Atec INA6502 User Manual

AUTOMATIC STEP TRANSFORMER
INA 6502
L U
IN
VAR
N
PE
16AT
N.C
N
To power network
PE
Instr. supply
Optional remote test + EUT power off
EUT power
rekaerb tiucriCffo/no
To Modula EUT power in
Power supply
X 3
Modula peripheral controller PC 6501
Plug X1 – System cable IN 25 way sub D – male
Plug X2 – System cable OUT 25 way sub D – female
0% 40%
70%
80%
System cable to Modula system Termination with Modula
interlock plug
The automatic step transformer type INA 6502 is a standard accessor y for the Teseq Modula 6000 instrumentation series. It enables the ac power supply voltage, which is fed to the EUT, to be set to 0%, 40%, 70% or 80% of UINunder program control.
The INA 6502 complies with the requirements of IEC/EN 61000-4-11(2004) for voltage dips and under-
voltages. A large number of product standard specifi cations can be fully met as they do not ask for
continuous and arbitrary voltage variations.
The device is connected to the Modula system via the system cable and is fully embedded in the
confi guration management concept, the software control and the report mechanism.
The voltage steps, UVAR, are programmed as a percentage of the power line voltage using either the mobile handheld controller or by means of WinModula. The updated version of WinModula includes preprogrammed tests using the INA 6502 for the basic standard, the generic standards and a consider­able number of product standards.
A circuit breaker is incorporated into the housing of the INA 6502. This allows the on/off switching of the
power to the EUT at any moment via the user interface or automatically under program control. A remote EUT-off hardware switch to cut power and high voltage pulse generation is available as an option.
The step transformer is designed into a sturdy aluminum housing fi tted with convenient handles. In
applications the INA 6502 can be used in any position and may even be placed under the test bench if the space on the set-up becomes limited.
Block diagram
Integrated function unit to
Modula system
Power line dips to 0%, 40%, 70%
and 80%
Complies with IEC/EN 61000-4-11 Includes EUT power control
switch
Teseq AG
Nordstrasse 11F 4542 Luterb ach Swit zerlan d T + 41 32 6 81 40 40 F + 41 32 681 40 48 sales@tese q.com www.teseq.com
AUTOMATIC STEP TRANSFORMER
INA 6502
691-04 0A July 2 007
Technical specifications:
Input voltage UIN: 0 to 250 VAC,50/60 Hz Output voltage UVAR: 4 steps:0% - 40% - 70% - 80% of UIN Accuracy: ±5% Voltage change with load: 100% output, 0 to 16 A less than 5% 80% output, 0 to 20 A less than 5% 70% output, 0 to 23 A less than 5% 40% output, 0 to 40 A less than 5% Output current capability at 230 V input voltage: 16 Arms at 100% output 20 Arms at 80% output, for 5 s 23 Arms at 70% output, for 5 s 40 Arms at 40% output, for 5 s Voltage selection: Software driven by Modula user interface EUT power on/off function: Front panel switch with power on indicator
circuit breaker operated by Modula user interface
Connectors: Harting type HAN3A,compatible with Modula
safe power linking concept
Possible extensions: Connector for optional remote EUT power-off
switch,interlock connection to door switch
Instrument supply: Selectable 100 –110 VAC, 220–250 VAC, 15 VA, 47 to 63 Hz Dimensions L x W x H: 460 x 200 x 160 mm (18.1 x 7.9 x 6.3”)
Weight: 15 kg (33 lbs) approx. Cable: Power network receptacle to INA 6502 ~2 m (~79”)
(uses input cable delivered with Modula mainframe),
INA 6502 to Modula EUT power input ~2 m (~79”)
Control cable: 2 m (79”), 25 way sub D, twisted pair, shielded (included in delivery) Modula system requirements: Master controller revision 2 (copper interface)
Master controller fi rmware 1.17 or higher
MHC software, ModPDA version 2.30 or higher WinModula version 2.30 or higher