定义9
User Responsibility 10
Precautions 10
For Customers in Canada 12
For Customers in the USA 12
For Customers in Europe 14
For All Customers 15
Instrument Description 16
Filter Leak Testing: The Most Common Application 17
How the Photometer Operates 18
Theory of Operation 18
Sampling System 18
Light Scattering Chamber (LSC) 19
Capabilities 19
Input / Output features 19
Utility Requirements 20
Packing List 21
Unpacking 21
Installation 22
Before you begin 22
Connecting Electrical Power 22
Connecting the Scanning Probe 23
Connecting the Interface Ports 24
USB Port 24
Printer Port 24
Front Panel General Overview 25
Rear Panel General Overview 26
Scanning Probe Overview 27
Thermal Printer Overview 28
Function Keys 29
Sampling Ports 29
3 Aerosol Photometer - 2i 1800224 H.1
CHAPTER 5
CHAPTER 6
CHAPTER 7
CHAPTER 8
Operating the 2i: Basic Operation 30
Before You Begin 30
Basic Operation 31
Reporting Modes 34
Summary Mode 34
Monitoring 36
Continuous 38
Alarms Modes and Set Point 38
Alarm Set Point 38
Alarm modes 39
Icons 40
Software Reference 41
Software Description 41
Start-up Screen 41
Making Entries 42
Entering Numerical Values 42
Turning Options Flags On or Off 43
Using the Alphanumerical Keypad 43
General Screen Structure 44
Menu Structure and Options 45
Menu Structure 45
Run Mode Screen 46
Top Level Menu Options 47
Alarm Set Point 48
Alarms 49
Set 100% to Upstream Concentration 49
Instructions 50
Measuring the Upstream Aerosol 50
Accepting the Result 51
Aerosol Concentration Result Too Low 51
Aerosol Concentration Result Too High 52
Set Internal Reference 52
Use Previous 100% Settings 53
Re-establish Zero 53
Setup 54
Aerosol Noise Suppression 54
Reporting Functions 56
Date and Time 57
Display 58
Error Messages 59
Application Notes 61
Abbreviations List 61
Aerosol Correction Factors 61
Maintaining the Photometer 62
Definitions and Features 62
Recommended Scheduled Maintenance 63
Daily 63
4 Aerosol Photometer - 2i 1800224 H.1
Annually 63
Replacing the Main Power Fuses 64
Cleaning the Lint Screens 64
CHAPTER 9
APPENDIX A Contacting ATI 67
APPENDIX B Warranty 68
APPENDIX C Declaration of Conformity 70
APPENDIX D Specifications 71
APPENDIX E Maintenance and Spare Items 72
APPENDIX F Accessories 73
APPENDIX G Manual Revision History 74
APPENDIX H Español – Precauciones 75
APPENDIX I Français – Précautions 77
APPENDIX J 中文 – 预防79
APPENDIX K Svenska – Försiktighetsåtgärder 81
Troubleshooting Guide 65
Instrument Will Not Power Up 65
Error E1 – Sample Line Blockage 65
Error E2 – Out of Temperature Range 66
Error E3 – Photomultiplier Tube (PMT) Inoperable 66
For Technical Support or Application Questions 67
For Customer Service 67
Main Unit Physical Characteristics 71
Operational Requirements 71
Operating Conditions 71
Aerosol Detection 71
Data Output Formats 71
Outputs 71
Scanning Probe 73
Thermal Printer Module (Optional) 73
Wheeled Transport Case (Optional) 73
Figure 13: Main Menu Screenshot 47
Figure 14: Setting the Alarm Set Point Screenshot 48
Figure 15: Setting the Alarms Screenshot 49
Figure 16: Set 100% to Upstream Concentration Initial Screenshot 50
Figure 17: Set 100% to Upstream Concentration Measurement Screenshot 51
Figure 18: Set 100% to Upstream Concentration Final Screenshot 51
Figure 19: Select Reagents Screenshot 52
Figure 20: Set Internal Reference Screenshot 52
Figure 21: Setup Menu Screenshot 54
Figure 22: Aerosol Noise Suppression Screenshot without Probe 55
Figure 23: Aerosol Noise Suppression Screenshot with Probe 55
Figure 24: Reporting Functions Screenshot 56
Figure 25: Date and Time Menu Screenshot 57
Figure 26: Set Time Screenshot 57
Figure 27: Set Date Screenshot 57
Figure 28: Set Date Format Screenshot 58
Figure 29: Display Setup Screenshot 58
Figure 30: Display Setup Screenshot 58
Figure 31: Error Message Screenshot 59
Tables
Table 1: Serial Port Settings ............................................................................................. 24
Table 17: Replacement Fuses for the 2i ............................................................................ 64
Table 18: Maintenance and Spare Items ........................................................................... 72
Table 19: Accessories List ................................................................................................ 73
Table 20: Manual Revision History .................................................................................. 74
6 Aerosol Photometer - 2i 1800224 H.1
CHAPTER 1
Introduction and Precautions
Scope of Manual
These instructions cover the specifications, features,
operation, maintenance, and troubleshooting for the 2i
Digital Aerosol Photometer.
These instructions also contain important information
required for the safe operation of the instrument.
Before using this instrument, all personnel associated
with the operation must read and understand this
entire manual and become familiar with the
terminology.
Failure to follow the specified procedures and
precautions could result in PERSONAL INJURY or
DAMAGE to the unit.
Definitions
The following defines the warnings, cautions and
notes used throughout this manual.
Warning
Indicates a strong possibility of severe personal injury or
death if instructions are not followed.
Caution
Indicates a possibility of equipment damage if
instructions are not followed.
Note
Indicates that helpful information is provided.
7 Aerosol Photometer - 2i 1800224 H.1
Definiciones
Las siguientes anotaciones definen los peligros,
cuidados y notas usadas a lo largo de este manual.
Advertencia
Indica una fuerte probabilidad de serias lesiones
personales o muerte si no son seguidas las instrucciones.
Cautela
Indica una posibilidad de daños al equipo si no son
seguidas las instrucciones.
Nota
Indica que se suministra información útil.
Définitions
Les informations suivantes, définissent les symboles,
précautions et notes présentes dans ce manuel.
Avertissement
Indique la possibilité de dommages corporels graves
pouvant entraîner la mort, si les instructions ne sont pas
suivies.
Attention
Indique la possibilité d’endommager l’équipement si les
instructions ne sont pas suivies.
Note
Indique que des informations utiles sont fournies.
8 Aerosol Photometer - 2i 1800224 H.1
Definitioner
Efterföljande text beträffande varningar, varsamhet
och noteringar är genomgånde för hela manualen.
Varning
Innebär att allvarlig personlig skada eller död kan inträffa
om instruktionerna inte följs.
Varsamhet
Innebär att skador på utrustning kan inträffa om
instruktionerna inte följs.
Notera
Innebär att användbar information ges.
定义
下面定义了在这本说明书中所有的警告,小心,注意.
警告
表示如果不按照指导操作.引起个人人身伤害甚至死亡
的可能性很高.
小心
表示如果不按照指导操作.可能会损害仪器
注意
表示提供了有用的信息.
9 Aerosol Photometer - 2i 1800224 H.1
User Responsibility
The user must:
1. read and comprehend the information contained in
this manual before using the product;
2. have an understanding of the electrical and
mechanical system principles used in the
operation of this photometer;
3. be trained in the proper use of electro-mechanical
equipment;
4. properly use this product for the intended purpose
and follow all regulations and procedures that
apply to the location where this product is used;
5. maintain the product as specified in this manual;
6. maintain in proper working condition any other
equipment associated with the operation of this
product.
Precautions
Read and understand all instructions. Failure to follow all
instructions listed below may result in electric shock, fire
or serious personal injury. The warnings, cautions, and
instructions discussed in this operation and maintenance
manual cannot cover all possible conditions and situations
that may occur. It must be understood by the operator that
common sense and caution are factors which cannot be
built into this equipment, but must be supplied by the
operator.
Warning – General Safety Rules
10 Aerosol Photometer - 2i 1800224 H.1
Caution – Shipping Hazard
When returning a 2i-N for service, the end user must
remove the potentially contaminated sampling train prior
to shipping the instrument. Refer to the 2i-N addendum for
the proper procedure.
Caution –Serviceability
There are no user-serviceable parts inside the instrument.
Refer all repair and maintenance to a qualified factoryauthorized technician.
Caution - Maintenance
Incorrect fuse replacement may damage internal
components of the photometer.
Warning – Electrical Hazard
Electrical hazard that can cause severe injury or death.
The electrical housing contains multiple high voltage
sources. Do not insert any objects under the cover
Warning – Damaged Equipment
Do not operate the equipment with a damaged cord or
plug or after the equipment has malfunctioned, or
been damaged in any way.
Contact ATI Service Department for advice on
examination, repair, electrical or mechanical
adjustment.
Failure to follow the prescribed procedures may result
in a hazardous situation.
11 Aerosol Photometer - 2i 1800224 H.1
For Customers in Canada
Note – Regulatory Information
This Class B digital apparatus complies with Canadian
ICES-003.
Cet appareil numérique de la Classe B est conforme à la
norme NMB-003 du Canada.
For Customers in the USA
Declaration of Conformity
Trade Name: ATI
Model Number: 2i
Responsible Party: Hamilton Associates Inc.
Address: 11403 Cronridge Drive
Owings Mills, MD 21117
USA
Telephone No: 410-363-9696
This device complies with Part 15 of the FCC Rules.
Operation is subject to the following two conditions: (1)
This device may not cause harmful interference, and (2)
this device must accept any interference received, including
interference that may cause undesired operation.
12 Aerosol Photometer - 2i 1800224 H.1
Note – Regulatory Information
This equipment has been tested and found to comply with
the limits for a Class A digital device, pursuant to Part 15
of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and
can radiate radio frequency energy and, if not installed in
accordance with the instructions, may cause harmful
interference to radio communications. However, there is
no guarantee that interference will not occur in a particular
installation. If this equipment does cause interference to
radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to
try to correct the interference by one or more of the
following measures:
Reorient or relocate the receiving antenna
Increase the separation between the equipment and
receiver
Connect the equipment into an outlet on a circuit
different from that to which the receiver is connected
Consult the dealer or an experienced radio/TV
technician for help
13 Aerosol Photometer - 2i 1800224 H.1
For Customers in Europe
Declaration of Conformity
Application of Council Directive(s): 2004/108/EC
Standard(s) to which conformity is declared:
EN 55011:2007; EN61000-3-2: 2005; EN61000-33:1995:A1 (2001):A2 (2005); EN 61326-1:2006
Manufacturer's Name: Air Techniques International
Manufacturer's Address: 11403 Cronridge Drive
Type of Equipment: Measurement Instrument
Model Number: 2i Photometer
Year of Manufacture: 2012
I, the undersigned, hereby declare that the equipment
specified above conforms to the above Directive(s)
and Standard(s).
will be accepted free of charge. This symbol is only
valid in the European Union. If you wish to discard
this product please contact your local authorities or
dealer and ask for the correct method of disposal.
This product shall not be treated as household waste.
Instead it shall be handed over to the applicable
collection point for recycling of electrical and
electronic equipment. By ensuring this product is
disposed of correctly, you will help prevent potential
negative consequences for the environment and
human health, which could otherwise be caused by
inappropriate waste handling of this product. The
recycling of materials will help to conserve natural
resources.
For more detailed information about recycling of this
product, please contact your local Civic Office, your
household waste disposal service or the Company
where you purchased this equipment.
This symbol on the product and / or
accompanying documents means that
used electrical and electronic products
should not be mixed with general
household waste. For proper
treatment, recovery and recycling,
please take this product(s) to
designated collection points where it
15 Aerosol Photometer - 2i 1800224 H.1
CHAPTER 2
Photometer Overview
Instrument Description
The 2i is a forward light-scattering, linear digital
photometer. It operates on 100 to 240 Volts, 50 or 60
Hz, adjusting automatically. It’s most basic function is
to sample air or other gases and report the mass
concentration of particulates in the sample. The
primary application for the 2i is integrity/leak testing of
HEPA/ULPA filtration systems.
The 2i is a compact and lightweight instrument. The
instrument case is constructed of rugged die cast
aluminum. The pressure-sensitive keypad and large,
bright LCD display and indicators provide ease of
operation and readability. The auto-ranging and onestep zeroing features assure the accuracy of all
readings.
Using the 2i is extremely intuitive through the use of a
basic menu structure. Operation of the instrument is
simple and straight forward. Use of the fully
compatible iProbe affords full unit control from the
scanning probe. The 2i provides easy access to realtime results via USB or optional printer, while
remaining versatile enough to accommodate special
testing needs that have become commonplace in
today’s filter testing industry.
When all of these elements come together as they
have in the 2i, the testing process ceases to be a
training intensive exercise and becomes a “set and
forget” secondary routine performed as part of the
daily work requirements in clean room testing.
16 Aerosol Photometer - 2i 1800224 H.1
Filter Leak Testing: The Most Common
Application
The most common application of the Aerosol
Photometer 2i is to detect leaks in high efficiency
filtration systems (HEPA & ULPA). To validate the
integrity of a filtration system, a known challenge
agent consisting of an airborne test aerosol is
generated and introduced upstream of the filter. The
challenge agent provides particulate matter upstream
of the filter to allow measurements downstream of the
filter.
The test aerosol is introduced into the upstream side
of the filter(s) as far from the filters as is practical to
insure adequate mixing. Ideally, a distance of 10 duct
diameters upstream is considered the minimum. A
sample of the aerosol-air mixture should be taken
from the upstream side, close to the center of the
filter/filter bank. This sample is used to establish a
100% base line for the upstream concentration. The
2i is adjusted as described in the Operating Section to
set the 100% reading and the stray light is adjusted
automatically. The stray light adjustment
compensates for light reflection off internal surfaces of
the scattering chamber. After these adjustments have
been made, the 2i instrument is ready to check for
filter leaks/integrity downstream.
The filter test is performed with the use of the
scanning probe. The filter and the perimeter of the
filter pack should be scanned by passing the probe in
slightly overlapping strokes so the entire area of the
filter is sampled. The end of the probe should be held
one inch from the filter surface. Separate passes
should be made around the entire periphery of the
filter, along the bond between the filter pack and the
frame, and around the seal of the filter. Readings on
the meter will indicate percent of penetration.
The display indicates the percent of leakage through
or around the filter. The iProbe is supplied with the
industry standard rectangular, blue isokinetic nozzle.
This nozzle is used for fast scanning and is accepted
by many standards, including NSF 49-2008.
17 Aerosol Photometer - 2i 1800224 H.1
Other optional probes are available, a round, black 1
inch (25 mm) in diameter nozzle, which complies with
NSF (National Sanitation Foundation) Standard 491992, and a round, red, isokinetic nozzle. The
isokinetic nozzles are designed for media face
velocities of 90 +/- 20 feet per minute (fpm) when
using a 1 cfm (28.3 L/min) sample rate.
How the Photometer Operates
Theory of Operation
When air or gas is drawn through the instrument,
particulate matter in the sample passes through the
focal point of the scattering chamber. Particles
scatter light which is detected by the photomultiplier
tube, which converts the light into an electrical signal.
The signal is amplified and digitized, then analyzed by
a microprocessor to determine the intensity of the
light scattered by the signal. This signal is then
compared to a known reference signal to provide an
output that is normalized to the reference response.
A photometer is ideally suited to detect particulate
matter in air or gas, reporting the mass concentration
encountered on the unit display. Particles from less
than 0.1 micron to approximately 600 microns can be
readily detected by the 2i. Since the photometer
reports concentration of particulate matter, many
applications are possible.
18 Aerosol Photometer - 2i 1800224 H.1
Sampling System
A vacuum pump in conjunction with a flow meter,
onboard sensors and closed feedback loop provides a
constant volumetric sample flow rate of 1 cfm (28.3
liters per minute). The 2i uses an oil-free, dual head,
diaphragm pump with a direct-coupled DC motor.
Solenoid actuated valving directs the airflow through
the sampling system to the scattering chamber from
three possible sources. The CLEAR position directs
particle free air from an internal 0 REFERENCE filter
to the scattering chamber for zeroing the instrument.
The UPSTREAM position permits sampling of the air
above the filter being challenged, and the
DOWNSTREAM position permits sampling of the air
that penetrates the filter.
Light Scattering Chamber (LSC)
The scattering chamber is not only an integral part of
the sampling system; it is a major component in itself.
The scattering chamber is a complex electro-optical
unit that consists of a pair of hollow cones connected
at the apexes. A pair of collimating lenses first
straightens the light emerging from the light source,
and then focuses it at the center of the sampling
cone. An aperture forms a dark cone around the
photomultiplier, preventing light from arriving directly
on the photomultiplier. A condensing lens opposite
the LED source focuses light scattered into this dark
cone onto the photomultiplier tube.
Capabilities
The Aerosol Photometer 2i will measure percentage
leakage or absolute aerosol concentration. By using
a 100% baseline of 100 micrograms per liter of
aerosol, it is possible to directly read the
concentrations of aerosols.
Note
Before attempting to operate this unit, become familiar
with the features and functions.
Input / Output features
Alarms:
The 2i, when used in conjunction with the iProbe,
provides the user three sensory alarms: audible,
visual, and vibratory. When a reading exceeds the
user selected alarm point the display will turn red, a
tone will sound, and the iProbe will vibrate, if enabled.
19 Aerosol Photometer - 2i 1800224 H.1
Data Output:
Leakage data is sent out the unit’s USB port. This
data may be logged using contemporary data
acquisition software in real-time. Multiple types of
data formatting are available to satisfy a wide range of
applications (see “Reporting Modes” page 34 for more
details). A “Hard Copy” of the data can also be printed
using the optional printer accessory when “Summary”
reporting mode is selected.
Utility Requirements
Power
The 2i is equipped with a “Universal AC” power
supply capable of handling 100 to 240 VAC, 50 or 60
Hz. A line power conditioner and surge protector is
recommended.
Ambient Air Conditions
While the instrument was design to be operated
through a wide range of environmental conditions, we
recommend its use in a controlled environment where
the temperature is between 5°C and 50°C (41°F to
122°F Fahrenheit) and the relative humidity is
between 5% and 95% non-condensing.
Note
A dedicated power line is recommended for stable
machine operation.
20 Aerosol Photometer - 2i 1800224 H.1
CHAPTER 3
Unpacking and Setting Up the
Photometer
Packing List
The 2i ships complete, ready to install and operate.
Any additional accessories purchased on the same
order will be listed as separate line items on the
shipment packing list. Please confirm receipt of all
line items prior to commissioning of the instrument.
Shipping with the standard 2i will be the following.
Individual items may be packaged separately or
installed:
Sturdy Shipping instrument case
“Operation and Maintenance” Manual
Calibration Report
iProbe Scanning Probe with umbilical
12 feet of clear “Upstream” sampling tube
One set of spare ‘O’-rings and lint screen
Power Cord (NEMA 5-15 125V or CEE 7/4
“Schuko” 250V configuration)
NSF 49:2008 Isokinetic (blue) rectangular
nozzle
Unpacking
21 Aerosol Photometer - 2i 1800224 H.1
Wherever possible, packing materials should be
retained for storage or future shipment and
transportation needs.
After unpacking, if anything is missing or appears to
be damaged, contact ATI Customer Service
immediately at (410) 363-9696 (see “Contacting ATI”
page 67).
Installation
Before you begin
Warning
If the unit is used in a manner not specified within the
user’s manual, the protection offered by the equipment
may be impaired.
You will need the following items to set up the Aerosol
Photometer Model 2i
Electrical outlet (100 VAC to 240 VAC)
2i Scanning Probe or PVC Sampling Tube
5°C to 50°C ambient temperature
Less than 95% RH, non-condensing.
Note
High ambient temperatures may create instability in the
readings.
Connecting Electrical Power
Voltage and current requirements for the Aerosol
Photometer Model 2i are:
100 to 240 VAC, 2 amps
The Aerosol Photometer automatically adjusts to
operate at the correct AC voltage for the destination
country (given this voltage is within the specifications
described in “APPENDIX D Specifications page 71).
This voltage is noted on a label attached to the back
panel of the instrument. The power cord contains a
plug which is specifically adapted to the destination
country.
You are responsible for plugging the power cord into
a matching receptacle. To connect the Aerosol
Photometer to electrical power, do the following:
22 Aerosol Photometer - 2i 1800224 H.1
1. Check to make sure the Aerosol Photometer is
turned off. (Refer to Figure 2: Rear Panel Overview,
for the location of the power switch.)
2. Plug the power cord into a matching power outlet.
Warning
Before connecting the power cord to the power outlet,
make sure the cord has not been cut or otherwise damaged
during shipment.
Connecting the Scanning Probe
If using the Aerosol Photometer 2i with the Scanning
Probe, you should first connect the iProbe to the
umbilical. To do so, connect the umbilical’s electrical
connector to the 7-pin connector on the end of the
probe. Connect the aerosol sampling line quickconnect fitting to the quick-connect connector on the
butt of the probe (see “Scanning Probe Overview”
page 27 for the probe description).
Note
The two pneumatic quick-connect fittings are not
interchangeable; the smaller one must be connected to the
iProbe, while the larger one must be connected to the 2i
base unit.
Once the umbilical is connected to the scanning
probe, connect the other end to the base unit.
Connect the umbilical’s electrical connector to the 7pin connector on the front panel of the 2i. Connect the
aerosol sampling line quick-connect fitting to the
quick-connect port of the photometer marked
“DOWNSTREAM” (see “Front Panel General
Overview” page 25 for the front panel component
description).
Note
Always check the lint screen located on both sampling
ports before instrument use (see “Maintaining the
Photometer” page 63 for the procedure).
23 Aerosol Photometer - 2i 1800224 H.1
When using the iProbe with the 2i, the probe should be
connected to the photometer before the power is turned on
if possible. Otherwise the iProbe display may show the
following message until the main unit has reached the
running mode:
“Waiting for User to complete current Base operation”
Connecting the Interface Ports
The 2i comes equipped with two standard Interface
Ports: a USB port and a printer port.
USB Port
The USB port is intended for interface with a
computer to provide serial data acquisition capability.
An industry standard USB Type ‘A’ Male to Type ‘B’
Male cable is required. The serial port connection
settings are listed in the following table.
Table 1: Serial Port Settings
Setting Value
Transfer Speed 9600
Data Bits 8
Stop Bits 1
Parity Control N
Flow Control None
Note
24 Aerosol Photometer - 2i 1800224 H.1
Printer Port
A 5-pin mini-DIN circular connector provides the
interface for the optional thermal printer. The
connector is keyed to prevent improper connection.
To connect, orient the arrow toward the top and insert
the connector in the receptacle.
Note – Port Usage
Do not use the Printer Port with equipment other than
those designed and supplied by ATI. Damage to the
equipment might result from such action
CHAPTER 4
Know Your Equipment
Front Panel General Overview
Figure 1: Front Panel Overview
LCD Display
Upstream
Sample Port
Downstream
Sample Port
Valve Selection with
LED Indicator
iProbe Connector
Description Function
LCD Display The color LCD display is the primary interface with the 2i.
During operation, percent leakage readings are displayed as
well as icons informing the user of the selection and status of
alarms, selections and connected peripherals.
Function Keys The function keys are used to access the settings and
operating parameters with the menu structure.
Valve Selection
with LED Indicator
Upstream Sample
Port
Downstream
Sample Port
iProbe Connector
Pressing one of the buttons will select the port as the source of
aerosol and a blue LED will indicate the selection.
Connects to the sample tubing that is used to measure the
upstream aerosol concentration.
Connects to the sample tubing that is used to measure the
downstream aerosol concentration.
A 7-pin circular connector with bayonet lock provides the
electrical connection for the scanning probe.
Function
Keys
25 Aerosol Photometer - 2i 1800224 H.1
Loading...
+ 57 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.