PIM 100
1.2/50us & 8/20us Combination Wave Impulse Module
IEC 61000-4-5 and ANSI C62.41 describe a
combination Wave or Hybrid generator with defined
open AND short circuit waveforms. The Combination
Wave is used to test AC and DC power systems.
Safety is paramount with surge impulses. The
PIM 100 High Voltage connections are all linked into
the System Safety Concept which prevents high
voltage plugs or cables from exposing operators to
potentially lethal voltages. This concept is extended
to Coupling/Decoupling Networks (CDNs) connected
to the PIM 100 output.
PIM 100 as part of the PSURGE 8000 Surge Platform
System can be used together with
Coupling/Decoupling Networks (CDNs) for
superimposition of the impulse on AC and DC power
lines. Single and three phase CDNs have multiplexed
inputs, allowing up to 3 impulse modules to be
connected. A full suite of test impulses can then be
performed completely automatically and without the
need to power down and reconnect the EUT.
PSURGE 8000 Surge Platform can perform all the
programming functions required to perform IEC and
ANSI testing without the need of a control computer.
Voltage and Current monitor circuits enable quick and
easy verification of the impulse shape and amplitude.
In addition, peak measurement circuits are used to
help determine EUT PASS/FAIL criterion.
The integration in the WinFEAT&R control and
reporting software package enhances an efficient
set-up and operation of this test system. Most
importantly, the test load can be transferred to a
computer freeing valuable resources.
Features
Floating high voltage output
Combination wave 1.2/50us - 8/20us
7.4kV impulse voltage
3.7kA impulse current
Impulse voltage & current monitors
Accurate phase angle synchronisation
Benefits
International application – Specifically designed
to meet and exceed the requirements of IEC, EN,
and ANSI tests for power line applications.
Accurate Impulse Synchronisation - The PIM
100 impulses can be applied to the AC power wave
with ±1° accuracy.
Safe and Easy - The interlocked HV section and the
integrated controller allow your operators to test
safely and easily.
Sturdy and Reliable – Careful component selection
ensures that the PIM 100 will continue to operate
under the most strenuous testing regime.
Report Generation - The unit can automatically
generate test reports without a computer. Add
WinFEAT&R control and reporting software to collect
and collate data in any format you like.
Faster completion of testing program - The
PIM 100 connects to an EUT via either a PCD 100 or
130. Both have multiplexed inputs, enabling testing
to continue with other pulses without having to
remove power from the EUT.
Applications
Single & Three phase power line systems
IEC 61000-4-5 Edition 1 and 2 Power & Data
lines
ANSI C62.41 Power lines
Many IEC & EN Product standards
Other international requirements for Hybrid
impulses
Technical Specifications
Impulse Voltage 0.20 – 7.4kV ±10% Impulse front time V= 1.2us ±30% & I= 8us ±20%
Impulse Current 0.10 – 3.7kA ±10% Impulse duration V= 50us ±20% & I= 20us ±20%
Source Impedance 2 Ohms Monitor Outputs BNC, volts(750/1) current(400A/V)
Repetition @ Umax 10 seconds / 6 per minute Impulse Polarity Positive and Negative
Line Synchronisation 1° steps ±1° @50Hz Impulse Output HV connector with safety circuit
Weights and Dimensions (W x H x D, net weight)
PIM 100 45 x 20 x 57 cm 44 lbs (29 kg)
PIM 100 Art. No. 249902 Scope of Supply
WinFEAT&R Control Window
Qty. 1 PIM 100 Impulse Module
Qty. 1 HV DC Bus cable 1m
Qty. 1 Haefely Bus cable 0.5m
Qty. 1 Earth bonding cable 1m
Qty. 1 Earth bonding cable 0.25m
Qty. 1 Users Manual
Options and Accessories
PCD 100 Single phase CDN 16Aac & dc.
Automatic coupling path switching.
Two multiplexed high voltage
inputs.
PCD 130 Three phase CDN 32Aac & dc
Automatic coupling path switching.
Three multiplexed high voltage
PCD 130 Three phase power line CDN
inputs.
PCD 121 Manual coupling network for
symmetrical data lines according to
IEC 61000-4-5
PCD 126 Manual coupling network for
asymmetrical networks according
to IEC 61000-4-5
PIM 110 100kHz Ring wave tester according
to ANSI C62.41 & IEC 61000-4-12
ADAPTERS Single phase input & output
adapters enable PCD 130 to be
used for single phase applications
WinFEAT&R Control and reporting software.
Runs under windows 98, NT, ME,
2000, XP
Rack Mounting Modules can be rack mounted for
greater mechanical stability and
mobility.
Headquarters
Haefely Test AG
Lehenmattstrasse 353
CH-4052, Basel
Switzerland
+ 41 61 373 41 11
+ 41 61 373 45 99
EMC-sales@haefely.com
Haefely has a policy of continuous product improvement. Therefore we reserve the right to change design and specification without notice. PIM100_2006-03-13.doc
Locate your local
sales representative at
www.haefelyEMC.com
North American Office
Hipotronics Inc.
Haefely EMC Division
1650 Route 22
Brewster, NY 10509
++1 845 279 3644 x264
++1 845 279 2467
EMCsales@hubbell-haefely.com