Atec WT200 User Manual

Specifications
Input
Input circuit type
Rated inputs (range)
Equipment loss (input resistance)
Instantaneous maximum allowable input for one cycle, 20 ms
Instantaneous maximum allowable input for 1 sec.
Continuous maximum allowable input
Continuous maximum common-mode voltage (at 50/60 Hz input)
Common-mode rejection ratio at 600 Vrms between input terminal and case
Input terminals A/D conversion Range switching Automatic range switching
Measurement mode switching
Notes: Direct input and external sensor input of current cannot be used simultaneously. When the input is switched for use, note that the ± terminals are shared.
Item Voltage V
Resistance voltage divider
15/30/60/150/300/600V
Input resistance: Approx. 2 M, Input capacitance: Approx. 13 pF
Peak voltage of 2.8 kV or rms of 2.0 kV, whichever is less
Peak voltage of 2.0 kV or rms of 1.5 kV, whichever is less
Peak voltage of 1.5 kV or rms of 1.0 kV, whichever is less
600 Vrms (with the output connector protective cover used) CAT II, 400 Vrms (with the output connector protective cover removed) CAT II
With voltage input terminals short and current input terminals open, 50/60 Hz, -80 dB or more (±0.01% of range or less) Reference value: 50 kHz max., ±{(maximum range rating)/(range rating) × 0.001 × 1% of rng} or less (voltage range, 0.5 A to 20 A current range) ±{(maximum range rating)/(range rating) × 0.0002 × 1% of rng} or less (WT200, 5 mA to 200 mA range)
0.01% or more. The unit of "f" is kHz. Binding posts Simultaneous conversion of voltage and current inputs, Resolution: 12 bits, Maximum conversion rate: Approx. 26 µs (at approx. 38 kHz) Range can be selected manually, automatically, or by communication control.
Range up: When the measured value exceeds 110% of the rated range or the peak value exceeds approximately 300% of the rated range Range down: When the measured value becomes less than 30% of the rated range and the peak value is less than approximately 300% of the subordinate range
One of the following modes can be set (manually or by communication control): RMS: True RMS measurements for both voltage and current; V MEAN: Rectified mean calibrated to an RMS sine wave measurement for voltage, and true RMS measurement for current; DC: Mean value measurement for both voltage and current
Floating input
Direct input: 5/10/20/50/100/200 mA (for WT200 only)
0.5/1/2/5/10/20 A (for WT200/WT130) External input (optional): 2.5/5/10 V or 50/100/200 mV
Direct input: Approx. 6 m + approx. 0.1 µH (0.5 to 20 A, WT200/WT130),Approx. 500 m (5 mA to 200 mA, WT200) External input: 2.5/5/10 V - approx. 100 k or 50/100/200 mV - approx. 20 k
0.5 to 20 A (WT200/WT130): Peak current of 450 A or rms of 300 A, whichever is less 5 to 200 mA (WT200): Peak current of 150 A or rms of 100 A, whichever is less
For external input, the peak value is equal to or less than 10 times range.
0.5 to 20 A (WT200/WT130): Peak current of 150 A or rms of 40 A, whichever is less 5 to 200 mA (WT200): Peak current of 30 A or rms of 20 A, whichever is less
For external input, the peak value is equal to or less than 10 times range.
0.5 to 20 A (WT200/WT130): Peak current of 100 A or rms of 25 A, whichever is less (WT200) Peak current of 100 A or rms of 30 A, whichever is less (WT130) 5 to 200 mA (WT200): Peak current of 30 A or rms of 20 A, whichever is less
For external input, the peak value is equal to or less than 5 times range.
Direct input: Large binding posts, External input: Safety terminals
Current A
Shunt input
Measurement Functions
Method Frequency range Crest factor
Display accuracy Accuracy (within 3 months after calibration) Conditions: Temperature: 23 ± 5°C Humidity: 30 to 75% R.H. Supply voltage: 100 V ± 5% Input waveform: Sine wave Common-mode voltage: 0 V DC Filter: ON at 200 Hz or less Scaling: OFF After CAL is performed, accuracy is assured by YOKOGAWA calibration system. Note: The unit of "f" in accuracy
Effect of power factor
Note: The unit of "f" in accuracy
Effective input range Accuracy ( Temperature coefficient Display update rate
Item Voltage/current
DC: 10Hz % f , 45Hz : 45Hz % f % 66Hz : 66Hz , f % 1kHz : 1kHz , f % 10kHz :
10kHz , f % 20kHz
Reference value
20kHz
, f %
50kHz :
expressions is kHz.
expressions is kHz.
within 12 months after calibration
*DC: ±0.2% of range is added if the 0.5/1 A range is selected (WT130 only)
For the input range of 10% to 110%, the above specified accuracy is valid. For the input range of 110% to 130%, the above specified reading accuracy increased by 0.5 times is added to the accuracy.
)
The above specified reading accuracy increased by 0.5 times is added to the accuracy (within 3 months after calibration). ±0.03% of range/°C at 5 to 18°C and 28 to 40°C 4 times/s
± (0.2 % of rdg + 0.2 % of rng) ± (0.3 % of rdg + 0.2 % of rng) ± (0.15 % of rdg + 0.1 % of rng) ± (0.3 % of rdg + 0.2 % of rng) ± (0.2 % of rdg + 0.3 % of rng) ± {(0.05 × )% of rdg} ± (0.5 % of rdg + 0.5 % of rng) ± [{0.15 × (f-10)}% of rdg]
± (0.5 % of rdg + 0.5 % of rng) ± [{0.15 × (f-10)}% of rdg]
Frequency Measurement
Input: One of V1, V2, V3, A1, A2, and A3 is selected. Operating principle: Reciprocal counting method Frequency range: 10 Hz to 50 kHz Accuracy: ±(0.1% of rdg + 1 digit)
Minimum input is more than 30% of rated range. When an input frequency is less than 200 Hz, FILTER must be ON to obtain the speci-
Minimum input frequency is more than 20% of frequency measurement range.
fied accuracy.
Digital sampling method and summation averaging method
DC, 10 Hz to 50 kHz
*
"3" at rated input
DC: 10Hz % f , 45Hz : 45Hz % f % 66Hz :
66Hz , f % 1kHz : 1kHz , f ^ 10kHz :
10kHz , f % 20kHz :
Reference value
20kHz
, f % 50kHz :
*DC: ±0.2% of range is added if the 0.5/1 A range is selected (WT130 only) cosϕ = 0 45 Hz % f % 66 Hz: Add ±0.25% of range to display accuracy. Reference data (50 kHz max.): ±{(0.23 + 0.4 x f)% of rng} 1 > cosϕ > 0 Add the value of the influence of cosϕ = 0 times tanϕ to display accuracy. Note that ϕ represents the phase angle between voltage and current.
Active power
± (0.3 % of rdg + 0.3 % of rng) ± (0.5 % of rdg + 0.3 % of rng) ± (0.2 % of rdg + 0.1 % of rng)(WT200) ± (0.25 % of rdg + 0.1 % of rng)(WT130) ± (0.5 % of rdg + 0.3 % of rng) ± (0.3 % of rdg + 0.5 % of rng) ± {(0.08 × f)% of rdg} ± (0.8 % of rdg + 0.8 % of rng) ± [{0.19 × (f-10)}% of rdg]
± (0.8 % of rdg + 0.8 % of rng) ± [{0.25 × (f-10)}% of rdg]
Note: "rdg" means reading and "rng" means range.
Communication
Either GP-IB or RS-232-C is selected. GP-IB Electrical and mechanical specifications: IEEE Std. 488-1978 (JIS C 1901-1987) Functional specifications: SH1, AH1, T5, L4, SR1, RL1, PR0, DC1, DT1, C0 Protocol: IEEE Std. 488.2-1987 Code used: ISO (ASCII) code Address: 0 to 30 talker/listener addresses are settable. RS-232-C Transmission mode: Start-stop synchronization Baud rate: 75, 150, 300, 600, 1200, 2400, 4800, 9600 bps
*
Active power
Apparent power
1-phase 2-wire
1-phase 3-wire
3-phase 3-wire (two­power­meter method)
W
Σ W =W
W
Σ W =W
(W)
W
i
i=1, 3
1+W3
i
i=1, 3
1+W3
(VA)
VA=V× A
VAi=Vi× A
i=1, 3
Σ VA =VA
+VA
1
VA
i=Vi
i=1, 3
Σ VA
3
=
×
2
× A
(VA1+VA3)
3-phase
W
Computation
3-wire
i=1, 2, 3
(three-
(Note that W
power-
does not have
meter
physical
method)
meaning.)
Σ W =W
3-phase
W
4-wire
i=1, 2, 3
Σ W
=W
Computing
Depends on
range
selected V and A ranges.
Display resolution
Computing accuracy (relative to the value calculated from measured values)
Note 1:The apparent power (VA), reactive power (var), power factor (PF), and phase angle
(deg) measurements in this instrument are computed digitally from the voltage, cur­rent, and active power. If the input is non-sinusoidal, the measured values may differ from those obtained with instruments employing different measurement principles.
Note 2:When the current or voltage is less than 0.5% of the range, VA and var will be dis-
played as 0, and PF/deg will be displayed as an error.
Note 3:The Lead and Lag are displayed for V and A input at 50% or more of the rated range.
The detected lead/lag accuracy is ±5 degrees over the frequency range of 20 Hz to 2 kHz.
Note 4:In a Σvar calculation, the var value of each phase is calculated as a negatively signed
value when the phase of the current input is advanced with respect to the voltage input, and is calculated as a positively signed value when the phase is lagging.
* The WT200 can provide 5-digit display (note that the resolution is 20000).
VAi=Vi× A
i
2
i=1, 2, 3
Σ VA
3
=
×
3
(VA1+VA2 +VA3)
1+W3
VAi=Vi× A
i
i=1, 2, 3
Σ VA
1+W2
=VA
+VA
1
+W
3
+VA
Depends on selected V and A ranges.
9999* 9999* 9999*
±0.005% of VA range
Reactive
(var)
(VA)2–W
var
i
i
= (VAi)2–W
i=1, 3
Σ var
3
=var
+var
1
var
i
i
= (VAi)2–W
i=1, 3
Σ var =var
1
var
i
i
= (VAi)2–W
i=1, 2, 3
Σ var =var
+var
1
var
i
i
= (VAi)2× W
i=1, 2, 3
Σ var
2
3
=var
+var2+var
1
Same as apparent power (var ^ 0)
±0.005% of var range
+var
3
3
3
2
2
i
2
i
2
i
2
i
3
i=1, 2, 3
Power factor
(PF)
W
VA
PF
i
W
i
=
VA
i
i=1, 3
Σ PF
Σ W
=
Σ VA
PF
i
W
i
=
VA
i
i=1, 3
Σ PF
Σ W
=
Σ VA
PF
i
W
i
=
VA
i
Σ PF
Σ W
=
Σ VA
PF
i
W
i
=
VA
i
i=1, 2, 3
Σ PF
Σ W
=
Σ VA
–1 to 0 to 1
± 1.000*
0.0005
Phase angle
(deg)
W
–1
()
cos
VA
ϕ
i
W
i
=cos–1()
VA
i
i=1, 3
Σϕ
Σ W
–1
=cos
()
ΣVA
ϕ
i
W
i
=cos–1()
VA
i
i=1, 3
Σϕ
Σ W
–1
=cos
()
ΣVA
ϕ
i
W
i
=cos–1()
VA
i
i=1, 2, 3
Σϕ
Σ W
–1
=cos
()
ΣVA
ϕ
i
W
i
=cos–1()
VA
i
i=1, 2, 3
Σϕ
Σ W
–1
=cos
()
ΣVA
–180 to 0 to 180
± 180.0
Resolution (power factor ±0.0005)
Display Function
Display type: 7-segment LED Number of displays: 3
Display
A
V, A, W, VA, var (each element), elapsed integrating time
B
V, A, W, PF, deg (each element), % (contents ratio in %, THD)
C
V, A, W, V · AHz, ±Wh, ±Ah (each element), Vpk, Apk, MATH
Displayed value
* In the WT200 either 4 or 5 digits for display can be selected.
Unit: m, k, M, V, A, W, VA, var, Hz, h±, deg, % Display update rate: 4 times/s Response time: Approx. 0.5 s (time for displayed value to settle within specified
Display scaling function
Significant digits: Selected automatically according to significant digits in the voltage
Setting range: 0.001 to 9999 (WT200), 0.001 to 1000 (WT130)
Averaging function:
Either of the following two algorithms can be selected: Exponential averaging Moving averaging Response can be set; for exponential averaging, the attenuation constant can be selected and for moving averaging, the number of averages (N) can be set to 8, 16, 32, or 64.
Peak over-range display
The alarm LED will light up if the rms value is greater than 140% of the range or the peak value is greater than 300% of the range.
MAX hold function (WT200 only)
Capable of storing the maximum values for V, A, W, VA, var, Vpk, and Apk.
accuracy of final value after step change from 0% to 10% or 100% to 0% of rated range)
and current ranges
Maximum Reading
V, A, W: 9999 Wh, Ah: 999999 V, AHz : 9999
MATH function
Method: When the DISPLAY C function is set to MATH, the efficiency
(WT130 only) and input crest factor can be measured, and the results of four arithmetic operations of measured values can be displayed on Displays A and B as well as the average active power after time-conversion of integrating power (WT200 only).
Integrator Function
Display resolution: Depending on integrated value, the resolution will be changed (for
Maximum display: –99999 to 999999 MWh/MAh Modes: Standard integration mode (timer mode)
Timer: When the timer is set, integration will be stopped automatically.
Type: Standard type Adds active power and current value of normal
Count overflow: If the integration count exceeds 999999 MWh/MAh or -999999
Accuracy: ± (display accuracy + 0.2% of rdg) Timer accuracy: ±0.02% Remote control: Start, stop, and reset can be remotely controlled by an external
WT200). Depending on elapsed time value, the resolution will be changed (for WT130).
Continuous integration mode (repeat mode) Manual integration mode
Setting range: 000 h:00 min:00 sec to 10000 h:00 min:00 sec (WT200) Setting range: 000 h:00 min to 999 h:59 min (WT130) (When set time is 0, the manual integration mode is automatically selected.)
measurements. Advanced type (WT200 only) Integrates active power and current values in short time intervals, not depending on input signal pe­riod.
MWh/MAh, integration stops and the elapsed time is held on the display.
contact signal. Note that this is available only when the /DA4 or / DA12 option has been installed.
Internal Memory Function
Measurement data Number of data that can be stored: WT200 (253421): 600 blocks
Writing interval: 250 ms, or 1 s to 99 h:59 min:59 sec Reading interval: 250 ms, or 1 s to 1 h (both intervals can be set in
Panel setup information: Four-pattern information can be written/read.
units of second)
WT130 (253502): 300 blocks WT130 (253503): 200 blocks
D/A Converter (Optional)
Output voltage: ±5 V DC FS (approx. ±7.5 V maximum) at rated value or range Number of output channels
Output data selection: Can be selected for each channel. Accuracy: ± (display accuracy + 0.2% of range) Update rate: Same as display update rate Temperature coefficient: ±0.05% of f.s./°C Output format
Frequency
Integration
Other items
: 12 when the /DA12 option is installed; 4 when the /DA4 option is
installed
D/A output
7.5V
5.0V
2.5V
0.5V
0
D/A output
7.0V
5.0V
0
Displayed value
140%
100%
0% –100% –140%
For PF and deg, a value between ±5 and ±7 V will not be output. If an error occurs, the output will be approx. 7.5 V.
10Hz 100Hz 1kHz 10kHz
140% input of the rated value
t0: Rated value setting time
Output
7.0V
5.0V 0V
–5.0V –7.0V
–100%–140%
Rated value input
D/A output
7.5V
7.0V
5.0V
0V
–5.0V –7.0V
–7.5V
Displayed value50kHz
t0 Integration time
Displayed value
140%100%
Loading...
+ 2 hidden pages