• Simultaneous front panel display
voltage and current
• Constant voltage or constant current operation
• Front and rear connectors
• Remote sense with 5 V line loss compensation
• LabVIEW® and LabWindows® drivers
The Sorensen XHR Series provides 1000 watts
of DC power in a compact half-rack package.
The supplies are designed for benchtop and
system use, and as an ideal companion for other
half-rack instruments in a test console. Its unique
size also eliminates the need for a blank panel to
preserve vertical rack space for OEM applications.
7.5–600 V
1.7–130 A
115230
The XHR is power factor corrected for low current
draw - only 11 amps at 120 volts AC for 1000
watts - and reduced generation of input current
harmonics. Zero voltage or “soft switching”
virtually eliminates switching transients for high
efciency, low noise and high reliability. It is also
stackable, with a small footprint, front panel
binding post connectors, and a low current
requirement with universal input, making the XHR
ideal for benchtop applications.
858.458.0223
sales@ProgrammablePower.com
123
XHR Series : Product Specications
Common
Switching Frequency7.5 V to 300 V models: nominal 125 kHz (250 kHz output ripple); 600 V model: nominal 62.5 kHz (125 kHz output ripple)
Time Delay4 sec maximum from power on until output stable
Voltage Mode Transient Response Time1 ms for output voltage to recover within 0.5% of its previous level after a step change in load current of up to 50% of rated output
Maximum Voltage Differential±600 Vdc from output to safety ground
Remote Start/Stop and Interlock2.5-15 V signal or TTL-compatible input, selectable logic
Remote Analog ProgrammingVoltage and current programming inputs (source must be isolated): 0-5 k, 0-10 k resistances; 0-5 V (default), 0-10 V voltage sources
Remote Analog MonitoringVoltage and current monitor outputs 0-5 V (default), 0-10 V ranges for 0-100% of output
Remote Programming &
Monitoring Accuracy
Front Panel Voltage and Current Control10-turn voltage and current potentiometers
Front Panel Voltage Control Resolution0.02% of maximum voltage
Main Output Connector7.5 to 40 V models: nickel-plated copper bus bars;
Protection FeaturesOver-voltage protection and Over-temperature protection
ApprovalsCE-marked units meet: EN61010-1, EN61000-6-2 and EN61000-6-4;
Environmental
Operating Temperature0°C to 40°C
Storage Temperature-40°C to 85°C
Humidity RangeUp to 80% RH, non-condensing
Physical
DimensionsWidth: 8.5” (216 mm)
WeightApproximately 14 lbs. ( 6.4 kg )
Input
Voltage Ranges85-250 VAC, 47-63 Hz, power factor corrected.
Phases
Power Factor0.99 minimum for full load and 120 Vac input
Current 13 A maximum at 100 Vac;
AC Input Connector TypeIEC 320 connector
1% zero to full scale output for the default range
60 to 600 V models: 4-terminal wire clamp connector for DC output and local sense
CSA C/US certied to UL61010-1B and CSA C22.2 No 1010.1;
Meets USA EMC standard: FCC, part 15B, Class A;
Meets Canadian EMC standard: ICES-001, Class A.
Height: 3.4” (86.4 mm)
Depth: 18.6” (472.2 mm)
Derate maximum output power to 900 W for AC input less than 95 V
11 A maximum at 120 Vac;
6 A maximum at 220 Vac
124
www.ProgrammablePower.com
XHR Series : Product Specications
Output
ModelVoltageCurrentPower
XHR 7.5-1300-7.50-130975 W
XHR 20-500-200-501000 W
XHR 33-330-330-331089 W
XHR 40-250-400-251000 W
XHR 60-180-600-181080 W
XHR 100-100-1000-101000 W
XHR 150-70-1500-71050 W
XHR 300-3.50-3000-3.51050 W
XHR 600-1.70-6000-1.71020 W
Output : At the front panel binding posts
Model
XHR 7.5-1300-7.50-1303 mV14 mA3 mV66 mA
XHR 20-500-200-504 mV6 mA4 mV26 mA
XHR 33-330-330-335 mV4.3 mA5 mV18 mA
XHR 40-250-400-258 mV3.5 mA6 mV14 mA
XHR 60-180-600-188 mV2.8 mA8 mV10 mA
XHR 100-100-1000-1012 mV2 mA12 mV6 mA
XHR 150-70-1500-717 mV1.7 mA17 mV4.5 mA
XHR 300-3.50-3000-3.532 mV1.3 mA32 mV3 mA
XHR 600-1.70-6000-1.762 mV1.2 mA62 mV2 mA
Voltage (VDC)Current (ADC)Voltage CurrentVoltageCurrent
Output RatingsLine Regulation
Meter AccuracyOutput Noise
Model
XHR 7.5-1300.09 V1.4 A70 mV10 mV3.75 mV130 mA
XHR 20-500.3 V0.6 A70 mV10 mV10 mV50 mA
XHR 33-330.43 V0.43 A75 mV7.5 mV16.5 mV33 mA
XHR 40-250.5 V0.35 A75 mV7.5 mV20 mV25 mA
XHR 60-180.7 V0.19 A75 mV10 mV30 mV18 mA
XHR 100-101.1 V0.11 A100 mV10 mV50 mV10 mA
XHR 150-71.6 V0.08 A150 mV20 mV75 mV7 mA
XHR 300-3.54 V0.05 A250 mV30 mV150 mV3.5 mA
XHR 600-1.77 V0.03 A500 mV120 mV300 mV1.7 mA
Voltage
(0.5% to 1% of Vmax + 1 count)
Current
(0.5% of Imax + 1 count)
(0-20 MHz)
Voltage (p-p)
2
Output Ripple
(rms)
Voltage
Load Regulation
Drift (8 hours)
Voltage
(0.05% of Vmax)
4
Current
(0.1% of Imax)
1 kW
3
3. For 0-100% load variation, with constant nominal line voltage. Measured at the rear panel output connector unless stated otherwise.
4. Maximum drift over 8 hours with constant line, load, and temperature, after 30-minute warm-up.
858.458.0223
sales@ProgrammablePower.com
125
XHR Series : Product Specications
5
Current
(0.03% of Imax/°C)
Maximum Remote
Sense
Line Drop Compensation
6
OVP Adjustment
Range
(5%to 110% of Vmax)
Efciency
Model
Temperature Coefcient
Voltage
(0.02% of Vmax/°C)
XHR 7.5-1301.5 mV39 mA3 V / line0.375-8.25 V81%
XHR 20-504 mV15 mA5 V / line1-22 V83%
XHR 33-336.6 mV9.9 mA5 V / line1.65-36.3 V83%
XHR 40-258 mV7.5 mA5 V / line2-44 V83%
XHR 60-1812 mV5.4 mA5 V / line3-66 V84%
XHR 100-1020 mV3 mA5 V / line5-110 V84%
XHR 150-730 mV2.1 mA5 V / line7.5-165 V85%
XHR 300-3.560 mV1.1 mA5 V / line15-330 V85%
XHR 600-1.7120 mV0.48 mA5 V / line30-660 V85%
XHR 1 kW Internal Interface Specications with RS-232 or GPIB Interface Installed