0.1% of selected range Warm-up Time 3 seconds to rated performance (electronics only) Supply Voltage Effects ± 0.05% span Instrument Response Time 60 seconds to 90% of step input at lowest damping
Equipment bearing this marking may not be discarded by traditional methods in the European community after August 12 2005 per EU Directive 2002/96/EC. End users must return old equipment to the manufacturer for proper disposal.
O&M Manual Rev F (8/17)
10
ATI Q46H/85 PAA System Part 2 – Analyzer Mounting
Part 2 – Analyzer Mounting
2.1 General
All Q46 Series instruments offer maximum mounting flexibility. A bracket is included with each unit that allows mounting to walls or pipes. In all cases, choose a location that is readily accessible for calibrations. Also consider that it may be necessary to utilize a location where solutions can be used during the calibration process. To take full advantage of the high contrast display, mount the instrument in a location where the display can be viewed from various angles and long distances.
Locate the instrument in close proximity to the point of sensor installation - this will allow easy access during calibration. The sensor-to-instrument distance should not exceed 100 feet. To maximize signal-to-noise ratio however, work with the shortest sensor cable possible. The standard cable length of the PAA sensor is 25 feet.
Due to the flexibility of the instrument design, some of the mounting features change based on the configuration that was ordered. For example, the 2-wire transmitter version is different for the 115/230 VAC controller because the rear of the enclosure is much deeper when the AC powered unit is used. In addition, the AC powered unit has an integrated panel mount flange requiring a single cutout for flush mounting.
Figure 4 – Q46 Enclosure Dimensions
11
O&M Manual Rev F (8/17)
ATI Q46H/85 PAA System Part 2 – Analyzer Mounting
2.2 Wall or Pipe Mount
A PVC mounting bracket with attachment screws is supplied with each transmitter as shown in Figure 5. The multi-purpose bracket is attached to the rear of the enclosure using the four flat head screws. The instrument is then attached to the wall using the four outer mounting holes in the bracket. These holes are slotted to accommodate two sizes of u-bolt that may be used to pipe mount the unit. Slots will accommodate u-bolts designed for 1½ “ or 2” pipe. The actual center to center dimensions for the u-bolts are shown in the drawing. Note that these slots are for u-bolts with ¼-20 threads. The 1½” pipe u-bolt (2” I.D. clearance) is available from ATI in type 304 stainless steel under part number (47-0005).
Figure 5 - Wall or Pipe mount Bracket
12
O&M Manual Rev F (8/17)
ATI Q46H/85 PAA System Part 2 – Analyzer Mounting
Figure 6 - Wall Mounting Diagram
Figure 7 - Pipe Mounting Diagram
13
O&M Manual Rev F (8/17)
ATI Q46H/85 PAA System Part 2 – Analyzer Mounting
2.3 Panel Mounting
Panel mounting uses the panel mounting flange molded into the rear section of the enclosure. Figure 8 provides dimensions for the panel cutout required for mounting.
The panel mounting bracket kit must be ordered separately (part number 05-0094). This kit contains a metal retainer bracket that attaches to the rear of the enclosure, 4 screws for attachment of this bracket, and a sealing gasket to insure that the panel mounted monitor provides a water tight seal when mounted to a panel.
The sealing gasket must first be attached to the enclosure. The gasket contains an adhesive on one side so that it remains in place on the enclosure. Remove the protective paper from the adhesive side of the gasket and slide the gasket over the back of the enclosure so that the adhesive side lines up with the back of the enclosure flange. Once in place, you can proceed to mount the monitor in the panel.
O&M Manual Rev F (8/17)
Figure 8 - Panel Mount Cutout
14
Part 3 – Sensor/Flowcell Mounting
3.1 General
Select a location within the maximum sensor cable length for mounting of the sensor flow cell.
3.2 Constant-Head Flowcell
PAA sensors are best used in a constant-head overflow chamber because variations in sample flow rate and pressure can cause unstable readings. When monitoring low concentrations (below
0.5 PPM), this method should always be used. Mechanical installation of the flow cell requires that it be mounted to a wall or other convenient flat surface. Alternatively, the mounting holes on the plate will accommodate a 2" U-bolt for mounting the plate to a 2" pipe. Figure 9 shows the dimensions and mounting hole locations for the flow cell. Be sure to allow enough clearance on the left side of the flow cell for insertion and removal of the sensor. About 12 inches clearance is recommended.
Once mounted, inlet and drain connections must be made. The flow cell contains a 1/8" MNPT inlet connection and a 3/8" MNPT drain connection. Hose barbs for the inlet and drain connections are supplied with the flow cell for use with flexible tubing. The inlet hose barb is used with ¼" I.D. tubing and the drain hose barb is used with ½" I.D. tubing.
O&M Manual Rev F (8/17)
Figure 9 - Constant Head Flowcell Details
15
ATI Q46H/85 PAA System Part 3 – Sensor/Flowcell Mounting
3.3 Sealed Flowcell
Applications where the sample inlet flow is well controlled can use a simpler sealed flowcell. Using this flowcell requires that flow be controlled externally to about 400 cc/min. Variable flow rate or variable pressure will cause unstable readings in this flowcell. ATI offers a special flow control element that can be used ahead of this flowcell on the incoming sample line. The flow control is part no. (55-0048). It will control the inlet flowrate at 400 cc/min. with inlet pressure variations from 5-150 PSIG. A 50 micron y-strainer ahead of the flow control element is recommended. The sealed flowcell provides a drain vent with check valve to avoid pulling a vacuum on the flow chamber.
O&M Manual Rev F (8/17)
Figure 10 - Sealed Flowcell Details
16
ATI Q46H/85 PAA System Part 3 – Sensor/Flowcell Mounting
O&M Manual Rev F (8/17)
Figure 11 - Sealed pH Flowcell Details
17
ATI Q46H/85 PAA System Part 3 – Sensor/Flowcell Mounting
3.4 Submersion Mounting
Some applications are much easier done using the submersible sensor. This method can sometimes be used where flow is reasonably constant, and hydraulic head does not vary appreciably. PAA sensors can never be used in completely stagnant conditions. A flow velocity of at least 0.3 feet per second is normally required for measurement. Any applications for a submersible PAA sensor should first be discussed with ATI. A trial of such installations may be necessary.
Submersible sensors are mounted to a 1" pipe using a standard 1" PVC thread by thread pipe coupling. The mounting pipe can be secured to standard 1½" pipe rail using a mounting bracket kit available from ATI (part number 00-0628) as shown in Figure 12 - Submersible Sensor Mounting
Assembly .
SENSOR CABLE
2" HANDRAIL
SWIVEL MOUNTING BRACKET WITH
HARDWARE, SUPPLIED BY ATI
SUPPLIED BY ATI
SENSOR, TYPICAL
2 - 3 FT. SUBMERGENCE
1" ALUMINUM CONDUIT, THREADED1" T x T PVC COUPLING
ONE END OR 1" SCHED 80 PVC PIPE
LENGTH AS REQUIRED
(SUPPLIED BY CUSTOMER)
Figure 12 - Submersible Sensor Mounting Assembly
O&M Manual Rev F (8/17)
18
Part 4 – Electrical Installation
4.1 General
The Q46 is powered in one of two ways, depending on the version purchased. The 12-24 VDC powered analyzer requires a customer supplied DC power supply. The 90-260 VAC version requires line power. Please verify the type of unit before connecting any power.
WARNING: Do not connect AC line power to a DC powered unit. Severe damage will
result.
Important Notes:
1. Use wiring practices that conform to all national, state and local electrical codes. For proper safety as well as stable measuring performance, it is important that the earth ground connection be made to a solid ground point from TB7. The AC power supply contains a single 630mA slo-blo fuse (Wickmann/Littlefuse #372-0630). The fuse F1 is located adjacent to TB7 and is easily replaceable.
2. Do NOT run sensor cables or instrument 4-20 mA output wiring in the same conduit that contains AC power wiring. AC power wiring should be run in a dedicated conduit to prevent electrical noise from coupling with the instrumentation signals.
3. This analyzer must be installed by specifically trained personnel in accordance with relevant local codes and instructions contained in this operating manual. Observe the analyzer's technical specifications and input ratings. Proper electrical disconnection means must be provided prior to the electrical power connected to this instrument, such as a circuit breaker - rated 250 VAC, 2 A minimum. If one line of the line power mains is not neutral, use a double-pole mains switch to disconnect the analyzer.
4. Repeated problems with lightning strikes damaging sensitive instrumentation are
often attributed to poorly bonded earth grounds in the instrument power source. The protection schemes incorporated into this analyzer cannot operate to maximum efficiency unless the ground connection is at its’ absolute lowest impedance.
There is no standard ground resistance universally recognized. Many agencies recommend a ground resistance value of 5 ohms or less. The NEC recommends an impedance to ground of less than 25 ohms, and less than 5 ohms where sensitive equipment is installed. Power sources feeding sensitive instruments like the Q46H/79PR should have the lowest possible impedance to ground.
4.2 Power Connection
Verify the AC power supply requirement before installing. Also verify that power is fully disconnected before attempting to wire.
Q46 systems are supplied with 5 cable gland fittings for sealing cable entries.
Connect HOT, NEUTRAL, and GROUND to the matching designations on terminal strip TB7.
O&M Manual Rev F (8/17)
19
ATI Q46H/85 PAA System Part 4 – Electrical Installation
WARNING
Disconnect line power voltage BEFORE connecting
line power wires to Terminal TB7 of the power supply.
wire
ies are
260 VAC operation at the factory at
time of order, and the power supply is labeled as
such. Do NOT connect voltages other than the
The two analog outputs for the standard system are present at terminalTB1. The loop-load limitation in this configuration is 500 Ohms maximum for each output. Also note that these two outputs are completely isolated from each other to insure that ground loops do not result from the connection of both outputs to the same device such as a PLC or DCS.
A ribbon cable connects the power supply assembly with the microprocessor assembly located in the front section of the enclosure. This cable may be unplugged from the front section of the monitor if service is needed, but should normally be left in place during installation.
The power supply accepts only standard three­single phase power. AC power suppl configured for 90-
labeled requirement to the input.
Figure 13 - Line Power Connection
The power strip, TB7, allows up to 12 AWG wire. A wire gauge of 16 AWG is recommended to allow for an easy pass-through into the ½” NPT ports when wiring.
20
O&M Manual Rev F (8/17)
ATI Q46H/85 PAA System Part 4 – Electrical Installation
4.3 Relay Connection
Three SPDT relays are provided on the power supply board. None of the relay contacts are powered. The user must supply the proper power to the contacts. For applications that require the same switched operating voltage as the Q46 (115 or 230 V), power may be jumpered from the power input terminals at TB7. Relay wiring is connected at TB4, TB5, and TB6 as shown below. Note that the relay contact markings are shown in the NORMAL mode. Programming a relay for “Failsafe” operation reverses the NO and NC positions in this diagram Figure 14.
O&M Manual Rev F (8/17)
Figure 14 - Optional Relay Board Wiring
21
Loading...
+ 46 hidden pages