Atec WAVEPRO-950, WAVEPRO-960 User Manual

WavePro Oscilloscopes
Specifications
Vertical System
Analog Bandwidth @ 50 Ohms (-3 dB)
Bandwidth Limiters 20 MHz, 200 MHz
Input Impedance 50 Ohms ± 1.5%; 10 MOhms // 11 pF typical (using PP005 probe)
Input Coupling 1 MOhms : AC, DC, GND; 50 Ohms : DC, GND
Maximum Input 50 Ohms : 5 Vrms; 1 MOhms : 100 Vmax (peak AC =5 kHz + DC)
Vertical Resolution 8 bits; up to 11 bits with enhanced resolution (ERES)
Sensitivity 50 Ohms : 1 mV – 1 V/div fully variable ; 1 MOhms : 1 mV – 2 V/div fully variable
DC Accuracy ± 2.0% full scale + 1.5% offset value @ gain > 10 mV
Offset Accuracy ± (1.5% + 0.5% of full scale + 1 mV)
Offset Range 50 Ohms or 1 MOhms : 1 mV – 4.99 mV/div: ±400 mV
Isolation — Channel-to­Channel
WavePro 960
DISCONTINUED
44 4
2 GHz* 1 GHz 500 MHz
50 Ohms : 5 mV – 99 mV/div: ±1 V; 0.1 V – 1 V/div: ±10 V
1 MOhms: 5 mV – 100 mV/div: ±1 V; 101 mV – 2 V/div: ±20 V
> 250:1 at same V/div settings
WavePro 950
DISCONTINUED
WavePro 940
DISCONTINUED
Timebase System
Timebases Main and up to four independent zoom traces simultaneously
Ranges 200 ps/div – 1000 s/div
Clock Accuracy =10 ppm
Interpolator Resolution 5 ps
External Clock Frequency 500 MHz maximum, 50 Ohms, or 1 MOhms impedance
Roll Mode – Operating Range time/div 500 ms – 1000 s/div or sample rate < 100 kS/s max
External Reference 10 MHz timebase reference clock available with input on rear panel
External Timebase Clock 500 MHz maximum external sample clock input on front panel EXT BNC
Acquisition System
Single-Shot Sample Rate
1 Channel Max. 16 GS/s 16 GS/s 8 GS/s
2 Channels Max. 8 GS/s 8 GS/s 8 GS/s
3 – 4 Channels Max. 4 GS/s 4 GS/s 4 GS/s
Maximum Acquisition Points/Ch
Standard 1M / 500k / 250k 1M / 500k / 250k 1M / 500k / 250k
M – Memory Option 4M / 2M / 1M 4M / 2M / 1M 4M / 2M / 1M
L – Memory Option 16M / 8M / 4M 16M / 8M / 4M 16M / 8M / 4M
VL – Memory Option 32M / 16M / 8M 32M / 16M / 8M 32M / 16M / 8M
XL – Memory Option 64M / 32M / 16M
Acquisition Modes
( 1 Ch) / (2 Ch) / ( 3 – 4 Ch)
Random Interleaved Sampling (RIS)
Single-Shot For transient and repetitive signals: 200 ps/div – 1000 s/div
Sequence 2 – 8000 segments
Intersegment Time Typically 30 µs
Acquisition Processing
Averaging
Enhanced Resolution (ERES) From 8.5 to 11 bits vertical resolution
Envelope (Extrema) Envelope, floor, roof for up to 106 sweeps
Triggering System
Modes Normal, Auto, Single, and Stop
Sources
Continuous averaging up to 106 sweeps with weighting range from 1:1 to 1:1023 (option).
Any input channel, external, Ext/5 or line; slope, level, and coupling unique to each source (except line trigger)
50 GS/s for repetitive signals: 200 ps/div – 1 µs/div
Summed averaging to 103 sweeps (standard).
Slope Positive, Negative, Window
Coupling modes DC, AC, HF, HFREJ, LFREJ
AC Cutoff Frequency
HFREJ, LFREJ
Pre-trigger delay 0 – 100% of horizontal time scale
Post-trigger delay 0 – 10000 divisions
Hold-off by time or events Up to 20s or from 1 to 99 999 999 events
Internal trigger range ±5 div
Max trigger frequency 1 GHz (DC, AC), >1 GHz (HF) on WavePro 950, >2.0 GHz (HF) on WavePro 960
External trigger input range ±0.5 ( ±2.5 V with Ext/5 selected )
Maximum ext. input @ 50 Ohms
7.5 Hz Typical
50 kHz typical
±5 V DC or 5Vrms
Loading...
+ 3 hidden pages