Atec RadMan User Manual

PERSONAL MONITOR
RadMan / RadMan XT
RadMan XT
DE Patent 19,726,138
US Patents 5,955,954 4,634,968
Monitoring electric and magnetic fields
radiated by broadcast transmitters, mobile phone base stations and radar systems
Wide frequency monitoring
from 1 MHz to 40 GHz
Loud warning buzzer with earphone
for noisy environments
Shaped frequency response matched to
national and international standards
Simultaneous E field and H field monitoring
for near field use
Detachable absorber cap to provide isotropic
response for monitoring signals apart from the body (e.g. for leak detection)
Data logger for permanent recording
(RadMan XT)
NSTS 1013-E0101M 1 / 5 Subject to change
PC Interface Kit (Option) For downloading and analyzing
logged data from RadMan XT For monitoring E field and H field
in real time via fiber optic cable, when the monitor is used off the body.
The Interface kit ESM-TS includes:
Windows® compatible User’s Software
O/E converter that connects to the USB port of a PC
Fiber optic cable to connect RadMan to O/E converter
Exposure versus time
Histogram analysis
DESCRIPTION
Early warning is required wherever humans are likely to be directly threatened by the presence of strong electromagnetic fields. Antenna and radar equipment and machines which use strong electromagnetic fields to heat, weld or bond materials are typical sources of dangerous fields. RadMan and RadMan XT provide safe and timely warning of electro­magnetic fields in these areas of application.
Warning and indication
A warning buzzer will produce a loud sound as soon as 50% of the maximum permissible exposure limit is exceeded. The included earphone is ideal for noisy environments. Four level indicator LEDs provide an approximate indication of the measured field strength in steps of 12.5%, 25%, 50% and 100% of the radiated power density.
Matching the standard
The shaped frequency response of the field sensors mirrors a particular
standard. There’s no more need to adjust the alarm limit to the frequency
of the field source or service. Thanks to the shaping the alarm limit is always matched – over the entire frequency range. Different versions of RadMan and RadMan XT are available to cover the different national and international standards.
Distance-independent operation
The fixed relationship between electric and magnetic field does not apply in the near field of radiation. RadMan and RadMan XT have both electric (E) and magnetic (H) field sensors – for reliable warning independent from distance to the source.
Minimized influence of the body
Personal monitors are mainly used “on the body”, best worn on the outside of clothing. To minimize reflections caused by the human body, the yellow cap of the RadMan and RadMan XT includes a specially designed RF absorber. The absorber cap can be repositioned on the opposite side of the housing to obtain isotropic (non-directional) response for use “off the body”.
Data logger included in RadMan XT
RadMan XT continuously records the measured values for both the E field and the H field and adds a time stamp to each record. The data logger is always on and is able to store more than 1600 records. When the circular memory is full, it simply stores the newest data in place of the oldest data.
NSTS 1013-E0101M 2 / 5 Subject to change
RadMan XT used as an instrument for leak detection
RadMan XT used as a personal monitor for radio frequency fields
APPLICATIONS
Monitoring RF signals (occupational limits)
Mobile, telecom, broadcast and industrial applications are covered by the standard version of RadMan and RadMan XT (with data logger). The E field and H field sensors are shaped to match the higher level of two­tier standards, i.e. the “Controlled”, “Occupational” or “RF Worker” limits.
Monitoring pulsed radar signals
The “Fast Response” versions of RadMan and RadMan XT are
designed to detect the peaks of sharp and narrow radar pulses. The ICNIRP standard, for example, requires peak detection when the ratio of peak to average power is greater than 30 dB. Peak detection is obtained by reducing the 1 second integration time for E-fields down to 30 milliseconds for all “Fast” models.
Monitoring workplaces
Fixed on a non-conductive tripod (optional accessory) the RadMan can be used for workplace radiation monitoring. This application is ideal for areas to be monitored occasionally. For permanent surveillance Narda offers the wall mountable Smarts IITM.
Searching for leaks
As a locating unit all RadMan and RadMan XT versions can be used to locate leaks on waveguides and coaxial screw connectors (picture on the right). A non-conductive extension with handle is available as an optional accessory.
NSTS 1013-E0101M 3 / 5 Subject to change
RadMan / RadMan XT
Frequency range
See Model Selection Guide
Type of frequency response
Shaped (see Model Selection Guide for available standards)
LED indicators
12.5%, 25%, 50% and 100% of Standard a
Alarm threshold
50 % of Standard b
CW damage level
20 dB above standard but not more than 10 kV/m or 26.5 A/m
Peak damage level
40 dB above standard for pulse widths < 10µs but not more than 100 kV/m or 265 A/m
Sensors
E field and H field Diode based design
Directivity
Isotropic (Tri-axial)
Sensitivity c
6 % of standard
Input signal integration time
1 s (30 ms for the E-field of all fast response models)
ELF Immunity @ 50/60 Hz
1 kV/m
Data logger (RadMan XT only) Number of records
e
Logging intervals f Logging time @ rate of 1/min
1638 1 s, 2 s, 5 s, 10 s, 1 min, 3 min (default: 1 min)
27.3 hours
UNCERTAINTY
Flatness of frequency response g
E-FIELD (electric field):
±3 dB (up to 3 GHz) +4/-3 dB (3 GHz to 10 GHz) +6/-3 dB (10 GHz to 18 GHz)
+6/-10 dB (18 GHz to 40 GHz)
H-FIELD (magnetic field): ±3 dB (up to 1 GHz)
Isotropic response d
+4/-2 dB (27 MHz to 500 MHz)
GENERAL SPECIFICATIONS
Calibration frequency
100 MHz
Recommended calibration interval
36 months
Battery type/ life (approx.)
2 x AAA Alkaline, 200 hrs. (with LEDs and alarm OFF, interface cover closed)
Temperature range Operating Non-operating (Transport)
-10 °C to +55 °C
-40 °C to +70 °C
Humidity
5 to 95%, non condensing 29 g/m³ absolute humidity (IEC 60721-3-2 class 7K2)
Size (with cap as absorber)
37 x 41 x 163 mm
Weight
130 g
Accessories (included)
Earphone, operating manual, belt bag, batteries
Country of origin
Germany
a The percent of standard ratings refer to equivalent power density b The alarm threshold is set to 50% of Standard ±1 dB at the calibration frequency c This value is only significant for data logging and online measurements d Uncertainty due to varying polarization (verified by type approval test). Ellipse ratio included e Each record includes the maximum, minimum and average values for both the E field and the H field f The logging interval can be selected via the ESM-TS software (optional) g The specified frequency response can be exceeded in the range 6.2 GHz to 8.8 GHz due to resonances in the H field sensor. No underestimation of the measured field will result from this in any case. This means that an alarm may occur at field strengths that are below the limit value.
SPECIFICATIONS
NSTS 1013-E0101M 4 / 5 Subject to change
® Names and Logo are registered trademarks of Narda Safety Test Solutions GmbH and L3 Communications Holdings, Inc. – Trade names are trademarks of the owners.
Narda Safety Test Solutions GmbH
Sandwiesenstrasse 7 72793 Pfullingen, Germany Phone: +49 (0) 7121 9732-0 Fax: +49 (0) 7121 9732-790 E-Mail: Support.Narda-de@L-3COM.com www.narda-sts.de
Narda Safety Test Solutions
435 Moreland Road Hauppauge, NY 11788, USA Phone: +1 631 231-1700 Fax: +1 631 231-1711 E-Mail: NardaSTS@L-3COM.com www.narda-sts.us
Narda Safety Test Solutions Srl
Via Leonardo da Vinci, 21/23 20090 Segrate (Milano), Italy Phone: +39 02 269987 1 Fax: +39 02 269987 00 E-mail: nardait.support@L-3COM.com www.narda-sts.it
STANDARD / GUIDANCE
Part Number
Frequency Range
RadMan XT (with data logger)
RadMan
E-FIELD
H-FIELD
BGV B11, 2001 EXP. 1 Occupational
2251/01 2251/51, fast response
2250/51 2250/01, fast response
1 MHz to 40 GHz
1 MHz to 1 GHz
Canada, Safety Code 6 99 EHD-237 RF Workers
2251/10
2250/60 2250/10, fast response
ICNIRP, 1998 a Occupational
2251/06 2251/56, fast response
2250/56 2250/06, fast response
27 MHz to 1 GHz
FCC 96-326 Occupational / Controlled
2251/02
2250/52 2250/02, fast response
3 MHz to 40 GHz
3 MHz to 1 GHz
Japan, RCR-38 Controlled
2251/03
2250/53 2250/03, fast response
RadMan / RadMan XT
Part Number
RadMan / RadMan XT Personal Monitor
See Model Selection Guide ACCESSORIES
ESM-TS, PC Interface Kit, USB for RadMan/XT
includes: Microsoft Windows compatible Software (supports COM 1...99), O/E Converter USB , Cable Fiber Optic Duplex 2m
2251/90.51
Tripod, Benchtop, 0.16m, Non-Conductive
2244/90.32
Test-Generator 27MHz, Hand-Held
2244/90.38
Handle, Non-Conductive Extension, 0.42 m
2250/92.02
Hardcase for RadMan
2250/92.03
Accessory Kit RadMan (Handle, Hardcase)
2250/92.05
Cable, Adapter USB 2.0 - RS232, 0.8m
2260/90.53
NATO STOCK NUMBERS
6625-12-355-2053
2251/06
6625-12-360-2005
2251/56
MODEL SELECTION GUIDE
a
ICNIRP versions are also compliant with standards ÖVE/ÖNORM E 8850 and ARPANSA RPS3
ORDERING INFORMATION
NSTS 1013-E0101M 5 / 5 Subject to change