ATI Technologies 2H, 2H-N Installation And Operation Manual

IQ/OQ
Installation / Operation
Qualification MANUAL
for the
2H and 2H-N
Aerosol Photometers
Air Techniques International
Division of Hamilton Associates, Inc
11403 Cronridge Drive
Owings Mills, MD 21117 USA
TEL: 410 363-9696 FAX: 410 363-9695
www.atitest.com
info@atitest.com
IQ/OQ 2H, 2H-N Revision 0
This page intentionally left blank
Revision A
2
Table of Contents
1 INSTALLATION QUALIFICATION 6
1.1 Requirements 6
1.2 Unpacking 6
2 SETUP 7
2.1 Locations & Functions 7
2.2 Front Panel Controls and Indicators 8
2.3 Scanning Probe connections 8
2.4 Rear Panel Connections 9
2.5 Side Connection 10
3 OPERATIONAL QUALIFICATION 11
3.1 Initialization 11
3.2 Setup Parameters 11
3.3 Changing Setup Parameters 12
3.4 Setting Up the Alarm Functions 17
3.5 Changing the Reference 17
3.6 UPSTREAM AEROSOL Measurements (DOP or PAO) 18
3.7 Measuring % Leakage - User Defined Reference (USER) 18
3.8 Testing with the 2H 19
3.9 Verifying the Upsream Concentration During Test 19
4 MAINTENANCE 20
4.1 Weekly 20
4.2 Annually 20
4.3 General Maintenance procedures 20
4.4 Recommended Spare Parts 20
4.5 Cleaning the TDA-2SP Scanning Probe Screens 21
4.6 Replacing the Main Power Fuses 21
Revision A
3
4.7 Replacing the HEPA Exhaust Filter 21
A-1. SPECIFICATIONS 22 A-2. TROUBLESHOOTING 23 A-3. QUICK SETUP REFERENCE 24 A-4. RS-232 COMMUNICATION 26
Revision A
4
Guidelines to the use of this manual
SYMBOLS
The following symbols are used throughout the manual to draw attention to items or procedures that require special notice or care.
Operator required action.
Note: Contains important information that, if ignored, can cause inaccurate readings.
Caution: Contains information that, if ignored, can cause equipment damage.
Warning: Contains information that, if ignored, can cause injury or death to those
handling the equipment.
Conventions
[Window] Indicates the information displayed in a window on the Control Panel. <Function> indicates a button on the Control Panel.
Revision A
5
1 INSTALLATION QUALIFICATION
1.1 REQUIREMENTS
NOTE: High ambient temperatures may create instability in the readings.
The 2H requires the following for proper operation:
Stable electrical power at 90/240 VAC, single phase, 50/60 Hz, 5 amps.
Environment Operating Range: 35° to105° Fahrenheit (1.6° to 40.6° Celsius)
with less than 75% relative humidity.
1.2 UNPACKING
NOTE: Save all packaging material for future use.
Carefully unpack and remove the 2H Aerosol Photometer and all accessories from its shipping container. If the instrument has been damaged in transit, notify the shipper immediately.
Make sure the following items are included with both the 2H (9300133) and the 2HHP (9300134):
1) 1 power cord, 6700001 (120 VAC) or T2E0-0063 (220 VAC)
2) 12 feet of clear PVC tubing, 5200116
3) Operating Manual, 1800110
4) Calibration certificate
The following items are optional. Check the order packing slip.
1) Shipping case, 9300145
2) 2H-SP Scanning Probe, complete, 9300144
a) 1 round isokinetic nozzle (red), T2E0-0572 b) 1 rectangular isokinetic nozzle (blue), T2E0-0798 c) 1 NSF 1” round nozzle (black), T2E0-0005 d) 1 Scanning probe, T2SP-0881
If any of these items are on the order packing slip but missing from the shipment, contact ATI immediately at:
Air Techniques International
Division of Hamilton Associates
11403 Cronridge Drive
Owings Mills, Maryland 21117-2247 U.S.A.
Tel: 410 363-9696 Fax: 410 363-9695
www.atitest.com
Revision A
6
2 SETUP
2.1 LOCATIONS & FUNCTIONS
1 9
8
2
10
3 4 5 6
7
Figure 1. Front View
1. Selector Valve. Selects the sample source.
2. Upstream Sample Port. Connects to the sample tubing that is used to measure the upstream
aerosol concentration.
3. Scanning Probe Ground Jack. Electrical ground connection for scanning probe shielding.
(Required for CE compliance)
4. Scanning Probe Connector. Electrical connection for the optional Scanning Probe.
NOTE: Connect the Scanning Probe before applying power to the 2H. Failure to do so
will result in the lack of operation of Scanning Probe’s display. If this occurs, simply turn off the power to the unit and restart the system.
5. Downstream Sample Port. Connects to the sample tubing or optional Scanning Probe that is
used to measure the downstream sample.
6. Optics indicator light. Corresponds to the cleanliness of the optics.
Green Light Less than 10,000mg* of aerosol
has moved through the scattering chamber.
Orange Light More than 10,000mg* but less
Revision A
7
than 17,000mg* of aerosol has moved through the scattering chamber.
Red Light More than 17,000mg* of aerosol
has moved through the scattering chamber. Servicing of the unit is highly recommended.
* Note: These values are approximate. Actual mass may vary with different airflow rates and aerosols.
7. Flow indicator light. Displays green when flow is within 5% of 1 CFM, yellow when flow is
from 5% to 10% of 1 CFM, and red when flow is more than 10% above or below 1 CFM.
8. Function Keys. Used for setting operating parameters and initiating program routines
9. Bar Graph Display. Displays an analog representation of the % leakage to aid in isolating
leaks.
10. Front Panel Display. Indicates % leakage readings and error messages.
2.2 FRONT PANEL CONTROLS AND INDICATORS
The front panel contains seven pressure-activated function keys. The <0>, <100>, <REF>, and <ALARM> keys each contain a red LED to indicate the state of the switch or to prompt the operator. The <Δ> and <> keys are used to scroll among selections as they are shown on the % LEAKAGE display. The <ENTER> key serves as the command key for sending information to
the processor and for initiating or stopping system routines. The <Δ>, <>, and <ENTER> keys contain no indicator LED.
The % LEAKAGE display is an array of high-visibility LEDs that forms a display screen. The Front Panel Display shows alphanumeric messages. Below it is a Bar Graph Display that gives a visual indication of internal photometer cycling or an analog representation of quantity or percentage.
The Scanning Probe contains an unlabeled display on the pistol grip that is a half-scale duplicate of the % LEAKAGE Indicator. The two displays are driven by the same electronics and will always read exactly the same.
The Selector Valve, the Scanning Probe connector, and the two barbed fittings are discussed later.
2.3 SCANNING PROBE CONNECTIONS
If using the 2H with the Scanning Probe, connect the probe’s electrical connector to the 12-pin connector on the front panel of the 2H before applying power to the photometer. Otherwise, the display on the Scanning Probe will not function. Connect the probe’s sampling hose to the barbed
DOWNSTREAM fitting to the right of the 12-pin connector. The cable and sampling hose are
Revision A
8
bound as a single, flexible umbilical. Select the appropriate probe nozzle and screw it onto the threaded end of the flexible portion of the Scanning Probe.
Note: When using the TDA-2SP Scanning Probe with the 2H, the probe must be connected to the photometer before the power is turned on. Otherwise the 2H will not detect the probe and no display will be observed on the pistol grip. If this occurs, switch the power off, verify that the probe is connected properly, and then switch the power back on.
2.4 REAR PANEL CONNECTIONS
3
1 5
2 4
6
Figure 2. Rear Panel
1. Cooling Fan – Maintains airflow through the unit’s enclosure to stabilize electronics.
2. Power Connector with –Connects to the Power Cord.
3. Fuse Block – Contains 2-amp fuse and spare fuse.
4. Power Switch – Rocker Switch. Turns system power on and off.
5. Serial Number Label – Lists the model and serial number.
6. Vacuum Pump Exhaust – Allows a filter to be installed to eliminate particulate
emissions.
The rear panel contains a recessed, 3-pronged male Power Connector, with an integral fuse block and power switch, and a Vacuum Pump Exhaust HEPA Filter.
Revision A
9
Loading...
+ 18 hidden pages