Atec XHR User Manual

Sorensen XHR Series
1 kW
DC Power Supply
• Universal input 85-250 Vac
• Power Factor Correction (PFC)
• Zero voltage “soft switching”
• Simultaneous front panel display voltage and current
• Constant voltage or constant current operation
• Front and rear connectors
• Remote sense with 5 V line loss compensation
• LabVIEW® and LabWindows® drivers
The Sorensen XHR Series provides 1000 watts of DC power in a compact half-rack package. The supplies are designed for benchtop and system use, and as an ideal companion for other half-rack instruments in a test console. Its unique size also eliminates the need for a blank panel to preserve vertical rack space for OEM applications.
7.5–600 V
1.7–130 A
115 230
The XHR is power factor corrected for low current draw - only 11 amps at 120 volts AC for 1000 watts - and reduced generation of input current harmonics. Zero voltage or “soft switching” virtually eliminates switching transients for high efciency, low noise and high reliability. It is also stackable, with a small footprint, front panel binding post connectors, and a low current requirement with universal input, making the XHR ideal for benchtop applications.
858.458.0223
sales@ProgrammablePower.com
123
XHR Series : Product Specications
Common
Switching Frequency 7.5 V to 300 V models: nominal 125 kHz (250 kHz output ripple); 600 V model: nominal 62.5 kHz (125 kHz output ripple)
Time Delay 4 sec maximum from power on until output stable
Voltage Mode Transient Response Time 1 ms for output voltage to recover within 0.5% of its previous level after a step change in load current of up to 50% of rated output
Maximum Voltage Differential ±600 Vdc from output to safety ground
Remote Start/Stop and Interlock 2.5-15 V signal or TTL-compatible input, selectable logic
Remote Analog Programming Voltage and current programming inputs (source must be isolated): 0-5 k, 0-10 k resistances; 0-5 V (default), 0-10 V voltage sources
Remote Analog Monitoring Voltage and current monitor outputs 0-5 V (default), 0-10 V ranges for 0-100% of output
Remote Programming & Monitoring Accuracy
Front Panel Voltage and Current Control 10-turn voltage and current potentiometers
Front Panel Voltage Control Resolution 0.02% of maximum voltage
Main Output Connector 7.5 to 40 V models: nickel-plated copper bus bars;
Protection Features Over-voltage protection and Over-temperature protection
Approvals CE-marked units meet: EN61010-1, EN61000-6-2 and EN61000-6-4;
Environmental
Operating Temperature 0°C to 40°C
Storage Temperature -40°C to 85°C
Humidity Range Up to 80% RH, non-condensing
Physical
Dimensions Width: 8.5” (216 mm)
Weight Approximately 14 lbs. ( 6.4 kg )
Input
Voltage Ranges 85-250 VAC, 47-63 Hz, power factor corrected.
Phases
Power Factor 0.99 minimum for full load and 120 Vac input
Current 13 A maximum at 100 Vac;
AC Input Connector Type IEC 320 connector
1% zero to full scale output for the default range
60 to 600 V models: 4-terminal wire clamp connector for DC output and local sense
CSA C/US certied to UL61010-1B and CSA C22.2 No 1010.1; Meets USA EMC standard: FCC, part 15B, Class A; Meets Canadian EMC standard: ICES-001, Class A.
Height: 3.4” (86.4 mm) Depth: 18.6” (472.2 mm)
Derate maximum output power to 900 W for AC input less than 95 V
11 A maximum at 120 Vac; 6 A maximum at 220 Vac
124
www.ProgrammablePower.com
XHR Series : Product Specications
Output
Model Voltage Current Power
XHR 7.5-130 0-7.5 0-130 975 W
XHR 20-50 0-20 0-50 1000 W
XHR 33-33 0-33 0-33 1089 W
XHR 40-25 0-40 0-25 1000 W
XHR 60-18 0-60 0-18 1080 W
XHR 100-10 0-100 0-10 1000 W
XHR 150-7 0-150 0-7 1050 W
XHR 300-3.5 0-300 0-3.5 1050 W
XHR 600-1.7 0-600 0-1.7 1020 W
Output : At the front panel binding posts
Model
XHR 7.5-130 0-7.5 0-130 3 mV 14 mA 3 mV 66 mA
XHR 20-50 0-20 0-50 4 mV 6 mA 4 mV 26 mA
XHR 33-33 0-33 0-33 5 mV 4.3 mA 5 mV 18 mA
XHR 40-25 0-40 0-25 8 mV 3.5 mA 6 mV 14 mA
XHR 60-18 0-60 0-18 8 mV 2.8 mA 8 mV 10 mA
XHR 100-10 0-100 0-10 12 mV 2 mA 12 mV 6 mA
XHR 150-7 0-150 0-7 17 mV 1.7 mA 17 mV 4.5 mA
XHR 300-3.5 0-300 0-3.5 32 mV 1.3 mA 32 mV 3 mA
XHR 600-1.7 0-600 0-1.7 62 mV 1.2 mA 62 mV 2 mA
Voltage (VDC) Current (ADC) Voltage Current Voltage Current
Output Ratings Line Regulation
Meter Accuracy Output Noise
Model
XHR 7.5-130 0.09 V 1.4 A 70 mV 10 mV 3.75 mV 130 mA
XHR 20-50 0.3 V 0.6 A 70 mV 10 mV 10 mV 50 mA
XHR 33-33 0.43 V 0.43 A 75 mV 7.5 mV 16.5 mV 33 mA
XHR 40-25 0.5 V 0.35 A 75 mV 7.5 mV 20 mV 25 mA
XHR 60-18 0.7 V 0.19 A 75 mV 10 mV 30 mV 18 mA
XHR 100-10 1.1 V 0.11 A 100 mV 10 mV 50 mV 10 mA
XHR 150-7 1.6 V 0.08 A 150 mV 20 mV 75 mV 7 mA
XHR 300-3.5 4 V 0.05 A 250 mV 30 mV 150 mV 3.5 mA
XHR 600-1.7 7 V 0.03 A 500 mV 120 mV 300 mV 1.7 mA
Voltage
(0.5% to 1% of Vmax + 1 count)
Current
(0.5% of Imax + 1 count)
(0-20 MHz)
Voltage (p-p)
2
Output Ripple
(rms)
Voltage
Load Regulation
Drift (8 hours)
Voltage
(0.05% of Vmax)
4
Current
(0.1% of Imax)
1 kW
3
3. For 0-100% load variation, with constant nominal line voltage. Measured at the rear panel output connector unless stated otherwise.
4. Maximum drift over 8 hours with constant line, load, and temperature, after 30-minute warm-up.
858.458.0223
sales@ProgrammablePower.com
125
XHR Series : Product Specications
5
Current
(0.03% of Imax/°C)
Maximum Remote
Sense
Line Drop Compensation
6
OVP Adjustment
Range
(5%to 110% of Vmax)
Efciency
Model
Temperature Coefcient
Voltage
(0.02% of Vmax/°C)
XHR 7.5-130 1.5 mV 39 mA 3 V / line 0.375-8.25 V 81%
XHR 20-50 4 mV 15 mA 5 V / line 1-22 V 83%
XHR 33-33 6.6 mV 9.9 mA 5 V / line 1.65-36.3 V 83%
XHR 40-25 8 mV 7.5 mA 5 V / line 2-44 V 83%
XHR 60-18 12 mV 5.4 mA 5 V / line 3-66 V 84%
XHR 100-10 20 mV 3 mA 5 V / line 5-110 V 84%
XHR 150-7 30 mV 2.1 mA 5 V / line 7.5-165 V 85%
XHR 300-3.5 60 mV 1.1 mA 5 V / line 15-330 V 85%
XHR 600-1.7 120 mV 0.48 mA 5 V / line 30-660 V 85%
XHR 1 kW Internal Interface Specications with RS-232 or GPIB Interface Installed
1, 8
Program Accuracy Readback Accuracy
Model
Voltage
(mV)
Current
(mA)
OVP
(mV)
Voltage Current
XHR 7.5-130 10 +0.12% 900 +0.1% 80 30 +0.12% 900 +0.1%
XHR 20-50 50 +0.12% 750 +0.1% 200 60 +0.12% 750 +0.1%
XHR 33-33 75 +0.12% 500 +0.1% 330 75 +0.12% 500 +0.1%
XHR 40-25 75 +0.3% 350 +0.15% 400 75 +0.3% 350 +0.1%
XHR 60-18 150 +0.25% 250 +0.1% 600 150 +0.25% 250 +0.1%
XHR 100-10 150 +0.35% 140 +0.15% 800 150 +0.35% 140 +0.15%
XHR 150-7 225 +0.35% 120 +0.1% 1500 225 +0.35% 120 +0.1%
XHR 300-3.5 225 +0.35% 80 +0.1% 3000 225 +0.35% 80 +0.1%
XHR 600-1.7 250 +0.35% 80 +0.1% 6000 300 +0.35% 80 +0.1%
7
Specications subject to change without notice.
1. Specications indicate typical performance at 25°C ± 5°C, nominal line input of 120 Vac.
5. Change in output per °C change in ambient temperature, with constant line and load.
6. Line drop is subtracted from total voltage available at supply output.
7. Typical efciency at 115 Vac input and rated output power.
8. Apply accuracy specications according to the following voltage program accuracy example:
Set a model 20-50 power supply to 10 V. The expected result will be within the range of 10 V ± 75 mV ± 0.12% of the set voltage of 10 V.
126
www.ProgrammablePower.com
XHR Series
XHR 40
Series
Voltage
25
Current
(MXX)
Options
Model Number Description
Options and Accessories
MGA * GPIB / IEEE 488.1
MRA * RS-232 interface card
MIA * ISOL interface card provides isolated analog control and readback
RM-XHR 19-inch Rack Mount Kit for up to two XHR power supplies
M13A Locking knobs for front panel controls
M22A No front binding post
* Options cannot be combined
© 2009 AMETEK Programmable Power All rights reserved. AMETEK Programmable Power is the trademark of AMETEK Inc., registered in the U.S. and other countries. Elgar, Sorensen, California Instruments, and Power Ten are trademarks of AMETEK Inc., registered in the U.S.
858.458.0223
sales@ProgrammablePower.com
127
Notes
128
www.ProgrammablePower.com
Loading...