Atec PMM-EP600 User Manual

EP-600 series
USB/RS232 Electric Field Probes
EP-600 series features EP-600, EP-601, EP-602, EP-603
The all-in-one that sets the standard for miniature optically coupled broadband E-field isotropic probes
Exceptionally small-sized spherical symmetrical configuration, lightweight and minia­turized electronics combined with excellent RF characteristics make the PMM EP-600 series RF Electric Field Probes the ideal solution for all EMC/EMI applications (cham­bers and TEM/GTEM cells), biology and materials research and wherever fast and ac­curate measurements with negligible or minimum interference of the probe to the field under measure are essential. Accredited calibration on request.
Main Features
• 5 kHz to 18 GHz frequency range
• 60 dB dynamic, single range
• Symmetrical dipole configuration
• Excellent isotropy (0,3 dB typical)
• Up to 40 meters communication by Fiber Optic Cable
• Up to 80 hours of operation before recharging
• High performance, high reliability Li-Mn battery
• PC direct connection via Optical to RS232/USB adapters
• Extremely lightweight: 25 g only!
EP-600, EP-601, EP-602, EP-603
USB/RS232 Electric Field Probes
COMMON SPECIFICATIONS
Resolution Sensors Measurement mode Sampling rate
Digital filtering Internal battery Operation time
Recharging time
Internal data memory
Communication Fiber optic connector Fiber optic length
Fiber optic to PC
PC software
Operating temperature Temperature reading resolution Battery voltage reading resolution Dimensions
Weight Probe holder tip
0,01 V/m Six monopoles X-Y-Z axis simultaneous sampling 22 to 0,03 Samples/s depending on filter settings Low-pass 2,3 to 28 Hz; 80 dB notch 50/60 Hz 3V - 5 mA/h, Li-Mn rechargeable 80 hours @ 0,4 S/s, 28 Hz filter 60 hours @ 5 S/s, 28Hz filter 48 hours for full operation time Partial charging allowed, e.g.: 8 hrs rech. = 12 hrs operation @ 5 S/s 1 hr rech. = 5 hrs operation @ 0,4 S/s Serial nr. - Calibration factors & date Firmware version Bidirectional fiber optic link HFBR-0500 10 m, standard 20/40 m options Fiber optic to RS232 converter RS232 to USB converter Displays field, temperature, battery Sampling, filters, averaging, frequency setting Manual data logging as text file
-10 °C + 50 °C 0,1 °C 10 mV 53 mm overall (body: 17 mm dia.; sensor, 17 mm each) 25 g, incl. 1 m fiber optic pigtail 1/4 20 UNC female
Ordering Information:
EP-600 Field probe 100 kHz - 9,25 GHz 0,14 - 140 V/m EP-601 Field probe 10 kHz - 9,25 GHz 0,5 - 500 V/m EP-602 Field probe 5 kHz - 9,25 GHz 1,5 - 1500 V/m EP-603 Field probe 300 kHz - 18 GHz 0,17 - 170 V/m
Supplied with:
• 10 m fiber optic cable
• Optical/RS232 adapter + RS232/USB adapter
• PC Utility WIN-EP600
• Charging fixture, A/C charger, tripod mounting tip
• Standard calibration certificate
Options:
• FO-EP600/20 Fiber optic cable, 20 m
• FO-EP600/40 Fiber optic cable, 40 m
• Accreditated Certificate
Main specifications (for full information please refer to the User Manual)
EP-600 EP-601 EP-602 EP-603
Frequency Range
Flatness
Dynamic Range
Overload
100 kHz - 9,25 GHz 10 kHz - 9,25 GHz 5 kHz - 9,25 GHz 300 kHz - 18 GHz
Frequency correction OFF
1 - 150 MHz: 0,8 dB
0,5 - 6000 MHz: 1,6 dB
0,3 - 7500 MHz: 3,2 dB
Frequency correction ON, typical
0,3 - 7500 MHz: 0,4 dB
0,14 - 140 V/m
60 dB single range
300 V/m 1000 V/m 3000 V/m 350 V/m
Frequency correction OFF
0,1 - 150 MHz: 0,4 dB
0,05 - 6000 MHz: 1,6 dB
0,03 - 7500 MHz: 3,2 dB
Frequency correction ON, typical
0,05 - 7500 MHz: 0,4 dB
0,5 - 500 V/m
60 dB single range
Frequency correction OFF
0,1 - 150 MHz: 0,4 dB
0,05 - 6000 MHz: 1,6 dB
0,03 - 7500 MHz: 3,2 dB
Frequency correction ON, typical
0,05 - 7500 MHz: 0,4 dB
1,5 - 1500 V/m
60 dB single range
Frequency correction OFF
3 - 8200 MHz: 1,4 dB
1 - 12000 MHz: 2,4 dB
0,6 - 1800 MHz: 3,8 dB
Frequency correction ON, typical
0,3 - 18000 MHz: 0,4 dB
0,17 - 170 V/m
60 dB single range
Narda Safety Test Solutions Srl
Via Leonardo da Vinci, 21/23 20090 Segrate (Milano) - ITALY Phone: +39 02 2699871 Fax: +39 02 26998700 E-Mail: support@narda-sts.it Internet: www.narda-sts.it
EP600-FEN-30200 - Specifications subject to changes without prior notice
Loading...