ATI Technologies B12-70 O & M Manual

O & M Manual
B12-70
2-Wire Wet Acid
Gas Transmitter
Home Office European Office
Analytical Technology, Inc. ATI (UK) Limited 6 Iron Bridge Drive Unit 1 & 2 Gatehead Business Park Collegeville, PA 19426 Delph New Road, Delph Phone: 800-959-0299 Saddleworth OL3 5DE 610-917-0991 Phone: +44 (0)1457-873-318 Fax: 610-917-0992 Fax: + 44 (0)1457-874-468 Email: sales@analyticaltechnology.com Email: sales@atiuk.com
Web: www.Analyticaltechnology.com
Fax: 610-917-0992 Fax: + 44 (0)1457-874-468
Model B12-70 Wet Acid Gas Transmitter
TABLE OF CONTENTS
TABLE OF CONTENTS ................................................................................................................... 2
INTRODUCTION .............................................................................................................................. 4
SPECIFICATIONS............................................................................................................................ 6
INSTALLATION ............................................................................................................................... 7
MECHANICAL MOUNTING .......................................................................................................... 7
ELECTRICAL CONNECTION .......................................................................................................... 9
WET ACID GAS SENSOR ASSEMBLY ....................................................................................... 11
OPERATION .................................................................................................................................. 13
TRANSMITTER TEST POINTS .................................................................................................. 13
DIGITAL DISPLAY OPTION ....................................................................................................... 14
CALIBRATION ............................................................................................................................... 15
ZERO ADJUSTMENT ................................................................................................................. 15
SPAN ADJUSTMENT ................................................................................................................. 15
SENSOR RESPONSE TEST ...................................................................................................... 16
MAINTENANCE ............................................................................................................................. 17
SPARE PARTS LIST ..................................................................................................................... 18
O & M Manual - 2 -
-Release E, 7/15
Model B12-70 Wet Acid Gas Transmitter
TABLE OF FIGURES
Figure 1 - Typical Insertion System Installation (ATI-0156) ................................................................... 4
Figure 2 - Typical Flow Through System Installation (ATI-0223) ........................................................... 5
Figure 3 - Nema 4X Transmitter Enclosure Dimensions (ATI-0198) ...................................................... 7
Figure 4 - Explosion-Proof Transmitter Enclosure Dimensions (ATI-0226) ......................................... 8
Figure 5 - Wet Acid Gas Sensor Dimensions (ATI-0276) ........................................................................ 8
Figure 6 - Electrical Connections (ATI-0181) ........................................................................................... 9
Figure 7 - Electrical Connections (ATI-0182) ......................................................................................... 10
Figure 8 - Wet Acid Gas Sensor Assembly (ATI-0265) ......................................................................... 11
Figure 9 - Transmitter Controls & Test Points (ATI-0227) .................................................................... 14
O & M Manual - 3 -
-Release E, 7/15
Model B12-70 Wet Acid Gas Transmitter
INTRODUCTION
The Series B12-70 is a two-wire transmitter system designed for use with a special acid gas sensor (for HF, HCl, or HBr) suitable for use in very wet environments. It provides a low power method of monitoring acid gas in applications where excessive moisture causes rapid blinding of standard acid gas sensors.
Transmitters are housed in a NEMA-4X polystyrene enclosure (Figure 3) suitable for installation in virtually any environment. The sensor cable is wired to the proper terminals in the transmitter through the Pg 11 cable gland located in one of the enclosure knockouts. One cable gland and one ½” conduit hub is supplied for customer installation. An optional LCD provides local indication of gas concentration on the front of the enclosure. The display is supplied only when specified on the order, and cannot be added in the field.
B12 Transmitters are also available in an explosion-proof version. The explosion-proof version is housed in a galvanized steel enclosure (Figure 4) rated for Class 1, Div. 1, Groups B, C, & D locations. LCD is not available on the XP version.
Figure 1 shows a typical insertion B12-70 gas monitoring system. The sensor is shown threaded into a blind flange of a vent stack.
Figure 1 - Typical Insertion System Installation (ATI-0156)
O & M Manual - 4 -
-Release E, 7/15
Model B12-70 Wet Acid Gas Transmitter
Figure 2 shows a typical flow through B12-70 gas monitoring system. The sensor is shown with the low volume flowcell in place. If used in insertion applications, the low volume flowcell must be removed.
Figure 2 - Typical Flow Through System Installation (ATI-0223)
O & M Manual - 5 -
-Release E, 7/15
Model B12-70 Wet Acid Gas Transmitter
SPECIFICATIONS
Type: Wet Acid Gases such as HF, HCl and HBr
Range: 0-10 PPM Min., 0-100 PPM Max.
Accuracy: ±0.2 PPM at constant temperature
Electronic Repeatability: ±1% of span
Electronic Linearity: ±0.5% of span
Output: Loop-powered 4-20 mA, 675 ohms maximum at 24 VDC
Power: 12-28 VDC
Display: None. Optional LCD available.
Enclosure: NEMA 4X polystyrene
Explosion-proof cast iron optional, Class 1, Div. 1, Groups B, C, & D.
Controls: Zero and span potentiometers mounted on transmitter circuit board
Operating Temperature: -5º to +55º C
Sensor: Model 70 Wet Acid Gas Sensor
Weight: 0.25 Lbs (0.12 Kg.)